| | Class 2 Device Recall 3M Prevena Plus 125 Therapy Unit and System Kit |  |
| Date Initiated by Firm | January 17, 2025 |
| Date Posted | February 14, 2025 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-1175-2025 |
| Recall Event ID |
96070 |
| 510(K)Number | K223263 |
| Product Classification |
Negative Pressure Wound Therapy device for reduction of wound complications - Product Code QFC
|
| Product | 3M Prevena Plus 125 Therapy Unit and System Kits
REF:
PRE3021US Prevena Plus Duo Peel and Place 20cm System Kit, US
PRE3201US Prevena Plus Peel and Place 35 cm system Kit, US
PRE3201 Prevena Plus Peel and Place 35 cm system Kit
PRE4000US Prevena Plus 125 Therapy Unit, US
PRE4001AU PREVENA PLUS SYSTEM KIT, AU
PRE4001CA PREVENA PLUS SYSTEM KIT, CA
PRE4001UK PREVENA PLUS SYSTEM KIT, UK
PRE4001US PREVENA PLUS SYSTEM KIT, US
PRE4001 PREVENA PLUS SYSTEM KIT
PRE4001ZA PREVENA PLUS SYSTEM KIT, ZA
PRE4010 PREVENA PLUS STANDALONE EXTENDED LIFE SYSTEM KIT
PRE5001 PREVENA RESTOR ARTHRO-FORM SYSTEM KIT 33X30 CM
PRE5101 PREVENA RESTOR ARTHRO-FORM SYSTEM KIT 46X30 CM
PRE5221 PREVENA RESTOR BELLA-FORM SYSTEM KIT 21X19 CM
PRE5321 PREVENA RESTOR BELLA-FORM SYSTEM KIT 24X22 CM
PRE5421 PREVENA RESTOR BELLA-FORM SYSTEM KIT 29X27 CM
PRE5501 PREVENA RESTOR AXIO-FORM SYSTEM KIT 29X28 CM |
| Code Information |
PRE3021US Prevena Plus Duo Peel and Place 20cm System Kit, US
UDI-DI code: 00849554006744
Assembly Item number: PRE3021USV005
Lot Numbers:
C16793V005
C17567V005
C18503V005
PRE3201US Prevena Plus Peel and Place 35 cm system Kit, US
UDI-DI code: 00849554006621
Assembly Item Number: PRE3201USV006
Lot Numbers:
C16794V006
C17165V006
C17569V006
C17954V006
C18502V006
C18610V006
C19238V006
Assembly Item Number: PRE3201USV007
Lot Numbers:
C18068V007
C18504V007
PRE3201 Prevena Plus Peel and Place 35 cm system Kit
UDI-DI code: 00849554006645
Assembly Item Number: PRE3201V008
Lot Numbers:
C17273V008
C17882V008
C18391V008
C19240V008
Assembly Item Number: PRE3201V009
Lot Numbers:
C16796V009
C17755V009
C17955V009
Assembly Item Numbers: PRE3201V010
Lot Numbers:
C16797V010
C17956V010
C18505V010
Assembly Item Number: PRE3201V011
Lot Numbers:
C16798V011
C17957V011
C18506V011
PRE4000US Prevena Plus 125 Therapy Unit, US
UDI-DI code: 00849554006768
Assembly Item Number: PRE4000USV005
Lot Numbers:
C17175V005
C17884V005
Assembly Item Number: PRE4000USV008
Lot Numbers:
C16801V008
C16874V008
C16998V008
C17030V008
C17133V008
C17183V008
C17405V008
C17541V008
C17574V008
C17757V008
C17779V008
C17878V008
C17951V008
C18057V008
C18149V008
C18208V008
C18392V008
C18444V008
C18719V008
C18723V008
C18810V008
C18860V008
C18956V008
C19159V008
C19232V008
PRE4001AU PREVENA PLUS SYSTEM KIT, AU
UDI-DI code: 00849554005877
Assembly Item Number: PRE4001AUV006
Lot Numbers:
C16680V006
C17886V006
C18461V006
C18813V006
PRE4001CA PREVENA PLUS SYSTEM KIT, CA
UDI-DI code: 00849554005891
Assembly Item Number: PRE4001CAV006
Lot Numbers:
C17683V006
C17818V006
C19162V006
PRE4001UK PREVENA PLUS SYSTEM KIT, UK
UDI-DI code: 00849554005907
Assembly Item Number: PRE4001UKV006
Lot Numbers:
C17024V006
C17177V006
C17300V006
C17958V006
PRE4001US PREVENA PLUS SYSTEM KIT, US
UDI-DI code: 00849554005365
Assembly Item Number: PRE4001USV005
Lot Number: C18959V005
Assembly Item Number: PRE4001USV006
Lot Numbers:
C16753V006
C16802V006
C16875V006
C16999V006
C17032V006
C17129V006
C17184V006
C17258V006
C17303V006
C17406V006
C17565V006
C17576V006
C17780V006
C17879V006
C17952V006
C18058V006
C18205V006
C18227V006
C18394V006
C18466V006
C18721V006
C18725V006
C18812V006
C18862V006
C18958V006
C19060V006
C19161V006
C19234V006
PRE4001 PREVENA PLUS SYSTEM KIT
