| | Class 2 Device Recall BD Pyxis Medbank |  |
| Date Initiated by Firm | September 30, 2025 |
| Date Posted | October 27, 2025 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0367-2026 |
| Recall Event ID |
97566 |
| Product Classification |
Cabinet, table and tray, anesthesia - Product Code BRY
|
| Product | Various models of BD Pyxis Medbank: Reference numbers: 1137-00
1145-00
1146-00
1147-00
1148-00
138902-01
138903-01
138905-01
138906-01
138907-01
138908-01
138909-01
138910-01
138911-01
138912-01
138913-01
138914-01
138915-01
138916-01
138917-01
138918-01
138919-01
138920-01
138921-01
138922-01
138923-01
138924-01
138926-01
138927-01
138928-01
138929-01
138930-01
138931-01
138932-01
138933-01
138934-01
138936-01
138937-01
138938-01
138939-01
138940-01
138941-01
138942-01
138943-01
138944-01
138945-01
138946-01
138947-01
138948-01
138949-01
138950-01
138951-01
138952-01
138953-01
138954-01
138955-01
138956-01
138957-01
138958-01
138959-01
138960-01
138961-01
138962-01
138963-01
138964-01
138965-01
138966-01
138967-01
138968-01
138969-01
138970-01
138971-01
138972-01
138973-01
138974-01
138975-01
138976-01
138977-01
138978-01
138989-01
138990-01
139000-01
139001-01
139002-01
139011-01
139013-01
139016-01
139018-01
139019-01
139020-01
139021-01
139022-01
139023-01
139028-01
139062-01
139063-01
139073-01
139082-01
139084-01
139085-01
139086-01
139091-01
139122-01
139125-01
139126-01
139127-01
139128-01
139131-01
139141-01
139142-01
139156-01
139163-01
139170-01
139171-01
139172-01
139174-01
139175-01
155081-01
155082-01
155083-01
155084-01
155085-01
155086-01
155087-01
155088-01
155089-01
155090-01
155091-01
155092-01
155093-01
155094-01
155095-01
155096-01
155097-01
155098-01
155099-01
155100-01
155101-01
155102-01
155103-01
155104-01
155105-01
155106-01
155107-01
155108-01
155109-01
155110-01
155111-01
155112-01
155113-01
155114-01
155115-01
155116-01
155117-01
155118-01
155119-01
155120-01
155121-01
155122-01
155123-01
155124-01
155125-01
155126-01
155127-01
155128-01
155129-01
155130-01
155131-01
155132-01
155133-01
155134-01
155135-01
155136-01
155137-01
155138-01
155139-01
155140-01
155141-01
155142-01
155143-01
155144-01
155145-01
155146-01
155147-01
155148-01
155149-01
155150-01
155151-01
155152-01
155153-01
155154-01
155155-01
155156-01
155157-01
155159-01
155160-01
155161-01
155163-01
155165-01
155166-01
155167-01
155168-01
155169-01
155171-01
155288-01
155293-01
155294-01
155299-01
155302-01
155303-01
155304-01
155305-01
155310-01
155311-01
169-03
169-04
169-05
169-06
169-07
169-08
169-09
169-10
169-100
169-101
169-102
169-103
169-104
169-105
169-106
169-107
169-108
169-109
169-11
169-110
169-111
169-112
169-113
169-114
169-115
169-12
169-120
169-124
169-125
169-127
169-128
169-129
169-13
169-131
169-132
169-137
169-138
169-139
169-14
169-140
169-141
169-142
169-143
169-144
169-145
169-15
169-150
169-151
169-152
169-155
169-156
169-159
169-16
169-160
169-161
169-162
169-163
169-164
169-18
169-19
169-20
169-21
169-22
169-23
169-24
169-25
169-26
169-27
169-28
169-29
169-30
169-31
169-32
169-33
169-34
169-35
169-36
169-37
169-38
169-39
169-40
