| | Class 2 Device Recall Vue Motion V12 |  |
| Date Initiated by Firm | March 05, 2026 |
| Date Posted | March 27, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-1692-2026 |
| Recall Event ID |
98520 |
| 510(K)Number | K151774 |
| Product Classification |
System, image processing, radiological - Product Code LLZ
|
| Product | Vue Motion V12. Product Number: 1017979. |
| Code Information |
Product Number: 1017979, Software Version: v12.2.0 - v12.2.8.500. UDI Numbers: (01)00884838100367(11)200221(10)12.2.5, (01)00884838100367(11)230223(10)12.2.8.300, (01)00884838100343(11)200423(10)12.2.5.0, (01)00884838100367(11)230615(10)12.2.8.310, (01)00884838100367(11)200903(10)12.2.5.1, (01)00884838100367(11)230720(10)12.2.8.400, (01)00884838100343(11)201125(10)12.2.5.100, (01)00884838100367(11)230921(10)12.2.8.410, (01)00884838100343(11)210217(10)12.2.5.200, (01)00884838100336(11)231221(10)12.2.8.420, (01)00884838100367(11)210628(10)12.2.5.300, (01)00884838100350(11)240116(10)12.2.8.421, (01)00884838100367(11)210628(10)12.2.5.300, (01)00884838100350(11)240204(10)12.2.8.422, (01)00884838100343(11)220525(10)12.2.5.400, (01)00884838100350(11)240226(10)12.2.8.424, (01)00884838100367(11)230223(10)12.2.5.405, (01)00884838100336(11)231221(10)12.2.8.420, (01)00884838100336(11)230301(10)12.2.5.406, (01)00884838100336(11)231221(10)12.2.8.420, (01)00884838100367(11)230320(10)12.2.5.407, (01)00884838100336(11)231221(10)12.2.8.420, (01)00884838100367(11)230403(10)12.2.5.408, (01)00884838100367#(11)231223(10)12.2.8.435, (01)00884838100343(11)240211(10)12.2.5.413,(01)00884838100367#(11)231228(10)12.2.8.435A, (01)00884838100343(11)20240923(10)12.2.5.414, (01)00884838100343(11)240917(10)12.2.8.436, (01)00884838100343(11)250728(10)12.2.5.416, (01)00884838100367(11)241014(10)12.2.8.437, (01)00884838100367(11)210308(10)12.2.8.0, (01)00884838100343(11)241128(10)12.2.8.438, (01)00884838100367(11)220127(10)12.2.8.100, (01)00884838100367(11)250122(10)12.2.8.439, (01)00884838100343(11)220417(10)12.2.8.130, (01)00884838100350(11)240407(10)12.2.8.440, (01)00884838100350(11)221213(10)12.2.8.200, (01)00884838100367(11)210308(10)12.2.8.0, (01)00884838100350(11)230301(10)12.2.8.210, (01)00884838100367(11)240616(10)12.2.8.500.
Equipment Number: 86183662, 118470289, 115532146, 115483951, 117699552, 86093591, 98297412, 86093647, 86241626, 86093616, 119288183, 86241420, 86326303, 86325905, 86325795, 86325794, 86325747, 86325221, 86093571, 108816071, 86093582, 86093358, 86093539, 86241707, 86241692, 86093595, 86241861, 86241751, 86241678, 86324616, 116468394, 106682968, 86241742, 87750382, 108664430, 118757545, 87751135, 87750882, 86325445, 86241699, 86241684, 86241669, 86241750, 86093376, 89412152, 86093502, 86369383, 86093605, 120505331, 86093623, 86093339, 86093663, 86093678, 86367969, 104418433, 86325612, 86093588, 86093356, 86324563, 86093674, 86093646, 86093566, 86093685, 86093659, 118007768, 107808460, 119061786, 86093547, 88168079, 86326302, 86325187, 86325123, 86093545, 86093435, 86093427, 95752135, 94477265, 92029774, 92029787, 87771660, 117645555, 117645550, 86241704, 86241689, 86241674, 107639091, 112190786, 112190691, 86093283, 106771488, 110432369, 110432367, 106771482, 119091432, 119091431, 118487430, 118487429, 97582725, 97582724, 97582723, 97582653, 97582719, 97582717, 97582716, 87940696, 107632484, 116468405, 106410735, 117979657, 86241787, 86241771, 107327494, 107327505, 118832692, 99601715, 99601713, 94677044, 92107086, 118239830, 117908479, 97223395, 96789905, 116645253, 116645227, 99601477, 99601462, 101939624, 101939623, 108423189, 108423187, 112368329, 94774385, 118905692, 115619712, 108420051, 86097812, 117795737, 95364241, 86097897, 119690483, 119690374, 119690229, 119690557, 108374575, 86097930, 86097807, 86367896, 94645462, 94645430, 94645536, 94644301, 94644110, 95364465, 86097900, 86097899, 86097870, 86097869, 86097867, 86097866, 86097865, 86097864, 86097872, 86097797, 93942911, 86097968, 95864471, 95863241, 95862554, 95861090, 95857876, 95857299, 95857238, 95856111, 116036959, 108387852, 108387818, 108387787, 108385798, 108385775, 108385742, 86097888, 92994681, 92994630, 92994624, 92994574, 92994546, 92994507, 92994368, 92993154, 92993151, 92993146, 92992839, 92992837, 92994642, 92993515, 86097956, 86097802, 86097795, 86097958, 86097957, 86097885, 86097884, 86097912, 86097911, 86097910, 86097874, 99156870, 86242184, 95635728, 86142456, 86539251, 98748118, 98746992, 100367081, 96233639, 100638306, 112044793, 112045596, 112045794, 117541969, 114535249, 98747092, 95635888, 97933599, 94881635, 86168167, 100428338, 98489592, 98489557, 106973031, 93903547, 100159454, 102820056, 102802637, 113204610, 95805302, 95805287, 95805271, 95805255, 95805242, 95805217, 94881654, 94881640, 98346586, 98346584, 98346582, 94799727, 98238281, 98238276, 98238267, 98238261, 98238253, 98238243, 98238237, 98238230, 98238226, 98238221, 98237322, 98237318, 116646058, 116646033, 116646026, 116645649, 116645624, 117182354, 116646037, 118828892, 118828856, 117097056, 117096664, 117096497, 117096075, 117095988, 117095970, 117094564, 117094474, 117094452, 117094420, 118828939, 118828867, 93042188, 93041325, 86095135, 86095074, 86095099, 104840334, 98948136, 105562169, 86094996, 118563482, 