| | Class 2 Device Recall BD PYxis MedBank CUBIE Replenishment Station |  |
| Date Initiated by Firm | January 08, 2025 |
| Date Posted | February 21, 2025 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-1227-2025 |
| Recall Event ID |
96124 |
| Product Classification |
Cabinet, table and tray, anesthesia - Product Code BRY
|
| Product | 138913-01 BD PYXIS MEDBANK MINI CR-2HH-1FH-P 169-114 BD PYXIS MEDBANK MINI CR-2HH-1FH-P 138911-01 BD PYXIS MEDBANK MINI CR-4HH-P 169-115 BD PYXIS MEDBANK MINI CR-4HH-P 139001-01 BD PYXIS MEDBANK MN 200 CR-2HH-1FH-P 139002-01 BD PYXIS MEDBANK MN 200 CR-2HH-1FM-P 139000-01 BD PYXIS MEDBANK MN 200 CR-4HH-P 139020-01 BD PYXIS MEDBANK MN 500 CR-2HH-1FM-2DD-P 139018-01 BD PYXIS MEDBANK MN 500 CR-2HH-5FM-P 139019-01 BD PYXIS MEDBANK MN 500 CR-4HH-1FH-3FM-P 139021-01 BD PYXIS MEDBANK MN 500 CR-4HH-4FM-P 139022-01 BD PYXIS MEDBANK MN 500 CR-6HH-3FM-P 138973-01 BD PYXIS MEDBANK TWR MN CR-10HH-2FH-P 155140-01 BD PYXIS MEDBANK TWR MN CR-10HH-2FH-P 169-82 BD PYXIS MEDBANK TWR MN CR-10HH-2FH-P 138975-01 BD PYXIS MEDBANK TWR MN CR-10HH-2FM-P 155141-01 BD PYXIS MEDBANK TWR MN CR-10HH-2FM-P 169-83 BD PYXIS MEDBANK TWR MN CR-10HH-2FM-P 139073-01 BD PYXIS MEDBANK TWR MN CR-12HH-1FH 155142-01 BD PYXIS MEDBANK TWR MN CR-12HH-1FH 169-84 BD PYXIS MEDBANK TWR MN CR-12HH-1FH 138976-01 BD PYXIS MEDBANK TWR MN CR-12HH-1FM-P 155143-01 BD PYXIS MEDBANK TWR MN CR-12HH-1FM-P 169-85 BD PYXIS MEDBANK TWR MN CR-12HH-1FM-P 138977-01 BD PYXIS MEDBANK TWR MN CR-14HH-P 155144-01 BD PYXIS MEDBANK TWR MN CR-14HH-P 169-86 BD PYXIS MEDBANK TWR MN CR-14HH-P 138951-01 BD PYXIS MEDBANK TWR MN CR-2HH-1FH-5FM 155145-01 BD PYXIS MEDBANK TWR MN CR-2HH-1FH-5FM 169-87 BD PYXIS MEDBANK TWR MN CR-2HH-1FH-5FM 139072-01 BD PYXIS MEDBANK TWR MN CR-2HH-6FH 155146-01 BD PYXIS MEDBANK TWR MN CR-2HH-6FH 169-88 BD PYXIS MEDBANK TWR MN CR-2HH-6FH 139085-01 BD PYXIS MEDBANK TWR MN CR-2HH-6FM 155147-01 BD PYXIS MEDBANK TWR MN CR-2HH-6FM 169-89 BD PYXIS MEDBANK TWR MN CR-2HH-6FM 138968-01 BD PYXIS MEDBANK TWR MN CR-4HH-1FH-4FM-P 155148-01 BD PYXIS MEDBANK TWR MN CR-4HH-1FH-4FM-P 169-90 BD PYXIS MEDBANK TWR MN CR-4HH-1FH-4FM-P 138964-01 BD PYXIS MEDBANK TWR MN CR-4HH-5FM-P 155149-01 BD PYXIS MEDBANK TWR MN CR-4HH-5FM-P 169-91 BD PYXIS MEDBANK TWR MN CR-4HH-5FM-P 138966-01 BD PYXIS MEDBANK TWR MN CR-6HH-1FH-3FM-P 155150-01 BD PYXIS MEDBANK TWR MN CR-6HH-1FH-3FM-P 169-92 BD PYXIS MEDBANK TWR MN CR-6HH-1FH-3FM-P 138967-01 BD PYXIS MEDBANK TWR MN CR-6HH-2FH-2FM-P 155151-01 BD PYXIS MEDBANK TWR MN CR-6HH-2FH-2FM-P 169-93 BD PYXIS MEDBANK TWR MN CR-6HH-2FH-2FM-P 138969-01 BD PYXIS MEDBANK TWR MN CR-6HH-4FM-P 155152-01 BD PYXIS MEDBANK TWR MN CR-6HH-4FM-P 169-94 BD PYXIS MEDBANK TWR MN CR-6HH-4FM-P 138965-01 BD PYXIS MEDBANK TWR MN CR-6HH-8HM-P 155153-01 BD PYXIS MEDBANK TWR MN CR-6HH-8HM-P 169-95 BD PYXIS MEDBANK TWR MN