| | Class 2 Device Recall MiniMed 620G Insulin Pump |  |
| Date Initiated by Firm | May 23, 2022 |
| Create Date | July 08, 2022 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-1352-2022 |
| Recall Event ID |
90370 |
| Product Classification |
Pump, infusion, insulin, to be used with invasive glucose sensor - Product Code OYC
|
| Product | MiniMed 620G Insulin Pump (O.U.S. version) |
| Code Information |
Product Number/CFN (O.U.S. version)
MiniMed 620G (MMT-1510, MMT-1710, MMT-1750)
UDI Code: Pending
Serial Numbers:
NG1003917H
NG1004461H
NG1010799H
NG1012557H
NG1020680H
NG1026110H
NG1028941H
NG1028949H
NG1028952H
NG1028973H
NG1034230H
NG1034247H
NG1034304H
NG1034314H
NG1034335H
NG1034338H
NG1034345H
NG1034358H
NG1034359H
NG1034364H
NG1034583H
NG1034584H
NG1034593H
NG1034595H
NG1034609H
NG1035490H
NG1035495H
NG1035505H
NG1035517H
NG1035520H
NG1035525H
NG1035526H
NG1035546H
NG1035552H
NG1035566H
NG1035572H
NG1035575H
NG1035581H
NG1035586H
NG1035593H
NG1035595H
NG1035915H
NG1035930H
NG1035933H
NG1035935H
NG1036645H
NG1036646H
NG1036671H
NG1036678H
NG1036688H
NG1036704H
NG1036713H
NG1036715H
NG1036725H
NG1036726H
NG1036731H
NG1036757H
NG1036785H
NG1036801H
NG1036809H
NG1036811H
NG1036813H
NG1036854H
NG1039152H
NG1040219H
NG1040230H
NG1040231H
NG1040237H
NG1040257H
NG1040262H
NG1040271H
NG1040290H
NG1040298H
NG1040305H
NG1040318H
NG1040344H
NG1040353H
NG1040380H
NG1040381H
NG1040383H
NG1040480H
NG1040483H
NG1040628H
NG1040637H
NG1040667H
NG1040668H
NG1040737H
NG1040999H
NG1041013H
NG1041041H
NG1041067H
NG1041078H
NG1041079H
NG1041134H
NG1041165H
NG1041169H
NG1041212H
NG1041229H
NG1041254H
NG1041262H
NG1043526H
NG1043542H
NG1043547H
NG1043549H
NG1043570H
NG1043571H
NG1043574H
NG1043585H
NG1043587H
NG1043589H
NG1043600H
NG1043605H
NG1043618H
NG1043619H
NG1043629H
NG1043637H
NG1043642H
NG1043656H
NG1043659H
NG1048215H
NG1051985H
NG1051990H
NG1052098H
NG1054459H
NG1054461H
NG1055929H
NG1055937H
NG1055959H
NG1055980H
NG1055987H
NG1056018H
NG1056021H
NG1056025H
NG1057684H
NG1057686H
NG1057690H
NG1059069H
NG1059076H
NG1059083H
NG1059107H
NG1059119H
NG1059125H
NG1059126H
NG1059151H
NG1059154H
NG1059178H
NG1059187H
NG1059297H
NG1059307H
NG1059313H
NG1059319H
NG1059326H
NG1059356H
NG1059367H
NG1059386H
NG1059411H
NG1059419H
NG1059424H
NG1059440H
NG1060999H
NG1066735H
NG1066747H
NG1066778H
NG1066783H
NG1066788H
NG1066803H
NG1066816H
NG1066821H
NG1066823H
NG1066827H
NG1066854H
NG1066863H
NG1066871H
NG1066891H
NG1066893H
NG1066894H
NG1066897H
NG1066899H
NG1066902H
NG1066904H
NG1066906H
NG1066909H
NG1066916H
NG1068481H
NG1068815H
NG1068819H
NG1068829H
NG1068831H
NG1068832H
NG1068838H
NG1068841H
NG1068842H
NG1068850H
NG1068892H
NG1073307H
NG1073320H
NG1073321H
NG1073336H
NG1073354H
NG1073371H
NG1073381H
NG1073392H
NG1073406H
NG1073412H
NG1073413H
NG1073415H
NG1076044H
NG1076047H
NG1076063H
NG1076079H
NG1076086H
NG1076103H
NG1076128H
NG1076132H
NG1076148H
NG1076153H
NG1076156H
NG1076170H
NG1076181H
NG1076187H
NG1076212H
NG1076217H
NG1082514H
NG1082516H
NG1082524H
NG1082527H
NG1082532H
NG1082534H
NG1082538H
NG1082559H
NG1082564H
NG1082567H
NG1082568H
NG1082571H
NG1082572H
NG1082587H
NG1082589H
NG1082594H
NG1082595H
NG1082597H
NG1082602H
NG1082641H
NG1082645H
NG1082650H
NG1082667H
NG1082672H
NG1082680H
NG1082682H
NG1082684H
NG1082694H
NG1082700H
NG1082702H
NG1082705H
NG1082737H
NG1082739H
NG1082741H
NG1084306H
NG1084330H
NG1084341H
NG1084348H
NG1084350H
NG1084360H
NG1084362H
NG1084368H
NG1084369H
NG1084370H
NG1084458H
NG1084462H
NG1084472H
NG1084477H
NG1084494H
NG1084498H
NG1084753H