96789997, 108853862, 97370939, 97370937, 104115975, 86094935, 116530988, 86094846, 86095283, 100324828, 100324783, 100324769, 100324760, 109708861, 109708838, 109708815, 109708792, 109708766, 109708740, 109708713, 109708685, 109708657, 103831916, 86474402, 104746873, 86474908, 103833462, 103833428, 86094304, 113894783, 113894781, 98792012, 91749792, 89339147, 91749587, 117060464, 117060439, 117060438, 114669457, 86093969, 117060475, 86474394, 116305411, 101521202, 101521343, 99365708, 86094754, 86095233, 95524915, 107326902, 86094659, 86095267, 105561308, 86095072, 86095003, 109457806, 99432277, 105560963, 86095055, 86094923, 86474387, 104840103, 105561126, 88008682, 86095076, 110163975, 86094926, 86095245, 86094955, 86095021, 86094999, 86095039, 86095286, 86095083, 118266668, 118737860, 88095608, 119056047, 86094889, 86094860, 95306362, 118133040, 116305085, 86095064, 100780848, 119677923, 86094948, 86095103, 86094850, 86095241, 118911878, 97013846, 117982532, 86094895, 118108165, 116179080, 116179074, 116179062, 116179061, 116179057, 116179055, 86095232, 86094921, 86094986, 86095248, 86095014, 86094854, 104838834, 116178956, 86095292, 86094836, 86094959, 86095139, 86094974, 86094838, 86094729, 86095236, 86095289, 95962665, 92958216, 86095024, 89339139, 103604138, 86094845, 110632504, 118007853, 111522070, 117645003, 103832464, 103834294, 110664739, 100324838, 86094111, 86094875, 86094874, 115177945, 86094380, 86095260, 95329429, 95329426, 88057857, 86094826, 97263587, 102239004, 103833517, 103833502, 115644537, 91749847, 116243803, 86094964, 117452752, 116195778, 116195776, 116195752, 117527403, 116468437, 97014506, 117587323, 117527386, 116244071, 107326906, 116107520, 114142262, 116243860, 116243824, 117183326, 98547390, 116304623, 91749780, 117928560, 91749843, 91749834, 116243771, 114142342, 120401436, 116304777, 120331900, 116304735, 108976759, 112538697, 97832433, 110432731, 120795489, 103604817, 103604785, 103604776, 103604682, 103604676, 103604673, 105559799, 114744006, 116304704, 114142329, 114142325, 106771491, 106771492, 114399298, 100782139, 103832327, 99073599, 117335599, 116530996, 117739845, 86094824, 120709292, 118266680, 113460356, 118831902, 118732069, 114142337, 116516398, 100315749, 118344445, 105135271, 115764493, 115764544, 99073536, 105560312, 115764281, 115764292, 116304578, 114399307, 115764433, 86365606, 86365067, 87934745, 105922777, 105922764, 95826001, 95826157, 101896585, 101896581, 91843284, 91841944, 97701367, 97701343, 97701385, 113134346, 119806987, 85169762, 85169499, 89763852, 89763985, 107041852, 95416751, 91843296, 93506127, 95415379, 109627884, 109627862, 109627866, 95625917, 95625840, 96023550, 119417234, 105580824, 101346371, 108664696, 116779621, 86095730, 99602288, 88949124, 86095781, 92711058, 104417373, 104417364, 120525935, 104838082, 106771479, 104838079, 102980168, 100928292, 100928298, 100576504, 100576470, 100576507, 116779626, 116779639, 116779630, 101349635, 118550980, 119391780, 120517450, 118484249, 118712209, 120514105, 120516850, 118726835, 118721885, 118483623, 120512460, 118550237, 119391366, 118483960, 118821140, 120513632, 119366996.118547417, 118521078, 99368906, 107631784, 87518262, 86096012, 99365741, 100928160, 86095999, 118088555, 118088550, 102388164, 102388165, 102236435, 102236439, 106683047, 117145699, 114744022, 112049919, 117145710, 117145707, 117145704, 117145695, 117145694, 115024870, 115024864, 115024855, 112049940, 112049889, 94460495, 115885945, 115177931, 115177935, 120014961, 120014957, 120014947, 120014937, 120014931, 99602079, 113461142, 113460948, 113461011, 93273587, 117373732, 94774372, 94774374, 86252456, 86167732, 86167725, 115252483, 89639007, 86098845, 86098930, 86098939, 86098938, 86098937, 86098936, 86098935, 86098934, 86098922, 86098829, 86098828, 86098827, 86098826, 86098825, 86098824, 86098823, 86098822, 91335350, 91335323, 93042747, 86096407, 116305249, 93390135, 108853850, 103461078, 96629405, 96629406, 116179160, 116179162, 117666256, 117666716, 95441681, 93740781, 92334050, 92333986, 92333968, 92333946, 86098635, 86251723, 107511097, 86098708, 86098711, 86098710, 92333976, 86098709, 107404506, 107404455, 107404447, 107401386, 118064316, 116214057, 116214030, 116214018, 116208322, 116208313, 96721834, 116981739, 116981737, 86137872, 107349095, 107348860, 103832777, 97881452, 97881429, 97881455, 97264192, 97264190, 97264179, 97264177, 118846696, 118846685, 118846667, 118846649, 94882588, 94882921, 94883607, 86138709, 86138708, 86138707, 86138706, 86138705, 99306437, 108866195, 86139833, 116901894, 116901874, 116901854, 105582742, 105581702, 105582720, 105581688, 86140309, 118973588, 116775930, 116775928, 116775922, 116775926, 116638428, 116638423, 116638405, 116638392, 116638389, 86140925, 92700403, 86139951, 86140414, 95375885, 86140360, 86140765, 86132915, 86132902, 86140325, 102127720, 86139774, 86140124, 86140122, 86382830, 86382829, 86382828, 86382826, 86334667, 86074674, 86139582, 86139854, 86139825, 86139680, 86139989, 86139424, 100617022, 93249258, 86334385, 86140659, 86140582, 86140020, 95825251, 86140587, 86140109, 86140918, 96702784, 86139133, 95412546, 105755879, 86074198, 86335016, 86080576, 101748281, 101748286, 86139491, 86139490, 86139488, 