| | Class 2 Device Recall CT/e and HiSpeed X/i computed Tomography systems |  |
| Date Initiated by Firm | December 01, 2008 |
| Date Posted | September 18, 2009 |
| Recall Status1 |
Terminated 3 on January 27, 2016 |
| Recall Number | Z-2224-2009 |
| Recall Event ID |
50528 |
| 510(K)Number | K021491 K944013 K980169 K991716 K993645 |
| Product Classification |
System, x-ray, tomography, computed - Product Code JAK
|
| Product | GE Healthcare CT/e and HiSpeed X/i, families of computed tomography Systems with Gantry.
The CT/e and HiSpeed X/i families CT Scanner Systems are indicated for head and whole body X-ray Computed Tomography applications. |
| Code Information |
MODEL NUMBER SERIAL NUMBER 2200997 00000474838YM0 2200997 00000474842YM2 2200997 00000474844YM8 2200997 00000474845YM5 2200997 00000474846YM3 2200997 00000474847YM1 2200997 00000474848YM9 2200997 00000474849YM7 2200997 00000474850YM5 2200997 00000474851YM3 2200997 00000474852YM1 2200997 00000474854YM7 2200997 00000474855YM4 2200997 00000474856YM2 2200997 00000474857YM0 2200997 00000474858YM8 2200997 00000474860YM4 2200997 00000474861YM2 2200997 00000474862YM0 2200997 00000474864YM6 2200997 00000474865YM3 2200997 00000474867YM9 2200997 00000474870YM3 2200997 00000474871YM1 2200997 00000474872YM9 2200997 00000474873YM7 2200997 00000474874YM5 2200997 00000474879YM4 2200997 00000474883YM6 2200997 00000474887YM7 2200997 00000474893YM5 2200997 00000474895YM0 2200997 00000474897YM6 2200997 00000474898YM4 2200997 00000474900YM8 2200997 00000474901YM6 2200997 00000474903YM2 2200997 00000474907YM3 2200997 00000474910YM7 2200997 00000474911YM5 2200997 00000474912YM3 2200997 00000474913YM1 2200997 00000474914YM9 2200997 00000474916YM4 2200997 00000474917YM2 2200997 00000474918YM0 2200997 00000474922YM2 2200997 00000474924YM8 2200997 00000474925YM5 2200997 00000474926YM3 2200997 00000474928YM9 2200997 00000474929YM7 2200997 00000474931YM3 2200997 00000474932YM1 2200997 00000474934YM7 2200997 00000522400YM1 2200997 00000522402YM7 2200997 00000522403YM5 2200997 00000522405YM0 2200997 00000522407YM6 2200997 00000522408YM4 2200997 00000522409YM2 2200997 00000522415YM9 2200997 00000522416YM7 2200997 00000522417YM5 2200997 00000522418YM3 2200997 00000522419YM1 2200997 00000522420YM9 2200997 00000522425YM8 2200997 00000522426YM6 2200997 00000522428YM2 2200997 00000522429YM0 2200997 00000522430YM8 2200997 00000522431YM6 2200997 00000522432YM4 2200997 00000522433YM2 2200997 00000522436YM5 2200997 00000522437YM3 2200997 00000522438YM1 2200997 00000522439YM9 2200997 00000522440YM7 2200997 00000522441YM5 2200997 00000522442YM3 2200997 00000522443YM1 2200997 00000522446YM4 2200997 00000522447YM2 2200997 00000522448YM0 2200997 00000522449YM8 2200997 00000522450YM6 2200997 00000522451YM4 2200997 00000522452YM2 2200997 00000522453YM0 2200997 00000522455YM5 2200997 00000522457YM1 2200997 00000522459YM7 2200997 00000522460YM5 2200997 00000522462YM1 2200997 00000522465YM4 2200997 00000522467YM0 2200997 00000522469YM6 2200997 00000522471YM2 2200997 00000522472YM0 2200997 00000522473YM8 2200997 00000522477YM9 2200997 00000522480YM3 2200997 00000522481YM1 2200997 00000522482YM9 2200997 00000522483YM7 2200997 00000522484YM5 2200997 00000522487YM8 2200997 00000522488YM6 2200997 00000522489YM4 2200997 00000522490YM2 2200997 00000522493YM6 2200997 00000522494YM4 2200997 00000522495YM1 2200997 00000522497YM7 2200997 00000522498YM5 2200997 00000544700YM8 2200997 00000544701YM6 2200997 00000544702YM4 2200997 00000544703YM2 2200997 00000544704YM0 2200997 00000544706YM5 2200997 00000544707YM3 2200997 00000544710YM7 2200997 00000544711YM5 2200997 00000544712YM3 2200997 00000544714YM9 2200997 00000544715YM6 2200997 00000544717YM2 2200997 00000544720YM6 2200997 00000544721YM4 2200997 00000544723YM0 2200997 00000544725YM5 2200997 00000544726YM3 2200997 00000544727YM1 2200997 00000544728YM9 2200997 00000544730YM5 2200997 00000544731YM3 2200997 00000544732YM1 2200997 00000544733YM9 2200997 00000544735YM4 2200997 00000544736YM2 2200997 00000544737YM0 2200997 00000544738YM8 2200997 00000544739YM6 2200997 00000544741YM2 2200997 00000544743YM8 2200997 00000544745YM3 2200997 00000544747YM9 2200997 00000544749YM5 2200997 00000544750YM3 2200997 00000544751YM1 2200997 00000544753YM7 2200997 00000544754YM5 2200997 00000544755YM2 2200997 00000544756YM0 2200997 00000544758YM6 2200997 00000544759YM4 2200997 00000544762YM8 2200997 00000544763YM6 2200997 00000544764YM4 2200997 00000544765YM1 2200997 00000544768YM5 2200997 00000544769YM3 2200997 00000544770YM1 2200997 00000544772YM7 2200997 00000544773YM5 2200997 00000544774YM3 2200997 00000544776YM8 2200997 00000544777YM6 2200997 00000544778YM4 2200997 00000544779YM2 2200997 00000544780YM0 2200997 00000544781YM8 2200997 00000544782YM6 2200997 00000544785YM9 2200997 00000544788YM3 2200997 00000544791YM7 2200997 00000544793YM3 2200997 00000544794YM1 2200997 00000544795YM8 2200997 00000544796YM6 