| | Class 1 Device Recall BiPAP V30 |  |
| Date Initiated by Firm | May 09, 2025 |
| Date Posted | July 02, 2025 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-2035-2025 |
| Recall Event ID |
96948 |
| Product Classification |
Ventilator, continuous, non-life-supporting - Product Code MNS
|
| Product | BiPAP V30 Auto Ventilator. Intended to provide non-invasive ventilatory support for pediatric and adult patients weighing over 10 kg. |
| Code Information |
Part Number 1111178; UDI: 0060959049635; Serial No. N2068642229E4, N206864244CD2, N206864255D5B, N206864277E49, N2068642886BE, N206864299737, N2068643102A7, N20686432303C, N2068643321B5, N20686434550A, N206864354483, N206864367618, N206864376791, N206864389F66, N206864398EEF, N206864405E26, N206864414FAF, N206864427D34, N206864441802, N20686445098B, N206864463B10, N206864472A99, N20686448D26E, N20686449C3E7, N2068645047FE, N206864515677, N2068645264EC, N206864537565, N2068645401DA, N206864551053, N2068645622C8, N206864573341, N20686458CBB6, N206864606D96, N206864617C1F, N206864624E84, N206864635F0D, N206864653A3B, N2068646608A0, N2070799636A7, N20708029041F, N20708103C229, N20708105A71F, N2070811974AB, N20708120C302, N20708137AE65, N20708141865B, N20708147E36D, N20713207A038, N20713212EE4D, N207132184117, N20713222C425, N20713232DDFD, N2071323872A7, N20713245E44A, N207132582677, N20713270998F, N207132810BCE, N2071329445BB, N2071329666A9, N207132977720, N207132988FD7, N207132999E5E, N20713309139A, N20713314D1A7, N207133190A42, N20713327C954, N207133328721, N2071333396A8, N2071335695D5, N20713363E810, N2071337594FE, N20713377B7EC, N2071338625AD, N207133972DFC, N207134088E16, N207134099F9F, N207134145DA2, N207134212067, N2071342654D8, N207134375C89, N20713442462C, N2071344357A5, N207134507CE6, N207134726C44, N2071347518FB, N207134762A60, N20713478C31E, N207134859B33, N205345571B7A, N20534558E38D, N20534559F204, N2053456045AD, N2053456266BF, N205345637736, N205345640389, N205345651200, N20534566209B, N205345673112, N20534568C9E5, N20534569D86C, N205345705C75, N205345714DFC, N205345727F67, N205345736EEE, N205345741A51, N205345750BD8, N206864266FC0, N20686430132E, N206864436CBD, N206864671929, N20686470744E, N2068647165C7, N20686472575C, N2068647346D5, N20686474326A, N206864761178, N2068647700F1, N20686478F806, N20686479E98F, N20686482D494, N20686483C51D, N20686484B1A2, N20686485A02B, N2068648692B0, N206864878339, N206864887BCE, N20713231EF66, N207132955432, N207133019FD2, N207134740972, N207134773BE9, N20713479D297, N205345615424, N20686469F057, N20686480F786, N207082010E5F, N20713263B2CC, N206864642BB2, N20686468E1DE, N20686481E60F, N20713350F0E3, N207134737DCD, N209923639729, N20992366C084, N2099236829FA, N209923693873, N20992371ADE3, N20992376D95C, N20992377C8D5, N2099237921AB, N209923812E2B, N209923920568, N2099239314E1, N2099239571D7, N20992396434C, N20992398AA32, N20992399BBBB, N209924043B43, N209924052ACA, N2099240709D8, N2099241064BF, N2099241247AD, N209924153312, N209924171000, N20992418E8F7, N209924237C4C, N20992425197A, N209924262BE1, N20992430570F, N20992432741D, N209924336594, N20992434112B, N209924401A07, N209924423915, N20992443289C, N2099245003DF, N209924511256, N209924533144, N2099245666E9, N209924599E1E, N2099246029B7, N209924620AA5, N209924646F93, N20992468A5FF, N20992469B476, N2099247121E6, N20992472137D, N2099247744D0, N20992479ADAE, N20992480B3A7, N20992481A22E, N209924007D67, N20992487C718, N209924892E66, N20992491BBF6, N20992492896D, N2099249398E4, N20992495FDD2, N20992496CF49, N20992508ABF3, N209925103E63, N209925112FEA, N20992390267A, N209924215F5E, N20992509BA7A, N209925075304, N20992506428D, N212172002D39, N212172013CB0, N21249204E30B, N21249205F282, N21249206C019, N21249207D190, N212492082967, N2124920938EE, N21197670550D, N211976714484, N21197672761F, N211976736796, N211976741329, N2119767502A0, N21197676303B, N2119767721B2, N21197679C8CC, N21197680D6C5, N21197682F5D7, N21197683E45E, N2119768490E1, N21197686B3F3, N211976885A8D, N211976894B04, N21197690CF1D, N21197691DE94, N21197692EC0F, N21197693FD86, N211976948939, N2119769598B0, N21197696AA2B, N21197697BBA2, N211976984355, N2119769952DC, N211977015350, N2119770261CB, N211977037042, N2119770404FD, N211977051574, N2119770627EF, N211977073666, N21197708CE91, N21197709DF18, N211977114A88, N211977127813, N211977141D25, N21249210BCF7, N21249211AD7E, N212492129FE5, N212492138E6C, N21249214FAD3, N21249215EB5A, N21249216D9C1, N2124921830BF, N212492192136, N21249220969F, N212492218716, N21249222B58D, N21249223A404, N21249224D0BB, N21249225C132, N21249226F3A9, N21249227E220, N212492281AD7, N212492290B5E, N212492308F47, N212492319ECE, N21249232AC55, N21249233BDDC, N21249234C963, N21249235D8EA, N21249236EA71, N21249237FBF8, N21249238030F, N21197678D945, N211977150CAC, N211977163E37, N211977172FBE, N21197718D749, N21197719C6C0, N211977207169, N2119772160E0, N21197722527B, N2119772343F2, N21197724374D, N2119772526C4, N21197726145F, N2119772705D6, N21197729ECA8, N212172031FA2, N211976858168, N2119770042D9, N211977105B01, N2135323639B0, N213532372839, N21353239C147, N21353240118E, N213532410007, N213532432315, N2135324457AA, N213532454623, N2135324674B8, N213532476531, N213532489DC6, N213532498C4F, N213532522B44, N213532533ACD, N213532544E72, N2139626309C2, N2139627647B7, N21396278AEC9, N21396279BF40, N21396290B891, N21396291A918, N21396296DDA7, N2139629834D9, N213963021647, N2139630307CE, N21437204656F, N21437206467D, N2143720757F4, N21437208AF03, N21437209BE8A, N214372103A93, N214372112B1A, N214372130808, N214372147CB7, N214372156D3E, N214372165FA5, N214372174E2C, N21437218B6DB, N21437219A752, N2143722010FB, N214372210172, N2143722233E9, N214372232260, N2143722456DF, N214372254756, N2143722675CD, N214372276444, N214372289CB3, N214372300923, N2143723118AA, N214372322A31, N214372333BB8, N214372344F07, N214372355E8E, N214372377D9C, N214372121981, N214372298D3A, N21448204E74A, N21448205F6C3, N21448206C458, N21448207D5D1, N214482082D26, N214482093CAF, N21448210B8B6, N21448211A93F, N214482129BA4, N214482138A2D, N21448214FE92, N21448215EF1B, N21448216DD80, N2144821834FE, N214482192577, N2144822092DE, N214482218357, N21448222B1CC, N21448223A045, N21448224D4FA, N21448225C573, N21448226F7E8, N21448227E661, N214482281E96, N214482290F1F, N214482308B06, N214482319A8F, N21448232A814, N21448233B99D, N21448234CD22, N21448235DCAB, N21448237FFB9, N21448238074E, N2144823916C7, N21448240C60E, N21448241D787, N21448242E51C, N21448243F495, N21448244802A, N2144824591A3, N21448246A338, N21448247B2B1, N214482484A46, N214372366C15, N21448217CC09, N21448236EE30, N21437384F173, N21437385E0FA, N21437386D261, N21437387C3E8, N214373883B1F, N214373892A96, N21437390AE8F, N21437391BF06, N214373928D9D, N214373939C14, N21437394E8AB, N21437395F922, N21437397DA30, N2143739822C7, N21437396CBB9, N2144943150ED, N2144943373FF, N214494340740, N2144943516C9, N214494362452, N214494411DE5, N214494444A48, N2144944778D3, N2144945236A6, N214494554219, N2144945899FC, N214494612E55, N214494545390, N214494674B63, N21449469A21D, N214494720516, N2144947571A9, N214494764332, N21449478AA4C, N21449481B445, N21449484E3E8, N21449451043D, N214494567082, N21449471378D, N214494746020, N214494598875, N214494630D47, N21449485F261, N2154157900EA, N215415801EE3, N215415810F6A, N215415832C78, N21541585494E, N215415867BD5, N2154158892AB, N215415898322, N21541590073B, N2154159116B2, N215415922429, N2154159335A0, N21541594411F, N215415955096, N21541596620D, N215415988B73, N215415999AFA, N215416003F47, N215416012ECE, N215416021C55, N215416030DDC, N215416047963, N2154160568EA, N215416065A71, N215416074BF8, N21541608B30F, N21541609A286, N21541610269F, N215416113716, N21541612058D, N215416131404, N21449438CD2C, N2154158458C7, N215415876A5C, N214494400C6C, N216765383093, N21676539211A, N21676540F1D3, N21676541E05A, N21676542D2C1, N21676543C348, N21676544B7F7, N21676545A67E, N2167654694E5, N21676547856C, N216765487D9B, N216765496C12, N21676550E80B, N21676551F982, N21676552CB19, N21676553DA90, N21676554AE2F, N21676555BFA6, N216765579CB4, N216765586443, N21676560C263, N21676561D3EA, N21676562E171, N21676563F0F8, N2167656595CE, N21676567B6DC, N216765684E2B, N216765695FA2, N21676570DBBB, N21676571CA32, N21676572F8A9, N21676573E920, N216765749D9F, N216765758C16, N21676576BE8D, N2167655975CA, N21676577AF04, N2167657857F3, N21676579467A, N216765805873, N216765827B61, N216765836AE8, N216765841E57, N216765850FDE, N216765863D45, N216765872CCC, N21676588D43B, N21676589C5B2, N2167659041AB, N216765915022, N2167659262B9, N216765937330, N21676594078F, N216765951606, N21676596249D, N216765973514, N21676598CDE3, N21676599DC6A, N2167660079D7, N21676601685E, N216766025AC5, N216766034B4C, N2167658149FA, N21676566A755, N214811493AF8, N21481150BEE1, N21481151AF68, N214811529DF3, N214811538C7A, N21481154F8C5, N21481155E94C, N21481156DBD7, N21481157CA5E, N2148115832A9, N214811592320, N214811609489, N214811618500, N21481162B79B, N21481163A612, N21481164D2AD, N21481165C324, N21481166F1BF, N21481167E036, N2148116818C1, N214811708D51, N214811719CD8, N21481173BFCA, N21481174CB75, N21481175DAFC, N21481176E867, N21481177F9EE, N214811780119, N2144916117E8, N214491622573, N2144916334FA, N214491644045, N2144916551CC, N214491666357, N2144916772DE, N214491688A29, N214491699BA0, N214491701FB9, N214491710E30, N214491723CAB, N21449174599D, N214491754814, N214491767A8F, N214491776B06, N2144917893F1, N214491798278, N214491809C71, N216768071D73, N21676809F40D, N216768107014, N21676811619D, N216768125306, N21676813428F, N216768161522, N2167681704AB, N21676818FC5C, N21676819EDD5, N216768205A7C, N216768214BF5, N21676822796E, N2167682368E7, N216768241C58, N216768250DD1, N216768272EC3, N21676828D634, N21676829C7BD, N2167683043A4, N21676831522D, N2167683260B6, N21676833713F, N216768340580, N216768351409, N216768362692, N21676837371B, N21676838CFEC, N21676839DE65, N216768400EAC, N216768411F25, N216768433C37, N216768444888, N216768455901, N216768466B9A, N216768477A13, N2167684882E4, N21676849936D, N216768501774, N2167685106FD, N216768523466, N2167685325EF, N216768545150, N2167685540D9, N216768567242, N2167685763CB, N216768589B3C, N216768598AB5, N216768603D1C, N216768621E0E, N216768630F87, N216768656AB1, N21676866582A, N2167686749A3, N21676868B154, N21676869A0DD, N2167687024C4, N21676871354D, N2167687207D6, N21676873165F, N2167687462E0, N214491732D22, N21676808E584, N216768143630, N2167681527B9, N216768263F4A, N216768422DBE, N216768757369, N21676877507B, N21676878A88C, N21676879B905, N21676880A70C, N21676882841E, N216768839597, N21481172AE43, N216768612C95, N216768647B38, N2167687641F2, N21676881B685, N21676884E128, N21449186F947, N21449187E8CE, N214491881039, N214491919420, N21449196E09F, N2144918901B0, N2144919085A9, N21449192A6BB, N21449193B732, N21449194C38D, N21449197F116, N2144919809E1, N214491991868, N21449200BDD5, N21449195D204, N21677157E36E, N216771590A10, N21677160BDB9, N21677161AC30, N216771629EAB, N216771638F22, N21677164FB9D, N21677165EA14, N21677166D88F, N2191827873A5, N219182816DAC, N219182825F37, N219182861913, N21918287089A, N21918288F06D, N21918289E1E4, N2191829246EF, N2191829423D9, N219182953250, N21918300E839, N21918301F9B0, N21918302CB2B, N21918304AE1D, N219183068D0F, N2191830975F8, N21918310F1E1, N21918314B7C5, N21918315A64C, N2191831694D7, N21918321CA00, N21918322F89B, N21918323E912, N219183258C24, N21918331D3D8, N21918332E143, N21918333F0CA, N219183348475, N219182807C25, N219182852B88, N2191829600CB, N219183086471, N21918337B6EE, N219183395F90, N219183408F59, N219183419ED0, N21918346EA6F, N21918347FBE6, N219183480