• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Philips Azurion

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Philips Azurionsee related information
Date Initiated by FirmAugust 01, 2025
Date PostedSeptember 03, 2025
Recall Status1 Open3, Classified
Recall NumberZ-2502-2025
Recall Event ID 97249
510(K)NumberK200917 K172822 K181830 
Product Classification Interventional fluoroscopic x-ray system - Product Code OWB
ProductPhilips Azurion System configured with a 1 Phase UPS. Azurion family (R1.x, R2.x). Labeled as the following with corresponding model numbers: 1. Azurion 3 M12, Model Number 722063. 2. Azurion 3 M15, Model Number 722064. 3. Azurion 7 B12, Model Number 722067. 4. Azurion 7 B20, Model Number 722068. 5. Azurion 7 M12, Model Number 722078. 6. Azurion 7 M20, Model Number 722079. 7. Azurion 3 M12, Model Number 722221. 8. Azurion 3 M15, Model Number 722222. 9. Azurion 7 M12, Model Number 722223. 10. Azurion 7 M20, Model Number 722224. 11. Azurion 7 B12, Model Number 722225. 12. Azurion 7 B20, Model Number 722226. 13. Azurion 5 M12, Model Number 722227. 14. Azurion 5 M20, Model Number 722228. 15. Azurion 7 M20, Model Number 722282.
Code Information 1. Model Number/System Code (UDI); Serial Numbers: 722063 (00884838085275); 25, 151, 91, 85, 90, 86, 107, 89, 88, 87, 120, 153, 96, 108. 2. Model Number/System Code (UDI); Serial Numbers: 722064 (00884838085282); 64, 260, 261, 335, 218, 267, 270, 258, 35, 10, 111, 82, 125, 94, 278, 36, 55, 12, 57, 69, 484, 451, 497, 155, 272, 490, 132, 70, 251, 226, 290, 186, 48, 373. 3. Model Number/System Code (UDI); Serial Numbers: 722067 (00884838085350); 150, 69, 221, 148, 42, 98, 123, 277, 92, 103, 41, 214, 173, 172, 163, 128, 151, 22, 23, 268, 112, 97, 243, 45, 108, 113, 127, 176, 165, 33, 30, 206, 203, 48, 149, 251, 262. 4. Model Number/System Code (UDI); Serial Numbers: 722068 (00884838085367); 213, 111, 85, 89, 320, 73, 308, 172, 295, 406, 285, 413, 322, 324, 184, 442, 279, 361, 490, 42, 103, 520, 470, 472, 457, 400, 547, 510, 239, 275, 265, 365, 481, 62, 552, 44, 158, 36, 276, 68, 551, 516, 372, 46, 185, 515, 347, 416, 242, 219, 299, 126, 211, 468, 238, 568, 583, 69, 60, 426, 427, 93, 303, 302, 141. 5. Model Number/System Code (UDI); Serial Numbers: 722078 (00884838085251); 465, 143, 278, 402, 290, 524, 412, 166, 539, 532, 291, 415, 753, 441, 305, 289, 803, 399, 536, 537, 478, 479, 224, 163, 719, 364, 239, 383, 829, 513, 661, 215, 677, 709, 410, 269, 185, 902, 360, 546, 294, 906, 603, 870, 523, 598, 317, 535, 359, 436, 165, 733, 218, 204, 205, 206, 888, 923, 197, 195, 797, 444, 222, 489, 361, 362, 30, 223, 584, 962, 893, 27, 977, 156, 909, 953, 182, 958, 847, 959, 610, 490, 103, 184, 506, 23, 112, 199, 1047, 1074, 850, 643, 319, 547, 879, 572, 38, 865, 881, 261, 472, 368, 369, 243, 244, 731, 202, 442, 351, 636, 777, 40, 477, 134, 341, 563, 633, 286, 336, 512, 158, 357, 397, 896, 281, 52, 366, 190, 617, 496, 461, 463, 270, 528, 638, 529, 431, 712, 714, 219, 203, 575, 498, 577, 574, 139, 178, 179, 265, 264, 266, 511, 526, 746, 1082, 426, 36, 242, 300, 301, 795, 32, 241, 49, 51, 50, 250, 145, 774, 785, 474, 475, 452, 789, 552, 157, 818, 460, 459, 352, 46, 39, 311, 312, 277, 455, 608, 878, 220, 221, 321, 533, 699, 877. 