| | Class 2 Device Recall BD PhaSeal & BD PhaSeal Optima |  |
| Date Initiated by Firm | June 16, 2025 |
| Date Posted | August 27, 2025 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-2453-2025 |
| Recall Event ID |
97070 |
| 510(K)Number | K123213 K130197 K140591 K181221 K201099 |
| Product Classification |
Closed antineoplastic and hazardous drug reconstitution and transfer system - Product Code ONB
|
| Product | BD PhaSeal Injector Luer (N30C), REF: 515001
BD PhaSeal Injector Luer Lock (N35), REF: 515003
BD PhaSeal Injector Luer Lock (N35C), REF: 515004
BD PhaSeal Injector Lu63er Lock (N35C Multi), REF: 515005
BD PhaSeal Protector (P14), REF: 515100
BD PhaSeal Protector (P21), REF: 515102
BD PhaSeal Protector (P21 Multi), REF: 515103
BD PhaSeal Protector (P28), REF: 515104
BD PhaSeal Protector (P50), REF: 515105
BD PhaSeal Protector (P50 Multi), REF: 515106
BD PhaSeal Protector (P53), REF: 515107
BD PhaSeal Protector (P55), REF: 515117
BD PhaSeal Connector Luer Lock (C35), REF: 515200
BD PhaSeal Connector Luer Lock (C45), REF: 515202
BD PhaSeal Y-site Connector (C80), REF: 515304
BD PhaSeal L Connector (C90), REF: 515305
BD PhaSeal Infusion Adapter (C100), REF: 515306
BD PhaSeal Infusion Adapter (C100 Multi), REF: 515307
BD PhaSeal Cap for Injector (M50), REF: 515404
BD PhaSeal Connector Cap (M70), REF: 515406
BD PhaSeal Optima Injector (N35-O), REF: 515052
BD PhaSeal Optima Injector (N35-O Multi), REF: 515053
BD PhaSeal Injector Locking (N40-O), REF: 515056
BD PhaSeal Injector Locking (N40-O Multi), REF: 515057
BD PhaSeal Optima Protector (P13-O), REF: 515060
BD PhaSeal Optima Protector (P13-O Multi), REF: 515062
BD PhaSeal Optima Protector (P28-O), REF: 515063
BD PhaSeal Optima Protector (P20-O), REF: 515064
BD PhaSeal Optima Protector (P20-O Multi), REF: 515065
BD PhaSeal Optima Connector (C35-O), REF: 515070
BD PhaSeal Optima Infusion Adapter (C100-O), 515078
BD PhaSeal Optima Infusion Adapter (C100-O Multi), REF: 515079 |
| Code Information |
REF/UDI-DI/Lot(Expiration):
515001/00382905150012.
515003/00382905150036/2402001(31-Jul-26), 2403002(31-Aug-26), 2403003(31-Aug-26), 2404006(30-Sep-26), 2405007(31-Oct-26), 2405014(31-Oct-26), 2408005(31-Jan-27), 2408006(31-Jan-27), 2409005(28-Feb-27), 2403713(31-Aug-26), 2404001(30-Sep-26), 2404002(30-Sep-26), 2404003(30-Sep-26), 2404004(30-Sep-26), 2404005(30-Sep-26), 2410001(31-Mar-27), 2410002(31-Mar-27), 2411001(30-Apr-27), 2411002(30-Apr-27), 2411003(30-Apr-27), 2411709(30-Apr-27), 2412002(31-May-27), 2501001(30-Jun-27), 2501002(30-Jun-27), 2501003(30-Jun-27), 2502001(31-Jul-27), 2502002(31-Jul-27), 2405004(31-Oct-26), 2405008(31-Oct-26), 2407001(31-Dec-26), 2412001(31-May-27), 2503001(31-Aug-27), 2503002(31-Aug-27), 2503003(31-Aug-27), 2504001(30-Sep-27), 2504002(30-Sep-27), 2504003(30-Sep-27);
515004/00382905150043/2405001(31-Oct-26), 2408018(31-Jan-27), 2411708(30-Apr-27), 2411727(30-Apr-27);