CR-6HH-8HM-P 138971-01 BD PYXIS MEDBANK TWR MN CR-8HH-1FH-2FM-P 155154-01 BD PYXIS MEDBANK TWR MN CR-8HH-1FH-2FM-P 169-96 BD PYXIS MEDBANK TWR MN CR-8HH-1FH-2FM-P 138972-01 BD PYXIS MEDBANK TWR MN CR-8HH-2FH-1FM-P 155155-01 BD PYXIS MEDBANK TWR MN CR-8HH-2FH-1FM-P 169-97 BD PYXIS MEDBANK TWR MN CR-8HH-2FH-1FM-P 138974-01 BD PYXIS MEDBANK TWR MN CR-8HH-3FH-P 155156-01 BD PYXIS MEDBANK TWR MN CR-8HH-3FH-P 169-98 BD PYXIS MEDBANK TWR MN CR-8HH-3FH-P 138970-01 BD PYXIS MEDBANK TWR MN CR-8HH-3FM-P 155157-01 BD PYXIS MEDBANK TWR MN CR-8HH-3FM-P 169-99 BD PYXIS MEDBANK TWR MN CR-8HH-3FM-P |
| Code Information |
MEDFLEX 1119-00 10885403512704
MEDFLEX 2.0 1139-00 10885403512704
MEDBANK TOWER MAIN 1145-00 10885403517167
MEDBANK MINI MAIN 1147-00 10885403517181
BD PYXIS MEDBANK MINI 2FH 138902-01 N/A
BD PYXIS MEDBANK MINI 2FM 138903-01 N/A
BD Pyxis MedBank Twr Mn 4HH-8HM-1FM 138905-01 N/A
BD PYXIS MEDBANK MINI 2HH-1FM 138906-01 N/A
BD PYXIS MEDBANK MINI 2HH-2HM 138907-01 N/A
BD PYXIS MEDBANK MINI 2HH-1FH 138908-01 N/A
BD PYXIS MEDBANK MINI 3HH-1HM 138909-01 N/A
BD PYXIS MEDBANK MINI 4HH 138910-01 N/A
BD PYXIS MEDBANK MINI CR-4HH-P 138911-01 N/A
BD PYXIS MEDBANK MINI 1HH-3HM 138912-01 N/A
BD PYXIS MEDBANK MINI CR-2HH-1FH-P 138913-01 N/A
BD Pyxis MedBank Twr Mn 5HH-9HM 138914-01 N/A
BD PYXIS MEDBANK MINI 4HM 138915-01 N/A
BD PYXIS MEDBANK TWR MN 4HM-5FM 138916-01 N/A
BD PYXIS MEDBANK TWR MN 4HH-5FH 138917-01 N/A
BD PYXIS MEDBANK TWR MN 4HH-5FM 138918-01 N/A
BD PYXIS MEDBANK TWR MN 2HH-6FM 138919-01 N/A
BD PYXIS MEDBANK TWR MN 7FH 138920-01 N/A
BD PYXIS MEDBANK MINI 1HH-1FH-1HM 138921-01 N/A
BD PYXIS MEDBANK TWR MN 6HH-4FM 138922-01 N/A
BD PYXIS MEDBANK TWR MN 8HH-3FM 138923-01 N/A
BD PYXIS MEDBANK TWR MN 10HH-2FM 138924-01 N/A
BD PYXIS MEDBANK TWR MN 12HH-1FM 138926-01 N/A
BD PYXIS MEDBANK TWR MN 14HH 138927-01 N/A
BD PYXIS MEDBANK TWR MN 4HH-2HM-4FM 138929-01 N/A
BD PYXIS MEDBANK TWR MN 4HH-2HM-4FM 138930-01 N/A
BD PYXIS MEDBANK TWR MN 4HH-2FH-3FM 138931-01 N/A
BD PYXIS MEDBANK TWR MN 2HH-1FH-5FM 138932-01 N/A
BD PYXIS MEDBANK TWR MN 3FH-4FM 138933-01 N/A
BD PYXIS MEDBANK TWR MN 4HH-1FH-4FM 138934-01 N/A
BD PYXIS MEDBANK TWR MN 6HH-1FH-3FM 138936-01 N/A
BD PYXIS MEDBANK TWR MN 7FM 138937-01 N/A
BD PYXIS MEDBANK TWR MN 2HH-4HM-4FF 138938-01 N/A
BD PYXIS MEDBANK TWR MN 8HH-2HM-2FM 138939-01 N/A
BD PYXIS MEDBANK TWR MN 4HH-4FH-1FM 138940-01 N/A
BD PYXIS MEDBANK TWR MN 4HH-6HM-2FM 138941-01 N/A
BD PYXIS MEDBANK TWR MN 8HH-1FH-2FM 138942-01 N/A
BD PYXIS MEDBANK TWR MN 10HH-1FH-1FM 138943-01 N/A
BD PYXIS MEDBANK TWR MN 6HH-2FH-2HM-1FM 138944-01 N/A
BD PYXIS MEDBANK TWR MN 4HH-2FH-2HM-2FM 138945-01 N/A
BD PYXIS MEDBANK TWR MN 4HH-3FH-2FM 138946-01 N/A
BD PYXIS MEDBANK TWR MN 4HH-1FH-6HM-1FM 138947-01 N/A
BD PYXIS MEDBANK TWR MN 8HH-3FF 138948-01 N/A
BD PYXIS MEDBANK TWR MN 8HH-1FH-2HM-1FM 138949-01 N/A