Assembly Item Number: PRE4001V006
Lot Numbers:
C17178V006
C18062V006
C18814V006
Assembly Item Number: PRE4001V008
Lot Number: C18523V008
PRE4001ZA PREVENA PLUS SYSTEM KIT, ZA
UDI-DI code: 00849554005884
Assembly Item Number: PRE4001ZAV006
Lot Numbers:
C18060V006
C18464V006
PRE4010 PREVENA PLUS STANDALONE EXTENDED LIFE SYSTEM KIT
UDI-DI code: 00849554007819
Assembly Item Number: PRE4010V001
Lot Numbers:
C16715V001
C16883V001
C17000V001
C17031V001
C17134V001
C17195V001
C17257V001
C17302V001
C17407V001
C17542V001
C17575V001
C17758V001
C17781V001
C17880V001
C17953V001
C18059V001
C18150V001
C18209V001
C18393V001
C18465V001
C18720V001
C18724V001
C18811V001
C18861V001
C18957V001
C19059V001
C19160V001
Assembly Item Number: PRE4010V003
Lot Number: C16878V003
Assembly Item Number: PRE4010V004
Lot Numbers:
C16879V004
C17684V004
Assembly Item Number: PRE4010V005
Lot Numbers:
C17007V005
C17534V005
C17959V005
Assembly Item Number: PRE4010V007
Lot Number: C18815V007
Assembly Number: PRE4010V016
UDI-DI codes: Shipper: 50849554014096, Primary Box: 00849554014091
Lot Numbers:
C16880V016
C17179V016
C17685V016
C17960V016
C18445V016
C18816V016
Assembly Item Number: PRE4010V017
UDI-DI Codes: Shipper: 50849554014102, Primary Box: 00849554014107
Lot Numbers:
C16881V017
C17180V017
C17304V017
C17686V017
C17961V017
C18446V017
C18817V017
Assembly Item Number: PRE4010V018
UDI-DI codes: PRE4010V018 - Shipper: 50849554014119, Primary Box: 00849554014114
Lot Numbers:
C18077V018
C18863V018
Assembly Item Number: PRE4010V019
UDI-DI codes: PRE4010V019 - Shipper: 50849554014126, Primary Box: 00849554014121
Lot Numbers:
C17181V019
C17687V019
C18447V019
Assembly Item Number: PRE4010V020
UDI-DI codes: PRE4010V020 - Shipper: 50849554014133, Primary Box: 00849554014138
Lot Numbers:
C16882V020
C17182V020
C17307V020
C17688V020
C18061V020
C18448V020
C18818V020
Assembly Item Number: PRE4010V021
UDI-DI codes: PRE4010V021 - Shipper: 50849554014140, Primary Box: 00849554014145
Lot Numbers:
C17305V021
C18819V021
Assembly Item Number: PRE4010V023
UDI-DI codes: PRE4010V023 - Shipper: 50849554014355, Primary Box: 00849554014350
Lot Number: C18820V023
PRE5001 PREVENA RESTOR ARTHRO-FORM SYSTEM KIT 33X30 CM
UDI-DI code: 00849554006805
Assembly Item Number: PRE5001V001
Lot Numbers:
C16928V001
C17196V001
C17759V001
C18578V001
C18625V001
C19061V001
Assembly Item Number: PRE5001V002
Lot Numbers:
C16748V002
C17760V002
Assembly Item Number: PRE5001V003
Lot Numbers:
C16749V003
C16929V003
Assembly Item Number: PRE5001V004
Lot Numbers:
C16750V004
C18626V004
C19062V004
Assembly Item Number: PRE5001V005
Lot Numbers:
C16930V005
C17761V005
C18069V005
PRE5101 PREVENA RESTOR ARTHRO-FORM SYSTEM KIT 46X30 CM
UDI-DI code: 00849554007116
Assembly Item Number: PRE5101V001
Lot Number: C17762V001
Assembly Item Number: PRE5101V002
Lot Number: C17763V002
Assembly Item Number: PRE5101V003
Lot Number: C17219V003
Assembly Item Number: PRE5101V004
Lot Numbers:
C17765V004
C18070V004
Assembly Item Number: PRE5101V005
Lot Numbers:
C18072V005
C18627V005
PRE5221 PREVENA RESTOR BELLA-FORM SYSTEM KIT 21X19 CM
UDI-DI code: 00849554007154
Assembly Item Number: PRE5221V001
Lot Numbers:
C18579V001
C18628V001
C19064V001
Assembly Item Number: PRE5221V002
Lot Numbers:
C16931V002
C17220V002
Assembly Item Number: PRE5221V004
Lot Numbers:
C18580V004
C18782V004
C19066V004
Assembly Item Number: PRE5221V005
Lot Numbers:
C17221V005
C17767V005
C18783V005
PRE5321 PREVENA RESTOR BELLA-FORM SYSTEM KIT 24X22 CM
UDI-DI code: 00849554007178
Assembly Item Number: PRE5321V001
Lot Numbers:
C18612V001
C18803V001
Assembly Item Number: PRE5321V002
Lot Numbers:
C18613V002
C18613V002
Assembly Item Numbers: PRE5321V003
