| | Class 2 Device Recall Edwards Vigilance II Monitor |  |
| Date Initiated by Firm | May 17, 2007 |
| Date Posted | November 24, 2007 |
| Recall Status1 |
Terminated 3 on October 09, 2009 |
| Recall Number | Z-0262-2008 |
| Recall Event ID |
45464 |
| 510(K)Number | K043103 |
| Product Classification |
Vigilance II Monitor - Product Code DXG
|
| Product | Edwards Vigilance II Continuous Cardiac Output/Oximetry/Volumetric (CCO/SvO2/CEDV) Monitor Model: VIG2/VIG2E Software Versions: 00.51,00.60, 00.61, 00.62, 00.63, 00.64, 00.67, 00.69, Edwards Lifesciences LLC., Irvine, CA 92614 |
| Code Information |
VG000034 VG000049 VG000114 VG000124 VG000125 VG000128 VG000128 VG000129 VG000131 VG000132 VG000133 VG000134 VG000135 VG000136 VG000137 VG000138 VG000139 VG000140 VG000142 VG000143 VG000144 VG000145 VG000146 VG000148 VG000149 VG000150 VG000151 VG000152 VG000153 VG000155 VG000157 VG000163 VG000164 VG000169 VG000170 VG000171 VG000172 VG000173 VG000175 VG000176 VG000177 VG000178 VG000180 VG000181 VG000183 VG000185 VG000187 VG000188 VG000189 VG000191 VG000192 VG000195 VG000199 VG000200 VG000201 VG000203 VG000204 VG000206 VG000207 VG000209 VG000210 VG000211 VG000212 VG000213 VG000214 VG000215 VG000216 VG000220 VG000224 VG000225 VG000229 VG000230 VG000231 VG000232 VG000237 VG000238 VG000239 VG000250 VG000251 VG000254 VG000255 VG000256 VG000257 VG000259 VG000280 VG000289 VG000295 VG000296 VG000297 VG000299 VG000302 VG000304 VG000308 VG000320 VG000322 VG000323 VG000326 VG000333 VG000335 VG000336 VG000340 VG000343 VG000345 VG000346 VG000348 VG000351 VG000352 VG000353 VG000354 VG000355 VG000360 VG000364 VG000368 VG000369 VG000370 VG000371 VG000372 VG000381 VG000382 VG000384 VG000387 VG000391 VG000393 VG000394 VG000400 VG000402 VG000418 VG000419 VG000424 VG000425 VG000426 VG000427 VG000428 VG000429 VG000433 VG000436 VG000438 VG000440 VG000442 VG000449 VG000450 VG000453 VG000455 VG000457 VG000458 VG000461 VG000462 VG000463 VG000464 VG000466 VG000467 VG000470 VG000471 VG000473 VG000475 VG000476 VG000477 VG000478 VG000479 VG000480 VG000481 VG000482 VG000483 VG000484 VG000485 VG000486 VG000487 VG000488 VG000490 VG000491 VG000492 VG000492 VG000494 VG000495 VG000498 VG000499 VG000500 VG000501 VG000503 VG000504 VG000508 VG000509 VG000510 VG000514 VG000515 VG000517 VG000518 VG000519 VG000520 VG000521 VG000522 VG000523 VG000524 VG000526 VG000527 VG000528 VG000529 VG000530 VG000531 VG000531a VG000532 VG000533 VG000534 VG000535 VG000535a VG000536 VG000537 VG000538 VG000539 VG000540 VG000541 VG000542 VG000543 VG000544 VG000545 VG000546 VG000547 VG000548 VG000549 VG000550 VG000551 VG000552 VG000553 VG000554 VG000557 VG000558 VG000560 VG000561 VG000562 VG000563 VG000565 VG000566 VG000567 VG000568 VG000570 VG000571 VG000573 VG000575 VG000576 VG000577 VG000578 VG000579 VG000581 VG000582 VG000585 VG000586 VG000587 VG000588 VG000589 VG000590 VG000591 VG000592 VG000593 VG000594 VG000595 VG000596 VG000597 VG000599 VG000600 VG000601 VG000602 VG000603 VG000604 VG000606 VG000608 VG000609 VG000612 VG000613 VG000614 VG000616 VG000618 VG000619 VG000620 VG000621 VG000623 VG000626 VG000627 VG000629 VG000631 VG000633 VG000635 VG000636 VG000637 VG000639 VG000640 VG000641 VG000642 VG000643 VG000644 VG000647 VG000648 VG000649 VG000650 VG000652 VG000653 VG000654 VG000656 VG000659 VG000660 VG000661 VG000661 VG000662 VG000664 VG000665 VG000667 VG000668 VG000671 VG000673 VG000674 VG000675 VG000677 VG000678 VG000679 