BD PYXIS MEDBANK TWR MN 4HH-1FH-4HM-2FM 138950-01 N/A
BD PYXIS MEDBANK TWR MN CR-2HH-1FH-5FM 138951-01 N/A
BD PYXIS MEDBANK TWR MN 4HH-4HM-3FM 138952-01 N/A
BD PYXIS MEDBANK TWR MN 6HH-1FH-2HM-2FM 138953-01 N/A
BD PYXIS MEDBANK TWR MN 4HH-1FH-2HM-3FM 138954-01 N/A
BD PYXIS MEDBANK TWR MN 10HH-2FH 138955-01 N/A
BD PYXIS MEDBANK TWR MN 8HH-3FH 138956-01 N/A
BD PYXIS MEDBANK TWR MN 6HH-4FH 138957-01 N/A
BD PYXIS MEDBANK TWR MN 6HH-8HM 138958-01 N/A
BD PYXIS MEDBANK TWR MN 4HH-10HM 138959-01 N/A
BD PYXIS MEDBANK TWR MN 12HH-1FF 138960-01 N/A
BD PYXIS MEDBANK TWR MN 12HH-1FH 138961-01 N/A
BD PYXIS MEDBANK TWR MN 6HH-2FH-2FM 138962-01 N/A
BD PYXIS MEDBANK TWR MN 2HH-2FM-4FF 138963-01 N/A
BD PYXIS MEDBANK TWR MN CR-4HH-5FM-P 138964-01 N/A
BD PYXIS MEDBANK TWR MN CR-6HH-8HM-P 138965-01 N/A
BD PYXIS MEDBANK TWR MN CR-6HH-1FH-3FM-P 138966-01 N/A
BD PYXIS MEDBANK TWR MN CR-6HH-2FH-2FM-P 138967-01 N/A
BD PYXIS MEDBANK TWR MN CR-4HH-1FH-4FM-P 138968-01 N/A
BD PYXIS MEDBANK TWR MN CR-6HH-4FM-P 138969-01 N/A
BD PYXIS MEDBANK TWR MN CR-8HH-3FM-P 138970-01 N/A
BD PYXIS MEDBANK TWR MN CR-8HH-1FH-2FM-P 138971-01 N/A
BD PYXIS MEDBANK TWR MN CR-8HH-2FH-1FM-P 138972-01 N/A
BD PYXIS MEDBANK TWR MN CR-10HH-2FH-P 138973-01 N/A
BD PYXIS MEDBANK TWR MN CR-8HH-3FH-P 138974-01 N/A
BD PYXIS MEDBANK TWR MN CR-10HH-2FM-P 138975-01 N/A
BD PYXIS MEDBANK TWR MN CR-12HH-1FM-P 138976-01 N/A
BD PYXIS MEDBANK TWR MN CR-14HH-P 138977-01 N/A
BD PYXIS MEDBANK MN 200 CR-4HH-P 139000-01 N/A
BD PYXIS MEDBANK MN 200 CR-2HH-1FH-P 139001-01 N/A
BD PYXIS MEDBANK MN 200 CR-2HH-1FM-P 139002-01 N/A
BD PYXIS MEDBANK MN 500 CR-2HH-5FM-P 139018-01 N/A
BD PYXIS MEDBANK MN 500 CR-4HH-1FH-3FM-P 139019-01 N/A
BD PYXIS MEDBANK MN 500 CR-2HH-1FM-2DD-P 139020-01 N/A
BD PYXIS MEDBANK MN 500 CR-4HH-4FM-P 139021-01 N/A
BD PYXIS MEDBANK MN 500 CR-6HH-3FM-P 139022-01 N/A
BD PYXIS MEDFLEX MN 1000 3FM 139038-01 N/A
BD PYXIS MEDFLEX MN 1000 6HH 139039-01 N/A
BD PYXIS MEDFLEX MN 1000 4HH-1FM 139040-01 N/A
BD PYXIS MEDFLEX MN 1000 2HH-2FM 139041-01 N/A
BD PYXIS MEDBANK TWR MN 2HM-6FM 139043-01 N/A
BD PYXIS MEDFLEX MN 1000 6FM 139044-01 N/A
BD PYXIS MEDBANK TWR MN 2HH-10HM-1FM 139045-01 N/A
BD PYXIS MEDFLEX MN 1000 2HH-5FM 139046-01 N/A
BD PYXIS MEDFLEX MN 1000 4HH-4FM 139049-01 N/A
BD PYXIS MEDFLEX MN 1000 6HH-3FM 139051-01 N/A
BD PYXIS MEDFLEX MN 1000 8HH-2FM 139052-01 N/A
BD PYXIS MEDFLEX MN 1000 2HH-2FH-3FM 139053-01 N/A
BD PYXIS MEDFLEX MN 1000 10HH-1FM 139054-01 N/A
BD PYXIS MEDFLEX MN 1000 12HH 139055-01 N/A
BD PYXIS MEDBANK TWR MN 2FH-2HM-4FM 139056-01 N/A
BD PYXIS MEDFLEX MN 2000 3FM 139058-01 N/A
BD PYXIS MEDFLEX MN 2000 6HH 139059-01 N/A
BD PYXIS MEDFLEX MN 2000 4HH-1FM 139060-01 N/A
BD PYXIS MEDFLEX MN 2000 2HH-2FM 139061-01 N/A
BD PYXIS MEDBANK TWR MN 8HH-2FH-1FM 139062-01 N/A
BD PYXIS MEDBANK TWR MN 6FH-1FM 139063-01 N/A
BD PYXIS MEDFLEX MN 2000 6FM 139064-01 N/A