311, N21918468BCA4, N2191847138BD, N219184720A26, N219184731BAF, N219184746F10, N219184757E99, N219184764C02, N219184775D8B, N21918478A57C, N21918479B4F5, N21918480AAFC, N21918481BB75, N2191848289EE, N219184839867, N21918484ECD8, N21918485FD51, N21918486CFCA, N21918487DE43, N2191848826B4, N219184702934, N21918469AD2D, N218927203A4D, N21892722195F, N2189272308D6, N218927247C69, N218927256DE0, N218927357438, N2189273646A3, N21892738AFDD, N218927406E9D, N218927605D2D, N218927636FB6, N218927650A80, N218927672992, N2189280152B3, N21892806260C, N21892809DEFB, N2189281279F0, N21892838E51A, N218928420748, N21892844627E, N2189284750E5, N218928733CA9, N218928766B04, N218928808DFA, N21892882AEE8, N21892910003E, N2189291332A5, N218929155793, N218929188C76, N218929407E86, N218929425D94, N2189294438A2, N21892948F2CE, N2189297054EE, N218929736675, N218929750343, N218929772051, N21893003ABD8, N21893006FC75, N218930191D5A, N21893020AAF3, N21893021BB7A, N218930239868, N21893007EDFC, N217949665E13, N217949674F9A, N21794968B76D, N21794969A6E4, N2179497022FD, N2179497201EF, N217949731066, N217949757550, N2179497647CB, N217949775642, N21794978AEB5, N21794979BF3C, N21794980A135, N21794981B0BC, N217949828227, N2179498393AE, N21794984E711, N21794985F698, N21794986C403, N21794987D58A, N217949893CF4, N21794990B8ED, N21794991A964, N217949938A76, N21794994FEC9, N21794995EF40, N21794996DDDB, N21794997CC52, N2179499834A5, N21794999252C, N217949713374, N2179497464D9, N217949882D7D, N217949929BFF, N21795001FED9, N21795002CC42, N21795003DDCB, N21795004A974, N22034419EC84, N220344205B2D, N220344214AA4, N22034422783F, N2203442369B6, N220344241D09, N220344250C80, N220344263E1B, N220344272F92, N22034428D765, N22034429C6EC, N2203443042F5, N22034431537C, N2203443261E7, N22034433706E, N2203443404D1, N220344351558, N2203443627C3, N22034437364A, N22034438CEBD, N22034439DF34, N220344411E74, N220344422CEF, N2203444449D9, N220344455850, N220344466ACB, N220344400FFD, N220344433D66, N220345003241, N2203450123C8, N220345021153, N2203450300DA, N220345047465, N2203450565EC, N220345065777, N2203450746FE, N22034508BE09, N22034509AF80, N220345102B99, N220345113A10, N22034512088B, N220345131902, N220345146DBD, N220345157C34, N220345164EAF, N22034518A7D1, N22034519B658, N2203452001F1, N220345211078, N2203452222E3, N22034523336A, N2203452447D5, N22034525565C, N220345175F26, N2203452664C7, N22034527754E, N220345288DB9, N220345299C30, N220345301829, N222265751280, N22226576201B, N222265773192, N22226578C965, N22226579D8EC, N22226580C6E5, N22226581D76C, N2222658480C1, N222265859148, N22226586A3D3, N22226587B25A, N222265884AAD, N222265895B24, N22226590DF3D, N22226591CEB4, N22226592FC2F, N22226593EDA6, N222265949919, N222265958890, N22226596BA0B, N22226597AB82, N2222659942FC, N22226601F6C8, N22226602C453, N22226603D5DA, N22226604A165, N22226605B0EC, N222266068277, N2222660793FE, N222266086B09, N222266097A80, N22226610FE99, N22226611EF10, N22226613CC02, N22226614B8BD, N222266169BAF, N222266178A26, N2222661872D1, N222266196358, N22226620D4F1, N225493051CE1, N225493062E7A, N225493073FF3, N22549308C704, N22549309D68D, N225493105294, N22549311431D, N225493127186, N22549313600F, N2254931414B0, N225493150539, N2254931637A2, N22549317262B, N22549318DEDC, N22549319CF55, N2254932078FC, N225493216975, N225493225BEE, N225493234A67, N225493243ED8, N225493252F51, N225493261DCA, N225493270C43, N22549328F4B4, N22549329E53D, N225493306124, N2254933170AD, N225493353689, N225493360412, N22549337159B, N22549338ED6C, N22549339FCE5, N225493402C2C, N225493413DA5, N225493420F3E, N225493431EB7, N225493446A08, N225493457B81, N22549346491A, N22226582E5F7, N22226615A934, N22226622F7E3, N22226623E66A, N2222662492D5, N22226625835C, N22226626B1C7, N222266294930, N22226630CD29, N22226631DCA0, N22226632EE3B, N22226633FFB2, N222266348B0D, N222266359A84, N22226636A81F, N22226637B996, N222266384161, N22226707C922, N2222670831D5, N22226709205C, N22226710A445, N22226711B5CC, N222267128757, N2222671396DE, N22226715F3E8, N22226716C173, N22226717D0FA, N22226718280D, N222267193984, N222267208E2D, N222267219FA4, N22226722AD3F, N22226723BCB6, N22226724C809, N22226725D980, N22226726EB1B, N22226727FA92, N222267280265, N2222672913EC, N2222673097F5, N22226731867C, N22226732B4E7, N22226733A56E, N22226734D1D1, N22226735C058, N22226736F2C3, N22226737E34A, N222267390A34, N22226740DAFD, N22226741CB74, N22226742F9EF, N22226743E866, N222267449CD9, N22226746BFCB, N22226747AE42, N2222674856B5, N2254933353BF, N22226583F47E, N222265985375, N22226600E741, N22226627A04E, N2222662858B9, N22226714E261, N222267381BBD, N222267458D50, N22226749473C, N22226750C325, N22226751D2AC, N22226752E037, N22226753F1BE, N222267548501, N222267559488, N22226756A613, N22226757B79A, N222267584F6D, N222267595EE4, N22226760E94D, N22226761F8C4, N22226762CA5F, N22226763DBD6, N22226764AF69, N225493324236, N225493342700, N219428299514, N21942830110D, N219428310084, N219428332396, N219428345729, N2194283546A0, N2194283765B2, N219428389D45, N219428398CCC, N219428414D8C, N219428427F17, N219428436E9E, N219428441A21, N219428450BA8, N219428463933, N2194284728BA, N21942848D04D, N21942849C1C4, N2194285045DD, N219428515454, N2194285266CF, N2194285403F9, N225594278B02, N2255942873F5, N22559429627C, N22559430E665, N22559431F7EC, N22559432C577, N22559433D4FE, N22559434A041, N22559435B1C8, N225594368353, N2255943792DA, N225594386A2D, N225594397BA4, N22559440AB6D, N22559441BAE4, N2255944399F6, N22559444ED49, N22559445FCC0, N22559446CE5B, N22559447DFD2, N225594482725, N22559450B2B5, N22559451A33C, N2255945291A7, N22559453802E, N22559454F491, N22559455E518, N22559456D783, N22559457C60A, N225594583EFD, N225594592F74, N2255946098DD, N225594618954, N219428537746, N2194285620EB, N21942858C995, N21942859D81C, N219428606FB5, N219428617E3C, N21942836743B, N22559442887F, N226593446D31, N226593457CB8, N226593475FAA, N22659348A75D, N2265935032CD, N2265935211DF, N226593530056, N2265935474E9, N226593556560, N226593574672, N22659358BE85, N22659359AF0C, N2265936018A5, N226593623BB7, N226593632A3E, N21942828849D, N219428405C05, N219428573162, N2265935657FB, N222271476420, N222271489CD7, N2222715118CE, N222271522A55, N222271533BDC, N222271544F63, N222271555EEA, N222271566C71, N222271577DF8, N22227158850F, N222271599486, N22227160232F, N2222716132A6, N22227162003D, N2222716311B4, N22227164650B, N222271657482, N222271664619, N222271675790, N22227168AF67, N22227169BEEE, N222271703AF7, N222271712B7E, N2222717219E5, N22227173086C, N222271756D5A, N222271765FC1, N222271774E48, N22227178B6BF, N22227179A736, N222271498D5E, N222271747CD3, N226597254B09, N226597267992, N22659727681B, N2265972890EC, N226597298165, N22659730057C, N2265973114F5, N22659732266E, N2265973337E7, N2265973552D1, N22659736604A, N2265973771C3, N2265973998BD, N226597404874, N2265974159FD, N226597426B66, N226597437AEF, N226597440E50, N222271500947, N227336228677, N2273362397FE, N22733624E341, N22733625F2C8, N22733626C053, N22733627D1DA, N22733628292D, N2273362938A4, N22733631AD34, N227336329FAF, N227336338E26, N22733634FA99, N22733635EB10, N22733636D98B, N22733637C802, N2273363830F5, N22733639217C, N22733640F1B5, N22733641E03C, N22733642D2A7, N22733643C32E, N22733644B791, N227336469483, N227336487DFD, N227336496C74, N22733650E86D, N22733651F9E4, N22733652CB7F, N22733653DAF6, N22733655BFC0, N227336568D5B, N226597388934, N22733630BCBD, N22733645A618, N22733654AE49, N22733647850A, N227874100FB5, N227874122CA7, N227874133D2E, N227874144991, N227874155818, N227874177B0A, N2278741883FD, N227874199274, N2278742025DD, N227874213454, N2278742206CF, N227874231746, N2278742463F9, N227874257270, N2278742640EB, N22787428A995, N22787429B81C, N227874303C05, N227874321F17, N227874330E9E, N227874347A21, N227874356BA8, N227874365933, N2278743748BA, N22787438B04D, N22787439A1C4, N22787440710D, N227874416084, N22787442521F, N227874434396, N227874443729, N2278744526A0, N22787446143B, N2278744705B2, N22787448FD45, N22787449ECCC, N2278745068D5, N227874524BC7, N227874560DE3, N227874571C6A, N22787458E49D, N2278746042BD, N227874615334, N2278746261AF, N227874637026, N227874640499, N227874651510, N22787466278B, N227874673602, N227874275162, N22787459F514, N227337833052, N2273378444ED, N227337855564, N2273378667FF, N227337877676, N227337888E81, N227337910A98, N22733793298A, N227337945D35, N227337954CBC, N227337967E27, N227337976FAE, N227337989759, N2273379986D0, N2273380086CE, N227338019747, N22733803B455, N22733805D163, N22733806E3F8, N22733807F271, N227338080A86, N227338091B0F, N227338109F16, N22733812BC04, N22733814D932, N22733815C8BB, N22733816FA20, N227337923803, N22733813AD8D, N22733817EBA9, N22733818135E, N2273381902D7, N22733820B57E, N22733821A4F7, N22733822966C, N22733824F35A, N22733825E2D3, N22733826D048, N227338283936, N2273382928BF, N22733830ACA6, N227338328FB4, N227338339E3D, N22733835FB0B, N22733836C990, N22733837D819, N2273383820EE, N227874535A4E, N227874553F78, N227874312D8C, N227337899F08, N227338118E9F, N2273382387E5, N22733827C1C1, N22733834EA82, N227874166A83, N227874542EF1, N22733804C0EA, N2281416300BB, N22814165658D, N22814169AFE1, N22814167469F, N22814168BE68, N228141713A71, N2281417208EA, N228141746DDC, N228141764ECE, N228141775F47, N22814178A7B0, N22814179B639, N22814180A830, N22814181B9B9, N228141828B22, N228141839AAB, N22814184EE14, N22814185FF9D, N22814186CD06, N22814187DC8F, N228141882478, N2281418935F1, N22814190B1E8, N22814191A061, N2281419292FA, N228141938373, N22814194F7CC, N22814195E645, N22814196D4DE, N22814197C557, N228141983DA0, N228141992C29, N228142008994, N228141665716, N228141702BF8, N22883770FB64, N22883772D876, N228837843E88, N22883788F4E4, N22883800FCAB, N22883801ED22, N228838078814, N22883810E573, N22883814A357, N2288381791CC, N22883818693B, N2288381978B2, N22883822EC09, N22883826AA2D, N22883827BBA4, N228838284353, N22883830D6C3, N22883833E458, N22883835816E, N215415823DF1, N228837924266, N22883802DFB9, N2288380870E3, N22883823FD80, N22883831C74A, N22883842B8D9, N22883845CC66, N228838481783, N22883851939A, N22883853B088, N22883856E725, N22883860A87B, N22883864EE5F, N22883865FFD6, N228838682433, N2288387292B1, N22883876D495, N228838792C62, N2288388300F0, N22883884744F, N22883888BE23, N228838957C1E, N228838975F0C, N22883901B7FE, N22883904E053, N22883907D2C8, N22883910BFAF, N228839138D34, N22883915E802, N22883917CB10, N22883925C26A, N22883928198F, N22883933BE84, N22883935DBB2, N22883936E929, N22883940C117, N228839448733, N22883946A421, N228839495CD6, N228839549EEB, N22883957AC70, N22883959450E, N22883962D1B5, N22883964B483, N228839669791, N228839687EEF, N22883970EB7F, N22883973D9E4, N22883974AD5B, N22883886575D, N22883914F98B, N2288386935BA, N22733831BD2F, N22883922B6D5, N22772168F3E3, N22772169E26A, N2277217177FA, N227721724561, N2277217354E8, N227721742057, N2277217531DE, N227721760345, N2277217712CC, N22772178EA3B, N22772179FBB2, N22772180E5BB, N22772181F432, N22772182C6A9, N22772183D720, N22772184A39F, N22772185B216, N22772186808D, N227721879104, N22772189787A, N22772190FC63, N22772191EDEA, N22772192DF71, N22772193CEF8, N22772194BA47, N22772195ABCE, N227721969955, N2277219788DC, N22772198702B, N2277219961A2, N22772200C41F, N22772201D596, N22772202E70D, N22772203F684, N230942042B1D, N230942053A94, N23094206080F, N227721706673, N2277218869F3, N229768729231, N2297687383B8, N22976874F707, N22976875E68E, N22976876D415, N22976877C59C, N229768783D6B, N229768792CE2, N2297688032EB, N229768812362, N2297688211F9, N229768830070, N2297688657DD, N229768874654, N22976888BEA3, N22976889AF2A, N229768902B33, N229768913ABA, N229768920821, N2297689319A8, N