2200997 00000544797YM4 2200997 00000544798YM2 2200997 00000544799YM0 2200997 00000601401YM3 2200997 00000601402YM1 2200997 00000601403YM9 2200997 00000601406YM2 2200997 00000601408YM8 2200997 00000601414YM6 2200997 00000601415YM3 2200997 00000601416YM1 2200997 00000601417YM9 2200997 00000601418YM7 2200997 00000601422YM9 2200997 00000601424YM5 2200997 00000601425YM2 2200997 00000601426YM0 2200997 00000601427YM8 2200997 00000601428YM6 2200997 00000601429YM4 2200997 00000601430YM2 2200997 00000601432YM8 2200997 00000601436YM9 2200997 00000601437YM7 2200997 00000601438YM5 2200997 00000601440YM1 2200997 00000601444YM3 2200997 00000601445YM0 2200997 00000601447YM6 2200997 00000601449YM2 2200997 00000601450YM0 2200997 00000601451YM8 2200997 00000601452YM6 2200997 00000601453YM4 2200997 00000601455YM9 2200997 00000601456YM7 2200997 00000601457YM5 2200997 00000601460YM9 2200997 00000601461YM7 2200997 00000601462YM5 2200997 00000601464YM1 2200997 00000601465YM8 2200997 00000601467YM4 2200997 00000601468YM2 2200997 00000601469YM0 2200997 00000601470YM8 2200997 00000601471YM6 2200997 00000601474YM0 2200997 00000601475YM7 2200997 00000601477YM3 2200997 00000601478YM1 2200997 00000601479YM9 2200997 00000601480YM7 2200997 00000601481YM5 2200997 00000601483YM1 2200997 00000601484YM9 2200997 00000601485YM6 2200997 00000601486YM4 2200997 00000601488YM0 2200997 00000601489YM8 2200997 00000601490YM6 2200997 00000714169YM7 2200997 00000714718YM4 2200997 00000714719YM2 2200997 00000714720YM0 2200997 00000714721YM8 2200997 00000714724YM2 2200997 00000714726YM7 2200997 00000714727YM5 2200997 00000714728YM3 2200997 00000714729YM1 2200997 00000714730YM9 2200997 00000714731YM7 2200997 00000714732YM5 2200997 00000714734YM1 2200997 00000714735YM8 2200997 00000714736YM6 2200997 00000714737YM4 2200997 00000714739YM0 2200997 00000714740YM8 2200997 00000714741YM6 2200997 00000714742YM4 2200997 00000714743YM2 2200997 00000714745YM7 2200997 00000714748YM1 2200997 00000714749YM9 2200997 00000714750YM7 2200997 00000714751YM5 2200997 00000714752YM3 2200997 00000714754YM9 2200997 00000714755YM6 2200997 00000714757YM2 2200997 00000714758YM0 2200997 00000714759YM8 2200997 00000714760YM6 2200997 00000714764YM8 2200997 00000714765YM5 2200997 00000714766YM3 2200997 00000714767YM1 2200997 00000714769YM7 2200997 00000714770YM5 2200997 00000714771YM3 2200997 00000714772YM1 2200997 00000714773YM9 2200997 00000714774YM7 2200999 00000474935YM4 2200999 00000474937YM0 2200999 00000474938YM8 2200999 00000474940YM4 2200999 00000474941YM2 2200999 00000474945YM3 2200999 00000474946YM1 2200999 00000474950YM3 2200999 00000474951YM1 2200999 00000714919YM8 2200999 00000714920YM6 2200999 00000714922YM2 2200999 00000714925YM5 2200999 00000714926YM3 2200999 00000884920YM4 2201000 00000000015313 2201000 00000006627HW4 2201000 00000006677HW9 2201000 00000033922HM6 2201000 00000475036YM0 2201000 00000475037YM8 2201000 00000475038YM6 2201000 00000475039YM4 2201000 00000475040YM2 2201000 00000475043YM6 2201000 00000475044YM4 2201000 00000475045YM1 2201000 00000475046YM9 2201000 00000475049YM3 2201000 00000475051YM9 2201000 00000475052YM7 2201000 00000475053YM5 2201000 00000475054YM3 2201000 00000475058YM4 2201000 00000475064YM2 2201000 00000475065YM9 2201000 00000475066YM7 2201000 00000475068YM3 2201000 00000475069YM1 2201000 00000475073YM3 2201000 00000475076YM6 2201000 00000475079YM0 2201000 00000475080YM8 2201000 00000475081YM6 2201000 00000475084YM0 2201000 00000475088YM1 2201000 00000475089YM9 2201000 00000475092YM3 2201000 00000475093YM1 2201000 00000475095YM6 2201000 00000475096YM4 2201000 00000475099YM8 2201000 00000475101YM2 2201000 00000475102YM0 2201000 00000475103YM8 2201000 00000475104YM6 2201000 00000475105YM3 2201000 00000475107YM9 2201000 00000475111YM1 2201000 00000475112YM9 2201000 00000475113YM7 2201000 00000475115YM2 2201000 00000475119YM4 2201000 00000475120YM2 2201000 00000475121YM0 2201000 00000475122YM8 2201000 00000475123YM6 2201000 00000475124YM4 2201000 00000475125YM1 2201000 00000475126YM9 2201000 00000475128YM5 2201000 00000475129YM3 2201000 00000475130YM1 2201000 00000475132YM7 2201000 00000475133YM5 2201000 00000475134YM3 2201000 00000532902YM4 2201000 00000532903YM2 2201000 00000532905YM7 2201000 00000532906YM5 2201000 00000532909YM9 2201000 00000532910YM7 2201000 00000532911YM5 2201000 00000532913YM1 2201000 00000532914YM9 2201000 00000532915YM6 2201000 00000532916YM4 2201000 00000532917YM2 2201000 00000532918YM0 2201000 00000532919YM8 2201000 00000532920YM6 2201000 00000532921YM4 2201000 00000532924YM8 2201000 00000532925YM5 2201000 00000532926YM3 2201000 00000532928YM9 2201000 00000532929YM7 2201000 00000532931YM3 2201000 00000532932YM1 2201000 00000532933YM9 2201000 00000532936YM2 2201000 00000532938YM8 2201000 00000532940YM4 2201000 00000532941YM2 2201000 00000532942YM0 2201000 00000532943YM8 2201000 00000532944YM6 2201000 00000532947YM9 2201000 00000532949YM5 2201000 