VG000679 VG000680 VG000681 VG000682 VG000684 VG000685 VG000687 VG000690 VG000691 VG000692 VG000694 VG000695 VG000697 VG000698 VG000700 VG000701 VG000702 VG000703 VG000704 VG000705 VG000706 VG000707 VG000708 VG000709 VG000710 VG000711 VG000712 VG000713 VG000714 VG000715 VG000717 VG000719 VG000722 VG000723 VG000724 VG000725 VG000728 VG000729 VG000731 VG000732 VG000733 VG000735 VG000736 VG000739 VG000740 VG000741 VG000742 VG000745 VG000746 VG000748 VG000749 VG000751 VG000753 VG000755 VG000757 VG000758 VG000759 VG000760 VG000761 VG000762 VG000763 VG000764 VG000765 VG000766 VG000768 VG000769 VG000770 VG000771 VG000772 VG000773 VG000774 VG000777 VG000778 VG000779 VG000780 VG000785 VG000786 VG000787 VG000789 VG000791 VG000792 VG000813 VG000857 VG000882 VG000883 VG000884 VG000885 VG000886 VG000887 VG000888 VG000889 VG000890 VG000891 VG000892 VG000894 VG000895 VG000896 VG000897 VG000898 VG000900 VG000901 VG000903 VG000904 VG000905 VG000906 VG000907 VG000908 VG000910 VG000911 VG000912 VG000914 VG000915 VG000918 VG000919 VG000920 VG000921 VG000922 VG000923 VG000925 VG000927 VG000928 VG000929 VG000930 VG000931 VG000932 VG000935 VG000936 VG000937 VG000938 VG000940 VG000941 VG000942 VG000947 VG000948 VG000951 VG000952 VG000953 VG000955 VG000957 VG000958 VG000961 VG000962 VG000963 VG000964 VG000965 VG000966 VG000967 VG000969 VG000970 VG000971 VG000972 VG000973 VG000974 VG000975 VG000976 VG000977 VG000978 VG000979 VG000980 VG000981 VG000982 VG000983 VG000984 VG000985 VG000986 VG000987 VG000988 VG000989 VG000990 VG000991 VG000992 VG000993 VG000994 VG000995 VG000997 VG000998 VG000999 VG001000 VG001001 VG001003 VG001004 VG001005 VG001007 VG001008 VG001012 VG001013 VG001014 VG001015 VG001017 VG001018 VG001019 VG001020 VG001021 VG001022 VG001023 VG001024 VG001025 VG001026 VG001027 VG001029 VG001031 VG001032 VG001034 VG001035 VG001038 VG001041 VG001042 VG001043 VG001044 VG001045 VG001048 VG001049 VG001050 VG001052 VG001053 VG001055 VG001056 VG001057 VG001059 VG001061 VG001063 VG001071 VG001072 VG001074 VG001076 VG001077 VG001078 VG001079 VG001080 VG001081 VG001082 VG001084 VG001087 VG001088 VG001089 VG001091 VG001092 VG001096 VG001097 VG001100 VG001101 VG001102 VG001104 VG001106 VG001108 VG001110 VG001112 VG001113 VG001116 VG001117 VG001118 VG001120 VG001122 VG001124 VG001125 VG001126 VG001127 VG001128 VG001129 VG001130 VG001131 VG001132 VG001134 VG001135 VG001136 VG001137 VG001138 VG001140 VG001141 VG001143 VG001144 VG001145 VG001146 VG001147 VG001148 VG001149 VG001150 VG001151 VG001152 VG001153 VG001154 VG001155 VG001156 VG001157 VG001158 VG001159 VG001160 VG001161 VG001162 VG001164 VG001165 VG001166 VG001167 VG001169 VG001170 VG001173 VG001174 VG001175 VG001176 VG001177 VG001179 VG001181 VG001185 VG001186 VG001188 VG001189 VG001190 VG001191 VG001192 VG001193 VG001194 VG001195 VG001197 VG001201 VG001202 VG001203 VG001204 VG001206 VG001208 VG001210 VG001211 VG001214 VG001215 VG001216 VG001217 VG001218 VG001220 VG001222 VG001223 VG001226 VG001227 VG001229 VG001230 VG001231 VG001234 VG001239 VG001241 VG001244 VG001245 VG001246 VG001247 VG001248 VG001249 VG001250 VG001251 VG001252 VG001253 VG001254 VG001255 VG001256 VG001257 VG001257 VG001258 VG001259 VG001260 VG001262 VG001263 VG001264 VG001265 VG001266 VG001267 VG001268 VG001269 VG001270 VG001272 VG001275 VG001277 VG001281 VG001282 VG001284 VG001287 VG001289 VG001291 VG001294 VG001309 VG001311 VG001313 VG001321 VG001322 VG001324 VG001325 VG001326 