| Date Initiated by Firm | November 03, 2005 |
| Date Posted | May 09, 2006 |
| Recall Status1 |
Terminated 3 on June 20, 2006 |
| Recall Number | Z-0855-06 |
| Recall Event ID |
34776 |
| 510(K)Number | K050810 |
| Product Classification |
computed radiography system digitizer - Product Code MQB
|
| Product | CR DX-S, DX-S, Computed radiography system (Digitizer) |
| Code Information |
CR DX-S: Units with Serial numbers SN1001-SN1079, Software versions: STR1102B and below. |
| FEI Number |
3001236302
|
Recalling Firm/ Manufacturer |
AGFA Corp.
10 S Academy St
Greenville
SC
29601-2632
|
| For Additional Information Contact | Ray B. Myers Ph.D 864-421-1827 |
Manufacturer Reason for Recall | Corrupted image after system start; Corrupted image after workflow interruption; An erasure unit problem can cause system to stop. |
FDA Determined Cause 2 | Other |
| Action | Consignees were notified at the time of upgrading. The software was upgraded on 11/3/2005, 12/19/2005 and 12/28/2005. A Mandatory Service Bulletin, No. 11, explaining the issues was shared with the accounts. |
| Quantity in Commerce | 3 units |
| Distribution | DC, MS, OR--three states |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|
| 510(K) Database | 510(K)s with Product Code = MQB
|