| | Class 2 Device Recall Philips Azurion system |  |
| Date Initiated by Firm | December 08, 2025 |
| Date Posted | January 14, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-1079-2026 |
| Recall Event ID |
98071 |
| 510(K)Number | K200917 |
| Product Classification |
Interventional fluoroscopic x-ray system - Product Code OWB
|
| Product | Philips Azurion Systems R3.0. Includes the following label descriptions and corresponding models:
1. Azurion 3 M12. Model Number: 722229.
2. Azurion 3 M15. Model Number: 722230.
3. Azurion 5 M12. Model Number: 722231.
4. Azurion 5 M20. Model Number: 722232.
5. Azurion 7 B12. Model Number: 722235.
6. Azurion 7 B20. Model Number: 722236.
7. Azurion 7 M12. Model Number: 722233.
8. Azurion 7 M20. Model Number: 722234. |
| Code Information |
Software Version: R3.0.
1. Model Number: 722229. UDI Numbers: (01)00884838116726(21)22, (01)00884838116726(21)15, (01)00884838116726(21)16, (01)00884838116726(21)6, (01)00884838116726(21)8, (01)00884838116726(21)4, (01)00884838116726(21)24, (01)00884838116726(21)26, (01)00884838116726(21)20, (01)00884838116726(21)25, (01)00884838116726(21)11, (01)00884838116726(21)21, (01)00884838116726(21)13, (01)00884838116726(21)7, (01)00884838116726(21)10, (01)00884838116726(21)2, (01)00884838116726(21)3, (01)00884838116726(21)29.
2. Model Number: 722230. UDI Numbers: (01)00884838116733(21)2, (01)00884838116733(21)40, (01)00884838116733(21)9, (01)00884838116733(21)1, (01)00884838116733(21)38, (01)00884838116733(21)10, (01)00884838116733(21)45, (01)00884838116733(21)7, (01)00884838116733(21)27, (01)00884838116733(21)28, (01)00884838116733(21)39, (01)00884838116733(21)4, (01)00884838116733(21)5, (01)00884838116733(21)23, (01)00884838116733(21)41, (01)00884838116733(21)20, (01)00884838116733(21)19, (01)00884838116733(21)13, (01)00884838116733(21)15, (01)00884838116733(21)12, (01)00884838116733(21)22.
3. Model Number: 722231. UDI Numbers: (01)00884838116740(21)13, (01)00884838116740(21)29, (01)00884838116740(21)50, (01)00884838116740(21)21, (01)00884838116740(21)19, (01)00884838116740(21)12, (01)00884838116740(21)25, (01)00884838116740(21)4, (01)00884838116740(21)1, (01)00884838116740(21)5, (01)00884838116740(21)2, (01)00884838116740(21)73, (01)00884838116740(21)72, (01)00884838116740(21)75, (01)00884838116740(21)33, (01)00884838116740(21)47, (01)00884838116740(21)60, (01)00884838116740(21)53, (01)00884838116740(21)45, (01)00884838116740(21)61, (01)00884838116740(21)34, (01)00884838116740(21)54, (01)00884838116740(21)35, (01)00884838116740(21)49, (01)00884838116740(21)32, (01)00884838116740(21)28, (01)00884838116740(21)27, (01)00884838116740(21)26, (01)00884838116740(21)51, (01)00884838116740(21)11, (01)00884838116740(21)20, (01)00884838116740(21)41, (01)00884838116740(21)52, (01)00884838116740(21)3.
4. Model Number: 722232. UDI Numbers: (01)00884838116757(21)61, (01)00884838116757(21)60, (01)00884838116757(21)5, (01)00884838116757(21)49, (01)00884838116757(21)2, (01)00884838116757(21)91, (01)00884838116757(21)33, (01)00884838116757(21)34, (01)00884838116757(21)48, (01)00884838116757(21)46, (01)00884838116757(21)23, (01)00884838116757(21)43, (01)00884838116757(21)36, (01)00884838116757(21)96, (01)00884838116757(21)95, (01)00884838116757(21)10, (01)00884838116757(21)76, (01)00884838116757(21)42, (01)00884838116757(21)6, (01)00884838116757(21)77, (01)00884838116757(21)3, (01)00884838116757(21)1, (01)00884838116757(21)56, (01)00884838116757(21)51, (01)00884838116757(21)44, (01)00884838116757(21)30, (01)00884838116757(21)29, (01)00884838116757(21)28, (01)00884838116757(21)102, (01)00884838116757(21)19, (01)00884838116757(21)8, (01)00884838116757(21)81, (01)00884838116757(21)20, (01)00884838116757(21)16, (01)00884838116757(21)11, (01)00884838116757(21)112, (01)00884838116757(21)63, (01)00884838116757(21)68, (01)00884838116757(21)32, (01)00884838116757(21)35, (01)00884838116757(21)122, (01)00884838116757(21)14, (01)00884838116757(21)82, (01)00884838116757(21)13, (01)00884838116757(21)50, (01)00884838116757(21)57, (01)00884838116757(21)58, (01)00884838116757(21)4, (01)00884838116757(21)105, (01)00884838116757(21)75, (01)00884838116757(21)121, (01)00884838116757(21)120, (01)00884838116757(21)71, (01)00884838116757(21)72, (01)00884838116757(21)73, (01)00884838116757(21)70, (01)00884838116757(21)69, (01)00884838116757(21)37.