BD PYXIS MEDFLEX MN 2000 2HH-5FM 139065-01 N/A
BD PYXIS MEDFLEX MN 2000 2HH-3FM-1DD 139066-01 N/A
BD PYXIS MEDFLEX MN 2000 4HH-4FM 139067-01 N/A
BD PYXIS MEDFLEX MN 2000 6HH-3FM 139068-01 N/A
BD PYXIS MEDFLEX MN 2000 8HH-2FM 139069-01 N/A
BD PYXIS MEDFLEX MN 2000 10HH-1FM 139070-01 N/A
BD PYXIS MEDFLEX MN 2000 12HH 139071-01 N/A
BD PYXIS MEDBANK TWR MN CR-2HH-6FH 139072-01 N/A
BD PYXIS MEDBANK TWR MN CR-12HH-1FH 139073-01 N/A
BD PYXIS MEDBANK TWR MN 1HH-1HM-2FM-4FF 139082-01 N/A
BD PYXIS MEDBANK TWR MN 2HH-2FH-4FM 139084-01 N/A
BD PYXIS MEDBANK TWR MN CR-2HH-6FM 139085-01 N/A
BD PYXIS MEDBANK TWR MN 7HH-1HM-3FM 139086-01 N/A
BD PYXIS MEDBANK MEDPASS MODULE 139088-01 N/A
BD PYXIS MEDBANK MINI 3HH-1FG 139091-01 N/A
BD PYXIS MEDBANK MINI 1HH-1HM-1FM 139122-01 N/A
BD PYXIS MEDBANK TWR MN 2HH-4HM-4FM 139125-01 N/A
BD Pyxis MedBank Twr Mn 2HH-6HM-3FM 139126-01 N/A
BD Pyxis MedBank Twr Mn 2HH-8HM-2FM 139127-01 N/A
BD Pyxis MedBank Twr Mn 6HH-6HM-1FM 139128-01 N/A
BD PYXIS MEDBANK MINI 2HM-1FM 139131-01 N/A
BD PYXIS MEDBANK TWR MN 5HH-3HM-3FM 139141-01 N/A
BD PYXIS MEDBANK TWR MN 3HH-7HM-2FM 139142-01 N/A
BD PYXIS MEDBANK TWR MN 9HH-1FG-2FF 139163-01 N/A
BD PYXIS MEDFLEX MN 1000 8HH-2FH 139164-01 N/A
BD PYXIS MEDFLEX MN 2000 8HH-2FH 139165-01 N/A
BD PYXIS MEDBANK TWR MN 10HH-1FH-1FM 155081-01 10885403503870
BD PYXIS MEDBANK TWR MN 10HH-2FH 155082-01 10885403503887
BD PYXIS MEDBANK TWR MN 10HH-2FM 155083-01 10885403503894
BD PYXIS MEDBANK TWR MN 12HH-1FF 155084-01 10885403503900
BD PYXIS MEDBANK TWR MN 12HH-1FH 155085-01 10885403503917
BD PYXIS MEDBANK TWR MN 12HH-1FM 155086-01 10885403503726
BD PYXIS MEDBANK TWR MN 14HH 155087-01 10885403503733
BD PYXIS MEDBANK TWR MN 1HH-1HM-2FM-4FF 155088-01 10885403503740
BD PYXIS MEDBANK TWR MN 2FH-2HM-4FM 155089-01 10885403503757
BD PYXIS MEDBANK TWR MN 2HH-10HM-1FM 155090-01 10885403503764
BD PYXIS MEDBANK TWR MN 2HH-1FH-5FM 155091-01 10885403503771
BD PYXIS MEDBANK TWR MN 2HH-2FH-4FM 155092-01 10885403503788
BD PYXIS MEDBANK TWR MN 2HH-2FM-4FF 155093-01 10885403503795
BD PYXIS MEDBANK TWR MN 2HH-4HM-4FF 155094-01 10885403503801
BD PYXIS MEDBANK TWR MN 2HH-4HM-4FM 155095-01 10885403503818
BD PYXIS MEDBANK TWR MN 2HH-6FM 155096-01 10885403503696
BD PYXIS MEDBANK TWR MN 2HH-6HM-3FM 155097-01 10885403504198
BD PYXIS MEDBANK TWR MN 2HH-8HM-2FM 155098-01 10885403503702
BD PYXIS MEDBANK TWR MN 2HM-6FM 155099-01 10885403504204
BD PYXIS MEDBANK TWR MN 3FH-4FM 155100-01 10885403504211
BD PYXIS MEDBANK TWR MN 3HH-7HM-2FM 155101-01 10885403504228
BD PYXIS MEDBANK TWR MN 4HH-10HM 155102-01 10885403504235
BD PYXIS MEDBANK TWR MN 4HH-1FH-2HM-3FM 155103-01 10885403504242
BD PYXIS MEDBANK TWR MN 4HH-1FH-4FM 155104-01 10885403504259
BD PYXIS MEDBANK TWR MN 4HH-1FH-4HM-2FM 155105-01 10885403503719
BD PYXIS MEDBANK TWR MN 4HH-1FH-6HM-1FM 155106-01 10885403504266