6. Model Number/System Code (UDI); Serial Numbers: 722079 (00884838085268); 942, 948, 702, 703465, 1316, 1042, 465, 674, 703413, 1142, 1090, 526, 439, 637, 455, 334, 875, 737, 599, 29, 704281, 801, 671, 1387, 981, 1211, 1213, 499, 1230, 708, 496, 903, 887, 912, 1041, 703324, 528, 703584, 738, 351, 820, 32, 1411, 537, 445, 428, 587, 1290, 251, 1402, 318, 273, 877, 854, 897, 896, 836, 891, 940, 297, 483, 689, 717, 396, 860, 840, 790, 862, 774, 703344, 1436, 629, 1049, 1355, 1414, 703404, 703387, 703317, 703540, 58, 810, 939, 634, 53, 57, 841, 1282, 703310, 1100, 703314, 54, 703408, 246, 890, 703409, 658, 776, 703495, 18, 314, 828, 1323, 645, 646, 325, 731, 262, 1252, 703418, 474, 481, 800, 1359, 1358, 1360, 809, 1258, 1006, 602, 335, 837, 472, 703559, 342, 344, 345, 343, 1232, 1226, 1237, 1236, 34, 1400, 863, 987, 703389, 1391, 1043, 905, 577, 714, 573, 805, 228, 30, 366, 543, 921, 1200, 1222, 353, 703379, 703402, 350, 673, 669, 1212, 1285, 1216, 416, 290, 272, 703306, 35, 43, 40, 1270, 274, 466, 1155, 1033, 703356, 284, 703685, 703550, 703380, 703381, 332, 331, 330, 679, 845, 846, 328, 329, 783, 1016, 1017, 74, 1077, 37, 293, 952, 312, 703351, 703384, 759, 557, 1169, 756, 1306, 703597, 703611, 1129, 234, 75, 316, 740, 1104, 703653, 803, 1388, 67, 1133, 319, 76, 760, 966, 632, 1082, 421, 704131, 704252, 704009, 704215, 1181, 703815, 703330, 782, 1372, 703713, 725, 907, 901, 960, 542, 703470, 1428, 1037, 215, 1035, 703463, 1261, 703321, 831, 832, 410, 224, 495, 340, 839, 728, 339, 282, 444, 1338, 529, 1121, 682, 703560, 727, 576, 630, 703424, 381, 1123, 703385, 71, 583, 703386, 1154, 748, 278, 876, 478, 787, 477, 283, 997, 784, 1116, 703753, 1278, 703313, 703872, 1217, 258, 260, 259, 395, 433, 451, 1405, 703355, 703449, 703654, 703699, 703870, 256, 498, 703638, 703841, 704101, 704171, 1347, 1264, 703834, 704277, 704284, 704150, 703740, 703935, 703995, 703966, 703824, 703961, 1317, 704258, 430, 934, 492, 36, 1202, 833, 834, 1223, 364, 720, 490, 1008, 703845, 1165, 1089, 1330, 743, 271, 791, 370, 1175, 703438, 1150, 72, 844, 786, 1239, 230, 796, 257, 522, 523, 338, 1272, 764. 7. Model Number/System Code (UDI); Serial Numbers: 722221 (00884838099203); 108, 19, 153, 49, 115, 171, 6, 8, 127, 59, 56, 126, 106, 166, 155. 8. Model Number/System Code (UDI); Serial Numbers: 722222 (00884838099210); 55, 23, 44, 45, 187, 223, 99, 100, 101, 332, 51, 65, 11, 268, 188, 77, 80, 245, 109, 83, 50, 244, 177, 280, 13, 300, 5, 9. 