169-41
169-42
169-43
169-44
169-45
169-46
169-47
169-48
169-49
169-50
169-51
169-52
169-53
169-54
169-55
169-56
169-57
169-58
169-59
169-60
169-61
169-62
169-63
169-64
169-65
169-66
169-67
169-68
169-69
169-70
169-71
169-72
169-73
169-74
169-75
169-76
169-77
169-78
169-79
169-80
169-81
169-82
169-83
169-84
169-85
169-86
169-87
169-88
169-89
169-90
169-91
169-92
169-93
169-94
169-95
169-96
169-97
169-98
169-99 |
| Code Information |
All Serial Numbers/Ref #/UDI: 1137-00 N/A,
1145-00 10885403517167,
1146-00 10885403517174,
1147-00 10885403517181,
1148-00 10885403517198,
138902-01 N/A,
138903-01 N/A,
138905-01 N/A,
138906-01 N/A,
138907-01 N/A,
138908-01 N/A,
138909-01 N/A,
138910-01 N/A,
138911-01 N/A,
138912-01 N/A,
138913-01 N/A,
138914-01 N/A,
138915-01 N/A,
138916-01 N/A,
138917-01 N/A,
138918-01 N/A,
138919-01 N/A,
138920-01 N/A,
138921-01 N/A,
138922-01 N/A,
138923-01 N/A,
138924-01 N/A,
138926-01 N/A,
138927-01 N/A,
138928-01 N/A,
138929-01 N/A,
138930-01 N/A,
138931-01 N/A,
138932-01 N/A,
138933-01 N/A,
138934-01 N/A,
138936-01 N/A,
138937-01 N/A,
138938-01 N/A,
138939-01 N/A,
138940-01 N/A,
138941-01 N/A,
138942-01 N/A,
138943-01 N/A,
138944-01 N/A,
138945-01 N/A,
138946-01 N/A,
138947-01 N/A,
138948-01 N/A,
138949-01 N/A,
138950-01 N/A,
138951-01 N/A,
138952-01 N/A,
138953-01 N/A,
138954-01 N/A,
138955-01 N/A,
138956-01 N/A,
138957-01 N/A,
138958-01 N/A,
138959-01 N/A,
138960-01 N/A,
138961-01 N/A,
138962-01 N/A,
138963-01 N/A,
138964-01 N/A,
138965-01 N/A,
138966-01 N/A,
138967-01 N/A,
138968-01 N/A,
138969-01 N/A,
138970-01 N/A,
138971-01 N/A,
138972-01 N/A,
138973-01 N/A,
138974-01 N/A,
138975-01 N/A,
138976-01 N/A,
138977-01 N/A,
138978-01 N/A,
138989-01 N/A,
138990-01 N/A,
139000-01 N/A,
139001-01 N/A,
139002-01 N/A,
139011-01 N/A,
139013-01 N/A,
139016-01 N/A,
139018-01 N/A,
139019-01 N/A,
139020-01 N/A,
139021-01 N/A,
139022-01 N/A,
139023-01 N/A,
139028-01 N/A,
139062-01 N/A,
139063-01 N/A,
139073-01 N/A,
139082-01 N/A,
139084-01 N/A,
139085-01 N/A,
139086-01 N/A,
139091-01 N/A,
139122-01 N/A,
139125-01 N/A,
139126-01 N/A,
139127-01 N/A,
139128-01 N/A,
139131-01 N/A,
139141-01 N/A,
139142-01 N/A,
139156-01 N/A,
139163-01 N/A,
139170-01 N/A,
139171-01 N/A,
139172-01 N/A,
139174-01 N/A,
139175-01 N/A,
155081-01 10885403503870,
155082-01 10885403503887,
155083-01 10885403503894,
155084-01 10885403503900,
155085-01 10885403503917,
155086-01 10885403503726,
155087-01 10885403503733,
155088-01 10885403503740,
155089-01 10885403503757,
155090-01 10885403503764,
155091-01 10885403503771,
155092-01 10885403503788,
155093-01 10885403503795,
155094-01 10885403503801,
155095-01 10885403503818,
155096-01 10885403503696,
155097-01 10885403504198,
155098-01 10885403503702,
155099-01 10885403504204,
155100-01 10885403504211,
155101-01 10885403504228,
155102-01 10885403504235,
155103-01 10885403504242,
155104-01 10885403504259,
155105-01 10885403503719,
155106-01 10885403504266,
155107-01 10885403504273,
155108-01 10885403504280,
155109-01 10885403504327,
155110-01 10885403504297,
155111-01 10885403504303,
155112-01 10885403504099,
155113-01 10885403504105,