104370859, 104370837, 86140275, 117218214, 117218172, 117218153, 117218103, 117218096, 117215833, 95030816, 101023477, 101023444, 101023485, 101023451, 99307349, 99307352, 86074647, 86074646, 86074645, 86074644, 86074643, 112408362, 112408352, 112408347, 112408340, 112408333, 112408089, 112408083, 112408077, 112408071, 112408065, 120887531, 120887443, 86075099, 107828243, 107827863, 107827861, 107827860, 107827859, 107827857, 107827844, 107827842, 107827841, 107827839, 107827836, 107827813, 107827812, 107827811, 107827810, 107827807, 105551378, 92247199, 95321669, 105756096, 107353443, 119072227, 97092933, 97740908, 97740950, 98729234, 102558519, 86074223, 86074224, 92494852, 92494860, 116620870, 116620868, 116620867, 114071257, 115843272, 109620063, 96056867, 108953901, 108953906, 95587207, 97881724, 101602194, 104138189, 104138325, 104806779, 108953707, 108953701, 105755664, 105755765, 105755789, 105755830, 105755884, 105755883, 105755934, 105755962, 105756032, 105756049, 105756059, 105756160, 108509964, 108509954, 116494637, 116494633, 108506963, 86142971, 100369386, 86142784, 96174422, 96174421, 96174423, 108560376, 108560658, 108560362, 115983469, 117129601, 108506692, 108506666, 108506645, 91679040, 86244951, 107608880, 114822314, 114822295, 100277088, 100277161, 100277076, 95782920, 95782918, 95782917, 97091030, 95152769, 95824856, 86142703, 86142702, 100367737, 86244907, 86244823, 114902649, 114902636, 96297256, 96297250, 86142745, 86142744, 86142743, 86142740, 86142981, 96174417, 102800923, 97689132, 102856223, 97792129, 89139861, 86142751, 97845471, 86142742, 96696337, 93034315, 95698027, 86142889, 97879038, 92849246, 93975136, 96020109, 94582178, 86142976, 86142916, 100335050, 100335042, 100928096, 100928064, 95782648, 95782645, 95782644, 115072213, 114929210, 115301106, 86142935, 86142905, 116513595, 115484922, 115484744, 97546353, 97546352, 97546351, 97546348, 111712541, 111712618, 111712442, 97879022, 118731615, 117044440, 117044418, 94530286, 94530268, 118636489, 118636398, 95697067, 89139925, 117501445, 117501207, 117501160, 117501211, 105535195, 109209723, 86143027, 89674846, 86167679, 86167704, 110146218, 86082865, 115983182, 115983048, 115982846, 115982582, 115100785, 115097719, 104661487, 86155173, 86155174, 93669460, 88472961, 88472905, 88472903, 88472811, 88472773, 102493004, 117295406, 92462767, 92883516, 92871813, 93069766, 92871428, 93432037, 92871196, 86366250, 108330581, 103286767, 116163232, 116163162, 103130324, 118622570, 104249484, 119317446, 118536819, 118534245, 118756363, 118756188, 89577119, 118757540, 96746296, 95159446, 95159421, 95159401, 95159434, 95159407, 95159443, 95159425, 95159428, 95159436, 95159419, 95159415, 95159405, 95159396, 95159438, 95159403, 95159449, 95159394, 95159417, 95159423, 95159440, 95159410, 95159432, 95159392, 92088275, 92088278, 118127348, 96746274, 97220852, 102277676, 102277591, 86155316, 86155306, 86152618, 86152617, 86149376, 86149375, 105039160, 105039086, 105039060, 102714733, 102714715, 104139671, 97883953, 93595973, 102431228, 102431188, 102431155, 117289555, 117289505, 117289280, 117289232, 117289008, 103773018, 103772965, 86154658, 86154652, 97690412, 97690376, 100351655, 100351637, 104628380, 104628297, 104627408, 110570140, 110570318, 110570329, 110570127, 100812857, 100812753, 100812751, 103831810, 86155433, 86155432, 86155421, 95696316, 95696295, 95695848, 102212738, 102212846, 108341057, 98147695, 98147684, 89360348, 86153882, 86153881, 86153880, 118341591, 88166467, 88163092, 106329656, 87221860, 86152275, 86152274, 86152273, 86152270, 86152238, 86152237, 86152236, 86151715, 86151711, 86151710, 86151709, 106329804, 106329870, 106329772, 106329714, 106329654, 106329638, 105603426, 105603495, 105603413, 86080636, 86153252, 114050991, 93185598, 93070532, 93452461, 93034712, 100426444, 100427011, 104139061, 102700604, 107663009, 107663075, 107662998, 99305867, 99305845, 114575889, 114575725, 114575502, 104659266, 104659154, 97322979, 97934035, 97934050, 106903970, 86153467, 86153466, 106903941, 106903875, 105549667, 102278248, 86155250, 91499857, 103771270, 103579166, 86081181, 86152996, 86152853, 86152826, 86152809, 86152767, 86152599, 86152594, 86152470, 86152460, 86152165, 86152076, 105558343, 105235936, 105235847, 105235804, 104805939, 104805758, 104805680, 104805578, 104805284, 104805153, 104805008, 104804821, 104804567, 104804354, 104634328, 104467868, 104288488, 86153813, 86153785, 86153741, 108292365, 108292431, 108292363, 120079800, 108810667, 108810678, 108810661, 103022083, 103022421, 103772099, 103772080, 101476340, 108357492, 108682560, 108682545, 108340786, 89842893, 106788754, 106788737, 117978477, 117978534, 117978494, 117978462, 117978426, 94759844, 94759763, 94762025, 103023771, 103022805, 100351735, 101476453, 91497248, 89743761, 93663748, 102607248, 100813323, 100813277, 102701612, 88236010, 89011948, 104510095, 117219108, 117219080, 110379127, 95752082, 103024259, 110541521, 110541500, 87914931, 102278615, 86155277, 88572195, 116414396, 96111759, 97323118, 88487094, 110379867, 88807278, 108340484, 108301034, 104628665, 103807427, 94760264, 98878597, 116414303, 88184694, 108320718, 108320715, 108320703, 94761912, 104139980, 115986819, 115985696, 