Lot Numbers:
C16751V003
C18614V003
Assembly Item Number: PRE5321V004
Lot Numbers:
C18615V004
C19538V004
Assembly Item Number: PRE5321V005
Lot Number: C18616V005
PRE5421 PREVENA RESTOR BELLA-FORM SYSTEM KIT 29X27 CM
UDI-DI code: 00849554007130
Assembly Item Number: PRE5421V001
Lot Numbers:
C16932V001
C17222V001
C17768V001
C18073V001
C18581V001
C18784V001
C19067V001
Assembly Item Number: PRE5421V002
Lot Numbers:
C16933V002
C17223V002
Assembly Item Number: PRE5421V003
Lot Numbers:
C17224V003
C17770V003
C18074V003
C18583V003
Assembly Item Number: PRE5421V004
Lot Numbers:
C16752V004
C16934V004
C17225V004
C17771V004
C18584V004
Assembly Item Number: PRE5421V005
Lot Numbers:
C18585V005
C19068V005
PRE5501 PREVENA RESTOR AXIO-FORM SYSTEM KIT 29X28 CM
UDI-DI code: 00849554007826
Assembly Item Number: PRE5501V001
Lot Numbers:
C16935V001
C18076V001
C18586V001
PRE6001 PREVENA RESTOR ADAPTI-FORM SYSTEM KIT
UDI-DI code: Pending
Assembly Item Number: Pending
Lot Numbers: Pending |
| FEI Number |
3009897021
|
Recalling Firm/ Manufacturer |
KCI USA, INC.
12930 W Interstate 10
San Antonio
TX
78249-2248
|
| For Additional Information Contact | Lissa Reitz 651-269-6250 |
Manufacturer Reason for Recall | Due to increase in complaints related to leak alarms |
FDA Determined Cause 2 | Device Design |
| Action | On 01/17/2025, the firm sent an "URGENT - MEDICAL DEVICE RECALL" Letter via US Postal to customers informing them that Solventum has identified that devices distributed between March 11, 2024 and November 18th, 2024 were released with a leak alarm threshold lower than the design specification. This results in the therapy unit prematurely alarming although there is no leak within the system and potentially continuing to alarm during therapy although there is no visible leak.
Customers are instructed to:
1.Screen their facility or other storage locations for the impacted products and lot numbers listed in Attachment 1.
2.Immediately cease use and quarantine any identified product.
3.Complete the Field Notice Receipt and Customer Response Form acknowledging that they have received a copy of this notice and select one of two options:
a.Acknowledge they do not have any product impacted by the field action.
b.Acknowledge they have product impacted by the field action and complete the catalog number, lot numbers and quantities they have identified and quarantined.
4.Return the form via email to kci3mfieldactionresponse@solventum.com.
5.Call Solventum customer service at 1-800-275-4524 to help coordinate the return shipment and receive return labels.
6.Products returned to Solventum will be assessed, reprogrammed, and returned or replaced.
a.Product skus and quantities shipped to Solventum will be returned or replaced with the same quantities.
b.Returned or replaced product will have a visual identifier confirming the assessment and reprogramming.
c.Returned or replaced product may have different lot numbers and expiry dates; Solventum plans to ship available product to help reduce wait times.
For any questions, please contact your local Solventum representative, visit Solventum.com or call 800-275-4524. |
| Quantity in Commerce | 135,268 |
| Distribution | Worldwide - U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV and WY. The countries of Australia, Austria, Bahrain, Belgium, Brazil, Canada, Chile, China, Colombia, Cyprus, Denmark, Finland, France, Germany, Greece, Ireland, Italy, Japan, Jordan, Kuwait, Mexico, Netherlands, New Zealand, Norway, Oman, Qatar, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Turkey, UAE and UK. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = QFC
|
|
|
|