5. Model Number: 722235. UDI Numbers: (01)00884838116788(21)88, (01)00884838116788(21)43, (01)00884838116788(21)12, (01)00884838116788(21)102, (01)00884838116788(21)127, (01)00884838116788(21)57, (01)00884838116788(21)55, (01)00884838116788(21)11, (01)00884838116788(21)76, (01)00884838116788(21)71, (01)00884838116788(21)59, (01)00884838116788(21)77, (01)00884838116788(21)89, (01)00884838116788(21)56, (01)00884838116788(21)99, (01)00884838116788(21)30, (01)00884838116788(21)84, (01)00884838116788(21)82, (01)00884838116788(21)83, (01)00884838116788(21)93, (01)00884838116788(21)31, (01)00884838116788(21)47, (01)00884838116788(21)3, (01)00884838116788(21)13, (01)00884838116788(21)10, (01)00884838116788(21)74, (01)00884838116788(21)98, (01)00884838116788(21)40, (01)00884838116788(21)81, (01)00884838116788(21)66, (01)00884838116788(21)63, (01)00884838116788(21)67, (01)00884838116788(21)64, (01)00884838116788(21)42, (01)00884838116788(21)65, (01)00884838116788(21)58, (01)00884838116788(21)53, (01)00884838116788(21)52, (01)00884838116788(21)49, (01)00884838116788(21)44, (01)00884838116788(21)35, (01)00884838116788(21)45, (01)00884838116788(21)39, (01)00884838116788(21)25, (01)00884838116788(21)34, (01)00884838116788(21)17, (01)00884838116788(21)24, (01)00884838116788(21)19, (01)00884838116788(21)33, (01)00884838116788(21)18, (01)00884838116788(21)9, (01)00884838116788(21)28, (01)00884838116788(21)8, (01)00884838116788(21)23, (01)00884838116788(21)21, (01)00884838116788(21)6, (01)00884838116788(21)7, (01)00884838116788(21)5, (01)00884838116788(21)108, (01)00884838116788(21)109, (01)00884838116788(21)107, (01)00884838116788(21)106, (01)00884838116788(21)4, (01)00884838116788(21)41, (01)00884838116788(21)72, (01)00884838116788(21)70, (01)00884838116788(21)54, (01)00884838116788(21)20, (01)00884838116788(21)14, (01)00884838116788(21)126, (01)00884838116788(21)105, (01)00884838116788(21)94.
