| Date Initiated by Firm | September 27, 2006 |
| Date Posted | October 24, 2006 |
| Recall Status1 |
Terminated 3 on October 05, 2009 |
| Recall Number | Z-0063-2007 |
| Recall Event ID |
36410 |
| Product Classification |
Hospital Bed. - Product Code LLI
|
| Product | AC-powered Adjustable Hospital Bed.
(ICCG Bed) Affected Model numbers:
IH820-3MDLX, IHSC900DLX, IH820-3MDLX-QS, IH820-3MPDLX-QS, SC900-80WDLX-QSP, SC900-80PDLX-QSP, IH820-3MDLX-AE-333PS, SC90080GDLX-AE-366PA, SC90080GDLX-AE-333PS, IH820-3MDLX-AE-366PS, IH820-3MDLX-AE-333PA, IH820-3MDLX-AE-366PA and SC90080GDLX-AE-366PS |
| Code Information |
Serial Numbers Affected : 06CF035033 06DF003462 06DF007408 06DF018274 06DF027564 06DF029581 06DF042349 06EF006839 06CF035034 06DF003483 06DF007409 06DF018275 06DF027565 06DF029582 06DF042350 06EF006840 06CF035035 06DF003484 06DF007410 06DF018276 06DF027566 06DF029583 06DF042351 06EF006841 06CF035036 06DF007274 06DF007411 06DF018277 06DF027567 06DF029584 06DF042352 06EF006842 06CF035037 06DF007275 06DF007412 06DF018278 06DF027568 06DF029585 06DF042353 06EF006843 06CF035038 06DF007276 06DF007413 06DF018279 06DF027569 06DF029586 06DF042354 06EF006844 06CF035039 06DF007277 06DF007414 06DF018280 06DF027570 06DF032354 06DF042355 06EF006845 06CF035040 06DF007278 06DF007415 06DF018281 06DF027571 06DF032355 06DF042356 06EF006846 06CF035041 06DF007279 06DF007416 06DF018282 06DF027572 06DF032356 06DF042357 06EF006847 06CF035042 06DF007280 06DF007417 06DF018283 06DF027573 06DF032357 06DF042358 06EF006848 06CF035043 06DF007281 06DF007418 06DF018284 06DF027574 06DF032358 06DF042359 06EF006849 06CF035044 06DF007282 06DF007419 06DF018285 06DF027575 06DF032359 06DF042360 06EF006850 06CF035045 06DF007283 06DF007420 06DF018286 06DF027576 06DF032360 06DF042361 06EF006851 06CF035046 06DF007284 06DF007421 06DF018287 06DF027577 06DF032361 06DF042362 06EF006852 06CF035047 06DF007285 06DF007422 06DF018288 06DF027578 06DF032362 06DF042363 06EF006853 06CF035048 06DF007286 06DF007423 06DF018289 06DF027579 06DF032363 06DF042364 06EF006854 06CF035049 06DF007287 06DF007424 06DF018290 06DF027580 06DF032364 06DF042365 06EF006855 06CF035050 06DF007288 06DF007425 06DF018291 06DF027581 06DF032365 06DF042366 06EF006856 06CF035051 06DF007289 06DF018186 06DF018292 06DF027582 06DF032366 06DF042367 06EF006857 06CF035052 06DF007290 06DF018187 06DF018293 06DF027583 06DF032367 06DF042368 06EF006858 06CF035053 06DF007291 06DF018188 06DF018294 06DF027584 06DF032368 06DF042369 06EF006859 06CF035054 06DF007292 06DF018189 06DF018295 06DF027585 06DF032369 06DF042370 06EF006860 06CF035055 06DF007293 06DF018190 06DF018296 06DF027586 06DF032370 06DF042371 06EF006861 06CF035056 06DF007294 06DF018191 06DF018297 06DF027618 06DF032371 06DF042372 06EF006862 06CF035057 06DF007295 06DF018192 06DF018298 06DF027619 06DF032372 06DF042373 06EF006863 06CF035058 06DF007296 06DF018193 06DF027474 06DF027620 06DF032373 06DF042374 06EF006864 06CF035059 06DF007297 06DF018194 06DF027475 06DF027621 06DF032374 06DF042375 06EF006865 06CF035060 06DF007298 06DF018201 