VG001327 VG001328 VG001329 VG001330 VG001332 VG001335 VG001337 VG001338 VG001339 VG001340 VG001341 VG001342 VG001344 VG001345 VG001346 VG001347 VG001348 VG001349 VG001350 VG001351 VG001358 VG001360 VG001362 VG001367 VG001368 VG001369 VG001370 VG001371 VG001373 VG001380 VG001391 VG001398 VG001399 VG001406 VG001407 VG001418 VG001423 VG001429 VG001452 VG001453 VG001454 VG001455 VG001456 VG001458 VG001460 VG001461 VG001462 VG001463 VG001464 VG001465 VG001467 VG001468 VG001469 VG001470 VG001471 VG001472 VG001473 VG001474 VG001475 VG001476 VG001477 VG001478 VG001479 VG001480 VG001481 VG001482 VG001483 VG001484 VG001485 VG001486 VG001487 VG001488 VG001489 VG001490 VG001491 VG001492 VG001493 VG001494 VG001495 VG001496 VG001497 VG001498 VG001499 VG001500 VG001501 VG001502 VG001503 VG001504 VG001505 VG001506 VG001507 VG001508 VG001509 VG001510 VG001511 VG001512 VG001513 VG001516 VG001521 VG001530 VG001540 VG001542 VG001543 VG001544 VG001545 VG001546 VG001547 VG001552 VG001555 VG001556 VG001558 VG001560 VG001561 VG001562 VG001565 VG001566 VG001567 VG001569 VG001571 VG001574 VG001576 VG001577 VG001578 VG001579 VG001583 VG001584 VG001587 VG001595 VG001595 VG001596 VG001598 VG001599 VG001599 VG001602 VG001605 VG001606 VG001607 VG001608 VG001609 VG001612 VG001612 VG001613 VG001614 VG001615 VG001616 VG001617 VG001618 VG001619 VG001622 VG001623 VG001624 VG001625 VG001627 VG001628 VG001629 VG001630 VG001631 VG001633 VG001634 VG001636 VG001637 VG001639 VG001640 VG001643 VG001644 VG001647 VG001650 VG001654 VG001657 VG001666 VG001667 VG001675 VG001679 VG001685 VG001690 VG001692 VG001694 VG001697 VG001699 VG001702 VG001703 VG001707 VG001708 VG001709 VG001711 VG001712 VG001714 VG001715 VG001716 VG001717 VG001719 VG001720 VG001721 VG001722 VG001724 VG001725 VG001727 VG001728 VG001729 VG001730 VG001731 VG001732 VG001736 VG001737 VG001738 VG001739 VG001740 VG001741 VG001742 VG001743 VG001744 VG001745 VG001746 VG001748 VG001749 VG001750 VG001751 VG001752 VG001753 VG001754 VG001755 VG001756 VG001758 VG001759 VG001761 VG001763 VG001764 VG001765 VG001766 VG001767 VG001768 VG001769 VG001771 VG001772 VG001773 VG001775 VG001776 VG001777 VG001779 VG001780 VG001782 VG001783 VG001784 VG001785 VG001786 VG001787 VG001788 VG001789 VG001790 VG001791 VG001792 VG001793 VG001794 VG001795 VG001796 VG001797 VG001798 VG001799 VG001800 VG001801 VG001802 VG001803 VG001804 VG001806 VG001807 VG001808 VG001809 VG001811 VG001812 VG001813 VG001814 VG001816 VG001818 VG001819 VG001820 VG001821 VG001822 VG001823 VG001824 VG001825 VG001826 VG001827 VG001828 VG001829 VG001830 VG001832 VG001833 VG001834 VG001835 VG001836 VG001838 VG001839 VG001840 VG001841 VG001842 VG001843 VG001844 VG001845 VG001846 VG001847 VG001849 VG001850 VG001851 VG001852 VG001853 VG001854 VG001855 VG001856 VG001857 VG001858 VG001859 VG001860 VG001861 VG001862 VG001863 VG001865 VG001866 VG001867 VG001868 VG001870 VG001871 VG001872 VG001874 VG001878 VG001879 VG001880 VG001881 VG001884 VG001887 VG001889 VG001890 VG001891 VG001892 VG001893 VG001894 VG001895 VG001896 VG001897 VG001898 VG001901 VG001902 VG001905 VG001906 VG001907 VG001909 VG001910 VG001911 VG001912 VG001913 VG001915 VG001916 VG001918 VG001919 VG001921 VG001926 VG001927 VG001934 VG001936 VG001938 VG001940 VG001942 VG001943 VG001945 VG001946 VG001947 VG001948 VG001949 VG001950 VG001951 VG001952 VG001953 VG001954 VG001956 VG001957 VG001958 VG001959 VG001960 VG001961 VG001962 VG001964 VG001965 