| | Class 2 Device Recall Alcon Custom Pak with NonLatex Symbol, Pak |  |
| Date Initiated by Firm | February 10, 2025 |
| Date Posted | March 06, 2025 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-1293-2025 |
| Recall Event ID |
96285 |
| 510(K)Number | K880961 |
| Product Classification |
General surgery tray - Product Code LRO
|
| Product | Alcon Custom Pak with Non-Latex Symbol, Pak: 12771-18, 18050-10, 18252-17, 12698-09, 15632-10, 3141-45, 12630-07, 18925-05, 19265-05, 13894-16, 17550-08, 12236-18, 11854-15, 8043-29, 12466-17, 15599-17, 14764-09, 13194-16, 17069-09 |
| Code Information |
UDI-DI: H5301ALCON1CPAK10.
Pak/Lot(Expiration):
12771-18/JZ5190507(3/31/2025);
18050-10/JZ5200922(5/31/2025), JZ5202199(9/30/2025);
18252-17/JZ5200852(5/31/2025);
12698-09/JZ5201264(6/30/2025);
15632-10, JZ5202447(6/30/2025), JZ5202445(6/30/2025);
3141-45/ JZ5201303(7/31/2025), JZ5201305(7/31/2025);
12630-07/ JZ5201295(8/31/2025);
18925-05/JZ5194816(9/30/2025), JZ5194818(9/30/2025), JZ5196180(9/30/2025), JZ5196182(9/30/2025), JZ5200323(11/30/2025);
19265-05/JZ5201146(9/30/2025), JZ5201148(9/30/2025), JZ5202207(9/30/2025);
13894-16/JZ5201181(9/30/2025), JZ5202143(9/30/2025);
17550-08/JZ5198232(10/31/2025);
12236-18/JZ5201054(11/30/2025);
11854-15/JZ5201561(11/30/2025);
8043-29/JZ5202201(11/30/2025);
12466-17/JZ5205311(12/31/2025), JZ5205313(12/31/2025), JZ5205915(1/31/2026);
15599-17/JZ5205311(12/31/2025), JZ5205313(12/31/2025), JZ5205915(1/31/2026);
14764-09/JZ5197158(2/28/2026);
13194-16/JZ5197142(3/31/2026);
17069-09/JZ5201931(9/30/2026)
|
| FEI Number |
1610287
|
Recalling Firm/ Manufacturer |
Alcon Research LLC
6201 South Fwy
Fort Worth
TX
76134-2099
|
| For Additional Information Contact | Kristen Kellerhals 817-551-4047 |
Manufacturer Reason for Recall | Customer-designed sterile surgical procedure packs contain latex components as specified; but, the content label sealed within the transparent sterile pouch contains the symbol for a latex-free product. If users or patients with latex sensitivity come into contact with latex, an allergic reaction may occur. |
FDA Determined Cause 2 | Software design |
| Action | On 2/10/2025, correction notices were mailed to customers who were asked to do the following:
1) Do not open packaging. Using the provided stickers, adhere one sticker to the outside of affected product near the label to identify it as containing latex.
2) Complete and return the response form electronically via: https://qrco.de/non-latex-symbol
Should you have any questions or concerns about this matter, please feel free to call the firm at A 1-800-862-5266
|
| Quantity in Commerce | 2511 |
| Distribution | US Nationwide distribution in the states of AK, AR, AZ, FL, MD, MN, NC, NE, OH, PA, TX, WA, WI.
|
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = LRO
|
|
|
|