06DF027476 06DF027622 06DF032375 06DF042376 06EF006866 06CF035061 06DF007299 06DF018202 06DF027477 06DF027623 06DF032376 06DF042377 06EF006867 06CF035062 06DF007300 06DF018203 06DF027478 06DF027624 06DF032377 06DF042378 06EF006868 06CF035063 06DF007301 06DF018204 06DF027479 06DF027625 06DF032378 06DF042379 06EF006869 06CF035064 06DF007302 06DF018205 06DF027480 06DF027626 06DF032379 06DF042380 06EF006870 06CF035065 06DF007303 06DF018206 06DF027481 06DF027627 06DF032380 06DF042381 06EF006871 06CF035066 06DF007304 06DF018207 06DF027482 06DF027628 06DF032381 06DF042382 06EF006872 06CF035067 06DF007342 06DF018208 06DF027483 06DF027629 06DF032382 06DF042383 06EF006873 06CF035068 06DF007343 06DF018209 06DF027484 06DF027630 06DF032385 06DF042384 06EF006874 06CF035069 06DF007344 06DF018210 06DF027485 06DF027631 06DF032388 06DF042385 06EF006875 06CF035070 06DF007345 06DF018211 06DF027486 06DF027632 06DF032389 06DF042386 06EF006876 06CF035071 06DF007346 06DF018212 06DF027502 06DF027633 06DF032390 06DF042387 06EF006877 06CF035072 06DF007347 06DF018213 06DF027503 06DF027634 06DF038967 06DF042388 06EF006878 06CF035073 06DF007348 06DF018214 06DF027504 06DF027635 06DF038968 06DF042389 06EF006879 06CF035074 06DF007349 06DF018215 06DF027505 06DF027636 06DF038969 06DF042390 06EF006880 06CF035075 06DF007350 06DF018216 06DF027506 06DF027637 06DF038970 06DF042391 06EF006881 06CF035076 06DF007351 06DF018217 06DF027507 06DF027638 06DF038971 06DF042392 06EF006882 06CF035077 06DF007352 06DF018218 06DF027508 06DF027639 06DF038972 06DF042393 06EF006883 06CF035078 06DF007353 06DF018219 06DF027509 06DF027640 06DF038973 06DF042394 06EF006884 06CF035079 06DF007354 06DF018220 06DF027510 06DF027641 06DF038974 06DF042395 06EF006885 06CF035080 06DF007355 06DF018221 06DF027511 06DF027642 06DF038975 06DF042396 06EF006886 06CF035081 06DF007356 06DF018222 06DF027512 06DF027643 06DF038976 06DF042397 06EF006887 06CF035082 06DF007357 06DF018223 06DF027513 06DF027644 06DF038977 06DF042398 06EF006888 06CF035083 06DF007358 06DF018224 06DF027514 06DF027645 06DF038978 06DF042399 06EF006889 06CF035084 06DF007359 06DF018225 06DF027515 06DF027646 06DF038979 06DF042400 06EF006890 06CF035085 06DF007360 06DF018226 06DF027516 06DF027647 06DF038980 06DF042401 06EF006891 06CF035086 06DF007361 06DF018227 06DF027517 06DF027648 06DF038981 06DF042402 06EF006892 06CF035087 06DF007362 06DF018228 06DF027518 06DF027649 06DF038982 06DF042403 06EF006893 06CF035088 06DF007363 06DF018229 06DF027519 06DF027650 06DF038983 06DF042404 06EF006894 06CF035089 06DF007364 06DF018230 06DF027520 06DF027651 06DF038984 06DF042405 06EF006895 06CF035090 06DF007365 06DF018231 06DF027521 06DF027652 06DF038985 06DF042406 06EF006896 06CF035091 06DF007366 06DF018232 06DF027522 06DF027653 06DF038986 06DF042407 06EF006897 06CF035092 06DF007367 06DF018233 06DF027523 06DF027654 06DF038987 06DF042408 06EF006898 06CF035093 06DF007368 06DF018234 06DF027524 06DF027655 06DF038988 06DF042409 06EF006899 06CF035094 06DF007369 06DF018235 06DF027525 06DF027656 06DF038989 06DF042410 06EF006900 