6. Model Number: 722236. UDI Numbers: (01)00884838116801(21)43, (01)00884838116801(21)42, (01)00884838116801(21)121, (01)00884838116801(21)50, (01)00884838116801(21)183, (01)00884838116801(21)32, (01)00884838116801(21)38, (01)00884838116801(21)197, (01)00884838116801(21)335, (01)00884838116801(21)325, (01)00884838116801(21)116, (01)00884838116801(21)191, (01)00884838116801(21)235, (01)00884838116801(21)22, (01)00884838116801(21)204, (01)00884838116801(21)249, (01)00884838116801(21)159, (01)00884838116801(21)40, (01)00884838116801(21)233, (01)00884838116801(21)194, (01)00884838116801(21)205, (01)00884838116801(21)276, (01)00884838116801(21)378, (01)00884838116801(21)20, (01)00884838116801(21)81, (01)00884838116801(21)51, (01)00884838116801(21)288, (01)00884838116801(21)25, (01)00884838116801(21)54, (01)00884838116801(21)186, (01)00884838116801(21)21, (01)00884838116801(21)68, (01)00884838116801(21)99, (01)00884838116801(21)208, (01)00884838116801(21)31, (01)00884838116801(21)230, (01)00884838116801(21)203, (01)00884838116801(21)180, (01)00884838116801(21)37, (01)00884838116801(21)342, (01)00884838116801(21)24, (01)00884838116801(21)221, (01)00884838116801(21)17, (01)00884838116801(21)252, (01)00884838116801(21)150, (01)00884838116801(21)60, (01)00884838116801(21)200, (01)00884838116801(21)226, (01)00884838116801(21)118, (01)00884838116801(21)127, (01)00884838116801(21)26, (01)00884838116801(21)36, (01)00884838116801(21)133, (01)00884838116801(21)345, (01)00884838116801(21)134, (01)00884838116801(21)131, (01)00884838116801(21)39, (01)00884838116801(21)231, (01)00884838116801(21)158, (01)00884838116801(21)48, (01)00884838116801(21)227, (01)00884838116801(21)148, (01)00884838116801(21)147, (01)00884838116801(21)272, (01)00884838116801(21)273, (01)00884838116801(21)100, (01)00884838116801(21)224, (01)00884838116801(21)164, (01)00884838116801(21)163, (01)00884838116801(21)170, (01)00884838116801(21)44, (01)00884838116801(21)46, (01)00884838116801(21)113, (01)00884838116801(21)216, (01)00884838116801(21)278, (01)00884838116801(21)102, (01)00884838116801(21)8, (01)00884838116801(21)169, (01)00884838116801(21)338, (01)00884838116801(21)117, (01)00884838116801(21)29, (01)00884838116801(21)34, (01)00884838116801(21)10, (01)00884838116801(21)126, (01)00884838116801(21)237, (01)00884838116801(21)125, (01)00884838116801(21)154, (01)00884838116801(21)157, (01)00884838116801(21)30, (01)00884838116801(21)67, (01)00884838116801(21)292, (01)00884838116801(21)289, (01)00884838116801(21)277.(01)00884838116801(21)244, (01)00884838116801(21)181, (01)00884838116801(21)264, (01)00884838116801(21)232, (01)00884838116801(21)207, (01)00884838116801(21)192, (01)00884838116801(21)132, (01)00884838116801(21)128, (01)00884838116801(21)160, (01)00884838116801(21)94, (01)00884838116801(21)86, (01)00884838116801(21)41, (01)00884838116801(21)27, (01)00884838116801(21)259, (01)00884838116801(21)251, (01)00884838116801(21)261, (01)00884838116801(21)57, (01)00884838116801(21)240, (01)00884838116801(21)107, (01)00884838116801(21)172, (01)00884838116801(21)142, (01)00884838116801(21)109, (01)00884838116801(21)55, (01)00884838116801(21)171, (01)00884838116801(21)155, (01)00884838116801(21)175, (01)00884838116801(21)111, (01)00884838116801(21)88, (01)00884838116801(21)104, (01)00884838116801(21)152, (01)00884838116801(21)124, (01)00884838116801(21)103, (01)00884838116801(21)90, (01)00884838116801(21)70, (01)00884838116801(21)153, (01)00884838116801(21)106, (01)00884838116801(21)151, (01)00884838116801(21)144, (01)00884838116801(21)141, (01)00884838116801(21)76, (01)00884838116801(21)112, (01)00884838116801(21)105, (01)00884838116801(21)85, (01)00884838116801(21)62, (01)00884838116801(21)52, (01)00884838116801(21)75, (01)00884838116801(21)92, (01)00884838116801(21)89, (01)00884838116801(21)108, (01)00884838116801(21)59, (01)00884838116801(21)73, (01)00884838116801(21)84, (01)00884838116801(21)69, (01)00884838116801(21)83, (01)00884838116801(21)80, (01)00884838116801(21)64, (01)00884838116801(21)74, (01)00884838116801(21)53, (01)00884838116801(21)49, (01)00884838116801(21)329, (01)00884838116801(21)130, (01)00884838116801(21)326, (01)00884838116801(21)168, (01)00884838116801(21)145, (01)00884838116801(21)146, (01)00884838116801(21)98, (01)00884838116801(21)96, (01)00884838116801(21)93, (01)00884838116801(21)258, (01)00884838116801(21)56, (01)00884838116801(21)91, (01)00884838116801(21)271, (01)00884838116801(21)201, (01)00884838116801(21)123, (01)00884838116801(21)162, (01)00884838116801(21)122, (01)00884838116801(21)114, (01)00884838116801(21)63, (01)00884838116801(21)374, (01)00884838116801(21)357, (01)00884838116801(21)348, (01)00884838116801(21)347, (01)00884838116801(21)352, (01)00884838116801(21)322, (01)00884838116801(21)330, (01)00884838116801(21)321, (01)00884838116801(21)188, (01)00884838116801(21)283, (01)00884838116801(21)285, (01)00884838116801(21)284, (01)00884838116801(21)269, (01)00884838116801(21)193, (01)00884838116801(21)173, (01)00884838116801(21)311, (01)00884838116801(21)156, (01)00884838116801(21)219, (01)00884838116801(21)167, (01)00884838116801(21)33, (01)00884838116801(21)268, (01)00884838116801(21)23, (01)00884838116801(21)361, (01)00884838116801(21)242, (01)00884838116801(21)223, (01)00884838116801(21)234, (01)00884838116801(21)182, (01)00884838116801(21)189, (01)00884838116801(21)165, (01)00884838116801(21)82, (01)00884838116801(21)66, (01)00884838116801(21)13.
