| | Class 2 Device Recall HeartMate 3 System Controllers |  |
| Date Initiated by Firm | June 25, 2024 |
| Date Posted | August 13, 2024 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-2583-2024 |
| Recall Event ID |
94973 |
| PMA Number | P160054 |
| Product Classification |
Ventricular (assist) bypass - Product Code DSQ
|
| Product | HeartMate 3 System Controllers provided within the following HeartMate 3 Left Ventricular Assist System (LVAS) Implant Kits or distributed separately:
HeartMate 3 LVAS Implant Kit, US, Model: 106524US;
HeartMate 3 LVAS Implant Kit, OUS, Model: 106524INT;
HeartMate 3 System Controller, US, Model: 106531US;
HeartMate 3 System Controller, OUS, Model: 106531INT;
HeartMate 3 System Controller Low Flow 2.0, Global, Model: 106531LF2 |
| Code Information |
Model/UDI-DI/System Controller Serial Numbers:
106524US/00813024013297/HSC-134168, HSC-134392, HSC-134393, HSC-134394, HSC-134395, HSC-134396, HSC-134397, HSC-134398, HSC-134399, HSC-134400, HSC-134401, HSC-134402, HSC-134403, HSC-134404, HSC-134405, HSC-134406, HSC-134407, HSC-134408, HSC-134409, HSC-134410, HSC-134411, HSC-134432, HSC-134433, HSC-134434, HSC-134436, HSC-134437, HSC-134438, HSC-134440, HSC-134452, HSC-134454, HSC-134502, HSC-134503, HSC-134504, HSC-134505, HSC-134506, HSC-134507, HSC-134508, HSC-134509, HSC-134510, HSC-134511, HSC-134512, HSC-134513, HSC-134514, HSC-134515, HSC-134516, HSC-134517, HSC-134518, HSC-134519, HSC-134520, HSC-134521, HSC-134572, HSC-134573, HSC-134574, HSC-134575, HSC-134576, HSC-134577, HSC-134578, HSC-134579, HSC-134580, HSC-134581, HSC-134582, HSC-134583, HSC-134584, HSC-134585, HSC-134586, HSC-134587, HSC-134588, HSC-134589, HSC-134590, HSC-134591, HSC-134602, HSC-134603, HSC-134604, HSC-134605, HSC-134606, HSC-134607, HSC-134608, HSC-134609, HSC-134610, HSC-134611, HSC-134612, HSC-134613, HSC-134614, HSC-134615, HSC-134616, HSC-134617, HSC-134618, HSC-134619, HSC-134620, HSC-134621, HSC-134662, HSC-134663, HSC-134664, HSC-134665, HSC-134666, HSC-134667, HSC-134668, HSC-134669, HSC-134670, HSC-134671, HSC-134672, HSC-134673, HSC-134674, HSC-134675, HSC-134676, HSC-134677, HSC-134678, HSC-134679, HSC-134680, HSC-134748, HSC-134749, HSC-134750, HSC-134806, HSC-134807, HSC-134808, HSC-134809, HSC-134810, HSC-134811, HSC-134812, HSC-134813, HSC-134814, HSC-134815, HSC-134816, HSC-134818, HSC-134819, HSC-134820, HSC-134821, HSC-134822, HSC-134823, HSC-134824, HSC-134825, HSC-134826, HSC-134827, HSC-134828, HSC-134829, HSC-134830, HSC-134831, HSC-134832, HSC-134833, HSC-134834, HSC-134835, HSC-134836, HSC-134837, HSC-134838, HSC-134839, HSC-134840, HSC-134841, HSC-134842, HSC-134843, HSC-134844, HSC-134845, HSC-134856, HSC-134857, HSC-134858, HSC-134859, HSC-134860, HSC-134861, HSC-134862, HSC-134863, HSC-134864, HSC-134865, HSC-134866, HSC-134867, HSC-134868, HSC-134869, HSC-134870, HSC-134871, HSC-134872, HSC-134873, HSC-134875, HSC-134876, HSC-134877, HSC-134878, HSC-134879, HSC-134880, HSC-134881, HSC-134882, HSC-134883, HSC-134885, HSC-134976, HSC-134977, HSC-134978, HSC-134979, HSC-134980, HSC-134981, HSC-134982, HSC-134983, HSC-134984.
