| | Class 2 Device Recall ABL 5 capillary adaptor |  |
| Date Initiated by Firm | June 09, 2005 |
| Date Posted | August 03, 2006 |
| Recall Status1 |
Terminated 3 on February 24, 2012 |
| Recall Number | Z-1320-06 |
| Recall Event ID |
35774 |
| 510(K)Number | K943882 |
| Product Classification |
in vitro diagnostic accessory - Product Code CHL
|
| Product | ABL 5 Capillary Adaptor (a connection that fits the capillary tube on to the aspirator on the ABL 5 & BPH 5 Analyzers), in vitro diagnostic accessory, Part Number: 906-018. |
| Code Information |
Part #906-018 lot 041643 to lot 045124 |
| FEI Number |
1523456
|
Recalling Firm/ Manufacturer |
Radiometer America Inc
810 Sharon Dr
Westlake
OH
44145-1598
|
| For Additional Information Contact | Vince Sigmund 800-736-0600 |
Manufacturer Reason for Recall | The dimensions and shape of the analyzer adaptor do not fit the analyzer tubes or probe, causing air to be aspirated with the blood sample and causing a deviation in the pO2 sample results. |
FDA Determined Cause 2 | Other |
| Action | The firm issued a letter to their customers dated 6/9/2005. The letter reports the defect with the ABL5 Capillary Adapters and informs the customers that Radiometer reps will be replacing all of the customers'' ABL5 adaptors ASAP. The letter is accompanied by a single bag of new adaptors for the customers'' ''immediate needs'' and also provides a faxed response form which the customer can use to request additional adapters as required. In addition, should the customer choose to use the old adapters in their possession, the letter stresses and reiterates the ''Precautions and Warnings'' applicable for use of the ABL5 should air bubbles be observed in the measuring chamber by the user/operator. |
| Quantity in Commerce | 155 bags |
| Distribution | Nationwide, including KS, OH, MD, FL, TX, CA, KY, MA, VA, WA, IL, AR and ME, and Canada. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|
| 510(K) Database | 510(K)s with Product Code = CHL
|
|
|
|