VG001966 VG001968 VG001969 VG001970 VG001971 VG001973 VG001974 VG001975 VG001976 VG001977 VG001978 VG001979 VG001980 VG001982 VG001983 VG001984 VG001985 VG001987 VG001988 VG001989 VG001990 VG001991 VG001992 VG001993 VG001994 VG001995 VG001996 VG001997 VG001999 VG002000 VG002001 VG002003 VG002004 VG002005 VG002006 VG002007 VG002009 VG002010 VG002011 VG002012 VG002015 VG002016 VG002017 VG002018 VG002019 VG002021 VG002022 VG002024 VG002028 VG002029 VG002030 VG002031 VG002033 VG002035 VG002038 VG002041 VG002050 VG002057 VG002065 VG002067 VG002068 VG002071 VG002072 VG002078 VG002083 VG002085 VG002089 VG002092 VG002095 VG002099 VG002100 VG002102 VG002105 VG002108 VG002109 VG002112 VG002113 VG002115 VG002118 VG002119 VG002120 VG002129 VG002130 VG002131 VG002132 VG002133 VG002134 VG002135 VG002136 VG002139 VG002140 VG002142 VG002143 VG002145 VG002146 VG002147 VG002149 VG002151 VG002152 VG002153 VG002155 VG002156 VG002157 VG002158 VG002159 VG002160 VG002161 VG002162 VG002163 VG002164 VG002172 VG002189 VG002191 VG002192 VG002193 VG002194 VG002195 VG002196 VG002197 VG002198 VG002199 VG002200 VG002201 VG002202 VG002203 VG002205 VG002206 VG002207 VG002208 VG002209 VG002211 VG002212 VG002213 VG002213 VG002214 VG002216 VG002217 VG002218 VG002219 VG002220 VG002222 VG002223 VG002224 VG002225 VG002225 VG002226 VG002227 VG002228 VG002229 VG002230 VG002231 VG002234 VG002235 VG002236 VG002236 VG002237 VG002238 VG002240 VG002241 VG002242 VG002243 VG002243 VG002244 VG002245 VG002246 VG002247 VG002248 VG002249 VG002250 VG002251 VG002252 VG002254 VG002255 VG002256 VG002258 VG002260 VG002261 VG002262 VG002263 VG002264 VG002265 VG002266 VG002267 VG002268 VG002269 VG002270 VG002271 VG002272 VG002273 VG002275 VG002276 VG002277 VG002278 VG002280 VG002281 VG002284 VG002285 VG002286 VG002287 VG002287 VG002288 VG002289 VG002290 VG002291 VG002293 VG002294 VG002294 VG002296 VG002298 VG002299 VG002300 VG002301 VG002306 VG002308 VG002309 VG002310 VG002319 VG002322 VG002323 VG002325 VG002326 VG002329 VG002331 VG002332 VG002333 VG002334 VG002335 VG002336 VG002337 VG002338 VG002339 VG002340 VG002341 VG002342 VG002343 VG002344 VG002345 VG002346 VG002347 VG002348 VG002349 VG002350 VG002351 VG002352 VG002353 VG002354 VG002355 VG002356 VG002357 VG002358 VG002359 VG002360 VG002362 VG002363 VG002364 VG002365 VG002366 VG002367 VG002368 VG002369 VG002370 VG002371 VG002372 VG002373 VG002374 VG002375 VG002376 VG002377 VG002378 VG002380 VG002381 VG002382 VG002383 VG002384 VG002386 VG002387 VG002388 VG002389 VG002391 VG002392 VG002393 VG002394 VG002395 VG002396 VG002397 VG002398 VG002400 VG002401 VG002402 VG002403 VG002404 VG002405 VG002412 VG002416 VG002423 VG002425 VG002426 VG002427 VG002428 VG002430 VG002431 VG002432 VG002433 VG002435 VG002436 VG002437 VG002438 VG002439 VG002440 VG002441 VG002442 VG002443 VG002444 VG002445 VG002446 VG002447 VG002448 VG002449 VG002450 VG002452 VG002453 VG002454 VG002455 VG002457 VG002458 VG002459 VG002460 VG002461 VG002462 VG002463 VG002464 VG002465 VG002466 VG002467 VG002468 VG002469 VG002472 VG002473 VG002474 VG002476 VG002478 VG002479 VG002481 VG002482 VG002485 VG002486 VG002488 VG002492 VG002493 VG002493 VG002496 VG002498 VG002499 VG002500 VG002503 VG002506 VG002507 VG002509 VG002515 VG002518 VG002519 VG002521 VG002523 VG002524 VG002526 VG002527 VG002528 VG002530 VG002531 VG002533 VG002535 VG002536 VG002537 VG002538 VG002539 VG002540 VG002541 VG002542 VG002543 VG002544 VG002545 