00000532950YM3 2201000 00000532951YM1 2201000 00000532953YM7 2201000 00000532954YM5 2201000 00000532955YM2 2201000 00000532956YM0 2201000 00000532957YM8 2201000 00000532958YM6 2201000 00000532960YM2 2201000 00000532962YM8 2201000 00000532964YM4 2201000 00000532965YM1 2201000 00000532966YM9 2201000 00000532967YM7 2201000 00000532970YM1 2201000 00000532971YM9 2201000 00000532972YM7 2201000 00000532974YM3 2201000 00000532975YM0 2201000 00000532976YM8 2201000 00000532977YM6 2201000 00000532978YM4 2201000 00000532979YM2 2201000 00000532980YM0 2201000 00000532981YM8 2201000 00000532982YM6 2201000 00000532984YM2 2201000 00000532985YM9 2201000 00000532986YM7 2201000 00000532987YM5 2201000 00000532988YM3 2201000 00000532989YM1 2201000 00000532990YM9 2201000 00000532992YM5 2201000 00000532994YM1 2201000 00000532995YM8 2201000 00000532999YM0 2201000 00000535928YM9 2201000 00000583700YM0 2201000 00000583701YM8 2201000 00000583703YM4 2201000 00000583705YM9 2201000 00000583706YM7 2201000 00000583708YM3 2201000 00000583709YM1 2201000 00000583710YM9 2201000 00000583711YM7 2201000 00000583715YM8 2201000 00000583716YM6 2201000 00000583716YN6 2201000 00000583717YM4 2201000 00000583718YM2 2201000 00000583720YM8 2201000 00000583721YM6 2201000 00000583722YM4 2201000 00000583724YM0 2201000 00000583725YM7 2201000 00000583726YM5 2201000 00000583727YM3 2201000 00000583728YM1 2201000 00000583729YM9 2201000 00000583730YM7 2201000 00000583732YM3 2201000 00000583735YM6 2201000 00000583736YM4 2201000 00000583738YM0 2201000 00000583739YM8 2201000 00000583740YM6 2201000 00000583741YM4 2201000 00000583743YM0 2201000 00000583744YM8 2201000 00000583745YM5 2201000 00000583746YM3 2201000 00000583747YM1 2201000 00000583750YM5 2201000 00000583751YM3 2201000 00000583753YM9 2201000 00000583754YM7 2201000 00000583755YM4 2201000 00000583756YM2 2201000 00000583757YM0 2201000 00000583759YM6 2201000 00000583760YM4 2201000 00000583761YM2 2201000 00000583762YM0 2201000 00000583763YM8 2201000 00000583764YM6 2201000 00000583765YM3 2201000 00000583766YM1 2201000 00000583767YM9 2201000 00000583770YM3 2201000 00000583771YM1 2201000 00000583772YM9 2201000 00000583773YM7 2201000 00000583774YM5 2201000 00000583775YM2 2201000 00000583777YM8 2201000 00000583778YM6 2201000 00000583781YM0 2201000 00000583782YM8 2201000 00000583785YM1 2201000 00000583786YM9 2201000 00000583787YM7 2201000 00000583788YM5 2201000 00000583789YM3 2201000 00000583791YM9 2201000 00000583792YM7 2201000 00000583793YM5 2201000 00000583794YM3 2201000 00000583795YM0 2201000 00000583796YM8 2201000 00000583797YM6 2201000 00000583798YM4 2201000 00000583799YM2 2201000 00000661253YM5 2201000 00000661254YM3 2201000 00000661255YM0 2201000 00000661256YM8 2201000 00000661258YM4 2201000 00000661259YM2 2201000 00000661260YM0 2201000 00000661263YM4 2201000 00000661264YM2 2201000 00000661265YM9 2201000 00000661266YM7 2201000 00000661267YM5 2201000 00000661268YM3 2201000 00000661270YM9 2201000 00000661271YM7 2201000 00000661273YM3 2201000 00000661274YM1 2201000 00000661275YM8 2201000 00000661276YM6 2201000 00000661277YM4 2201000 00000661278YM2 2201000 00000661279YM0 2201000 00000661280YM8 2201000 00000661281YM6 2201000 00000661282YM4 2201000 00000661284YM0 2201000 00000661286YM5 2201000 00000661287YM3 2201000 00000661288YM1 2201000 00000661289YM9 2201000 00000661291YM5 2201000 00000661292YM3 2201000 00000661294YM9 2201000 00000661297YM2 2201000 00000661298YM0 2201000 00000661300YM4 2201000 00000661301YM2 2201000 00000661302YM0 2201000 00000661304YM6 2201000 00000661306YM1 2201000 00000661308YM7 2201000 00000661309YM5 2201000 00000661311YM1 2201000 00000661314YM5 2201000 00000661315YM2 2201000 00000661316YM0 2201000 00000661317YM8 2201000 00000661318YM6 2201000 00000661319YM4 2201000 00000661322YM8 2201000 00000661323YM6 2201000 00000661324YM4 2201000 00000661327YM7 2201000 00000661328YM5 2201000 00000661329YM3 2201000 00000661330YM1 2201000 00000661331YM9 2201000 00000661332YM7 2201000 00000661333YM5 2201000 00000661334YM3 2201000 00000661335YM0 2201000 00000661336YM8 2201000 00000661337YM6 2201000 00000661339YM2 2201000 00000661340YM0 2201000 00000661341YM8 2201000 00000661342YM6 2201000 00000661345YM9 2201000 00000661346YM7 2201000 00000661348YM3 2201000 00000661349YM1 2201000 00000661350YM9 2201000 00000661351YM7 2201000 00000661352YM5 2201000 00000661353YM3 2201000 00000661354YM1 2201000 00000661355YM8 2201000 00000661356YM6 2201000 00000661358YM2 2201000 00000661359YM0 2201000 00000661361YM6 2201000 00000661362YM4 2201000 00000661363YM2 2201000 00000661364YM0 2201000 00000715118YM6 2201000 00000715121YM0 2201000 00000715125YM1 2201000 00000715126YM9 2201000 00000715129YM3 2201000 00000715130YM1 2201000 00000715131YM9 2201000 00000715133YM5 2201000 00000715134YM3 2201000 00000715135YM0 2201000 00000715136YM8 2201000 00000715137YM6 2201000 00000715139YM2 2201000 00000715141YM8 2201000 00000715144YM2 2201000 00000715146YM7 2201000 00000715147YM5 