| | Class 2 Device Recall Plate 17 Microcatheter |  |
| Date Initiated by Firm | February 01, 2023 |
| Date Posted | March 13, 2023 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-1249-2023 |
| Recall Event ID |
91676 |
| 510(K)Number | K210601 |
| Product Classification |
Catheter, percutaneous - Product Code DQY
|
| Product | Plato 17 Microcatheter: Straight Tip, PL17-160-000 ; 45 degree Tip, PL17-160-045; 90 degree Tip, PL17-160-090 |
| Code Information |
All lots, REF/UDI-DI/Lot(Expiration):
PL17-160-000/00818075010455/ 900164(8/31/2022), 900165(8/31/2022), 900172(8/31/2022), 900177(9/30/2022), 900180(9/30/2022), 900186(10/31/2022), 900187(10/31/2022), 900188(10/31/2022), 900193(10/31/2022), 900196(12/31/2022), 900199(12/31/2022), 900202(12/31/2022), 900221(2/28/2023), 900222(2/28/2023), 900225(2/28/2023), 900226(3/31/2023), 20441(11/30/2023), 20476(11/30/2023), 20650(12/31/2023);
PL17-160-045/00818075010462/900167(8/31/2022), 900168(8/31/2022), 900169(8/31/2022), 900174(8/31/2022), 900175(8/31/2022), 900179(9/30/2022), 900181(9/30/2022), 900183(10/31/2022), 900185(10/31/2022), 900205(1/31/2023), 900206(1/31/2023), 900210(1/31/2023), 900211(1/31/2023), 900215(1/31/2023), 900220(1/31/2023), 900219(1/31/2023), 900217(2/28/2023), 900223(2/28/2023), 900224(2/28/2023);
PL17-160-090/00818075010479/900170(8/31/2022), 900171(8/31/2022), 900178(9/30/2022), 900182(9/30/2022), 900184(10/31/2022), 900192(11/30/2022), 900195(11/30/2022), 900209(1/31/2023), 900228(3/31/2023)
|
| FEI Number |
3009546466
|
Recalling Firm/ Manufacturer |
Scientia Vascular, Inc.
2460 S 3270 W
West Valley City
UT
84119-1116
|
| For Additional Information Contact | 888-385-9016 |
Manufacturer Reason for Recall | Microcatheters for embolic coil and diagnostic agent introduction to peripheral/neuro vasculature, due to manufacturing non-conformance, and due to flush, guidewire insertion and manipulation prior to placement, may have loose particles within catheter lumens that may be introduced to patients, which could cause embolic event, small vessel occlusion, and subacute to chronic inflammatory response. |
FDA Determined Cause 2 | Device Design |
| Action | On 02/01/23, recall notices were mailed and emailed to customers who were told the following:
Actions for Customer/User: Identify and segregate affected devices.
Actions for Recalling Firm: Sales representatives will visit and assist with completing the acknowledgement and receipt form and the collection of affected devices. If devices have been transferred from one facility to another the additional consignees will be contacted upon identification.
Sales reps received the recall notice and a response form they would take to customers to obtain additional information and the affected devices.
Customers with questions can contact customer support at 1-888-385-9016 9016 between the hours of 9:30 AM and 6:30 PM ET (7:30 AM and 4:30 PM MST), Monday through Friday, or via email to regulatory@scientiavascular.com |
| Quantity in Commerce | 353 |
| Distribution | US Nationwide distribution in the states of IL, AL, FL, NC, WI, GA, SC, IN, KS, CA, OH, NY, NH, CO, MA, NV, UT, TX. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|