| Date Initiated by Firm | December 17, 2004 |
| Date Posted | March 10, 2005 |
| Recall Status1 |
Terminated 3 on January 27, 2006 |
| Recall Number | Z-0580-05 |
| Recall Event ID |
30803 |
| Product Classification |
Heart-Valve, Allograft - Product Code MIE
|
| Product | CryoValve, Pulmonary Valve & Conduit |
| Code Information |
Donor #62654, Serial #7816169, Model #PV00 |
| FEI Number |
3001451326
|
Recalling Firm/ Manufacturer |
Cryolife Inc
1655 Roberts Blvd Nw
Kennesaw
GA
30144-3632
|
| For Additional Information Contact | Mr. P. Tyler Cochran 770-419-3355 |
Manufacturer Reason for Recall | Records noted that the infant donor's mother had tested positive for Hepatitis B. |
FDA Determined Cause 2 | Other |
| Action | Physicians were notified by letter on 12/17/2004. |
| Quantity in Commerce | 1 tissue |
| Distribution | CA, FL |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|