• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Plate 17 Microcatheter

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Plate 17 Microcathetersee related information
Date Initiated by FirmFebruary 01, 2023
Date PostedMarch 13, 2023
Recall Status1 Open3, Classified
Recall NumberZ-1249-2023
Recall Event ID 91676
510(K)NumberK210601 
Product Classification Catheter, percutaneous - Product Code DQY
ProductPlato 17 Microcatheter: Straight Tip, PL17-160-000 ; 45 degree Tip, PL17-160-045; 90 degree Tip, PL17-160-090
Code Information All lots, REF/UDI-DI/Lot(Expiration): PL17-160-000/00818075010455/ 900164(8/31/2022), 900165(8/31/2022), 900172(8/31/2022), 900177(9/30/2022), 900180(9/30/2022), 900186(10/31/2022), 900187(10/31/2022), 900188(10/31/2022), 900193(10/31/2022), 900196(12/31/2022), 900199(12/31/2022), 900202(12/31/2022), 900221(2/28/2023), 900222(2/28/2023), 900225(2/28/2023), 900226(3/31/2023), 20441(11/30/2023), 20476(11/30/2023), 20650(12/31/2023); PL17-160-045/00818075010462/900167(8/31/2022), 900168(8/31/2022), 900169(8/31/2022), 900174(8/31/2022), 900175(8/31/2022), 900179(9/30/2022), 900181(9/30/2022), 900183(10/31/2022), 900185(10/31/2022), 900205(1/31/2023), 900206(1/31/2023), 900210(1/31/2023), 900211(1/31/2023), 900215(1/31/2023), 900220(1/31/2023), 900219(1/31/2023), 900217(2/28/2023), 900223(2/28/2023), 900224(2/28/2023); PL17-160-090/00818075010479/900170(8/31/2022), 900171(8/31/2022), 900178(9/30/2022), 900182(9/30/2022), 900184(10/31/2022), 900192(11/30/2022), 900195(11/30/2022), 900209(1/31/2023), 900228(3/31/2023)
FEI Number 3009546466
Recalling Firm/
Manufacturer
Scientia Vascular, Inc.
2460 S 3270 W
West Valley City UT 84119-1116
For Additional Information Contact
888-385-9016
Manufacturer Reason
for Recall
Microcatheters for embolic coil and diagnostic agent introduction to peripheral/neuro vasculature, due to manufacturing non-conformance, and due to flush, guidewire insertion and manipulation prior to placement, may have loose particles within catheter lumens that may be introduced to patients, which could cause embolic event, small vessel occlusion, and subacute to chronic inflammatory response.
FDA Determined
Cause 2
Device Design
ActionOn 02/01/23, recall notices were mailed and emailed to customers who were told the following: Actions for Customer/User: Identify and segregate affected devices. Actions for Recalling Firm: Sales representatives will visit and assist with completing the acknowledgement and receipt form and the collection of affected devices. If devices have been transferred from one facility to another the additional consignees will be contacted upon identification. Sales reps received the recall notice and a response form they would take to customers to obtain additional information and the affected devices. Customers with questions can contact customer support at 1-888-385-9016 9016 between the hours of 9:30 AM and 6:30 PM ET (7:30 AM and 4:30 PM MST), Monday through Friday, or via email to regulatory@scientiavascular.com
Quantity in Commerce353
DistributionUS Nationwide distribution in the states of IL, AL, FL, NC, WI, GA, SC, IN, KS, CA, OH, NY, NH, CO, MA, NV, UT, TX.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database510(K)s with Product Code = DQY
-
-