106524INT/00813024011712/HSC-134262, HSC-134263, HSC-134264, HSC-134265, HSC-134266, HSC-134267, HSC-134268, HSC-134269, HSC-134270, HSC-134271, HSC-134272, HSC-134274, HSC-134275, HSC-134276, HSC-134277, HSC-134278, HSC-134279, HSC-134280, HSC-134281, HSC-134282, HSC-134283, HSC-134284, HSC-134285, HSC-134286, HSC-134287, HSC-134288, HSC-134289, HSC-134290, HSC-134291, HSC-134292, HSC-134293, HSC-134294, HSC-134295, HSC-134296, HSC-134297, HSC-134298, HSC-134299, HSC-134300, HSC-134301, HSC-134352, HSC-134353, HSC-134356, HSC-134357, HSC-134358, HSC-134359, HSC-134360, HSC-134361, HSC-134362, HSC-134364, HSC-134365, HSC-134368, HSC-134369, HSC-134371, HSC-134532, HSC-134533, HSC-134534, HSC-134535, HSC-134536, HSC-134537, HSC-134538, HSC-134539, HSC-134540, HSC-134541, HSC-134542, HSC-134543, HSC-134544, HSC-134545, HSC-134546, HSC-134547, HSC-134548, HSC-134549, HSC-134550, HSC-134551, HSC-134552, HSC-134553, HSC-134554, HSC-134555, HSC-134556, HSC-134557, HSC-134558, HSC-134559, HSC-134560, HSC-134561, HSC-134562, HSC-134563, HSC-134564, HSC-134565, HSC-134566, HSC-134567, HSC-134568, HSC-134569, HSC-134570, HSC-134571, HSC-134706, HSC-134707, HSC-134708, HSC-134709, HSC-134710, HSC-134711, HSC-134712, HSC-134713, HSC-134714, HSC-134715, HSC-134716, HSC-134717, HSC-134718, HSC-134719, HSC-134720, HSC-134721, HSC-134722, HSC-134723, HSC-134724, HSC-134725, HSC-134746, HSC-134747, HSC-134778, HSC-134779, HSC-134780, HSC-134781, HSC-134782, HSC-134783, HSC-134784, HSC-134785, HSC-134846, HSC-134848, HSC-134849, HSC-134850, HSC-134851, HSC-134852, HSC-134853, HSC-134854, HSC-134855, HSC-134957, HSC-134958, HSC-134959, HSC-134960, HSC-134961, HSC-134962, HSC-134964, HSC-134965, HSC-134966, HSC-134968, HSC-134969, HSC-134970, HSC-134971, HSC-134972, HSC-134973, HSC-134974, HSC-134975.