115985183, 116414304, 103807846, 116413862, 116413858, 94760358, 93363451, 108340607, 108301355, 86079186, 96368352, 112350356, 94760170, 86081933, 86081932, 86079691, 86079695, 89338607, 106731717, 104509776, 86146559, 104836754, 86159650, 97883143, 96235484, 86082574, 111644281, 108340710, 97933831, 103918668, 97264346, 86081238, 86081236, 86081234, 86079244, 86079240, 86079238, 86079028, 86079020, 112768214, 112761350, 112418830, 86079018, 86082066, 86082062, 86082060, 86082058, 89577778, 86079673, 86079677, 86081386, 86082070, 86081400, 86081395, 86081389, 86081383, 86081354, 86081247, 86158269, 86158232, 86158231, 86158228, 104661727, 86158227, 86158226, 86158225, 100944075, 86080160, 100944077, 86301457, 86301431, 86301503, 93360179, 86082160, 86079694, 86150895, 86150820, 86080579, 95409796, 95409777, 94760125, 86082097, 95719962, 88025943, 86080584, 88891036, 86301824, 86154922, 107150667, 86080570, 97221519, 97221466, 97221558, 95807901, 100334813, 86081037, 86155935, 86155913, 86153866, 86153857, 86153848, 105719741, 105719724, 105719702, 105719686, 105719671, 105719655, 86147845, 86147844, 86147359, 86147358, 86147357, 108319484, 109178520, 109178498, 109178483, 109178390, 89577368, 89576425, 86150478, 101476149, 101476114, 101476147, 109313185, 109313141, 100945030, 100945008, 98586392, 98586238, 89173337, 86155737, 86155736, 107355427, 107355194, 107354852, 105552144, 105552088, 86080468, 86080574, 86080572, 91641659, 91641647, 91697962, 119224399, 103720577, 102735934, 108356564, 118193656, 118572712, 117333491, 99155686, 99155143, 109601392, 115705785, 109641625, 93266520, 93266266, 97563253, 108982826, 97838711, 118330388, 118330352, 114059054, 110515243, 107733404, 109205651, 100425243, 89436953, 92535806, 86167426, 102323602, 108299362, 104137044, 104137011, 113195057, 113194408, 96514712, 96514402, 96514221, 104672768, 86163623, 106341583, 99401519, 100982159, 104647226, 93929870, 93929825, 88277265, 98903574, 96094785, 96094769, 105340504, 105340492, 105340406, 109642470, 117234595, 117234567, 96381743, 95522688, 93196722, 93196704, 102421392, 88233595, 95578220, 94468561, 112173309, 112170683, 118553947, 102317987, 92658797, 113843432, 91761457, 108211334, 108211289, 108389735, 99605260, 108243109, 108242761, 118586504, 93527408, 116460352, 104136481, 93222897, 86326332, 87109166, 86077810, 86077845, 86076064, 86167564, 89750598, 105379212, 87453905, 96247510, 107450721, 86323978, 86076746, 99121954, 86367975, 86077234, 86367958, 86077315, 86077311, 87345031, 118653837, 86076321, 86076766, 86325458, 86441407, 86076563, 86077089, 86890152, 86077229, 86077227, 88802514, 119504300, 86076157, 89768454, 86077093, 86076941, 86441349, 86077317, 114741567, 98773691, 98726606, 98726233, 87784682, 101445747, 110515414, 110515395, 104278123, 102324467, 102324033, 91950075, 98463123, 103673535, 102872250, 108233250, 108233230, 110429788, 108854388, 99024407, 101868558, 86366534, 86326455, 104956752, 86183736, 100615323, 86273056, 86273038, 116879981, 104787421, 104768170, 102429097, 117948831, 117948830, 117948818, 117948758, 117948752, 117948751, 117948685, 86240916, 86240887, 88472863, 88472720, 88471413, 86365665, 86275188, 86275128, 86275124, 86240872, 86240871, 86365675, 86365673, 110355227, 115041760, 115013232, 115041550, 119006025, 119155342, 115057624, 86325620, 101616590, 118567700, 118567215, 118567196, 116563496, 116563485, 104723096, 105167176, 95888800, 95888797, 109455379, 109455345, 97262971, 115331347, 115331318, 112190095, 96116472, 96116153, 105449561, 106410712, 106684033, 106410722, 106410720, 106410718, 98947535, 98492613, 98015174, 97832767, 97832759, 97370970, 97223021, 97014757, 95516216, 94676513, 103459479, 99242504, 98947564, 97832732, 97382213, 101521086, 100928094, 100577014, 100576816, 100576307, 108093607, 120709284, 107638328, 105579663, 107638074, 116586216, 112539279, 112539202, 109140786, 109140743, 109140683, 102652012, 102651978, 102388242, 102388237, 102388234, 101526484, 101526470, 101526367, 101526363, 101526353, 101526347, 101526332, 101526323, 101526304, 101526281, 100782187, 100782184, 99601393, 101526293, 101526267, 101526253, 100782173, 100323353, 116901682, 116658897, 116658894, 116658892, 116658887, 86093108, 98297674, 113834281, 86093031, 105696881, 96025660, 86093039, 86092137, 116901678, 86092643, 86252262, 86252282, 86092500, 86092637, 86092311, 94461283, 86092087, 86092091, 86092809, 104723044, 86092627, 86092112, 86092104, 94080067, 92404650, 86092101, 86092175, 86092290, 95307171, 86092432, 116621960, 116621958, 114967241, 114967240, 116621953, 96214808, 118975072, 98947927, 98947818, 98947827, 108134947, 108134892, 111984150, 109707994, 119477261, 119477228, 118958578, 118958541, 118958538, 118939305, 118939302, 118939296, 118540651, 118487414, 118487400, 117416959, 116901692, 108134956, 108134955, 108134954, 108134953, 107326950, 120709289, 120383147, 118239996, 117249179, 115065793, 115065792, 114512333, 112114642, 86366271, 86083840, 86083839, 117801532, 108303254, 108303228, 86083846, 89448269, 89448170, 89448049, 86141889, 86141888, 86083715, 86083714, 86083635, 86083634, 86083881, 95110566, 86401484, 86401480, 86401479, 86401478, 103560911, 103560507, 103577829, 107492896, 107493129, 