| | Class 2 Device Recall VITROS Chemistry Products Performance Verifier I |  |
| Date Initiated by Firm | June 10, 2025 |
| Date Posted | July 17, 2025 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-2120-2025 |
| Recall Event ID |
97047 |
| Product Classification |
Colorimetry, cholinesterase - Product Code DIH
|
| Product | Brand Name: VITROS Chemistry Products
Product Name: VITROS Chemistry Products Performance Verifier I
Model/Catalog Number: 806 7324
Software Version: Not Applicable
Product Description: VITROS Chemistry Products Performance Verifier is an assayed control used to monitor performance on VITROS 250/350/5,1 FS/4600/XT 3400 Chemistry Systems and the VITROS 5600/XT 7600 Integrated Systems.
VITROS Performance Verifier is prepared from processed human serum to which enzymes, electrolytes, stabilizers,mpreservatives and other organic analytes have been added.
Component: No, not a component |
| Code Information |
Lot Code: Product Code: 806 7324 (UDI: 10758750004317)
Lots currently within expiry:
V1665 - Exp.8/14/2025
B2245 - Exp 3/25/2026
Y2110 - Exp. 3/25/2026
E2828 - Exp. 2/3/2027
|
| FEI Number |
1000136573
|
Recalling Firm/ Manufacturer |
Ortho-Clinical Diagnostics, Inc.
100 Indigo Creek Dr
Rochester
NY
14626-5101
|
| For Additional Information Contact | Laurie O'Riordan 954-6239546 |
Manufacturer Reason for Recall | Quality control (QC) drift when certain lots of Performance Verifiers (PVs) are used with certain lots of cholinesterase enzyme (CHE) slides. This drift occurs when the PVs are used 2-7 days after reconstitution; however, the label stability claims for the PV fluids is 7 days following reconstitution. This may lead to inaccurate QC results. The use of the PVs when the QC results are outside of the customer s established ranges could result in delayed results for patients due to the need to repeat QC testing or reconstitute new PV fluids and repeat testing. It could also result in erroneous patient results (small positive bias). |
FDA Determined Cause 2 | Under Investigation by firm |
| Action | On June 10, 2025, "IMPORTANT PRODUCT CORRECTION NOTIFICATION" letters were sent to distributors and customers.
CUSTOMER LETTERS: REQUIRED ACTIONS
" Monitor VITROS CHE Slides QC results to detect a drift in VITROS Performance Verifiers I and II below 2x Within-lab SD limits. If QC results are below 2x WLSD, perform QC again using freshly reconstituted vials of VITROS Performance Verifiers I and II.
" Complete the enclosed Confirmation of Receipt form no later than June 18, 2025.
" Save this notification with your User Documentation or post this notification near your laboratory's fluid reconstitution location until the issue has been resolved.
" Please forward this notification if the affected products were distributed outside of your facility.
" If your laboratory has experienced the issue described in this notification and you have not already done so, please report the occurrence to your local Global Services Organization.
QuidelOrtho's investigation is ongoing, and we are working to identify root cause. We will communicate again after root cause has been determined and the issue has been resolved. Please note that until this issue is resolved, future lots of VITROS Performance Verifiers I and II may not meet the 7-day reconstituted stability for CHE.
DISTRIBUTORS: REQUIRED ACTIONS
" Please send the enclosed customer letter and customer Confirmation of Receipt form to all customers who have been shipped VITROS Chemistry Products Performance Verifiers AND VITROS CHE Slides from August 1, 2023, to date.
" Complete the enclosed distributor Confirmation of Receipt form no later than June 18, 2025.
" Please forward this notification if the affected products were distributed outside of your facility.
If you have further questions, please contact the Global Services Organization at 1-800-421-3311. |
| Quantity in Commerce | 37429 units (US)/11,703 units (OUS) |
| Distribution | Worldwide Distribution: US (nationwide) to states of: AZ, CA, FL, GA, IL, MD, MI, MO, MS, NY, TN, TX, VA; and OUS (International) to countries of: Australia, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Netherlands, Norway, Portugal, Russia, Singapore, Spain, Sweden, and United Kingdom. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
|
|
|