06CF035095 06DF007370 06DF018236 06DF027526 06DF027657 06DF038990 06DF042411 06EF006901 06CF035096 06DF007371 06DF018237 06DF027527 06DF027658 06DF038991 06EF006785 06EF006902 06CF035097 06DF007372 06DF018238 06DF027528 06DF027659 06DF038992 06EF006786 06EF006903 06CF035098 06DF007373 06DF018239 06DF027529 06DF027660 06DF038993 06EF006787 06EF006904 06CF035099 06DF007374 06DF018240 06DF027530 06DF027661 06DF038994 06EF006788 06EF006905 06CF035100 06DF007375 06DF018241 06DF027531 06DF027662 06DF038995 06EF006789 06EF006906 06CF035101 06DF007376 06DF018242 06DF027532 06DF027663 06DF038996 06EF006800 06EF006907 06CF035102 06DF007377 06DF018243 06DF027533 06DF027664 06DF038997 06EF006801 06EF006908 06CF035103 06DF007378 06DF018244 06DF027534 06DF027665 06DF038998 06EF006802 06EF006909 06CF046827 06DF007379 06DF018245 06DF027535 06DF027666 06DF038999 06EF006803 06EF006910 06CF046828 06DF007380 06DF018246 06DF027536 06DF027667 06DF039000 06EF006804 06EF006911 06CF046829 06DF007381 06DF018247 06DF027537 06DF027668 06DF039001 06EF006805 06EF006912 06CF046830 06DF007382 06DF018248 06DF027538 06DF027669 06DF039002 06EF006806 06EF006913 06CF046831 06DF007383 06DF018249 06DF027539 06DF027670 06DF039003 06EF006807 06EF006914 06CF046832 06DF007384 06DF018250 06DF027540 06DF027671 06DF039004 06EF006808 06EF006915 06CF046833 06DF007385 06DF018251 06DF027541 06DF027672 06DF039005 06EF006809 06EF006916 06CF046834 06DF007386 06DF018252 06DF027542 06DF027673 06DF039006 06EF006810 06EF006917 06CF046835 06DF007387 06DF018253 06DF027543 06DF027674 06DF039007 06EF006811 06EF006918 06CF046836 06DF007388 06DF018254 06DF027544 06DF027675 06DF039008 06EF006812 06EF006919 06CF046837 06DF007389 06DF018255 06DF027545 06DF027676 06DF039009 06EF006813 06EF006920 06CF046838 06DF007390 06DF018256 06DF027546 06DF027677 06DF039010 06EF006814 06EF006921 06CF046839 06DF007391 06DF018257 06DF027547 06DF027678 06DF039011 06EF006815 06EF006922 06CF046840 06DF007392 06DF018258 06DF027548 06DF027679 06DF039012 06EF006816 06EF006923 06CF046841 06DF007393 06DF018259 06DF027549 06DF027680 06DF039013 06EF006817 06EF006924 06CF046842 06DF007394 06DF018260 06DF027550 06DF029567 06DF039014 06EF006818 06EF006925 06CF046843 06DF007395 06DF018261 06DF027551 06DF029568 06DF039015 06EF006819 06EF006926 06CF046844 06DF007396 06DF018262 06DF027552 06DF029569 06DF039016 06EF006820 06EF006927 06CF046845 06DF007397 06DF018263 06DF027553 06DF029570 06DF039017 06EF006821 06EF006928 06CF046846 06DF007398 06DF018264 06DF027554 06DF029571 06DF039018 06EF006822 06EF006929 06CF046847 06DF007399 06DF018265 06DF027555 06DF029572 06DF039019 06EF006823 06EF006930 06CF046848 06DF007400 06DF018266 06DF027556 06DF029573 06DF039020 06EF006824 06EF006931 06CF046849 06DF007401 06DF018267 06DF027557 06DF029574 06DF039021 06EF006832 06EF006932 06CF046850 06DF007402 06DF018268 06DF027558 06DF029575 06DF039022 06EF006833 06EF006933 06CF046851 06DF007403 06DF018269 06DF027559 06DF029576 06DF039023 06EF006834 06EF006934 06CF046852 06DF007404 06DF018270 06DF027560 06DF029577 06DF039024 