107492952, 97843757, 97843303, 97843626, 94804014, 94803938, 94803796, 103746263, 103723229, 103746974, 106609386, 106609262, 106608861, 92286108, 92286077, 92286045, 103077104, 103076453, 103076408, 103076677, 99159171, 99159099, 99158802, 107492615, 116313013, 100352039, 100352041, 114755754, 106815772, 106815771, 106815770, 106815764, 86141805, 86141766, 86855403, 100437088, 100437083, 100437068, 109308791, 106096109, 86083721, 86083681, 86083607, 86083706, 105354181, 91738036, 102305390, 102305388, 102305386, 99009179, 98989485, 103139586, 86098565, 86098564, 86098561, 86098560, 86098558, 86098557, 86098549, 86098488, 86098487, 86272819, 86272683, 86272650, 86271783, 86098541, 86098540, 86098462, 86098461, 97319696, 97712904, 97712165, 97713437, 86365953, 86252583, 87343545, 87343808, 95833909, 114967246, 114904062, 108179819, 108179813, 108179812, 108179810, 108179753, 108179750, 108183653, 108179739, 116195642, 109130901, 86096238, 97832849, 104418772, 86096202, 99074463, 86096215, 86096221, 86096217, 92236072, 86096274, 86096206, 103601605, 94233764, 94233765, 86096270, 97383627, 86096200, 86096266, 91460853, 97832850, 97832858, 89046511, 87577897, 96669258, 104869132, 98311069, 86325488, 118082104, 88467153, 118037043, 93200210, 86083054, 86083052, 86083049, 86083017, 104869746, 86083002, 86082999, 86082996, 86082994, 86083026, 86083023, 86365551, 86365114, 92480489, 97885672, 97885656, 104870247, 104248646, 100900394, 117782127, 102493186, 110326932, 86148421, 119205735, 87973542, 91852619, 87503730, 86148306, 87244983, 95269077, 105081082, 103116602, 103116684, 100915545, 86100260, 114084645, 93994644, 93994560, 86253045, 86252836, 86253018, 86252838, 86252638, 86253084, 86253362, 86252804, 86253352, 86253357, 86252752, 86253266, 86252660, 86253381, 87804505, 86100122, 86100140, 86100135, 86100132, 86100137, 86100136, 86100273, 86100306, 100915309, 100915317, 98235497, 109181699, 102219161, 101525598, 104118634, 115521230, 113224750, 109181493, 105135814, 86252375, 86252455, 86252446, 88036114, 105135733, 86252457, 86096385, 109140869, 112191109, 115521173, 98614965, 118128292, 89589267, 89589257, 95010219, 88472473, 88472446, 88472421, 88472412, 88472398, 88472383, 88472363, 88472354, 88472341, 88472322, 88472309, 88472292, 88472281, 88472271, 88472255, 88472235, 88472228, 88472215, 88472194, 88472181, 88472167, 88472150, 88472126, 88472094, 86365957, 86365261, 98528071, 98595702, 104265451, 104265421, 104267279, 104267264, 104265732, 101005814, 101005667, 101005754, 118851115, 118850995, 108697743, 111705377, 110000762, 110000760, 110000761, 115858621, 116881812, 104265545, 104265514, 119323616, 101094597, 119130611, 104267092, 104267332, 115747513, 115747507, 103770562, 103770455, 113946965, 113946851, 96099643, 86274865, 86274863, 86274821, 86274819, 86274693, 86274651, 86274522, 86274479, 86274227, 86274056, 86273877, 86273850, 86273658, 86273593, 94901371, 94881740, 94881739, 91384817, 89985128, 94883588, 88412316, 91319408, 88413001, 88413000, 88412990, 88412989, 88412983, 88412982, 88412975, 88412974, 88412968, 88412961, 100456158, 95010181, 102386760, 88412960, 88412954, 88412952, 88412946, 88412939, 88412938, 88412932, 88412931, 88412925, 88412924, 88412918, 88412917, 88412911, 88412910, 88412904, 88412903, 88412897, 88412896, 88412890, 88412889, 88412883, 88412882, 88412876, 88412875, 88412869, 88412868, 88412862, 88412861, 88412853, 88412852, 88412844, 88412843, 88412834, 88412833, 88412824, 88412823, 88412815, 88412814, 88412805, 88412804, 88412796, 88412795, 88412783, 88412781, 88412773, 88412772, 88412764, 88412763, 88412755, 88412754, 88412743, 88412742, 88412733, 88412732, 88412722, 88412721, 88412863, 91515398, 103115137, 103115105, 95010202, 116076728, 116312664, 108505360, 115202252, 115201907, 88471781, 104869059, 104869083, 108487598, 108487599, 108487601, 108487633, 108487631, 108487718, 108487596, 108487746, 108487712, 108487696, 108487723, 108487720, 108487652, 108487706, 104868931, 114834995, 114835013, 114835038, 114835060, 114835073, 114835226, 94493575, 94493566, 94493555, 108487801, 108487433, 86323788, 93765576, 116998574, 86083152, 86083158, 86083157, 86083156, 86083171, 86083162, 86083161, 86083160, 86083155, 86083146, 109338081, 109338082, 109338075, 109338077, 109338023, 86272990, 88471613, 117465735, 97337986, 86183802, 117996230, 117996115, 117996006, 117927824, 117926378, 118015756, 118015691, 117877105, 117877042, 117827782, 86183828, 86183824, 86183770, 86183771, 86275022, 86275017, 86275002, 86274992, 86274988, 86274983, 86274916, 86274911, 86274898, 86274897, 86274891, 86274887, 86183816, 91365541, 89589282, 89589275, 89589270, 103792554, 103792545, 100882853, 100700656, 100700305, 103115075, 103115066, 103115025, 104087434, 116881580, 94783312, 97563157, 116303434, 86099474, 86249267, 107555376, 100664820, 86251144, 94782574, 86099764, 86099631, 86099543, 119754056, 115176959, 115176734, 94783918, 98183872, 98183877, 86251017, 114100170, 117931639, 95751410, 95751453, 95751444, 95751435, 95751386, 95751384, 95751448, 95751446, 95751404, 95751400, 86099622, 86246039, 114917049, 104000415, 104000414, 86099280, 86248748, 86248580, 114085118, 86246280, 94781788, 94781731, 86250320, 86099405, 86250878, 