9. Model Number/System Code (UDI); Serial Numbers: 722223 (00884838099241); 580, 669, 164, 127, 570, 131, 816, 365, 390, 352, 340, 242, 205, 627, 229, 262, 214, 458, 256, 728, 937, 944, 945, 951, 180, 439, 631, 681, 47, 126, 210, 336, 206, 367, 357, 572, 801, 197, 192, 87, 282, 177, 882, 208, 388, 581, 189, 479, 211, 186, 187, 566, 37, 370, 387, 116, 154, 99, 100, 598, 619, 854, 325, 787, 553, 185, 38, 859, 313, 314, 201, 1083, 860, 382, 645, 647, 646, 648, 30, 296, 297, 41, 914, 461, 808, 165, 33, 559, 563, 564, 558, 114, 176, 321, 661, 913, 246, 245, 422, 2, 639, 91, 333, 1056, 761, 21, 95, 531, 1017, 1037, 392, 358, 618, 590, 65, 295, 840, 12, 68, 465, 547, 702, 34, 450, 376, 451, 792, 876, 1077, 140, 988, 49, 442, 157, 539, 726, 342, 232, 492, 322, 591, 587, 812, 875, 904, 1167, 971, 532, 82, 109, 170, 4, 641, 640, 542, 543, 805, 121, 122, 374, 118, 13, 428, 983, 623, 1009, 515, 660, 712, 305, 179, 1007, 769, 754, 720, 721, 753, 755, 674, 344, 620, 768, 475, 474, 378, 294, 163, 815, 497, 362, 544, 707, 535, 536. 10. Model Number/System Code (UDI); Serial Numbers: 722224 (00884838099258); 606, 1828, 1829, 1830, 1201, 612, 503, 277, 261, 75, 182, 415, 876, 1488, 2957, 266, 512, 1160, 766, 2156, 2138, 250, 226, 360, 790, 452, 1147, 1278, 88, 1142, 792, 595, 892, 814, 813, 2995, 3120, 189, 2139, 2140, 108, 100, 591, 385, 260, 1827, 850, 338, 113, 438, 45, 444, 39, 554, 434, 3207, 696, 1499, 42, 841, 2656, 56, 2331, 421, 589, 2520, 793, 129, 84, 1810, 362, 327, 124, 125, 2280, 239, 1324, 950, 946, 381, 443, 252, 441, 621, 622, 435, 433, 99, 291, 332, 422, 1917, 2248, 2247, 2229, 2246, 1376, 1377, 1514, 351, 1960, 2045, 2403, 80, 183, 366, 101, 564, 396, 120, 454, 300, 1507, 1510, 329, 328, 1336, 307, 953, 1252, 1869, 1458, 904, 940, 1486, 165, 216, 527, 2376, 220, 176, 188, 466, 1112, 2913, 2426, 190, 490, 593, 469, 273, 333, 306, 432, 439, 244, 417, 414, 770, 516, 645, 1409, 565, 1625, 78, 267, 268, 789, 788, 478, 346, 298, 1299, 289, 1873, 2417, 1484, 1335, 1611, 657, 476, 251, 517, 507, 2018, 2239, 1555, 163, 240, 467, 468, 539, 128, 1126, 1246, 1259, 1255, 1256, 900, 191, 64, 1493, 1032, 2222, 1251, 1356, 194, 254, 1403, 440, 246, 2184, 2436, 613, 1565, 229, 286, 1508, 1461, 1512, 1511, 640, 1524, 836, 1692, 141, 574, 109, 2729, 46, 1063, 571, 174, 1490, 2569, 275, 262, 295, 272, 178, 228, 651, 462, 1049, 867, 1954, 1956, 116, 471, 233, 1995, 2185, 2860, 833, 874, 875, 1247, 1478, 330, 649, 173, 172, 379, 1477, 2752, 1672, 35, 1570, 406, 204, 206, 2081, 1042, 1591, 544, 2680, 1298, 513, 2349, 184, 2079, 545, 2080, 2082, 908, 573, 827, 1027, 1026, 1812, 781, 1204, 1184, 221, 208, 633, 3380, 816, 732, 647, 2510, 1388, 145, 2534, 2527, 2124, 1325, 2638, 632, 1766, 2214, 2269, 2575, 1169, 2515, 2337, 2338, 2823, 