• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Cardiac Workstation

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Cardiac Workstationsee related information
Date Initiated by FirmSeptember 12, 2025
Date PostedOctober 16, 2025
Recall Status1 Open3, Classified
Recall NumberZ-0165-2026
Recall Event ID 97612
510(K)NumberK241556 
Product Classification Electrocardiograph - Product Code DPS
ProductCardiac Workstation 5000; Model Number: 860439;
Code Information Model Number: 860439; UDI-DI: 00884838094826; Serial Numbers:
FEI Number 3006648320
Recalling Firm/
Manufacturer
Philips North America
222 Jacobs St
Cambridge MA 02141-2296
For Additional Information ContactPhilips Customer Services
800-722-9377
Manufacturer Reason
for Recall
Failure of Environmental Stress Testing from a pinched power module wire within the housing of the Cardiac Workstation device which resulted in a short.
FDA Determined
Cause 2
Process control
ActionOn September 12, 2025 URGENT MEDICAL DEVICE CORRECTION letters were sent to customers. Actions to be taken: Share this communication with all clinical staff to review and undedrstand. Please this letter with the device. Pass the notice along to all those who need to be aware within the organization or to any organization where affected devices were transferred. Ensure technical solution established by Philips is implemented as soon as available within the timeframe communicated by Philips. In case any device cannot be corrected, inform Philips of the reason why the Technical Solution could not be implemented. Action By Philips: A Philips representative will contact customers to arrange a replacement of the device's hardware on affected devices with the pinched power module wire within the housing of the Cardiac Workstation.
Quantity in Commerce28 units
DistributionUS Nationwide distribution in the states of CA, IA, TN, TX, VA.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database510(K)s with Product Code = DPS
-
-