• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall CalciCare Calcium Alginate Dressing

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall CalciCare Calcium Alginate Dressingsee related information
Date Initiated by FirmOctober 01, 2024
Date PostedDecember 04, 2024
Recall Status1 Open3, Classified
Recall NumberZ-0617-2025
Recall Event ID 95613
Product Classification Dressing, wound, hydrophilic - Product Code NAC
ProductHollister CalciCare Calcium Alginate Dressing REF 529937R calcium alginate dressings are indicated for moderate to heavily exuding chronic and acute wounds such as: pressure ulcers leg ulcers cavity wounds lacerations post-operative surgical wounds
Code Information Model Number: 529937R UDI-DI code: 0060075029530 Lot Number: W00070134
FEI Number 3003080333
Recalling Firm/
Manufacturer
Advanced Medical Solutions Ltd.
Unit A
33 Road One
Winsford United Kingdom
For Additional Information ContactJames Bartlett
44 01 606545510
Manufacturer Reason
for Recall
Due to defects on primary packaging pouch that compromise the device's ability to maintain a sterile barrier.
FDA Determined
Cause 2
Packaging process control
ActionOn 10/22/2024, the firm emailed a "Field Corrective Action - Field Notice" to customers. On 11/19/2024, an "URGENT: MEDICAL DEVICE RECALL" Letter was distributed via email to affected customers informing them that the firm has become aware of defects on primary packaging pouches, in which minor missing patches of Polyethylene have been detected. These defects could compromise the device's ability to maintain a sterile barrier. Customers are instructed: Distributor/Logistic Centers - 1. Stop/cease distribution of affected products. Immediately inspect internal inventory for the aforementioned packaging defect and quarantine all affected Product pending safe destruction. 2. As soon as possible, and no later than 14 days after receipt of this Recall Notice, complete the attached APPENDIX 1 - DISTRIBUTOR / LOGISTIC CENTRES FORM and return it to AMS by post or by email. 3. As soon as possible, and no later than 14 days after receipt of this Recall Notice, complete the attached APPENDIX 3 - CERTIFICATE OF DESTRUCTION FORM and return it to AMS by post or by email. 4. Immediately distribute this Recall Notice to all affected end user customers/Healthcare facilities alongside the attached APPENDIX 2 - CUSTOMER REPLY FORM and APPENDIX 3 - CERTIFICATE OF DESTRUCTION . Advise them to execute the actions and collect the forms from customers. 5. END USERS Ensure product is inspected at point of care in line with the aforementioned packaging defect and execute actions in accordance with this Notice 6. The Recall Notice does not need to be communicated to patients. There is no action to take with patients. 7. Relevant product that has passed inspection without defect presence can be removed from quarantine and returned to distribution. 8. Defect product will be replaced free of charge upon receipt of certificate of destruction (Appendix 3). 9. Customer Services Contact Number: +44 1606 545617 Email: Customer.Support@admedsol.com
Quantity in Commerce3900 boxes
DistributionU.S. Nationwide distribution in the states of IL, TX, and VA.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
-
-