155114-01 10885403504112,
155115-01 10885403504129,
155116-01 10885403504136,
155117-01 10885403504143,
155118-01 10885403504150,
155119-01 10885403504334,
155120-01 10885403504167,
155121-01 10885403504174,
155122-01 10885403504181,
155123-01 10885403504082,
155124-01 10885403504075,
155125-01 10885403503658,
155126-01 10885403503665,
155127-01 10885403503672,
155128-01 10885403503689,
155129-01 10885403504341,
155130-01 10885403504358,
155131-01 10885403504365,
155132-01 10885403504372,
155133-01 10885403503542,
155134-01 10885403503559,
155135-01 10885403503566,
155136-01 10885403503573,
155137-01 10885403503597,
155138-01 10885403503603,
155139-01 10885403503610,
155140-01 10885403504440,
155141-01 10885403503580,
155142-01 10885403504457,
155143-01 10885403504464,
155144-01 10885403504471,
155145-01 10885403504488,
155146-01 10885403504754,
155147-01 10885403504761,
155148-01 10885403504778,
155149-01 10885403504785,
155150-01 10885403504792,
155151-01 10885403504709,
155152-01 10885403504716,
155153-01 10885403504723,
155154-01 10885403504730,
155155-01 10885403504747
155156-01 10885403504624,
155157-01 10885403504631,
155159-01 10885403504655,
155160-01 10885403504662,
155161-01 10885403504679,
155163-01 10885403504686,
155165-01 10885403504693,
155166-01 10885403504570,
155167-01 10885403504587,
155168-01 10885403504594,
155169-01 10885403504600,
155171-01 10885403504617,
155288-01 10885403508721,
155293-01 10885403509926,
155294-01 10885403509933,
155299-01 10885403510526,
155302-01 10885403510762,
155303-01 10885403510779,
155304-01 10885403510786,
155305-01 10885403510793,
155310-01 10885403510830,
155311-01 10885403510847,
169-03 10885403512575,
169-04 10885403512575,
169-05 10885403512575,
169-06 10885403512575,
169-07 10885403512575,
169-08 10885403512575,
169-09 10885403512575,
169-10 10885403512575,
169-100 10885403512568,
169-101 10885403512544,
169-102 10885403512544,
169-103 10885403512544,
169-104 10885403512544,
169-105 10885403512544,
169-106 10885403512544,
169-107 10885403512544,
169-108 10885403512544,
169-109 10885403512544,
169-11 10885403512575,
169-110 10885403512544,
169-111 10885403512544,
169-112 10885403512544,
169-113 10885403512544,
169-114 10885403512544,
169-115 10885403512544,
169-12 10885403512575,
169-120 10885403512551,
169-124 10885403512551,
169-125 10885403512551,
169-127 10885403512575,
169-128 10885403512575,
169-129 10885403512575,
169-13 10885403512575,
169-131 10885403512568,
169-132 10885403512568,
169-137 10885403512544,
169-138 10885403512568,
169-139 10885403512568,
169-14 10885403512575,
169-140 10885403512568,
169-141 10885403512575,
169-142 10885403512568,
169-143 10885403512568,
169-144 10885403512568,
169-145 10885403512568,
169-15 10885403512575,
169-150 10885403512544,
169-151 10885403512544,
169-152 10885403512544,
169-155 10885403512568,
169-156 10885403512568,
169-159 10885403512544,
169-16 10885403512575,
169-160 10885403512544,
169-161 10885403512544,
169-162 10885403512544,
169-163 10885403512544,
169-164 10885403512544,
169-18 10885403512575,
169-19 10885403512575,
169-20 10885403512575,
