• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall IMed Products

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall IMed Productssee related information
Date Initiated by FirmDecember 13, 2024
Date PostedJanuary 22, 2025
Recall Status1 Open3, Classified
Recall NumberZ-1012-2025
Recall Event ID 96068
Product Classification Dressing change tray - Product Code OXQ
ProductBrand Name: IMed Products Product Name: Dressing Change Kit with ChloraPrep 3mL & HaloGUARD Model/Catalog Number: IM46062 Software Version: N/A Product Description: Dressing change kit, case of 30, lot #600022, 66 cases affected. Component: No
Code Information Model No IM46062, UDI-DI 10853155008769, Lot Number 600022
FEI Number 1000123297
Recalling Firm/
Manufacturer
Trinity Sterile, Inc.
201 Kiley Dr
Salisbury MD 21801-2249
For Additional Information ContactDouglas B. Kraemer
1-410-8605123
Manufacturer Reason
for Recall
Potential for hole in package compromising sterility of the medical kit.
FDA Determined
Cause 2
Process control
ActionOn December 13, 2024 URGENT MEDICAL DEVICE RECALL letters were sent to direct accounts. Action required: Our records indicate you have may have received and/or distributed product that is subject to this recall. We are now notifoing you to take the following actions: 1, Inspect your inventory for the affected product lots. If you have affected product lots in inventory, discontinue use immediately. 2. Notify Additional Users. If you have further distributed any of the items referenced in this notification, you must notify those downstream customers of this recall. Have those users return any affected product to you. Please include these additional users'quantities on your response form. 3. Notify Trinity. Please complete the enclosed Recall Acknowledgement form as soon as possible, indicating whether you do or do not have inventory and fax it to 410-860-2913 or e-mail it to customerseruice@tri nitysterile.com. 4. Return the affected product. Once Trinity receives your completed Recall Acknowledgement form, a customer seryice representative will contact you with a Return Materials Authorization (RMA) number covering your inventory and will provide instruction for the return of products. Trinity is committed to providing high-quality, safe, and effective products. We sincerely apologize for any inconvenience this action may cause your operations. If you have experienced an adverse event associated with this product, please call Customer Service and/or report the event to FDA's MedWatch Adverse Event Reporting program online. If you have any questions, please contact Customer Service at 410-860-5123.
Quantity in Commerce1,980 units
DistributionUS Nationwide distribution in the states of MD, PA, TN and TX.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
-
-