7. Model Number: 722233. UDI Numbers: (01)00884838116764(21)118, (01)00884838116764(21)96, (01)00884838116764(21)259, (01)00884838116764(21)169, (01)00884838116764(21)108, (01)00884838116764(21)212, (01)00884838116764(21)267, (01)00884838116764(21)214, (01)00884838116764(21)147, (01)00884838116764(21)153, (01)00884838116764(21)111, (01)00884838116764(21)122, (01)00884838116764(21)124, (01)00884838116764(21)127, (01)00884838116764(21)59, (01)00884838116764(21)172, (01)00884838116764(21)64, (01)00884838116764(21)170, (01)00884838116764(21)41, (01)00884838116764(21)103, (01)00884838116764(21)144, (01)00884838116764(21)219, (01)00884838116764(21)218, (01)00884838116764(21)192, (01)00884838116764(21)30, (01)00884838116764(21)189, (01)00884838116764(21)186, (01)00884838116764(21)17, (01)00884838116764(21)9, (01)00884838116764(21)245, (01)00884838116764(21)57, (01)00884838116764(21)32, (01)00884838116764(21)203, (01)00884838116764(21)205, (01)00884838116764(21)204, (01)00884838116764(21)184, (01)00884838116764(21)162, (01)00884838116764(21)18, (01)00884838116764(21)83, (01)00884838116764(21)84, (01)00884838116764(21)88, (01)00884838116764(21)76, (01)00884838116764(21)137, (01)00884838116764(21)85, (01)00884838116764(21)86, (01)00884838116764(21)87, (01)00884838116764(21)46, (01)00884838116764(21)47, (01)00884838116764(21)26, (01)00884838116764(21)234, (01)00884838116764(21)200, (01)00884838116764(21)250, (01)00884838116764(21)201, (01)00884838116764(21)123, (01)00884838116764(21)235, (01)00884838116764(21)68, (01)00884838116764(21)14, (01)00884838116764(21)82, (01)00884838116764(21)27, (01)00884838116764(21)38, (01)00884838116764(21)37, (01)00884838116764(21)8, (01)00884838116764(21)145, (01)00884838116764(21)112, (01)00884838116764(21)95, (01)00884838116764(21)173, (01)00884838116764(21)202, (01)00884838116764(21)101, (01)00884838116764(21)165, (01)00884838116764(21)166, (01)00884838116764(21)97, (01)00884838116764(21)81, (01)00884838116764(21)61, (01)00884838116764(21)31, (01)00884838116764(21)29, (01)00884838116764(21)109, (01)00884838116764(21)279, (01)00884838116764(21)260, (01)00884838116764(21)221, (01)00884838116764(21)178, (01)00884838116764(21)181, (01)00884838116764(21)159, (01)00884838116764(21)117, (01)00884838116764(21)141, (01)00884838116764(21)120, (01)00884838116764(21)273, (01)00884838116764(21)163, (01)00884838116764(21)272, (01)00884838116764(21)174, (01)00884838116764(21)249, (01)00884838116764(21)78, (01)00884838116764(21)133, (01)00884838116764(21)102, (01)00884838116764(21)106, (01)00884838116764(21)104, (01)00884838116764(21)80, (01)00884838116764(21)77, (01)00884838116764(21)206, (01)00884838116764(21)67, (01)00884838116764(21)39, (01)00884838116764(21)15, (01)00884838116764(21)13, (01)00884838116764(21)179, (01)00884838116764(21)180, (01)00884838116764(21)136, (01)00884838116764(21)129, (01)00884838116764(21)222, (01)00884838116764(21)149, (01)00884838116764(21)148, (01)00884838116764(21)146, (01)00884838116764(21)5, (01)00884838116764(21)51, (01)00884838116764(21)19, (01)00884838116764(21)7, (01)00884838116764(21)6, (01)00884838116764(21)231, (01)00884838116764(21)107, (01)00884838116764(21)91, (01)00884838116764(21)89, (01)00884838116764(21)90, (01)00884838116764(21)71, (01)00884838116764(21)48, (01)00884838116764(21)74, (01)00884838116764(21)62, (01)00884838116764(21)55, (01)00884838116764(21)50, (01)00884838116764(21)43, (01)00884838116764(21)40, (01)00884838116764(21)11, (01)00884838116764(21)35, (01)00884838116764(21)60, (01)00884838116764(21)28, (01)00884838116764(21)20, (01)00884838116764(21)12, (01)00884838116764(21)138, (01)00884838116764(21)139, (01)00884838116764(21)58, (01)00884838116764(21)257, (01)00884838116764(21)160, (01)00884838116764(21)24, (01)00884838116764(21)268, (01)00884838116764(21)284, (01)00884838116764(21)291, (01)00884838116764(21)238, (01)00884838116764(21)225, (01)00884838116764(21)217, (01)00884838116764(21)211, (01)00884838116764(21)16, (01)00884838116764(21)110, (01)00884838116764(21)183, (01)00884838116764(21)143, (01)00884838116764(21)134, (01)00884838116764(21)44, (01)00884838116764(21)197.
