| | Class 2 Device Recall Avance |  |
| Date Initiated by Firm | April 25, 2022 |
| Create Date | June 24, 2022 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-1285-2022 |
| Recall Event ID |
90137 |
| 510(K)Number | K172045 |
| Product Classification |
Gas-machine, anesthesia - Product Code BSZ
|
| Product | Back up batteries in Avance CS2 and Avance CS2 Pro anesthesia devices manufactured
on or after April 1, 2019 |
| Code Information |
Model Number: 1009-9050-000; UDI/DI: 00840682102322; Serial Numbers: APKB00323, APKB00324, APKB00325, APKB00326, APKB00327, APKB00328, APKZ01002, APKB00236, APKB00237, APKB00238, APKB00239, APKB00240, APKA00595, APKA00596, APKA01159, APKB00159, APKB00241, APKB00242, APKB00243, APKB00244, APKB00245, APKB00246, APKB00247, APKB00248, APKB00249, APKB00250, APKB00308, APKB00309, APKY01153, APKY01219, APKY02040, APKY02041, APKY02401, APKY02402, APKY02403, APKY02404, APKY02405, APKY02406, APKY02407, APKY02408, APKY02409, APKY02410, APKY02411, APKY02412, APKY02413, APKY02414, APKY02415, APKY02416, APKY02417, APKY02418, APKY02419, APKY02420, APKY02421, APKY02422, APKY02423, APKY02424, APKY02425, APKY02426, APKY02427, APKY02428, APKY02429, APKY02430, APKY02431, APKY02432, APKY02433, APKY02434, APKY02435, APKY02436, APKY02437, APKY02438, APKY02439, APKY02440, APKY02441, APKY02442, APKY02443, APKY02444, APKY02445, APKY02446, APKY02447, APKY02448, APKY02449, APKY02450, APKY02451, APKY02452, APKY02453, APKY02454, APKY02455, APKY02456, APKY02457, APKY02458, APKY02459, APKY02460, APKY02461, APKZ00107, APKZ00108, APKZ00109, APKZ00110, APKZ00111, APKZ00112, APKZ00113, APKZ00516, APKZ00517, APKZ01733, APKZ01734, APKZ01735, APKZ01736, APKZ01437, APKZ00843, APKZ00844, APKZ00845, APKZ00846, APKZ00847, APKZ00848, APKZ00889, APKZ00890, APKZ00891, APKZ00892, APKZ00893, APKZ00894, APKA00876, APKA01882, APKA01883, APKA01884, APKY00920, APKB00257, APKB00258, APKB00259, APKB00260, APKB00261, APKB00262, APKB00263, APKB00264, APKB00265, APKA00316, APKY02395, APKY02396, APKA00404, APKA01969, APKB00330, APKB00331, APKB00332, APKB00333, APKB00334, APKB00335, APKB00336, APKB00337, APKY02107, APKY02509, APKY02510, APKY02511, APKY02512, APKY02513, APKY02514, APKY02573, APKY02574, APKZ00398, APKZ01001, APKZ01844, APKY01718, APKY01719, APKY01720, APKZ00637, APKZ01007, APKZ01153, APKA00188, APKA00189, APKZ01703, APKZ01703, APKZ01704, APKZ01704, APKZ01705, APKZ01705, APKZ01706, APKZ01706, APKZ01707, APKZ01707, APKZ01708, APKZ01708, APKZ01709, APKZ01709, APKZ01710, APKZ01710, APKZ01960, APKA01148, APKA01149, APKA00283, APKA00284, APKA00503, APKA01086, APKA01087, APKA01088, APKA01089, APKA01090, APKZ00175, APKZ01614, APKZ02013, APKZ02014, APKZ02015, APKA00285, APKB00130, APKB00131, APKB00132, APKB00133, APKB00134, APKB00135, APKB00136, APKB00137, APKB00445, APKZ02336, APKZ02337, APKY02545, APKY01540, APKY01541, APKY01542, APKA01073, APKA01074, APKA01075, APKA01076, APKA01077, APKA01078, APKZ02320, APKZ01994, APKZ01995, APKY00913, APKY00914, APKY00915, APKY01006, APKY01007, APKZ01665, APKZ01761, APKZ01820, APKA00427, APKB00164, APKB00165, APKB00166, APKB00167, APKB00168, APKB00169, APKB00404, APKB00300, APKB00301, APKB00251, APKB00253, APKB00254, APKB00255, APKB00256, APKB00288, APKB00289, APKB00310, APKY02640, APKY01224, APKY01225, APKY01226, APKY01227, APKY01228, APKY02663, APKY02664, APKY02665, APKY02666, APKY02667, APKY02668, APKY02669, APKY02670, APKY02671, APKY02672, APKY02673, APKA00807, APKA00808, APKA00809, APKA00810, APKA00811, APKA00812, APKA00813, APKA00814, APKB00369, APKB00402, APKB00403, APKB00405, APKB00406, APKB00407, APKZ00696, APKZ00943, APKZ00944, APKY01623, APKZ02151, APKZ02152, APKZ02409, APKZ02410, APKZ02411, APKZ00146, APKZ00147, APKB00252, APKB00287, APKB00290, APKB00291, APKB00292, APKB00293, APKZ02333, APKZ02482, APKZ02483, APKZ02484, APKZ02485, APKZ02486, APKZ02487, APKZ02488, APKZ01800, APKZ01801, APKZ01802, APKZ01803, APKZ01804, APKZ01758, APKZ01759, APKB00297, APKB00298, APKY01252, APKY01253, APKY01254, APKY01255, APKY01256, APKY01624, APKY01625, APKY01834, APKY01883, APKY01884, APKY01887, APKY01888, APKY01835, APKY01928, APKY01929, APKY01930, APKY01931, APKA01246, APKZ00441, APKZ00572, APKZ00573, APKZ00574, APKA00322, APKB00172, APKZ00115, APKB00163, APKZ01996, APKA01727, APKA01926, APKA01927, APKY02569, APKY02570, APKY02571, APKY02572, APKA00400, APKA00401, APKA00402, APKA00403, APKA01002, APKA01003, APKA01004, APKA01005, APKA01006, APKA01007, APKZ00606, APKA00313, APKA00314, APKA00327, APKA00328, APKA00329, APKA00330, APKZ00607, APKY00787, APKZ00329, APKZ00256, APKB00113, APKB00114, APKB00115, APKB00116, APKY01387, APKY01388, APKY01512, APKY01513, APKY01514, APKY01515, APKY01516, APKY01517, APKY01518, APKY01519, APKY01520, APKY01521, APKY01522, APKY01755, APKY01756, APKB00311, APKB00312, APKB00313, APKB00314, APKZ02470, APKZ02471, APKY02544, APKY01904, APKY01905, APKZ02499, APKZ02500, APKZ02501, APKZ02502, APKZ02503, APKZ02504, APKA00362, APKA00618, APKA01720, APKB00383, APKY01783, APKY01784, APKZ00192, APKZ00193, APKZ00194, APKZ00195, APKZ00481, APKZ00486, APKZ02455, APKZ02456, APKZ02457, APKZ02458, APKZ02459, APKZ02460, APKZ02461, APKZ02462, APKZ02463, APKZ02464, APKZ02465, APKZ02466, APKZ00706, APKZ00718, APKZ00720, APKZ00721, APKZ00722, APKA01211, APKA01211, APKA01659, APKB00206, APKB00207, APKB00208, APKB00209, APKY01384, APKY02273, APKY02394, APKZ00258, APKZ00259, APKZ00260, APKZ00406, APKZ00408, APKZ00413, APKZ00482, APKZ00483, APKZ00492, APKA01661, APKA01938, APKA01939, APKB00187, APKY00750, APKY00751, APKA00571, APKA00572, APKA00573, APKA00574, APKA00575, APKA00576, APKY01349, APKY01350, APKY01351, APKY01352, APKY01353, APKY01354, APKY01355, APKY01356, APKY01357, APKY01358, APKY01359, APKY01360, APKY01361, APKY01362, APKY01363, APKY01364, APKY01365, APKY01366, APKY01367, APKY01368, APKY01369, APKY01370, APKY02387, APKZ00415, APKA01682, APKA01683, APKA01684, APKA01685, APKA01686, APKA01687, APKA01688, APKA01689, APKA01690, APKA01691, APKB00472, APKY01246, APKY01247, APKY01248, APKY00999, APKY01000, APKA01319, APKY02467, APKY02468, APKY02469, APKY02470, APKY02471, APKY02472, APKY02473, APKA00801, APKA00310, APKA00311, APKA01670, APKA01671, APKA01672, APKA01673, APKA01674, APKA01675, APKA01676, APKA01677, APKA01678, APKA01679, APKA01680, APKA01681, APKY01249, APKY01250, APKY01251, APKZ00679, APKZ00680, APKZ00681, APKZ01435, APKZ01436, APKB00147, APKB00148, APKB00149, APKB00150, APKB00151, APKB00152, APKB00153, APKB00154, APKB00155, APKB00156, APKB00157, APKB00450, APKY01267, APKY00852, APKY00853, APKY00854, APKY00855, APKA01266, APKA01267, APKA01268, APKA01273, APKZ01283, APKZ01431, APKZ02063, APKZ02064, APKA00705, APKY02675, APKY02676, APKY02677, APKY02678, APKZ00409, APKZ00493, APKA00347, APKA00348, APKA00349, APKA00672, APKA00673, APKA00674, APKY01158, APKZ00988, APKZ00989, APKA01316, APKA01317, APKA01395, APKA01396, APKA01397, APKA01398, APKA01399, APKA01400, APKA01376, APKA01377, APKA01378, APKA01379, APKA01380, APKA01381, APKA01382, APKY01159, APKY01923, APKY01924, APKB00126, APKB00127, APKB00128, APKB00174, APKB00175, APKB00176, APKB00338, APKB00339, APKB00340, APKB00341, APKB00342, APKB00343, APKB00344, APKB00345, APKB00346, APKB00347, APKB00348, APKY02039, APKZ00286, APKZ00287, APKZ01129, APKZ01130, APKZ01131, APKZ01132, APKZ01133, APKZ01134, APKZ01135, APKZ02300, APKZ02301, APKZ02302, APKZ02303, APKZ02304, APKZ02305, APKZ02306, APKZ02307, APKZ02308, APKZ02309, APKZ02310, APKZ02311, APKZ02312, APKZ02313, APKZ02314, APKZ02315, APKZ02553, APKZ02554, APKZ02555, APKZ02556, APKZ02557, APKZ02558, APKZ02559, APKZ02560, APKZ02561, APKZ02562, APKZ02563, APKZ02564, APKZ02565, APKZ02566, APKZ02567, APKZ02568, APKZ02569, APKY01101, APKZ00490, APKZ00491, APKY02674, APKZ02338, APKA00545, APKA00546, APKA00547, APKA00619, APKA00620, APKY00951, APKY00952, APKY00953, APKY00954, APKY00955, APKY00956, APKY00957, APKY00958, APKY00959, APKY00960, APKY00961, APKY00962, APKY00963, APKY00964, APKY00965, APKY00966, APKY01636, APKZ00233, APKZ00234, APKZ00235, APKZ00236, APKZ00237, APKZ00238, APKZ00239, APKZ00240, APKZ00241, APKZ00248, APKZ00249, APKZ00250, APKZ00251, APKY01280, APKY01281, APKY01282, APKY01283, APKY01284, APKY01285, APKZ00693, APKY01591, APKY01592, APKA00244, APKA00245, APKA00246, APKY01529, APKY01530, APKY01959, APKY02488, APKZ01003, APKZ01004, APKZ01005, APKZ01006, APKZ01100, APKZ01101, APKZ01102, APKB00272, APKB00273, APKB00274, APKB00275, APKY02504.APKY02505, APKY02506, APKZ01103, APKA01970, APKA01971, APKA01972, APKY02557, APKY02558, APKY02559, APKY02560, APKY02561, APKY02562, APKY02563, APKY02564, APKA01001, APKA01001, APKY01094, APKY02605, APKY02606, APKZ01920, APKA01791, APKA01792, APKA01793, APKA01794, APKZ00536, APKZ00538, APKZ00539, APKZ00549, APKZ00551, APKZ00552, APKY00262, APKZ01915, APKZ02405, APKZ02406, APKA00776, APKA00777, APKA00778, APKA00779, APKA00781, APKA00782, APKA00829, APKY01580, APKA00465, APKA00466, APKA00467, APKA00468, APKB00271, APKZ01267, APKZ01268, APKY00736, APKY00737, APKY01376, APKY01376, APKY01876, APKY01877, APKY01878, APKY01879, APKY01880, APKY01881, APKY01882, APKY02117, APKZ00252, APKZ00253, APKZ00254, APKZ00255, APKA00742, APKA00978, APKA00979, APKA00981, APKA00982, APKA00983, APKA00984, APKA00985, APKA00986, APKA00987, APKA00988, APKA01012, APKA00430, APKZ00433, APKA00741, APKA00980, APKA01008, APKA01009, APKA01010, APKA01011, APKA01013, APKA01014, APKA01015, APKA01016, APKA01017, APKA01018, APKA01019, APKA01020, APKA01021, APKA01022, APKA01023, APKA01024, APKZ00435, APKZ00436, APKA00469, APKA00470, APKA00471, APKA00472, APKA00473, APKA00474, APKY01392, APKY01393, APKA00740, APKA00786, APKA00911, APKA00912, APKA00913, APKA00914, APKA00915, APKA00916, APKA00917, APKA00918, APKA00919, APKA00920, APKA00921, APKA00922, APKA01025, APKA01026, APKA01027, APKA01028, APKA01029, APKA01030, APKA01031, APKA01032, APKA01033, APKA01034, APKA01035, APKA01036, APKA01037, APKA01038, APKA01039, APKA01040, APKA01041, APKA01042, APKA01043, APKA01044, APKA01045, APKA01046, APKA01047, APKA01048, APKA01049, APKA01050, APKA01051, APKA00492, APKA00493, APKA00494, APKA00594, APKZ00330, APKZ00331, APKZ00332, APKZ00333, APKZ00541, APKA00429, APKA00479, APKA01150, APKB00462, APKB00511, APKB00512, APKB00513, APKB00515, APKB00516, APKY01567, APKY01568, APKY01569, APKY01570, APKY01571, APKY01572, APKY01573, APKZ00669, APKZ02105, APKZ02106, APKA00363, APKA00364, APKY02393, APKZ02157, APKZ02158, APKZ02159, APKZ02160, APKZ02161, APKZ02162, APKZ02163, APKZ02164, APKZ02165, APKZ02166, APKZ02167, APKZ02168, APKB00466, APKB00467, APKY01457, APKA01262, APKY02052, APKY02053, APKY02054, APKY02055, APKY00639, APKY00640, APKY00641, APKY00642, APKY00643, APKY00644, APKY00645, APKY00646, APKY00647, APKY00648, APKY00649, APKY00650, APKY00651, APKY00652, APKY00653, APKY00654, APKY00655, APKY00656, APKY00657, APKY00658, APKY00659, APKY00660, APKY00661, APKY00662, APKY00663, APKY00664, APKY00665, APKY00666, APKA01185, APKA01186, APKA01335, APKB00485, APKY02592, APKY02593, APKY02594, APKY02595, APKY02596, APKY02597, APKY02598, APKY02599, APKY02600, APKY02601, APKY02602, APKY02603, APKY02604, APKY02688, APKY02689, APKY02690, APKY02700, APKA01052, APKA01053, APKA01054, APKA01441, APKA01442, APKA01443, APKA01444, APKA01445, APKA01460, APKA01144, APKA01145, APKY01545, APKY01546, APKY01547, APKY01548, APKY01549, APKY01550, APKY01551, APKA01719, APKB00182, APKZ00322, APKZ00323, APKA00343, APKA00344, APKZ01428, APKZ01429, APKA00783, APKA00784, APKB00140, APKB00141, APKB00142, APKB00143, APKA01193, APKA01194, APKA01195, APKA01196, APKA01197, APKA01198, APKA01199, APKA01200, APKA01201, APKA01202, APKB00111, APKB00170, APKB00171, APKB00415, APKZ00601, APKZ00602, APKZ01757, APKY00912, APKZ02282, APKZ02283, APKZ02284, APKZ02285, APKZ02286, APKZ02287, APKZ02288, APKZ01689, APKZ01750, APKZ01751, APKZ01753, APKZ01825, APKY02591, APKA00333, APKA00334, APKA00335, APKA00336, APKA00337, APKA00338, APKA00339, APKA00340, APKA00385, APKA00386, APKA00387, APKA00388, APKA00389, APKA00390, APKA00391, APKA01152, APKZ00149, APKY00686, APKZ00488, APKA00831, APKA00696, APKY01702, APKY02078, APKA00713, APKA00714, APKA00715, APKA00716, APKA00717, APKA00718, APKA00885, APKA01153, APKA01154, APKA01155, APKA01156, APKA01157, APKB00295, APKB00296, APKY01109, APKY01110, APKY01111, APKY01112, APKY02061, APKZ01146, APKZ01147, APKY01966, APKA01171, APKA01172, APKA01173, APKA01174, APKA01175, APKA01176, APKA01177, APKA01178, APKA01179, APKA01976, APKA01977, APKY01257, APKY01258, APKY01259, APKY01874, APKY02588, APKY02589, APKY02590, APKZ00380, APKZ00381, APKZ02535, APKZ02536, APKZ02537, APKZ02538, APKZ02539, APKY01759, APKY01760, APKY01626, APKA00210, APKA00211, APKA00212, APKA00834, APKA01237, APKA01238, APKA01239, APKA01240, APKA01241, APKA01242, APKA01243, APKA01244, APKB00229, APKB00230, APKB00231, APKB00361, APKA00977, APKZ01696, APKZ01697, APKZ02065, APKZ00432, APKZ00434, APKA01831, APKY02528, APKY02568, APKY01401, APKY01402, APKY01403, APKY01404, APKY01405, APKY01406, APKZ00349, APKZ00350, APKZ00351, APKZ00352, APKZ00353, APKZ00354, APKZ00355, APKZ00356, APKZ00357, APKZ00358, APKA01058, APKA01059, APKA01060, APKA01061, APKA01062, APKA01063, APKA01064, APKA01065, APKA01066, APKA01067, APKA01068, APKA01069, APKA01070, APKA01071, APKA01072, APKY01377, APKY01378, APKY01379, APKY01380, APKY01932, APKY02626, APKZ00514, APKZ00546, APKZ01041, APKZ01042, APKZ01043, APKZ01044, APKZ01045, APKZ01046, APKZ01047, APKZ01048, APKZ01049, APKZ01050, APKZ01081, APKZ01082, APKZ01083, APKZ01084, APKZ01085, APKZ01109, APKZ01110, APKZ01111, APKZ01112, APKZ01113, APKZ02297, APKZ02366, APKZ02367, APKZ02368, APKA00616, APKA00617, APKA00656, APKA00657, APKA00658, APKA00659, APKA00660, APKA00661, APKY01371, APKY01372, APKY01373, APKY01374, APKZ01790, APKZ01791, APKZ01792, APKZ01793, APKA01263, APKA00206, APKZ00506, APKZ01997, APKZ02011, APKZ02012, APKZ01425, APKZ01425, APKZ01426, APKZ01426, APKZ01427, APKZ01427, APKZ02088, APKZ02236, APKZ02237, APKZ02238, APKZ02239, APKY02038, APKA00242, APKA00243, APKZ00247, APKZ00247, APKB00496, APKB00497, APKB00498, APKB00499, APKB00500, APKB00501, APKB00502, APKB00503, APKB00504, APKB00505, APKZ02090, APKZ02091, APKZ02092, APKZ02093, APKZ02094, APKZ02095, APKA00183, APKA00184, APKZ02292, APKZ02293, APKZ02294, APKZ02295, APKZ02296, APKZ02298, APKZ02299, APKZ02365, APKA01271, APKA01272, APKA01767, APKZ00515, APKA01276, APKA01277, APKZ02280, APKZ02281, APKY01230, APKY01231, APKY01232, APKY01233, APKY01234, APKA00698, APKZ02321, APKA00937, APKA00938, APKA01158, APKZ02467, APKZ02468, APKZ02469, APKA01924, APKZ01199, APKZ00511, APKZ01104, APKZ02376, APKZ02377, APKZ02378, APKA01341, APKZ01805, APKZ01806, APKZ01807, APKZ01808, APKZ01809, APKZ01810, APKZ00320, APKZ00416, APKZ01416, APKZ01417, APKA00877, APKY01458, APKY02149, APKB00390, APKB00391, APKA00499, APKA00500, APKA00367, APKA00368, APKA00369, APKA00370, APKA00371, APKA01349, APKA01350, APKA01351, APKY01218, APKZ00139, APKZ00185, APKZ00186, APKA00317, APKA01885, APKA01886, APKA01887, APKA01888, APKA01889, APKA01890, APKA01891, APKA01892, APKA01893, APKA01894, APKY02462, APKY02463, APKY02464, APKY02465, APKY02466, APKZ01481, APKZ00456, APKA00855, APKA00856, APKA00857, APKA00858, APKY02119, APKY02120, APKY02121, APKY02122, APKY02123, APKY02124, APKY00873, APKY00874, APKY00875, APKY00876, APKY00877, APKZ00570, APKZ00571, APKZ01941, APKA01925, APKA00221, APKY01972, APKY01973, APKY01974, APKY01975, APKZ01739, APKZ01740, APKZ01741, APKZ00216, APKZ00217, APKZ00218, APKZ00219, APKZ00220, APKA00936, APKA01258, APKA01852, APKA01853, APKA01854, APKA01855, APKA01856, APKA01857, APKA01858, APKY01154, APKY01155, APKY02479, APKZ00298, APKZ01262, APKZ01263, APKZ01264, APKZ01265, APKZ01266, APKA00933, APKA00934, APKA00935, APKY00776, APKY00777, APKY00778, APKZ00505, APKY01260, APKA00931, APKA00932, APKZ01328, APKA01554, APKA01555, APKA01556, APKA01557, APKA01558, APKA01559, APKB00464, APKB00465, APKB00468, APKB00469, APKB00470, APKZ01269, APKY02086, APKY02087, APKY02088, APKY02089, APKZ00388, APKZ00389, APKZ00725, APKZ00727, APKZ00729, APKZ00731, APKZ00732, APKZ00733, APKA01245, APKA01336, APKA01337, APKA01338, APKA01339, APKA01340, APKZ02322, APKZ02323, APKZ02324, APKZ02386, APKZ02387, APKZ02388, APKZ02389, APKZ02390, APKZ02391, APKZ02392, APKZ02393, APKZ02394, APKZ02395, APKZ02396, APKZ02397, APKZ02398, APKZ02399, APKZ02400, APKZ02401, APKZ02402, APKZ02403, APKZ02414, APKZ02415, APKZ02416, APKZ02417, APKZ02418, APKZ02419, APKZ02420, APKZ02421, APKZ02422, APKZ02425, APKZ02426, APKZ02427, APKZ02428, APKZ02429, APKZ02430, APKZ02431, APKZ02432, APKZ02433, APKZ02434, APKZ02435, APKZ02436, APKZ02437, APKZ02438, APKZ02439, APKZ02440, APKZ02441, APKZ02442, APKZ02443, APKZ02444, APKZ02445, APKZ02446, APKZ02447, APKZ02448, APKZ02449, APKZ02450, APKZ02451, APKZ02452, APKZ02453, APKZ02454, APKZ02505, APKZ02506, APKZ02507, APKZ02508, APKB00285, APKB00286, APKY00667, APKY00668, APKY00669, APKY01589, APKY01590, APKY02042, APKY02043, APKZ00451, APKZ00452, APKZ00540, APKZ00703, APKZ00704, APKZ00705, APKZ00707, APKZ00708, APKZ00709, APKZ00710, APKZ00711, APKZ00712, APKZ00713, APKZ00714, APKZ00715, APKZ00716, APKZ00717, APKZ00719, APKZ00670, APKZ00671, APKZ00672, APKZ00673, APKZ00724, APKZ00728, APKZ00791, APKZ00792, APKZ00793, APKZ00878, APKZ00879, APKZ00895, APKZ00896, APKZ00897, APKZ00898, APKZ01086, APKZ01087, APKZ01088, APKZ01089, APKZ01165, APKZ01166, APKZ01173, APKZ01180, APKZ01181, APKZ01120, APKZ01121, APKZ01122, APKZ01123, APKZ01124, APKZ01125, APKZ01126, APKZ02153, APKZ02154, APKY02622, APKY02623, APKY01102, APKY01103, APKY01104, APKY01105, APKY01106, APKY01107, APKZ01611, APKA00162, APKB00201, APKB00202, APKB00203, APKY01554, APKY01555, APKZ02107, APKA01428, APKA01429, APKZ01203, APKZ01812, APKZ01685, APKZ01686, APKZ00631, APKZ00632, APKZ00633, APKZ00634, APKZ00635, APKZ00636, APKZ00723, APKZ00726, APKZ00730, APKZ00734, APKB00393, APKB00394, APKB00395, APKZ01821, APKZ01822, APKZ01823, APKY02583, APKY02584, APKY02585, APKY02586, APKY02587, APKZ01105, APKZ01106, APKZ01107, APKZ01108, APKY02548, APKY02549, APKY02550, APKZ01479, APKZ01482, APKZ01483, APKZ01484, APKZ01485, APKA01079, APKA01080, APKA01081, APKA01235, APKA01759, APKB00234, APKY01939, APKA00158, APKZ02540, APKA01265, APKY01229, APKZ01811, APKA00392, APKA00393, APKA00394, APKA00395, APKA00396, APKB00329, APKZ01344, APKY00969, APKY01394, APKY01395, APKY01396, APKY01397, APKY01398, APKY01399, APKY01419, APKY01420, APKY01421, APKY02474, APKY02475, APKY02476, APKY02477, APKY02478, APKA01342, APKA01343, APKA01344, APKA01345, APKA01346, APKA01347, APKA01348, APKZ00625, APKZ00626, APKZ00627, APKZ00622, APKZ00623, APKZ00624, APKB00442, APKB00443, APKB00444, APKZ01