 Class 3 Device Recall AxSYM & IMx CA 125 Reagent Packssee related information
Date Initiated by FirmNovember 25, 2003
Date PostedJanuary 10, 2004
Recall Status1 Terminated 3 on July 26, 2004
Recall NumberZ-0318-04
Recall Event ID 27834
510(K)NumberK964020 K964185 
Product Classification unknown device name - Product Code LTK--
ProductAxSYM CA 125 Reagent Pack, list 3B41-22 and IMx CA 125 Reagent Pack, list 7A89-22; 100 test kits; Abbott Laboratories, Abbott Park, IL 60064
Code Information all lots
FEI Number 1415939
Recalling Firm/
Manufacturer
Abbott Laboratories HPD/ADD/GPRD
100/200 Abbott Park Road
Abbott Park IL 60064
For Additional Information ContactAbbott Customer Support
877-422-2688
Manufacturer Reason
for Recall
The CA 125 assay product inserts list the incorrect storage instructions for the CA 125 Calibrators and Controls, and the product information letter (PI-P) shipped with the products to correct the storage conditions, 69-6349/R2, refers to an outdated control package insert, 39-1642/R3, instead of the current control package insert 34-1047/R4
FDA Determined
Cause 2
Other
ActionField Correction recall by letter dated 11/25/03 to all customers receiving the affected reagents. The accounts were informed of the incorrect control package insert referenced in the product information letter shipped with each CA 125 product. The letters contained the correct control package insert number, 34-1047/R4, and stated that all CA 125 assay product inserts will be updated shortly to include the revised storage instructions for the CA 125 Calibrators and Controls.
DistributionNationwide and internationally through Abbott subsidiaries in Argentina, Aruba, Australia, Bermuda, Brazil, Canada, Cayman Islands, Chile, Colombia, Ecuador, Germany, Peru, Hong Kong, India, Jamaica, Japan, Korea, New Zealand, Pakistan, Puerto Rico, Singapore, Taiwan, Thailand, United Kingdom and Uruguay.

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
510(K) Database510(K)s with Product Code = LTK--
510(K)s with Product Code = LTK--
-
-