229768957C9E, N229768964E05, N229768975F8C, N22976899B6F2, N22976900A6F7, N22976901B77E, N2297690285E5, N22976903946C, N22976904E0D3, N22976905F15A, N22976906C3C1, N22976907D248, N229769082ABF, N229769093B36, N229768856546, N229768946D17, N22976898A77B, N2297688474CF, N231653880701, N231653891688, N231653909291, N231653918318, N23165392B183, N23165393A00A, N23165394D4B5, N23165395C53C, N23165396F7A7, N23165397E62E, N231653981ED9, N231653990F50, N23165400C98C, N23165401D805, N23165402EA9E, N23165403FB17, N231654048FA8, N231654059E21, N23165407BD33, N2316540845C4, N23165409544D, N23165410D054, N23165411C1DD, N23165412F346, N23165413E2CF, N231654149670, N2316541587F9, N23165416B562, N23165417A4EB, N231654185C1C, N231654194D95, N23165420FA3C, N23165422D92E, N23165423C8A7, N23165424BC18, N23165425AD91, N231654278E83, N231654287674, N2316542967FD, N23165430E3E4, N23165432C0F6, N23165433D17F, N23165434A5C0, N23165435B449, N2316543686D2, N23165437975B, N231654386FAC, N23165440AEEC, N23165441BF65, N231654428DFE, N231654439C77, N23165444E8C8, N23165445F941, N23165446CBDA, N232183295CD6, N23218330D8CF, N23218332FBDD, N23218333EA54, N232183358F62, N23218336BDF9, N23218337AC70, N232183385487, N23218339450E, N2321834095C7, N23218341844E, N23218342B6D5, N23218343A75C, N23218345C26A, N23218346F0F1, N23218347E178, N232183490806, N232183508C1F, N232183519D96, N23218352AF0D, N23218353BE84, N23218354CA3B, N23218355DBB2, N232183580057, N2321835911DE, N23218360A677, N23218361B7FE, N232183628565, N2321836394EC, N23218364E053, N23218365F1DA, N232183682A3F, N232183693BB6, N23218371AE26, N232183729CBD, N232183738D34, N23218376DA99, N23218377CB10, N2321837833E7, N23218379226E, N232183803C67, N232183812DEE, N232183830EFC, N232183847A43, N232183856BCA, N232183865951, N2321838748D8, N23218388B02F, N23218389A1A6, N232183913436, N2321839206AD, N232183931724, N23218394639B, N232183957212, N232183964089, N232183975100, N23218398A9F7, N23218399B87E, N232184007EA2, N232184016F2B, N232184025DB0, N23165406ACBA, N23165421EBB5, N231654269F0A, N231654397E25, N23218331C946, N232183349EEB, N23218344D3E3, N23218348198F, N23218357F8A0, N23218366C341, N23218367D2C8, N23218370BFAF, N23218374F98B, N23218375E802, N232183821F75, N232187496B67, N23218750EF7E, N23218751FEF7, N23218753DDE5, N23218754A95A, N232187579BC1, N2321875972BF, N23218760C516, N23218761D49F, N23218762E604, N23218763F78D, N232187648332, N23218766A020, N23218767B1A9, N2321876958D7, N23218770DCCE, N23218771CD47, N23218772FFDC, N23218773EE55, N232187749AEA, N232187758B63, N23218776B9F8, N23218777A871, N232187785086, N23218779410F, N232187805F06, N232187827C14, N232187836D9D, N232187841922, N2321878508AB, N232187872BB9, N23218788D34E, N23218789C2C7, N23218752CC6C, N232187586336, N232187568A48, N2321876592BB, N23218768495E, N232187814E8F, N232187863A30, N2321879046DE, N232187915757, N2321879265CC, N232187937445, N2321879400FA, N232187951173, N2321879623E8, N232187973261, N23218798CA96, N23218800DB01, N23218802F813, N23218803E99A, N232188049D25, N23218806BE37, N23218807AFBE, N2321880946C0, N23218810C2D9, N23218812E1CB, N23218813F042, N2321881484FD, N232188159574, N23218816A7EF, N23218817B666, N232188184E91, N23218820E8B1, N23218821F938, N23218822CBA3, N23218823DA2A, N23218824AE95, N23218825BF1C, N232188279C0E, N2321882864F9, N232188297570, N23218830F169, N23218831E0E0, N23218832D27B, N23218833C3F2, N23218834B74D, N23218835A6C4, N23218836945F, N2321883785D6, N232188387D21, N232188396CA8, N23218840BC61, N23218841ADE8, N232188429F73, N232188438EFA, N23218844FA45, N23218845EBCC, N23218846D957, N23218847C8DE, N2321884921A0, N23218799DB1F, N232188058CAC, N232188085749, N23218811D350, N232188195F18, N232188483029, N23218850A5B9, N23218851B430, N2321885286AB, N232188539722, N23218854E39D, N23218855F214, N23218856C08F, N23218857D106, N232188593878, N232188608FD1, N232188619E58, N23218862ACC3, N23218863BD4A, N23218865D87C, N23291870354A, N2329187124C3, N232918721658, N2329187307D1, N2329187562E7, N23291876507C, N23291878B902, N23291879A88B, N23291880B682, N23291881A70B, N232918829590, N232918838419, N23291884F0A6, N23291885E12F, N23291887C23D, N232918883ACA, N232918892B43, N23291890AF5A, N23291891BED3, N232918928C48, N232918939DC1, N23291894E97E, N23291895F8F7, N23291896CA6C, N2321885829F1, N23218864C9F5, N23291874736E, N23291886D3B4, N23291897DBE5, N232918982312, N23291899329B, N23291900229E, N232919013317, N23291902018C, N232919031005, N2329190464BA, N2329190647A8, N232919075621, N23291908AED6, N23291909BF5F, N232919103B46, N232919112ACF, N232919121854, N2329191309DD, N232919147D62, N232919156CEB, N232919165E70, N232919174FF9, N23291918B70E, N23291919A687, N23291920112E, N2329192100A7, N23291922323C, N2329192323B5, N23291924570A, N232919254683, N232919267418, N232919276591, N232919289D66, N232919298CEF, N2329193008F6, N23291931197F, N232919322BE4, N232919333A6D, N232919344ED2, N232919355F5B, N232919366DC0, N232919377C49, N2329193884BE, N232919399537, N2329194045FE, N232919415477, N2329194266EC, N232919437765, N2329194403DA, N232919451253, N2329194620C8, N232919473141, N23291948C9B6, N23291949D83F, N232919505C26, N232919514DAF, N232919527F34, N232919536EBD, N232919541A02, N232919550B8B, N232919563910, N232919572899, N234324192867, N234324209FCE, N23432422BCDC, N23432423AD55, N23432424D9EA, N23432425C863, N23432427EB71, N234324281386, N23432429020F, N234324308616, N23432431979F, N23432432A504, N23432433B48D, N23432434C032, N23432435D1BB, N23432436E320, N23432437F2A9, N234324380A5E, N234324391BD7, N23432440CB1E, N23432441DA97, N23432442E80C, N23432443F985, N234324448D3A, N23432446AE28, N23432447BFA1, N234324484756, N2343244956DF, N23432450D2C6, N23432451C34F, N23432452F1D4, N23432453E05D, N23432455856B, N23432456B7F0, N23432457A679, N234324585E8E, N234324594F07, N23432460F8AE, N23432461E927, N23432462DBBC, N23432463CA35, N23432465AF03, N234324669D98, N234324678C11, N2343246874E6, N23432469656F, N23432470E176, N23432471F0FF, N2343245494E2, N2329187741F5, N2343241839EE, N234324218E47, N23432426FAF8, N234324459CB3, N23432464BE8A, N23432472C264, N237682649ED4, N237682658F5D, N23768266BDC6, N23768267AC4F, N2376826854B8, N237682694531, N23768270C128, N23768271D0A1, N23768272E23A, N23768273F3B3, N23768274870C, N237682759685, N23768276A41E, N23768277B597, N237682784D60, N237682795CE9, N2376828042E0, N237682815369, N2376828261F2, N23768283707B, N2376828404C4, N23768285154D, N2376828627D6, N23768287365F, N23768288CEA8, N23768289DF21, N237682905B38, N237682914AB1, N23768292782A, N2376829369A3, N237682941D1C, N237682950C95, N237682963E0E, N237682972F87, N23768298D770, N23768299C6F9, N23768300D6FC, N23768301C775, N23768302F5EE, N23768345E631, N23768346D4AA, N23768347C523, N237683483DD4, N237683492C5D, N23768350A844, N23768351B9CD, N237683528B56, N237683539ADF, N23768354EE60, N23768355FFE9, N23768356CD72, N23768357DCFB, N23768358240C, N237683593585, N23768360822C, N2376836193A5, N23768362A13E, N23768363B0B7, N23768364C408, N23768365D581, N237683680E64, N237683691FED, N237683709BF4, N237683718A7D, N23768373A96F, N23768376FEC2, N23768377EF4B, N2376837817BC, N237683790635, N2376838109B5, N237683823B2E, N237683832AA7, N237683867D0A, N237683876C83, N237683889474, N23768367F693, N23768380183C, N237683845E18, N2376838985FD, N2376839001E4, N23768391106D, N2376839222F6, N23768393337F, N2376839447C0, N237683955649, N23768397755B, N237683988DAC, N237683999C25, N237684005AF9, N2376840279EB, N234328101005, N23432811018C, N234328123317, N23432813229E, N234328145621, N2343281547A8, N234328167533, N2343281764BA, N234328189C4D, N234328198DC4, N234328203A6D, N234328212BE4, N23432822197F, N2343282308F6, N234328247C49, N234328256DC0, N234328265F5B, N234328274ED2, N23432828B625, N23432829A7AC, N2343283023B5, N23432831323C, N2343283200A7, N23432833112E, N234328346591, N23432837570A, N23432838AFFD, N23432839BE74, N23768303E467, N23768372B8E6, N2376839664D2, N237684014B70, N23899345ADFF, N238993469F64, N238993478EED, N238993496793, N23899350E38A, N23899351F203, N23899352C098, N23899353D111, N23899354A5AE, N2389935686BC, N238993586FC2, N23899360C9E2, N23899362EAF0, N23899363FB79, N238993648FC6, N238993659E4F, N23899366ACD4, N2389936845AA, N238993695423, N23899370D03A, N23899371C1B3, N23899372F328, N23899373E2A1, N23899374961E, N238993758797, N23899376B50C, N2389938270E0, N238993836169, N23899385045F, N2389938636C4, N23899387274D, N23899388DFBA, N23899389CE33, N23768375CC59, N234328364683, N23563074FAF7, N23563075EB7E, N23563076D9E5, N23563077C86C, N23563078309B, N235630792112, N235630803F1B, N235630812E92, N235630821C09, N235630830D80, N23563084793F, N2356308568B6, N235630865A2D, N235630874BA4, N23563088B353, N23563089A2DA, N23563091374A, N2356309205D1, N235630931458, N2356309460E7, N2356309643F5, N23563097527C, N23563098AA8B, N23563099BB02, N23563101BA8E, N235631028815, N23563103999C, N23563104ED23, N23563105FCAA, N23563106CE31, N23563107DFB8, N23563108274F, N2356310936C6, N23563110B2DF, N23563111A356, N2356311291CD, N23563114F4FB, N23563115E572, N23563116D7E9, N23563117C660, N235631183E97, N235631192F1E, N2356312098B7, N23563122BBA5, N23563123AA2C, N23563124DE93, N23563126FD81, N23563127EC08, N2356312814FF, N235631290576, N23563130816F, N2356313190E6, N23563132A27D, N23563133B3F4, N23563134C74B, N237683854F91, N23899367BD5D, N23563135D6C2, N23563136E459, N23563137F5D0, N235631380D27, N235631391CAE, N23563140CC67, N23563141DDEE, N23563142EF75, N23563143FEFC, N235631448A43, N235631459BCA, N23563146A951, N23563147B8D8, N23563148402F, N2356314951A6, N23563151C436, N23563152F6AD, N23563153E724, N23563154939B, N235631558212, N23563156B089, N235630738E48, N23649814AEBE, N236498168DAC, N236498179C25, N2364981864D2, N23649819755B, N23649821D37B, N23649822E1E0, N23649823F069, N23649825955F, N23649827B64D, N236498284EBA, N236498295F33, N23649831CAA3, N23649832F838, N23649833E9B1, N236498358C87, N23649836BE1C, N23649837AF95, N236498385762, N2364983946EB, N236498409622, N2364984187AB, N23649842B530, N238993794DFB, N23649815BF37, N23649844D006, N23649845C18F, N23649846F314, N23649847E29D, N236498481A6A, N236498490BE3, N236498508FFA, N236498519E73, N23649852ACE8, N23649853BD61, N23649854C9DE, N23649855D857, N23649856EACC, N23649857FB45, N23649859123B, N23649860A592, N23649861B41B, N236498628680, N236498639709, N23649864E3B6, N23649865F23F, N23649866C0A4, N23649867D12D, N2364986829DA, N236498693853, N23649870BC4A, N23649871ADC3, N236498738ED1, N23649874FA6E, N23649875EBE7, N23649876D97C, N23649877C8F5, N236498783002, N23649879218B, N236498803F82, N236498812E0B, N236498821C90, N236498830D19, N2364988479A6, N23649885682F, N236498865AB4, N236498874B3D, N23649888B3CA, N23649889A243, N23649890265A, N236498920548, N2364989314C1, N23899361D86B, N2389938415D6, N2356309026C3, N23563150D5BF, N2364989137D3, N23563100AB07, N23649820C2F2, N23649826A7C4, N236498349D0E, N2364985803B2, N238993597E4B, N238993785C72, N23563121893E, N23649830DB2A, N236498729F58, N235630729FC1, N23899348761A, N239001181136, N23900125E0BB, N23900126D220, N23900127C3A9, N239001283B5E, N239001292AD7, N23900130AECE, N23900131BF47, N239001328DDC, N239001339C55, N23900134E8EA, N23900135F963, N23900136CBF8, N23900137DA71, N239001382286, N23900140E3C6, N23900141F24F, N23900142C0D4, N23900144A5E2, N23900145B46B, N2390014686F0, N239001479779, N239001486F8E, N239001497E07, N23900150FA1E, N23900151EB97, N23900152D90C, N23900153C885, N23900154BC3A, N23900155ADB3, N239001578EA1, N239001587656, N23900160D076, N23900161C1FF, N23900162F364, N23900163E2ED, N239001649652, N2390016587DB, N23900167A4C9, N239001685C3E, N239001694DB7, N23900