BD PYXIS MEDBANK TWR MN 4HH-2FH-2HM-2FM 155107-01 10885403504273
BD PYXIS MEDBANK TWR MN 4HH-2FH-3FM 155108-01 10885403504280
BD PYXIS MEDBANK TWR MN 4HH-2HM-4FM 155109-01 10885403504327
BD PYXIS MEDBANK TWR MN 4HH-3FH-2FM 155110-01 10885403504297
BD PYXIS MEDBANK TWR MN 4HH-4FH-1FM 155111-01 10885403504303
BD PYXIS MEDBANK TWR MN 4HH-4HM-3FM 155112-01 10885403504099
BD PYXIS MEDBANK TWR MN 4HH-5FH 155113-01 10885403504105
BD PYXIS MEDBANK TWR MN 4HH-5FM 155114-01 10885403504112
BD PYXIS MEDBANK TWR MN 4HH-6HM-2FM 155115-01 10885403504129
BD PYXIS MEDBANK TWR MN 4HH-8HM-1FM 155116-01 10885403504136
BD PYXIS MEDBANK TWR MN 4HM-5FM 155117-01 10885403504143
BD PYXIS MEDBANK TWR MN 5HH-3HM-3FM 155118-01 10885403504150
BD PYXIS MEDBANK TWR MN 5HH-9HM 155119-01 10885403504334
BD PYXIS MEDBANK TWR MN 6FH-1FM 155120-01 10885403504167
BD PYXIS MEDBANK TWR MN 6HH-1FH-2HM-2FM 155121-01 10885403504174
BD PYXIS MEDBANK TWR MN 6HH-1FH-3FM 155122-01 10885403504181
BD PYXIS MEDBANK TWR MN 6HH-2FH-2FM 155123-01 10885403504082
BD PYXIS MEDBANK TWR MN 6HH-2FH-2HM-1FM 155124-01 10885403504075
BD PYXIS MEDBANK TWR MN 6HH-4FH 155125-01 10885403503658
BD PYXIS MEDBANK TWR MN 6HH-4FM 155126-01 10885403503665
BD PYXIS MEDBANK TWR MN 6HH-4HM-2FM 155127-01 10885403503672
BD PYXIS MEDBANK TWR MN 6HH-6HM-1FM 155128-01 10885403503689
BD PYXIS MEDBANK TWR MN 6HH-8HM 155129-01 10885403504341
BD PYXIS MEDBANK TWR MN 7FH 155130-01 10885403504358
BD PYXIS MEDBANK TWR MN 7FM 155131-01 10885403504365
BD PYXIS MEDBANK TWR MN 7HH-1HM-3FM 155132-01 10885403504372
BD PYXIS MEDBANK TWR MN 8HH-1FH-2FM 155133-01 10885403503542
BD PYXIS MEDBANK TWR MN 8HH-1FH-2HM-1FM 155134-01 10885403503559
BD PYXIS MEDBANK TWR MN 8HH-2FH-1FM 155135-01 10885403503566
BD PYXIS MEDBANK TWR MN 8HH-2HM-2FM 155136-01 10885403503573
BD PYXIS MEDBANK TWR MN 8HH-3FF 155137-01 10885403503597
BD PYXIS MEDBANK TWR MN 8HH-3FH 155138-01 10885403503603
BD PYXIS MEDBANK TWR MN 8HH-3FM 155139-01 10885403503610
BD PYXIS MEDBANK TWR MN CR-10HH-2FH-P 155140-01 10885403504440
BD PYXIS MEDBANK TWR MN CR-10HH-2FM-P 155141-01 10885403503580
BD PYXIS MEDBANK TWR MN CR-12HH-1FH 155142-01 10885403504457
BD PYXIS MEDBANK TWR MN CR-12HH-1FM-P 155143-01 10885403504464
BD PYXIS MEDBANK TWR MN CR-14HH-P 155144-01 10885403504471
BD PYXIS MEDBANK TWR MN CR-2HH-1FH-5FM 155145-01 10885403504488
BD PYXIS MEDBANK TWR MN CR-2HH-6FH 155146-01 10885403504754
BD PYXIS MEDBANK TWR MN CR-2HH-6FM 155147-01 10885403504761
BD PYXIS MEDBANK TWR MN CR-4HH-1FH-4FM-P 155148-01 10885403504778
BD PYXIS MEDBANK TWR MN CR-4HH-5FM-P 155149-01 10885403504785
BD PYXIS MEDBANK TWR MN CR-6HH-1FH-3FM-P 155150-01 10885403504792
BD PYXIS MEDBANK TWR MN CR-6HH-2FH-2FM-P 155151-01 10885403504709
BD PYXIS MEDBANK TWR MN CR-6HH-4FM-P 155152-01 10885403504716
BD PYXIS MEDBANK TWR MN CR-6HH-8HM-P 155153-01 10885403504723