515005/00382905150050/2403006(31-Aug-26), 2404008(30-Sep-26), 2404009(30-Sep-26), 2404010(30-Sep-26), 2404011(30-Sep-26), 2404012(30-Sep-26), 2405002(31-Oct-26), 2407010(31-Dec-26), 2409011(28-Feb-27), 2409012(28-Feb-27), 2409016(28-Feb-27), 2410007(31-Mar-27), 2410008(31-Mar-27), 2410010(31-Mar-27), 2411725(30-Apr-27), 2405003(31-Oct-26), 2407011(31-Dec-26), 2407012(31-Dec-26), 2408008(31-Jan-27), 2408009(31-Jan-27), 2409007(28-Feb-27), 2409008(28-Feb-27), 2409009(28-Feb-27), 2409010(28-Feb-27), 2501006(30-Jun-27), 2503738(31-Aug-27), 2504004(30-Sep-27);
515100/00382905151002/2504117(31-Mar-30), 2404103(31-Mar-29), 2404122(31-Mar-29), 2404123(31-Mar-29), 2406112(31-May-29), 2406113(31-May-29), 2404124(31-Mar-29), 2405130(30-Apr-29), 2407112(30-Jun-29), 2409113(31-Aug-29), 2410113(30-Sep-29), 2410114(30-Sep-29), 2411115(31-Oct-29), 2412109(30-Nov-29), 2502116(31-Jan-30), 2503112(28-Feb-30);
515102/00382905151026/2403101(28-Feb-29), 2403102(28-Feb-29), 2403103(28-Feb-29), 2403104(28-Feb-29), 2403141(28-Feb-29), 2403707(28-Feb-29), 2404130(31-Mar-29), 2404131(31-Mar-29), 2404132(31-Mar-29), 2404141(31-Mar-29), 2404143(31-Mar-29), 2409601(31-Aug-29), 2403140(28-Feb-29), 2404125(31-Mar-29), 2404126(31-Mar-29), 2404127(31-Mar-29), 2404128(31-Mar-29), 2404129(31-Mar-29), 2404142(31-Mar-29), 2409108(31-Aug-29), 2409109(31-Aug-29), 2410701(30-Sep-29), 2412108(30-Nov-29), 2412114(30-Nov-29), 2501101(31-Dec-29), 2501102(31-Dec-29), 2501103(31-Dec-29), 2501104(31-Dec-29), 2501105(31-Dec-29), 2504108(31-Mar-30), 2504109(31-Mar-30), 2504119(31-Mar-30.
515103/00382905151033/2404133(31-Mar-29), 2404134(31-Mar-29), 2404140(31-Mar-29), 2410101(30-Sep-29), 2410102(30-Sep-29), 2410103(30-Sep-29), 2410104(30-Sep-29), 2504111(31-Mar-30).
515104/00382905151040/2410119(30-Sep-29), 2410107(30-Sep-29).
515105/00382905151057/2402112(31-Jan-29), 2403122(28-Feb-29), 2403123(28-Feb-29), 2403124(28-Feb-29), 2403125(28-Feb-29), 2403127(28-Feb-29), 2404113(31-Mar-29), 2404115(31-Mar-29), 2404117(31-Mar-29), 2405101(30-Apr-29), 2405102(30-Apr-29), 2405106(30-Apr-29), 2405107(30-Apr-29), 2405108(30-Apr-29), 2405152(30-Apr-29), 2406104(31-May-29), 2406105(31-May-29), 2406106(31-May-29), 2406107(31-May-29), 2406109(31-May-29), 2407103(30-Jun-29), 2407104(30-Jun-29), 2407106(30-Jun-29), 2407107(30-Jun-29), 2407108(30-Jun-29), 2407109(30-Jun-29), 2409103(31-Aug-29), 2409602(31-Aug-29), 2411123(31-Oct-29), 2411125(31-Oct-29), 2411126(31-Oct-29), 2411707(31-Oct-29), 2412105(30-Nov-29), 2403126(28-Feb-29), 2404102(31-Mar-29), 2404110(31-Mar-29), 2404111(31-Mar-29), 2404112(31-Mar-29), 2404114(31-Mar-29), 2404116(31-Mar-29), 2404118(31-Mar-29), 2404119(31-Mar-29), 2405103(30-Apr-29), 2405104(30-Apr-29), 2405105(30-Apr-29), 2407102(30-Jun-29), 2407105(30-Jun-29), 2409104(31-Aug-29), 2409105(31-Aug-29), 2411113(31-Oct-29), 2411124(31-Oct-29), 2501112(31-Dec-29), 2501113(31-Dec-29), 2501114(31-Dec-29), 2501115(31-Dec-29), 2502106(31-Jan-30), 2502107(31-Jan-30), 2501111(31-Dec-29), 2502108(31-Jan-30), 2502109(31-Jan-30), 2502110(31-Jan-30), 2502111(31-Jan-30), 2502112(31-Jan-30), 2503101(28-Feb-30), 2503102(28-Feb-30), 2503103(28-Feb-30), 2503801(28-Feb-30), 2504101(31-Mar-30), 2504102(31-Mar-30), 2504103(31-Mar-30), 2504104(31-Mar-30), 2504105(31-Mar-30), 2504106(31-Mar-30).