304, APKZ01305, APKA00998, APKA00566, APKA00567, APKA00568, APKA00569, APKA00570, APKB00193, APKB00194, APKB00195, APKB00196, APKZ02039, APKZ02040, APKZ02041, APKZ02042, APKZ01432, APKZ01433, APKA00216, APKA00217, APKA00222, APKA00223, APKA00224, APKA00225, APKA00226, APKA00227, APKA00228, APKA00229, APKA00230, APKA00231, APKA00232, APKA00233, APKA01264, APKB00384, APKY00847, APKY00848, APKY00849, APKY00850, APKY00851, APKY02077, APKZ00269, APKZ00270, APKZ00271, APKZ00272, APKZ00273, APKZ00274, APKZ00275, APKZ00276, APKZ00277, APKZ00278, APKZ00581, APKZ01174, APKZ01175, APKZ01176, APKZ01177, APKZ01178, APKZ01179, APKZ02289, APKZ02331, APKB00426, APKB00427, APKY01599, APKY01600, APKY01601, APKZ01421, APKZ01422, APKZ01423, APKZ02359, APKA00491, APKA00622, APKY01121, APKY01122, APKY01123, APKY01124, APKY01125, APKY01126, APKY01127, APKY01128, APKY01129, APKY01130, APKA01352, APKA01353, APKA01354, APKZ01679, APKA00592, APKA00593, APKA01250, APKA01251, APKA01252, APKA01253, APKA01254, APKA01255, APKA01256, APKA01257, APKY01216, APKY01217, APKY01685, APKY01686, APKY02520, APKY02607, APKY02608, APKY02609, APKA01401, APKA01402, APKA01403, APKA01404, APKA01405, APKA01406, APKA01407, APKA01408, APKA01409, APKA01410, APKA01411, APKA01412, APKA01413, APKA01414, APKA01415, APKA01424, APKA01425, APKA01426, APKA00428, APKA00478, APKA00480, APKA00481, APKY01064, APKY01065, APKY02610, APKY02611, APKY02612, APKA01209, APKB00506, APKY01847, APKB00181, APKZ00297, APKZ01784, APKZ01700, APKZ01897, APKB00392, APKY01071, APKY01072, APKZ01542, APKZ01543, APKZ01742, APKZ01743, APKZ01744, APKZ01745, APKZ01746, APKZ01747, APKZ01748, APKZ01749, APKY00929, APKZ01969, APKZ01970, APKZ01971, APKZ01972, APKZ01973, APKZ01975, APKZ01976, APKZ01977, APKZ01978, APKZ01979, APKZ02316, APKZ02317, APKZ02318, APKZ02319, APKA01223, APKA01416, APKA01417, APKA01418, APKA01419, APKA01420, APKA01421, APKA01422, APKA01423, APKZ01974, APKA00138, APKA00139, APKA00140, APKA00141, APKA00253, APKA00254, APKA00255, APKA00256, APKA00257, APKA00989, APKA00990, APKA00991, APKA00992, APKA01315, APKB00507, APKB00508, APKB00509, APKB00510, APKY02397, APKY02398, APKY02399, APKY02400, APKA00518, APKA00519, APKA00520, APKA00521, APKA00522, APKA00523, APKA00524, APKA00525, APKA00526, APKA00527, APKA00528, APKA00529, APKA00530, APKA00531, APKA00532, APKA00533, APKA00534, APKA00535, APKA00536, APKA00537, APKA00538, APKA00539, APKA00540, APKA00541, APKA00542, APKA00543, APKA00544, APKA00585, APKY01833, APKY01889, APKY01890, APKY01891, APKY01892, APKY01893, APKY01894, APKY01895, APKY01896, APKY01897, APKY01898, APKY01899, APKZ00467, APKZ00468, APKZ00469, APKZ00470, APKZ00471, APKZ00472, APKZ00473, APKZ00474, APKZ00475, APKZ00476, APKZ00477, APKZ00478, APKA01655, APKA01180, APKA01181, APKA01182, APKA01183, APKA01184, APKA01394, APKA01439, APKA01440, APKZ00374, APKZ00375, APKZ00376, APKA00405, APKA00406, APKA00407, APKA00408, APKA00409, APKA00410, APKA00411, APKA00412, APKA00413, APKA00414, APKA00415, APKA00416, APKA00417, APKA00418, APKA00419, APKA00420, APKY01389, APKY01390, APKY01391, APKB00451, APKB00452, APKA00240, APKA01817, APKA01818, APKA01819, APKA01820, APKA01821, APKA01822, APKA01823, APKA01824, APKA01825, APKA01826, APKA01827, APKA01828, APKA01829, APKA01830, APKA01224, APKA01225, APKA01270, APKA01455, APKA01456, APKA01457, APKA01458, APKA01459, APKA01506, APKY02619, APKY02620, APKY02621, APKZ00125, APKZ00126, APKZ00127, APKZ00128, APKZ00129, APKZ00130, APKZ00131, APKA00163, APKY02021, APKY01058, APKY01059, APKY01060, APKY01061, APKY01875, APKA00794, APKB00299, APKY02059, APKZ00735, APKZ00736, APKZ00737, APKZ00738, APKZ00739, APKZ00740, APKZ00741, APKZ00742, APKZ00743, APKZ00744, APKZ00745, APKZ00746, APKZ00747, APKZ00748, APKZ00749, APKZ00750, APKZ00751, APKZ00752, APKZ00753, APKZ00754, APKZ00755, APKZ00756, APKZ00757, APKZ00758, APKZ00759, APKZ00760, APKZ00761, APKZ00762, APKZ00763, APKZ00764, APKZ00765, APKZ00766, APKZ00767, APKZ00768, APKZ00769, APKZ00770, APKZ00771, APKZ00772, APKZ00773, APKZ00774, APKZ00796, APKZ00797, APKZ00798, APKZ00799, APKZ00800, APKZ00801, APKZ00802, APKZ00803, APKZ00804, APKZ00805, APKZ00806, APKZ00807, APKZ00808, APKZ00809, APKZ00810, APKZ00811, APKZ00812, APKZ00813, APKZ00814, APKZ00815, APKZ00816, APKZ00817, APKZ00818, APKZ00819, APKZ00820, APKZ00821, APKZ00822, APKZ00823, APKZ00824, APKZ00825, APKZ00826, APKZ00827, APKZ00828, APKZ00829, APKZ00830, APKZ00831, APKZ00832, APKZ00833, APKZ00834, APKZ00835, APKZ00836, APKZ00837, APKZ02196, APKZ02197, APKZ02198, APKZ02199, APKZ02200, APKZ02201, APKZ02202, APKZ02203, APKZ02326, APKY02516, APKY02517, APKY02518, APKY02519, APKZ01351, APKZ01352, APKZ01353, APKZ01354, APKA01921, APKZ01699, APKY01245, APKA01320, APKA01321, APKA01322, APKA01323, APKA01324, APKA01325, APKA01326, APKA01327, APKA01328, APKA01329, APKA01330, APKA01331, APKA01332, APKA01333, APKA01334, APKZ00143, APKZ00144, APKA01355, APKA00859, APKA00860, APKA00861, APKA00862, APKY01235, APKY01236, APKY01237, APKY01238, APKY01239, APKY01240, APKY01241, APKY01242, APKY01243, APKY01244, APKB00434, APKB00302, APKB00303, APKB00304, APKB00305, APKB00306, APKB00307, APKY02491, APKY02492, APKZ00101, APKY01066, APKY02515, APKZ01698, APKB00269, APKB00270, APKA01261, APKA00791, APKB00408, APKB00409, APKB00410, APKB00411, APKB00412, APKB00413, APKB00414, APKA00123, APKA00124, APKA00125, APKA00178, APKA00179, APKA00180, APKA00399, APKY02618, APKZ00699, APKZ00700, APKZ00701, APKZ00702, APKY01215, APKZ00639, APKZ00640, APKZ00641, APKB00382, APKA01942, APKA01943, APKA01944, APKA01945, APKY02386, APKA00120, APKA00121, APKA00122, APKY01622, APKA00655, APKY01543, APKY01544, APKZ00204, APKB00177, APKB00178, APKB00179, APKB00180, APKY01323, APKY01324, APKY01325, APKY01326, APKY01327, APKY01328, APKY01467, APKY01468, APKY01469, APKY01471, APKY01472, APKY01474, APKY01475, APKY01476, APKY01477, APKY01478, APKA00309, APKY01906, APKY01907, APKY02679, APKY02680, APKZ01773, APKZ01774, APKZ01775, APKZ01776, APKZ01777, APKZ01778, APKA00171, APKA00172, APKA00173, APKA00174, APKA00185, APKA00186, APKZ02408, APKZ00479, APKZ00484, APKZ00485, APKA00187, APKY01470, APKY01473, APKA00802, APKZ00299, APKZ00556, APKZ00557, APKZ00558, APKZ00559, APKZ00560, APKZ00561, APKZ00562, APKZ00563, APKZ00564, APKZ00565, APKA01712, APKB00400, APKZ00494, APKZ00495, APKZ00496, APKZ00497, APKZ00498, APKZ00499, APKZ00500, APKZ00501, APKZ00502, APKZ00503, APKZ01375, APKZ01376, APKZ01728, APKA01453, APKA01454, APKA01981, APKA01982, APKY00977, APKY00978, APKY01004, APKY01005, APKZ02341, APKZ02342, APKZ02343, APKZ02423, APKZ02424, APKA00199, APKA00200, APKA00201, APKA00202, APKA00203, APKA00204, APKA00205, APKA00588, APKA01949, APKA01950, APKA01951, APKA01952, APKA01953, APKA01954, APKA01955, APKA01956, APKY01009, APKY01010, APKY01011, APKY01012, APKY01013, APKY01014, APKY01015, APKY01016, APKY01017, APKY01018, APKY01019, APKY01020, APKY01021, APKY01022, APKY01023, APKY01024, APKY01025, APKY01026, APKY01027, APKY01028, APKY01029, APKY01030, APKY01031, APKY01032, APKY01033, APKY01034, APKY01035, APKY01036, APKY01037, APKY01038, APKY01039, APKY01040, APKY01041, APKY01042, APKZ02345, APKZ02346, APKZ02347, APKZ02348, APKZ02349, APKZ02350, APKZ02351, APKZ02352, APKZ02353, APKY01330, APKY01331, APKZ00326, APKZ00327, APKA01983, APKA01984, APKA01846, APKA01847, APKA00925, APKA00926, APKA00927, APKA00928, APKA00929, APKA00930, APKB00478, APKB00480, APKB00517, APKZ02072, APKZ02073, APKZ02074, APKZ02075, APKZ02076, APKZ02077, APKZ02078, APKZ02079, APKZ02080, APKZ02081, APKZ02082, APKZ02083, APKZ02084, APKZ02085, APKZ02086, APKA01957, APKZ02489, APKY01073, APKY01074, APKY01075, APKY01076, APKY00815, APKY00816, APKY00817, APKY00818, APKY00819, APKY00820, APKY00821, APKY00822, APKY00823, APKY00824, APKY00825, APKY00826, APKY00827, APKY00828, APKA00421, APKA00422, APKY01063, APKY01575, APKY01576, APKY01577, APKY01578, APKA00699, APKA00700, APKA00701, APKA00702, APKZ00150, APKZ00153, APKZ00154, APKZ00155, APKZ00156, APKZ00157, APKZ00158, APKZ02240, APKZ02241, APKZ02242, APKZ02243, APKZ02244, APKY00945, APKA00840, APKA00841, APKA00842, APKA00843, APKA00844, APKA00845, APKA00846, APKA00847, APKA00848, APKA00849, APKA00850, APKA00851, APKA00852, APKA00853, APKA00854, APKY00976, APKA01231, APKA01232, APKA01233, APKA01234, APKZ00615, APKA01259, APKA01260, APKZ01018, APKZ01019, APKZ02089, APKA00969, APKA00970, APKA00681, APKA00682, APKA00683, APKA00684, APKA00685, APKA00686, APKA00687, APKY00923, APKY00924, APKY00925, APKY00926, APKY02651, APKY02652, APKY02653, APKY02654, APKY02655, APKY02656, APKY02657, APKY02658, APKY02659, APKY02660, APKZ01150, APKZ01151, APKZ00529, APKA00101, APKA00711, APKZ01720, APKZ01721, APKB00401, APKZ02046, APKA00482, APKA00483, APKA00484, APKA00485, APKA00486, APKA00487, APKA00488, APKA00489, APKA00490, APKA00351, APKA00352, APKA00353, APKA00354, APKA00355, APKA00356, APKA00357, APKA00358, APKA00359, APKA00695, APKY00797, APKY00798, APKY00866, APKY00950, APKZ00510, APKB00284, APKY01908, APKY01909, APKA00377, APKA00378, APKA00379, APKA00380, APKA01187, APKA01188, APKA00799, APKB00446, APKB00447, APKZ00442, APKZ01223, APKA01973, APKA01974, APKA01975, APKB00473, APKZ01937, APKZ01938, APKZ01939, APKA00373, APKA00374, APKA00375, APKA00376, APKA00381, APKA00381, APKA00382, APKA00382, APKA00383, APKA00383, APKA00384, APKA00384, APKZ02360, APKZ02360, APKB00218