NG1084760H
NG1084768H
NG1084770H
NG1084771H
NG1084773H
NG1084776H
NG1084777H
NG1084795H
NG1084800H
NG1084801H
NG1084805H
NG1084806H
NG1084812H
NG1084816H
NG1084820H
NG1084827H
NG1084842H
NG1084852H
NG1084867H
NG1084881H
NG1084884H
NG1084886H
NG1084891H
NG1084905H
NG1084906H
NG1084911H
NG1084926H
NG1084935H
NG1085346H
NG1085359H
NG1085366H
NG1085372H
NG1085385H
NG1085392H
NG1085407H
NG1085408H
NG1085413H
NG1085424H
NG1085428H
NG1085444H
NG1085463H
NG1085464H
NG1085476H
NG1085478H
NG1085493H
NG1085494H
NG1085497H
NG1085510H
NG1085525H
NG1085527H
NG1085528H
NG1085529H
NG1085537H
NG1093874H
NG1093885H
NG1093897H
NG1093905H
NG1093907H
NG1099997H
NG1100000H
NG1100001H
NG1100007H
NG1100011H
NG1100020H
NG1100025H
NG1100027H
NG1100035H
NG1100044H
NG1100080H
NG1100087H
NG1100096H
NG1100110H
NG1100123H
NG1100215H
NG1100219H
NG1100221H
NG1100235H
NG1100237H
NG1100256H
NG1100271H
NG1100295H
NG1100303H
NG1100305H
NG1100315H
NG1100316H
NG1100326H
NG1100336H
NG1107782H
NG1107788H
NG1107791H
NG1107801H
NG1107803H
NG1107809H
NG1107812H
NG1107833H
NG1107835H
NG1107848H
NG1107874H
NG1107896H
NG1107904H
NG1107907H
NG1107916H
NG1107930H
NG1113607H
NG1113629H
NG1113630H
NG1113631H
NG1113653H
NG1113672H
NG1113678H
NG1113679H
NG1113683H
NG1113692H
NG1113697H
NG1113700H
NG1113707H
NG1113719H
NG1113736H
NG1113739H
NG1113767H
NG1113775H
NG1113782H
NG1113788H
NG1113801H
NG1113802H
NG1113805H
NG1113810H
NG1113812H
NG1113813H
NG1113832H
NG1113842H
NG1113843H
NG1113893H
NG1113898H
NG1113913H
NG1113916H
NG1113927H
NG1113933H
NG1113940H
NG1113951H
NG1113955H
NG1113964H
NG1113975H
NG1113985H
NG1126343H
NG1126448H
NG1126449H
NG1126463H
NG1126476H
NG1126488H
NG1126494H
NG1126499H
NG1126533H
NG1126551H
NG1126552H
NG1126555H
NG1126593H
NG1126608H
NG1126612H
NG1126625H
NG1126627H
NG1126646H
NG1126655H
NG1126723H
NG1126735H
NG1126745H
NG1126750H
NG1126754H
NG1126772H
NG1126800H
NG1126810H
NG1126830H
NG1126832H
NG1126845H
NG1144128H
NG1144143H
NG1144158H
NG1144177H
NG1144181H
NG1144183H
NG1144185H
NG1148631H
NG1148647H
NG1148654H
NG1153641H
NG1159764H
NG1159778H
NG1159781H
NG1159782H
NG1159790H
NG1159798H
NG1159799H
NG1159816H
NG1159827H
NG1159828H
NG1159836H
NG1159848H
NG1159850H
NG1159853H
NG1159863H
NG1159866H
NG1159886H
NG1165870H
NG1171819H
NG1171821H
NG1171827H
NG1171834H
NG1171837H
NG1171839H
NG1171841H
NG1171846H
NG1171859H
NG1171863H
NG1171865H
NG1171897H
NG1171907H
NG1173122H
NG1173512H
NG1175448H
NG1175463H
NG1176432H
NG1176440H
NG1176447H
NG1176454H
NG1176472H
NG1176476H
NG1176485H
NG1176488H
NG1176489H
NG1176493H
NG1176514H
NG1176528H
NG1176536H
NG1176545H
NG1176550H
NG1176565H
NG1176571H
NG1176574H
NG1176585H
NG1176610H
NG1176612H
NG1176617H
NG1176630H
NG1176639H
NG1176640H
NG1176650H
NG1176651H
NG1176652H
NG1176654H
NG1176668H
NG1176672H
NG1177264H
NG1177917H
NG1180905H
NG1182525H
NG1182682H
NG1182707H
NG1183396H
NG1183414H
NG1190399H
NG1190731H
NG1192060H
NG1192104H
NG1192115H
NG1192128H
NG1192140H
NG1192145H
NG1192147H
NG1192158H
NG1192160H
NG1192161H
NG1192162H
NG1192165H
NG1192170H
NG1195230H
NG1197048H
NG1204061H
NG1204075H
NG1204090H
NG1204107H
NG1204112H
NG1204116H
NG1204120H
NG1204130H
NG1204131H
NG1204137H
NG1204139H
NG1204628H
NG1205370H
NG1205704H
NG1205712H
NG1217864H
NG1217872H
NG1217883H
NG1217891H
NG1217892H
NG1217893H
NG1217899H
NG1217900H
NG1217904H
NG1217912H
NG1217923H
NG1217944H
NG1217952H
NG1217953H
NG1217957H
NG1217966H
NG1217972H
NG1217984H
NG1217998H
NG1218011H
NG1218029H
NG1218037H
NG1218041H
NG1218051H
NG1218073H
NG1218080H
NG1218081H
NG1218087H
NG1218090H
NG1224836H
NG1225113H
