| Date Initiated by Firm | August 16, 2004 |
| Date Posted | September 02, 2004 |
| Recall Status1 |
Terminated 3 on March 31, 2005 |
| Recall Number | Z-1437-04 |
| Recall Event ID |
29876 |
| Product Classification |
Catheter, Electrode Recording, Or Probe, Electrode Recording - Product Code DRF
|
| Product | Deflectable Tip Catheter for cardiac mapping.
Mfg Part # Catalog # EU Part #
D-1078-35-S 107835S SW1078-035;
D-1078-63-S D5S06AL252RT 36A-07Q;
D-1078-78-S 107878S N/A;
D-1097-511-S 1097511S SW1097-511;
D-1097-554-S D708DL002RT 36F-32Q. |
| Code Information |
112 lots |
| FEI Number |
3003113629
|
Recalling Firm/ Manufacturer |
Biosense Webster, Inc.
15715 Arrow Hwy
Irwindale
CA
91706-2006
|
| For Additional Information Contact | Mary McDermott 909-839-8720 |
Manufacturer Reason for Recall | Potential that product is incorrectly assembled with monel wire attached to the tip dome rather than copper wire. |
FDA Determined Cause 2 | Other |
| Action | Sales Representative to provide letter to each of the 25 US customers explaining the action. Sales Representatives will remove any affected product from customers inventory. Certification records will be completed for each customer confirming the removal action.
International affiliates to conduct similar actions. |
| Quantity in Commerce | 848 |
| Distribution | Nationwide and Japan, Europe, & Uruguay |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|