06EF006835 06EF006935 06CF046853 06DF007405 06DF018271 06DF027561 06DF029578 06DF042346 06EF006836 06EF006936 06CF046854 06DF007406 06DF018272 06DF027562 06DF029579 06DF042347 06EF006837 06EF006937 06DF003461 06DF007407 06DF018273 06DF027563 06DF029580 06DF042348 06EF006838 06EF006938 06EF006939 06EF010455 06EF014740 06EF016729 06EF024142 06EF030019 06EF041599 06FF009829 06EF006940 06EF010456 06EF014741 06EF016730 06EF024143 06EF030020 06EF041600 06FF009830 06EF006941 06EF010457 06EF014742 06EF016731 06EF024144 06EF030021 06EF041601 06FF009831 06EF006942 06EF010458 06EF014743 06EF016732 06EF024145 06EF030022 06EF041602 06FF009832 06EF006943 06EF010459 06EF014744 06EF016733 06EF024146 06EF030023 06EF041603 06FF009833 06EF006944 06EF010460 06EF014746 06EF016734 06EF024147 06EF030024 06EF041605 06FF009834 06EF006945 06EF010461 06EF014747 06EF016735 06EF024148 06EF030025 06EF041606 06FF012780 06EF006946 06EF010462 06EF015792 06EF016736 06EF024149 06EF030026 06EF041607 06FF012781 06EF006947 06EF010463 06EF015793 06EF016737 06EF024150 06EF030027 06EF041608 06FF012782 06EF006948 06EF010464 06EF015794 06EF016738 06EF024151 06EF030028 06EF041609 06FF012783 06EF006949 06EF010465 06EF015795 06EF016739 06EF024152 06EF030029 06EF041610 06FF012784 06EF006950 06EF010466 06EF015796 06EF016740 06EF024153 06EF030030 06EF041611 06FF012785 06EF006951 06EF010467 06EF015797 06EF016741 06EF024154 06EF030031 06EF041612 06FF012786 06EF006952 06EF010468 06EF015798 06EF016742 06EF024155 06EF030032 06EF041613 06FF012787 06EF006953 06EF010469 06EF015799 06EF016743 06EF024156 06EF030033 06EF041614 06FF012788 06EF006954 06EF010470 06EF015800 06EF016744 06EF024157 06EF030034 06EF041615 06FF012789 06EF006955 06EF010501 06EF015801 06EF016745 06EF024158 06EF030035 06EF041616 06FF012790 06EF006956 06EF010502 06EF015802 06EF016746 06EF024159 06EF030036 06EF041617 06FF012791 06EF006957 06EF010503 06EF015803 06EF016748 06EF024160 06EF030037 06EF041618 06FF015278 06EF006958 06EF010504 06EF015804 06EF016749 06EF024161 06EF030038 06EF041619 06FF015279 06EF006959 06EF010505 06EF015805 06EF016750 06EF024162 06EF030039 06EF041620 06FF015280 06EF006960 06EF010507 06EF015806 06EF016751 06EF024163 06EF030040 06EF041621 06FF015281 06EF006961 06EF010508 06EF015807 06EF022487 06EF024164 06EF030041 06EF041632 06FF015282 06EF006962 06EF010509 06EF015808 06EF022488 06EF024165 06EF030042 06EF041633 06FF015283 06EF006963 06EF010510 06EF015809 06EF022489 06EF024166 06EF030043 06EF041634 06FF015284 06EF006964 06EF010511 06EF015810 06EF022490 06EF024167 06EF030044 06EF041635 06FF015285 06EF006965 06EF010512 06EF015811 06EF022491 06EF024168 06EF030045 06EF041636 06FF015286 06EF006966 06EF010513 06EF015812 06EF022492 06EF024169 06EF030046 06EF041637 06FF015287 06EF006967 06EF010514 06EF015813 06EF022493 06EF024170 06EF034192 06EF041638 06FF017482 06EF006968 06EF010515 06EF015814 06EF022494 06EF024171 06EF034193 06EF041639 06FF017483 06EF006969 06EF010516 06EF015815 06EF022495 06EF024172 06EF034194 06EF041640 06FF017484 06EF006970 06EF010518 06EF015816 