New Search Back to Search Results
 Class 2 Device Recall VIRATA" PREASSEMBLED POLYAXIAL SCREWSsee related information
Date Initiated by FirmSeptember 19, 2025
Date PostedNovember 07, 2025
Recall Status1 Open3, Classified
Recall NumberZ-0478-2026
Recall Event ID 97689
510(K)NumberK250908 
Product Classification Thoracolumbosacral pedicle screw system - Product Code NKB
ProductVirata Preassembled Polyaxial Screw, utilized in the Virata Spinal Fixation System. The following sizes are affected: 1. POLYAXIAL SCREW, CC, 5.5mm X 30mm; Item Number: TL1-10755030. 2. POLYAXIAL SCREW, CC, 5.5mm X 35mm; Item Number: TL1-10755035. 3. POLYAXIAL SCREW, CC, 5.5mm X 40mm; Item Number: TL1-10755040. 4. POLYAXIAL SCREW, CC, 5.5mm X 45mm; Item Number: TL1-10755045. 5. POLYAXIAL SCREW, CC, 5.5mm X 50mm; Item Number: TL1-10755050. 6. POLYAXIAL SCREW, CC, 6.5mm X 30mm; Item Number: TL1-10765030. 7. POLYAXIAL SCREW, CC, 6.5mm X 35mm; Item Number: TL1-10765035. 8. POLYAXIAL SCREW, CC, 6.5mm X 40mm; Item Number: TL1-10765040. 9. POLYAXIAL SCREW, CC, 6.5mm X 45mm; Item Number: TL1-10765045. 10. POLYAXIAL SCREW, CC, 6.5mm X 50mm; Item Number: TL1-10765050. 11. POLYAXIAL SCREW, CC, 6.5mm X 55mm; Item Number: TL1-10765055. 12. POLYAXIAL SCREW, CC, 6.5mm X 60mm; Item Number: TL1-10765060. 13. POLYAXIAL SCREW, CC, 7.5mm X 30mm; Item Number: TL1-10775030. 14. POLYAXIAL SCREW, CC, 7.5mm X 35mm; Item Number: TL1-10775035. 15. POLYAXIAL SCREW, CC, 7.5mm X 40mm; Item Number: TL1-10775040. 16. POLYAXIAL SCREW, CC, 7.5mm X 45mm; Item Number: TL1-10775045. 17. POLYAXIAL SCREW, CC, 7.5mm X 50mm; Item Number: TL1-10775050. 18. POLYAXIAL SCREW, CC, 7.5mm X 55mm; Item Number: TL1-10775055. 19. POLYAXIAL SCREW, CC, 7.5mm X 60mm; Item Number: TL1-10775060. 20. POLYAXIAL SCREW, CC, 8.5mm X 35mm; Item Number: TL1-10785035. 21. POLYAXIAL SCREW, CC, 8.5mm X 40mm; Item Number: TL1-10785040. 22. POLYAXIAL SCREW, CC, 8.5mm X 45mm; Item Number: TL1-10785045. 23. POLYAXIAL SCREW, CC, 8.5mm X 50mm; Item Number: TL1-10785050. 24. POLYAXIAL SCREW, CC, 8.5mm X 55mm; Item Number: TL1-10785055. 25. POLYAXIAL SCREW, CC, 8.5mm X 60mm; Item Number: TL1-10785060.
Code Information 1. Item Number: TL1-10755030. UDI-DI: 10889981335229. Lot Number: JP2572B. 2. Item Number: TL1-10755035. UDI-DI: 10889981336165. Lot Number: JP2757B. 3. Item Number: TL1-10755040. UDI-DI: 10889981335717. Lot Number: JP2758B. 4. Item Number: TL1-10755045. UDI-DI: 10889981335953. Lot Numbers: JP2759B, JP2799B, JP2800B, JP2880B. 5. Item Number: TL1-10755050. UDI-DI: 10889981336318. Lot Numbers: JP2760B, JP2801B. 6. Item Number: TL1-10765030. UDI-DI: 10889981335885. Lot Number: JP2655B. 7. Item Number: TL1-10765035. UDI-DI: 10889981335540. Lot Number: JP2656B. 8. Item Number: TL1-10765040. UDI-DI: 10889981336332. Lot Number: JP2657B, JP2791B. 9. Item Number: TL1-10765045. UDI-DI: 10889981335526. Lot Numbers: JP2658B, JP2659B, JP2792B. 10. Item Number: TL1-10765050. UDI-DI: 10889981335342. Lot Numbers: JP2660B, JP2794B, JP2795B. 11. Item Number: TL1-10765055. UDI-DI: 10889981335168. Lot Number: JP2661B. 12. Item Number: . UDI-DI: . Lot Number: Item Number: TL1-10765060. UDI-DI: 10889981335816. Lot Number: JP2662B. 13. Item Number: TL1-10775030. UDI-DI: 10889981335427. Lot Number: JP2663B. 14. Item Number: TL1-10775035. UDI-DI: 10889981335632. Lot Number: JP2668B. 15. Item Number: TL1-10775040. UDI-DI: 10889981335113. Lot Number: JP2669B. 16. Item Number: TL1-10775045. UDI-DI: 10889981335403. Lot Number: JP2670B. 17. Item Number: TL1-10775050. UDI-DI: 10889981335830. Lot Number: JP2671B. 18. Item Number: TL1-10775055. UDI-DI: 10889981335250. Lot Numbers: JP2672B, JP2887B. 19. Item Number: TL1-10775060. UDI-DI: 10889981335076. Lot Number: JP2673B. 20. Item Number: TL1-10785035. UDI-DI: 10889981335410. Lot Number: JP2674B. 21. Item Number: TL1-10785040. UDI-DI: 10889981335779. Lot Number: JP2675B. 22. Item Number: TL1-10785045. UDI-DI: 10889981335243. Lot Number: JP2664B. 23. Item Number: TL1-10785050. UDI-DI: 10889981335847. Lot Number: JP2665B. 24. Item Number: TL1-10785055. UDI-DI: 10889981335106. Lot Number: JP2666B. 25. Item Number: TL1-10785060. UDI-DI: 10889981334949. Lot Number: JP2667B.
FEI Number 2183449
Recalling Firm/
Manufacturer
Orthofix U.S. LLC
3451 Plano Pkwy
Lewisville TX 75056-9453
For Additional Information ContactSeaSpine Complaints
214-937-2000
Manufacturer Reason
for Recall
Polyaxial screws do not meet performance standards.
FDA Determined
Cause 2
Process change control
ActionOrthofix notified consignees via email on 09/19/2025. Consignees were instructed to immediately discontinue use of the affected devices, quarantine all affected inventory, and return units to Orthofix using the provided Return Goods Authorization (RGA) process. Surgeons were instructed to monitor affected patients according to standard clinical practice. Customers were advised not to implant or distribute any remaining affected product. All consignees were requested to complete and return the response form.
Quantity in Commerce624 units
DistributionUS Nationwide distribution in the states of California, North Carolina, and Texas.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database510(K)s with Product Code = NKB
-
-