170C9AE, N23900171D827, N23900172EABC, N23900173FB35, N239001748F8A, N239001759E03, N24107161709E, N241071624205, N241071660421, N241071752F62, N23900139330F, N24107169FCD6, N241071725BDD, N241071761DF9, N24107178F487, N24107179E50E, N24107181EA8E, N24107182D815, N24107183C99C, N241071869E31, N24107188774F, N2410718966C6, N24107191F356, N24107193D044, N2410719687E9, N24107200DAA3, N24107201CB2A, N24342206F484, N243422090C73, N24342210886A, N24342213BAF1, N24342215DFC7, N24342216ED5C, N2434221915AB, N24342220A202, N243422228110, N24342224E426, N24342227D6BD, N243422293FC3, N243422338941, N24342234FDFE, N243422383792, N24342240F6D2, N24342242D5C0, N24342244B0F6, N243422487A9A, N24342250EF0A, N24342251FE83, N24342253DD91, N243422579BB5, N243422586342, N24342261D4EB, N243422648346, N24342268492A, N24342270DCBA, N24342273EE21, N24342276B98C, N23900143D15D, N2390015967DF, N23900166B540, N24107163538C, N24107184BD23, N24342236DEEC, N24342266A054, N24107180FB07, N24107203E838, N24354643E7F6, N243546449349, N2435464582C0, N24354646B05B, N24354647A1D2, N243546485925, N2435464948AC, N24354650CCB5, N24354651DD3C, N24354652EFA7, N24354653FE2E, N243546548A91, N243546559B18, N24354656A983, N24354657B80A, N2435465840FD, N243546595174, N24354660E6DD, N24354661F754, N24354662C5CF, N24354663D446, N24354665B170, N2435466683EB, N243546679262, N243546686A95, N243546697B1C, N24354670FF05, N24354671EE8C, N24354672DC17, N24354674B921, N24354675A8A8, N243546778BBA, N24354678734D, N2435467962C4, N243546807CCD, N243546816D44, N243546825FDF, N243546843AE9, N243546852B60, N2435468619FB, N243546870872, N24354688F085, N24354689E10C, N24354691749C, N2435469532B8, N2430974575C7, N24309746475C, N2430975218A0, N243097530929, N24309758B7FA, N2430976011DA, N2430976457FE, N243097689D92, N24309771198B, N24309838A9ED, N243098417924, N243098471C12, N24309849F56C, N2430985526D8, N2430985705CA, N24309859ECB4, N243098605B1D, N243098641D39, N243098663E2B, N24309873705E, N24309877367A, N24309882E21F, N24309887B5B2, N243098895CCC, N24309893EA4E, N243098949EF1, N24309898549D, N243098994514, N243546924607, N24354693578E, N24354703DA4A, N243547097510, N24354710F109, N24354711E080, N24354712D21B, N24354713C392, N24354714B72D, N24354716943F, N243547187D41, N24309778844A, N2430977995C3, N24309782A8D8, N24309783B951, N24309789160B, N24309796F724, N243097981E5A, N243098022CDF, N243098033D56, N243098088385, N24309809920C, N2430981541B8, N243098189A5D, N243098203C7D, N243098221F6F, N24309826594B, N24309829A1BC, N243098357208, N239001569F28, N24354664A0F9, N24354673CD9E, N243546834E56, N243546906515, N243546942331, N243547079C6E, N243547086499, N24309813248E, N243431433796, N243431444329, N2434314552A0, N24343146603B, N2434314771B2, N243431488945, N2434314998CC, N243431510D5C, N243431523FC7, N243431532E4E, N243431554B78, N24343157686A, N24343158909D, N2434316036BD, N243431612734, N2434316215AF, N243431647099, N243431656110, N24343166538B, N24343168BAF5, N24343169AB7C, N243431702F65, N243431713EEC, N243431720C77, N2434317578C8, N243431764A53, N243431775BDA, N24343178A32D, N24343179B2A4, N24343180ACAD, N24343181BD24, N243431828FBF, N2430976232C8, N24309791839B, N24309871534C, N243098848729, N243431839E36, N24343184EA89, N24343185FB00, N24343186C99B, N24343187D812, N2434318820E5, N24343190B575, N24343191A4FC, N2434319387EE, N2434315679E3, N243431598114, N24343194F351, N24343196D043, N24343197C1CA, N24343198393D, N2434319928B4, N243432008D09, N243432019C80, N24343202AE1B, N24343203BF92, N246282042427, N246282060735, N2462820716BC, N246282107BDB, N246282116A52, N2462821258C9, N246282134940, N246282143DFF, N246282161EED, N246282170F64, N24628218F793, N24628219E61A, N2462822051B3, N24628221403A, N2462822272A1, N243431545AF1, N24343189316C, N24343195E2D8, N2462820535AE, N24628208EE4B, N243431731DFE, N243431929667, N24628209FFC2, N247147085925, N24714716A983, N247147195174, N247147297B1C, N24714733CD9E, N247147369A33, N247147378BBA, N247147438096, N24714744F429, N24714753994E, N24714754EDF1, N2471476081BD, N24714766E48B, N247147691C7C, N24714773AAFE, N24714778142D, N2471477905A4, N247147845D89, N2471478897E5, N24714789866C, N247147944451, N2471479555D8, N247147988E3D, N24714803AD31, N24714730FF05, N24714746D73B, N24714758279D, N2471479113FC, N24715039F27A, N24715048AEFB, N24715049BF72, N247150503B6B, N247150512AE2, N247150521879, N247150547D4F, N247150556CC6, N247150565E5D, N247150574FD4, N24715058B723, N24715059A6AA, N247150601103, N24715061008A, N247150623211, N247150632398, N247150645727, N2471506546AE, N247150667435, N2471506765BC, N247150689D4B, N247150698CC2, N2471507008DB, N247150711952, N246283896BDE, N24628390EFC7, N24628391FE4E, N24628392CCD5, N24628393DD5C, N24628394A9E3, N24628395B86A, N246283968AF1, N24628398638F, N246283997206, N24628400B4DA, N24628401A553, N2462840297C8, N246284038641, N24628404F2FE, N24628405E377, N24628406D1EC, N24628407C065, N24628409291B, N24628410AD02, N24628411BC8B, N246284128E10, N246284139F99, N24628414EB26, N24628415FAAF, N24628416C834, N24628417D9BD, N24628418214A, N2462841930C3, N2462842196E3, N24628422A478, N24628423B5F1, N24628424C14E, N24628425D0C7, N24628426E25C, N24628427F3D5, N246283979B78, N24628420876A, N246284280B22, N2478486907DF, N2478487083C6, N24784871924F, N24784872A0D4, N24784873B15D, N24784874C5E2, N24784875D46B, N24784876E6F0, N24911230914D, N249112390C8C, N24911243EEDE, N249112449A61, N249112559230, N24911256A0AB, N24911259585C, N2491126863BD, N24911271E7A4, N24911274B009, N24911276931B, N249112787A65, N246284083892, N246286552877, N246286561AEC, N246286570B65, N24628658F392, N24628659E21B, N2462866055B2, N24628661443B, N2462866276A0, N246286636729, N246286641396, N24628665021F, N246286663084, N24628667210D, N24628668D9FA, N24628669C873, N246286704C6A, N246286715DE3, N246286726F78, N246286737EF1, N246286740A4E, N246286751BC7, N24628676295C, N2462867738D5, N24628678C022, N24628679D1AB, N24628680CFA2, N24628681DE2B, N24628682ECB0, N24628683FD39, N246286848986, N24628685980F, N24628686AA94, N24628687BB1D, N2462868843EA, N246286895263, N24628690D67A, N24628691C7F3, N24628692F568, N24628693E4E1, N24628694905E, N24628696B34C, N24628697A2C5, N246286994BBB, N246287005BBE, N246287014A37, N2462870278AC, N246287036925, N246287041D9A, N246287050C13, N246287063E88, N246287072F01, N24628708D7F6, N24628709C67F, N24784877F779, N247848780F8E, N247848791E07, N24784880000E, N247848811187, N24784882231C, N247848833295, N24784884462A, N2478488557A3, N247848866538, N2478488774B1, N247848888C46, N247848899DCF, N2478489019D6, N24784891085F, N247848923AC4, N247848932B4D, N247848945FF2, N247848954E7B, N247848967CE0, N247848976D69, N24784898959E, N247848998417, N247849009412, N24784901859B, N24784902B700, N24784903A689, N249112494184, N249112796BEC, N2462869581D7, N244640983E99, N244640992F10, N244641003F15, N244641012E9C, N244641021C07, N244641030D8E, N244641047931, N2446410568B8, N244641065A23, N244641074BAA, N24464108B35D, N24464109A2D4, N2446411026CD, N244641113744, N2446411205DF, N244641131456, N2446411460E9, N244641157160, N2446411643FB, N244641175272, N24464118AA85, N24464119BB0C, N244641200CA5, N244641211D2C, N244641222FB7, N244641233E3E, N244641244A81, N244641255B08, N244641266993, N24464127781A, N2446412880ED, N244641299164, N24464130157D, N2446413104F4, N24464132366F, N2446413327E6, N244641345359, N246287271CB1, N24628728E446, N24628729F5CF, N2462873071D6, N24628731605F, N2462873252C4, N24628733434D, N2462873437F2, N24628735267B, N243190681AD3, N243190690B5A, N243190719ECA, N24319072AC51, N24319073BDD8, N24319075D8EE, N24319076EA75, N24319077FBFC, N24319078030B, N243190791282, N243190800C8B, N243190833E10, N243190844AAF, N243190855B26, N2431908669BD, N2431908880C3, N2431909104DA, N243190923641, N243190945377, N2431909542FE, N243190967065, N24319098991B, N243190998892, N243191009897, N24319101891E, N24319102BB85, N249733509DAF, N249733518C26, N24973354DB8B, N24973355CA02, N24973356F899, N24973359006E, N24973360B7C7, N24973363855C, N24973364F1E3, N24973366D2F1, N249733692A06, N249733728D0D, N249733739C84, N24973376CB29, N24973377DAA0, N249733831F4C, N249733857A7A, N2497338648E1, N24973388A19F, N24973390340F, N2497339740B0, N24973399A9CE, N249734006F12, N249734060A24, N249734071BAD, N24973408E35A, N2497341076CA, N2497341613FC, N249734170275, N243190708F43, N24319074C967, N243190811D02, N243190822F99, N243190877834, N24973418FA82, N249734214D2B, N249734227FB0, N249734236E39, N249734263994, N2497343154F3, N2497343377E1, N24973436204C, N2497343731C5, N250813302053, N2508133131DA, N250813320341, N2508133312C8, N250813346677, N2508133577FE, N250813364565, N2508133754EC, N25081338AC1B, N25081339BD92, N250813406D5B, N250813417CD2, N250813424E49, N250813435FC0, N250813442B7F, N250813453AF6, N2508134719E4, N25081348E113, N25081349F09A, N250813507483, N25081351650A, N250813525791, N250813534618, N2508135432A7, N25081355232E, N2508135611B5, N25081358F8CB, N25081359E942, N250813605EEB, N250813614F62, N250813636C70, N2508136418CF, N250813650946, N250813663BDD, N250813672A54, N25081368D2A3, N25081369C32A, N2508137156BA, N250813726421, N2508137375A8, N250813740117, N25081375109E, N25081378CB7B, N25081379DAF2, N25081380C4FB, N25081381D572, N25081382E7E9, N25081383F660, N25081386A1CD, N25081387B044, N2508138848B3, N25081389593A, N249733965139, N2431918364CC, N243191841073, N2431918501FA, N243191863361, N2431918722E8, N24319188DA1F, N24319189CB96, N243191904F8F, N243191915E06, N243191926C9D, N243191937D14, N2431919409AB, N243191951822, N243191973B30, N24319199D24E, N2431920077F3, N24319201667A, N2431920254E1, N243192034568, N2514620432F9, N251462052370, N2514620611EB, N251462070062, N25146209E91C, N251462106D05, N251462117C8C, N251462124E17, N251462135F9E, N251462153AA8, N251462160833, N2514621719BA, N25146218E14D, N25146219F0C4, N25146220476D, N2514622156E4, N25146222647F, N2514622375F6, N251462240149, N25146226225B, N2514622733D2, N25146229DAAC, N251462305EB5, N251462314F3C, N251462341891, N251462350918, N251462372A0A, N25146238D2FD, N25146239C374, N2514624013BD, N251462410234, N251462432126, N251462445599, N251462454410, N25146246768B, N251462489FF5, N251462498E7C, N251462500A65, N251462511BEC, N251462522977, N251462544C41, N251462566F53, N251462577EDA, N243191962AB9, N25081346086D, N25081377338C, N250813859356, N250813627DF9, N2508138482DF, N251462142B21, N24319198C3C7, N246286985A32, N246891955DC1, N246891977ED3, N2468919997AD, N246892003210, N246892012399, N246892021102, N24689203008B, N249734152167, N25081357003C, N250813704733, N2514622510C0, N25146228CB25, N251462327DA7, N251462336C2E, N2514624230AF, N2514625338FE, N251462555DC8, N252362040E58, N252362051FD1, N252362062D4A, N252362073CC3, N25236208C434, N2523621051A4, N25236211402D, N2523621272B6, N25236213633F, N252362141780, N252362150609, N252362163492, N25236217251B, N25236218DDEC, N25236219CC65, N252362207BCC, N252362216A45, N246891966F5A, N246891988624, N251468985CF7, N251468994D7E, N251469005D7B, N251469014CF2, N251469027E69, N251469036FE0, N251469041B5F, N251469050AD6, N25146906384D, N2514690729C4, N25146908D133, N25146909C0BA, N2514691044A3, N25146911552A, N2514691267B1, N251469137638, N251469140287, N25146915130E, N251469162195, N25146917301C, N25146919D962, N251469206ECB, N251469217F42, N251469224DD9, N251469235C50, N2514692428EF, N251469253966, N251469271A74, N25146928E283, N25146929F30A, N251469307713, N25146931669A, N251469325401, N251469334588, N251469343137, N2514693520BE, N251469361225, N2514693703AC, N25146938FB5B, N25146939EAD2, N251469403A1B, N251469412B92, N251469421909, N251469430880, N251469447C3