515106/00382905151064/2403109(28-Feb-29), 2403110(28-Feb-29), 2403111(28-Feb-29), 2403112(28-Feb-29), 2403113(28-Feb-29), 2403114(28-Feb-29), 2403115(28-Feb-29), 2403116(28-Feb-29), 2403117(28-Feb-29), 2403118(28-Feb-29), 2403119(28-Feb-29), 2403120(28-Feb-29), 2403121(28-Feb-29), 2404101(31-Mar-29), 2404104(31-Mar-29), 2404105(31-Mar-29), 2404106(31-Mar-29), 2404107(31-Mar-29), 2404108(31-Mar-29), 2405119(30-Apr-29), 2405120(30-Apr-29), 2405121(30-Apr-29), 2405122(30-Apr-29), 2405123(30-Apr-29), 2405124(30-Apr-29), 2405125(30-Apr-29), 2405126(30-Apr-29), 2405127(30-Apr-29), 2405128(30-Apr-29), 2405129(30-Apr-29), 2406101(31-May-29), 2406102(31-May-29), 2406103(31-May-29), 2407101(30-Jun-29), 2408101(31-Jul-29), 2409102(31-Aug-29), 2410110(30-Sep-29), 2409101(31-Aug-29), 2410108(30-Sep-29), 2410109(30-Sep-29), 2411101(31-Oct-29), 2411102(31-Oct-29), 2411103(31-Oct-29), 2411104(31-Oct-29), 2412106(30-Nov-29), 2412107(30-Nov-29), 2501110(31-Dec-29), 2502102(31-Jan-30), 2502103(31-Jan-30), 2503107(28-Feb-30), 2503108(28-Feb-30), 2503118(28-Feb-30), 2503119(28-Feb-30), 2504107(31-Mar-30).
515107/00382905151071/2407115(30-Jun-29).
515117/00382905151170/2405131(30-Apr-29), 2405132(30-Apr-29), 2405133(30-Apr-29), 2405134(30-Apr-29), 2407113(30-Jun-29), 2407114(30-Jun-29), 2409115(31-Aug-29), 2409120(31-Aug-29), 2409121(31-Aug-29), 2409122(31-Aug-29), 2411117(31-Oct-29), 2503111(28-Feb-30), 2504115(31-Mar-30), 2504116(31-Mar-30).
515200/00382905152009/2403715(29-Feb-28), 2404208(31-Mar-28), 2405203(30-Apr-28), 2407204(30-Jun-28), 2409202(31-Aug-28), 2411202(31-Oct-28), 2501201(31-Dec-28), 2405204(30-Apr-28), 2407203(30-Jun-28), 2409201(31-Aug-28), 2410201(30-Sep-28), 2410202(30-Sep-28), 2411201(31-Oct-28), 2411203(31-Oct-28), 2501202(31-Dec-28), 2502201(31-Jan-29), 2503201(28-Feb-29), 2503202(28-Feb-29), 2504201(31-Mar-29), 2504202(31-Mar-29).
515202/00382905152023/2404207(31-Mar-28), 2404202(31-Mar-28), 2407201(30-Jun-28), 2412202(30-Nov-28), 2501222(31-Dec-28), 2503206(28-Feb-29), 2504203(31-Mar-29).
515304/00382905153044/2406206(30-Sep-26), 2412212(31-Jan-27), 2503802(30-Apr-27).
515305/00382905153051/2403220(29-Feb-28), 2405221(30-Apr-28), 2405222(30-Apr-28), 2409218(31-Aug-28), 2404201(31-Mar-28), 2404206(31-Mar-28), 2502215(31-Jan-29).
