Class 2 Device Recall Tube Stand for the Dynarad Phantom Portable XRay System

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Tube Stand for the Dynarad Phantom Portable XRay Systemsee related information
Date Initiated by FirmMarch 10, 2005
Date PostedMay 11, 2005
Recall Status1 Terminated 3 on November 14, 2005
Recall NumberZ-0816-05
Recall Event ID 31456
Product Classification Tube Mount, X-Ray, Diagnostic - Product Code IYB
ProductTube Stand for the Dynarad Phantom Mobile X-Ray System; Del Medical Systems, 11550 West King Street, Franklin Park, IL 60131
Code Information All Phantom Model Tube Stands with squared corners on all four corners of the stand. Catalog #501031, serial numbers 00743-0204, 00778-0204, 00997-0304, 01001-0304, 01103-0304, 01293-0304, 01371-0304, 01425-0404, 01452-0404, 01470-0404, 01475-0404, 01575-0404, 01592-0404, 01700-0404, 01704-0404, 01790-0404, 01841-0404, 01851-0404, 01898-0504, 02019-0504, 02023-0504, 02051-0504, 02096-0504, 02178-0504, 02188-0504, 02310-0604, 02357-0604, 02429-0604, 02517-0604, 02534-0604, 02584-0604, 02671-0604, 02752-0704, 02760-0704, 02776-0704, 02839-0704, 02891-0704, 02928-0704, 02938-0704, 02945-0704, 02959-0704, 02980-0704, 02991-0704, 03076-0704, 03084-0704, 03122-0704, 03375-0804, 03455-0804, 03476-0804, 03518-0804, 03543-0804, 03570-0904, 03579-0904, 03636-0904, 03700-0904, 03708-0904, 03722-0904, 03740-0904, 03952-0904, 04037-0904, 04134-1004, 04183-1004, 04347-1004, 04380-1004, 04407-1004, 04477-1004, 04481-1004, 04491-1004, 04524-1004, 04534-1004, 04592-1104, 04600-1104, 04605-1104, 04648-1104, 04694-1104, 04699-1104, 04703-1104, 0439-1104, 04783-1104, 04862-1104, 05190-1204, 05194-1204, 05198-1204, 05216-1204, 05364-1204, 05414-1204, 05418-1204, 05439-1204, 00283-0105, 00340-0105, 00481-0105, 00571-0205, 00575-0205, 00617-0205, 00649-0205, 00763-0205, 00841-0205, 00858-0205, 00891-0205, 00900-0205, 00913-0205
FEI Number 1418964
Recalling Firm/
Manufacturer
Del Medical Systems Group
11550 West King Street
Franklin Park IL 60131-1330
For Additional Information ContactMs. Veronica Meridith
800-800-6006
Manufacturer Reason
for Recall
The Phantom Portable X-Ray System Stand may have a suspect weld supporting the bracket holding the locking cylinder assembly. It the weld fails, the tubehead arm assembly could fall and contact the patient or operator.
FDA Determined
Cause 2
Other
ActionDel Medical sent Field Correction Recall letters dated 3/10/05 to the direct accounts, informing them of the potential for a defective weld supporting the braket holding the locking cylinder assembly. If the weld fails, the tubehead arm assembly could potentially fall and contact the patient or operator. The accounts were requested to locate the affected systems and take them out of service until they are inspected and, if needed, repaired.
Quantity in Commerce31 stands
DistributionNationwide, Mexico and Thailand.