2201000 00000715148YM3 2201000 00000715150YM9 2201000 00000715151YM7 2201000 00000715152YM5 2201000 00000715153YM3 2201000 00000715155YM8 2201000 00000715156YM6 2201000 00000715157YM4 2201000 00000715158YM2 2201000 00000715159YM0 2201000 00000715160YM8 2201000 00000715161YM6 2201000 00000715163YM2 2201000 00000715164YM0 2201000 00000715166YM5 2201000 00000715167YM3 2201000 00000715168YM1 2201000 00000715169YM9 2201000 00000715170YM7 2201000 00000715172YM3 2201000 00000715173YM1 2201000 00000715176YM4 2201000 00000715177YM2 2201000 00000715181YM4 2201000 00000715182YM2 2201000 00000715184YM8 2201000 00000715185YM5 2201000 00000715186YM3 2201000 00000715187YM1 2201000 00000715190YM5 2201000 00000715191YM3 2201000 00000715192YM1 2201000 00000715193YM9 2201000 00000715195YM4 2201000 UNK22010000000 2244227 00000007237HW1 2244227 00000007310HW6 2244227 00000007318HW9 2244227 00000007326HW2 2244227 00000007388HM2 2244227 00000007395HM7 2244227 00000007402HM1 2244227 00000007413HM8 2244227 00000007414HM6 2244227 00000007455HM9 2244227 00000007456HM7 2244227 00000007520HM0 2244227 00000007529HM1 2244227 00000007530HM9 2244227 00000007589HM5 2244227 00000007590HM3 2244227 00000007614HM1 2244227 00000007615HM8 2244227 00000007676HM0 2244227 00000007712HM3 2244227 00000007710HM7 2244227 00000007711HM5 2244227 00000007713HM1 2244227 00000007714HM9 2244227 00000007715HM6 2244227 00000007819HM6 2244227 00000007820HM4 2244227 00000007839HM4 2244227 00000007840HM2 2244227 00000007900HM4 2244227 00000007901HM2 2244227 00000007967HM3 2244227 00000007968HM1 2244227 00000007969HM9 2244227 00000008006HM9 2244227 00000008007HM7 2244227 00000008008HM5 2244227 00000008009HM3 2244227 00000008074HM7 2244227 00000008075HM4 2244227 00000008077HM0 2244227 00000008124HM0 2244227 00000008125HM7 2244227 00000008154HM7 2244227 00000008155HM4 2244227 00000008183HM6 2244227 00000008184HM4 2244227 00000008185HM1 2244227 00000008244HM6 2244227 00000008245HM3 2244227 00000008282HM6 2244227 00000008283HM4 2244227 00000008284HM2 2244227 00000008285HM9 2244227 00000008460HM8 2244227 00000008461HM6 2244227 00000008490HM5 2244227 00000008491HM3 2244227 00000008512HM6 2244227 00000008513HM4 2244227 00000008514HM2 2244227 00000008515HM9 2244227 00000008641HM3 2244227 00000008642HM1 2244227 00000008757HM7 2244227 00000008758HM5 2244227 00000008760HM1 2244227 00000008795HM7 2244227 00000008796HM5 2244227 00000008972HM2 2244227 00000008973HM0 2244227 00000008974HM8 2244227 00000008975HM5 2244227 00000009028HM2 2244227 00000009029HM0 2244227 00000009030HM8 2244227 00000009031HM6 2244227 00000009118HM1 2244227 00000009168HM6 2244227 00000009169HM4 2244227 00000009170HM2 2244227 00000009236HM1 2244227 00000009236HM9 2244227 00000009314HM6 2244227 00000009315HM3 2244227 00000009238HM7 2244227 00000009409HM4 2244227 00000009437HM5 2244227 00000009438HM3 2244227 00000009439HM1 2244227 00000009440HM9 2244227 00000009520HM8 2244227 00000009521HM6 2244227 00000009522HM4 2244227 00000009557HM0 2244227 00000009558HM8 2244227 00000009615HM6 2244227 00000009617HM2 2244227 00000009618HM0 2244227 00000009653HM7 2244227 00000009654HM5 2244227 00000009675HM0 2244227 00000009676HM8 2244227 00000009677HM6 2244227 00000009678HM4 2244227 00000009749HM3 2244227 00000009860HM8 2244227 00000009862HM4 2244227 00000009863HM2 2244227 00000009928HM3 2244227 00000009929HM1 2244227 00000009948HM1 2244227 00000009949HM9 2244227 00000009968HM9 2244227 00000009969HM7 2244227 00000009985HM3 2244227 0000009995HM2 2244227 00000010058HM6 2244227 00000010059HM4 2244227 00000010089HM1 2244227 00000010090HM9 2244227 00000010091HM7 2244227 00000010129HM5 2244227 00000010130HM3 2244227 00000010155HM0 2244227 00000010171HM7 2244227 00000010172HM5 2244227 00000010173HM3 2244227 00000010204HM6 2244227 00000010205HM3 2244227 00000010219HM4 2244227 00000010218HM6 2244227 00000010356HM4 2244227 00000010357HM2 2244227 00000010358HM0 2244227 00000010394HM5 2244227 00000010395HM2 2244227 00000010396HM0 2244227 00000010397HM8 2244227 00000010435HM6 2244227 00000010436HM4 2244227 00000010437HM2 2244227 00000010523HM9 2244227 00000010524HM7 2244227 00000010564HM3 2244227 00000010566HM8 2244227 00000010628HM6 2244227 00000010721HM9 2244227 00000010722HM7 2244227 00000010723HM5 2244227 00000010724HM3 2244227 00000010761HM5 2244227 00000010763HM1 2244227 00000010764HM9 2244227 00000010802HM7 2244227 00000010803HM5 2244227 00000010831HM6 2244227 00000011106HM2 2244227 00000011107HM 2244227 00000011108HM8 2244227 00000011109HM6 2244227 00000011144HM3 2244227 00000011145HM0 2244227 00000011167HM4 2244227 00000011168HM2 2244227 00000011169HM0 2244227 00000011231HM8 2244227 00000011232HM6 2244227 00000011251HM6 2244227 00000011252HM4 2244227 00000011286HM2 2244227 00000011288HM8 2244227 00000011289HM6 2244227 00000011411HM6 2244227 00000011412HM4 2244227 00000011432HM2 2244227 00000011450HM4 2244227 00000011502HM2 2244227 00000011503HM0 2244227 