BD PYXIS MEDBANK TWR MN CR-8HH-1FH-2FM-P 155154-01 10885403504730
BD PYXIS MEDBANK TWR MN CR-8HH-2FH-1FM-P 155155-01 10885403504747
BD PYXIS MEDBANK TWR MN CR-8HH-3FH-P 155156-01 10885403504624
BD PYXIS MEDBANK TWR MN CR-8HH-3FM-P 155157-01 10885403504631
BD PYXIS MEDBANK MINI 1HH-1FH-1HM 155159-01 10885403504655
BD PYXIS MEDBANK MINI 1HH-1HM-1FM 155160-01 10885403504662
BD PYXIS MEDBANK MINI 1HH-3HM 155161-01 10885403504679
BD PYXIS MEDBANK MINI 2FM 155163-01 10885403504686
BD PYXIS MEDBANK MINI 2HH-1FM 155165-01 10885403504693
BD PYXIS MEDBANK MINI 2HH-2HM 155166-01 10885403504570
BD PYXIS MEDBANK MINI 2HM-1FM 155167-01 10885403504587
BD PYXIS MEDBANK MINI 3HH-1FG 155168-01 10885403504594
BD PYXIS MEDBANK MINI 3HH-1HM 155169-01 10885403504600
BD PYXIS MEDBANK MINI 4HM 155171-01 10885403504617
BD PYXIS MEDBANK MINI 1FH-1FM 155288-01 10885403508721
BD PYXIS MEDBANK TWR MN 1FH-10HM-1FM 155293-01 10885403509926
BD PYXIS MEDBANK TWR MN 2HH-4FH-2FM 155294-01 10885403509933
BD PYXIS MEDBANK TWR MN 3FH-8HH 155299-01 10885403510526
BD PYXIS MEDBANK TWR MN 4FH-4HH-1FM 155302-01 10885403510762
BD PYXIS MEDBANK TWR MN 1FM-10HH-1FH 155303-01 10885403510779
BD PYXIS MEDBANK TWR MN 2FM-8HH-1FH 155304-01 10885403510786
BD PYXIS MEDBANK TWR MN 2FH-1FM-4HH-2FM 155305-01 10885403510793
BD PYXIS MEDBANK MINI 4HH 155310-01 10885403510830
BD PYXIS MEDBANK MINI 1FH-2HH 155311-01 10885403510847
BD PYXIS MEDBANK MINI 1HH-1FH-1HM 169-101 10885403512544
BD PYXIS MEDBANK MINI 1HH-1HM-1FM 169-102 10885403512544
BD PYXIS MEDBANK MINI 1HH-3HM 169-103 10885403512544
BD PYXIS MEDBANK MINI 2FH 169-104 10885403512544
BD PYXIS MEDBANK MINI 2FM 169-105 10885403512544
BD PYXIS MEDBANK MINI 2HH-1FH 169-106 10885403512544
BD PYXIS MEDBANK MINI 2HH-1FM 169-107 10885403512544
BD PYXIS MEDBANK MINI 2HH-2HM 169-108 10885403512544
BD PYXIS MEDBANK MINI 2HM-1FM 169-109 10885403512544
BD PYXIS MEDBANK MINI 3HH-1FG 169-110 10885403512544
BD PYXIS MEDBANK MINI 3HH-1HM 169-111 10885403512544
BD PYXIS MEDBANK MINI 4HH 169-112 10885403512544
BD PYXIS MEDBANK MINI 4HM 169-113 10885403512544
BD PYXIS MEDBANK MINI CR-2HH-1FH-P 169-114 10885403512544
BD PYXIS MEDBANK MINI CR-4HH-P 169-115 10885403512544
BD PYXIS MEDBANK MINI 1FH-1FM 169-137 10885403512544
BD PYXIS MEDBANK TWR MN 1FH-10HM-1FM 169-138 10885403512568
BD PYXIS MEDBANK TWR MN 2HH-4FH-2FM 169-139 10885403512568
BD PYXIS MEDBANK TWR MN 3FH-8HH MEDPASS 169-140 10885403512568
BD PYXIS MEDBANK MINI 4HH MEDPASS 169-150 10885403512544
BD PYXIS MEDBANK MINI 1FH-2HH MEDPASS 169-151 10885403512544
BD PYXIS MEDBANK MINI 2HH-1FM MEDPASS 169-152 10885403512544
BD PYXIS MEDBANK TWR MN 10HH-1FH-1FM 169-23 10885403512568
BD PYXIS MEDBANK TWR MN 10HH-2FH 169-24 10885403512568
BD PYXIS MEDBANK TWR MN 10HH-2FM 169-25 10885403512568
BD PYXIS MEDBANK TWR MN 12HH-1FF 169-26 10885403512568
BD PYXIS MEDBANK TWR MN 12HH-1FH 169-27 10885403512568
