| | Class 2 Device Recall SureStep and Bardex |  |
| Date Initiated by Firm | May 09, 2024 |
| Date Posted | June 18, 2024 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-2154-2024 |
| Recall Event ID |
94722 |
| 510(K)Number | K040658 |
| Product Classification |
Catheter, urological (antimicrobial) and accessories - Product Code MJC
|
| Product | Bard Medical SureStep Foley Tray System, 16 Fr., Catalog #A319416AM; and Bard Medical Bardex IC Complete Care Infection Control Temperature-Sensing Foley Catheter Tray with Bard Hydrogel and Bacti-Guard Silver Alloy Coating, 350 ml Urine Meter, 16 Fr, Catalog #319416AM. Intended for use in the drainage and/or collection and/or measurement of urine. |
| Code Information |
SureStep Foley Tray System - Lot #NGHZ1757, exp. 31Jan2026, UDI-DI 00801741073861 (eaches); 10801741073868 (case).
Bardex IC Complete Care Infection Control Temperature-Sensing Foley Catheter Tray, Lot #NGHX2932, exp. 31May2025, UDI-DI 00801741024450 (each); 10801741024457 (case). |
| FEI Number |
1018233
|
Recalling Firm/ Manufacturer |
C.R. Bard Inc
8195 Industrial Blvd Ne
Covington
GA
30014-1497
|
Manufacturer Reason for Recall | Some of the devices may function incorrectly resulting in no temperature display or erratic temperatures being displayed on the temperature monitoring devices. |
FDA Determined Cause 2 | Component design/selection |
| Action | The recalling firm issued letters on and dated 5/9/2024 via FedEx and email.
The letter to medical facilities explained the issue, contained a copy of the label for each product and showed where to find the catalog number and lot number, provided the clinical impact, and advised clinical users to review and follow instructions provided in the field action communication to avoid potential health consequences in the future. The letter also said that if the device is currently in use and functioning appropriately, the device does not need to be replaced. Actions to be taken by the medical facilities were: (1) Immediately discontinue use of unused inventory; (2) Check all inventory locations for the affected products and immediately quarantine; (3) Record the inventory totals on the attached Customer Response Form then discard all affected product within the facility's control per the facility's procedures; (4) Share the notice with any users of the product within the facility or with any interfacility users where product was transferred to ensure they are also aware of the recall; (5) Complete and return the attached Customer Response Form to the BD contact noted on the form to confirm acknowledgment of the correction whether or not the consignee has any affected product.
The letter to the distributors was similar to the medical facility letter except they were instructed to identify all customers within their distribution network who purchased the affected product and provide a copy of the attached customer letter to all customers to advise them of the field action notification. A Distributor Response Form was also attached to acknowledge receipt of the recall notification, indicate the quantity of product identified at the distributor, and confirm the inventory was destroyed. |
| Quantity in Commerce | 6,420 devices |
| Distribution | Distribution was nationwide. There was no government/military/foreign distribution. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = MJC
|
|
|
|