F, N251469456DB6, N251469465F2D, N251469474EA4, N25146948B653, N25146918C8EB, N251469260BFD, N25148068B7BF, N25148069A636, N25148070222F, N25148074640B, N251480757582, N251480775690, N25148086C4D1, N25148087D558, N251480882DAF, N25148091A9B6, N25148094FE1B, N25148095EF92, N2514810035FB, N251481012472, N2514810216E9, N251481074144, N25148108B9B3, N251481102C23, N251481131EB8, N251481146A07, N251481157B8E, N25148120064B, N2514812334D0, N25148124406F, N251481299B8A, N251481323C81, N2514813459B7, N25148140529B, N251481436000, N2514814414BF, N251481450536, N2514814637AD, N251481472624, N25148148DED3, N25148149CF5A, N251481504B43, N251481515ACA, N251481526851, N2514815379D8, N251481540D67, N2534232530A6, N25342326023D, N2534232713B4, N25342328EB43, N25342329FACA, N253423316F5A, N253423325DC1, N2534233438F7, N25342335297E, N253423361BE5, N253423370A6C, N25342338F29B, N2534234033DB, N2534234210C9, N2534234475FF, N253423456476, N2534234656ED, N253423474764, N25342349AE1A, N253423502A03, N253423513B8A, N253423520911, N253423531898, N253423546C27, N253423557DAE, N253423564F35, N253423575EBC, N25342358A64B, N25342359B7C2, N25342360006B, N2534236111E2, N25342364464F, N25062961786D, N250629624AF6, N250629635B7F, N250629642FC0, N250629653E49, N250629660CD2, N250629671D5B, N25062968E5AC, N25062969F425, N25062970703C, N2506297161B5, N2506297342A7, N250629743618, N250629752791, N25062976150A, N250629770483, N25062978FC74, N25062979EDFD, N25062980F3F4, N25062981E27D, N25062982D0E6, N25062984B5D0, N25062985A459, N250629887FBC, N250629896E35, N25062990EA2C, N25062991FBA5, N25062992C93E, N25062993D8B7, N25062994AC08, N25062995BD81, N250629968F1A, N250629979E93, N250629986664, N2506299977ED, N25063000BD91, N25063001AC18, N250630029E83, N250630038F0A, N25063004FBB5, N25063005EA3C, N25063006D8A7, N2506300831D9, N25063010A449, N25063011B5C0, N25063012875B, N2506301396D2, N25063014E26D, N25063015F3E4, N25063016C17F, N25063017D0F6, N250630182801, N250630193988, N250630208E21, N250630219FA8, N25063022AD33, N25063023BCBA, N25063024C805, N25063026EB17, N25063027FA9E, N2506303097F9, N250630318670, N25063032B4EB, N25063033A562, N25063034D1DD, N25063035C054, N25063036F2CF, N25063037E346, N250630381BB1, N250630390A38, N25063040DAF1, N25062983C16F, N25063025D98C, N250630280269, N25342348BF93, N255855182F19, N255855193E90, N255855208939, N2558552198B0, N25585522AA2B, N25585523BBA2, N25585524CF1D, N25585525DE94, N25585526EC0F, N25585527FD86, N255855280571, N2558552914F8, N2558553090E1, N255855318168, N25585532B3F3, N25585534D6C5, N25585535C74C, N25585536F5D7, N25585537E45E, N255855381CA9, N25585540DDE9, N25585541CC60, N25585543EF72, N255855449BCD, N255855458A44, N25585546B8DF, N25585547A956, N2558554851A1, N255855494028, N25585550C431, N25585551D5B8, N25585552E723, N25585553F6AA, N25585556A107, N25342654559A, N253426554413, N253426576701, N253426589FF6, N253426598E7F, N25342661285F, N253426621AC4, N253426630B4D, N253426647FF2, N253426656E7B, N253426665CE0, N253426674D69, N25342668B59E, N25342669A417, N25342670200E, N25342672031C, N253426731295, N25342674662A, N2534267577A3, N253426764538, N2534267754B1, N25342678AC46, N25342679BDCF, N25342680A3C6, N25342681B24F, N2534268280D4, N25342683915D, N25342684E5E2, N25342685F46B, N25342686C6F0, N25342687D779, N253426882F8E, N25585557B08E, N255855584879, N2558555959F0, N25585560EE59, N25585561FFD0, N25585562CD4B, N25585563DCC2, N25585564A87D, N25585565B9F4, N255855668B6F, N255855679AE6, N255855697398, N25585570F781, N25585571E608, N25585572D493, N25585573C51A, N25585574B1A5, N253426893E07, N25342692990C, N253426938885, N25342694FC3A, N25342695EDB3, N25342696DF28, N25342697CEA1, N253426983656, N2534270037DA, N253427012653, N2534270214C8, N253427030541, N2534270652EC, N253427074365, N25342708BB92, N250630092050, N2506302913E0, N253423622379, N2534287567B8, N253428765523, N2534287744AA, N25342878BC5D, N25342879ADD4, N25342880B3DD, N25342881A254, N253428838146, N25342884F5F9, N25342885E470, N25342886D6EB, N25342887C762, N253428883F95, N253428892E1C, N25342894EC21, N25342895FDA8, N25342898264D, N25342909BA00, N253429103E19, N25342918B251, N2534292326EA, N253429267147, N253429322EBB, N253429355A04, N2534293881E1, N2534294263B3, N253429440685, N253429462597, N25342947341E, N2534295148F0, N253429527A6B, N253429536BE2, N25342958D531, N25342959C4B8, N253429616298, N253429661627, N2534296707AE, N25342968FF59, N253429735852, N253429753D64, N253429771E76, N25342979F708, N25342981F888, N25342982CA13, N25342991E150, N25342992D3CB, N25342993C242, N25342994B6FD, N253429978466, N253429996D18, N25343005F0C9, N25343006C252, N25343007D3DB, N253430129DAE, N253430138C27, N25343014F898, N25343017CA03, N25343021855D, N25343022B7C6, N25343027E06B, N25343028189C, N253430290915, N25343035DAA1, N25343037F9B3, N2534304597A9, N253430495DC5, N25343050D9DC, N25343051C855, N25343052FACE, N25343053EB47, N253430549FF8, N253430558E71, N25343060F3B4, N25343063C12F, N25343064B590, N25343070EA6C, N25343072C97E, N25343073D8F7, N2534307977AD, N25343081782D, N253430824AB6, N253430835B3F, N2562352193FF, N25623522A164, N25623523B0ED, N25623524C452, N25623526E740, N256235280E3E, N256235309BAE, N256235318A27, N25623532B8BC, N25623533A935, N25623536FE98, N25623537EF11, N25623539066F, N25623540D6A6, N25623541C72F, N25623542F5B4, N25623546B390, N25623545810B, N25623547A219, N256235485AEE, N256235494B67, N25623550CF7E, N25623551DEF7, N25623553FDE5, N25623554895A, N2562355598D3, N25623557BBC1, N256235584336, N2562355952BF, N25623560E516, N25623561F49F, N25623562C604, N25623563D78D, N25623564A332, N25623565B2BB, N256235668020, N2562356791A9, N25623568695E, N2562356978D7, N25623570FCCE, N256240854637, N256240876525, N256240898C5B, N256240900842, N256240922B50, N256240933AD9, N256240944E66, N256240955FEF, N256240966D74, N256240977CFD, N25624098840A, N256240999583, N256241008586, N25624101940F, N25624102A694, N261547718BBD, N26154772B926, N26154773A8AF, N26154774DC10, N26154775CD99, N26154776FF02, N26154777EE8B, N26154778167C, N2615477907F5, N2615478019FC, N26407280375F, N2640728126D6, N26407282144D, N2640728305C4, N26407284717B, N2640728560F2, N26407332AA39, N26407333BBB0, N26407334CF0F, N26407337FD94, N2640733914EA, N26407340C423, N264073448207, N26407345938E, N26407350DDFB, N26407351CC72, N26407356B8CD, N2640735851B3, N26407360F793, N26407361E61A, N26438359EF0F, N2643836058A6, N26438361492F, N264383627BB4, N264383636A3D, N264383650F0B, N264383663D90, N264383672C19, N26438368D4EE, N26438370417E, N2643837150F7, N260661300B1C, N260661311A95, N260661333987, N260661355CB1, N260661366E2A, N260661388754, N2606613996DD, N260661404614, N26066141579D, N260661426506, N26066143748F, N260661440030, N2606614511B9, N2606614732AB, N26066148CA5C, N260661505FCC, N260661514E45, N260661527CDE, N260661536D57, N2606615419E8, N260661550861, N260661563AFA, N260661572B73, N26066158D384, N26066159C20D, N2606616075A4, N26066161642D, N2606616256B6, N26066163473F, N260661652209, N260661661092, N26066167011B, N26066168F9EC, N264072865269, N2640728743E0, N26407288BB17, N26407289AA9E, N264072902E87, N264072913F0E, N264072931C1C, N26407295792A, N264072964BB1, N26407307D7FC, N26407314FCBF, N26407315ED36, N26407316DFAD, N26407317CE24, N2640732090F3, N26407321817A, N26407322B3E1, N26407324D6D7, N26407325C75E, N260661289E8C, N260661643380, N26066169E865, N260661706C7C, N260661717DF5, N260661724F6E, N260661735EE7, N260661742A58, N260661753BD1, N26066176094A, N2606617718C3, N26066178E034, N26066179F1BD, N26066180EFB4, N26066181FE3D, N26066184A990, N26066185B819, N260661868A82, N260661879B0B, N2606618863FC, N260661897275, N26066190F66C, N26066191E7E5, N26066192D57E, N26066193C4F7, N26066194B048, N26066195A1C1, N26066196935A, N2606619782D3, N260661987A24, N260661996BAD, N26066200CE10, N26066201DF99, N26066202ED02, N26066203FC8B, N2640729468A3, N264072975A38, N26407298A2CF, N26407299B346, N26407300A343, N26407301B2CA, N264073028051, N2640730391D8, N26407306C675, N264073093E82, N26407310BA9B, N264073129989, N264073138800, N2640731836D3, N26407323A268, N264073281CBB, N264073290D32, N2640733198A2, N264072792956, N26407304E567, N26407305F4EE, N26407326F5C5, N26407327E44C, N26407362D481, N264383471FA9, N26438348E75E, N26438349F6D7, N2643835072CE, N264383516347, N2643835251DC, N264383534055, N2643835434EA, N264383552563, N2643835617F8, N264383570671, N263075304323, N263075326031, N2630753371B8, N263075340507, N26307535148E, N263075362615, N26307537379C, N26307538CF6B, N26307539DEE2, N263075411FA2, N26307544480F, N263075466B1D, N263075488263, N26307551067A, N2630755234E1, N2630755451D7, N26307557634C, N263075612C12, N263075621E89, N263075647BBF, N2630756658AD, N2630757135CA, N263075720751, N263075746267, N26307580A78B, N263075828499, N263075839510, N263075882BC3, N26307590BE53, N263075929D41, N26307597CAEC, N26307598321B, N263075992392, N26307603B4B4, N26307605D182, N26307607F290, N263076080A67, N26438783934C, N26438785F67A, N26438786C4E1, N26438787D568, N264387882D9F, N26438790B80F, N26438796DD39, N264387983447, N26438803174B, N26438805727D, N2643880640E6, N264388112D81, N264388130E93, N264388156BA5, N264388201660, N2643882107E9, N2643882324FB, N264388245044, N2643882541CD, N2643882762DF, N264388333D23, N2643883883F0, N264388399279, N264388355815, N26438843702B, N264388462786, N264388505B68, N26438852787A, N264388541D4C, N26438859C6A9, N264388616089, N26438863439B, N26438869ECC1, N2643887068D8, N264388724BCA, N264388771C67, N26438878E490, N26438884AD34, N26438885BCBD, N264388879FAF, N26438893C053, N26438895A565, N2643889697FE, N264389007F0C, N264389016E85, N264389043928, N264389070BB3, N26438909E2CD, N2643891066D4, N26438913544F, N2643891420F0, N264389153179, N26438919FB15, N264389204CBC, N264389273803, N26438929D17D, N264389305564, N2643893367FF, N264389341340, N2643893502C9, N26438939C8A5, N26438940186C, N264389454FC1, N264389467D5A, N26438951103D, N26438953332F, N264389555619, N264389599C75, N264389602BDC, N2643896208CE, N264389646DF8, N264389664EEA, N264389675F63, N26438969B61D, N26438971238D, N26438973009F, N25623534DD8A, N260661462322, N26066149DBD5, N264072920D95, N264073082F0B, N264383641E82, N26438845151D, N26600481AF78, N266004829DE3, N26600484F8D5, N26600485E95C, N26600486DBC7, N26600487CA4E, N2660048832B9, N266004892330, N26600490A729, N26600491B6A0, N26600492843B, N2660049395B2, N26600496C21F, N266005147511, N266005165603, N26600519AEF4, N266005300085, N26600531110C, N266005399D44, N266005404D8D, N266005415C04, N26600548C1C5, N26600549D04C, N2660055366CE, N266005607E3D, N266005670A82, N26600568F275, N26600569E3FC, N26600578EBAD, N26600580E42D, N26600582C73F, N26600583D6B6, N26600590FDF5, N26600594BBD1, N26600600C589, N266006059224, N266006095848, N266006158BFC, N266006194190, N26600622D52B, N26600625A194, N26600630EFE1, N26600631FE68, N26600632CCF3, N26600633DD7A, N26600634A9C5, N26600644E4CD, N26600645F544, N266006583779, N2660065926F0, N26600663A3C2, N26600664D77D, N26600665C6F4, N266006719908, N26600673BA1A, N26600674CEA5, N2660068339D2, N266006844D6D, N266006877FF6, N266006888701, N266007018EDC, N26600702BC47, N26600703ADCE, N266007090294, N26600712A59F, N266007191B4C, N26600720ACE5, N26600721BD6C, N26600724EAC1, N26600727D85A, N26600732962F, N2660073387A6, N26600734F319, N266007478C8A, N2660074965F4, N26600750E1ED, N256235208276, N260661298F05, N265342048391, N266004838C6A, N26600730B53D, N25343032AE1E, N2562353817E6, N25623552EC6C, N256240889DD2, N26066132280E, N260661377FA3, N26407319275A, N266006855CE4, N25342690BA1E, N2534269927DF, N253429121D0B, N253429311C20, N25623527F6C9, N263075400E2B, N26600559C994, N2