106531US/00813024013235/HSC-134242, HSC-134244, HSC-134245, HSC-134246, HSC-134247, HSC-134252, HSC-134253, HSC-134254, HSC-134255, HSC-134256, HSC-134257, HSC-134258, HSC-134259, HSC-134260, HSC-134261, HSC-134302, HSC-134303, HSC-134304, HSC-134305, HSC-134306, HSC-134307, HSC-134309, HSC-134310, HSC-134311, HSC-134312, HSC-134313, HSC-134314, HSC-134315, HSC-134316, HSC-134317, HSC-134318, HSC-134319, HSC-134320, HSC-134321, HSC-134322, HSC-134323, HSC-134324, HSC-134325, HSC-134326, HSC-134327, HSC-134328, HSC-134329, HSC-134330, HSC-134331, HSC-134332, HSC-134333, HSC-134334, HSC-134335, HSC-134336, HSC-134337, HSC-134338, HSC-134339, HSC-134340, HSC-134341, HSC-134342, HSC-134343, HSC-134344, HSC-134345, HSC-134346, HSC-134347, HSC-134348, HSC-134349, HSC-134350, HSC-134351, HSC-134372, HSC-134373, HSC-134374, HSC-134375, HSC-134376, HSC-134377, HSC-134378, HSC-134379, HSC-134380, HSC-134381, HSC-134382, HSC-134383, HSC-134384, HSC-134385, HSC-134386, HSC-134387, HSC-134388, HSC-134389, HSC-134390, HSC-134391, HSC-134412, HSC-134413, HSC-134414, HSC-134415, HSC-134416, HSC-134417, HSC-134418, HSC-134419, HSC-134420, HSC-134422, HSC-134423, HSC-134424, HSC-134425, HSC-134426, HSC-134427, HSC-134428, HSC-134429, HSC-134430, HSC-134431, HSC-134442, HSC-134443, HSC-134444, HSC-134445, HSC-134446, HSC-134447, HSC-134448, HSC-134449, HSC-134450, HSC-134451, HSC-134462, HSC-134463, HSC-134464, HSC-134465, HSC-134466, HSC-134467, HSC-134468, HSC-134469, HSC-134470, HSC-134471, HSC-134472, HSC-134473, HSC-134474, HSC-134475, HSC-134476, HSC-134477, HSC-134478, HSC-134479, HSC-134480, HSC-134481, HSC-134482, HSC-134483, HSC-134484, HSC-134485, HSC-134486, HSC-134487, HSC-134488, HSC-134489, HSC-134490, HSC-134491, HSC-134492, HSC-134493, HSC-134494, HSC-134495, HSC-134496, HSC-134497, HSC-134498, HSC-134499, HSC-134500, HSC-134501, HSC-134522, HSC-134523, HSC-134524, HSC-134525, HSC-134526, HSC-134527, HSC-134528, HSC-134529, HSC-134530, HSC-134531, HSC-134593, HSC-134622, HSC-134623, HSC-134624, HSC-134625, HSC-134626, HSC-134627, HSC-134628, HSC-134629, HSC-134630, HSC-134631, HSC-134632, HSC-134633, HSC-134634, HSC-134635, HSC-134636, HSC-134637, HSC-134638, HSC-134639, HSC-134640, HSC-134641, HSC-134642, HSC-134643, HSC-134644, HSC-134645, HSC-134646, HSC-134647, HSC-134648, HSC-134649, HSC-134650, HSC-134651, HSC-134652, HSC-134653, HSC-134654, HSC-134655, HSC-134656, HSC-134657, HSC-134658, HSC-134659, HSC-134660, HSC-134661, HSC-134682, HSC-134683, HSC-134684, HSC-134685, HSC-134686, HSC-134687, HSC-134688, HSC-134689, HSC-134690, HSC-134691, HSC-134692, HSC-134693, HSC-134694, HSC-134695, HSC-134696, HSC-134697, HSC-134698, HSC-134699, HSC-134700, HSC-134701, HSC-134730, HSC-134736, HSC-134737, HSC-134738, HSC-134739, HSC-134740, HSC-134741, HSC-134742, HSC-134743, HSC-134744, HSC-134745, HSC-134755, HSC-134756, HSC-134757, HSC-134758, HSC-134759, HSC-134760, HSC-134761, HSC-134762, HSC-134763, HSC-134764, HSC-134765, HSC-134766, HSC-134767, HSC-134768, HSC-134769, HSC-134770, HSC-134771, HSC-134772, HSC-134773, HSC-134774, HSC-134775, HSC-134776, HSC-134786, HSC-134787, HSC-134788, HSC-134789, HSC-134790, HSC-134791, HSC-134792, HSC-134793, HSC-134794, HSC-134795, HSC-134796, HSC-134797, HSC-134798, HSC-134799, HSC-134800, HSC-134801, HSC-134802, HSC-134803, HSC-134804, HSC-134805, HSC-134886, HSC-134887, HSC-134888, HSC-134889, HSC-134890, HSC-134891, HSC-134892, HSC-134893, HSC-134894, HSC-134895, HSC-134896, HSC-134897, HSC-134898, HSC-134899, HSC-134900, HSC-134901, HSC-134902, HSC-134903, HSC-134904, HSC-134905, HSC-134906, HSC-134907, HSC-134908, HSC-134909, HSC-134910, HSC-134911, HSC-134912, HSC-134913, HSC-134914, HSC-134915, HSC-134916, HSC-134917, HSC-134918, HSC-134919, HSC-134920, HSC-134921, HSC-134922, HSC-134923, HSC-134924, HSC-134925, HSC-134927, HSC-134933, HSC-134934, HSC-134936, HSC-134937, HSC-134938, HSC-134939, HSC-134940, HSC-134942, HSC-134943, HSC-134944, HSC-134945, HSC-134985, HSC-134996, HSC-134997, HSC-134998, HSC-134999.