NG1226007H
NG1226994H
NG1228066H
NG1228072H
NG1230937H
NG1230969H
NG1231486H
NG1231493H
NG1231753H
NG1235372H
NG1236922H
NG1236932H
NG1236940H
NG1236947H
NG1236956H
NG1236958H
NG1236960H
NG1236967H
NG1236982H
NG1236983H
NG1236984H
NG1236988H
NG1236990H
NG1236991H
NG1236999H
NG1237001H
NG1237004H
NG1237010H
NG1237014H
NG1237017H
NG1237021H
NG1237025H
NG1237026H
NG1237575H
NG1237579H
NG1237582H
NG1237596H
NG1237597H
NG1237599H
NG1237609H
NG1237615H
NG1237617H
NG1237621H
NG1237629H
NG1237651H
NG1237654H
NG1237660H
NG1237668H
NG1237669H
NG1237678H
NG1237679H
NG1237688H
NG1237692H
NG1237699H
NG1237702H
NG1237712H
NG1237713H
NG1237715H
NG1237717H
NG1237723H
NG1237735H
NG1237739H
NG1237751H
NG1237761H
NG1237772H
NG1237787H
NG1237791H
NG1237794H
NG1237799H
NG1237805H
NG1237808H
NG1237810H
NG1237813H
NG1237816H
NG1237835H
NG1237840H
NG1237861H
NG1237874H
NG1237881H
NG1237882H
NG1243231H
NG1243265H
NG1244124H
NG1244132H
NG1244133H
NG1244141H
NG1244155H
NG1244161H
NG1244171H
NG1251589H
NG1252740H
NG1252960H
NG1254821H
NG1254840H
NG1254841H
NG1254842H
NG1270483H
NG1270484H
NG1270490H
NG1270492H
NG1270493H
NG1270494H
NG1270499H
NG1270504H
NG1270508H
NG1270515H
NG1270516H
NG1270521H
NG1270530H
NG1270538H
NG1270539H
NG1270549H
NG1270553H
NG1270572H
NG1270578H
NG1270585H
NG1270588H
NG1270591H
NG1270597H
NG1270600H
NG1270616H
NG1270618H
NG1270624H
NG1270633H
NG1270634H
NG1270642H
NG1270649H
NG1270661H
NG1270662H
NG1274532H
NG1274538H
NG1274553H
NG1274568H
NG1274569H
NG1274598H
NG1274614H
NG1274616H
NG1274617H
NG1275250H
NG1279314H
NG1279565H
NG1279573H
NG1279610H
NG1279611H
NG1279615H
NG1279620H
NG1279623H
NG1279635H
NG1279647H
NG1279655H
NG1279659H
NG1279660H
NG1279666H
NG1279674H
NG1279678H
NG1279684H
NG1279695H
NG1279696H
NG1279699H
NG1279708H
NG1279713H
NG1279715H
NG1279725H
NG1283014H
NG1283024H
NG1283026H
NG1283027H
NG1283032H
NG1283043H
NG1283048H
NG1283052H
NG1283056H
NG1283059H
NG1283078H
NG1283079H
NG1283085H
NG1283109H
NG1283110H
NG1283114H
NG1283115H
NG1283119H
NG1283121H
NG1283126H
NG1284382H
NG1286292H
NG1286297H
NG1286300H
NG1286305H
NG1286313H
NG1286321H
NG1286323H
NG1286347H
NG1286350H
NG1286355H
NG1286362H
NG1286371H
NG1286379H
NG1286380H
NG1286383H
NG1286385H
NG1286395H
NG1286400H
NG1286403H
NG1286405H
NG1286409H
NG1286410H
NG1286418H
NG1286425H
NG1286433H
NG1290705H
NG1292792H
NG1292794H
NG1292799H
NG1292800H
NG1292802H
NG1292806H
NG1292807H
NG1292808H
NG1292809H
NG1292819H
NG1292826H
NG1292827H
NG1292836H
NG1292841H
NG1292848H
NG1292863H
NG1292873H
NG1292875H
NG1292886H
NG1292905H
NG1292914H
NG1292916H
NG1292921H
NG1292924H
NG1292926H
NG1292930H
NG1292934H
NG1293219H
NG1301055H
NG1301070H
NG1301078H
NG1301080H
NG1301091H
NG1301096H
NG1301103H
NG1301105H
NG1307875H
NG1312042H
NG1325956H
NG1325996H
NG1326028H
NG1326032H
NG1326036H
NG1326039H
NG1326041H
NG1326052H
NG1326075H
NG1326083H
NG1326099H
NG1326114H
NG1326122H
NG1326128H
NG1326130H
NG1326131H
NG1326141H
NG1326148H
NG1326150H
NG1326151H
NG1326155H
NG1326165H
NG1326203H
NG1326206H
NG1326227H
NG1326228H
NG1326229H
NG1326232H
NG1326235H
NG1326236H
NG1326840H
NG1326900H
NG1327305H
NG1327613H
NG1334413H
NG1334420H
NG1334422H
NG1334424H
NG1334425H
NG1334429H
NG1334435H
NG1334436H
NG1334439H
NG1334441H
NG1334464H
NG1334479H
NG1334480H
NG1334481H
NG1334482H
NG1334486H
NG1334489H
NG1334512H
NG1334520H
NG1334530H
NG1334534H
NG1334535H
NG1334546H
NG1334550H
NG1334571H
NG1334579H
NG1334584H
NG1334586H
NG1334609H
NG1334610H
NG1334611H
NG1334614H
NG1334616H
NG1334620H
NG1334623H
NG1334631H
NG1334636H
NG1334637H
NG1334655H