534236332F0, N25342691AB97, N2562408457BE, N2562409119CB, N26494862898E, N26494866CFAA, N2649486826D4, N26600715D120, N25343047B4BB, N256240832301, N26494869375D, N26494870B344, N26494871A2CD, N26494874F560, N26494875E4E9, N26494877C7FB, N264948783F0C, N264948792E85, N26494880308C, N264948812105, N26494882139E, N264948830217, N2649488476A8, N264948856721, N2649488655BA, N264948874433, N26494888BCC4, N26494889AD4D, N264948902954, N2649489138DD, N264948920A46, N264948931BCF, N264948946F70, N264948957EF9, N264948964C62, N264948975DEB, N26494898A51C, N26494899B495, N26494900A490, N26494901B519, N264949028782, N26494904E2B4, N26494905F33D, N26494907D02F, N2649490828D8, N264949093951, N26494916D87E, N26494917C9F7, N264949192089, N264949308EF8, N264949319F71, N26494932ADEA, N264949391339, N26494940C3F0, N26494941D279, N26494946A6C6, N26494947B74F, N264949484FB8, N26494962D352, N26494964B664, N26494974AFBC, N26494975BE35, N2649497865D0, N264949842C74, N264949871EEF, N26494988E618, N26494999EE49, N264950002435, N264950020727, N264950130F76, N2649501658DB, N2649502451A1, N2649502672B3, N26495027633A, N253426567688, N253426713187, N253427056077, N255855686211, N25623535CC03, N25623543E43D, N2562408674AC, N26407813A1A6, N26407814D519, N26407815C490, N26407816F60B, N26407817E782, N264078181F75, N264078190EFC, N26407820B955, N264078229A47, N26407824FF71, N26407825EEF8, N26407826DC63, N26407827CDEA, N26407828351D, N264078292494, N26407830A08D, N26407832839F, N264078339216, N26407834E6A9, N26407835F720, N26407837D432, N264078382CC5, N264078393D4C, N26407840ED85, N26407841FC0C, N26407842CE97, N26407844ABA1, N26407845BA28, N2640784688B3, N26407847993A, N264078497044, N26407850F45D, N26407851E5D4, N26407852D74F, N26407853C6C6, N26407854B279, N26407862FD27, N26407863ECAE, N264078649811, N264078658998, N26407866BB03, N26407867AA8A, N26407868527D, N2640786943F4, N26407870C7ED, N26407871D664, N26407872E4FF, N26407873F576, N2640787481C9, N264078759040, N26407876A2DB, N26407877B352, N264078784BA5, N264078795A2C, N264078804425, N2640788155AC, N264078826737, N2640788376BE, N264078840201, N264078851388, N26407887309A, N26407888C86D, N26407889D9E4, N264078905DFD, N264078927EEF, N264078936F66, N264078950A50, N2640789638CB, N264078972942, N26407898D1B5, N26407899C03C, N26407900D039, N26407901C1B0, N26407902F32B, N26407903E2A2, N26407904961D, N264079058794, N26407906B50F, N26407907A486, N264079085C71, N264079094DF8, N26407910C9E1, N26407911D868, N26407912EAF3, N26407913FB7A, N264079148FC5, N264079159E4C, N26407916ACD7, N26407917BD5E, N2640791845A9, N264079195420, N26407920E389, N26407921F200, N26753274AA76, N26753275BBFF, N267532768964, N2675327798ED, N26753278601A, N267532797193, N267532925550, N2675329521EF, N26753305AC2B, N267533069EB0, N267533078F39, N26753315B5F3, N2675331796E1, N267533186E16, N26753327BC89, N2675332955F7, N26753332F2FC, N26753345CB4B, N2675334810AE, N26753357F181, N267533639DCD, N26753375E123, N267533792B4F, N267533803546, N267533865070, N2675338741F9, N267533902C9E, N2675340431A7, N26753405202E, N2675340612B5, N267534160B6D, N26753419F39A, N2675342155BA, N267534305DEB, N267534314C62, N267534327EF9, N26753447645C, N267534489CAB, N267534498D22, N26753450093B, N267534664665, N2675346757EC, N2676044028E2, N267604420BF0, N267604431A79, N267604446EC6, N267604457F4F, N267604464DD4, N267604475C5D, N26760448A4AA, N26760449B523, N26760450313A, N2676045120B3, N267604521228, N26760454771E, N267604556697, N26760456540C, N267604574585, N26760458BD72, N26760459ACFB, N267604601B52, N267604623840, N2676046329C9, N267604645D76, N267604654CFF, N267604667E64, N267604698693, N26760470028A, N267604711303, N263075598A32, N26438881FA99, N26760468971A, N267604722198, N2676047444AE, N267604755527, N2676047667BC, N267604777635, N267604788EC2, N267604799F4B, N267604808142, N26760482A250, N26760483B3D9, N26760484C766, N267605312EBF, N267605347912, N26760535689B, N267605365A00, N267605374B89, N26760538B37E, N26760539A2F7, N26760540723E, N2676054163B7, N26760542512C, N26760544341A, N267605452593, N267605461708, N267605470681, N26760549EFFF, N267605506BE6, N267605517A6F, N2676055248F4, N26760553597D, N267605553C4B, N267605560ED0, N267605571F59, N26760558E7AE, N26760559F627, N26760560418E, N267605615007, N26760562629C, N2676056407AA, N2534266039D6, N2534292105F8, N26066127667B, N26407330892B, N26407843DF1E, N2640784861CD, N26407861CFBC, N264078914C74, N26438834499C, N264389423B7E, N264389476CD3, N26494865FD31, N264949860F66, N266005322397, N26600601D400, N266006866E7F, N267533580976, N26760441396B, N267604733011, N267605542DC2, N267605651623, N2676056624B8, N267605673531, N26760569DC4F, N267605705856, N267605727B44, N267605736ACD, N267605750FFB, N267605763D60, N267605772CE9, N26760578D41E, N26760579C597, N26760580DB9E, N2640781182B4, N264078238BCE, N26407836C5BB, N26494864ECB8, N26535178F72A, N26535179E6A3, N26535180F8AA, N26535189656B, N26535191F0FB, N26535193D3E9, N265351997CB3, N266005643819, N266006439072, N2685920734C2, N2685922435E9, N268592252460, N2685922616FB, N268592512F4C, N268592530C5E, N268592556968, N268592733FEE, N268592755AD8, N26859293A5FE, N26859295C0C8, N268592990AA4, N268593231B8A, N268593264C27, N26859329B4D0, N26859347097E, N26859349E000, N268593714620, N268593736532, N2685938949A0, N26859392EEAB, N268593948B9D, N268594090B65, N26407821A8DC, N264383460E20, N264950121EFF, N267532961374, N2675335194B7, N267604676FED, N267605321C24, N2676054340A5, N26760548FE76, N26760581CA17, N26760582F88C, N26760583E905, N267605858C33, N26760587AF21, N26760589465F, N26760590C246, N26760591D3CF, N26760592E154, N26760593F0DD, N267605948462, N26760596A770, N26760597B6F9, N267605984E0E, N267605995F87, N26760600FA3A, N26760601EBB3, N26760602D928, N26760603C8A1, N26760604BC1E, N267606069F0C, N267606078E85, N267606087672, N2676060967FB, N26760610E3E2, N268593452A6C, N26859413BDE7, N26859437C873, N268594648450, N268594695FB5, N26859472F8BE, N26859473E937, N268594758C01, N26760612C0F0, N26760613D179, N26760614A5C6, N26760615B44F, N2676061686D4, N26760617975D, N2630753152AA, N264388415339, N2649487381DF, N26760568CDC6, N267605741E72, N268594108F7C, N26407352FEE9, N26407357A944, N26407831B104, N264078862113, N26438797CCB0, N26535201C887, N2660059871BD, N2676048190CB, N26760586BEA8, N2685937057A9, N26970204ED99, N26970205FC10, N26970206CE8B, N26970207DF02, N2697020827F5, N26970209367C, N26970211A3EC, N269702129177, N2697021380FE, N26970215E5C8, N26970216D753, N269702218984, N26970223AA96, N26970224DE29, N26970225CFA0, N26970226FD3B, N26970227ECB2, N269702281445, N2697022905CC, N2697023081D5, N26970231905C, N26970232A2C7, N26970233B34E, N26970234C7F1, N26970235D678, N26970236E4E3, N26970237F56A, N269702391C14, N26970240CCDD, N26970241DD54, N26970242EFCF, N26970243FE46, N269702448AF9, N269702459B70, N26970246A9EB, N26970247B862, N269702484095, N26970249511C, N26970251C48C, N26970252F617, N26970253E79E, N269702549321, N2697025582A8, N26970256B033, N26970257A1BA, N2697025948C4, N267605849DBA, N2676058857D6, N2676059595EB, N26760611F26B, N26859228FF85, N26970258594D, N270264669BEA, N270264678A63, N270264687294, N27026471F68D, N27026472C416, N27026473D59F, N27026474A120, N27026475B0A9, N270264768232, N2702647793BB, N270264786B4C, N270264797AC5, N2702648064CC, N270264817545, N2702648422E8, N270264853361, N2702648601FA, N270264871073, N27026488E884, N27026489F90D, N270264907D14, N270264916C9D, N270264934F8F, N270264943B30, N270264952AB9, N270264961822, N2702649709AB, N27026498F15C, N27026499E0D5, N27026500F0D0, N27026502D3C2, N27026503C24B, N27026504B6F4, N25623556AA48, N26970210B265, N26970214F441, N26970217C6DA, N269702183E2D, N26970220980D, N26970222BB1F, N26970250D505, N270264925E06, N27026505A77D, N2702650695E6, N27026507846F, N270265087C98, N27026511F881, N27026512CA1A, N27026513DB93, N27026514AF2C, N27026515BEA5, N270265179DB7, N270265186540, N2702651974C9, N27026520C360, N27026521D2E9, N27026523F1FB, N2702652594CD, N27026526A656, N270265284F28, N270265295EA1, N27026530DAB8, N27026531CB31, N27026532F9AA, N27026533E823, N270265349C9C, N270265358D15, N27026536BF8E, N27026537AE07, N270265394779, N2702654097B0, N270265418639, N27026542B4A2, N27026543A52B, N27026544D194, N27026545C01D, N27026546F286, N270265481BF8, N270265490A71, N270265519FE1, N27026553BCF3, N26942619687F, N26942620DFD6, N26942621CE5F, N26942622FCC4, N26942623ED4D, N2694262499F2, N26942625887B, N26942626BAE0, N26942627AB69, N26942628539E, N269426294217, N26942630C60E, N26942631D787, N26942632E51C, N26942633F495, N26942634802A, N2694263591A3, N26942636A338, N26942637B2B1, N269426384A46, N269426408B06, N269426419A8F, N26942645DCAB, N26942642A814, N26942643B99D, N26942646EE30, N2694264916C7, N269426518357, N26942653A045, N269426581E96, N26942660B8B6, N269426629BA4, N26942664FE92, N26942666DD80, N26942670A16E, N26942672827C, N26942674E74A, N26942676C458, N269426782D26, N2694268022A6, N269426864790, N269426875619, N26942688AEEE, N269426903B7E, N26942692186C, N269426947D5A, N269426965E48, N26942698B736, N26942699A6BF, N26942700B6BA, N2694270295A8, N269427038421, N26942705E117, N26942707C205, N269427083AF2, N26942710AF62, N269427139DF9, N26942715F8CF, N26942723B791, N26942724C32E, N26942725D2A7, N26942726E03C, N269427280942, N26942737E86D, N26942738109A, N26942742F2C8, N2694274497FE, N26942746B4EC, N26942751D98B, N26942752EB10, N26942753FA99, N26942760E26A, N26942761F3E3, N26942762C178, N26942763D0F1, N26942764A44E, N26942765B5C7, N26942766875C, N2694276796D5, N26760605AD97, N27026465A971, N27026469631D, N27026470E704, N2702648247DE, N270265168C3E, N27026522E072, N270265248544, N27026527B7DF, N27026547E30F, N270265508E68, N27026552AD7A, N269702192FA4, N270264835657, N27026501E159, N27026510E908, N269427309CD2, N27204460D974, N27204461C8FD, N27204463EBEF, N272044649F50, N272044658ED9, N27204467ADCB, N27204468553C, N2720446944B5, N27204470C0AC, N27204471D125, N27204472E3BE, N27204473F237, N272044759701, N27204476A59A, N27204477B413, N272044784CE4, N272044795D6D, N272044804364, N2720448152ED, N272044826076, N2720448371FF, N272044840540, N2720448514C9, N272044862652, N2720448737DB, N27204488CF2C, N272044905ABC, N272044914B35, N272044936827, N272044972E03, N2721477462A8, N272147757321, N2721477641BA, N27214778A8C4, N27214779B94D, N27214780A744, N27214781B6CD, N272147828456, N2721478395DF, N27214784E160, N27214785F0E9, N27214786C272, N27214787D3FB, N272147882B0C, N272147893A85, N27214790BE9C, N27214791AF15, N272147929D8E, N272147938C07, N27214794F8B8, N27214795E931, N27214796DBAA, N27214797CA23, N2721479832D4, N27214799235D, N272148002343, N2721480132CA, N272148020051, N2721480311D8, N272148046567, N2721480574EE, N272148064675, N2721480757FC, N27214809BE82, N272148103A9B, N272148112B12, N272148121989, N272148130800, N272148147CBF, N272044950D11, N272044963F8A, N27204499C77D, N27204501C6F1, N27204503E5E3, N27204504915C, N2720450580D5, N27204506B24E, N27204507A3C7, N272045085B30, N272045094AB9, N27204510CEA0, N27204530FD10, N27204531EC99, N27204532DE02, N27204534BB34, N27204535AABD, N272045369826, N2720453789AF, N272045387158, N2720453960D1, N27204540B018, N27204541A191, N27204542930A, N272045438283, N27204544F63C, N27204545E7B5, N27204546D52E, N27204547C4A7, N27204462FA66, N27204466BC42, N272044748688, N2720449279AE, N27204500D778, N272147775033, N27214808AF0B, N27204533CF8B, N2730664173A8, N273066424133, N2730664350BA, N273066442405, N27306645358C, N27306648EE69, N27306649FFE0, N273066507BF9, N273066516A70, N2730665258EB, N273066534962, N273066543DDD, N273245529871, N2732455389F8, N27324554FD47, N27324555ECCE, N27324556DE55, N27324557