 Class 2 Device Recall DRG: HYBRiDXL Insulinsee related information
Date Initiated by FirmOctober 07, 2024
Date PostedJanuary 15, 2025
Recall Status1 Open3, Classified
Recall NumberZ-0921-2025
Recall Event ID 95785
Product Classification Radioimmunoassay, immunoreactive insulin - Product Code CFP
ProductDRG:HYBRiD-XL Insulin. Model/Catalog Number: HYE-5361
Code Information UDI: (01)04048474053616(17)20251130(10)429K084-1. Lot Number: 429K084-1
FEI Number 2245285
Recalling Firm/
Manufacturer
DRG International, Inc.
841 Mountain Ave
Springfield NJ 07081-3437
For Additional Information ContactIna Hairston
973-564-7555
Manufacturer Reason
for Recall
The user of the assay may either face a failure message during re-calibration of the Hybrid.XL instrument using the re-calibration R1 and R2 of the device or find controls out of specification (failed Control High) which causes a delay in results.
FDA Determined
Cause 2
Nonconforming Material/Component
ActionDRG International began notifying consignees on 10/07/2024 via letter. Consignees were instructed to check inventory for affected units, cease use of them, and quarantine all product. Arrange return of those units. Consignees were also instructed to notify customers if units were further distributed.
Quantity in Commerce10 units
DistributionWorldwide - US Nationwide distribution in the state of New Jersey and the countries of Czechia and Romania.
Total Product Life CycleTPLC Device Report