8. Model Number: 722234. UDI Numbers: (01)00884838116771(21)626, (01)00884838116771(21)320, (01)00884838116771(21)256, (01)00884838116771(21)321, (01)00884838116771(21)437, (01)00884838116771(21)101, (01)00884838116771(21)616, (01)00884838116771(21)418, (01)00884838116771(21)426, (01)00884838116771(21)87, (01)00884838116771(21)420, (01)00884838116771(21)827, (01)00884838116771(21)716, (01)00884838116771(21)607, (01)00884838116771(21)183, (01)00884838116771(21)345, (01)00884838116771(21)520, (01)00884838116771(21)662, (01)00884838116771(21)327, (01)00884838116771(21)524, (01)00884838116771(21)191, (01)00884838116771(21)201, (01)00884838116771(21)449, (01)00884838116771(21)435, (01)00884838116771(21)657, (01)00884838116771(21)171, (01)00884838116771(21)124, (01)00884838116771(21)123, (01)00884838116771(21)79, (01)00884838116771(21)409, (01)00884838116771(21)70, (01)00884838116771(21)55, (01)00884838116771(21)68, (01)00884838116771(21)780, (01)00884838116771(21)794, (01)00884838116771(21)500, (01)00884838116771(21)450, (01)00884838116771(21)471, (01)00884838116771(21)459, (01)00884838116771(21)489, (01)00884838116771(21)617, (01)00884838116771(21)874, (01)00884838116771(21)92, (01)00884838116771(21)698, (01)00884838116771(21)758, (01)00884838116771(21)585, (01)00884838116771(21)300, (01)00884838116771(21)502, (01)00884838116771(21)216, (01)00884838116771(21)493, (01)00884838116771(21)358, (01)00884838116771(21)85, (01)00884838116771(21)274, (01)00884838116771(21)804, (01)00884838116771(21)484, (01)00884838116771(21)519, (01)00884838116771(21)346, (01)00884838116771(21)469, (01)00884838116771(21)853, (01)00884838116771(21)852, (01)00884838116771(21)60, (01)00884838116771(21)59, (01)00884838116771(21)94, (01)00884838116771(21)485, (01)00884838116771(21)813, (01)00884838116771(21)433, (01)00884838116771(21)432, (01)00884838116771(21)168, (01)00884838116771(21)223, (01)00884838116771(21)67, (01)00884838116771(21)492, (01)00884838116771(21)71, (01)00884838116771(21)535, (01)00884838116771(21)77, (01)00884838116771(21)288, (01)00884838116771(21)383, (01)00884838116771(21)332, (01)00884838116771(21)762, (01)00884838116771(21)284, (01)00884838116771(21)620, (01)00884838116771(21)76, (01)00884838116771(21)232, (01)00884838116771(21)286, (01)00884838116771(21)267, (01)00884838116771(21)269, (01)00884838116771(21)619, (01)00884838116771(21)260, (01)00884838116771(21)162, (01)00884838116771(21)84, (01)00884838116771(21)161, (01)00884838116771(21)165, (01)00884838116771(21)107, (01)00884838116771(21)233, (01)00884838116771(21)419, (01)00884838116771(21)515, (01)00884838116771(21)212, (01)00884838116771(21)248, (01)00884838116771(21)655, (01)00884838116771(21)674, (01)00884838116771(21)139, (01)00884838116771(21)348, (01)00884838116771(21)586, (01)00884838116771(21)252, (01)00884838116771(21)633, (01)00884838116771(21)689, (01)00884838116771(21)395, (01)00884838116771(21)560, (01)00884838116771(21)263, (01)00884838116771(21)257, (01)00884838116771(21)194, (01)00884838116771(21)403, (01)00884838116771(21)331, (01)00884838116771(21)262, (01)00884838116771(21)362, (01)00884838116771(21)175, (01)00884838116771(21)580, (01)00884838116771(21)877, (01)00884838116771(21)186, (01)00884838116771(21)329, (01)00884838116771(21)266, (01)00884838116771(21)372, (