CFDC, N27324558372B, N2732455926A2, N27324560910B, N273245618082, N27324562B219, N27324563A390, N27324564D72F, N27324565C6A6, N27324566F43D, N27324567E5B4, N273245681D43, N273245690CCA, N2732457088D3, N27324571995A, N27324572ABC1, N27324573BA48, N27324574CEF7, N27324575DF7E, N27324576EDE5, N27324577FC6C, N27324578049B, N273245791512, N273245811A92, N273245822809, N273245833980, N273245844D3F, N273245855CB6, N273245866E2D, N273245877FA4, N273245888753, N2732458996DA, N2732459012C3, N27324591034A, N2732459231D1, N273245932058, N2732459454E7, N27324595456E, N2732459677F5, N27324597667C, N273245989E8B, N273245998F02, N273246002ABF, N273246013B36, N2732460209AD, N273246031824, N273246046C9B, N273246064F89, N273246075E00, N27324608A6F7, N27324609B77E, N273246103367, N2732461122EE, N273246121075, N2732461301FC, N273246147543, N2732461564CA, N273246165651, N27324619AEA6, N27324620190F, N273246210886, N273246232B94, N273246245F2B, N273246254EA2, N273246267C39, N273246276DB0, N273246289547, N27324631115E, N2732463223C5, N27324633324C, N2732463446F3, N27324635577A, N273246377468, N273246388C9F, N273246404DDF, N273246415C56, N273246426ECD, N273246437F44, N273246440BFB, N273246451A72, N2732464628E9, N273066552C54, N273245800B1B, N273246057D12, N2732461747D8, N273246223A1D, N2732462984CE, N2732463665E1, N273246399D16, N27324618BF2F, N27695394CDEE, N27695395DC67, N27695396EEFC, N27695397FF75, N276953980782, N27695399160B, N27695400D0D7, N27695401C15E, N2769540496F3, N27695405877A, N276954085C9F, N276954094D16, N27695410C90F, N27695411D886, N27695412EA1D, N27695413FB94, N276954148F2B, N27695416AC39, N27695417BDB0, N276954184547, N2769541954CE, N27695421F2EE, N27695422C075, N27695423D1FC, N27695424A543, N27695425B4CA, N276954268651, N2769542797D8, N27695430FABF, N27695431EB36, N27695432D9AD, N27695433C824, N27695434BC9B, N276954369F89, N2778132677D7, N27781327665E, N277813289EA9, N277813298F20, N27695393B951, N27695402F3C5, N27695406B5E1, N27695407A468, N276954159EA2, N27695420E367, N276954297EA6, N27695435AD12, N276954378E00, N277813311AB0, N27781332282B, N2778133339A2, N277813355C94, N277813366E0F, N277813377F86, N277813388771, N277813462307, N27781347328E, N27781348CA79, N277813607581, N2778136610B7, N277813300B39, N276955596972, N27695560DEDB, N27695561CF52, N27695562FDC9, N27695563EC40, N276955658976, N27695566BBED, N27695567AA64, N276955685293, N27695569431A, N27695570C703, N27695571D68A, N27695573F598, N27695576A235, N27695577B3BC, N276955784B4B, N27695572E411, N2769556498FF, N276955748127, N27681972381A, N276819745D2C, N276819754CA5, N276819767E3E, N276819776FB7, N276819789740, N2768197986C9, N2768198098C0, N276819818949, N27681982BBD2, N27681984DEE4, N27681985CF6D, N27681986FDF6, N27681987EC7F, N276819881488, N276819890501, N276819919091, N27681992A20A, N27681993B383, N27681994C73C, N276820068AE5, N27682008639B, N27682022FF71, N27682023EEF8, N276820249A47, N276820258BCE, N27682026B955, N27682031D432, N27682033F720, N2768203849F3, N2768204088B3, N27682041993A, N27682065ECAE, N27682068374B, N276819732993, N276819908118, N27682034839F, N276820590CAA, N276819710A81, N27681983AA5B, N27696054AEE7, N27696055BF6E, N276960568DF5, N276960579C7C, N27696058648B, N276960597502, N27696060C2AB, N27696061D322, N27696062E1B9, N27696063F030, N27696064848F, N276960659506, N27696066A79D, N27696067B614, N276960684EE3, N276960695F6A, N27696070DB73, N27696071CAFA, N27696072F861, N27696073E9E8, N276960749D57, N276960758CDE, N27696076BE45, N27696077AFCC, N2769607946B2, N2769608058BB, N276960814932, N276960827BA9, N276960836A20, N276960841E9F, N276960850F16, N276960872C04, N27696088D4F3, N27696089C57A, N276960904163, N2769609150EA, N276960926271, N2769609373F8, N2769609516CE, N276960962455, N2769609735DC, N27696098CD2B, N27696100CCA7, N27696101DD2E, N27696102EFB5, N27696103FE3C, N27817120802E, N2781712191A7, N27817122A33C, N27817123B2B5, N27817124C60A, N27696078573B, N276960863D8D, N27817126E518, N27817127F491, N278171280C66, N276960940747, N27696099DCA2, N27817125D783, N279036676277, N279036689A80, N279036698B09, N279036700F10, N279036711E99, N279036722C02, N279036733D8B, N279036744934, N2790367558BD, N279036766A26, N279036777BAF, N279036788358, N2790367992D1, N279036808CD8, N279036819D51, N27903682AFCA, N27903684CAFC, N27903685DB75, N27903686E9EE, N27903687F867, N279036918489, N27903692B612, N27903693A79B, N27903694D324, N279036981948, N2790369908C1, N27903701094D, N279037023BD6, N279037032A5F, N279037045EE0, N279037054F69, N279037067DF2, N279037076C7B, N27903708948C, N279037098505, N27903710011C, N279037111095, N27903712220E, N279037144738, N2790371556B1, N27903716642A, N2790371775A3, N279037199CDD, N279037202B74, N279037213AFD, N2790372319EF, N279036880090, N27903695C2AD, N27903697E1BF, N279037133387, N279037188D54, N279037246D50, N279037257CD9, N279037264E42, N279036891119, N279037220866, N27903683BE43, N279036909500, N27903696F036, N2790370018C4, N286646442B98, N286646453A11, N28664649F07D, N2866465165ED, N286646525776, N2866465346FF, N286646543240, N286646561152, N2866465700DB, N28664658F82C, N28664659E9A5, N286646605E0C, N286646614F85, N286646627D1E, N286646636C97, N2866466509A1, N286646672AB3, N28664668D244, N28664669C3CD, N2866467047D4, N28664671565D, N2866467264C6, N28664673754F, N2866467401F0, N286646751079, N28664677336B, N28664678CB9C, N28664680C41C, N28743112F991, N28743113E818, N287431158D2E, N28743116BFB5, N2874311856CB, N287431194742, N286646641828, N286646663B3A, N2866467622E2, N28664679DA15, N28664681D595, N28664682E70E, N28664683F687, N286646848238, N2866468593B1, N28664686A12A, N28664687B0A3, N2866468959DD, N28664690DDC4, N28664691CC4D, N28664692FED6, N28664693EF5F, N286646949BE0, N286646958A69, N28664696B8F2, N28664697A97B, N28664698518C, N286646994005, N286647005000, N286647014189, N286647027312, N28664703629B, N286647041624, N2866470507AD, N2866470724BF, N286646471903, N2866465523C9, N287431149CA7, N286646435F27, N286646507464, N286646884854, N288707750AAF, N288707763834, N2887077729BD, N28870778D14A, N28870779C0C3, N28870780DECA, N28870781CF43, N28870782FDD8, N2887078498EE, N288707858967, N28870786BBFC, N288707885282, N28870789430B, N28870791D69B, N28870792E400, N28870793F589, N288707948136, N28870796A224, N28870797B3AD, N288707984B5A, N288707995AD3, N288708005ACD, N2887080279DF, N288708036856, N288708041CE9, N288708050D60, N288708063FFB, N28870808D685, N28870809C70C, N288708104315, N28870811529C, N288708126007, N28870813718E, N288708140531, N2887081514B8, N288708162623, N2887081737AA, N28870818CF5D, N28870819DED4, N28870820697D, N2887082178F4, N288708224A6F, N288708235BE6, N288708242F59, N288708253ED0, N288708260C4B, N288708271DC2, N28870828E535, N28870829F4BC, N2887083070A5, N28870831612C, N2887083253B7, N28870833423E, N288708343681, N288708352708, N288708361593, N28870838FCED, N28870839ED64, N288708403DAD, N288708412C24, N288708421EBF, N288708430F36, N288708447B89, N288708456A00, N28870846589B, N288708474912, N28870848B1E5, N28870849A06C, N28743117AE3C, N28870790C712, N2887079590BF, N288710497A15, N28871061C5ED, N28871062F776, N28871063E6FF, N288710649240, N2887106583C9, N288710748B98, N28871076A88A, N288710873ACB, N28871088C23C, N2887109057AC, N28871101CBE1, N28871102F97A, N288711085620, N288711148594, N28870837041A, N28871116A686, N28871125BE75, N288711268CEE, N29084312EF3C, N2908431951EF, N29084322C554, N29084323D4DD, N29084324A062, N290843268370, N2908432792F9, N290843286A0E, N29084330FF9E, N29084332DC8C, N29084333CD05, N29084334B9BA, N290843369AA8, N2908433873D6, N29084339625F, N29084341A31F, N290843429184, N29084343800D, N29084344F4B2, N29084345E53B, N29084346D7A0, N29084347C629, N290843483EDE, N290843492F57, N29084352885C, N2908435399D5, N29084354ED6A, N29084356CE78, N29084359368F, N290843608126, N2908436190AF, N29084362A234, N29084364C702, N29084366E410, N29084367F599, N290843691CE7, N29084350AB4E, N29084513286C, N29084520309F, N290845256732, N29084528BCD7, N290845357EEA, N290845364C71, N29084539B486, N2908454175C6, N2908454356D4, N2908454533E2, N29084548E807, N290845525E85, N290845552A3A, N2908455618A1, N290845570928, N29084558F1DF, N29084559E056, N2908456057FF, N290845614676, N2908456274ED, N2908456411DB, N2908456632C9, N290845672340, N29084568DBB7, N29084569CA3E, N290845704E27, N290845715FAE, N290845726D35, N290845737CBC, N290845740803, N29084575198A, N290845762B11, N290845773A98, N29084578C26F, N29084579D3E6, N29084580CDEF, N29084581DC66, N29084582EEFD, N29084583FF74, N290845848BCB, N290845859A42, N29084586A8D9, N2908458841A7, N29084589502E, N29084590D437, N29084592F725, N29084593E6AC, N29084595839A, N29084596B101, N29084597A088, N2908459949F6, N29084602CF59, N29084603DED0, N29084604AA6F, N29084605BBE6, N2908460798F4, N290846086003, N29084609718A, N29084610F593, N29084611E41A, N29084612D681, N29084613C708, N2908461690A5, N29084617812C, N290846196852, N29084620DFFB, N29084621CE72, N29084622FCE9, N29084623ED60, N2908462499DF, N29084626BACD, N29084627AB44, N2908462853B3, N29084629423A, N29084631D7AA, N28871124AFFC, N290843184066, N29084320E646, N29084325B1EB, N290843297B87, N290843378B21, N29084340B296, N29084357DFF1, N29084363B3BD, N2908461879DB, N286647063536, N28870783EC51, N28871073FF27, N28871113F12B, N29084351BAC7, N29084355FCE3, N290843582706, N29084365D68B, N290843680D6E, N290845230204, N28888622D550, N28888623C4D9, N28888624B066, N28888625A1EF, N288886287A0A, N28888631FE13, N28888632CC88, N28888633DD01, N28888634A9BE, N28888635B837, N288886368AAC, N288886379B25, N2888863863D2, N28888639725B, N28888640A292, N28888641B31B, N288886428180, N288886439009, N28888644E4B6, N28888645F53F, N28888646C7A4, N290952392F7C, N29095240FFB5, N29095241EE3C, N29095242DCA7, N29095243CD2E, N29095244B991, N29095245A818, N290952469A83, N290952478B0A, N2909524873FD, N290952496274, N29095250E66D, N29095251F7E4, N29095252C57F, N29095253D4F6, N29095255B1C0, N29095256835B, N2909525792D2, N290952586A25, N290952597BAC, N29095260CC05, N29095261DD8C, N29095262EF17, N29095263FE9E, N290952648A21, N290952659BA8, N29095266A933, N29095267B8BA, N29095268404D, N2909526951C4, N29095270D5DD, N29095271C454, N29095272F6CF, N29095273E746, N2909527493F9, N290952785995, N29095279481C, N290952805615, N29095281479C, N290952841031, N2909528501B8, N290952863323, N2909528722AA, N29095288DA5D, N29084591C5BE, N290952758270, N29095276B0EB, N29095283648E, N29084331EE17, N290845636564, N290845650052, N29084630C623, N2888862782FD, N29084587B950, N29095254A049, N290952827507, N292282316701, N29228232559A, N292282334413, N2922823430AC, N292282352125, N2922823613BE, N29228239EB49, N292282403B80, N292282412A09, N29228243091B, N292282456C2D, N292282465EB6, N29228248B7C8, N29228249A641, N292282502258, N29228252014A, N2922825310C3, N29228254647C, N2922825575F5, N2922825756E7, N29228258AE10, N29228259BF99, N292282600830, N2922826119B9, N292282633AAB, N292282644E14, N292282655F9D, N292282666D06, N292282677C8F, N292282688478, N2922826995F1, N2922827011E8, N292282710061, N2922827232FA, N292282622B22, N292282732373, N2922827457CC, N292282754645, N292282789DA0, N292282809220, N29228282B132, N29228283A0BB, N29228284D404, N29228285C58D, N29228286F716, N29228287E69F, N292282881E68, N292282890FE1, N29228292A8EA, N29228293B963, N29228294CDDC, N29228295DC55, N29228297FF47, N292282991639, N2922830117B5, N2922830334A7, N292283044018, N292283055191, N29084313FEB5, N29084606897D, N29084614B3B7, N29084615A23E, N29228238FAC0, N2922825133D1, N2922827674DE, N292282798C29, N292282919A71, N2922829807B0, N292282447DA4, N292282776557, N2922828183A9, N292282908BF8, N29084601FDC2, N292282370237, N29228256476E