NG1334662H
NG1334665H
NG1334676H
NG1334677H
NG1334689H
NG1334705H
NG1334707H
NG1334710H
NG1336263H
NG1340644H
NG1350013H
NG1350022H
NG1350023H
NG1350030H
NG1350035H
NG1350039H
NG1350043H
NG1350045H
NG1350078H
NG1350085H
NG1350086H
NG1350093H
NG1350102H
NG1350107H
NG1350132H
NG1350135H
NG1350136H
NG1350145H
NG1350150H
NG1350154H
NG1350186H
NG1350191H
NG1350198H
NG1350205H
NG1354773H
NG1357523H
NG1357536H
NG1357538H
NG1357539H
NG1357541H
NG1357552H
NG1357570H
NG1357576H
NG1357584H
NG1358576H
NG1361984H
NG1363118H
NG1363120H
NG1363121H
NG1363124H
NG1363128H
NG1363131H
NG1363134H
NG1363135H
NG1363136H
NG1363162H
NG1363164H
NG1363166H
NG1363167H
NG1363173H
NG1363179H
NG1363181H
NG1363190H
NG1363193H
NG1363202H
NG1363207H
NG1364612H
NG1364733H
NG1364762H
NG1367165H
NG1367182H
NG1367195H
NG1380477H
NG1380857H
NG1383676H
NG1388461H
NG1388476H
NG1388486H
NG1388497H
NG1388513H
NG1388531H
NG1388542H
NG1388543H
NG1388544H
NG1388545H
NG1388548H
NG1388556H
NG1388559H
NG1388572H
NG1388588H
NG1388589H
NG1388595H
NG1388598H
NG1388601H
NG1388603H
NG1388607H
NG1388621H
NG1388627H
NG1388652H
NG1388684H
NG1388691H
NG1388692H
NG1388701H
NG1388702H
NG1388706H
NG1388713H
NG1388717H
NG1388726H
NG1388773H
NG1388774H
NG1388793H
NG1388798H
NG1388816H
NG1388820H
NG1388821H
NG1388822H
NG1388832H
NG1388833H
NG1389570H
NG1391450H
NG1392235H
NG1392738H
NG1396849H
NG1400113H
NG1400641H
NG1401069H
NG1401325H
NG1413039H
NG1413058H
NG1413852H
NG1415236H
NG1416325H
NG1416405H
NG1416865H
NG1416952H
NG1419435H
NG1420232H
NG1420250H
NG1420258H
NG1420264H
NG1420283H
NG1420286H
NG1420289H
NG1420292H
NG1420293H
NG1420313H
NG1420326H
NG1420340H
NG1420357H
NG1420367H
NG1420375H
NG1420385H
NG1420387H
NG1420398H
NG1420409H
NG1420411H
NG1420417H
NG1420424H
NG1420470H
NG1420471H
NG1420475H
NG1420476H
NG1420482H
NG1420505H
NG1420508H
NG1420511H
NG1420516H
NG1420529H
NG1420530H
NG1420531H
NG1420543H
NG1420560H
NG1420569H
NG1420578H
NG1420579H
NG1420588H
NG1420592H
NG1420597H
NG1420600H
NG1420613H
NG1420635H
NG1420640H
NG1420647H
NG1420650H
NG1420656H
NG1420663H
NG1420667H
NG1420672H
NG1420690H
NG1420693H
NG1420703H
NG1420706H
NG1420713H
NG1420731H
NG1420733H
NG1420735H
NG1420742H
NG1420752H
NG1420754H
NG1420762H
NG1420763H
NG1420765H
NG1420768H
NG1420770H
NG1420786H
NG1420788H
NG1420789H
NG1420798H
NG1420801H
NG1420822H
NG1420844H
NG1420849H
NG1420858H
NG1420859H
NG1420861H
NG1420868H
NG1420872H
NG1420873H
NG1420875H
NG1420884H
NG1420914H
NG1420915H
NG1420923H
NG1420925H
NG1420936H
NG1420941H
NG1420943H
NG1420959H
NG1420963H
NG1420967H
NG1423793H
NG1424796H
NG1425157H
NG1425930H
NG1439713H
NG1441411H
NG1442256H
NG1442617H
NG1448616H
NG1448631H
NG1448643H
NG1448646H
NG1448661H
NG1448666H
NG1448678H
NG1448694H
NG1448699H
NG1448712H
NG1448726H
NG1448730H
NG1448731H
NG1448732H
NG1448744H
NG1448750H
NG1448755H
NG1448756H
NG1448758H
NG1448768H
NG1448770H
NG1448773H
NG1453245H
NG1459850H
NG1460054H
NG1472962H
NG1473174H
NG1478485H
NG1478499H
NG1478504H
NG1478527H
NG1478547H
NG1478656H
NG1478705H
NG1478826H
NG1478870H
NG1478946H
NG1478959H
NG1478965H
NG1478971H
NG1478975H
NG1478987H
NG1479010H
NG1479029H
NG1479032H
NG1479043H
NG1479054H
NG1479063H
NG1479067H
NG1479072H
NG1479077H
NG1479079H
NG1479082H
NG1479085H
NG1479093H
NG1479098H
NG1479100H
NG1479143H
NG1479144H
NG1479146H
NG1479149H
NG1479152H
NG1479159H
NG1479161H
NG1479163H
NG1479651H
NG1482500H
NG1482503H
NG1482513H
NG1482535H
NG1482537H
NG1482545H
NG1482553H
NG1482566H
NG1482572H
NG1482585H
NG1482586H
NG1482597H
NG1482599H
NG1482608H
NG1482614H
NG1482617H
NG1482621H