169-21 10885403512575,
169-22 10885403512575,
169-23 10885403512568,
169-24 10885403512568,
169-25 10885403512568,
169-26 10885403512568,
169-27 10885403512568,
169-28 10885403512568,
169-29 10885403512568,
169-30 10885403512568,
169-31 10885403512568,
169-32 10885403512568,
169-33 10885403512568,
169-34 10885403512568,
169-35 10885403512568,
169-36 10885403512568,
169-37 10885403512568,
169-38 10885403512568,
169-39 10885403512568,
169-40 10885403512568,
169-41 10885403512568,
169-42 10885403512568,
169-43 10885403512568,
169-44 10885403512568,
169-45 10885403512568,
169-46 10885403512568,
169-47 10885403512568,
169-48 10885403512568,
169-49 10885403512568,
169-50 10885403512568,
169-51 10885403512568,
169-52 10885403512568,
169-53 10885403512568,
169-54 10885403512568,
169-55 10885403512568,
169-56 10885403512568
169-57 10885403512568,
169-58 10885403512568,
169-59 10885403512568,
169-60 10885403512568,
169-61 10885403512568,
169-62 10885403512568,
169-63 10885403512568,
169-64 10885403512568,
169-65 10885403512568
169-66 10885403512568,
169-67 10885403512568,
169-68 10885403512568,
169-69 10885403512568,
169-70 10885403512568,
169-71 10885403512568,
169-72 10885403512568,
169-73 10885403512568,
169-74 10885403512568,
169-75 10885403512568,
169-76 10885403512568,
169-77 10885403512568,
169-78 10885403512568,
169-79 10885403512568,
169-80 10885403512568,
169-81 10885403512568,
169-82 10885403512568,
169-83 10885403512568,
169-84 10885403512568,
169-85 10885403512568,
169-86 10885403512568,
169-87 10885403512568,
169-88 10885403512568,
169-89 10885403512568,
169-90 10885403512568,
169-91 10885403512568,
169-92 10885403512568,
169-93 10885403512568,
169-94 10885403512568,
169-95 10885403512568,
169-96 10885403512568,
169-97 10885403512568,
169-98 10885403512568,
169-99 10885403512568
|
| FEI Number |
2016493
|
Recalling Firm/ Manufacturer |
CareFusion 303, Inc.
10020 Pacific Mesa Blvd
San Diego
CA
92121-4386
|
| For Additional Information Contact | BD Technical Support 1800-727-6102 |
Manufacturer Reason for Recall | Potential fluid ingress of anesthesia station or med station may result in smoke, system downtime and/or fire. |
FDA Determined Cause 2 | Under Investigation by firm |
| Action | On September 30, 2025, BD issue a Urgent Medical Device Recall Expansion notification to affected consignees via FedEx. BD is asking consignees to take the following actions:
1. Affected products can continue to be used per existing Manuals and Guides.
2. In the event of an accidental fluid spill on the device or in a drawer, users should following instructions outlined within the customer notification.
3. Review the customer letter attachment which provides warnings, cautions and recommendations regarding fluid ingress.
4. Complete the attached Customer Response Form and return to the BD contact noted on the form whether or not you have any of the impacted material.
|
| Quantity in Commerce | 8,262 units |
| Distribution | Worldwide distribution - US Nationwide and the countries of Canada, United Arab Emirates, Australia. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
|
|
|