VG002546 VG002547 VG002548 VG002549 VG002550 VG002551 VG002552 VG002554 VG002555 VG002556 VG002557 VG002558 VG002559 VG002560 VG002562 VG002563 VG002564 VG002565 VG002566 VG002567 VG002568 VG002569 VG002570 VG002571 VG002572 VG002573 VG002574 VG002575 VG002576 VG002577 VG002579 VG002580 VG002581 VG002582 VG002584 VG002585 VG002586 VG002589 VG002591 VG002592 VG002593 VG002596 VG002597 VG002598 VG002599 VG002600 VG002602 VG002603 VG002604 VG002605 VG002606 VG002607 VG002608 VG002609 VG002611 VG002612 VG002613 VG002615 VG002628 VG002631 VG002632 VG002637 VG002638 VG002639 VG002640 VG002642 VG002643 VG002644 VG002647 VG002648 VG002649 VG002650 VG002651 VG002652 VG002654 VG002659 VG002660 VG002662 VG002667 VG002668 VG002669 VG002673 VG002675 VG002679 VG002680 VG002681 VG002685 VG002686 VG002689 VG002692 VG002692 VG002696 VG002697 VG002698 VG002699 VG002701 VG002706 VG002707 VG002708 VG002708 VG002709 VG002710 VG002715 VG002716 VG002717 VG002719 VG002720 VG002721 VG002722 VG002724 VG002725 VG002726 VG002727 VG002729 VG002730 VG002731 VG002732 VG002733 VG002734 VG002735 VG002736 VG002738 VG002739 VG002740 VG002741 VG002742 VG002745 VG002746 VG002749 VG002750 VG002751 VG002752 VG002753 VG002754 VG002755 VG002756 VG002757 VG002759 VG002760 VG002761 VG002764 VG002765 VG002766 VG002768 VG002769 VG002770 VG002771 VG002772 VG002773 VG002774 VG002775 VG002776 VG002777 VG002780 VG002781 VG002783 VG002784 VG002785 VG002786 VG002787 VG002788 VG002789 VG002791 VG002793 VG002793 VG002794 VG002795 VG002796 VG002798 VG002799 VG002800 VG002801 VG002802 VG002804 VG002805 VG002806 VG002807 VG002808 VG002809 VG002810 VG002811 VG002812 VG002814 VG002816 VG002818 VG002819 VG002821 VG002822 VG002824 VG002825 VG002826 VG002827 VG002828 VG002829 VG002831 VG002832 VG002833 VG002834 VG002835 VG002836 VG002838 VG002839 VG002840 VG002841 VG002842 VG002843 VG002844 VG002845 VG002848 VG002849 VG002850 VG002852 VG002853 VG002854 VG002855 VG002857 VG002858 VG002859 VG002860 VG002861 VG002862 VG002863 VG002864 VG002866 VG002868 VG002869 VG002870 VG002871 VG002872 VG002873 VG002874 VG002875 VG002876 VG002877 VG002878 VG002880 VG002882 VG002883 VG002885 VG002886 VG002888 VG002890 VG002891 VG002898 VG002901 VG002903 VG002905 VG002907 VG002908 VG002909 VG002913 VG002923 VG002930 VG002932 VG002933 VG002936 VG002937 VG002938 VG002939 VG002940 VG002941 VG002942 VG002943 VG002944 VG002945 VG002946 VG002948 VG002949 VG002950 VG002951 VG002952 VG002953 VG002955 VG002956 VG002957 VG002958 VG002959 VG002960 VG002961 VG002962 VG002963 VG002964 VG002966 VG002967 VG002968 VG002969 VG002970 VG002971 VG002972 VG002973 VG002974 VG002976 VG002977 VG002978 VG002979 VG002980 VG002981 VG002982 VG002983 VG002984 VG002985 VG002986 VG002987 VG002988 VG002989 VG002990 VG002991 VG002992 VG002993 VG002994 VG002995 VG002996 VG002997 VG002998 VG002999 VG003000 VG003001 VG003002 VG003003 VG003004 VG003005 VG003007 VG003008 VG003009 VG003010 VG003011 VG003012 VG003014 VG003015 VG003016 VG003017 VG003018 VG003019 VG003020 VG003021 VG003022 VG003023 VG003024 VG003025 VG003026 VG003027 VG003029 VG003030 VG003031 VG003032 VG003033 VG003034 VG003035 VG003037 VG003040 VG003042 VG003043 VG003047 VG003049 VG003052 VG003053 VG003055 VG003056 VG003057 VG003058 VG003060 VG003062 VG003064 VG003072 VG003073 VG003082 VG003084 VG003087 VG003088 VG003089 VG003090 VG003091 VG003092 VG003093 VG003094 VG003095 VG003096 VG003097 VG003099 VG003100 VG003101 