| Date Initiated by Firm | June 12, 2024 |
| Date Posted | July 12, 2024 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-2350-2024 |
| Recall Event ID |
94827 |
| 510(K)Number | K013153 |
| Product Classification |
Unit, electrosurgical, endoscopic (with or without accessories) - Product Code KNS
|
| Product | Autotome RX 20mm CUT WIRE Cannulating Sphincterotome, Material Number (UPN) M00545170. The device is used in selective cannulation of the bile duct. |
| Code Information |
UDI-DI (GTIN) 08714729444749
Lot 33280747
Expiration Date: January 23, 2027 |
| FEI Number |
3005099803
|
Recalling Firm/ Manufacturer |
Boston Scientific Corporation
100 Boston Scientific Way
Marlborough
MA
01752-1234
|
| For Additional Information Contact | Rebecca KinKead Rubio 508-683-4000 |
Manufacturer Reason for Recall | Foreign material may be present on the enclosed section of the cutting wire which may lead to inability to optimally bow the sphincterotome. This issue may lead to prolongation of procedure. |
FDA Determined Cause 2 | Process control |
| Action | All customers notified via courier on June 12, 2024 - notice was titled "Urgent Medical Device Removal."
Customers were informed to cease further use or distribution of any remaining affected product. Product should be segregated and returned to the firm.
For more information, please reach out to your local Boston Scientific representative. |
| Quantity in Commerce | 256 US, 39 OUS |
| Distribution | Domestic distribution nationwide. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = KNS
|