99175367, 101998316, 101763584, 101763562, 101763556, 101745125, 101745120, 101745118, 101654077, 101653973, 101653968, 101653965, 101653943, 101651771, 101651767, 101651762, 101651753, 101651748, 108567309, 104939097, 86248559, 86099482, 86099483, 86249259, 106274099, 94106482, 94547572, 86249098, 86248929, 86249952, 86245895, 86246581, 105040421, 86249743, 86250293, 86246905, 86250539, 86249660, 86247787, 86247725, 86247890, 86249090, 86248921, 86248753, 86245918, 86247962, 86246047, 86247173, 86245882, 86247916, 95890320, 86900911, 95504166, 86099665, 94783143, 86099533, 86250463, 89576384, 86099667, 86099798, 86099801, 86099329, 86099336, 86099333, 86250671, 86250501, 86099334, 86099769, 95890338, 86099781, 86099780, 119754009, 96638239, 96638231, 86099870, 105061675, 100713492, 100713951, 105252665, 118469737, 108567736, 86098385, 89654364, 93816382, 86098341, 86098310, 86253772, 86253764, 86253756, 86098387, 89653660, 89002914, 86098365, 86098378, 86142577, 86242770, 115639723, 115639699, 86242835, 86242810, 98749078, 98670009, 98669985, 94101517, 98669997, 98669947, 101717722, 101604572, 101604575, 114823743, 114823622, 86142626, 93903996, 86142593, 86142560, 86142559, 86142596, 86142595, 98485758, 98485757, 94106146, 94106139, 94106164, 94106154, 110356117, 110328976, 116820160, 116820134, 116820094, 116820073, 116819837, 101814295, 101814290, 101814217, 101814212, 101814207, 101814302, 101814222, 115784997, 115784992, 86142592, 86142548, 86243339, 86242608, 86242582, 86242556, 110140477, 110142434, 86142639, 86242926, 103747870, 103747801, 101182977, 101182966, 117560210, 117560175, 86142635, 86142563, 87958062, 86142681, 86142676, 86243106, 86242797, 86242772, 86243326, 86242595, 86242569, 86242542, 86243054, 86242764, 86242958, 86242929, 86242883, 86242854, 86242619, 96299693, 109934527, 86242581, 103429487, 98486648, 101561078, 86243075, 86242897, 86142542, 86142553, 86142524, 86142549, 88139938, 86142672, 106709827, 88545728, 88545703, 108667554, 108667559, 100828693, 100828686, 93069894, 93069891, 93069886, 93069879, 92534170, 92534166, 92534158, 98879043, 116679525, 116679331, 116679004, 116678831, 116678558, 116678339, 116678078, 116677640, 116677155, 116658619, 116658600, 116658489, 116656298, 116656237, 116656189, 116655776, 116655688, 116655568, 116655298, 116655192, 116655737, 116655475, 116655273, 116655257, 116655215, 102024191, 108462017 , 108462017, 102088934, 102088771, 102111219, 102110377, 103598387, 103598148, 102739579, 103252115, 102759874, 102749994, 102739882, 102023892, 102734168, 102733950, 95127345, 102024073, 102761529, 64086021, 103212419, 103252135, 102978877, 103210917, 103211356, 103252408, 102761064, 103696069, 102752729, 102309545, 100913986, 100913981, 100913969, 100913960, 100913897, 100913582, 100913574, 100913569, 100913334, 100913138, 100913130, 100913123, 100913104, 104043689, 104043480, 94247320, 94247290, 107452033, 107452031, 107452028, 107452025, 106613466, 106613458, 106613450, 106613425, 117246093, 117246083, 117238119, 117238104, 116139863, 116139816, 116115900, 116115897, 116115763, 98949472, 85163083, 85163082, 85163084, 85163079, 86033326, 116512147, 116511508, 116511502, 116511501, 85162367, 85162358, 108732712, 109380598, 109339269, 108540287, 101588227, 101588218, 101587384, 101587373, 101200499, 108507986, 89577724, 109196371, 109339325, 116148434, 85167291, 93741011, 93741004, 98878291, 85163227, 85163226, 85163228, 109257586, 85162509, 85162482, 85162481, 94566740, 85162518, 85162527, 85162444, 85162427, 92535871, 92535872, 85162904, 116777383, 116777376, 116737896, 116737894, 116148438, 109258935, 109251807, 109339822, 109258309, 109179074, 88208195, 101202172, 109381249, 104836354, 104633423, 104633208, 104835556, 104835544, 104629666, 103079101, 91679368, 108464568, 85162734, 85162723, 85286140, 88208185, 88208183, 88207889, 88207888, 85256839, 85167205, 85167187, 85167180, 85162471, 93904600, 86223020, 86223017, 86223004, 86223003, 85162570, 109071779, 118734581, 116113383, 116113378, 87805616, 87805604, 91587041, 103622326, 103622306, 95580271, 95580241, 85256842, 107353280, 107353274, 107353270, 107353268, 85167482, 85167469, 85167467, 85167466, 85167464, 85167254, 85167239, 101231366, 101231325, 101231230, 101231323, 101231214, 101231209, 103811041, 103811018, 85167221, 98238891, 98238890, 98238717, 98238715, 85167212, 85162423, 85162755, 98065673, 98065669, 96295935, 96295866, 87439271, 91679889, 91679887, 91679370, 91679366, 91679294, 109339485, 117785374, 117785366, 94307880, 94307875, 94307869, 94307866, 117760577, 117760457, 101854322, 107349826, 107349824, 107349822, 107349819, 119187937, 119187913, 117979653, 117979647, 117979581, 117928777, 117928730, 117928677, 85162546, 109252568, 85162673, 85162594, 85162587, 101203124, 85162378, 85162563, 118954466, 85162525, 85162524, 94567206, 94567197, 97295668, 97295643, 97295624, 86177966, 109381513, 93106498, 93106349, 117183071, 117182649, 117183027, 117182858, 117182838, 117182826, 117182788, 117182516, 117182501, 117182489, 117182459, 100162027, 100161995, 86338928, 85162572, 85162580, 93492543, 93472089, 88299635, 93568414, 116165677, 116165570, 116166276, 116166240, 87408440, 97666018, 97665997, 97666059, 101590297, 