, APKB00219, APKB00220, APKB00221, APKB00222, APKB00223, APKB00224, APKZ01962, APKB00210, APKB00211, APKB00212, APKB00213, APKB00214, APKB00215, APKB00216, APKB00217, APKZ00535, APKZ00537, APKZ00548, APKZ00550, APKZ00553, APKZ00580, APKZ01456, APKZ01458, APKZ01460, APKZ01461, APKZ01462, APKZ01463, APKZ01464, APKZ01465, APKZ01466, APKZ01467, APKZ01468, APKZ01469, APKZ01470, APKZ01471, APKA00679, APKA00680, APKY01053, APKY01054, APKY01055, APKY01056, APKY01552, APKY01553, APKY02037, APKZ00931, APKZ00932, APKZ00933, APKZ00934, APKZ00935, APKZ00458, APKZ00459, APKZ00460, APKZ00461, APKZ00462, APKZ00463, APKZ00464, APKZ00465, APKZ00466, APKY00986, APKY00987, APKY00988, APKY00989, APKY00990, APKY00991, APKY00992, APKZ01788, APKA00647, APKA00648, APKA00649, APKA00650, APKA00651, APKA00652, APKA00653, APKA00654, APKA00824, APKA00825, APKA00826, APKA00827, APKA00828, APKA01838, APKA01839, APKA01840, APKA01841, APKA01842, APKA01843, APKA01844, APKA01844, APKA01845, APKA01845, APKB00232, APKB00233, APKZ00480, APKZ01284, APKZ01285, APKZ01286, APKZ01287, APKZ01288, APKY01593, APKY01594, APKY01595, APKY01596, APKA00706, APKY00865, APKA00130, APKA00131, APKA00132, APKA00181, APKA00214, APKA00215, APKA00605, APKA00606, APKA00607, APKA00608, APKA00609, APKA00610, APKA00611, APKA00612, APKA00613, APKA00623, APKA00624, APKA00625, APKA00626, APKA00627, APKA00628, APKA00629, APKA00630, APKA00631, APKA00632, APKA00633, APKA00634, APKA00667, APKA00668, APKA00697, APKB00349, APKB00350, APKB00351, APKB00422, APKB00423, APKB00424, APKB00425, APKB00435, APKY01416, APKY01417, APKY01418, APKY01556, APKY01557, APKY01558, APKY01559, APKY01560, APKY01561, APKY01562, APKY01563, APKY01564, APKY01565, APKY01566, APKY01657, APKY01658, APKY01659, APKY01660, APKY01661, APKY01662, APKY01663, APKY01664, APKY01665, APKY01666, APKY01667, APKY02575, APKY02576, APKY02577, APKY02578, APKY02579, APKY02580, APKY02581, APKY02582, APKY02687, APKZ00334, APKZ02048, APKZ02049, APKZ02050, APKZ02051, APKZ02052, APKZ02053, APKZ02054, APKZ02055, APKZ02056, APKY01627, APKY01628, APKA01513, APKA01514, APKA01516, APKA01517, APKA01518, APKA01519, APKA01520, APKA01522, APKA01523, APKA01524, APKA01531, APKA01532, APKB00123, APKB00124, APKA00194, APKA00195, APKA00196, APKA00197, APKA00198, APKA00463, APKA00464, APKY00927, APKY01725, APKY01726, APKY01727, APKY01728, APKZ00148, APKA00318, APKA00319, APKA00372, APKA00870, APKA00871, APKA00872, APKA00873, APKA00874, APKA00875, APKA01715, APKA01716, APKA01832, APKA01833, APKA01834, APKA01835, APKA01836, APKA01837, APKY02022, APKY02023, APKY02024, APKY02027, APKY02029, APKY02031, APKY02032, APKY02056, APKY02057, APKY02058, APKY02060, APKA00126, APKA01537, APKA01538, APKA01539, APKA01540, APKA01541, APKA01543, APKA01544, APKA01545, APKA01546, APKA01547, APKA01548, APKA01549, APKA01550, APKA01551, APKA01552, APKA01553, APKY02026, APKY02028, APKY02030, APKZ00694, APKZ01457, APKZ01459, APKA01560, APKB00486, APKB00487, APKZ01942, APKZ01943, APKZ01944, APKZ01945, APKZ01946, APKZ01947, APKZ01948, APKA01356, APKY02152, APKY02153, APKY02154, APKY02155, APKY02156, APKY02157, APKA00315, APKB00160, APKB00161, APKY01261, APKY01262, APKY01763, APKA00102, APKA00103, APKA00104, APKA00105, APKA01692, APKA01693, APKA01694, APKA01695, APKA01696, APKA01697, APKA01698, APKA01699, APKA01700, APKA01701, APKA01702, APKY00878, APKZ01949, APKZ01950, APKY01157, APKZ01182, APKZ01183, APKZ01184, APKZ01185, APKZ01186, APKZ01187, APKZ01188, APKZ01189, APKZ01190, APKZ01191, APKZ01192, APKZ01193, APKZ01194, APKZ01195, APKZ01196, APKZ01197, APKZ01198, APKZ01722, APKY02165, APKZ00528, APKZ01524, APKZ01525, APKZ01526, APKZ01527, APKZ01528, APKZ01529, APKZ01530, APKZ01531, APKZ01532, APKZ01533, APKZ01534, APKZ01535, APKZ01536, APKZ01537, APKZ01538, APKZ01539, APKZ01540, APKZ01541, APKA01283, APKA01284, APKA01285, APKA01286, APKA01374, APKA01375, APKA01427, APKA01650, APKA01651, APKA01652, APKA01653, APKA01922, APKA01923, APKB00277, APKB00278, APKB00279, APKB00280, APKB00281, APKB00282, APKB00283, APKY00981, APKY00982, APKY00983, APKY00984, APKY01220, APKY01221, APKY01222, APKY01301, APKY01302, APKY01303, APKY01304, APKY01721, APKY01722, APKY01723, APKY01724, APKY01821, APKY01822, APKY01900, APKY01901, APKY01902, APKY01903, APKY01911, APKY01912, APKY01913, APKY01914, APKY01998, APKY01999, APKY02261, APKY02262, APKY02263, APKY02480, APKY02565, APKZ00124, APKZ01008, APKZ01009, APKZ01010, APKZ01011, APKZ01012, APKZ01013, APKZ01295, APKZ01296, APKZ01297, APKZ01298, APKZ01299, APKZ01510, APKZ01511, APKY00788, APKY00789, APKY00790, APKY00791, APKY00792, APKY00793, APKY00794, APKY00795, APKY00796, APKZ00937, APKZ00938, APKZ00939, APKZ00940, APKZ00941, APKA01385, APKA01386, APKA01387, APKA01388, APKZ02097, APKZ02098, APKZ02099, APKZ02100, APKZ02101, APKZ02102, APKA01383, APKA01384, APKA01094, APKA01095, APKA01096, APKA01097, APKA01098, APKA01099, APKA01100, APKA01101, APKA01102, APKA01103, APKA01104, APKA01105, APKY01603, APKY01604, APKY01605, APKY01606, APKY01607, APKY01608, APKY01609, APKY01610, APKY01611, APKY01612, APKY01613, APKY01614, APKY01615, APKZ01390, APKZ01391, APKZ01392, APKZ01393, APKZ01394, APKZ01678, APKA00273, APKA00910, APKY00928, APKY01156, APKA00295, APKA00296, APKA00297, APKA00298, APKA00299, APKA00300, APKA00301, APKA00302, APKA00303, APKA00304, APKA00305, APKA00306, APKA00307, APKA00308, APKA00923, APKA00924, APKB00183, APKY02000, APKY02001, APKZ01507, APKZ01508, APKZ01509, APKA00803, APKZ01329, APKZ01330, APKZ01331, APKZ01332, APKZ01333, APKZ01334, APKZ01335, APKZ01336, APKZ01337, APKZ01338, APKZ01339, APKA01389, APKA01390, APKA01391, APKA01392, APKA01393, APKA01869, APKA01870, APKA01871, APKA01872, APKA01873, APKA01874, APKA01875, APKA01876, APKA01877, APKZ00866, APKZ00867, APKZ00868, APKZ00869, APKZ00870, APKZ00871, APKZ00872, APKZ00873, APKZ00874, APKZ00875, APKZ00876, APKZ00877, APKZ01690, APKZ01691, APKZ01692, APKZ01693, APKZ01694, APKZ01695, APKY01152, APKZ01630, APKZ01631, APKZ01632, APKZ01633, APKZ01634, APKZ01780, APKZ01781, APKZ01782, APKZ01783, APKZ02332, APKA00971, APKA00972, APKA00973, APKA00974, APKA00975, APKA00976, APKA01228, APKA01229, APKA01230, APKY01115, APKA01167, APKA01168, APKA01169, APKA01170, APKY01967, APKY01968, APKY01969, APKY01970, APKZ00858, APKZ00859, APKZ00862, APKZ00863, APKZ00864, APKZ00865, APKZ00899, APKZ00900, APKZ00901, APKZ00902, APKY01084, APKY01085, APKY01086, APKY01087, APKY01088, APKY01089, APKY01531, APKZ01167, APKZ01168, APKZ01169, APKZ01170, APKZ01171, APKZ01172, APKZ02361, APKZ02362, APKZ02363, APKZ02364, APKZ00860, APKZ00861, APKA00449, APKA00450, APKA00451, APKA00452, APKA00453, APKA00454, APKA00455, APKA00456, APKA00457, APKA00458, APKA00459, APKA00460, APKY01116, APKY02169, APKZ02264, APKZ02265, APKZ02266, APKZ02267, APKZ02268, APKZ02269, APKZ02270, APKZ02271, APKZ00378, APKZ00379, APKB00436, APKB00437, APKB00438, APKZ01154, APKZ01155, APKA01867, APKA01868, APKA01878, APKA01879, APKA01880, APKA01881, APKY02076, APKB00471, APKZ01896, APKZ00512, APKZ01430, APKZ01657, APKZ01658, APKZ01659, APKZ01660, APKY01300, APKZ00513, APKZ02155, APKZ02156, APKZ02025, APKZ02026, APKZ02027, APKZ02028, APKZ02029, APKZ02030, APKZ02031, APKZ02032, APKZ01152, APKB00110, APKB00381, APKY01629, APKZ00566, APKZ00567, APKZ00568, APKZ00569, APKA01851, APKY00944, APKA00504, APKA00505, APKA00506, APKA00507, APKA00508, APKB00367, APKB00370, APKB00374, APKB00375, APKY01848, APKY01849, APKY01910, APKZ02334, APKZ02335, APKB00225, APKB00226, APKB00227, APKB00228, APKY01915, APKY01916, APKA01735, APKZ01789, APKZ01138, APKZ01140, APKZ01143, APKA01734, APKA00882, APKA00883, APKA00884, APKB00397, APKB00398, APKB00399, APKA01106, APKA01940, APKA01941, APKZ01341, APKB00396, APKB00431, APKB00432, APKB00433, APKY01321, APKY01322, APKB00365, APKB00366, APKB00368, APKB00371, APKB00372, APKB00373, APKB00376, APKB00377, APKB00378, APKB00379, APKB00380, APKY02166, APKZ00440, APKZ01136, APKZ01137, APKZ01139, APKZ01141, APKZ01142, APKZ01144, APKZ01145, APKA01515, APKA01521, APKA01654, APKA01733, APKA02009, APKA02010, APKA02011, APKA02012, APKA02013, APKY00362, APKA00838, APKA00839, APKY01444, APKZ01680, APKZ01681, APKZ01682, APKZ01683, APKZ01826, APKZ01827, APKZ01828, APKZ01829, APKZ01830, APKZ01831, APKZ01832, APKZ01833, APKZ01834, APKZ01835, APKZ01836, APKZ01837, APKZ01838, APKZ01839, APKZ01840, APKZ01841, APKZ01842, APKZ01843, APKA00209, APKZ01156, APKZ01157, APKZ01158, APKZ01159, APKY01885, APKY01886, APKY01961, APKY01962, APKY01963, APKY01964, APKZ01816, APKZ01817, APKZ01818, APKZ01819, APKZ01853, APKZ01854, APKZ01855, APKZ01856, APKZ01857, APKZ01858, APKZ01859, APKZ01860, APKZ01861, APKZ01862, APKZ01863, APKZ01864, APKZ01865, APKZ01866, APKB00360, APKB00199, APKB00200, APKA00241, APKZ00386, APKZ02216, APKZ02217, APKZ02218, APKZ02219, APKZ02220, APKZ02221, APKZ02222, APKZ02223, APKZ02224, APKZ02225, APKZ02226, APKZ02227, APKZ02228, APKZ00487, APKA00552, APKA00553, APKA00554, APKA00555, APKA00556, APKA00557, APKA00558, APKA00559, APKA00560, APKA00561, APKA00562, APKA00563, APKA00564, APKA00578, APKA01722, APKA01723, APKA01724, APKA01725, APKA01726, APKB00276, APKZ00366, APKA00274, APKA00275, APKA00276, APKA00277, APKA00278, APKA00279, APKA00280, APKB00235, APKB00416, APKB00417, APKY00780, APKA01435, APKA01436, APKA01437, APKA01438, APKZ02131, APKZ02132, APKY02025, APKA00165, APKB00315, APKB00316, APKB00317, APKB00318, APKA00448, APKA01226, APKY01424, APKA01848, APKA01849, APKA01850, APKZ00437, APKZ00438, APKZ00439, APKZ02181, APKZ02182, APKZ02183, APKZ02184, APKZ02185, APKZ02186, APKZ02235, APKZ02340, APKY01223, APKY01425, APKY01426, APKY01427, APKY01501, APKY01502, APKY01503, APKY01504, APKY01505, APKY01506, APKZ02245, APKZ02246, APKZ02247, APKZ02248, APKZ02249, APKZ02250, APKY01584, APKY01585, APKY01586, APKY01587, APKY01588, APKA01859, APKA01860, APKA01861, APKA01862, APKB00205, APKA00636, APKA00637, APKA00638, APKA00639, APKA00640, APKA00641, APKA00642, APKA00643, APKA00644, APKA00645, APKA00646, APKA00787, APKA00788, APKA01666, APKB00173, APKB00188, APKZ00582, APKZ01845, APKZ01846, APKZ01847, APKZ01848, APKZ01849, APKZ01850, APKZ01851, APKZ01852, APKZ02472, APKZ02473, APKZ02474, APKZ02475, APKZ02476, APKZ02477, APKZ02478, APKA01656, APKA01657, APKA00780, APKA00785, APKA00793, APKA00830, APKA01084, APKA01247, APKA01932, APKA01933, APKA01934, APKA01935, APKA01936, APKA01937, APKA01958, APKY01337, APKY01338, APKY01339, APKY01340, APKY01341, APKY02615, APKY02616, APKY02627, APKY02628, APKY02631, APKY02632, APKZ01148, APKZ01149, APKA00238, APKA00239, APKA00800, APKA00423, APKA00424, APKA00425, APKA00426, APKA01085, APKY02105, APKY02546, APKY02547, APKZ02480, APKZ02481, APKA01511, APKA01512, APKY01965, APKZ00554, APKZ00555, APKA00669, APKZ01911, APKZ01912, APKZ01913, APKZ01914, APKZ02128, APKZ02129, APKZ02130, APKY00970, APKY01703, APKY01704, APKY01705, APKZ02214, APKZ02215, APKZ02229, APKZ02230, APKZ02231, APKZ02232, APKZ02233, APKA00792, APKA01959, APKY00809, APKY00810, APKY00811, APKY00812, APKY00813, APKY00814, APKY00871, APKY00872, APKY02260, APKZ02194, APKZ02195, APKZ02383, APKZ02384, APKZ02385, APKZ02479, APKA01658, APKZ02407, APKY02661, APKY02662, APKZ01114, APKZ01115, APKZ01116, APKZ01117, APKZ01118, APKZ01119, APKY01917, APKY01918, APKY01919, APKA01269, APKA01274, APKA01275, APKA00220, APKA01212, APKA01213, APKA01214, APKA01215, APKA01216, APKA01217, APKA01218, APKA01219, APKA01220, APKA01221, APKA01222, APKY02073, APKY02074, APKA00323, APKA00324, APKA00325, APKA00326, APKA01736, APKA01737, APKA01738, APKA01739, APKA01740, APKA01741, APKA01742, APKA01743, APKA01744, APKA01745, APKA01746, APKA01747, APKA01748, APKA01749, APKA01750, APKA01751, APKA01752, APKA01753, APKA01754, APKA01755, APKA01756, APKA01946, APKA01947, APKA01948, APKA01960, APKA01961, APKA02027, APKB00101, APKB00102, APKB00103, APKB00104, APKB00105, APKB00106, APKB00107, APKB00108, APKB00117, APKY00902, APKY01263, APKY01264, APKY01265, APKY01266, APKY01428, APKY01429, APKY01430, APKY01431, APKY01432, APKY02126, APKY02127, APKY02128, APKY02129, APKY02130, APKY02131, APKY02132, APKY02134, APKY02135, APKY02136, APKY02137, APKY02138, APKY02139, APKY02140, APKY02141, APKY02142, APKY02143, APKY02144, APKY02145, APKY02146, APKY02147, APKY02174, APKZ00118, APKZ00119, APKZ00120, APKZ00121, APKZ00122, APKZ00123, APKZ00222, APKZ00224, APKZ00225, APKZ00226, APKZ00227, APKZ00228, APKZ00295, APKZ00300, APKZ00324, APKZ00335, APKZ00936, APKZ01160, APKZ01161, APKZ01162, APKZ01163, APKZ01164, APKZ01224, APKZ01225, APKZ01226, APKZ01227, APKZ01228, APKZ01229, APKZ01342, APKZ01343, APKZ01661, APKZ01662, APKZ01663, APKZ01664, APKA00709, APKY00279, APKY01958, APKY02082, APKY02083, APKY02084, APKY02085, APKZ01395, APKZ01396, APKZ01397, APKZ01398, APKZ01399, APKZ01400, APKZ01401, APKZ01402, APKZ01404, APKZ01405, APKZ01418, APKZ01419, APKZ01420, APKZ01504, APKZ01505, APKZ01506, APKZ02120, APKZ02121, APKZ02122, APKZ02123, APKZ02124, APKZ02125, APKZ02126, APKZ02127, APKA01210, APKA01210, APKZ01403, APKA01288, APKA01289, APKA01290, APKA01291, APKA01292, APKA01293, APKA01965, APKA01966, APKA01967, APKA01968, APKZ00407, APKY02613, APKY02614, APKB00184, APKY01946, APKY01947, APKY01948, APKY01949, APKY01950, APKY01951, APKY01952, APKY01953, APKY01954, APKY01955, APKY01956, APKY01957, APKA01660, APKA01525, APKA01146, APKA01147, APKY02482, APKY02483, APKY02566, APKY02567, APKY02641, APKY02642, APKY02643, APKY02644, APKY02645, APKY02646, APKY02647, APKY02648, APKY02649, APKY02650, APKZ00508, APKZ00509, APKZ01738, APKA01728, APKA01729, APKA01730, APKA01731, APKA01732, APKZ00382, APKZ01512, APKZ01513, APKA01508, APKA01509, APKA01510, APKA01768, APKZ02380, APKZ02381, APKZ02382, APKA00816, APKA00817, APKA00818, APKA00819, APKA00820, APKA00821, APKA00822, APKA00823, APKA00603, APKA00604, APKY02148, APKY02630, APKZ00208, APKZ00209, APKZ00210, APKZ00211, APKZ00212, APKZ00213, APKZ00214, APKZ00215, APKZ02033, APKZ02034, APKZ02035, APKZ02036, APKZ02037, APKZ02038, APKZ00221, APKZ00223, APKB00362, APKZ01289, APKZ01290, APKZ01291, APKZ01292, APKZ01293, APKZ01294, APKA01298, APKA01299, APKA01300, APKA01301, APKA01302, APKA01303, APKA01304, APKA01446, APKA01447, APKA01448, APKA01449, APKA01450, APKA01451, APKA01452, APKZ00296, APKZ00325, APKZ00397, APKA01863, APKA01864, APKA01865, APKA01866, APKA01294, APKA01295, APKA01296, APKA01297, APKA00168, APKA01507, APKB00120, APKX02608, APKZ00697, APKZ00698, APKB00189, APKB00190, APKB00449, APKB00138, APKZ00114, APKZ02354, APKZ02355, APKZ02356, APKZ02357, APKZ02358, APKZ02379, APKZ02490, APKZ02491, APKZ02492, APKZ02493, APKZ02494, APKZ02495, APKZ02496, APKB00385, APKB00386, APKB00387, APKB00388, APKB00389, APKZ01349, APKZ01350, APKZ01355, APKZ01356, APKZ01357, APKZ01358, APKZ01368, APKZ01369, APKZ01370, APKZ01371, APKZ01372, APKZ01373, APKB00145, APKB00146, APKY01385, APKY01386, APKB00204, APKB00118, APKY01868, APKY01869, APKY01870, APKY01871, APKY01872, APKY01873, APKB00448, APKA01979, APKA01980, APKA01985, APKZ00173, APKZ00174, APKY01618, APKY01619, APKY01620, APKY01621, APKA00501, APKA00502, APKZ00453, APKA00234, APKA00235, APKA00236, APKA00237, APKA00119, APKA00366, APKA00712, APKB00139, APKB00191, APKB00192, APKY01920, APKY01921, APKZ00583, APKZ02096, APKZ01984, APKZ01985, APKZ01986, APKZ01987, APKZ01988, APKZ01989, APKZ01990, APKZ01991, APKZ01992, APKZ01993, APKA00445, APKA00446, APKA00447, APKB00162, APKB00185, APKB00186, APKY02063, APKY02064, APKY02065, APKY02066, APKY02067, APKY02068, APKY02069, APKY02070, APKY02071, APKY02072, APKA01227, APKB00319, APKB00320, APKB00321, APKB00322, APKZ00695, APKZ00367, APKZ00387, APKZ00390, APKZ02150, APKZ00391, APKZ00978, APKZ00979, APKZ00980, APKZ00981, APKZ00982, APKZ00983, APKZ00984, APKZ00985, APKZ00986, APKZ00987, APKZ00489, APKZ01020, APKA01189, APKA01190, APKA01191, APKY02617, APKA00614, APKA00159, APKA00160, APKA00161, APKA00167, APKA00365, APKA00760, APKA00761, APKA00762, APKA00763, APKA00878, APKA00879, APKA00880, APKA00881, APKA01166, APKA01430, APKA01431, APKA01432, APKA01433, APKA01703, APKA01704, APKA01705, APKA01706, APKA01707, APKA01708, APKB00112, APKB00144, APKY01933, APKY01934, APKY01935, APKY01936, APKY01937, APKY01938, APKY02358, APKY02507, APKY02508, APKZ00414, APKZ00530, APKZ02062, APKZ02369, APKY02133, APKY02170, APKY02171, APKY02172, APKY02173, APKY02175, APKY02629, APKY02692, APKA01713, APKA01714, APKY02388, APKY02389, APKZ02325, APKA00670, APKA00671, APKB00197, APKB00198, APKY01925, APKY01926, APKY01927, APKA00127, APKA00128, APKA00129, APKZ00614, APKA01160, APKA01161, APKA01162, APKA01163, APKA01164, APKA01165, APKA01278, APKA01279, APKA01280, APKA01281, APKA01282, APKY01971, APKZ00152, APKZ02020, APKZ02021, APKZ02022, APKZ02023, APKZ02024, APKZ02066, APKZ02067, APKZ02370, APKZ02371, APKZ02372, APKZ02373, APKZ02374, APKA00495, APKA00496, APKA00497, APKA00498, APKZ00151, APKA00113, APKA00114, APKA00115, APKA00116, APKA00117, APKA00118, APKZ01201, APKZ01202, APKY00856, APKY00857, APKY00858, APKY00859, APKY00860, APKZ00302, APKZ00303, APKZ00304, APKZ00305, APKZ00306, APKZ00307, APKZ00308, APKZ00309, APKZ00310, APKZ00311, APKZ00312, APKZ00313, APKZ00314, APKZ00315, APKZ00316, APKZ00317, APKZ00261, APKZ00262, APKZ00263, APKZ00264, APKZ00265, APKZ00266, APKZ01200, APKZ01434, APKZ01794, APKZ01795, APKZ01796, APKZ01797, APKZ01798, APKZ01799, APKZ00504, APKZ01388, APKZ01389, APKA00582, APKA00583, APKA00584, APKY02390, APKB00363, APKB00364, APKY01316, APKY01317, APKY01318, APKY02624, APKY02625, APKA01357, APKA01358, APKA01359, APKA01360, APKA01361, APKA01362, APKA01363, APKA01364, APKA01365, APKA01366, APKA01367, APKA01368, APKA01369, APKA01370, APKA01371, APKA01372, APKA01373, APKZ01752, APKZ01671, APKZ01672, APKZ01754, APKZ01755, APKZ01756, APKZ02272, APKB00441, APKY01960, APKZ00116, APKZ01729, APKZ01872, APKZ01873, APKZ01874, APKZ01346, APKY01857, APKY01858, APKY00733, APKY00980, APKZ01347, APKA00164, APKA00548, APKY01991, APKZ02045, APKY01413, APKY01414, APKY01415, APKZ02043, APKZ02044, APKZ01712, APKZ01713, APKZ01714, APKZ01715, APKZ01716, APKZ01717, APKZ01718, APKZ01719, APKY02272, APKA01092, APKA01929, APKA01930, APKA01931, APKY00779, APKY00846, APKZ00594, APKZ00595, APKZ00596, APKZ00597, APKZ00598, APKZ00666, APKZ00667, APKZ00668, APKZ02009, APKZ02010, APKY01410, APKY01459, APKY02535, APKY02110, APKY02111, APKY02112, APKY02113, APKZ00257, APKY01691, APKY00835, APKY00836, APKY00837, APKY00838, APKY00839, APKY00840, APKY00841, APKY00842, APKY00843, APKY00844, APKY00845, APKY00903, APKY00904, APKY01859, APKY02536, APKY01982, APKY01983, APKY01344, APKY01345, APKY01411, APKY01412, APKY02253, APKZ00841, APKZ00842, APKY01863, APKY01864, APKY01865, APKA00312, APKY02683, APKY02684, APKY02685, APKB00125, APKY00781, APKY00783, APKY02391, APKY02392, APKA00662, APKA01562, APKA01564, APKA01565, APKB00129, APKY01532, APKB00352, APKB00353, APKB00354, APKB00355, APKB00356, APKY01780, APKA00461, APKA00462, APKA00835, APKA00836, APKY01767, APKY01768, APKY01769, APKY01770, APKY01779, APKY01830, APKZ02103, APKZ02104, APKA00743, APKA00744, APKA00745, APKA00746, APKA00747, APKA00748, APKA00749, APKA00750, APKA00751, APKA00752, APKA00753, APKA00754, APKA00755, APKA00756, APKA00757, APKA00758, APKA00759, APKY01762, APKZ01340, APKA01612, APKA01613, APKA01614, APKA01615, APKA01616, APKA01617, APKA01618, APKY01001, APKY01002, APKY01003, APKY01008, APKY01697, APKY01698, APKY01699, APKZ00616, APKZ01684, APKA01561, APKY02378, APKY02379, APKY01535, APKY01536, APKA00663, APKA00664, APKA00665, APKA00666, APKA00832, APKY01043, APKY01790, APKY01791, APKY01792, APKY01793, APKY01794, APKY01795, APKY01796, APKY01797, APKY01798, APKZ00321, APKA01485, APKY00994, APKY00995, APKY01701, APKY01700, APKY01990, APKY02360, APKY02361, APKY02362, APKY02363, APKY02364, APKY02365, APKY02366, APKY02367, APKY02368, APKY02369, APKY02370, APKY02371, APKY02372, APKY02373, APKY02374, APKY02375, APKY02377, APKZ02149, APKY00782, APKB00121, APKA01717, APKA00272, APKA00805, APKA00806, APKA01563, APKA01566, APKA01567, APKA01568, APKY00973, APKY00974, APKY00975, APKY01048, APKY01049, APKY01050, APKY01051, APKY01077, APKY01078, APKY01079, APKY01772, APKY01773, APKY01774, APKY01775, APKZ00856, APKZ00857, APKZ00925, APKZ00926, APKZ00974, APKZ00975, APKZ00976, APKZ00977, APKZ00991, APKZ01615, APKZ01616, APKZ01617, APKZ01618, APKZ01619, APKZ01620, APKZ01621, APKZ01622, APKZ01623, APKZ01624, APKZ01625, APKZ01626, APKZ01723, APKZ02251, APKZ02252, APKZ02253, APKZ02254, APKZ02344, APKA00968, APKZ00449, APKZ00450, APKA00190, APKA00191, APKA00192, APKA00271, APKZ01221, APKZ01222, APKY01771, APKA00804, APKY01052, APKY01533, APKY01534, APKY01789, APKZ00855, APKZ00973, APKZ00990, APKY01537, APKY01538, APKZ00619, APKY01067, APKY01068, APKY01069, APKY01070, APKA00833, APKZ01627, APKY01695, APKY01696, APKY01761, APKZ00620, APKA00586, APKA00587, APKA01117, APKA01795, APKA01796, APKA01797, APKA01798, APKZ00368, APKZ00369, APKZ00370, APKZ00371, APKZ00372, APKZ00373, APKZ00957, APKZ00958, APKZ00959, APKZ01035, APKZ01036, APKZ01037, APKZ01038, APKZ01039, APKZ01040, APKZ01204, APKZ01205, APKZ01206, APKZ01207, APKZ01208, APKZ01209, APKZ01210, APKZ01211, APKZ01212, APKZ02542, APKY02308, APKZ01870, APKZ01871, APKY02315, APKY02316, APKY02317, APKY02318, APKY02319, APKY02320, APKA00615, APKZ00674, APKZ02212, APKA00516, APKA00517, APKY01507, APKY01508, APKY01509, APKY01510, APKY01511, APKZ00132, APKZ02017, APKZ02018, APKZ02019, APKY00985, APKY01940, APKY01941, APKY01942, APKY01943, APKY01944, APKY01945, APKZ00628, APKZ00629, APKZ00630, APKY02020, APKA00341, APKA00342, APKZ01022, APKZ01024, APKZ01025, APKZ01026, APKY02283, APKY02285, APKY02292, APKY02294, APKY02297, APKY02301, APKY02302, APKY02303, APKY02304, APKY02305, APKY00911, APKZ00199, APKY00907, APKZ01345, APKA00707, APKA00708, APKY00908, APKY00909, APKY01095, APKY01096, APKY01097, APKY01098, APKY01099, APKY01100, APKY02094, APKY02095, APKY02096, APKY02097, APKY02098, APKY02100, APKY02101, APKY02102, APKZ00201, APKZ00202, APKZ00243, APKZ00244, APKZ00301, APKZ01023, APKZ01214, APKZ01217, APKZ01601, APKZ01603, APKZ01604, APKZ01605, APKZ01606, APKZ01898, APKZ01899, APKZ01900, APKY00910, APKZ00288, APKX02476, APKX02477, APKZ00200, APKZ00383, APKZ00384, APKZ00385, APKZ01213, APKY02274, APKY02275, APKY02276, APKY02277, APKY02278, APKY02279, APKY02280, APKY02281, APKZ00880, APKZ01021, APKZ01215, APKZ01216, APKZ01218, APKZ01599, APKZ01600, APKZ01602, APKY02284, APKY02286, APKY02289, APKY02290, APKY02291, APKY02295, APKY02296, APKY02298, APKY02299, APKY02300, APKY02306, APKY02307, APKY02099, APKY02103, APKY02287, APKY02288, APKY02293, APKA00193, APKZ02003, APKZ02004, APKZ02005, APKZ02006, APKY01307, APKZ01701, APKZ01501, APKZ01502, APKA00550, APKY00734, APKY01445, APKZ00996, APKZ00997, APKZ00998, APKY01637, APKA01807, APKZ01955, APKZ01956, APKZ01957, APKZ01958, APKZ01959, APKZ01677, APKY01348, APKA00635, APKB00495, APKA01469, APKA01470, APKY01150, APKY01151, APKY01992, APKY00861, APKZ01359, APKA01487, APKA01488, APKZ01374, APKZ02191, APKZ02192, APKA00350, APKY02484, APKY00863, APKY01046, APKY01047, APKZ00692, APKY00802, APKZ02119, APKA01806, APKY00670, APKZ02290, APKY01319, APKY01320, APKA00360, APKY02264, APKA00286, APKY02493, APKY02494, APKY02495, APKY02496, APKY02497, APKY02498, APKY02499, APKY02500, APKY02501, APKY02502, APKY02503, APKZ01885, APKZ01886, APKZ01887, APKZ01888, APKZ01889, APKZ01890, APKZ01891, APKZ01892, APKZ01893, APKZ01894, APKZ01895, APKZ01963, APKZ01964, APKZ01965, APKZ01966, APKZ01967, APKZ01968, APKZ02059, APKZ02060, APKZ02061, APKA00182, APKZ00393, APKZ01629, APKA01569, APKB00490, APKB00491, APKB00492, APKA00398, APKZ00547, APKY01132, APKY01133, APKY01134, APKY01135, APKY01136, APKY01137, APKY01160, APKY01161, APKY01162, APKY01163, APKY01164, APKY01165, APKY01166, APKY01167, APKY01168, APKY01169, APKY01170, APKY01171, APKY01172, APKY01173, APKY01174, APKY01175, APKY01176, APKY01177, APKY01178, APKY01179, APKY01180, APKY01181, APKY01182, APKY01211, APKY01212, APKY01213, APKY01781, APKA00213, APKA00475, APKA00476, APKA00477, APKA01467, APKZ00839, APKZ00840, APKZ02273, APKZ02274, APKA01978, APKA00251, APKA00265, APKA00266, APKA00267, APKA00268, APKA00269, APKY01438, APKY02343, APKZ02291, APKA00320, APKA00321, APKY00862, APKY01113, APKY01114, APKY01149, APKY01687, APKY01688, APKY01689, APKY01690, APKY02342, APKY02359, APKZ00454, APKZ00786, APKZ00794, APKY00972, APKY01147, APKY01148, APKY01976, APKY02542, APKY02543, APKA00939, APKA00940.APKA00941, APKA00942, APKA01769, APKA01770, APKA01780, APKZ00787, APKZ00788, APKZ01236, APKZ01237, APKZ01238, APKZ01239, APKZ01240, APKZ01241, APKZ01242, APKZ01243, APKZ01244, APKZ01245, APKZ01246, APKZ01247, APKZ01248, APKZ01249, APKZ01250, APKZ01251, APKZ01252, APKZ01253, APKZ01254, APKZ01255, APKZ01256, APKZ01257, APKZ01258, APKZ01259, APKZ01260, APKZ01261, APKA00602, APKA01248, APKA01249, APKY00993, APKY02686, APKZ00525, APKZ00526, APKZ00916, APKZ00917, APKZ00918, APKZ00919, APKZ00920, APKZ00921, APKZ00922, APKZ00923, APKY00784, APKY00785, APKY00786, APKY00429, APKY00430, APKZ00785, APKY00431, APKY01852, APKZ00160, APKZ00161, APKY00738, APKY00738, APKY00739, APKY00739, APKY00740, APKY00740, APKY00741, APKY00741, APKY00742, APKY00742, APKY00743, APKY00743, APKY01853, APKY02118, APKY01437, APKY01758, APKZ00117, APKY01837, APKY01838, APKY01839, APKY01840, APKY00948, APKY00949, APKY00971, APKY01617, APKY01692, APKY01693, APKY01694, APKZ00145, APKY01860, APKZ00455, APKA00218, APKA00247, APKA00248, APKA00249, APKA00250, APKA00252, APKA01463, APKA01464, APKA01465, APKA01466, APKY01851, APKZ00795, APKZ01553, APKZ01554, APKZ01555, APKZ01556, APKZ01557, APKZ01558, APKZ01559, APKZ01560, APKZ01561, APKZ01562, APKZ01563, APKZ01564, APKZ01565, APKZ01566, APKZ01567, APKZ01568, APKZ01569, APKZ01570, APKZ01571, APKZ01572, APKZ01573, APKZ01574, APKZ01575, APKZ01576, APKZ01577, APKZ01578, APKZ01579, APKZ01580, APKZ01581, APKZ01582, APKZ01583, APKZ01584, APKZ01585, APKZ01586, APKZ01587, APKZ01588, APKZ01589, APKZ01590, APKZ01591, APKZ01592, APKZ01593, APKZ01594, APKZ01595, APKZ01596, APKA00600, APKA00601, APKA00270, APKY01861, APKZ02524, APKZ02525, APKZ02526, APKA00281, APKA00282, APKZ00203, APKZ00285, APKA00599, APKA01781, APKZ00245, APKZ00246, APKZ00521, APKZ00522, APKZ00523, APKZ00524, APKY02109, APKY01305, APKY01306, APKY01649, APKY01650, APKY01651, APKZ00775, APKZ00776, APKZ00777, APKZ00778, APKZ00779, APKZ00781, APKY01643, APKY01646, APKY01647, APKA01131, APKA01132, APKA01133, APKY01422, APKZ00784, APKY01652, APKY01656, APKA00136, APKY00943, APKZ01999, APKY00942, APKY00938, APKY00934, APKY00935, APKY00936, APKY00937, APKY01757, APKA01129, APKA00169, APKA00170, APKA02021, APKA02022, APKA02023, APKA02024, APKA00345, APKA00346, APKA00867, APKA00868, APKA00869, APKA00945, APKA02018, APKA02019, APKY01108, APKZ01408, APKZ01922, APKA00108, APKA00109, APKA00110, APKA00111, APKA00112, APKA00137, APKA00866, APKA00943, APKA00944, APKA00946, APKA00947, APKA00948, APKA00949, APKA00950, APKA00951, APKA00952, APKA01124, APKA01125, APKA01126, APKA01127, APKA01128, APKA01139, APKA01140, APKA01141, APKA01461, APKA01462, APKA01502, APKA01503, APKA02016, APKA02017, APKA02020, APKA02025, APKZ01409, APKZ01411, APKZ01412, APKZ01415, APKZ00664, APKZ00665, APKZ01998, APKY01655, APKY01654, APKA01130, APKY02681, APKY02701, APKA01208, APKY01642, APKY02108, APKZ00656, APKZ00657, APKZ01407, APKZ01410, APKZ01413, APKZ02213, APKY01648, APKZ00780, APKA00133, APKA00134, APKZ00658, APKZ01406, APKY00432, APKY02682, APKZ00928, APKY00939, APKZ00782, APKZ01414, APKZ02008, APKZ00783, APKZ00661, APKZ00663, APKA00135, APKY01645, APKZ00660, APKZ02007, APKY00940, APKY00941, APKY01644, APKY01653, APKZ00662, APKZ00659, APKA01585, APKA01586, APKA01587, APKA01588, APKA01589, APKA01590, APKA01591, APKA01592, APKA01593, APKA01594, APKA01595, APKA01596, APKA01597, APKA01598, APKA01599, APKA01600, APKA01601, APKA01602, APKA01603, APKA01604, APKA01605, APKA01606, APKA01607, APKA01608, APKA01609, APKA01610, APKA01611, APKA01623, APKA01624, APKA01625, APKA01626, APKA01627, APKA01628, APKA01629, APKA01630, APKA01631, APKA01632, APKA01633, APKA01634, APKA01635, APKA