515306/00382905153068/2402228(31-Jan-29), 2404210(31-Mar-28), 2404211(31-Mar-28), 2404212(31-Mar-28), 2404213(31-Mar-28), 2404214(31-Mar-28), 2404215(31-Mar-28), 2404216(31-Mar-28), 2406204(31-May-28), 2410212(30-Sep-28), 2410213(30-Sep-28), 2412216(30-Nov-28), 2404209(31-Mar-28), 2405224(30-Apr-28), 2502213(31-Jan-29), 2502214(31-Jan-29), 2406202(31-May-28), 2406203(31-May-28), 2410217(30-Sep-28), 2411206(31-Oct-28), 2411207(31-Oct-28), 2411721(31-Oct-28), 2411722(31-Oct-28), 2411723(31-Oct-28), 2412205(30-Nov-28), 2412206(30-Nov-28), 2412217(30-Nov-28), 2501213(31-Dec-28), 2501214(31-Dec-28), 2503207(28-Feb-29), 2503208(28-Feb-29), 2503209(28-Feb-29), 2504204(31-Mar-29), 2504205(31-Mar-29), 2504206(31-Mar-29).
515307/00382905153075 2403714(29-Feb-28), 2502212(31-Jan-29), 2403601(28-Feb-26), 2405225(30-Apr-28), 2410216(30-Sep-28).
515404/00382905154041/2404203(31-Mar-29), 2409220(31-Aug-29), 2405201(30-Apr-29), 2405202(30-Apr-29), 2406205(31-May-29).
515406/00382905154065/2404217(31-Mar-29).
515052/00382905150524/2403303(31-Aug-26), 2403304(31-Aug-26), 2403306(31-Aug-26), 2403307(31-Aug-26), 2404301(30-Sep-26), 2406303(30-Nov-26), 2406702(30-Nov-26), 2406801(30-Nov-26), 2408305(31-Jan-27), 2409302(28-Feb-27), 2410304(31-Mar-27), 2410306(31-Mar-27), 2411301(30-Apr-27), 2411302(30-Apr-27), 2412305(31-May-27), 2501303(30-Jun-27), 2501304(30-Jun-27), 2501305(30-Jun-27), 2502303(31-Jul-27), 2502304(31-Jul-27), 2502305(31-Jul-27), 2503301(31-Aug-27), 2405301(31-Oct-26), 2405302(31-Oct-26), 2405303(31-Oct-26), 2406302(30-Nov-26), 2406310(30-Nov-26), 2406803(30-Nov-26), 2407301(31-Dec-26), 2407302(31-Dec-26), 2407303(31-Dec-26), 2408304(31-Jan-27), 2408306(31-Jan-27), 2408307(31-Jan-27), 2408308(31-Jan-27), 2409301(28-Feb-27), 2409303(28-Feb-27), 2409304(28-Feb-27), 2410305(31-Mar-27), 2410307(31-Mar-27), 2503716(31-Aug-27), 2503717(31-Aug-27), 2503721(31-Aug-27), 2503722(31-Aug-27), 2504303(30-Sep-27), 2504304(30-Sep-27), 2504305(30-Sep-27), 2504306(30-Sep-27), 2504307(30-Sep-27).
515053/00382905150531/2404302(30-Sep-26), 2404303(30-Sep-26), 2404304(30-Sep-26), 2409305(28-Feb-27), 2409306(28-Feb-27), 2409307(28-Feb-27), 2405304(31-Oct-26), 2405305(31-Oct-26), 2410308(31-Mar-27), 2412303(31-May-27), 2412304(31-May-27), 2501301(30-Jun-27), 2501302(30-Jun-27), 2406701(30-Nov-26), 2406703(30-Nov-26), 2407304(31-Dec-26), 2407305(31-Dec-26), 2407306(31-Dec-26), 2407307(31-Dec-26), 2407308(31-Dec-26), 2407802(31-Dec-26), 2408309(31-Jan-27), 2408310(31-Jan-27), 2411303(30-Apr-27), 2501601(30-Jun-27), 2502701(31-Jul-27), 2502702(31-Jul-27), 2502703(31-Jul-27), 2502704(31-Jul-27), 2503723(31-Aug-27), 2503724(31-Aug-27), 2503725(31-Aug-27), 2504308(30-Sep-27).