00000011504HM8 2244227 00000011505HM5 2244227 00000011548HM5 2244227 00000011558HM4 2244227 00000011697HM0 2244227 00000011696HM2 2244227 00000011716HM8 2244227 00000011717HM6 2244227 00000011740HM8 2244227 00000011741HM6 2244227 00000011743HM2 2244227 00000011744HM0 2244227 00000011745HM7 2244227 00000011876HM0 2244227 00000011942HM0 2244227 00000011943HM8 2244227 00000011962HM8 2244227 00000011963HM6 2244227 00000011985HM9 2244227 00000011986HM7 2244227 00000011987HM5 2244227 00000011988HM3 2244227 00000011989HM1 2244227 00000012116HM0 2244227 00000012117HM8 2244227 00000012118HM6 2244227 00000012119HM4 2244227 00000012154HM1 2244227 00000012155HM8 2244227 00000012176HM4 2244227 00000012177HM2 2244227 00000011944HM6 2244227 00000011945HM3 2244227 00000012260HM6 2244227 00000012261HM4 2244227 00000012278HM8 2244227 00000012279HM7 2244227 00000012406HM5 2244227 00000012407HM3 2244227 00000012499HM0 2244227 00000012518HM7 2244227 00000012519HM5 2244227 00000012555HM9 2244227 00000012660HM7 2244227 00000012673HM0 2244227 00000012674HM8 2244227 00000012803HM3 2244227 00000012844HM7 2244227 00000012845HM4 2244227 00000012976HM7 2244227 00000012977HM5 2244227 00000012994HM0 2244227 00000013017HM9 2244227 00000013018HM7 2244227 00000013019HM5 2244227 00000013081HM5 2244227 00000013189HM6 2244227 00000013199HM5 2244227 00000013205HM0 2244227 00000013264HM7 2244227 00000013355HM3 2244227 00000013366HM0 2244227 00000013367HM8 2244227 00000013391HM8 2244227 00000013418HM9 2244227 00000013419HM7 2244227 00000013441HM1 2244227 00000013442HM9 2244227 00000013443HM7 2244227 00000013476HM7 2244227 00000013477HM5 2244227 00000013560HM8 2244227 00000013356HM6 2244227 00000013562HM4 2244227 00000013590HM5 2244227 00000013608HM5 2244227 00000013618HM4 2244227 00000013628HM3 2244227 00000013639HM0 2244227 00000013649HM9 2244227 00000013659HM8 2244227 00000013669HM7 2244227 00000013679HM6 2244227 00000013689HM5 2244227 00000013699HM4 2244227 00000013709HM1 2244227 00000013719HM0 2244227 00000013761HM2 2244227 00000013771HM1 2244227 00000013781HM0 2244227 00000013794HM3 2244227 00000013802HM4 2244227 00000013860HM2 2244227 00000013870HM1 2244227 00000013880HM0 2244227 00000014073HM1 2244227 00000014120HM0 2244227 00000014121HM8 2244227 00000014122HM6 2244227 00000014123HM4 2244227 00000014164HM8 2244227 00000014169HM7 2244227 00000014170HM5 2244227 00000014171HM3 2244227 00000013956HM8 2244227 00000014231HM5 2244227 00000014243HM0 2244227 00000014261HM2 2244227 00000014309HM9 2244227 00000014310HM7 2244227 00000014328HM9 2244227 00000014338HM8 2244227 00000014348HM7 2244227 00000014456HM8 2244227 00000014468HM3 2244227 00000014469HM1 2244227 00000014470HM9 2244227 00000014548HM2 2244227 00000014552HM4 2244227 00000014566HM4 2244227 00000014567HM2 2244227 00000014568HM0 2244227 00000014570HM6 2244227 00000014616HM7 2244227 00000014617HM5 2244227 00000014618HM3 2244227 00000014619HM1 2244227 00000014620HM9 2244227 00000014663HM9 2244227 00000014664HM7 2244227 00000014682HM9 2244227 00000014693HM6 2244227 00000014694HM4 2244227 00000014821HM3 2244227 00000014822HM1 2244227 00000014823HM9 2244227 00000014824HM7 2244227 00000014825HM4 2244227 00000014826HM2 2244227 00000014827HM0 2244227 00000014828HM8 2244227 00000014969HM0 2244227 00000014983HM1 2244227 00000014984HM9 2244227 00000014985HM6 2244227 00000014986HM4 2244227 00000014987HM2 2244227 00000015190HM2 2244227 00000015191HM0 2244227 00000015192HM8 2244227 00000015193HM6 2244227 00000015259HM5 2244227 00000015295HM9 2244227 00000015306HM4 2244227 00000015325HM4 2244227 00000015344HM5 2244227 00000015354HM4 2244227 00000015364HM3 2244227 00000015517HM6 2244227 00000015518HM4 2244227 00000015519HM2 2244227 00000015545HM7 2244227 00000015577HM0 2244227 00000015668HM7 2244227 00000015668HM7 2244227 00000015669HM5 2244227 00000015687HM7 2244227 00000015701HM6 2244227 00000015702HM4 2244227 00000015697HM6 2244227 00000015729HM7 2244227 00000015730HM5 2244227 00000015731HM3 2244227 00000015787HM5 2244227 00000015799HM0 2244227 00000015307HM2 2244227 00000015819HM6 2244227 00000015920HM2 2244227 00000015921HM0 2244227 00000015922HM8 2244227 00000015958HM2 2244227 00000015959HM0 2244227 00000015980HM6 2244227 00000015981HM4 2244227 00000015982HM2 2244227 00000015983HM0 2244227 00000016019HM2 2244227 00000016020HM0 2244227 00000016021HM8 2244227 00000016060HM6 2244227 00000016209HM9 2244227 00000016210HM7 2244227 00000016211HM5 2244227 00000016237HM0 2244227 00000016311HM3 2244227 00000016324HM6 2244227 00000016325HM3 2244227 00000016326HM1 2244227 00000016327HM9 2244227 00000016361HM8 2244227 00000016387HM3 2244227 00000016389HM9 2244227 00000016390HM7 2244227 00000016391HM5 2244227 00000016392HM3 2244227 00000016393HM1 2244227 00000016507HM6 2244227 00000016529HM0 2244227 00000016530HM8 2244227 00000016549HM8 2244227 00000016550HM6 2244227 00000016569HM6 2244227 00000016570HM4 2244227 00000016591HM0 2244227 00000016592HM8 