BD PYXIS MEDBANK TWR MN 12HH-1FM 169-28 10885403512568
BD PYXIS MEDBANK TWR MN 14HH 169-29 10885403512568
BD PYXIS MEDBANK TWR MN 1HH-1HM-2FM-4FF 169-30 10885403512568
BD PYXIS MEDBANK TWR MN 2FH-2HM-4FM 169-31 10885403512568
BD PYXIS MEDBANK TWR MN 2HH-10HM-1FM 169-32 10885403512568
BD PYXIS MEDBANK TWR MN 2HH-1FH-5FM 169-33 10885403512568
BD PYXIS MEDBANK TWR MN 2HH-2FH-4FM 169-34 10885403512568
BD PYXIS MEDBANK TWR MN 2HH-2FM-4FF 169-35 10885403512568
BD PYXIS MEDBANK TWR MN 2HH-4HM-4FF 169-36 10885403512568
BD PYXIS MEDBANK TWR MN 2HH-4HM-4FM 169-37 10885403512568
BD PYXIS MEDBANK TWR MN 2HH-6FM 169-38 10885403512568
BD PYXIS MEDBANK TWR MN 2HH-6HM-3FM 169-39 10885403512568
BD PYXIS MEDBANK TWR MN 2HH-8HM-2FM 169-40 10885403512568
BD PYXIS MEDBANK TWR MN 2HM-6FM 169-41 10885403512568
BD PYXIS MEDBANK TWR MN 3FH-4FM 169-42 10885403512568
BD PYXIS MEDBANK TWR MN 3HH-7HM-2FM 169-43 10885403512568
BD PYXIS MEDBANK TWR MN 4HH-10HM 169-44 10885403512568
BD PYXIS MEDBANK TWR MN 4HH-1FH-2HM-3FM 169-45 10885403512568
BD PYXIS MEDBANK TWR MN 4HH-1FH-4FM 169-46 10885403512568
BD PYXIS MEDBANK TWR MN 4HH-1FH-4HM-2FM 169-47 10885403512568
BD PYXIS MEDBANK TWR MN 4HH-1FH-6HM-1FM 169-48 10885403512568
BD PYXIS MEDBANK TWR MN 4HH-2FH-2HM-2FM 169-49 10885403512568
BD PYXIS MEDBANK TWR MN 4HH-2FH-3FM 169-50 10885403512568
BD PYXIS MEDBANK TWR MN 4HH-2HM-4FM 169-51 10885403512568
BD PYXIS MEDBANK TWR MN 4HH-3FH-2FM 169-52 10885403512568
BD PYXIS MEDBANK TWR MN 4HH-4FH-1FM 169-53 10885403512568
BD PYXIS MEDBANK TWR MN 4HH-4HM-3FM 169-54 10885403512568
BD PYXIS MEDBANK TWR MN 4HH-5FH 169-55 10885403512568
BD PYXIS MEDBANK TWR MN 4HH-5FM 169-56 10885403512568
BD PYXIS MEDBANK TWR MN 4HH-6HM-2FM 169-57 10885403512568
BD PYXIS MEDBANK TWR MN 4HH-8HM-1FM 169-58 10885403512568
BD PYXIS MEDBANK TWR MN 4HM-5FM 169-59 10885403512568
BD PYXIS MEDBANK TWR MN 5HH-3HM-3FM 169-60 10885403512568
BD PYXIS MEDBANK TWR MN 5HH-9HM 169-61 10885403512568
BD PYXIS MEDBANK TWR MN 6FH-1FM 169-62 10885403512568
BD PYXIS MEDBANK TWR MN 6HH-1FH-2HM-2FM 169-63 10885403512568
BD PYXIS MEDBANK TWR MN 6HH-1FH-3FM 169-64 10885403512568
BD PYXIS MEDBANK TWR MN 6HH-2FH-2FM 169-65 10885403512568
BD PYXIS MEDBANK TWR MN 6HH-2FH-2HM-1FM 169-66 10885403512568
BD PYXIS MEDBANK TWR MN 6HH-4FH 169-67 10885403512568
BD PYXIS MEDBANK TWR MN 6HH-4FM 169-68 10885403512568
BD PYXIS MEDBANK TWR MN 6HH-4HM-2FM 169-69 10885403512568
BD PYXIS MEDBANK TWR MN 6HH-6HM-1FM 169-70 10885403512568
BD PYXIS MEDBANK TWR MN 6HH-8HM 169-71 10885403512568
BD PYXIS MEDBANK TWR MN 7FH 169-72 10885403512568
BD PYXIS MEDBANK TWR MN 7FM 169-73 10885403512568
BD PYXIS MEDBANK TWR MN 7HH-1HM-3FM 169-74 10885403512568
BD PYXIS MEDBANK TWR MN 8HH-1FH-2FM 169-75 10885403512568
BD PYXIS MEDBANK TWR MN 8HH-1FH-2HM-1FM 169-76 10885403512568
BD PYXIS MEDBANK TWR MN 8HH-2FH-1FM 169-77 10885403512568