515056/00382905150562/2405306(31-Oct-26), 2405307(31-Oct-26), 2405308(31-Oct-26), 2405312(31-Oct-26), 2408301(31-Jan-27), 2406301(30-Nov-26), 2409310(28-Feb-27), 2410301(31-Mar-27), 2410303(31-Mar-27), 2411304(30-Apr-27), 2411305(30-Apr-27), 2412301(31-May-27), 2501306(30-Jun-27), 2501307(30-Jun-27), 2501308(30-Jun-27), 2502301(31-Jul-27), 2502302(31-Jul-27), 2407309(31-Dec-26), 2407310(31-Dec-26), 2408302(31-Jan-27), 2409309(28-Feb-27), 2410302(31-Mar-27), 2410309(31-Mar-27), 2411306(30-Apr-27), 2503736(31-Aug-27), 2503737(31-Aug-27), 2504301(30-Sep-27), 2504801(30-Sep-27).
515057/00382905150579/2405311(31-Oct-26), 2409308(28-Feb-27), 2411308(30-Apr-27), 2412302(31-May-27), 2501309(30-Jun-27), 2501310(30-Jun-27), 2408303(31-Jan-27), 2411307(30-Apr-27), 2504302(30-Sep-27).
515060/00382905150609/2403413(28-Feb-26), 2403415(28-Feb-26), 2404408(31-Mar-26), 2404409(31-Mar-26), 2405409(30-Apr-26), 2405410(30-Apr-26), 2405412(30-Apr-26), 2405413(30-Apr-26), 2405414(30-Apr-26), 2405415(30-Apr-26), 2406407(31-May-26), 2407805(30-Jun-26), 2407806(30-Jun-26), 2407807(30-Jun-26), 2403414(28-Feb-26), 2404407(31-Mar-26), 2404410(31-Mar-26), 2405411(30-Apr-26), 2407803(30-Jun-26), 2407804(30-Jun-26), 2408406(31-Jul-26), 2408407(31-Jul-26), 2408408(31-Jul-26), 2409411(31-Aug-26), 2409413(31-Aug-26), 2410402(30-Sep-26), 2410408(30-Sep-26), 2410409(30-Sep-26), 2410410(30-Sep-26), 2410416(30-Sep-26), 2410417(30-Sep-26), 2410602(30-Sep-26), 2411413(31-Oct-26), 2411414(31-Oct-26), 2412407(30-Nov-26), 2412408(30-Nov-26), 2412409(30-Nov-26), 2412410(30-Nov-26), 2412411(30-Nov-26), 2412701(30-Nov-26), 2501406(31-Dec-26), 2409412(31-Aug-26), 2409414(31-Aug-26), 2409415(31-Aug-26), 2409416(31-Aug-26), 2412412(30-Nov-26), 2412414(30-Nov-26), 2501407(31-Dec-26), 2501408(31-Dec-26), 2504409(31-Mar-27).
515062/00382905150623.
515063/00382905150630/2502706(31-Mar-27), 2502705(31-Mar-27).
515064/00382905150647/2402405(31-Jan-26), 2403401(28-Feb-26), 2403404(28-Feb-26), 2403405(28-Feb-26), 2403408(28-Feb-26), 2403409(28-Feb-26), 2403410(28-Feb-26), 2403411(28-Feb-26), 2403412(28-Feb-26), 2403703(28-Feb-26), 2403704(28-Feb-26), 2404403(31-Mar-26), 2404404(31-Mar-26), 2404405(31-Mar-26), 2404411(31-Mar-26), 2404412(31-Mar-26), 2404416(31-Mar-26), 2405401(30-Apr-26), 2405403(30-Apr-26), 2405407(30-Apr-26), 2405408(30-Apr-26), 2405417(30-Apr-26), 2406402(31-May-26), 2406403(31-May-26), 2406404(31-May-26), 2406411(31-May-26), 2406412(31-May-26), 2406413(31-May-26), 2406414(31-May-26), 2406415(31-May-26), 2406416(31-May-26), 2407402(30-Jun-26), 2407403(30-Jun-26), 2407404(30-Jun-26), 2408410(31-Jul-26), 2408413(31-Jul-26), 2408414(31-Jul-26), 2405402(30-Apr-26), 2407415(30-Jun-26), 2408403(31-Jul-26), 2408404(31-Jul-26), 2408411(31-Jul-26), 2408412(31-Jul-26), 2409401(31-Aug-26), 2409402(31-Aug-26), 2409403(31-Aug-26), 2409405(31-Aug-26), 2411403(31-Oct-26), 2411404(31-Oct-26), 2411411(31-Oct-26), 2411412(31-Oct-26), 2412401(30-Nov-26), 2412402(30-Nov-26), 2412403(30-Nov-26), 2501403(31-Dec-26), 2501404(31-Dec-26), 2501415(31-Dec-26), 2501416(31-Dec-26), 2502401(31-Jan-27), 2502403(31-Jan-27), 2502404(31-Jan-27), 2502408(31-Jan-27), 2502410(31-Jan-27), 2502707(31-Jan-27), 2410403(30-Sep-26), 2410407(30-Sep-26), 2410413(30-Sep-26), 2410419(30-Sep-26), 2410420(30-Sep-26), 2410601(30-Sep-26), 2411410(31-Oct-26), 2501405(31-Dec-26), 2502708(31-Jan-27), 2502709(31-Jan-27), 2503703(28-Feb-27), 2503705(28-Feb-27), 2503706(28-Feb-27), 2503707(28-Feb-27), 2503708(28-Feb-27), 2504405(31-Mar-27), 2504406(31-Mar-27), 2504407(31-Mar-27), 2504408(31-Mar-27), 2505404(31-Mar-27).