2244227 00000016593HM6 2244227 00000016648HM8 2244227 00000016629HM8 2244227 00000016630HM6 2244227 00000016594HM4 2244227 00000016706HM4 2244227 00000016728HM8 2244227 00000016729HM6 2244227 00000016730HM4 2244227 00000016731HM2 2244227 00000016839HM3 2244227 00000016840HM1 2244227 00000016841HM9 2244227 00000016842HM7 2244227 00000016876HM5 2244227 00000016890HM6 2244227 00000016891HM4 2244227 00000016892HM2 2244227 00000017035HM7 2244227 00000017036HM5 2244227 00000017054HM8 2244227 00000017064HM7 2244227 00000017075HM3 2244227 00000017076HM1 2244227 00000017095HM1 2244227 00000017096HM9 2244227 00000017114HM0 2244227 00000017125HM6 2244227 00000017131HM4 2244227 00000017132HM2 2244227 00000017156HM1 2244227 00000017153HM8 2244227 00000017509HM1 2244227 00000017566HM1 2244227 00000017567HM9 2244227 00000018274HM1 2244227 00000021802HM4 2244227 00000021960HM0 2244227 00000025744HM4 2244227 00000030447HM7 2244227 00000030891HM6 2244227 00000030893HM2 2244227 00000033517HM4 2244227 00000034107HM3 2244227 00000034889HM6 2244227 00000039887HM5 2244227 00000040696HM7 2244227 00000041770HM9 2244227 00000042514HM0 2244227 00000045488HM4 2244227 00000047811HM5 2244227 00000052773HM9 2244227 00000055682HM9 2244227 00000057477HM2 2244227 00000059358HM2 2244227 00000060230HM0 2244227 00000062682HM0 2244227 00000064036HM7 2244227 00000070508HM7 2244227 00000021560HM8 2244227 00000710822YM8 2244227 00000713319YM2 2244227 00000713320YM0 2244227 00000713321YM8 2244227 00000713322YM6 2244227 00000713323YM4 2244227 00000713324YM2 2244227 00000713325YM9 2244227 00000713326YM7 2244227 00000713327YM5 2244227 00000713328YM3 2244227 00000713329YM1 2244227 00000713330YM9 2244227 00000713331YM7 2244227 00000713332YM5 2244227 00000713333YM3 2244227 00000713334YM1 2244227 00000713335YM8 2244227 00000713337YM4 2244227 00000713338YM2 2244227 00000713339YM0 2244227 00000713340YM8 2244227 00000713341YM6 2244227 00000713342YM4 2244227 00000713343YM2 2244227 00000713344YM0 2244227 00000713345YM7 2244227 00000713346YM5 2244227 00000713347YM3 2244227 00000745528YM0 2244227 00000745529YM8 2244227 00000745530YM6 2244227 00000745531YM4 2244227 00000745532YM2 2244227 00000745533YM0 2244227 00000745534YM8 2244227 00000023260HM3 2244227 00000007294HW2 2244227 00000002552HM1 2244227 00000008076HM2 2244227 00000008759HM3 2244227 00000009316HM1 2244227 00000009616HM4 2244227 00000010156HM8 2244227 00000010359HM8 2244227 00000010563HM5 2244227 00000010762HM3 2244227 00000012155HM8 2244227 00000012554HM2 2244227 00000013016HM1 2244227 00000013091HM4 2244227 00000015557HM2 2244227 00000016371HM7 2244227 00000016388HM1 2244227 00000016518HM3 2244227 00000016569HM6 2244227 00000017154HM6 2244227 00000017155HM3 2244227 00000017190HM0 2244227 00000017200HM7 2244227 00000017261HM9 2244227 00000017262HM7 2244227 00000017351HM8 2244227 00000017352HM6 2244227 00000017467HM2 2244227 00000017468HM0 2244227 00000017469HM8 2244227 00000017470HM6 2244227 00000017507HM5 2244227 00000017508HM3 2244227 00000017537HM2 2244227 00000017539HM8 2244227 00000017688HM3 2244227 00000017689HM1 2244227 00000017707HM1 2244227 00000017708HM9 2244227 00000017709HM7 2244227 00000017733HM7 2244227 00000017734HM5 2244227 00000017752HM7 2244227 00000017762HM6 2244227 00000017772HM5 2244227 00000017783HM2 2244227 00000017784HM0 2244227 00000017805HM3 2244227 00000017825HM1 2244227 00000017826HM9 2244227 00000017844HM2 2244227 00000017855HM8 2244227 00000017856HM6 2244227 00000017874HM9 2244227 00000017885HM5 2244227 00000017886HM3 2244227 00000017904HM4 2244227 00000018093HM5 2244227 00000018094HM3 2244227 00000018095HM0 2244227 00000018122HM2 2244227 00000018143HM8 2244227 00000018161HM0 2244227 00000018171HM9 2244227 00000018182HM6 2244227 00000018247HM7 2244227 00000018248HM5 2244227 00000018307HM9 2244227 00000018308HM7 2244227 00000018309HM5 2244227 00000018310HM3 2244227 00000018311HM1 2244227 00000018312HM9 2244227 00000018364HM0 2244227 00000018365HM7 2244227 00000018366HM5 2244227 00000018393HM9 2244227 00000018394HM7 2244227 00000018412HM7 2244227 00000018592HM6 2244227 00000018613HM0 2244227 00000018614HM8 2244227 00000018615HM5 2244227 00000018616HM3 2244227 00000018651HM0 2244227 00000018652HM8 2244227 00000018700HM5 2244227 00000018711HM2 2244227 00000018712HM0 2244227 00000018747HM6 2244227 00000018748HM4 2244227 00000018770HM8 2244227 00000018771HM6 2244227 00000018772HM4 2244227 00000018773HM2 2244227 00000018774HM0 2244227 00000018816HM9 2244227 00000018833HM4 2244227 00000018834HM2 2244227 00000018837HM5 2244227 00000018838HM3 2244227 00000018896HM1 2244227 00000018906HM8 2244227 00000018916HM7 2244227 00000018926HM6 2244227 00000019077HM7 2244227 00000019078HM5 2244227 00000019079HM3 2244227 00000019109HM8 2244227 00000019110HM6 2244227 00000019111HM4 2244227 00000019112HM2 2244227 00000019113HM0 2244227 00000019155HM1 2244227 00000019176HM7 2244227 00000019177HM5 2244227 00000019198HM1 2244227 