01)00884838116771(21)155, (01)00884838116771(21)779, (01)00884838116771(21)696, (01)00884838116771(21)352, (01)00884838116771(21)667, (01)00884838116771(21)797, (01)00884838116771(21)217, (01)00884838116771(21)509, (01)00884838116771(21)675, (01)00884838116771(21)335, (01)00884838116771(21)347, (01)00884838116771(21)330, (01)00884838116771(21)292, (01)00884838116771(21)43, (01)00884838116771(21)225, (01)00884838116771(21)442, (01)00884838116771(21)322, (01)00884838116771(21)726, (01)00884838116771(21)209, (01)00884838116771(21)333, (01)00884838116771(21)475, (01)00884838116771(21)163, (01)00884838116771(21)624, (01)00884838116771(21)628, (01)00884838116771(21)629, (01)00884838116771(21)108, (01)00884838116771(21)198, (01)00884838116771(21)462, (01)00884838116771(21)768, (01)00884838116771(21)222, (01)00884838116771(21)730, (01)00884838116771(21)287, (01)00884838116771(21)174, (01)00884838116771(21)606, (01)00884838116771(21)97, (01)00884838116771(21)270, (01)00884838116771(21)255, (01)00884838116771(21)88, (01)00884838116771(21)359, (01)00884838116771(21)208, (01)00884838116771(21)102, (01)00884838116771(21)207, (01)00884838116771(21)463, (01)00884838116771(21)711, (01)00884838116771(21)340, (01)00884838116771(21)465, (01)00884838116771(21)140, (01)00884838116771(21)774, (01)00884838116771(21)119, (01)00884838116771(21)565, (01)00884838116771(21)413, (01)00884838116771(21)772, (01)00884838116771(21)54, (01)00884838116771(21)511, (01)00884838116771(21)510, (01)00884838116771(21)483, (01)00884838116771(21)547, (01)00884838116771(21)516, (01)00884838116771(21)447, (01)00884838116771(21)638, (01)00884838116771(21)714, (01)00884838116771(21)185, (01)00884838116771(21)253, (01)00884838116771(21)501, (01)00884838116771(21)178, (01)00884838116771(21)467, (01)00884838116771(21)98, (01)00884838116771(21)404, (01)00884838116771(21)554, (01)00884838116771(21)411, (01)00884838116771(21)344, (01)00884838116771(21)460, (01)00884838116771(21)309, (01)00884838116771(21)133, (01)00884838116771(21)122, (01)00884838116771(21)159, (01)00884838116771(21)461, (01)00884838116771(21)640, (01)00884838116771(21)639, (01)00884838116771(21)99, (01)00884838116771(21)803, (01)00884838116771(21)641, (01)00884838116771(21)513, (01)00884838116771(21)416, (01)00884838116771(21)430, (01)00884838116771(21)514, (01)00884838116771(21)319, (01)00884838116771(21)334, (01)00884838116771(21)181, (01)00884838116771(21)82, (01)00884838116771(21)281, (01)00884838116771(21)508, (01)00884838116771(21)497, (01)00884838116771(21)775, (01)00884838116771(21)873, (01)00884838116771(21)490, (01)00884838116771(21)498, (01)00884838116771(21)782, (01)00884838116771(21)792, (01)00884838116771(21)826, (01)00884838116771(21)295, (01)00884838116771(21)294, (01)00884838116771(21)243, (01)00884838116771(21)114, (01)00884838116771(21)42, (01)00884838116771(21)21, (01)00884838116771(21)399, (01)00884838116771(21)192, (01)00884838116771(21)12, (01)00884838116771(21)13, (01)00884838116771(21)567, (01)00884838116771(21)572, (01)00884838116771(21)566, (01)00884838116771(21)264, (01)00884838116771(21)307, (01)00884838116771(21)308, (01)00884838116771(21)443, (01)00884838116771(21)305, (01)00884838116771(21)304, (01)00884838116771(21)296, (01)00884838116771(21)446, (01)00884838116771(21)187, (01)00884838116771(21)249, (01)00884838116771(21)190, (01)00884838116771(21)200, (01)00884838116771(21)341, (01)00884838116771(21)448, (01)00884838116771(21)558, (01)00884838116771(21)679, (01)00884838116771(21)598, (01)00884838116771(21)385, (01)00884838116771(21)247, (01)00884838116771(21)755, (01)00884838116771(21)205, (01)00884838116771(21)737, (01)00884838116771(21)757, (01)00884838116771(21)582, (01)00884838116771(21