2825, 1539, 463, 368, 2415, 383, 563, 519, 1686, 2089, 2574, 1367, 988, 2163, 2725, 2724, 93, 1671, 1594, 1595, 2568, 1401, 2851, 1092, 534, 86, 1572, 634, 782, 1406, 3223, 19, 2230, 2480, 1962, 2218, 227, 337, 822, 1981, 1953, 1864, 749, 1952, 382, 4, 37, 1778, 3135, 648, 971, 2983, 861, 2482, 31, 1523, 59, 898, 866, 2730, 52, 29, 853, 810, 1722, 724, 1431, 171, 413, 2310, 2804, 596, 1737, 801, 937, 1326, 1330, 1349, 1350, 1369, 1609, 2244, 2245, 2373, 3099, 731, 1505, 1552, 280, 764, 1065, 910, 142, 1421, 721, 1430, 2068, 654, 397, 557, 1188, 2134, 2424, 1814, 2101, 2103, 2618, 2343, 2260, 2259, 2261, 2267, 2706, 2707, 2714, 2758, 2813, 3288, 2812, 2828, 1800, 1799, 2263, 2300, 2299, 2705, 2745, 2704, 2819, 2905, 2788, 2817, 2889, 3287, 3286, 3295, 1815, 2102, 2105, 2749, 2751, 1930, 2715, 1139, 156, 1562, 803, 671, 1538, 1730, 387, 13, 399, 2151, 424, 16, 2007, 3, 1331, 7, 863, 8, 358, 3259, 658, 1051, 1050, 1052, 1053, 1054, 1080, 1348, 2277, 604, 3271, 542, 729, 203, 1728, 1736, 610, 2344, 48, 635, 2848, 1272, 2402, 799, 800, 85, 2050, 537, 1566, 1588, 1656, 1649, 1447, 2046, 1602, 2013, 18, 1536, 1580, 1387, 1433, 1957, 1678, 2011, 2076, 2099, 1703, 1933, 1837, 2275, 1196, 1391, 1393, 1404, 1402, 2916, 1645, 1854, 1359, 1443, 1587, 1392, 1399, 1422, 1677, 1691, 1710, 2009, 1345, 2014, 1600, 1620, 2028, 2071, 1425, 1340, 1333, 1855, 2153, 1585, 1372, 1408, 1432, 2049, 1586, 2055, 1429, 1535, 1537, 1544, 1543, 1865, 1597, 1398, 1462, 1596, 1603, 1598, 1327, 2056, 2175, 2330, 1968, 1936, 1351, 2016, 1980, 1495, 2002, 2087, 1666, 1584, 2069, 2398, 1806, 1969, 1601, 1621, 1589, 1699, 1734, 1758, 2305, 2339, 1835, 1583, 1868, 1805, 1644, 2021, 2104, 727, 2057, 213, 509, 1932, 2621, 3263, 984, 344, 829, 1590, 2017, 1522. 11. Model Number/System Code (UDI); Serial Numbers: 722225 (00884838099265); 259, 55, 268, 147, 340, 72, 71, 216, 299, 291, 289, 249, 110, 298, 46, 29, 122, 233, 157, 149, 237, 167, 364, 4, 398, 30, 231, 38, 219, 170, 171, 90, 89, 77, 369, 161, 163, 137, 162, 40, 91, 74, 125, 190, 303, 377, 211, 329, 280, 86, 8, 31, 384, 87, 283, 421, 422, 267, 241. 12.Model Number/System Code (UDI); Serial Numbers: 722226 (00884838099272); 219, 120, 272, 105, 144, 319, 204, 255, 153, 350, 442, 183, 78, 467, 108, 143, 360, 555, 114, 209, 174, 687, 281, 554, 181, 147, 379, 468, 173, 271, 182, 419, 472, 122, 123, 124, 125, 126, 127, 403, 294, 194, 249, 432, 892, 87, 790, 261, 514, 766, 941, 228, 539, 950, 489, 622, 628, 637, 491, 345, 625, 93, 238, 299, 166, 364, 5, 506, 557, 227, 188, 827, 453, 883, 171, 708, 738, 655, 382, 106, 56, 196, 236, 770, 73, 302, 402, 210, 304, 916, 904, 40, 525, 653, 157, 657, 583, 540, 216, 456, 357, 596, 733, 439, 118, 263, 24, 840, 88, 224, 231, 233, 407, 598, 21, 762, 882, 505, 542, 507, 497, 485, 490, 494, 643, 644, 543, 636, 849, 633, 634, 504, 550, 551, 484, 631, 316, 639, 640, 642, 563, 547, 591, 911, 277, 278, 207, 773, 832, 501, 1000, 810. 