06EF022496 06EF024173 06EF034195 06EF041641 06FF017485 06EF006971 06EF010519 06EF015817 06EF023973 06EF024174 06EF034196 06EF041642 06FF017486 06EF006972 06EF010520 06EF015818 06EF023974 06EF024175 06EF034197 06EF041643 06FF017487 06EF006973 06EF010521 06EF015819 06EF023975 06EF024746 06EF034198 06FF006314 06FF017488 06EF006974 06EF010522 06EF015820 06EF023976 06EF024747 06EF034199 06FF006315 06FF017489 06EF006975 06EF010523 06EF015821 06EF023977 06EF024748 06EF034200 06FF006316 06FF017490 06EF006976 06EF010524 06EF015822 06EF023978 06EF024749 06EF034201 06FF006317 06FF018153 06EF006977 06EF010525 06EF015823 06EF023979 06EF024750 06EF035665 06FF006318 06FF018154 06EF006978 06EF010526 06EF015824 06EF023980 06EF024751 06EF035666 06FF006319 06FF018155 06EF006979 06EF010528 06EF015825 06EF023981 06EF024752 06EF035667 06FF006322 06FF018156 06EF006980 06EF010529 06EF015826 06EF023982 06EF024753 06EF035668 06FF006323 06FF018157 06EF006981 06EF010530 06EF015827 06EF023983 06EF024754 06EF035669 06FF006324 06FF018158 06EF006982 06EF010531 06EF015828 06EF023984 06EF024755 06EF035671 06FF007328 06FF018159 06EF006983 06EF010532 06EF015829 06EF023985 06EF024756 06EF035672 06FF007329 06FF018160 06EF006984 06EF010533 06EF015830 06EF023986 06EF024757 06EF035673 06FF007330 06FF018161 06EF006985 06EF010534 06EF015831 06EF023987 06EF024758 06EF035674 06FF007331 06FF018162 06EF006986 06EF010535 06EF015832 06EF023988 06EF024759 06EF038008 06FF007332 06FF018163 06EF006987 06EF010536 06EF015833 06EF023989 06EF024760 06EF038009 06FF007333 06FF018164 06EF006988 06EF010537 06EF015834 06EF023990 06EF024761 06EF038010 06FF007334 06FF018165 06EF006989 06EF010538 06EF015835 06EF023991 06EF024762 06EF038011 06FF007335 06FF018166 06EF006990 06EF010539 06EF015836 06EF023992 06EF024763 06EF038012 06FF007336 06FF018167 06EF006991 06EF010540 06EF015837 06EF023993 06EF024764 06EF038043 06FF007337 06FF018168 06EF006992 06EF010541 06EF015838 06EF023994 06EF024765 06EF038044 06FF007338 06FF018169 06EF006993 06EF010542 06EF015839 06EF023995 06EF024766 06EF038045 06FF007339 06FF018170 06EF006994 06EF010543 06EF015840 06EF023996 06EF024767 06EF038046 06FF007340 06FF018171 06EF006995 06EF010544 06EF015841 06EF023997 06EF024768 06EF038047 06FF007341 06FF018172 06EF006996 06EF014690 06EF015842 06EF023998 06EF024769 06EF038048 06FF007342 06FF018173 06EF006997 06EF014691 06EF015843 06EF023999 06EF024770 06EF038049 06FF007343 06FF018174 06EF006998 06EF014692 06EF015844 06EF024000 06EF027907 06EF038050 06FF007344 06FF018175 06EF006999 06EF014693 06EF015845 06EF024001 06EF027908 06EF038051 06FF007345 06FF018176 06EF007000 06EF014694 06EF015846 06EF024002 06EF027909 06EF038052 06FF007346 06FF018177 06EF007001 06EF014695 06EF015847 06EF024003 06EF027910 06EF038053 06FF007347 06FF018178 06EF007002 06EF014696 06EF015848 06EF024004 06EF027911 06EF038054 06FF007348 06FF018179 06EF007003 06EF014697 06EF015849 06EF024005 06EF027912 06EF038055 06FF007349 06FF018180 06EF007004 06EF014698 06EF015850 06EF024006 06EF027913 06EF038056 06FF007350 06FF018181 06EF007005 