101589994, 100944697, 100944695, 100944694, 100944693, 100944692, 100944690, 100944666, 100944665, 100944664, 100944662, 100944661, 100944659, 100944674, 100944671, 118912244, 118912232, 117607614, 117607611, 117607607, 117607604, 117607584, 117607576, 117607572, 117607569, 117607566, 107361339, 107356902, 107356715, 119993246, 119993240, 119993238, 119993227, 119993204, 119993202, 118912266, 118912257, 117607621, 117607586, 86069058, 116160078, 115153129, 115153122, 115153102, 115153096, 115153055, 115153049, 107980774, 101589275, 101588314, 113132350, 113132272, 86070022, 86072973, 86069984, 86069986, 107299006, 108699903, 108699895, 108699891, 108699878, 113747458, 113747358, 116028059, 116028064, 116028051, 116028043, 101841184, 101841032, 105754238, 105754097, 99366621, 99366614, 99366608, 109217770, 109217769, 109217768, 109217757, 109217756, 109217738, 109217737, 109217736, 102737628, 101596985, 104676414, 105754728, 105754720, 105754738, 95469687, 95469078, 95469684, 95469660, 95469324, 95469319, 95469072, 95469070, 95468502, 95468497, 95470716, 95469361, 95468661, 98688528, 98688525, 118066927, 118066916, 118066912, 118066895, 118066891, 116382868, 116382862, 116382899, 116382823, 116382818, 116382931, 116382830, 104334310, 104334393, 104334221, 95223253, 112055766, 112055638, 112055072, 95408435, 113770564, 113770522, 113749274, 101840130, 101839968, 97677867, 97677860, 97677876, 98932877, 98932851, 98932826, 101587398, 101587095, 117627396, 117145872, 117145870, 86069542, 86069541, 91538194, 113462283, 113462184, 107455943, 107455676, 107455674, 107455668, 107455635, 107455532, 107455530, 107455529, 107455496, 107455491, 113160105, 113159999, 86072777, 86071757, 86071499, 86071378, 86071141, 110461965, 106685090, 106685041, 101840758, 101840426, 96374083, 96374081, 96374080, 96374076, 96373670, 96373622, 96329997, 96329683, 103964017, 103963812, 101853545, 101853399, 122697820, 107810176, 107810156, 107810129, 95825764, 95825731, 95825673, 106281409, 106281331, 106281293, 106281269, 106281231, 106281209, 106281598, 100325901, 100325939, 95222488, 95222478, 106389536, 106389497, 106389753, 103964379, 103964326, 103964461, 103964371, 103964359, 103964350, 103964340, 86072885, 110649438, 110649391, 110649379, 98712738, 98712483, 98712471, 98689324, 98689315, 93114729, 93114179, 98154845, 93114737, 96206094, 96211128, 96211018, 96210547, 106711210, 106711204, 106711192, 106711179, 106711159, 106711105, 106711090, 106711072, 106711036, 116709570, 106711144, 106711124, 106711018, 106711003, 116709572, 99366534, 99366529, 99366523, 114983369, 114983366, 114983361, 109412366, 109412360, 106732882, 106732711, 106733489, 106733445, 106732860, 106732852, 86071675, 86070335, 86070348, 86070347, 89997671, 89997662, 89997681, 116042677, 116042669, 98995644, 98995646, 98995645, 115602925, 115602924, 115602922, 115602921, 115602920, 115602913, 115602899, 115602898, 115602897, 115602896, 115602895, 99104407, 99104408, 94446577, 94457163, 94457158, 94457155, 94456661, 94446729, 94446633, 94446625, 94446593, 116414356, 107773258, 107773254, 98999361, 98999360, 98999299, 98999298, 98999355, 98999354, 98999353, 98999352, 98999351, 98999350, 98999297, 98999296, 98999295, 98999294, 98814395, 98814388, 98999369, 98999368, 98834700, 98834595, 98834265, 98834228, 98834198, 98834191, 116180997, 116180996, 116180415, 116180407, 116181123, 116180994, 116180924, 116180923, 116180405, 116180404, 116180402, 116180595, 108182208, 108182197, 108182174, 108182165, 108182146, 108182141, 106713860, 106712789, 106712473, 106712191, 106712156, 106712099, 106712089, 106712082, 106712045, 106712146, 106712138, 106712131, 106712123, 106712107, 120826531, 105840488, 105840487, 105840486, 105840465, 105840464, 105840444, 105840502, 105840482, 117183151, 117183150, 117183148, 115989036, 115988185, 115970408, 115969801, 115261742, 115261741, 115261736, 115261735, 115261718, 115261716, 94773326, 117467054, 117467052, 117467049, 117467046, 104069494, 86069027, 116810399, 116810404, 117024264, 117024257, 117024251, 117024263, 117024256, 117024250, 94554410, 94554431, 113743775, 113743542, 104472704, 104472570, 104472191, 117375167, 117375159, 101973676, 101973634, 101973428, 101973357, 101973187, 101973733, 95605666, 94678943, 94678937, 94678909, 94678370, 94678366, 94678355, 94678315, 94678306, 94678278, 94678256, 94678170, 113748704, 113750391, 113750355, 98778460, 98778212, 98778469, 95888494, 95888489, 95888434, 95888425, 95888421, 95888412, 95888410, 95888402, 95888400, 95888362, 95888359, 95888326, 100221460, 100221457, 112063563, 112063526, 112063489, 112063373, 116337904, 116263480, 116337874, 116263408, 116263461, 116263466, 116263413, 117627383, 106792616, 106792615, 106792592, 106792591, 106792579, 106792601, 106792600, 106792313, 106792607, 106792606, 106792580, 115256425, 115256424, 115256421, 115256413, 115256411, 115256410, 115256409, 115256408, 106792599, 106792598, 106792552, 106792452, 106775040, 101249167, 101248851, 106687221, 106686322, 106686311, 106686250, 106687219, 106687218, 106687177, 106686617, 106686190, 106686181, 106686096, 106685887, 95678383, 95629360, 95630253, 96972337, 96972310, 96971757, 96971744, 106465988, 106465460, 106179014, 106178993, 106178900, 