| Date Initiated by Firm | August 30, 2007 |
| Date Posted | December 12, 2007 |
| Recall Status1 |
Terminated 3 on September 08, 2008 |
| Recall Number | Z-0459-2008 |
| Recall Event ID |
38897 |
| 510(K)Number | K033734 |
| Product Classification |
Capillaries - Product Code JJF
|
| Product | Roche Diagnostics LightCycler Capillaries (100uL) for use in conjunction with the LightCycler analyzers; Catalog #: 03337090001, plus any capillaries received at the time of installation of the LightCycler instruments or systems **, Roche Diagnostics Corporation, Indianapolis 46250.(**LightCycler 2.0 System, Catalog Number: 04391039001; Batch Number Range: 1415471-1415933; LightCycler 2.0 Instrument-Laptop; Catalog Number: 03603946001; Batch Number Range; 1415434-1415734; LightCycler 2.0 Instrument-Desktop; Catalog Number: 03531414201; Batch Number Range: 1415271-1415732 |
| Code Information |
Lots 34979900, exp. 2/28/08; 35290100, exp. 2/28/08; 35293300 and 35296200 plus any capillaries received at the time of installation of the LightCycler instruments or systems**. (**LightCycler 2.0 System, Catalog Number: 04391039001; Batch Number Range: 1415471-1415933; LightCycler 2.0 Instrument-Laptop; Catalog Number: 03603946001; Batch Number Range; 1415434-1415734; LightCycler 2.0 Instrument-Desktop; Catalog Number: 03531414201; Batch Number Range: 1415271-1415732) |
| FEI Number |
1823260
|
Recalling Firm/ Manufacturer |
Roche Diagnostics Corp.
9115 Hague Rd
Indianapolis
IN
46250-0416
|
| For Additional Information Contact | 800-526-1247 |
Manufacturer Reason for Recall | False Negative Results: Air bubbles in the lens or the area surrounding the lens of the LightCycler 100 uL capillaries may cause false negative results. The affected capillaries showed up to a 40% reduction in signal |
FDA Determined Cause 2 | Process design |
| Action | Consignees were notified by Urgent Medical Device Recall letter dated 8/30/07, to discontinue use of the 100uL capillaries, plus all of those received at installation, and to disposed of them per local guidelines. |
| Quantity in Commerce | 119 boxes of 96 capillaries each |
| Distribution | Nationwide |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|
| 510(K) Database | 510(K)s with Product Code = JJF
|