13. Model Number/System Code (UDI); Serial Numbers: 722227 (00884838099227); 88, 135, 261, 19, 133, 134, 241, 164, 82, 32, 16, 12, 72, 64, 242, 176, 2, 75, 145, 143, 120, 112, 74, 188, 147, 146, 77, 141, 234, 14, 15, 47, 95, 163, 48, 3, 1, 123, 92, 173, 174, 121, 204, 57, 282, 8, 9. 14. Model Number/System Code (UDI); Serial Numbers: 722228 (00884838099234); 187, 328, 113, 205, 204, 150, 159, 23, 127, 24, 10, 36, 6, 213, 64, 20, 26, 47, 149, 62, 502, 18, 125, 391, 395, 229, 12, 408, 309, 310, 557, 98, 261, 179, 536, 143, 46, 70, 7, 13, 191, 259, 218, 11, 223, 370, 389, 14, 71, 39, 465, 552. 15. Model Number/System Code (UDI); Serial Numbers: 722282 ((01)00884838110571(21)8); 8.
FEI Number 3002807880
Recalling Firm/
Manufacturer
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Veenpluis 4-6
Best Netherlands
Manufacturer Reason
for Recall
Philips Azurion and Allura systems configured with a 1 Phase Uninterruptible Power Supply (UPS) have the potential of shutting down and/or not starting up, if power to the system is lost.
FDA Determined
Cause 2
Under Investigation by firm
ActionPhilips notified consignees on about 08/01/2025 via letter. Consignees were informed that systems may still be utilized in accordance with their Instructions for Use (IFU), notify all applicable personnel of this recall, maintain the Urgent Medical Device Correction letter with documentation of the system until Philips corrects the system, and establish an emergency protocol prior to all applicable diagnostics, interventional and minimally invasive procedures to manage the situation should you experience the issue during a procedure. If an affected system has been transferred to another organization, please send a copy of this Urgent Medical Device Correction letter to that organization and inform Philips about this transfer through your local Philips representative. Additionally, consignees were requested to complete and return the provided response form and report any issue experienced to Philips.
Quantity in Commerce1,861 units
DistributionWorldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, China, Colombia, Cyprus, Czech Republic, Dominican Republic, Ecuador, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Iran, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Korea, Republic of, Kosovo, Latvia, Lithuania, Macedonia, Maldives, Malta, Mexico, Moldova, Netherlands, New Caledonia, New Zealand, Nicaragua, Norway, Panama, Paraguay, Peru, Philippines, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Tunisia, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Viet Nam.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database510(K)s with Product Code = OWB
-
-