, N293193302908, N293193663286, N29319428309D, N29319445BFA8, N293194748449, N2931949149F4, N29319521F780, N2931968969DC, N293197152E74, N29319732697B, N29319812100C, N293199423468, N293199544886, N2932000884E3, N29320119D66E, N293201352FB2, N29320225D90E, N293203169B21, N293203509977, N293205266790, N293205301B7E, N2932053329E5, N293205504FAE, N29320568E98E, N293205986A46, N29320628618A, N29320655F767, N29320726D228, N29320755ADBB, N29320789D7AF, N29320802ED7B, N293208901971, N29320971C834, N29303855D439, N2930388210FE, N29303888BFA4, N29311930BFAA, N29311945A50F, N29311951FAF3, N29311966A424, N293197006001, N29319728DFF9, N29321118DB5C, N29320798DFFE, N293450908A7C, N29345092A96E, N29345096EF4A, N2934509917BD, N2934510224AA, N293451101E60, N293451110FE9, N293451212581, N29345122171A, N293451346BF4, N29345139B011, N2934514060D8, N2934514243CA, N29345159E4C1, N29345164154C, N2934517269A2, N29345195870D, N293451985CE8, N29345201F955, N29348218F6C3, N29348219E74A, N29348221416A, N293482351E96, N29348238C573, N29348265602E, N2934827579F6, N293483504737, N293484139EC9, N29348416C964, N293484778C3D, N293484852CE7, N29348574E47A, N293485974CF1, N293486044616, N29348682EDE0, N293522381299, N29352243E142, N29352293513A, N293119841D26, N293196997004, N29319790B719, N29320867EE06, N2932113247B6, N293451740C94, N293451859ED5, N293482630518, N2934826471A7, N2934864176DB, N293451715B39, N29348407C135, N29348414EA76, N29348441C363, N29348443E071, N293484458547, N293487370439, N2934909769F7, N29547847A9F5, N295435436242, N295435551EAC, N295478942B16, N2954346039B5, N296851125EC5, N296855994DB7, N29685923A3E6, N297223976DB2, N29722568C054, N29722597BB6B, N2972261973B1, N29722665F4D5, N297226874DD7, N29722706C842, N29722742E906, N29722821C58A, N297228802973, N297228956706, N297230047BEE, N297230120700, N29723068E552, N29723098669A, N29723159846F, N297231757DB3, N297231828AC4, N29724234B31F, N29724265DC2E, N29724366B469, N29724372EB95, N29724479D943, N29724702C5FC, N29724711EEBF, N29724726B068, N297248696538, N297250993C89, N29725202E0FA, N297248916AB8, N2972499202FF, N297251886E04, N29726311B4A8, N29726316C017, N29726324C96D, N296854776405, N29685752EF7C, N2968637924B3, N29686584AA47, N29722644D6EC, N297244994353, N2972511307CF, N297243339D7C, N29722854DF2F, N297230056A67, N29835976F4BD, N298360720B4A, N29836091A3C1, N2983610479A8, N29841352B063, N29841377D47E, N29841389BEC8, N29841411699D, N29841725EAB5, N298417497409, N29841767AEC7, N29842128C244, N298421500304, N298421765582, N29842196CF92, N2984343072D2, N298434421CC8, N298434600C6A, N298435110A37, N29843867DD76, N298438946C25, N29843910BE1D, N29843928183D, N29844272B758, N2984428672B4, N29844316FF70, N298444471C44, N29844471531A, N29844487B5E4, N298445027655, N29845193CF1E, N29845202E6EB, N29846210F9EC, N29846221C20D, N29846357B0EF, N29846370F7E0, N298463796A21, N2984638346B3, N29846409AB2C, N298464164A03, N29846458C41D, N29846481E9A4, N298465466E67, N298465665DD7, N29846580A2F1, N29846637DD82, N298620986AB8, N2986213735E3, N298621532717, N29862184E3D0, N29863213B3A8, N29865688F298, N2986578573A1, N298657921EC6, N298658129AC1, N29865817CD6C, N298635365BB0, N29865302AABF, N29865343DC56, N298653486285, N29865422150A, N29865426532E, N29865428BA50, N298655206CC4, N298655224FD6, N298655235E5F, N2986552609F2, N298655600BA4, N29865563393F, N298655699665, N298655745458, N298656219229, N29865627F71F, N298656309A78, N298656861BE6, N29865732E3B6, N299289115032, N29928916248D, N29928930720B, N299289316382, N299289465A35, N29928949A2C2, N298464707BE5, N29846616FFBB, N29939926191A, N29939957459B, N2993996329D7, N299399755539, N29939984C778, N29939993AA1F, N299400434FC6, N29940053561E, N2994006408C9, N299400750098, N29940079CAF4, N29940082F7EF, N29940083E666, N299400959A88, N29940097B99A, N29940105174C, N299401105939, N299401136BA2, N299401172D86, N2994012162D8, N299401261667, N29940132499B, N29940145702C, N2994014642B7, N299401521D4B, N299401530CC2, N29940159A398, N299401601431, N299401667107, N299401689879, N2994016989F0, N299401722EFB, N2994017881A1, N299401808E21, N29940186EB17, N2994018913E0, N2994019097F9, N29940002392F, N298658028319, N300040129AFE, N3000405498A8, N30004120C958, N300041639CA3, N30004202369E, N300042275283, N30004258E77C, N300042959539, N300043243AC4, N300043530373, N30004383B30B, N30004411CB04, N30004445F398, N30004527DE86, N300046341AA1, N30004868BE6E, N30004879B63F, N29844250A7FA, N29865634DC5C, N29928934342F, N299289421C11, N29940091DCAC, N2994014461A5, N299401799028, N300044811C1C, N30004690A1F5, N300049012DA3, N300050128645, N30005161A30A, N3000519745F4, N3000532352C0, N300053421799, N30005412E524, N3000542798E1, N30005446DDB8, N300056236B7D, N30005697C9F1, N30005792C480, N300059002091, N30005927679E, N30016285BED3, N30016316018C, N300163401A02, N30016378BC22, N300164284E9F, N30016457FB60, N3002980877B9, N30029847E82E, N300298591888, N30029864E905, N300298738462, N30029875E154, N3002989649DF, N30029901B0A4, N30029902823F, N30029905F680, N300299093CEC, N3002991834BD, N30029923A006, N300299281ED5, N300299290F5C, N30029941D7C4, N3002994591E0, N30029962D6EF, N30029964B3D9, N30004994AD16, N300163588F92, N300300386934, N30030050B1AC, N300300583DE4, N300300618A4D, N30030066FEF2, N300300719395, N30030076E72A, N3003008001D4, N30030081105D, N300300855679, N300300945E28, N30030136DA96, N30030142D1BA, N300301478617, N300301579FCF, N3003016596B5, N30030170D8C0, N300302007AAC, N300342055FED, N300342066D76, N300342189DD0, N3003422218E2, N298632684073, N299401711C60, N3008459237A3, N30084791B080, N30084879EC1F, N30085093032C, N30085110A5AB, N300851381A53, N300851861785, N30086286DD2C, N300864925409, N30086567DAB0, N30086588B857, N30086714109B, N300867172200, N300867216D5E, N30086759AC1E, N300867698676, N300868101C78, N300869036DE7, N300869677F13, N30087084573D, N300162344BF2, N300301915359, N300846978F6A, N300847124CDB, N30087073A04A, N301258665908, N301259164EDC, N301260950B54, N30126103E3A6, N3012611844AD, N3012615823CD, N301261709C35, N30126176F903, N3012617901F4, N301261944001, N301272103B3A, N301272165E0C, N3012722100DB, N301272289D1A, N30127249D843, N301272505C5A, N301272514DD3, N301272536EC1, N301272541A7E, N301272607632, N3012726167BB, N30127267028D, N30127268FA7A, N301272717E63, N301272771B55, N30127300783E, N3012730169B7, N301273025B2C, N30127318EDAE, N30127319FC27, N301273262EB8, N301273341472, N3012733505FB, N30127344597A, N3012734548F3, N301273467A68, N301273476BE1, N301273500686, N301273657B43, N301273765000, N30127378B97E, N30127379A8F7, N30145132C612, N30152218E9F9, N301522476FB6, N301522721273, N301523322FCF, N3015236340FE, N30152378E7F5, N30152391E024, N30152406CF86, N30152422BA12, N30152435D775, N301524495011, N30152464B844, N301524779307, N301524925EBA, N301525946150, N301526067A3E, N3015261663E6, N3015263741DF, N3015265061B0, N301526517039, N3015265242A2, N30152653532B, N30152681C041, N30152686B4FE, N301527114D85, N3015271728B3, N30152752187E, N30152758B724, N301527645720, N301527711955, N301527918345, N301527981E84, N301528044937, N301528077BAC, N301528132450, N300868331D53, N30126155F828, N3012731242F4, N3012721309A1, N3015272344FF, N301527464782, N301527777C63, N301528230E38, N3012733360CD, N301831931641, N301831957377, N3018319641EC, N301831975065, N30183198A892, N30183199B91B, N301832001CA6, N301527051279, N3018319462FE, N301832010D2F, N301832023FB4, N301832032E3D, N302019063F1D, N30201915145E, N30201920699B, N30201938FC0B, N302019421E59, N30201955733E, N3020196448DF, N302019666BCD, N302019745107, N30201982B7F9, N30201983A670, N3020199910F2, N30202005B455, N30202010FA20, N30202011EBA9, N30202012D932, N30202013C8BB, N302020408498, N302020491959, N302020802D38, N302020857A95, N3020209472C4, N30202095634D, N3020209651D6, N302021029A36, N30202124CCB0, N30202125DD39, N30202126EFA2, N30202130934C, N30202135C4E1, N30202136F67A, N30202156A2AA, N30202160EDF4, N30202161FC7D, N30202162CEE6, N30202037BD2F, N302021559031, N304507466A73, N304507510714, N304507532406, N304507554130, N304508550BF7, N30450869EBF3, N3045094548F3, N30450951170F, N3045095663B0, N304510248D0F, N30451032F1E1, N30451037A64C, N304510913E0A, N30451095782E, N30451098A3CB, N30451101B3CE, N30451159413E, N30451164B0B3, N304511678228, N30451170EF4F, N3045521752A8, N304552266949, N30455231042E, N304552770678, N30455280F10F, N30455283C394, N304552869439, N30455308E95B, N30455318F083, N304553554C06, N30455359866A, N30455365666E, N30455386CEE5, N30505287A7CF, N305053547FF0, N305053632127, N305053711BED, N30505383BB37, N30505387FD13, N305053909074, N30505395C7D9, N30505978F556, N30506053C1E6, N30506054B559, N30506055A4D0, N3050605787C2, N305060587F35, N305060596EBC, N30506060D915, N30506063EB8E, N305538189694, N30553830296C, N3055384533C9, N30553891C595, N30553893E687, N3055389583B1, N30553896B12A, N305539430C23, N30553949A379, N3055395570CD, N305539919F49, N30553996EBF6, N305539991301, N30554041E40E, N30554043C71C, N3055404879CF, N30554051FDD6, N305539881B50, N30592512AD3A, N305538357EC1, N305539474A07, N306372146A8B. |
| FEI Number |
2518422
|
Recalling Firm/ Manufacturer |
Philips Respironics, Inc.
1001 Murry Ridge Ln
Murrysville
PA
15668-8517
|
| For Additional Information Contact | Philips Customer Service 800-345-6443 |
Manufacturer Reason for Recall | This device does not indicate for use in patients with respiratory failure. |
FDA Determined Cause 2 | Device Design |
| Action | Philips sent an "URGENT Medical Device Recall Notice", for All BiPAP A30 and BiPAP A40 devices, to consignees via mail 5/9/2025. Note, the letter does not address BiPAP V30 devices. The Notice informs consignees that for patients who cannot tolerate interruption or loss of therapy an alternative ventilator is required. Philips Respironics recommends that physicians and healthcare professionals assess whether the patients under their care who are using affected devices are able to tolerate interruptions to therapy. For patients who can and cannot tolerate interruptions in therapy, Philips Respironics provides recommendations for the clinicians and patients to consider. Patients who cannot tolerate interruptions in therapy are directed to contact their physician to expedite transition to an alternative ventilator. If a Ventilator Inoperative alarm occurs and an alternative is not available, follow Appendix A3 on how to perform a hard reboot.
Consignees with any questions are to call 1-800-345-6443, prompts 4/5 or email respironics.clinical@philips.com.
An additional communication was sent to consignees on 9/12/25 with further guidance for users. |
| Quantity in Commerce | 5,958 units |
| Distribution | Worldwide distribution: US (Nationwide)including Guam and Puerto Rico; OUS (Foreign) to countries of: United Arab Emirates (AE), Argentina (AR), Austria (AT), Australia (AU), Bosnia and Herzegovina (BA), Belgium (BE), Brazil (BR), Canada (CA), Chile (CL), China (CN), Colombia (CO), Costa Rica (CR), Germany (DE), Denmark (DK), Spain (ES), France (FR), United Kingdom (GB), Hong Kong (HK), Hungary (HU), Indonesia (ID), Ireland (IE), India (IN), Italy (IT), Jordan (JO), Japan (JP), Korea, Republic of (KR), Lebanon (LB), Morocco (MA), Monaco (MC), Montenegro (ME), Mexico (MX), Malaysia (MY), Netherlands (NL), Norway (NO), Nepal (NP), Peru (PE), Philippines (PH), Pakistan (PK), Poland (PL), Portugal (PT), Qatar (QA), Romania (RO), Serbia (RS), Russian Federation (RU), Saudia Arabia (SA), Singapore (SG), Slovakia (SK), Thailand (TH), Turkey (TR), Taiwan (TW), Viet Nam (VN), South Africa (ZA). |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
|
|
|