NG1482629H
NG1482630H
NG1482634H
NG1482642H
NG1482643H
NG1482646H
NG1482648H
NG1482650H
NG1482653H
NG1482656H
NG1482660H
NG1482662H
NG1482663H
NG1482669H
NG1485938H
NG1492513H
NG1492521H
NG1492530H
NG1492541H
NG1492542H
NG1492543H
NG1492552H
NG1492603H
NG1492609H
NG1492618H
NG1492619H
NG1492625H
NG1492675H
NG1492693H
NG1492695H
NG1492707H
NG1492710H
NG1492716H
NG1492727H
NG1492732H
NG1492734H
NG1492740H
NG1492741H
NG1492743H
NG1492788H
NG1492789H
NG1492796H
NG1492798H
NG1492802H
NG1492811H
NG1492813H
NG1492815H
NG1492823H
NG1492834H
NG1492840H
NG1492857H
NG1492863H
NG1492870H
NG1492874H
NG1492879H
NG1492892H
NG1501878H
NG1501899H
NG1501910H
NG1501921H
NG1502303H
NG1502316H
NG1502317H
NG1502329H
NG1502333H
NG1502335H
NG1502360H
NG1502361H
NG1502362H
NG1502366H
NG1502368H
NG1502372H
NG1502378H
NG1502379H
NG1502390H
NG1502421H
NG1502437H
NG1502439H
NG1502445H
NG1502448H
NG1509701H
NG1509773H
NG1509782H
NG1509817H
NG1525571H
NG1525859H
NG1525906H
NG1525968H
NG1533310H
NG1533342H
NG1534019H
NG1539920H
NG1539934H
NG1540431H
NG1540644H
NG1545915H
NG1546316H
NG1546381H
NG1546411H
NG1546699H
NG1546943H
NG1547336H
NG1547538H
NG1547540H
NG1547589H
NG1555406H
NG1555424H
NG1555572H
NG1555591H
NG1556438H
NG1558043H
NG1558559H
NG1558713H
NG1558751H
NG1558771H
NG1558978H
NG1561553H
NG1561705H
NG1577153H
NG1577206H
NG1577320H
NG1582527H
NG1582964H
NG1583182H
NG1587333H
NG1587358H
NG1587362H
NG1587363H
NG1587460H
NG1587517H
NG1587674H
NG1587679H
NG1587693H
NG1587750H
NG1587864H
NG1588054H
NG1588127H
NG1588242H
NG1589041H
NG1589062H
NG1589063H
NG1595797H
NG1595832H
NG1595833H
NG1595837H
NG1595840H
NG1595843H
NG1595844H
NG1595845H
NG1595846H
NG1595848H
NG1595849H
NG1595850H
NG1595851H
NG1595853H
NG1595858H
NG1595860H
NG1595861H
NG1595862H
NG1595863H
NG1595865H
NG1595870H
NG1595871H
NG1595872H
NG1595873H
NG1595875H
NG1595878H
NG1595879H
NG1595881H
NG1595885H
NG1595887H
NG1595888H
NG1595889H
NG1595890H
NG1595893H
NG1595894H
NG1595895H
NG1595897H
NG1595898H
NG1595899H
NG1595900H
NG1595901H
NG1595902H
NG1595904H
NG1595905H
NG1595906H
NG1595907H
NG1595908H
NG1595909H
NG1595910H
NG1595921H
NG1595924H
NG1595926H
NG1595931H
NG1595937H
NG1595938H
NG1595940H
NG1595941H
NG1595942H
NG1595944H
NG1595945H
NG1595947H
NG1595949H
NG1595956H
NG1595958H
NG1595959H
NG1595964H
NG1595969H
NG1595971H
NG1595972H
NG1595973H
NG1595976H
NG1595978H
NG1595980H
NG1595981H
NG1595984H
NG1595987H
NG1595990H
NG1595994H
NG1595996H
NG1595997H
NG1596000H
NG1596003H
NG1596004H
NG1596005H
NG1596006H
NG1596007H
NG1596009H
NG1596010H
NG1596011H
NG1596012H
NG1596013H
NG1596014H
NG1596015H
NG1596016H
NG1596017H
NG1596018H
NG1596019H
NG1596020H
NG1596022H
NG1596024H
NG1596028H
NG1596035H
NG1596040H
NG1596041H
NG1596047H
NG1596051H
NG1596056H
NG1596058H
NG1596059H
NG1596060H
NG1596062H
NG1596064H
NG1596065H
NG1596068H
NG1596069H
NG1596073H
NG1596097H
NG1596100H
NG1596101H
NG1596103H
NG1596104H
NG1596106H
NG1596107H
NG1596108H
NG1596109H
NG1596112H
NG1596115H
NG1596116H
NG1596117H
NG1596118H
NG1596119H
NG1596123H
NG1596125H
NG1596126H
NG1596129H
NG1596130H
NG1596131H
NG1596132H
NG1596133H
NG1596134H
NG1596139H
NG1596143H
NG1596144H
NG1596147H
NG1596149H
NG1596150H
NG1596151H
NG1596152H
NG1596165H
NG1596168H
NG1596169H
NG1596173H
NG1596176H
NG1596177H
NG1596179H
NG1596180H
NG1596183H
NG1596185H
NG1596186H
NG1596192H
NG1596195H
NG1596200H
NG1596202H
NG1596203H
NG1596208H
NG1596211H
NG1596212H
NG1596221H
NG1596224H
NG1596227H
NG1596231H
NG1596233H
NG1596234H
NG1596237H
NG1596239H
NG1596245H
NG1596246H
NG1596258H
NG1596260H
NG1596265H
NG1596267H
