| Date Initiated by Firm | September 27, 2002 |
| Date Posted | February 01, 2003 |
| Recall Status1 |
Terminated 3 on January 12, 2005 |
| Recall Number | Z-0526-03 |
| Recall Event ID |
25248 |
| 510(K)Number | K953910 |
| Product Classification |
System, X-Ray, Mobile - Product Code IZL
|
| Product | Philips BV300 Mobile X-Ray System |
| Code Information |
Part numbers 9896 000 06801 and 9806 000 06802 with serial numbers: DBXXXX - all DBXXXX - all DDXXXX - all DEXXXX - all DG0001-DG0129 DH0001-DH0033 DJ0001 - DJ0152 DK0001 - DK 0076 |
| FEI Number |
1000524572
|
Recalling Firm/ Manufacturer |
Philips Medical Systems
22100 Bothell Everett Highway
Bothell
WA
98041
|
| For Additional Information Contact | Sarah Baxter 425-487-7665 |
Manufacturer Reason for Recall | The C-bow support can bend and break allowing C-arm to fall from support |
FDA Determined Cause 2 | Other |
| Action | The firm issued a letter dated 9/27/02 to their customers. The letter advised of the problem and that a service representative will visit to inspect unit and correct. |
| Quantity in Commerce | 229 devices |
| Distribution | The firm distributed to hospitals, x-ray centers and medical centers throughout the U.S. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|
| 510(K) Database | 510(K)s with Product Code = IZL
|