515065/00382905150654/2402404(31-Jan-26), 2402411(31-Jan-26), 2402412(31-Jan-26), 2403403(28-Feb-26), 2403406(28-Feb-26), 2403416(28-Feb-26), 2403417(28-Feb-26), 2404401(31-Mar-26), 2404402(31-Mar-26), 2404406(31-Mar-26), 2405404(30-Apr-26), 2405405(30-Apr-26), 2405406(30-Apr-26), 2406401(31-May-26), 2406405(31-May-26), 2406406(31-May-26), 2406408(31-May-26), 2406409(31-May-26), 2406410(31-May-26), 2407401(30-Jun-26), 2407405(30-Jun-26), 2407406(30-Jun-26), 2407407(30-Jun-26), 2407408(30-Jun-26), 2407409(30-Jun-26), 2407416(30-Jun-26), 2407417(30-Jun-26), 2408401(31-Jul-26), 2408405(31-Jul-26), 2408417(31-Jul-26), 2409404(31-Aug-26), 2409407(31-Aug-26), 2409408(31-Aug-26), 2404417(31-Mar-26), 2408402(31-Jul-26), 2409406(31-Aug-26), 2410406(30-Sep-26), 2411401(31-Oct-26), 2411402(31-Oct-26), 2411418(31-Oct-26), 2412404(30-Nov-26), 2501401(31-Dec-26), 2501413(31-Dec-26), 2501418(31-Dec-26), 2410401(30-Sep-26), 2502405(31-Jan-27), 2410404(30-Sep-26), 2410405(30-Sep-26), 2502409(31-Jan-27), 2502415(31-Jan-27), 2410414(30-Sep-26), 2410415(30-Sep-26), 2410418(30-Sep-26), 2412405(30-Nov-26), 2412601(30-Nov-26), 2501402(31-Dec-26), 2501414(31-Dec-26), 2501417(31-Dec-26), 2502402(31-Jan-27), 2502413(31-Jan-27), 2502414(31-Jan-27), 2503405(28-Feb-27), 2503601(28-Feb-27), 2503701(28-Feb-27), 2503702(28-Feb-27), 2503704(28-Feb-27), 2503709(28-Feb-27), 2503710(28-Feb-27), 2503711(28-Feb-27), 2503712(28-Feb-27), 2503713(28-Feb-27), 2503715(28-Feb-27), 2503741(28-Feb-27), 2503742(28-Feb-27), 2504403(31-Mar-27), 2504404(31-Mar-27), 2504601(31-Mar-27), 2504701(31-Mar-27), 2504702(31-Mar-27).
515070/00382905150708/2403701(28-Feb-26), 2403705(28-Feb-26), 2404507(31-Mar-26), 2404508(31-Mar-26), 2404509(31-Mar-26), 2405505(30-Apr-26), 2405506(30-Apr-26), 2406507(31-May-26), 2407506(30-Jun-26), 2407507(30-Jun-26), 2404506(31-Mar-26), 2406508(31-May-26), 2409501(31-Aug-26), 2409502(31-Aug-26), 2409506(31-Aug-26), 2410505(30-Sep-26), 2410506(30-Sep-26), 2411503(31-Oct-26), 2411504(31-Oct-26), 2412503(30-Nov-26), 2412504(30-Nov-26), 2501506(31-Dec-26), 2501507(31-Dec-26), 2501508(31-Dec-26), 2502501(31-Jan-27), 2501518(31-Dec-26), 2502502(31-Jan-27), 2503726(28-Feb-27), 2503727(28-Feb-27), 2504501(31-Mar-27), 2504502(31-Mar-27), 2504503(31-Mar-27).