06EF014705 06EF015851 06EF024007 06EF027914 06EF038057 06FF007351 06FF018182 06EF007006 06EF014706 06EF016696 06EF024008 06EF027915 06EF038058 06FF007352 06FF018183 06EF007007 06EF014707 06EF016697 06EF024009 06EF027916 06EF038059 06FF007353 06FF018184 06EF007008 06EF014708 06EF016698 06EF024010 06EF029988 06EF038063 06FF007354 06FF018185 06EF007009 06EF014709 06EF016699 06EF024011 06EF029989 06EF039050 06FF007355 06FF018186 06EF007010 06EF014710 06EF016700 06EF024012 06EF029990 06EF039051 06FF007356 06FF018187 06EF007011 06EF014711 06EF016701 06EF024013 06EF029991 06EF039052 06FF007357 06FF018188 06EF007012 06EF014712 06EF016702 06EF024014 06EF029992 06EF039053 06FF007358 06FF018189 06EF007013 06EF014713 06EF016703 06EF024015 06EF029993 06EF039054 06FF007359 06FF018190 06EF007014 06EF014714 06EF016704 06EF024016 06EF029994 06EF039055 06FF007360 06FF018191 06EF007015 06EF014715 06EF016705 06EF024017 06EF029995 06EF039056 06FF007361 06FF018192 06EF007016 06EF014716 06EF016706 06EF024018 06EF029996 06EF039057 06FF007362 06FF018193 06EF007017 06EF014717 06EF016707 06EF024019 06EF029997 06EF039058 06FF007363 06FF018194 06EF007018 06EF014718 06EF016708 06EF024020 06EF029998 06EF039059 06FF007364 06FF018195 06EF009212 06EF014719 06EF016709 06EF024021 06EF029999 06EF039060 06FF007365 06FF018196 06EF009213 06EF014720 06EF016710 06EF024022 06EF030000 06EF039061 06FF007366 06FF018197 06EF009214 06EF014721 06EF016711 06EF024023 06EF030001 06EF039062 06FF007367 06FF018198 06EF009215 06EF014722 06EF016712 06EF024024 06EF030002 06EF039935 06FF007368 06FF018199 06EF009216 06EF014723 06EF016713 06EF024025 06EF030003 06EF041579 06FF007369 06FF018200 06EF009217 06EF014724 06EF016714 06EF024026 06EF030004 06EF041580 06FF007370 06FF018201 06EF009218 06EF014725 06EF016715 06EF024027 06EF030005 06EF041581 06FF007371 06FF018202 06EF009219 06EF014726 06EF016716 06EF024028 06EF030006 06EF041582 06FF007372 06FF018203 06EF009220 06EF014727 06EF016717 06EF024029 06EF030007 06EF041583 06FF007373 06FF018204 06EF009221 06EF014728 06EF016718 06EF024030 06EF030008 06EF041584 06FF007374 06FF018205 06EF009222 06EF014729 06EF016719 06EF024031 06EF030009 06EF041585 06FF007375 06FF018206 06EF009223 06EF014730 06EF016720 06EF024032 06EF030010 06EF041586 06FF007376 06FF018207 06EF009224 06EF014731 06EF016721 06EF024033 06EF030011 06EF041588 06FF007377 06FF018208 06EF009225 06EF014732 06EF016722 06EF024034 06EF030012 06EF041591 06FF007378 06FF018209 06EF009226 06EF014734 06EF016723 06EF024035 06EF030013 06EF041593 06FF007379 06FF018210 06EF009227 06EF014735 06EF016724 06EF024036 06EF030014 06EF041594 06FF007380 06FF018211 06EF010442 06EF014736 06EF016725 06EF024037 06EF030015 06EF041595 06FF007381 06FF018212 06EF010443 06EF014737 06EF016726 06EF024139 06EF030016 06EF041596 06FF009816 06FF018213 06EF010444 06EF014738 06EF016727 06EF024140 06EF030017 06EF041597 06FF009817 06FF018214 06EF010445 06EF014739 06EF016728 06EF024141 06EF030018 06EF041598 06FF009818 06FF018216 06FF018217 06FF033953 06FF038551 06FF047441 06GF016272 06FF018218 