BD PYXIS MEDBANK TWR MN 8HH-2HM-2FM 169-78 10885403512568
BD PYXIS MEDBANK TWR MN 8HH-3FF 169-79 10885403512568
BD PYXIS MEDBANK TWR MN 8HH-3FH 169-80 10885403512568
BD PYXIS MEDBANK TWR MN 8HH-3FM 169-81 10885403512568
BD PYXIS MEDBANK TWR MN CR-10HH-2FH-P 169-82 10885403512568
BD PYXIS MEDBANK TWR MN CR-10HH-2FM-P 169-83 10885403512568
BD PYXIS MEDBANK TWR MN CR-12HH-1FH 169-84 10885403512568
BD PYXIS MEDBANK TWR MN CR-12HH-1FM-P 169-85 10885403512568
BD PYXIS MEDBANK TWR MN CR-14HH-P 169-86 10885403512568
BD PYXIS MEDBANK TWR MN CR-2HH-1FH-5FM 169-87 10885403512568
BD PYXIS MEDBANK TWR MN CR-2HH-6FH 169-88 10885403512568
BD PYXIS MEDBANK TWR MN CR-2HH-6FM 169-89 10885403512568
BD PYXIS MEDBANK TWR MN CR-4HH-1FH-4FM-P 169-90 10885403512568
BD PYXIS MEDBANK TWR MN CR-4HH-5FM-P 169-91 10885403512568
BD PYXIS MEDBANK TWR MN CR-6HH-1FH-3FM-P 169-92 10885403512568
BD PYXIS MEDBANK TWR MN CR-6HH-2FH-2FM-P 169-93 10885403512568
BD PYXIS MEDBANK TWR MN CR-6HH-4FM-P 169-94 10885403512568
BD PYXIS MEDBANK TWR MN CR-6HH-8HM-P 169-95 10885403512568
BD PYXIS MEDBANK TWR MN CR-8HH-1FH-2FM-P 169-96 10885403512568
BD PYXIS MEDBANK TWR MN CR-8HH-2FH-1FM-P 169-97 10885403512568
BD PYXIS MEDBANK TWR MN CR-8HH-3FH-P 169-98 10885403512568
BD PYXIS MEDBANK TWR MN CR-8HH-3FM-P 169-99 10885403512568
BD PYXIS MEDBANK TWR MN 1FM-10HH-1FH 169-143 n/a
BD PYXIS MEDBANK TWR MN 1FM-10HH-1FH 169-145 n/a
BD PYXIS MEDBANK TWR MN 2FM-8HH-1FH 169-144 n/a
BD PYXIS MEDBANK TWR MN 4FH-4HH-1FM 169-142 n/a
MEDBANK GUIDED SELLING 1144-00 n/a
|
| FEI Number |
2016493
|
Recalling Firm/ Manufacturer |
CareFusion 303, Inc.
10020 Pacific Mesa Blvd
San Diego
CA
92121-4386
|
Manufacturer Reason for Recall | Labeling update to include a contraindication statement against the use of medicine cabinets in procedure and operating rooms. |
FDA Determined Cause 2 | Under Investigation by firm |
| Action | On January 8, 2025, CareFusion 303 issued a "Urgent: Medical Device Correction" Notification to affected customer via mail or Email. CareFusion asked consignees to take the following actions:
1. Customers should immediately ensure that MedBank" devices are not installed in the procedural or operating space and at least 1.5 meters from patients.
2. Report any issues associated with the above products to BD Pyxis" MedBank" Product Support Line via phone at 866-930-9251.
3. Report any adverse health consequences experienced with the use of this product to the FDA's MedWatch Adverse Event Reporting program.
4. Please post these notices with the device until labeling has been updated.
5. Complete the attached Customer Response Form and return to BD. |
| Quantity in Commerce | 6,075 units |
| Distribution | US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY.
|
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
|
|
|