00000019199HM9 2244227 00000019257HM5 2244227 00000019258HM3 2244227 00000019259HM1 2244227 00000019260HM9 2244227 00000019261HM7 2244227 00000019312HM8 2244227 00000019313HM6 2244227 00000019331HM8 2244227 00000019434HM0 2244227 00000019435HM7 2244227 00000019436HM5 2244227 00000019463HM9 2244227 00000019464HM7 2244227 00000019582HM6 2244227 00000019635HM2 2244227 00000019636HM0 2244227 00000019637HM8 2244227 00000019638HM6 2244227 00000019639HM4 2244227 00000019640HM2 2244227 00000019694HM9 2244227 00000019695HM6 2244227 00000019696HM4 2244227 00000019697HM2 2244227 00000019698HM0 2244227 00000019740HM0 2244227 00000019887HM9 2244227 00000019888HM7 2244227 00000019889HM5 2244227 00000019890HM3 2244227 00000019891HM1 2244227 00000019934HM9 2244227 00000019935HM6 2244227 00000019975HM2 2244227 00000019976HM0 2244227 00000019977HM8 2244227 00000020003HM0 2244227 00000020052HM7 2244227 00000020053HM5 2244227 00000020054HM3 2244227 00000020055HM0 2244227 00000020098HM0 2244227 00000020099HM8 2244227 00000020119HM4 2244227 00000020120HM2 2244227 00000020121HM0 2244227 00000020350HM5 2244227 00000020351HM3 2244227 00000020368HM7 2244227 00000020378HM6 2244227 00000020444HM6 2244227 00000020456HM0 2244227 00000020457HM8 2244227 00000020458HM6 2244227 00000020679HM7 2244227 00000020680HM5 2244227 00000020681HM3 2244227 00000020682HM1 2244227 00000020683HM9 2244227 00000020684HM7 2244227 00000020742HM3 2244227 00000020752HM2 2244227 00000020762HM1 2244227 00000020772HM0 2244227 00000020833HM0 2244227 00000020843HM9 2244227 00000020965HM0 2244227 00000020967HM6 2244227 00000021000HM5 2244227 00000021001HM3 2244227 00000021020HM3 2244227 00000021021HM1 2244227 00000021040HM1 2244227 00000021041HM9 2244227 00000021061HM7 2244227 00000021062HM5 2244227 00000021063HM3 2244227 00000021399HM1 2244227 00000021437HM9 2244227 00000021456HM9 2244227 00000021457HM7 2244227 00000021458HM5 2244227 00000021459HM3 2244227 00000021461HM9 2244227 00000021462HM7 2244227 00000021463HM5 2244227 00000021464HM3 2244227 00000021465HM0 2244227 00000021547HM5 2244227 00000021559HM0 2244227 00000021587HM1 2244227 00000021599HM6 2244227 00000021600HM2 2244227 00000021601HM0 2244227 00000021630HM9 2244227 00000021631HM7 2244227 00000021632HM5 2244227 00000021633HM3 2244227 00000021873HM5 2244227 00000021883HM4 2244227 00000021895HM8 2244227 00000021896HM6 2244227 00000021897HM4 2244227 00000021925HM3 2244227 00000021926HM1 2244227 00000021927HM9 2244227 00000021957HM6 2244227 00000021958HM4 2244227 00000021959HM2 2244227 00000022086HM3 2244227 00000022087HM1 2244227 00000022088HM9 2244227 00000022089HM7 2244227 00000022135HM8 2244227 00000022136HM6 2244227 00000022137HM4 2244227 00000022138HM2 2244227 00000022139HM0 2244227 00000022185HM3 2244227 00000022186HM1 2244227 00000022187HM9 2244227 00000022188HM7 2244227 00000022189HM5 2244227 00000022312HM3 2244227 00000022313HM1 2244227 00000022339HM6 2244227 00000022340HM4 2244227 00000022341HM2 2244227 00000022342HM0 2244227 00000022343HM8 2244227 00000022344HM6 2244227 00000022346HM1 2244227 00000022347HM9 2244227 00000022423HM8 2244227 00000022442HM8 2244227 00000022443HM6 2244227 00000023135HM7 2244227 00000023141HM5 2244227 00000023147HM2 2244227 00000023152HM2 2244227 00000023176HM1 2244227 00000023182HM9 2244227 00000023188HM6 2244227 00000023194HM4 2244227 00000023240HM5 2244227 00000023250HM4 2244227 00000023532HM5 2244227 00000023545HM7 2244227 00000023546HM5 2244227 00000023547HM3 2244227 00000023584HM6 2244227 00000023585HM3 2244227 00000023586HM1 2244227 00000023613HM3 2244227 00000023614HM1 2244227 00000023633HM1 2244227 00000023634HM9 2244227 00000023950HM9 2244227 00000023951HM7 2244227 00000023952HM5 2244227 00000023978HM0 2244227 00000023988HM9 2244227 00000024015HM0 2244227 00000024016HM8 2244227 00000024017HM6 2244227 00000024018HM4 2244227 00000024019HM2 2244227 00000024020HM0 2244227 00000024021HM8 2244227 00000024022HM6 2244227 00000024027HM5 2244227 00000024028HM3 2244227 00000024029HM1 2244227 00000024030HM9 2244227 00000024110HM9 2244227 00000024111HM7 2244227 00000024112HM5 2244227 00000024113HM3 2244227 00000024300HM6 2244227 00000024302HM2 2244227 00000024485HM5 2244227 00000024495HM4 2244227 00000024507HM6 2244227 00000024508HM4 2244227 00000024589HM4 2244227 00000024920HM1 2244227 00000024921HM9 2244227 00000024922HM7 2244227 00000024923HM5 2244227 00000024990HM4 2244227 00000024991HM2 2244227 00000025013HM4 2244227 00000025014HM2 2244227 00000025015HM9 2244227 00000025016HM7 2244227 00000025017HM5 2244227 00000025061HM3 2244227 00000025063HM9 2244227 00000025550HM5 2244227 00000025551HM3 2244227 00000025552HM1 2244227 00000025585HM1 2244227 00000025586HM9 2244227 00000025588HM5 2244227 00000025589HM3 2244227 00000025590HM1 2244227 00000025591HM9 2244227 00000025592HM7 2244227 00000025711HM3 2244227 00000025723HM8 2244227 00000025743HM6 2244227 00000025763HM4 2244227 00000025764HM2 2244227 00000025807HM9 