 Class 3 Device Recall Hamilton microliter syringesee related information
Date Initiated by FirmMarch 28, 2006
Date PostedJune 22, 2006
Recall Status1 Terminated 3 on December 21, 2006
Recall NumberZ-1169-06
Recall Event ID 35292
Product Classification syringe HPLC - Product Code JQW
ProductSyringe, 62RNR 2.5 Microliter Syringe, with needle point style 3 (blunt needle point for use with HPLC injection valves and for sample pipetting) Part #: 87942.
Code Information Lot 340007 and 342081
FEI Number 2915796
Recalling Firm/
Manufacturer
Hamilton Co
4970 Energy Way
Reno NV 89502-4123
For Additional Information ContactWayne McAuliffe
775-858-3000
Manufacturer Reason
for Recall
There is an error in the marking of the graduated divisions on syringes in affected lots that shows an amount that is double the actual volume of the syringe
FDA Determined
Cause 2
Other
ActionOn 3/28/06, the firm initiated the recall and its notification was via letters explaining the reason for the recall and requesting the product be returned.
Quantity in Commerce63 units
DistributionWorldwide-The product was distributed to 28 U.S. consignees and 3 foreign consignees. (Canada, Switzerland, Japan)
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
-
-