01636, APKA01637, APKA01638, APKA01639, APKA01640, APKA01641, APKA01642, APKA01643, APKA01644, APKA01645, APKA01646, APKA01647, APKA01648, APKA01649, APKZ02047, APKA01056, APKA01057, APKY01735, APKY01736, APKY01737, APKY01738, APKY01739, APKY01740, APKY01741, APKY01742, APKY01743, APKY01744, APKY01745, APKY01746, APKY01747, APKY01748, APKY01749, APKY01750, APKY01751, APKY01752, APKY01753, APKY01754, APKY01799, APKY01800, APKY01801, APKY01802, APKY01803, APKY01804, APKY01805, APKY01806, APKY01807, APKY01808, APKY01809, APKY01810, APKY01811, APKY01812, APKY01813, APKY01814, APKY01815, APKY01816, APKY01817, APKY01818, APKY01819, APKY01820, APKY02018, APKY02019, APKY02150, APKZ00102, APKZ00103, APKZ00531, APKZ00532, APKZ00533, APKZ00534, APKZ02204, APKZ02205, APKZ00404, APKZ02206, APKZ02207, APKY00745, APKZ00995, APKZ00591, APKZ00968, APKZ00969, APKZ00970, APKZ00971, APKY02015, APKY01714, APKY01616, APKY01708, APKY01709, APKY01710, APKY01712, APKY01713, APKY01715, APKY01716, APKY01717, APKY02090, APKZ00906, APKZ00942, APKZ00377, APKY01994, APKY01995, APKY01996, APKY01997, APKY02002, APKY02003, APKY02004, APKY02005, APKZ00967, APKZ00592, APKZ00593, APKY01333, APKY01334, APKY01335, APKY01336, APKZ00136, APKZ00137, APKZ00966, APKA01504, APKA01505, APKY02006, APKY02007, APKY02008, APKY02009, APKY02010, APKY02011, APKY02013, APKY02014, APKZ00176, APKZ00177, APKZ00178, APKZ00179, APKZ00180, APKZ01687, APKZ01688, APKZ02173, APKZ02169, APKY01706, APKY01707, APKY01711, APKA01123, APKY02490, APKZ00575, APKZ00576, APKZ00577, APKZ00578, APKZ00579, APKZ02170, APKZ02208, APKZ02209, APKZ02521, APKY02016, APKY02017, APKY02486, APKY02487, APKZ00183, APKZ00589, APKZ00590, APKZ00965, APKZ02171, APKZ02172, APKA01474, APKA01475, APKA01476, APKA01477, APKA01478, APKA01479, APKY00808, APKY01120, APKY01294, APKY01295, APKZ00999, APKZ01000, APKZ01472, APKY01764, APKY01765, APKZ02255, APKZ02256, APKY02381, APKZ01128, APKA01500, APKA01501, APKZ00605, APKY01630, APKZ00106, APKY01329, APKY01776, APKY01777, APKA00798, APKY02376, APKY02380, APKZ00284, APKZ02176, APKZ02177, APKZ02178, APKZ02179, APKZ02180, APKA01236, APKZ02513, APKY01062, APKZ00242, APKY01638, APKZ02174, APKZ02175, APKY02012, APKZ01961, APKA00289, APKA00290, APKZ00956, APKA00207, APKA00208, APKY01993, APKA01192, APKZ01666, APKY00711, APKZ02514, APKY01375, APKY01845, APKY00968, APKZ01824, APKA00703, APKA01782, APKZ01628, APKZ00402, APKZ00403, APKA00993, APKY01057, APKZ01127, APKZ01730, APKZ01731, APKZ01732, APKA00764, APKA00765, APKA00766, APKA00767, APKA00675, APKA00676, APKA00677, APKA00678, APKA01472, APKA01473, APKZ00399, APKZ00400, APKA00710, APKA01111, APKA01489, APKY02534, APKY02344, APKY02345, APKZ00417, APKZ00418, APKZ00419, APKZ00420, APKZ00421, APKZ00683, APKZ02520, APKZ00545, APKY00905, APKA00864, APKA01490, APKA01491, APKA01492, APKA01495, APKY01117, APKY01118, APKY01119, APKY01602, APKY02310, APKY02311, APKY02312, APKY02313, APKY02314, APKZ00281, APKZ00282, APKZ00283, APKZ00927, APKZ01324, APKZ01325, APKZ01326, APKZ01327, APKZ02541, APKA00704, APKA01434, APKY00906, APKZ02515, APKZ02516, APKZ02517, APKZ02518, APKZ02519, APKY02537, APKY02538, APKY01782, APKY00867, APKZ00162, APKZ00163, APKZ00164, APKZ00165, APKZ00166, APKZ00167, APKZ00168, APKZ00169, APKZ00170, APKZ00171, APKZ00172, APKA00863, APKA00865, APKA01112, APKA01113, APKA01114, APKA01121, APKA01122, APKA01480, APKA01481, APKA01482, APKA01483, APKA01484, APKA01493, APKA01494, APKY00868, APKY00869, APKY00870, APKY00921, APKY01274, APKY01278, APKY01279, APKY01841, APKY01842, APKY01843, APKY01844, APKY02044, APKZ00395, APKZ00924, APKZ00929, APKZ00930, APKZ01724, APKZ01725, APKZ01726, APKZ01727, APKZ01940, APKA01305, APKA01306, APKA01307, APKA01308, APKA01309, APKA01310, APKA01311, APKA01312, APKA01313, APKA01314, APKZ02108, APKZ02109, APKZ02110, APKZ02111, APKZ02112, APKZ02113, APKZ02114, APKZ02115, APKZ02116, APKZ02117, APKA00739, APKY02349, APKY02350, APKY02351, APKY02352, APKZ00187, APKZ00188, APKZ00189, APKZ00190, APKZ00191, APKZ02510, APKA00591, APKA00999, APKA01000, APKY01346, APKZ00133, APKZ00134, APKZ01270, APKZ01271, APKZ01272, APKZ01273, APKZ01274, APKZ01275, APKZ01276, APKZ01277, APKZ01278, APKZ01279, APKZ01280, APKZ01281, APKZ01282, APKZ01312, APKZ01313, APKZ01314, APKZ01315, APKZ01316, APKZ01544, APKZ01545, APKZ01546, APKZ01547, APKZ01548, APKZ01549, APKZ01550, APKZ01551, APKZ01552, APKA00292, APKA00293, APKA00294, APKY00916, APKY00917, APKZ02057, APKZ00542, APKZ00543, APKZ00544, APKZ01597, APKZ01598, APKZ01613, APKZ01668, APKZ01669, APKZ01670, APKZ00881, APKZ00882, APKZ00883, APKZ00884, APKZ00885, APKZ00886, APKZ00887, APKZ00888, APKZ01318, APKZ01319, APKZ01320, APKZ01321, APKZ01322, APKZ01323, APKY00746, APKY01400, APKZ00422, APKZ02016, APKY02385, APKZ00401, APKZ01219, APKZ01220, APKZ02531, APKY01828, APKZ02532, APKY01984, APKY01985, APKY02034, APKZ01814, APKZ01815, APKZ02002, APKY00979, APKY01981, APKZ01951, APKZ02529, APKZ02530, APKZ00914, APKY01579, APKY01831, APKY01832, APKZ01953, APKZ01954, APKY02045, APKY02046, APKY02047, APKY02048, APKY02049, APKY02050, APKY02051, APKY01409, APKY00735, APKZ00396, APKZ00915, APKZ01813, APKY01145, APKY01146, APKZ02533, APKA01142, APKA01143, APKZ02534, APKY01980, APKZ01952, APKY01978, APKY01979, APKZ00361, APKZ00362, APKZ00363, APKZ00364, APKZ00365, APKZ00789, APKZ00790, APKA00886, APKA00887, APKA00888, APKA00889, APKA00890, APKA00891, APKA00892, APKA00893, APKA00894, APKA00895, APKA00896, APKA00897, APKA00898, APKA00899, APKA00900, APKA00901, APKA00902, APKA00903, APKA00904, APKA00905, APKA00906, APKA00907, APKA00908, APKA00909, APKA00965, APKA00966, APKA00967, APKY01138, APKY01139, APKY01140, APKY01141, APKY01142, APKY01143, APKY01144, APKY01187, APKY01188, APKY01189, APKY01190, APKY01191, APKY01192, APKY01193, APKY01194, APKY01195, APKY01196, APKY01197, APKY01198, APKY01199, APKY01200, APKY01201, APKY01202, APKY01203, APKY01204, APKY01205, APKY01206, APKY01207, APKY01208, APKY01209, APKY01210, APKY02187, APKY02188, APKY02189, APKY02190, APKY02191, APKY02192, APKY02193, APKY02194, APKY02195, APKY02196, APKY02197, APKY02198, APKY02199, APKY02200, APKY02201, APKY02202, APKY02203, APKY02204, APKY02205, APKY02206, APKY02207, APKY02208, APKY02209, APKY02210, APKY02211, APKY02212, APKY02213, APKY02214, APKY02215, APKY02216, APKY02217, APKY02218, APKY02219, APKY02220, APKY02221, APKY02222, APKY02223, APKY02224, APKY02225, APKY02226, APKY02227, APKY02228, APKY02229, APKY02230, APKY02231, APKY02232, APKY02233, APKY02234, APKY02235, APKY02236, APKY02237, APKY02238, APKY02239, APKY02240, APKY02241, APKY02242, APKY02243, APKY02244, APKY02245, APKY02246, APKY02254, APKY02255, APKY02256, APKY02257, APKY02258, APKY02259, APKA01816, APKZ00289, APKZ00290, APKZ00291, APKZ00292, APKZ00293, APKA00291, APKA00431, APKA00432, APKA00433, APKA00434, APKA00435, APKA00436, APKA00437, APKA00438, APKA00439, APKA00440, APKA00441, APKA00442, APKA01808, APKA01809, APKA01811, APKA01812, APKA01813, APKA01928, APKZ02509, APKY02091, APKY02106, APKY02481, APKZ00617, APKZ01090, APKZ01091, APKZ01092, APKZ01093, APKZ01094, APKZ01095, APKZ01096, APKZ01097, APKZ01098, APKZ01099, APKY01668, APKY01669, APKY01670, APKY01671, APKY01672, APKY01673, APKY01674, APKY01675, APKY01676, APKY01677, APKY01678, APKY01679, APKY01680, APKY01681, APKY01682, APKY01683, APKY01684, APKZ02375, APKA01774, APKA01775, APKA01776, APKA00287, APKA00288, APKZ00423, APKZ01647, APKZ01648, APKZ01649, APKZ01650, APKZ01652, APKZ01653, APKZ01654, APKZ01655, APKZ01656, APKA00443, APKA00444, APKY01381, APKY01382, APKY01383, APKZ02193, APKY01446, APKY02033, APKY02062, APKA00258, APKA00964, APKA00953, APKA00954, APKA00955, APKA00956, APKA00957, APKA00958, APKA00959, APKA00960, APKA00961, APKA00962, APKA00996, APKA00997, APKA00963, APKY01766, APKZ00279, APKZ00280, APKZ01014, APKZ01015, APKZ01016, APKZ01017, APKZ01034, APKA01471, APKA01206, APKA01207, APKY01090, APKY01091, APKY01092, APKY02539, APKZ01031, APKZ01032, APKZ01033, APKZ01785, APKZ01786, APKZ01787, APKY00799, APKY00800, APKY00801, APKY01093, APKZ02277, APKZ02278, APKY00922, APKZ02118, APKZ00655, APKZ01029, APKZ01030, APKA00689, APKA01118, APKA01119, APKA01120, APKY01343, APKY01778, APKZ01027, APKZ01028, APKA00688, APKA00690, APKA00691, APKA00692, APKA00693, APKA00694, APKY01977, APKY02035, APKY02036, APKZ00267, APKZ00268, APKZ00392, APKZ00600, APKZ02279, APKZ00618, APKY01433, APKY01434, APKY01435, APKY01436, APKY01487, APKY01488, APKY01489, APKY01490, APKY01491, APKY01492, APKY01493, APKY01494, APKY01495, APKY01496, APKY01497, APKY01498, APKY01499, APKY01500, APKZ01921, APKA01486, APKZ01867, APKZ01868, APKZ02087, APKZ02275, APKZ02276, APKZ02339, APKA00994, APKA00995, APKY02353, APKY02354, APKY02355, APKY02356, APKY02357, APKY01523, APKY01524, APKY01525, APKY01526, APKY01527, APKY01528, APKY02348, APKY01183, APKY02092, APKZ02069, APKZ02070, APKZ02071, APKY02485, APKY02489, APKA01108, APKA01109, APKY00803, APKY00804, APKY00805, APKY00806, APKY00807, APKY00829, APKY00830, APKY00831, APKY00832, APKY00833, APKY00834, APKY02533, APKZ00850, APKZ00851, APKZ00852, APKA00597, APKA00598, APKA00551, APKA01107, APKA01497, APKZ00960, APKZ00961, APKZ00962, APKZ00963, APKZ00964, APKY01787, APKY01788, APKY02346, APKY02347, APKZ01300, APKZ01317, APKZ01301, APKA01115, APKY01347, APKA00549, APKZ00181, APKY01850, APKY01296, APKY01297, APKZ00903, APKZ00904, APKZ00905