)625, (01)00884838116771(21)687, (01)00884838116771(21)401, (01)00884838116771(21)486, (01)00884838116771(21)36, (01)00884838116771(21)37, (01)00884838116771(21)46, (01)00884838116771(21)478, (01)00884838116771(21)479, (01)00884838116771(21)632, (01)00884838116771(21)480, (01)00884838116771(21)242, (01)00884838116771(21)316, (01)00884838116771(21)195, (01)00884838116771(21)250, (01)00884838116771(21)93, (01)00884838116771(21)95, (01)00884838116771(21)18, (01)00884838116771(21)111, (01)00884838116771(21)840, (01)00884838116771(21)650, (01)00884838116771(21)795, (01)00884838116771(21)589, (01)00884838116771(21)754, (01)00884838116771(21)634, (01)00884838116771(21)424, (01)00884838116771(21)635, (01)00884838116771(21)717, (01)00884838116771(21)665, (01)00884838116771(21)491, (01)00884838116771(21)523, (01)00884838116771(21)614, (01)00884838116771(21)439, (01)00884838116771(21)412, (01)00884838116771(21)438, (01)00884838116771(21)445, (01)00884838116771(21)408, (01)00884838116771(21)324, (01)00884838116771(21)645, (01)00884838116771(21)534, (01)00884838116771(21)821, (01)00884838116771(21)482, (01)00884838116771(21)339, (01)00884838116771(21)360, (01)00884838116771(21)677, (01)00884838116771(21)142, (01)00884838116771(21)724, (01)00884838116771(21)505, (01)00884838116771(21)604, (01)00884838116771(21)303, (01)00884838116771(21)364, (01)00884838116771(21)73, (01)00884838116771(21)313, (01)00884838116771(21)374, (01)00884838116771(21)365, (01)00884838116771(21)406, (01)00884838116771(21)388, (01)00884838116771(21)387, (01)00884838116771(21)378, (01)00884838116771(21)239, (01)00884838116771(21)241, (01)00884838116771(21)245, (01)00884838116771(21)229, (01)00884838116771(21)135, (01)00884838116771(21)127, (01)00884838116771(21)47, (01)00884838116771(21)58, (01)00884838116771(21)57, (01)00884838116771(21)41, (01)00884838116771(21)19, (01)00884838116771(21)279, (01)00884838116771(21)152, (01)00884838116771(21)301, (01)00884838116771(21)33, (01)00884838116771(21)156, (01)00884838116771(21)197, (01)00884838116771(21)26, (01)00884838116771(21)31, (01)00884838116771(21)32, (01)00884838116771(21)20, (01)00884838116771(21)29, (01)00884838116771(21)158, (01)00884838116771(21)481, (01)00884838116771(21)275, (01)00884838116771(21)45, (01)00884838116771(21)468, (01)00884838116771(21)425, (01)00884838116771(21)361, (01)00884838116771(21)219, (01)00884838116771(21)39, (01)00884838116771(21)27, (01)00884838116771(21)16, (01)00884838116771(21)721, (01)00884838116771(21)78, (01)00884838116771(21)338, (01)00884838116771(21)235, (01)00884838116771(21)236, (01)00884838116771(21)237, (01)00884838116771(21)115, (01)00884838116771(21)160, (01)00884838116771(21)881, (01)00884838116771(21)690, (01)00884838116771(21)23, (01)00884838116771(21)132, (01)00884838116771(21)189, (01)00884838116771(21)44, (01)00884838116771(21)52, (01)00884838116771(21)51, (01)00884838116771(21)38, (01)00884838116771(21)148, (01)00884838116771(21)49, (01)00884838116771(21)91, (01)00884838116771(21)611, (01)00884838116771(21)120, (01)00884838116771(21)177, (01)00884838116771(21)678, (01)00884838116771(21)596, (01)00884838116771(21)850, (01)00884838116771(21)839, (01)00884838116771(21)504, (01)00884838116771(21)767, (01)00884838116771(21)649, (01)00884838116771(21)588, (01)00884838116771(21)506, (01)00884838116771(21)503, (01)00884838116771(21)688, (01)00884838116771(21)733, (01)00884838116771(21)464, (01)00884838116771(21)739, (01)00884838116771(21)720, (01)00884838116771(21)65, (01)00884838116771(21)499, (01)00884838116771(21)474, (01)00884838116771(21)238, (01)00884838116771(21)282, (01)00884838116771(21)143, (01)00884838116771(21)180, (01)00884838116771(21)134, (01)00884838116771(21)224, (01)00884