NG1596270H
NG1596274H
NG1596286H
NG1596290H
NG1596292H
NG1596294H
NG1596295H
NG1596296H
NG1596297H
NG1596299H
NG1596305H
NG1596308H
NG1596317H
NG1596338H
NG1596343H
NG1596344H
NG1596348H
NG1596353H
NG1596368H
NG1596369H
NG1596375H
NG1596380H
NG1628540H
NG1628697H
NG1628795H
NG1628982H
NG1630028H
NG1630554H
NG1634539H
NG1634879H
NG1635361H
NG1635383H
NG1648220H
NG1648232H
NG1648238H
NG1648244H
NG1648264H
NG1648268H
NG1648271H
NG1648273H
NG1648281H
NG1648286H
NG1648287H
NG1648288H
NG1648307H
NG1648330H
NG1648348H
NG1648380H
NG1648415H
NG1648419H
NG1648430H
NG1648443H
NG1648457H
NG1648462H
NG1648463H
NG1648465H
NG1648472H
NG1648481H
NG1648490H
NG1648888H
NG1655795H
NG1655815H
NG1655825H
NG1655837H
NG1655838H
NG1655841H
NG1655844H
NG1655845H
NG1655896H
NG1655898H
NG1656060H
NG1656069H
NG1656147H
NG1656176H
NG1656232H
NG1656246H
NG1656252H
NG1656282H
NG1656306H
NG1656329H
NG1656411H
NG1656412H
NG1656475H
NG1656489H
NG1656536H
NG1656765H
NG1656796H
NG1656815H
NG1659538H
NG1688728H
NG1688731H
NG1688732H
NG1688741H
NG1688743H
NG1688745H
NG1688746H
NG1688750H
NG1688752H
NG1688754H
NG1688758H
NG1688761H
NG1688767H
NG1688769H
NG1688776H
NG1688786H
NG1688791H
NG1688797H
NG1688798H
NG1688818H
NG1688841H
NG1688846H
NG1688852H
NG1688854H
NG1688856H
NG1688857H
NG1688860H
NG1688863H
NG1688868H
NG1688870H
NG1688871H
NG1688875H
NG1688879H
NG1688881H
NG1688883H
NG1688885H
NG1688887H
NG1688888H
NG1688889H
NG1688891H
NG1688893H
NG1688894H
NG1688896H
NG1688897H
NG1688899H
NG1688909H
NG1688930H
NG1688934H
NG1688936H
NG1688938H
NG1688940H
NG1688943H
NG1688945H
NG1688947H
NG1688949H
NG1688954H
NG1688956H
NG1688988H
NG1688990H
NG1688992H
NG1689000H
NG2253581H
NG2253584H
NG2253587H
NG2253589H
NG2253591H
NG2253593H
NG2253595H
NG2253597H
NG2253599H
NG2253601H
NG2253603H
NG2253606H
NG2253608H
NG2253610H
NG2253612H
NG2253614H
NG2253619H
NG2253621H
NG2253623H
NG2253626H
NG2253628H
NG2253630H
NG2253633H
NG2253635H
NG2253643H
NG2253645H
NG2253646H
NG2253649H
NG2253650H
NG2253652H
NG2253654H
NG2253657H
NG2253662H
NG2253664H
NG2253666H
NG2253670H
NG2253674H
NG2253677H
NG2253681H
NG2253683H
NG2253685H
NG2253686H
NG2253687H
NG2253688H
NG2253690H
NG2253692H
NG2253695H
NG2253697H
NG2253700H
NG2253702H
NG2253706H
NG2253708H
NG2253710H
NG2253712H
NG2253718H
NG2253719H
NG2253721H
NG2253724H
NG2253726H
NG2253728H
NG2253731H
NG2253737H
NG2253739H
NG2253743H
NG2253745H
NG2253747H
NG2253749H
NG2253751H
NG2253753H
NG2253758H
NG2253760H
NG2253762H
NG2253763H
NG2253764H
NG2253765H
NG2253767H
NG2253768H
NG2253769H
NG2253770H
NG2253771H
NG2253773H
NG2253774H
NG2253775H
NG2253776H
NG2253777H
NG2253778H
NG2253779H
NG2253781H
NG2253782H
NG2253783H
NG2253784H
NG2253785H
NG2253786H
NG2253787H
NG2253788H
NG2253790H
NG2253791H
NG2253792H
NG2253793H
NG2394724H
NG2394726H
NG2394727H
NG2394728H
NG2394729H
NG2394731H
NG2394732H
NG2394733H
NG2394734H
NG2394737H
NG2394740H
NG2394742H
NG2394744H
NG2394747H
NG2394756H
NG2394757H
NG2394758H
NG2394759H
NG2394762H
NG2394763H
NG2394764H
NG2394765H
NG2394766H
NG2394767H
NG2394768H
NG2394769H
NG2394770H
NG2394771H
NG2394772H
NG2394773H
NG2394775H
NG2394776H
NG2394777H
NG2394779H
NG2394780H
NG2394781H
NG2394782H
NG2394783H
NG2394784H
NG2394785H
NG2394787H
NG2394789H
NG2394792H
NG2394793H
NG2394796H
NG2394799H
NG2394802H
NG2394803H
NG2394804H
NG2394805H
NG2394806H
NG2394807H
NG2394808H
NG2394809H
NG2394810H
NG2394811H
NG2394813H
NG2394815H
NG2394817H
NG2394820H
NG2394822H
NG2394829H
NG2394833H
NG2394834H
NG2394837H
NG2394839H
NG2394841H
NG2394842H
NG2394843H
NG2394844H
NG2394845H
NG2394847H