VG003102 VG003103 VG003104 VG003106 VG003107 VG003108 VG003109 VG003110 VG003111 VG003112 VG003113 VG003114 VG003115 VG003116 VG003117 VG003118 VG003119 VG003120 VG003121 VG003122 VG003123 VG003125 VG003126 VG003127 VG003128 VG003129 VG003130 VG003132 VG003133 VG003134 VG003135 VG003137 VG003139 VG003141 VG003143 VG003148 VG003149 VG003150 VG003152 VG003154 VG003155 VG003156 VG003157 VG003158 VG003159 VG003161 VG003163 VG003164 VG003165 VG003167 VG003171 VG003173 VG003174 VG003178 VG003183 VG003184 VG003185 VG003187 VG003188 VG003190 VG003192 VG003193 VG003194 VG003195 VG003197 VG003199 VG003200 VG003201 VG003203 VG003204 VG003206 VG003207 VG003208 VG003210 VG003211 VG003213 VG003215 VG003216 VG003217 VG003218 VG003219 VG003220 VG003221 VG003222 VG003223 VG003225 VG003227 VG003228 VG003230 VG003231 VG003232 VG003234 VG003235 VG003238 VG003239 VG003240 VG003241 VG003254 VG003263 VG003266 VG003270 VG003271 VG003273 VG003275 VG003281 VG003282 VG003283 VG003285 VG003287 VG003289 VG003291 VG003294 VG003297 VG003300 VG003302 VG003307 VG003309 VG003313 VG003314 VG003319 VG003322 VG003323 VG003324 VG003325 VG003326 VG003327 VG003330 VG003331 VG003332 VG003333 VG003334 VG003335 VG003336 VG003337 VG003338 VG003339 VG003340 VG003341 VG003342 VG003343 VG003346 VG003347 VG003348 VG003351 VG003353 VG003355 VG003356 VG003357 VG003358 VG003360 VG003361 VG003363 VG003364 VG003366 VG003367 VG003369 VG003373 VG003374 VG003375 VG003377 VG003378 VG003380 VG003382 VG003382 VG003385 VG003386 VG003387 VG003388 VG003390 VG003391 VG003392 VG003394 VG003398 VG003399 VG003400 VG003401 VG003402 VG003403 VG003404 VG003405 VG003406 VG003407 VG003409 VG003410 VG003412 VG003413 VG003414 VG003415 VG003416 VG003417 VG003418 VG003419 VG003421 VG003422 VG003423 VG003424 VG003426 VG003428 VG003429 VG003429 VG003431 VG003432 VG003433 VG003435 VG003436 VG003437 VG003439 VG003440 VG003441 VG003442 VG003443 VG003444 VG003445 VG003446 VG003447 VG003448 VG003449 VG003450 VG003451 VG003452 VG003453 VG003454 VG003455 VG003456 VG003458 VG003459 VG003460 VG003461 VG003463 VG003464 VG003465 VG003466 VG003467 VG003468 VG003469 VG003470 VG003471 VG003472 VG003473 VG003474 VG003475 VG003476 VG003478 VG003479 VG003480 VG003481 VG003482 VG003484 VG003485 VG003486 VG003487 VG003488 VG003489 VG003490 VG003491 VG003492 VG003493 VG003495 VG003496 VG003497 VG003498 VG003499 VG003501 VG003502 VG003505 VG003506 VG003507 VG003508 VG003509 VG003513 VG003515 VG003517 VG003519 VG003520 VG003523 VG003525 VG003528 VG003531 VG003535 VG003537 VG003543 VG003545 VG003546 VG003547 VG003549 VG003550 VG003551 VG003552 VG003553 VG003554 VG003557 VG003558 VG003559 VG003562 VG003563 VG003564 VG003565 VG003566 VG003567 VG003568 VG003569 VG003571 VG003571 VG003572 VG003573 VG003574 VG003575 VG003576 VG003577 VG003578 VG003579 VG003580 VG003581 VG003582 VG003584 VG003585 VG003586 VG003587 VG003588 VG003589 VG003590 VG003591 VG003592 VG003593 VG003595 VG003596 VG003597 VG003598 VG003599 VG003600 VG003601 VG003602 VG003604 VG003606 VG003607 VG003610 VG003611 VG003612 VG003613 VG003615 VG003616 VG003617 VG003618 VG003619 VG003620 VG003622 VG003623 VG003624 VG003625 VG003627 VG003629 VG003630 VG003631 VG003632 VG003634 VG003637 VG003638 VG003639 VG003641 VG003642 VG003644 VG003645 VG003647 VG003648 VG003649 VG003650 VG003653 VG003654 VG003655 VG003656 VG003658 VG003660 VG003663 VG003664 VG003665 VG003666 VG003667 VG003668 VG003669 VG003674 VG003675 VG003677 