06FF033954 06FF038552 06FF048159 06GF016273 06FF018219 06FF033955 06FF038553 06FF048160 06GF016274 06FF018220 06FF033956 06FF038554 06FF048161 06GF016275 06FF018221 06FF033957 06FF038555 06FF048162 06GF016276 06FF018222 06FF033958 06FF038559 06FF048163 06GF016277 06FF018223 06FF033959 06FF038560 06GF000401 06GF016278 06FF018224 06FF037043 06FF038561 06GF000402 06GF016279 06FF018225 06FF037044 06FF038562 06GF000403 06GF016281 06FF018226 06FF037045 06FF038563 06GF000404 06GF016287 06FF018227 06FF037046 06FF038564 06GF000405 06GF016288 06FF018228 06FF037047 06FF038565 06GF000406 06GF016289 06FF018229 06FF037048 06FF038570 06GF000407 06GF016290 06FF018230 06FF037049 06FF038571 06GF000408 06GF016291 06FF018231 06FF037050 06FF038572 06GF000409 06GF016292 06FF018232 06FF037051 06FF038573 06GF000410 06GF016293 06FF018233 06FF037052 06FF038574 06GF000411 06GF016294 06FF018234 06FF037053 06FF038575 06GF000412 06GF022752 06FF018235 06FF037054 06FF038576 06GF000413 06GF022753 06FF018236 06FF038470 06FF038577 06GF000414 06GF022754 06FF018237 06FF038471 06FF038578 06GF000415 06GF022755 06FF018238 06FF038472 06FF038579 06GF000416 06GF022756 06FF018239 06FF038473 06FF038580 06GF000417 06GF022757 06FF018240 06FF038474 06FF038581 06GF000418 06GF022770 06FF018241 06FF038475 06FF038582 06GF000419 06GF022771 06FF018242 06FF038476 06FF038583 06GF000420 06GF022772 06FF023914 06FF038477 06FF038584 06GF008944 06GF022773 06FF023915 06FF038478 06FF038585 06GF010962 06GF022774 06FF023916 06FF038479 06FF038586 06GF010963 06GF022775 06FF023917 06FF038480 06FF038587 06GF010964 06GF022776 06FF023918 06FF038481 06FF038588 06GF010966 06GF022777 06FF023919 06FF038482 06FF038589 06GF010967 06GF022778 06FF023920 06FF038483 06FF038590 06GF010968 06GF022779 06FF023921 06FF038484 06FF038591 06GF010969 06GF022780 06FF023922 06FF038485 06FF038592 06GF010970 06GF02278106FF023923 06FF038486 06FF038593 06GF010971 06GF022782 06FF023924 06FF038487 06FF038594 06GF010972 06GF022783 06FF023925 06FF038488 06FF038595 06GF010974 06GF022784 06FF023926 06FF038489 06FF038596 06GF010975 06GF022785 06FF023927 06FF038490 06FF038597 06GF010976 06GF022786 06FF023928 06FF038491 06FF038598 06GF010977 06GF022787 06FF023929 06FF038492 06FF038599 06GF010978 06GF027976 06FF029435 06FF038493 06FF038600 06GF012758 06GF027977 06FF029436 06FF038494 06FF038601 06GF012759 06GF027978 06FF029437 06FF038495 06FF038602 06GF012760 06GF027979 06FF029438 06FF038496 06FF038603 06GF012761 06GF027980 06FF029439 06FF038497 06FF038604 06GF012762 06GF027981 06FF029440 06FF038498 06FF038605 06GF012763 06GF027982 06FF029441 06FF038499 06FF038606 06GF012764 06GF027983 06FF029442 06FF038500 06FF038607 06GF012765 06GF027984 06FF029443 06FF038501 06FF038608 06GF012766 06GF027985 06FF029444 06FF038502 06FF038609 06GF012767 06GF027986 06FF029445 06FF038503 06FF041453 06GF012768 06GF027987 06FF029446 06FF038504 06FF041455 06GF012769 06GF027988 06FF029447 06FF038505 06FF041457 06GF012770 06GF027989 06FF029448 06FF038506 06FF041458 06GF012771 06GF027990 06FF029449 06FF038507 06FF041459 