| | Class 2 Device Recall Sizer Set |  |
| Date Initiated by Firm | October 10, 2022 |
| Date Posted | December 11, 2024 |
| Recall Status1 |
Completed |
| Recall Number | Z-0669-2025 |
| Recall Event ID |
95781 |
| Product Classification |
Guide, surgical, instrument - Product Code FZX
|
| Product | Hologic Sizer Set- sterilization tray and 6 individual sizers, intended as a Surgical instrument guide
Model/Catalog Number: F0101 |
| Code Information |
Model No F0101. UDI 15420045515390. Lot Numbers: 186728, SS1-200727, SS1-181121, SS1-190301, SS1-190501, 175517, 175516, 162427, 140718-01, 171056, 175515.
|
| FEI Number |
1000120743
|
Recalling Firm/ Manufacturer |
Hologic, Inc
250 Campus Dr
Marlborough
MA
01752-3020
|
| For Additional Information Contact | Breast Health Customer Support +1-877-3714372 |
Manufacturer Reason for Recall | IFU was not provided with the Sizer Set sterilization tray Revised Sterilization Instructions for Sizer Set |
FDA Determined Cause 2 | Labeling Change Control |
| Action | Hologic issued Letter-Revised Sterilization Instructions for Sizer Set on 10/10/22 via Fed'X.
The letter states reason for recall, health risk and action to take:
Effective immediately, please steam autoclave F0101 in accordance with such IFU, subject to the following parameters. Any additional parameters contained in Hologic s IFU for F0101 should no longer be utilized.
link to their Instructions for Use (IFU) below:
https://summitmedicalusa.com/wordpress/wp-content/uploads/2021/05/40192-01_RevZ_IS-IFU_Multilingual.pdf
Please also note that when sterilizing F0101, one (1) layer of wrapping material (or one layer of 1 ply wrap) is necessary to provide a sterile barrier for each sterilization tray. Please do not utilize 2 layers or one layer of 2-ply wrap. In the future, Hologic will be updating the IFU supplied with the sizers within F0101, to reflect this change.
mailto:BreastHealth.Support@hologic.com |
| Quantity in Commerce | 178 units |
| Distribution | US Nationwide distribution. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
|
|
|
|