, APKZ00838, APKA01116, APKA01526, APKA01527, APKA01528, APKA01529, APKA01530, APKA01662, APKA01663, APKA01664, APKA01665, APKA01718, APKA01765, APKY01080, APKY01081, APKY01082, APKY01083, APKY02282, APKZ00638, APKZ00675, APKZ00676, APKZ01980, APKZ02511, APKZ02512, APKA00789, APKA00790, APKA00795, APKA00796, APKA00797, APKA01138, APKZ01737, APKA01815, APKA01667, APKA01668, APKA01669, APKY02162, APKY02164, APKZ02234, APKY00946, APKY00947, APKY00997, APKY00998, APKA01578, APKA01579, APKA01580, APKA01581, APKA01582, APKA01583, APKA01584, APKA01721, APKA01757, APKA01758, APKY01862, APKY01289, APKY01290, APKY01291, APKY01292, APKZ00410, APKZ01711, APKZ00328, APKA01577, APKZ01901, APKZ01902, APKZ01903, APKZ01904, APKZ01905, APKZ01906, APKZ01907, APKZ01908, APKZ01909, APKZ01910, APKZ01923, APKZ01924, APKZ01925, APKZ01926, APKZ01927, APKZ01928, APKZ00184, APKZ00608, APKZ00609, APKZ00610, APKZ01424, APKY01631, APKY01632, APKY01633, APKY01635, APKZ01302, APKZ01303, APKY01986, APKZ00599, APKA01203, APKA01204, APKA01205, APKY01634, APKA01134, APKA01135, APKA01136, APKA01137, APKZ00849, APKA01093, APKA01574, APKA01575, APKA01576, APKA01572, APKA01573, APKA01814, APKZ00677, APKZ00678, APKY01987, APKY02176, APKY02177, APKY02178, APKY02179, APKY02180, APKY02181, APKY02182, APKY02183, APKY02184, APKY02185, APKY02186, APKY01286, APKY01287, APKY01288, APKY02163, APKY01273, APKZ00182, APKY01407, APKY01408, APKY02270, APKY02271, APKZ01360, APKZ01383, APKA02026, APKY02125, APKZ01361, APKZ01362, APKZ01363, APKZ01364, APKZ01365, APKZ01366, APKZ01367, APKZ01382, APKZ01384, APKZ01385, APKY01308, APKY01309, APKY01310, APKY01311, APKY01312, APKY01313, APKY01314, APKY01315, APKY02075, APKZ00159, APKZ00907, APKZ00908, APKZ01503, APKA00175, APKA00176, APKY02530, APKZ00405, APKZ00642, APKZ00643, APKZ00644, APKZ00684, APKZ00685, APKZ00686, APKZ00687, APKZ00688, APKZ00689, APKZ00690, APKZ00691, APKY02114, APKY02115, APKY02116, APKY02321, APKY02322, APKY02323, APKY02324, APKY02325, APKY02326, APKA00509, APKA00589, APKA00590, APKA01766, APKB00119, APKY01332, APKY01342, APKY01829, APKY02382, APKY02383, APKY02384, APKY02540, APKY02541, APKZ00207, APKZ00229, APKZ00230, APKZ00231, APKZ00909, APKZ00910, APKZ00911, APKZ00912, APKZ00913, APKZ01607, APKZ01608, APKZ01609, APKZ01610, APKZ01612, APKZ01635, APKZ01636, APKZ01637, APKZ01638, APKZ01779, APKZ01982, APKZ01983, APKZ02187, APKZ02188, APKZ02189, APKZ02190, APKA01777, APKA01778, APKA01779, APKA00107, APKY01214, APKA01761, APKZ01667, APKA00815, APKY02691, APKY01866, APKY01867, APKZ00105, APKA01760, APKA01764, APKY01988, APKY01989, APKY02526, APKZ00613, APKZ00972, APKY00996, APKA01762, APKA01763, APKY00967, APKY01045, APKY01275, APKY01276, APKY02167, APKY02168, APKY02532, APKY02556, APKZ00360, APKZ00682, APKY01131, APKZ00104, APKY02531, APKZ02210, APKZ02211, APKZ00992, APKZ00993, APKZ00994, APKA00565, APKY00744, APKY01184, APKY01185, APKY01186, APKY02529, APKY02093, APKZ02404, APKY01823, APKY01824, APKZ00138, APKZ00394, APKY00918, APKY00919, APKA00361, APKY02521, APKA01496, APKY01423, APKZ00603, APKZ00604, APKZ02068, APKY02309, APKA01621, APKA01619, APKA01772, APKA01773, APKB00158, APKA01620, APKA01622, APKA01771, APKY01268, APKY01269, APKY01270, APKY01271, APKY01272, APKY01277, APKY01293, APKY01581, APKY01582, APKY01583, APKZ01673, APKZ01674, APKZ01675, APKZ01676, APKY01539, APKZ01500, APKZ01702, APKY02104, APKZ00359, APKY00884, APKY01447, APKY01448, APKY01449, APKY01450, APKY01451, APKY01452, APKY01453, APKY01454, APKY01455, APKY01456, APKY01479, APKY01480, APKY01481, APKY01482, APKY01483, APKY01484, APKA00730, APKA00732, APKA00733, APKA00738, APKA01784, APKB00357, APKB00358, APKB00359, APKY00891, APKY00896, APKY00897, APKY00898, APKY00899, APKY01460, APKY01461, APKY02336, APKY02337, APKZ01230, APKZ01231, APKZ01232, APKZ01233, APKZ01234, APKZ01235, APKZ01495, APKZ01497, APKZ01498, APKZ01514, APKZ01515, APKZ01516, APKZ01522, APKZ01523, APKZ01929, APKZ01930, APKZ01931, APKZ01932, APKZ01933, APKZ00945, APKA00581, APKA01785, APKA01787, APKA01788, APKA01789, APKA01790, APKA01799, APKA01800, APKA01801, APKA01802, APKA01803, APKA01805, APKZ01936, APKZ02263, APKZ00955, APKY00883, APKZ01520, APKZ02546, APKA00734, APKZ02551, APKY00889, APKY00890, APKY00895, APKZ00948, APKZ02548, APKY02341, APKA00719, APKZ01518, APKZ00424, APKA00729, APKA01804, APKA00731, APKY00886, APKY01443, APKY02331, APKY02339, APKZ00947, APKZ01496, APKZ02547, APKZ02549, APKZ02552, APKA00726, APKY00887, APKY00888, APKY01439, APKY01466, APKZ00426, APKZ00950, APKZ01521, APKZ02260, APKZ02261, APKZ02262, APKZ02543, APKA00736, APKZ00431, APKA00735, APKY01485, APKZ01935, APKZ01517, APKY02329, APKZ00953, APKZ00954, APKY00879, APKY00881, APKY00892, APKY00893, APKY02332, APKZ00430, APKY02335, APKY00894, APKY01465, APKA00580, APKY02333, APKZ02257, APKA00720, APKA00721, APKY01441, APKY01442, APKY02340, APKZ01499, APKA00727, APKA00728, APKA01783, APKY01486, APKY02334, APKZ02550, APKY00901, APKY02327, APKA01786, APKZ00425, APKZ00427, APKZ00428, APKZ00429, APKY00885, APKZ00946, APKZ00951, APKZ01934, APKZ02258, APKY01440, APKY01462, APKY01464, APKY00880, APKY00900, APKZ01519, APKZ00952, APKA00737, APKY00882, APKZ02259, APKY02338, APKZ02544, APKZ02545, APKA00579, APKA00722, APKA00723, APKA00724, APKA00725, APKY01463, APKZ00949, APKY02328, APKY02330, APKA01110, APKZ01444, APKZ01452, APKZ00527, APKY02267, APKY02268, APKY02269, APKY02551, APKZ01438, APKZ01441, APKZ01448, APKZ01449, APKY01854, APKY01855, APKY01856, APKY00752, APKY00753, APKY00754, APKY00755, APKY00756, APKY00757, APKY00758, APKY00759, APKY00760, APKY00761, APKY00762, APKY00763, APKY00764, APKY00765, APKY00766, APKY00767, APKY00768, APKY00769, APKY00770, APKY00771, APKY00772, APKY00773, APKY00774, APKY00775, APKA01498, APKA01499, APKY00747, APKY00748, APKY00749, APKZ01446, APKZ01447, APKY02158, APKY02159, APKY02160, APKY02161, APKZ01440, APKZ01443, APKZ01455, APKY01729, APKY01730, APKY01731, APKY01732, APKY01733, APKY01734, APKY02552, APKY02553, APKY02554, APKY02555, APKZ00142, APKZ00232, APKZ00444, APKZ00645, APKZ00646, APKZ00647, APKZ00648, APKZ00649, APKZ00650, APKZ00651, APKZ00652, APKZ00653, APKZ00654, APKZ01442, APKZ01445, APKZ01450, APKZ01453, APKZ00135, APKZ00443, APKZ00446, APKZ01439, APKZ01451, APKZ01454, APKY00931, APKY00933, APKY02266, APKY00930, APKY00932, APKZ00445, APKZ00457, APKA00331, APKA00332, APKA01570, APKA01571, APKA00577, APKY02081, APKZ00140, APKZ00141, APKZ00854, APKA00621, APKA01055, APKA01709, APKA01710, APKA01711, APKY01836, APKY02265, APKZ00447, APKZ00448, APKZ00853, APKZ02058, APKZ02527, APKZ02528, APKZ01386, APKZ01387, APKA00142, APKA00143, APKA00144, APKA00145, APKA00146, APKA00147, APKA00148, APKA00149, APKA00150, APKA00151, APKA00152, APKA00153, APKA00154, APKA00155, APKA00156, APKA00157, APKA00259, APKA00260, APKA00261, APKA00262, APKA00263, APKA00264, APKY02079, APKY02080, APKZ00196, APKZ00197, APKZ00198, APKZ00205, APKZ00206, APKA01091, APKZ00518, APKZ00519, APKZ00520, APKZ02522, APKZ02523, APKZ00584, APKZ00585, APKZ00586, APKZ00587, APKA00397, APKY01639, APKY01640, APKY01641, APKA01318, APKZ00336, APKZ01477, APKZ01478, APKZ01639, APKZ01640, APKZ01641, APKZ01642, APKZ01643, APKZ01644, APKZ01645, APKZ01646, APKZ01651, APKZ02327, APKZ02328, APKZ02329, APKZ02330, APKA00510, APKA00511, APKA00512, APKA00513, APKA00514, APKA00515, APKZ00411, APKZ00412, APKY02522, APKY02523, APKY02524, APKY02525, APKY01597, APKY01598, APKY00864, APKZ01348, APKZ01306, APKZ01307, APKZ01308, APKZ01309, APKZ01310, APKZ01311, APKY01574, APKY01825, APKY01826, APKY02151, APKY02247, APKY02248, APKY02249, APKY02250, APKY02251, APKY02252, APKY01827, APKY01846, APKY02527, APKZ00337, APKZ00338, APKZ00339, APKZ00340, APKZ00341, APKZ00342, APKZ00343, APKZ00344, APKZ00345, APKZ00346, APKZ00347, APKZ00348, APKZ00294
|
| FEI Number |
2126677
|
Recalling Firm/ Manufacturer |
GE Healthcare, LLC
3000 N Grandview Blvd
Waukesha
WI
53188-1615
|
| For Additional Information Contact | GE Healthcare Service 800-437-1171 |
Manufacturer Reason for Recall | Backup batteries can fail earlier than their estimated life. |
FDA Determined Cause 2 | Under Investigation by firm |
| Action | An Urgent Medical Device Correction letter dated 04/25/2022 was sent to customers with affected devices. The letter mentions the users can continue to use the ventilator while connected to an AC mains power source that is supported by backup emergency power. The letter provides a series of recommendations for the consignee to follow. The letter also requests that the consignee complete the medical device notification acknowledgement response form and send it back to the firm. |
| Quantity in Commerce | 6,612 devices |
| Distribution | Worldwide distribution - US Nationwide and the countries of Algeria, Australia, Austria, Bangladesh, Belgium, Brazil, China, Columbia, Costa Rica, Cyprus, Czech Republic, Denmark, Egypt, France, Germany, Greece, Hong Kong, India, Indonesia, Iraq, Italy, Japan, Kazakhstan, Kenya, Korea, Kuwait, Lao PDR, Latvia, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Mexico, Morocco, Nepal, New Zealand, Norway, Oman, Panama, Peru, Philippines, Poland, Portugal, Romania, Russia, Saudi Arabia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, Ukraine, United Arab Emirates, United Kingdom, and Vietnam. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = BSZ
|
|
|
|