| | Class 2 Device Recall uDR 592h, uDR 596i |  |
| Date Initiated by Firm | July 30, 2019 |
| Create Date | May 12, 2022 |
| Recall Status1 |
Terminated 3 on May 17, 2022 |
| Recall Number | Z-1075-2022 |
| Recall Event ID |
90104 |
| 510(K)Number | K181413 |
| Product Classification |
System, x-ray, stationary - Product Code KPR
|
| Product | Digital Medical X-ray Imaging System, Model: uDR 596i, uDR 592h |
| Code Information |
DI/GTIN: 06971576833023, Serial Numbers: 270012, 270013,270014, 270015 |
| FEI Number |
3011015597
|
Recalling Firm/ Manufacturer |
Shanghai United Imaging Healthcare Co., Ltd.
No. 2258 Chengbei Rd, Jiading Ind.
Shanghai
China
|
Manufacturer Reason for Recall | X-ray imaging system positioning image and protocol label is reversed for Flexion and Extension on C-Spine and L-Spine, this will cause the image to be incorrectly labeled and may cause the patient to have a repeat exposure. |
FDA Determined Cause 2 | Software change control |
| Action | On 07/30/19, Customer Notification Letters started being distributed to customers. Customers were asked to contact the firm if they found an inconsistency between the protocol name and the positioning image. The following phone numbers were provided: (855) 221-1552 (toll free) or (832) 699-2799 (direct).
The firm issued Field Correction Instructions to Service Engineers to modify the protocol name of Flexion and Extension on C-Spine & L-Spine with the corrected guidelines. In addition, the firm committed to permanently fix impacted systems with a software update.
|
| Quantity in Commerce | 4 |
| Distribution | US Nationwide distribution in the state of TX. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|