 Class 2 Device Recall Clea Stand Ceiling used with the Integris XRay System H5000, Allura 12/15, Allura 9Csee related information
Date Initiated by FirmOctober 10, 2003
Date PostedJuly 20, 2004
Recall Status1 Terminated 3 on January 13, 2005
Recall NumberZ-0099-04
Recall Event ID 27485
510(K)NumberK971365 
Product Classification Unit, X-Ray, Extraoral With Timer - Product Code EHD
ProductClea Stand Ceiling used with the Integris X-Ray System H5000, Allura 12/15, Allura 9C
Code Information Part Numbers 986 000 75921 + 9896 000 75922, 9896 001 32411 + 9896 001 32412, 9896 001 31981 + 9896 001 31982. The firm uses site numbers instead of serial numbers. Site numbers: 102674, 105731, 105514, 102009, 105491, 13093, 86561, 10105, 103382, 100434, 100328, 103287, 85633, 85635, 86456, 38728, 85090, 38592, X1732, X1733, 103290, 103150, 104565, 82612, 103957, 76499, 10439, 10440, 13117, 86289, 104335, 103494, 101135, 101136, 101040, X1596, 102983, 82528, 41461, 100504, 103554, 85126, 85131, 86594, 87324, 13118, 83016, 103860, 101948, 82692, 103282, 86135, 101589, 103379, 100708, 103002, 103214, 104412, 104975, X1695, 87339, 101971, 10464, 100685, 102723, 103557, 86196, 10482, 102348, 101965, 86398, 38373, 76511, 87012, 102564, 102179, 104865, 102664, 103456, 104416, 104978, 13289, 103026, 38925, 102521, 104974, 82526, 101938, 86367, X1408, X1540, 86585, 86599, 100689, 83313, 82576, 38795, 41480, 104425, 104426, X1687, 105511, 105686, 85498, 38358, 102855, 102856, 85014, 100503, 86253, 100701, 86144, 10478, 82950, 85676, 101239, 102493, 86571, 102793, 86914, 102863, 103712, 104580, 104581, 38356, 104169, 104170, 101139, X1738, X1736, 86935, 76830, 104511, 86066, 100262, 101606, 101607, 103711, 85488, 100657, 86434, 83015, 83996, 13109, 86981, 76962, 100811, 76931, 76932, 76933, X1690, 100683, X1327, 103221, 103991, 86577, 50222, 50223, 83807, 13107, 13997, 100612, 76545, 102808, 102810, 101612, 10391, 101873, 103891, 76986, 86136, 76559, 41288, 52472, 86061, 102809, 104427, 101582, 101583, 103630, 100658, 102193, 102502, 86575, 103721, 86319, 103029, 101036, 85195, 102868, 84443, 82823, 10104, 85093, 104402, 102872, 101465, 104056, 100948, 87603, 100949, 100950, 82522, 102248, 104062, 85106, 100839, 86321, 103248, 84444, 84466, 84468, 100607, 101864, 76528, 87081, 83994, 62940, 45077, 102161, 104529, 103714, 100945, 103246, 104424, 50399, 100643, 10454, 85416, 100913, 76556, 76557, 101566, 102612, 86502, 105299, 104076, 102945, 102953, 13883, 105309, 105254, 100840, 85129, 26117, 86145, 76917, 84789, 85028, 86063, X1168, 41139, 103746, 86290, 103327, 76514, 100316, 105098, 105294, 100611, 102675, 86008, 103335, 86341, 38146, 87532, 103144, 102791, 84447, 100865, 82575, 26161, 82822, 103133, 41145, 101928, 102156, 103650, 86411, 86989, 103371, 85503, 85639, 85010, 85011, 103901, 105389, 103336, 76987, 102096, 103004, 87300, 86576, 102589, 85107, 86420, 86495, 103709, 102826, 83924, 103776, 104976, 14147, 76527, X1584, 103879, 82517, 82523, 86295, 86537, 87316, 82699, 83024, 102890, 84489, 41460, X1724, X1699, 104038, 103777, 102635, 100587, 86320, 100810, 101617, 101618, 86982, 102811, X1262, X1282, X1284, X1389, X1697, 102944, 38261, 85499, 82537, 86433, 102776, 103875, 76577, 100470, 87197, 103169, 104797, 87528, 86435, 103926, 10441, 100645, 82824, X1688, 85445, 86422, 86088, 103231, 103240, 41104, 100783, 83861, 83862, 50371, 50373, 102572, 104539, X1146, 102098, 87004, 86994, 103587, 52463, 104077, 103034, 59703, 87337, 14168, 100585, 102418, 76525, 100869, 102094, 102487, 103819, X1528, 104145, 101576, 82696, 84445, 86189, 101077, 102180, 100475, 100477, 84721, 38425, 102817, 102871, 85128, 13114, 100485, 100486, 100487, 104418, 38702, X1476, 86421, 100866, 103372, 85013, 100630, 100631, 103793, 105733, 14145, 50433, 101078, 102842, 100959, 102631, 102160, 102162, 83042, 100521, 38201, 13296, 100656, 86972, 13292, 102166, X1717, X1718, 103266, 85169, 103032, X1716, 10442, 85041, 86975, 91650, 100696, 102001, 103900, 76918, 87319, 76505, 50438, 82520, 82719, 104885, 84713, 103828, 38276, 85135, 86183, 82825, 104337, 103182, 103183, 86360, 104410, 103512, 50339, 86536, 86542, 76867, 102823, 76883, 73884, 76885, 76886, 104933, 102083, 102995, 104798, X1685, X1686, 102037, 87366, 87069, 104541, 102632, 13290, 85094, 84482, 86296, 104187, 103929, 104513, 103584, 103612, 50374, 101208, 86506, 76510, 85196, 86065, 102558, 102563, 104430, 87013, 103070, 86581, 101966, 103237, 80330, 102934, 100788, 103193, 85133, 101485, 100613, 76880, 76882, 86933, 86953, 82815, 86578, 103594, 100319, 83838, 104060, 86437, 87364, 103823, 100714, 103987, 86172, 86272, 87179, 101973, 83326, 83375, 101309, 50335, 100432, 104422.
FEI Number 1000524572
Recalling Firm/
Manufacturer
Philips Medical Systems Sales & Service Region No. America
22100 Bothell Everett Highway
Bothell WA 98041
For Additional Information Contact
425-487-7000
Manufacturer Reason
for Recall
The cable carrier and cables may fall after repeat collisions with OR lamps, radiation shields or other supported accessories.
FDA Determined
Cause 2
Other
ActionThe firm sent a letter dated, October 10, 2003 indicating the firm''s service representatives will upgrade the cable carrier by mounting a safety catch inside the cable carrier.
Quantity in Commerce536 units
DistributionThe firm distributes to hospitals and medical centers nationwide.
Total Product Life CycleTPLC Device Report