106178781, 96211591, 96211589, 96773800, 96773797, 97435876, 97435868, 96981434, 96981433, 105754530, 105754488, 103086159, 103086154, 103086153, 103086148, 103086119, 103086117, 103086116, 103083908, 103083831, 103083824, 102132965, 102132939, 102132787, 102132770, 102737938, 102737931, 102737926, 102737921, 102737910, 102737897, 102737892, 102737884, 100934807, 100934804, 100934779, 100934778, 100934774, 96641338, 96641327, 86070742, 114902841, 114902840, 114902830, 114902828, 107535024, 107535022, 106539077, 106539065, 114399168, 96376345, 96376484, 96376473, 86070795, 86069029, 86116202, 86106052, 86071527, 86122487, 86241237, 86106280, 86108876, 86112566, 86271382, 92928283, 86108290, 86105309, 93458771, 93458758, 86112046, 86069994, 86108715, 94259122, 94258847, 86072848, 86069107, 86069675, 86124702, 86108487, 86071165, 86114635, 86112724, 86071206, 86069889, 86108297, 86107836, 86111624, 86121527, 86110248, 91768771, 91768697, 86109051, 86111795, 86113910, 86111724, 94478785, 94478757, 86111756, 86109527, 86072033, 120712618, 86070010, 86072021, 116042666, 86110701, 86108681, 86111302, 86271221, 86101694, 86072603, 86271284, 86069998, 86113409, 86069897, 86069579, 86116157, 94678271, 95721750, 95721739, 86105188, 86107936, 96773785, 86071392, 102132666, 102737874, 86107968, 86112905, 86071739, 86072909, 86070205, 86103470, 86116232, 86071156, 86069156, 86111903, 86111692, 86071755, 86111582, 86070950, 86069964, 86120097, 86108357, 98547836, 98547833, 98547830, 98532315, 98532259, 98532168, 86070203, 86108195, 86069463, 86128008, 86102922, 96641619, 86072285, 86070759, 86069867, 86069822, 100165727, 86072028, 86070711, 86326258, 86069076, 86069074, 86072322, 86072985, 86069106, 86072824, 86072823, 86072822, 86072467, 86072466, 86072465, 86072464, 86072436, 86071579, 86071498, 115727184, 86071142, 86071139, 86071130, 86069556, 86072855, 105848281, 105848234, 105848247, 105848246, 105848245, 105848244, 105848242, 105848199, 105848198, 105848197, 105848196, 105848194, 105848251, 105848250, 105848249, 105848248, 105848201, 105848200, 120871738, 120871737, 120871734, 120871721, 120871719, 120871716, 86072632, 86070209, 86070890, 86070897, 86072336, 86241261, 86072974, 86241232, 86072920, 86072917, 86241229, 86241228, 86071372, 96641126, 96641118, 86071977, 86071973, 96641657, 96641685, 86070684, 86071163, 86326077, 86070751, 86070750, 86071903, 86070556, 86070555, 86071008, 86072933, 86072916, 86071057, 86070713, 86070712, 86326462, 86071421, 86069959, 86071376, 86072988, 86072135, 86072133, 86071370, 86070020, 86072921, 86070700, 114785236, 86072991, 89121552, 86326155, 86241451, 86241436, 89579112, 86093650, 86093336, 87129101, 117633052, 94491370, 88467413, 87976723, 89552593, 89552537, 86365080, 86365075, 95010778 |
| FEI Number |
3002807880
|
Recalling Firm/ Manufacturer |
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Veenpluis 4-6
Best
Netherlands
|
| For Additional Information Contact | Informatics Customer Support 877-3282808 Ext. 4 |
Manufacturer Reason for Recall | Potential that mis-ordered frames in Vue Motion during dynamic cine runs may cause images frames to display out of sequence. |
FDA Determined Cause 2 | Software design |
| Action | Philips Medical notified consignees on about 03/02/2026 via URGENT Medical Device Correction letter. Consignees were instructed to first identify, and how to identify, if the affected products are on hand. If so, they were further instructed that if they do NOT have thumbnail enabled, they will NOT experience this issue described, but if they HAVE thumbnail enabled, when reviewing multi frame studies, always review the frame number displayed on the screen as the correct frame number is visible on the screen. Consignees were told that they may continue to use your system(s) in accordance with the intended use and by provided the recommendation in the letter, to circulate the letter to all users of this device so that they are aware of the potential issue retain this letter with your system(s) until a solution is installed on your system; ensure the letter is in a place likely to be seen/viewed. Consignees were also requested to complete and return the provided response form.
A Philips representative will contact each consignee to schedule a time to install a software solution on the affected system(s) to resolve the issue (reference C&R 2025-EI-RI-003). |
| Quantity in Commerce | 3,552 units |
| Distribution | Worldwide distribution - US Nationwide and the countries of Algeria, Andorra, Angola, Argentina, Australia, Austria, Belgium, Bolivia, Brazil, Cambodia, Canada, Chile, China, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Finland, France, Germany, Gibraltar, Greece, Guernsey, Holy See (Vatican City State), Honduras, Hungary, Iceland, India, Indonesia, Ireland, Israel, Italy, Japan, Jersey, Jordan, Kuwait, Lebanon, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Norway, Oman, Palestine, State of, Panama, Papua New Guinea, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, Romania, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Thailand, United Arab Emirates, United Kingdom, Uruguay, Viet Nam, Zambia.
|
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = LLZ
|
|
|
|