515078/00382905150784/2402505(31-Jan-26), 2403505(28-Feb-26), 2403507(28-Feb-26), 2404501(31-Mar-26), 2404502(31-Mar-26), 2405501(30-Apr-26), 2405502(30-Apr-26), 2405503(30-Apr-26), 2405504(30-Apr-26), 2406504(31-May-26), 2406505(31-May-26), 2406506(31-May-26), 2407501(30-Jun-26), 2407502(30-Jun-26), 2407801(30-Jun-26), 2403504(28-Feb-26), 2404505(31-Mar-26), 2408501(31-Jul-26), 2409504(31-Aug-26), 2409505(31-Aug-26), 2410501(30-Sep-26), 2410502(30-Sep-26), 2410507(30-Sep-26), 2411501(31-Oct-26), 2411502(31-Oct-26), 2411508(31-Oct-26), 2412502(30-Nov-26), 2501501(31-Dec-26), 2501502(31-Dec-26), 2501503(31-Dec-26), 2502505(31-Jan-27), 2502506(31-Jan-27), 2502507(31-Jan-27), 2503728(28-Feb-27), 2504506(31-Mar-27), 2504507(31-Mar-27).
515079/00382905150791/2402601(31-Jan-26), 2403702(28-Feb-26), 2403706(28-Feb-26), 2404503(31-Mar-26), 2406501(31-May-26), 2406502(31-May-26), 2406503(31-May-26), 2408502(31-Jul-26), 2408503(31-Jul-26), 2408504(31-Jul-26), 2408505(31-Jul-26), 2410503(30-Sep-26), 2410504(30-Sep-26), 2403508(28-Feb-26), 2409503(31-Aug-26), 2411506(31-Oct-26), 2411507(31-Oct-26), 2412501(30-Nov-26), 2501504(31-Dec-26), 2501505(31-Dec-26), 2502503(31-Jan-27), 2502504(31-Jan-27), 2503504(28-Feb-27), 2503505(28-Feb-27), 2503506(28-Feb-27), 2504504(31-Mar-27), 2504505(31-Mar-27).
|
| FEI Number |
3012798826
|
Recalling Firm/ Manufacturer |
BD SWITZERLAND SARL
Terre Bonne Park A4
Route De Crassier 17
Eysins
Switzerland
|
Manufacturer Reason for Recall | Closed system drug transfer devices were shipped to the U.S. market without a U.S. version of the instructions for use, which could be an issue if a healthcare professional is prompted to review it. |
FDA Determined Cause 2 | Error in labeling |
| Action | On 6/16/2025, Medical Device Product Advisory notices were mailed and emailed to customers who were asked to do the following:
1) The US IFU to be attached to your product can be accessed and downloaded from the following link by searching under the applicable product code: The US IFU to be attached to your product can be accessed and downloaded from the following link by searching under the applicable product code: eIFU.bd.com
2) Circulate this notice to all those who need to be aware within your organization or to any organization where the potentially affected products have been transferred.
3) Complete the attached Customer Response Form and return via email to BDRC13@bd.com
Distributors were asked to do the following: Identify all customers within your distribution network that purchased any affected product as defined in this notification. Provide a copy of the attached customer letter to all customers to advise them of this product advisory on recalling firm's behalf.
Report any complaints experienced with the use of this product to firm via the North American Regional Complaint Center:
Phone: 1-844-823-5433 Say "Product Complaints" when prompted Mon-Fri 8:00am to 5:00pm CT, Email: productcomplaints@bd.com
|
| Quantity in Commerce | 3,884,670 |
| Distribution | US Nationwide distribution including in the states of CA, WA, AL, OH, NC, GA, MA, IL, MO, AZ, PA, MS, MN, MI, TX, VA, IN, NV, CO, FL, WV, WI, NY, NJ, UT, MD, MT, NM, TN, HI, OK, OR, AK, DC, IA, LA, ME, SC.
|
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = ONB 510(K)s with Product Code = ONB
|
|
|
|