| | Class 2 Device Recall iSTAT CG8 cartridge |  |
| Date Initiated by Firm | August 21, 2025 |
| Date Posted | September 17, 2025 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-2583-2025 |
| Recall Event ID |
97443 |
| Product Classification |
Electrode measurement, blood-gases (pco2, po2) and blood ph - Product Code CHL
|
| Product | i-STAT CG8+ cartridge. List Number: 03P88-25.
510(k) numbers: K894914, K940918, K223710, K230275, K230285, K230300 |
| Code Information |
List Number: 03P88-25. All lot numbers. |
| FEI Number |
2245578
|
Recalling Firm/ Manufacturer |
Abbott Point Of Care Inc.
400 College Rd E
Princeton
NJ
08540-6607
|
| For Additional Information Contact | Ellen Wichman 224-667-8522 |
Manufacturer Reason for Recall | Lack of a 510(k) premarket clearance for the i-STAT CG8+, EG7+ and EG6+ cartridges to reflect updated sample type for potassium and ionized calcium and measurement ranges for pH and PO2. |
FDA Determined Cause 2 | No Marketing Application |
| Action | Abbott notified consignees via email and letter on about 08/21/2025. Consignees were instructed to download, review and implement the updated i-STAT CG8+, EG7+ and EG6+ cartridge Instructions for Use (IFU), discontinue use of i-STAT CG8+, EG7+ and EG6+ cartridges for testing of capillary whole blood samples for potassium and ionized calcium, and use venous or arterial whole blood or an alternate method for testing capillary potassium and ionized calcium. Update the pH and PO2 reportable ranges on the i-STAT 1 analyzers running i-STAT CG8+, EG7+ or EG6+ cartridges. As this feature is not customizable through the i STAT 1 handheld keypad, please see the i STAT DE Quick Reference Guide - Customizing Reportable Ranges (ART 770547-00 Revision A) for more information. To use the Custom Reportable Range feature, you must have the i-STAT 1 handheld and i STAT DE version 2.3 or higher. Changes to the reportable range may require a change to the Laboratory Information System (LIS) interface. Additionally, consignees were requested to confirm receipt and understanding of this communication by responding to the business reply card included with the letter. If consignees have forwarded any i-STAT CG8+, EG7+ or EG6+ cartridges to another facility, they were requested to notify them. |
| Quantity in Commerce | 8,962,450 units |
| Distribution | US Nationwide distribution. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
|
|
|