106531INT/00813024011842/HSC-134946, HSC-134947, HSC-134948, HSC-134949, HSC-134951, HSC-134952, HSC-134954, HSC-134955, HSC-134986, HSC-134987, HSC-134988, HSC-134989, HSC-134990, HSC-134991, HSC-134992, HSC-134993, HSC-134994.
106531LF2/05415067031105/HSC-133326, HSC-134726, HSC-134729, HSC-134733, HSC-134735, HSC-134752
|
| FEI Number |
2916596
|
Recalling Firm/ Manufacturer |
Thoratec LLC
6035 Stoneridge Dr
Pleasanton
CA
94588-3270
|
| For Additional Information Contact | Shelley Lange 612-346-3514 |
Manufacturer Reason for Recall | Left ventricular assist system controller UI membrane/screen may lift along the edge of the controller housing, and if controller is exposed to fluid, ingress may occur, which may damage printed circuit board assemblies, which could result in unexpected and false positive alarms, loss of visual alarms, loss of user interface(use of buttons, loss of visual alarms,etc.), loss of power, and pump stop |
FDA Determined Cause 2 | Process design |
| Action | On 6/25/24, verbal and email communications were initiated informing customers of the following:
1) Carefully inspect the controller at the edge near the display button to confirm the UI membrane is not lifted (refer to provided visual examples).
2) Report Controllers with lifted UI membrane by contacting Firm's Technical Support at 1-800-456-1477.
3) Reinforce requirement as listed on Page 2-11 "System Controller Warnings and Cautions" in the IFU to patients to protect controller from water. Advise them to follow instructions to avoid direct exposure to water; and communicate risk of continuous, steady exposure to water.
4) Reinforce requirement to patient to perform the daily self-test to verify the display works appropriately as listed on Page 2-29 "Performing a System Controller Self Test". Patients should contact the hospital in the event of display issues.
Firm will follow-up with a formal notification about the lifted UI membrane.
|
| Quantity in Commerce | 694 |
| Distribution | US: MN, PA, FL, IL, NY, DE, AR, NJ, NC, GA, WI, AZ, TN, MO, TX, MA, OH, VA, NE, CA, OK, CO, DC, IN, MI, CT, UT, IA, MD, OR, NM, ME, SC, KY, LA, WA, NV, HI, AL, MS, KS, WV.
OUS: Algeria, Australia, Austria, Bahrain, Bangladesh, Belgium, Bosnia & Herzegovina, Brazil, Bulgaria, Canada, Cayman Islands, Chile, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Germany, Greece, Hong Kong, India, Iran, Islamic Republic of, Iraq, Ireland, Israel, Italy, Korea, Kuwait, Lebanon, Lithuania, Macedonia, Morocco, Netherlands, New Zealand, Norway, Oman, Palestine, Poland, Portugal, Qatar, Romania, Russian Federation, Saudi Arabia, Serbia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Tunisia, Turkey, United Arab Emirates, United Kingdom, Uzbekistan
|
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| PMA Database | PMAs with Product Code = DSQ
|
|
|
|