NG2394848H
NG2394849H
NG2394851H
NG2394852H
NG2394853H
NG2394854H
NG2394855H
NG2394856H
NG2394857H
NG2394858H
NG2394859H
NG2394860H
NG2394863H
NG2394864H
NG2394866H
NG2394868H
NG2394870H
NG2394872H
NG2394874H
NG2394877H
NG2394879H
NG2394880H
NG2394881H
NG2394883H
NG2394884H
NG2394886H
NG2394888H
NG2394891H
NG2394892H
NG2394918H
NG2394920H
NG2394921H
NG2394922H
NG2394924H
NG2394925H
NG2394926H
NG2394927H
NG2394928H
NG2394929H
NG2833969H
NG2834029H
NG2834031H
NG2834033H
NG2834037H
NG2834039H
NG2834044H
NG2834045H
NG2834047H
NG2834049H
NG2834053H
NG2834055H
NG2834057H
NG2834058H
NG2834060H
NG2834064H
NG2834065H
NG2834066H
NG2834069H
NG2834070H
NG2834075H
NG2834076H
NG2834077H
NG2834078H
NG2834079H
NG2834081H
NG2834082H
NG2834084H
NG2834086H
NG2834088H
NG2834089H
NG2834101H
NG2834103H
NG2834104H
NG2834105H
NG2834106H
NG2834108H
NG2834110H
NG2834115H
NG2834117H
NG2834125H
NG2834131H
NG2834133H
NG2834135H
NG2834139H
NG2834141H
NG2834142H
NG2834145H
NG2834154H
NG2834182H
NG2834183H
NG2834184H
NG2834186H
NG2834187H
NG2834190H
NG2834193H
NG2834197H
NG2834199H
NG2834210H
NG2834213H
NG2834219H
NG2834221H
NG2834223H
NG2834228H
NG2834233H
NG2834238H
NG2834244H
NG2834247H
NG2834249H
NG2834252H
NG2834254H
NG2834258H
NG2834261H
NG2834263H
NG2834274H
NG2834282H
NG2834287H
NG2834289H
NG2834292H
NG2834294H
NG2834296H
NG2834297H
NG2834299H
NG2834301H
NG2834303H
NG2834306H
NG2834310H
NG2834312H
NG2834313H
NG2834316H
NG2834318H
NG2834320H
NG2834322H
NG2834324H
NG2834327H
NG2834338H
NG2834340H
NG2834343H
NG2834347H
NG2834350H
NG2834351H
NG2834356H
NG2834360H
NG2834364H
NG2834369H
NG2834371H
NG2834376H
NG2834379H
NG2834404H
NG2834407H
NG2834411H
NG2834420H
NG2834443H
NG2834444H
NG2834445H
NG2834446H
NG2834448H
NG2834449H
NG2834450H
NG2834451H
NG2834454H
NG2834456H
NG2834459H
NG2834461H
NG2834470H
NG2834475H
NG2834478H
NG2834480H
NG2834483H
NG2834495H
NG2834497H
NG2834499H
NG2834503H
NG2834505H
NG2834507H
NG2834509H
NG2834511H
NG2834515H
NG2834522H
NG2834529H
NG2834531H
NG2834536H
NG2834539H
NG2834540H
NG2834541H
NG2834547H
NG2834548H
NG2834550H
NG2834551H
NG2834552H
NG2834562H
NG2834564H
NG2834566H
NG2834569H
NG2834571H
NG2834581H
NG2834585H
NG2834587H
NG2834591H
NG2834595H
NG2834611H
NG2834615H
NG2834622H
NG2834631H
NG2834634H
NG2834640H
NG2834644H
NG2834647H
NG2834652H
NG2834654H
NK5001796H
NK5001801H
NK5001805H
NK5001812H
NK5001817H
NK5001818H
NK5001820H
NK5001821H
NK5001823H
NK5001846H
NK5001849H
NK5001852H
NK5001853H
NK5001859H
NK5001883H
NK5001884H
NK5001885H
NK5001887H
NK5001889H
NK5001890H
NK5001891H
NK5001894H
NK5001902H
NK5001904H
NK5001906H
NK5001913H
NK5001986H
NK5001991H
NK5002091H
NK5002092H
NK5002104H
NK5002112H
NK5002140H
NK5002146H
NK5002154H
NK5002158H
NK5002164H
NK5002168H
NK5002170H
NK5002173H
NK5002174H
NK5002243H
NK5002251H
NK5002260H
NK5002280H
NK5002281H
NK5002312H
NK5002321H
NK5002348H
NK5002412H
NK5005055H
NK5005076H
NK5005085H
NK5005088H
NK5005089H
NK5005090H
NK5005091H
NK5005092H
NK5005116H
NK5005117H
NK5005121H
NK5005356H
NK5005365H
NK5005367H
NK5005368H
NK5005375H
NK5005385H
NK5005386H
NK5005387H
NK5005859H
NK5006032H
NK5006044H
NK5006045H
NK5006049H
NK5006082H
NK5006097H
NK5006846H
NK5007563H
NK5007564H
NK5007597H
NK5007610H
NK5007617H
NK5007642H
NK5007643H
NK5009160H
NK5009172H
NK5009177H
NK5009344H
NK5009358H
NK5009361H
NK5009365H
NK5009371H
NK5009375H
NK5009389H
NK5009390H
NK5009391H
NK5009392H
NK5009399H
NK5009415H
NK5009426H
NK5009478H
NK5009772H
NK5009774H
NK5009813H
NK5010156H
NK5010172H
NK5010344H
NK5010362H
NK5010363H
NK5010364H
NK5010373H
NK5010396H
