• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Baxter 6060 & Sabratek 6060 Homerun MultiTherapy Infusion Pump

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Baxter 6060 & Sabratek 6060 Homerun MultiTherapy Infusion Pumpsee related information
Date Initiated by FirmNovember 14, 2005
Date PostedJanuary 28, 2006
Recall Status1 Terminated 3 on June 19, 2007
Recall NumberZ-0412-06
Recall Event ID 34020
510(K)NumberK002679 
Product Classification Pump, Infusion - Product Code FRN
ProductBaxter 6060 Multi-Therapy Infusion Pump, product codes 2M9832, 2M9832P and 2M9832R, and Sabratek 6060 Homerun Infusion Pumps, product codes 606000, 606000-40, 606000-40L, and 606000-40I; Manufactured by an affiliate of Baxter Healthcare Corp., Deerfield, IL 60013 USA, made in Singapore
Code Information All serial numbers of all product codes
FEI Number 1416980
Recalling Firm/
Manufacturer
Baxter Healthcare Corp.
Rt. 120 & Wilson Rd
Round Lake IL 60073
For Additional Information ContactCenter for One Baxter
800-422-9837
Manufacturer Reason
for Recall
Failures within the PCA profile as well as incidents resulting in interruptions of therapy in various profiles.
FDA Determined
Cause 2
Other
ActionBaxter sent recall letters dated 11/14/05 to all of their 6060 pump customers on the same date via DHL overnight delivery. The accounts were informed of the reports of failures within the PCA profile and interruptions of therapy in various profiles. Due to the obsolescence of certain critical components, Baxter is conducting a controlled removal of all 6060 pumps from the market. Baxter will coordinate with customers individually to ensure a smooth transition to a substitute device. Baxter provided interim instructions to discontinue use of the PCA profile to avoid over infusion or non-delivery conditions, and to discontinue the use of the pump for therapies where interruption of the infusion could cause immediate patient harm. Baxter representatives telephoned each account within 48 hours of the letter, and visited them within a week to help the customers transition to other infusion pumps. A press release issued on 11/15/05. Baxter sent a follow-up recall letter to the accounts on 6/20/06, informing them that service repairs for the 6060 Infusion Pump will cease after 12/31/06, and compensation requests cannot be processed for pumps received after 12/31/06. Production of sets for the 6060 will cease on 10/1/06, and will not be available after 12/31/06. The accounts were requested to complete and fax back to Baxter the enclosed reply form, indicating the current inventory of 6060 pumps remaining at the facility.
Quantity in Commerce34,000 pumps
DistributionNationwide and internationally to Belgium, Germany, Finland, France, the United Kingdom, Ireland, Sweden, Norway, Canada, Japan, New Zealand and Australia.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
510(K) Database510(K)s with Product Code = FRN
-
-