838116771(21)138, (01)00884838116771(21)149, (01)00884838116771(21)179, (01)00884838116771(21)137, (01)00884838116771(21)131, (01)00884838116771(21)106, (01)00884838116771(21)53, (01)00884838116771(21)30, (01)00884838116771(21)17, (01)00884838116771(21)531, (01)00884838116771(21)259, (01)00884838116771(21)254, (01)00884838116771(21)64, (01)00884838116771(21)83, (01)00884838116771(21)90, (01)00884838116771(21)66, (01)00884838116771(21)452, (01)00884838116771(21)682, (01)00884838116771(21)719, (01)00884838116771(21)494, (01)00884838116771(21)578, (01)00884838116771(21)691, (01)00884838116771(21)544, (01)00884838116771(21)302, (01)00884838116771(21)56, (01)00884838116771(21)145, (01)00884838116771(21)103, (01)00884838116771(21)40, (01)00884838116771(21)627, (01)00884838116771(21)552, (01)00884838116771(21)328, (01)00884838116771(21)658, (01)00884838116771(21)631.
|
| FEI Number |
3002807880
|
Recalling Firm/ Manufacturer |
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Veenpluis 4-6
Best
Netherlands
|
Manufacturer Reason for Recall | Nine (9) identified software issues which may result in the following: loss of X-ray after Pedal Tap, loss of X-ray after Phase Fault, missing live X-ray on the FlexVision monitor, system continues to restart, AMD triple drive, C-Partition of Suite PC running out of free space, system remains in continuous restart mode after start-up, misalignment of Marker Tool Overlay, longitudinal position error. |
FDA Determined Cause 2 | Software design |
| Action | Philips notified consignees on about 12/08/2025 via emailed URGENT Medical Device Correction letter. The letter informed consignees that they have identified nine (9) software issues affecting Azurion R3.0 systems that may result in loss of imaging (X-ray) functionality, and/or loss of motorized movement, and/or incorrect image content, and/or loss of data. Consignees were instructed to identify all affected units on hand, ensure all users of the affected units are aware of the issue and notify any customers if the unit has been transferred or distributed. Consignees were also instructed to keep the notification with documentation of the system until the issue is resolved, and complete and return the response form. If any of the nine issues are encountered, consignees were asked to contact their local Philips representative and the letter provided recommended actions.
***Updated May 8, 2026***
Previously, Philips planned to address issues 1-7 in software update 3.1.5, while issues 8 and 9 would be addressed separately.
Update 3.1.5 has been canceled.
Issues 1-9 will now be addressed through software release 3/1/15. Anticipated release date for software version 3.1.15 is Q3 2026.
Local Philips representatives will contact each consignee to schedule visits to install the software updates once available. |
| Quantity in Commerce | 990 units |
| Distribution | Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Argentina, Australia, Austria, Bangladesh, Belgium, Brazil, Brunei Darussal, Bulgaria, Cambodia, Canada, Chile, Colombia, Cote de'Ivore, Croatia, Czech Republic, Denmark, Dominican Rep, Ecuador, Egypt, El Salvador, Estonia, Finland, France, Germany, Greece, Honduras, Hong Kong, India, Indonesia, Iraq, Israel, Italy, Japan, Kazakhstan, Kosovo, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Mauritius, Morocco, Nepal, Netherlands, New Zealand, Nicaragua, Norway, Oman, Palestine, Philippines, Poland, Portugal, R¿union, Romania, Saudi Arabia, Serbia, Singapore, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkiye, Ukraine, United Kingdom, Utd.Arab.Emir., Uzbekistan, Vietnam, Yemen.
|
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = OWB
|
|
|
|