VG003679 VG003681 VG003682 VG003683 VG003684 VG003686 VG003687 VG003688 VG003689 VG003692 VG003694 VG003695 VG003697 VG003700 VG003703 VG003710 VG003711 VG003712 VG003713 VG003714 VG003715 VG003716 VG003717 VG003718 VG003719 VG003720 VG003721 VG003722 VG003723 VG003724 VG003725 VG003726 VG003727 VG003728 VG003729 VG003730 VG003731 VG003732 VG003734 VG003735 VG003737 VG003738 VG003739 VG003740 VG003741 VG003742 VG003743 VG003747 VG003748 VG003749 VG003750 VG003751 VG003751 VG003753 VG003754 VG003755 VG003756 VG003758 VG003759 VG003760 VG003761 VG003762 VG003763 VG003764 VG003765 VG003766 VG003767 VG003768 VG003769 VG003770 VG003772 VG003773 VG003774 VG003775 VG003776 VG003777 VG003778 VG003779 VG003780 VG003781 VG003782 VG003783 VG003785 VG003786 VG003787 VG003788 VG003789 VG003790 VG003790 VG003792 VG003793 VG003794 VG003796 VG003797 VG003798 VG003800 VG003801 VG003802 VG003803 VG003804 VG003805 VG003808 VG003809 VG003811 VG003812 VG003813 VG003814 VG003815 VG003817 VG003818 VG003819 VG003823 VG003826 VG003831 VG003834 VG003835 VG003836 VG003837 VG003838 VG003839 VG003840 VG003842 VG003844 VG003845 VG003848 VG003852 VG003853 VG003856 VG003859 VG003862 VG003863 VG003864 VG003866 VG003867 VG003868 VG003869 VG003870 VG003871 VG003872 VG003873 VG003874 VG003875 VG003877 VG003878 VG003879 VG003880 VG003881 VG003882 VG003883 VG003884 VG003885 VG003886 VG003887 VG003888 VG003889 VG003890 VG003891 VG003892 VG003893 VG003894 VG003895 VG003896 VG003897 VG003898 VG003899 VG003900 VG003902 VG003906 VG003907 VG003908 VG003910 VG003917 VG003921 VG003928 VG003931 VG003932 VG003941 VG003942 VG003944 VG003945 VG003946 VG003948 VG003949 VG003950 VG003951 VG003953 VG003954 VG003955 VG003956 VG003957 VG003958 VG003959 VG003960 VG003961 VG003962 VG003963 VG003964 VG003965 VG003967 VG003968 VG003969 VG003970 VG003971 VG003972 VG003973 VG003977 VG003978 VG003979 VG003980 VG003981 VG003982 VG003983 VG003984 VG003985 VG003987 VG003989 VG003990 VG003991 VG003992 VG003994 VG003995 VG003996 VG003997 VG003998 VG003999 VG004000 VG004002 VG004003 VG004005 VG004006 VG004007 VG004008 VG004010 VG004011 VG004013 VG004015 VG004016 VG004017 VG004018 VG004019 VG004020 VG004021 VG004022 VG004023 VG004024 VG004026 VG004028 VG004029 VG004030 VG004031 VG004032 VG004033 VG004034 VG004035 VG004036 VG004037 VG004038 VG004039 VG004040 VG004041 VG004042 VG004043 VG004044 VG004045 VG004046 VG004049 VG004050 VG004051 VG004052 VG004053 VG004054 VG004055 VG004056 VG004056 VG004057 VG004058 VG004059 VG004060 VG004062 VG004063 VG004064 VG004065 VG004066 VG004067 VG004068 VG004069 VG004070 VG004071 VG004072 VG004073 VG004074 VG004076 VG004077 VG004078 VG004079 VG004080 VG004081 VG004082 VG004083 VG004084 VG004085 VG004086 VG004087 VG004088 VG004089 VG004090 VG004091 VG004092 VG004093 VG004094 VG004095 VG004096 VG004098 VG004099 VG004100 VG004101 VG004102 VG004103 VG004104 VG004105 VG004107 VG004107 VG004108 VG004109 VG004110 VG004111 VG004112 VG004115 VG004116 VG004118 VG004119 VG004120 VG004121 VG004122 VG004123 VG004124 VG004125 VG004126 VG004128 VG004131 VG004133 VG004134 VG004135 VG004137 VG004138 VG004139 VG004140 VG004141 VG004143 VG004144 VG004146 VG004147 VG004148 VG004149 VG004150 VG004152 VG004153 VG004154 VG004155 VG004156 VG004157 VG004158 VG004159 VG004160 VG004162 VG004163 VG004164 VG004165 VG004167 VG004168 VG004169 VG004170 VG004172 VG004173 VG004175 VG004176 VG004177 VG004178 VG004179 VG004180 VG004181 