 Class 2 Device Recall Roversee related information
Date Initiated by FirmAugust 31, 2022
Date PostedSeptember 23, 2022
Recall Status1 Open3, Classified
Recall NumberZ-1786-2022
Recall Event ID 90839
510(K)NumberK211423 
Product Classification System, x-ray, mobile - Product Code IZL
ProductMICRO-X ROVER MOBILE X-RAY SYSTEM, MXU-RV19
Code Information Model Number: MXU-RV19 UDI Codes: (01)09357123000013(11)210326(21)00222 (01)09357123000013(11)210628(21)00234 (01)09357123000013(11)210629(21)00235 (01)09357123000013(11)210629(21)00236 (01)09357123000013(11)210630(21)00237 (01)09357123000013(11)210708(21)00238 (01)09357123000013(11)210721(21)00240 (01)09357123000013(11)210910(21)00261 (01)09357123000013(11)210913(21)00262 (01)09357123000013(11)210914(21)00263 (01)09357123000013(11)210921(21)00264 (01)09357123000013(11)210922(21)00265 (01)09357123000013(11)210923(21)00266 (01)09357123000013(11)210927(21)00267 (01)09357123000013(11)211006(21)00268 (01)09357123000013(11)211008(21)00269 (01)09357123000013(11)211011(21)00270 (01)09357123000013(11)211011(21)00271 (01)09357123000013(11)211012(21)00272 (01)09357123000013(11)211013(21)00273 (01)09357123000013(11)211013(21)00274 (01)09357123000013(11)211210(21)00288 (01)09357123000013(11)211213(21)00289 (01)09357123000013(11)211213(21)00290 (01)09357123000013(11)220329(21)00305 (01)09357123000013(11)220330(21)00306 (01)09357123000013(11)220331(21)00307 (01)09357123000013(11)220401(21)00308 (01)09357123000013(11)220404(21)00309 (01)09357123000013(11)220405(21)00310 (01)09357123000013(11)220406(21)00311 (01)09357123000013(11)220404(21)00312 Serial Numbers: 222 234 235 236 237 238 240 261 262 263 264 265 266 267 268 269 270 271 272 273 274 288 289 290 305 306 307 308 309 310 311 312
FEI Number 3012855798
Recalling Firm/
Manufacturer
Micro-X Ltd.
A14 6 MAB Eastern Promenade
Tonsley Australia
Manufacturer Reason
for Recall
Software calibration error with product equip with a Dose Area Product (DAP) meter. This results in the product displaying the incorrect DAP meter values which may impede radiation exposure management decisions
FDA Determined
Cause 2
Process control
ActionOn or about 09/31/2022, the firm sent an "URGENT: MEDICAL DEVICE CORRECTION" Letter via email/telephone to customers informing them that during an internal data analysis on 08/18/2022, Micro-X Ltd identified software calibration errors with the product equipped with a Dose Area Product (DAP) meter. This resulted in the product displaying incorrect DAP meter values. The DICOM image produced by the system will report a DAP measurement in microgray/square meter that is 8 to 10 times lower than the actual dose delivered. Customers are instructed to: 1. Inspect their stock immediately and place the Recall Notification Letter on the affected product and quarantine affected product on hand to prevent further use. 2. Contact the nominated Service representative to arrange the re-calibration of the DAP Meter. Email: support-us@micro-x.com or Phone 1-206-249-8764. 3.Previous version of Service Documentation (ROV208-2 Acceptance and Compliance Testing Instructions) which includes inaccurate DAP calibration processes are to be removed from use. 4. Complete and return Acknowledgement and Receipt Form immediately to Micro-X LTD in Australia: support@micro-x.com and copy djerkovic@micro-x.com 5. Ensure relevant staff members are informed of the Recall and if affected units have been supplied/transferred to another facility, inform them immediately. For question or further assistance, contact Regulatory Affairs Manager at djerkovic@micro-x.com
Quantity in Commerce42 units
DistributionWorldwide distribution - U.S. Nationwide distribution in the states of CA, NV, NY, OH, TX and WA and the countries of Australia, France, Israel, Japan, and United Arab Emirates.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database510(K)s with Product Code = IZL
-
-