| Date Initiated by Firm | October 27, 2025 |
| Date Posted | December 09, 2025 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0597-2026 |
| Recall Event ID |
98052 |
| 510(K)Number | K200917 |
| Product Classification |
System, x-ray, fluoroscopic, image-intensified - Product Code JAA
|
| Product | Azurion R3.0.
Model Number 722234. Commercial Name: Azurion 7 M20. INTERVENTIONAL FLUOROSCOPIC XRAY SYSTEM |
| Code Information |
Model Number 722234
Commercial Name: Azurion 7 M20
How to identify: Philips Reference C&R 2025-IGT-BST-016 |
| FEI Number |
3006648320
|
Recalling Firm/ Manufacturer |
Philips North America
222 Jacobs St
Cambridge
MA
02141-2296
|
| For Additional Information Contact | 978-482-2309 |
Manufacturer Reason for Recall | Between July 2024 and June 2025, Philips Azurion R3.0 systems with a FlexArm
stand were shipped without the Source-to-Skin Distance (SSD) Spacer. The SSD spacer is necessary to meet the FDA minimum source-skin distance requirement of 38 cm for standard fluoroscopy applications. |
FDA Determined Cause 2 | Radiation Control for Health and Safety Act |
| Action | Philips will send a customer notification to the users of the affected systems to alert them of the issue
and provide the actions that should be taken to prevent risks to patients.
Philips will contact customers to schedule a time for a Philips Field Service Engineer (FSE) to visit their site and install the SSD spacer to resolve the issue. |
| Quantity in Commerce | 110 |
| Distribution | U.S. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = JAA
|