 Class 2 Device Recall Percept " RC Rechargeable Implantable Neurostimulator with BrainSense " Technologysee related information
Date Initiated by FirmSeptember 24, 2024
Date PostedOctober 23, 2024
Recall Status1 Open3, Classified
Recall NumberZ-0206-2025
Recall Event ID 95489
PMA NumberP960009 
Product Classification Stimulator, electrical, implanted, for parkinsonian tremor - Product Code MHY
ProductMedtronic Percept RC Implantable Neurostimulators (INS), Model B35300.
Code Information GTIN 00763000813468, Serial Numbers: NND504236H, NND504252H; GTIN 00763000813475, Serial Numbers: NND504165H, NND504169H, NND504174H, NND504181H, NND504208H, NND504212H, NND504215H, NND504216H, NND504217H, NND504218H, NND504221H, NND504222H, NND504223H, NND504224H, NND504229H, NND504230H, NND504232H, NND504234H, NND504235H, NND504243H, NND504255H, NND504156H, NND504157H, NND504159H, NND504160H; GTIN 00763000813529, Serial Numbers: NND504167H, NND504178H, NND504179H, NND504180H, NND504182H, NND504227H, NND504228H; GTIN A7630008135201, Serial Numbers: NND504167H, NND504178H, NND504180H; GTIN A7630008134701, Serial Numbers: NND504181H, NND504243H
FEI Number 2182207
Recalling Firm/
Manufacturer
Medtronic Neuromodulation
7000 Central Ave Ne
Minneapolis MN 55432-3568
For Additional Information ContactNaomi Rodiles
763-427-5521
Manufacturer Reason
for Recall
A limited number of Percept RC Implantable Neurostimulators (INS), Model B35300 will not be able to communicate with the A610 DBS Clinician Programmer Application upon use and an "Invalid Device" message will appear.
FDA Determined
Cause 2
Under Investigation by firm
ActionMedtronic neuromodulation issued an URGENT MEDICAL DEVICE RECALL notice to its consignees of 09/24/2024 via letter. The notice explained the issue, potential risk, and requested the affected devices in stock be returned. The notice is to be shared with those to whom the device was distributed. If the device has been implanted, the consignee was directed to contact their Medtronic Representative to determine if there are additional steps that may establish communication with the INS. On 07-Oct-24 and 08-Oct-24, a Medtronic representative successfully performed an update using a manufacturing software tool to reestablish communication to the affected devices. Consignees were provided a consignee notification regarding this action.
Quantity in Commerce39 units
DistributionInternational distribution to the countries of Germany, UK, and Israel.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.