06GF012772 06GF027991 06FF029450 06FF038508 06FF041460 06GF012773 06GF027992 06FF029451 06FF038509 06FF041461 06GF012774 06GF027993 06FF029452 06FF038510 06FF041462 06GF012775 06GF027994 06FF029453 06FF038511 06FF041463 06GF012776 06GF027997 06FF029454 06FF038512 06FF041464 06GF012777 06GF027998 06FF029455 06FF038513 06FF041465 06GF012778 06GF028002 06FF029456 06FF038514 06FF041466 06GF012779 06GF028003 06FF030551 06FF038515 06FF041467 06GF012780 06GF028004 06FF030552 06FF038516 06FF041468 06GF012781 06FF030553 06FF038517 06FF041469 06GF012782 06FF030554 06FF038518 06FF041470 06GF012783 06FF030555 06FF038519 06FF041471 06GF012784 06FF030556 06FF038520 06FF042745 06GF012785 06FF030557 06FF038521 06FF042746 06GF012786 06FF030558 06FF038522 06FF042747 06GF012787 06FF030559 06FF038523 06FF042748 06GF016241 06FF030560 06FF038524 06FF042749 06GF016242 06FF030561 06FF038525 06FF042750 06GF016243 06FF030962 06FF038526 06FF042751 06GF016244 06FF030963 06FF038527 06FF042752 06GF016245 06FF030964 06FF038528 06FF042753 06GF016246 06FF030965 06FF038529 06FF042754 06GF016247 06FF030966 06FF038530 06FF042755 06GF016248 06FF030967 06FF038531 06FF042756 06GF016250 06FF030968 06FF038532 06FF043261 06GF016251 06FF030969 06FF038533 06FF043262 06GF016252 06FF030970 06FF038534 06FF043263 06GF016253 06FF030971 06FF038535 06FF043264 06GF016254 06FF030972 06FF038536 06FF045037 06GF016255 06FF030973 06FF038537 06FF045038 06GF016258 06FF030974 06FF038538 06FF045039 06GF016259 06FF033940 06FF038539 06FF045040 06GF016260 06FF033941 06FF038540 06FF045041 06GF016261 06FF033943 06FF038541 06FF045042 06GF016262 06FF033944 06FF038542 06FF045043 06GF016263 06FF033945 06FF038543 06FF045044 06GF016264 06FF033946 06FF038544 06FF045045 06GF016265 06FF033947 06FF038545 06FF045046 06GF016266 06FF033948 06FF038546 06FF045047 06GF016267 06FF033949 06FF038547 06FF045048 06GF016268 06FF033950 06FF038548 06FF045049 06GF016269 06FF033951 06FF038549 06FF045050 06GF016270 06FF033952 06FF038550 06FF045051 06GF016271 |
| FEI Number |
1031452
|
Recalling Firm/ Manufacturer |
Invacare Corporation
2101 E. Lake Mary Blvd.
Sanford
FL
32773-7141
|
Manufacturer Reason for Recall | The incorrect caution label that describes washing instructions was applied to beds distributed in the US, i.e. the specified label is P/N 1116654 (English translation); the label that was used is P/N 1119165 (French translation). |
FDA Determined Cause 2 | Other |
| Action | E-mails and letters regarding this correction provided to the consignees on September 27, 2006, and to appropriate Invacare personnel for information purposes. Consignees are being notified via delivery of a field correction package. Once the packages have been sent, a notice will be posted on the Invacare website. A follow-up contact will be scheduled in 2 to 3 months to try to reach those who have not returned their tracking sheet indicating completion of their corrective action. |
| Quantity in Commerce | 2065 |
| Distribution | United States and Canada. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|