| | Class 2 Device Recall VITROS Immunodiagnostic Products |  |
| Date Initiated by Firm | March 01, 2024 |
| Date Posted | June 13, 2024 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-2142-2024 |
| Recall Event ID |
94679 |
| Product Classification |
Test, hepatitis b (b core, be antigen, be antibody, b core igm) - Product Code LOM
|
| Product | VITROS Immunodiagnostic Products Anti HBs Calibrators. The VITROS Anti-HBs Calibrator Pack contains 3 X labelled Calibrator tube (1 per level: C1 - green, C2 - yellow, C3 - red). |
| Code Information |
Model No. 6801926; GTIN# 10758750001699; Lot 9010 (Manufactured 13-Oct-2023; Expiry 15-Jul-2024) and 9030 (Manufactured 09-Dec-2023; Expiry 26-Aug-2024). |
| FEI Number |
1000136573
|
Recalling Firm/ Manufacturer |
Ortho-Clinical Diagnostics, Inc.
100 Indigo Creek Dr
Rochester
NY
14626-5101
|
| For Additional Information Contact | Joe Falvo 585-4533452 |
Manufacturer Reason for Recall | Potential for positively biased results due to signal reduction over shelf life. |
FDA Determined Cause 2 | Under Investigation by firm |
| Action | Consignees were mailed an URGENT PRODUCT CORRECTION NOTIFICATION, dated 5/14/24. In this notice consignees are instructed to discontinue using, render unusable, and discard affected product. The provided notification is to be retained and posted by each VITROS System until the issue has been resolved; copies of the recall notification are to be forwarded if affected product was distributed outside your facility. Consignees are to complete and return the provided Confirmation of Receipt form no later than 5/22/24. Consignees with any questions are to contact Quidel Ortho's Global Services Organization at 1-800-421-3311. |
| Quantity in Commerce | 481 units |
| Distribution | Worldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, FL, GA, IA, IL, IN, KS, LA, MA, MD, ME, MI, MT, NC, NJ, NY, OH, OK, OR, PA, SC, SD, TN, TX, VA, VT, WI, WV and the countries of Bermuda, Colombia, Singapore. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
|
|
|