NK5010399H
NK5010407H
NK5010414H
NK5010419H
NK5010433H
NK5010434H
NK5010435H
NK5010436H
NK5010467H
NK5010751H
NK5010969H
NK5010975H
NK5010988H
NK5011007H
NK5011008H
NK5011011H
NK5011012H
NK5011025H
NK5011026H
NK5011027H
NK5011351H
NK5011358H
NK5011457H
NK5011459H
NK5011462H
NK5011588H
NK5011897H
NK5013814H
NK5013819H
NK5014606H
NK5014653H
NK5014678H
NK5014688H
NK5014706H
NK5014720H
NK5014721H
NK5015607H
NK5015610H
NK5015624H
NK5015629H
NK5015652H
NK5015653H
NK5015670H
NK5015672H
NK5015726H
NK5015728H
NK5016288H
NK5016296H
NK5016297H
NK5016309H
NK5016337H
NK5016343H
NK5016391H
NK5018596H
NK5018604H
NK5018626H
NK5018763H
NK5018765H
NK5019916H
NK5020781H
NK5020802H
NK5020803H
NK5020862H
NK5020863H
NK5021061H
NK5021064H
NK5021181H
NK5021201H
NK5021209H
NK5021442H
NK5021446H
NK5021449H
NK5021456H
NK5021458H
NK5021482H
NK5021504H
NK5021524H
NK5021540H
NK5021551H
NK5021553H
NK5021554H
NK5021555H
NK5021574H
NK5021579H
NK5021919H
NK5021931H
|
| FEI Number |
3003166194
|
Recalling Firm/ Manufacturer |
Medtronic MiniMed
18000 Devonshire St
Northridge
CA
91325-1219
|
| For Additional Information Contact | Medtronic Diabetes Technical Support 800-646-4633 Ext. 1 |
Manufacturer Reason for Recall | Due to battery cap deterioration, this may potentially result in an incomplete battery circuit (similar to when battery cap is not attached) and a loss of power/functionality of the insulin infusion pumps. |
FDA Determined Cause 2 | Device Design |
| Action | On 05/24/2022, Medtronic began communicating an "URGENT MEDICAL DEVICE CORRECTION" Letter to customers informing them of the battery cap issue where if the metal contact becomes loose or falls off from the battery cap, it can result in an incomplete battery connection, leading to no power source to the insulin pump.
Actions Required by customers:
Before beginning: Do not remove the battery cap unless you have a new battery available. If you have a spare undamaged battery cap, ensure it is available nearby.
During routine battery replacement, check the metal contact on your pump battery cap to see if it is loose, damaged, or missing. You can view animated instructions on how to inspect the battery cap contact through our website at www.medtronicdiabetes.com/battery-cap-guide.
" If the battery cap contact is damaged, immediately replace it with a spare cap that you received with original pump shipment, and discard the damaged cap. If no spare cap, stop using pump and revert to a back-up plan per healthcare provider s recommendations. Then, request a spare battery cap online at www.medtronicdiabetes.com/battery-cap-form, or contact Medtronic 24-Hour
Technical Support at 877-496-7933.
" If unsure that the battery cap contact are damaged, replace it with a spare cap or request a spare battery cap online at www.medtronicdiabetes.com/battery-cap-form, or contact Medtronic 24-Hour Technical Support at 877-496-7933.
" If the battery cap contact is not damaged, continue to use your pump and monitor for cap damage during battery replacement, and complete and return the attached confirmation form by mail. Medtronic will send you a spare cap in the coming months.
" Always pay close attention to the pump and pump battery status after inserting the new
battery.
For questions, contact Medtronic 24-Hour Technical Support at 877-496-7933.
|
| Quantity in Commerce | 2,287 pumps |
| Distribution | U.S.: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VI, VT, WA, WI, WV, and WY
O.U.S.: Not provided |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
|
|
|