 Class 2 Device Recall Sizer Setsee related information
Date Initiated by FirmOctober 10, 2022
Date PostedDecember 11, 2024
Recall Status1 Completed
Recall NumberZ-0669-2025
Recall Event ID 95781
Product Classification Guide, surgical, instrument - Product Code FZX
ProductHologic Sizer Set- sterilization tray and 6 individual sizers, intended as a Surgical instrument guide Model/Catalog Number: F0101
Code Information Model No F0101. UDI 15420045515390. Lot Numbers: 186728, SS1-200727, SS1-181121, SS1-190301, SS1-190501, 175517, 175516, 162427, 140718-01, 171056, 175515.
FEI Number 1000120743
Recalling Firm/
Manufacturer
Hologic, Inc
250 Campus Dr
Marlborough MA 01752-3020
For Additional Information ContactBreast Health Customer Support
+1-877-3714372
Manufacturer Reason
for Recall
IFU was not provided with the Sizer Set sterilization tray Revised Sterilization Instructions for Sizer Set
FDA Determined
Cause 2
Labeling Change Control
ActionHologic issued Letter-Revised Sterilization Instructions for Sizer Set on 10/10/22 via Fed'X. The letter states reason for recall, health risk and action to take: Effective immediately, please steam autoclave F0101 in accordance with such IFU, subject to the following parameters. Any additional parameters contained in Hologic s IFU for F0101 should no longer be utilized. link to their Instructions for Use (IFU) below: https://summitmedicalusa.com/wordpress/wp-content/uploads/2021/05/40192-01_RevZ_IS-IFU_Multilingual.pdf Please also note that when sterilizing F0101, one (1) layer of wrapping material (or one layer of 1 ply wrap) is necessary to provide a sterile barrier for each sterilization tray. Please do not utilize 2 layers or one layer of 2-ply wrap. In the future, Hologic will be updating the IFU supplied with the sizers within F0101, to reflect this change. mailto:BreastHealth.Support@hologic.com
Quantity in Commerce178 units
DistributionUS Nationwide distribution.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
-
-