| Date Initiated by Firm | June 20, 2005 |
| Date Posted | August 30, 2005 |
| Recall Status1 |
Terminated 3 on August 30, 2005 |
| Recall Number | Z-1422-05 |
| Recall Event ID |
32714 |
| Product Classification |
System, Multipurpose For In Vitro Coagulation Studies - Product Code JPA
|
| Product | STA-R blood coagulation analyzer. Catalog Number 57160. |
| Code Information |
All units distributed under Catalog Number 57160. |
| FEI Number |
2245451
|
Recalling Firm/ Manufacturer |
Diagnostica Stago, Inc.
5 Century Dr
Parsippany
NJ
07054-4607
|
| For Additional Information Contact | James Barrow 973-631-1200 |
Manufacturer Reason for Recall | Concerning the STA-R blood coagulation analyzer, there is a possible failure mode associated with patient samples receiving prolonged exposure time to reagents during the cuvette roll change process. |
FDA Determined Cause 2 | Other |
| Action | Alert on the problem was sent out on 6/20/2005 to all customers alerting the problem and instructions on how to prevent the problem. The units are not be to returned. |
| Quantity in Commerce | 215 units |
| Distribution | Units were distributed to clinics and hospitals nationwide. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|