VG004182 VG004183 VG004184 VG004185 VG004186 VG004187 VG004188 VG004189 VG004190 VG004191 VG004192 VG004194 VG004195 VG004196 VG004197 VG004198 VG004201 VG004203 VG004204 VG004205 VG004207 VG004208 VG004209 VG004210 VG004211 VG004212 VG004214 VG004217 VG004218 VG004219 VG004220 VG004223 VG004224 VG004225 VG004226 VG004227 VG004228 VG004229 VG004230 VG004232 VG004233 VG004234 VG004235 VG004236 VG004237 VG004238 VG004240 VG004241 VG004242 VG004243 VG004244 VG004245 VG004248 VG004249 VG004251 VG004252 VG004253 VG004254 VG004255 VG004256 VG004257 VG004258 VG004259 VG004260 VG004261 VG004262 VG004263 VG004267 VG004268 VG004275 VG004277 VG004278 VG004280 VG004284 VG004286 VG004290 VG004292 VG004907 |
| FEI Number |
2015691
|
Recalling Firm/ Manufacturer |
Edwards Lifesciences, Llc
1 Edwards Way
Irvine
CA
92614-5688
|
| For Additional Information Contact | Carla Sivak, RN 949-250-2024 |
Manufacturer Reason for Recall | MAP offset errors: If users are "slaving" mean arterial pressure (MAP) from the bedside monitor into the Vigilance II Monitor for the purpose of the continuous calculation of systemic vascular resistance (SVR), at times, on some bedside monitors, the Vigilance II monitor may display a MAP offset of as much as 25mmHg. Since the MAP is used in calculations of various patient parameters such as SVR, |
FDA Determined Cause 2 | Software design |
| Action | An Urgent Customer notification letter dated May 17, 2007 was mailed to all Vigilance II customers advising that If they are "slaving" mean arterial pressure (MAP) from their bedside monitor into the Vigilance II Monitor for the purpose of the continuous calculation of systemic vascular resistance (SVR), to be aware that at times, on some bedside monitors, the Vigilance II monitor may display a MAP offset of as much as 25 mmHg. Since the MAP is used in calculations of various patient parameters such as SVR, these may also not be accurate. A table is provided showing those bedside monitors with the problem: GE Solar, Phillips (HP) M1006A, M1006B, as well as, those bedside monitors verified to not have the problem: SpaceLabs Model 90402 and Nihon Kohden Model BSM 8000. The letter also serves to remind the customer of the Warning in the Vigilance II Directions for Use which is also provided in the letter that instructs user to verify slaved MAP values with those displayed on bedside monitor. A temporary work around for this offset problem is to manually enter the MAP based on the readings from the bedside monitor. All Vigilance II calculations from the MAP have been re-verified to be correct when the MAP is accurate. The problem is in the transfer of the MAP between the bedside monitor and the Vigilance II. Customers are also advised that if they have the GE Solar or the Phillips (HP) models listed in the letter or any monitor not listed as verified to be compatible with the Vigilance II (Space labs and Nihon Kohden models listed above) Edwards recommends verification of MAP accuracy prior to treating based on parameters from a Vigilance II in a "slave" configuration. If there is a difference, they should enter the MAP into the Vigilance II manually for a static calculation of SVR. Edwards will be supplying a software update to correct this problem. |
| Quantity in Commerce | 2,680 units in North America distribution |
| Distribution | Worldwide: USA and Canada |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|