2244227 00000026498HM6 2244227 00000026499HM4 2244227 00000026500HM9 2244227 00000026501HM7 2244227 00000026561HM1 2244227 00000026562HM9 2244227 00000026563HM7 2244227 00000026564HM5 2244227 00000026956HM3 2244227 00000026957HM1 2244227 00000026978HM7 2244227 00000026979HM5 2244227 00000026980HM3 2244227 00000027019HM9 2244227 00000027020HM7 2244227 00000027021HM5 2244227 00000027022HM3 2244227 00000027023HM1 2244227 00000027462HM1 2244227 00000027463HM9 2244227 00000027481HM1 2244227 00000027493HM6 2244227 00000027494HM4 2244227 00000027495HM1 2244227 00000027523HM0 2244227 00000027524HM8 2244227 00000027525HM5 2244227 00000027554HM5 2244227 00000027556HM0 2244227 00000027934HM9 2244227 00000027935HM6 2244227 00000027955HM4 2244227 00000027956HM2 2244227 00000027957HM0 2244227 00000027987HM7 2244227 00000027988HM5 2244227 00000027989HM3 2244227 00000027990HM1 2244227 00000027991HM9 2244227 00000028380HM4 2244227 00000028381HM2 2244227 00000028382HM0 2244227 00000028383HM8 2244227 00000028384HM6 2244227 00000028385HM3 2244227 00000028386HM1 2244227 00000028387HM9 2244227 00000028456HM2 2244227 00000028457HM0 2244227 00000028458HM8 2244227 00000028459HM6 2244227 00000028805HM0 2244227 00000028806HM8 2244227 00000028888HM6 2244227 00000028889HM4 2244227 00000028890HM2 2244227 00000028927HM2 2244227 00000028928HM0 2244227 00000028954HM6 2244227 00000028955HM3 2244227 00000028956HM1 2244227 00000028957HM9 2244227 00000028958HM7 2244227 00000028959HM5 2244227 00000028960HM3 2244227 000000289J3HV8 2244227 00000029435HM5 2244227 00000029436HM3 2244227 00000029458HM7 2244227 00000029471HM0 2244227 00000029472HM8 2244227 00000029473HM6 2244227 00000029474HM4 2244227 00000029526HM1 2244227 00000029527HM9 2244227 00000029547HM7 2244227 00000029715HM0 2244227 00000029730HM9 2244227 00000029731HM7 2244227 00000029732HM5 2244227 00000029733HM3 2244227 00000029735HM8 2244227 00000029873HM7 2244227 00000030385HM9 2244227 00000030386HM7 2244227 00000030387HM5 2244227 00000030388HM3 2244227 00000030389HM1 2244227 00000030443HM6 2244227 00000030445HM1 2244227 00000030446HM9 2244227 00000030825HM4 2244227 00000030826HM2 2244227 00000030828HM8 2244227 00000030829HM6 2244227 00000030873HM4 2244227 00000030902HM1 2244227 00000030903HM9 2244227 00000030923HM7 2244227 00000030924HM5 2244227 00000030925HM2 2244227 00000030952HM6 2244227 00000030953HM4 2244227 00000030972HM4 2244227 00000030973HM2 2244227 00000031362HM7 2244227 00000031363HM5 2244227 00000031366HM8 2244227 00000031674HM5 2244227 00000031692HM7 2244227 00000031702HM4 2244227 00000031717HM2 2244227 00000031718HM0 2244227 00000031719HM8 2244227 00000031720HM6 2244227 00000032005HM1 2244227 00000032017HM6 2244227 00000032018HM4 2244227 00000032019HM2 2244227 00000032338HM6 2244227 00000032351HM9 2244227 00000032352HM7 2244227 00000032353HM5 2244227 00000032354HM3 2244227 00000032389HM9 2244227 00000032390HM7 2244227 00000032626HM4 2244227 00000032627HM2 2244227 00000032628HM0 2244227 00000032629HM8 2244227 00000032664HM5 2244227 00000032665HM2 2244227 00000032952HM4 2244227 00000032953HM2 2244227 00000032972HM2 2244227 00000032973HM0 2244227 00000033186HM8 2244227 00000033187HM6 2244227 00000033188HM4 2244227 00000033189HM2 2244227 00000033190HM0 2244227 00000033234HM6 2244227 00000033235HM3 2244227 00000033236HM1 2244227 00000033389HM8 2244227 00000033496HM1 2244227 00000033516HM6 2244227 00000033538HM0 2244227 00000033539HM8 2244227 00000033541HM4 2244227 00000034097HM6 2244227 00000034117HM2 2244227 00000034758HM3 2244227 00000034759HM1 2244227 00000034777HM3 2244227 00000034891HM2 2244227 00000034998HM5 2244227 00000035000HM9 2244227 00000035001HM7 2244227 00000035035HM5 2244227 00000035052HM0 2244227 00000035053HM8 2244227 00000035054HM6 2244227 00000035055HM3 2244227 00000035056HM1 2244227 00000035057HM9 2244227 00000035058HM7 2244227 00000035253HM4 2244227 00000035266HM6 2244227 00000035267HM4 2244227 00000035268HM2 2244227 00000035269HM0 2244227 00000035600HM6 2244227 00000035601HM4 2244227 00000035602HM2 2244227 00000035628HM7 2244227 00000035995HM0 2244227 00000036010HM7 2244227 00000036011HM5 2244227 00000036012HM3 2244227 00000036014HM9 2244227 00000036015HM6 2244227 00000036106HM3 2244227 00000036124HM6 2244227 00000036204HM6 2244227 00000036205HM3 2244227 00000036206HM1 2244227 00000036207HM9 2244227 00000036208HM7 2244227 00000036250HM9 2244227 00000036525HM4 2244227 00000036526HM2 2244227 00000036527HM0 2244227 00000036528HM8 2244227 00000036530HM4 2244227 00000036531HM2 2244227 00000036532HM0 2244227 00000036533HM8 2244227 00000036534HM6 2244227 00000036616HM1 2244227 00000036626HM0 2244227 00000036637HM7 2244227 00000036638HM5 2244227 00000037012HM2 2244227 00000037013HM0 2244227 00000037042HM9 2244227 00000037043HM7 2244227 00000037044HM5 2244227 00000037070HM0 2244227 00000037081HM7 2244227 00000037082HM5 2244227 00000037103HM9 2244227 00000037104HM7 2244227 00000037105HM4 2244227 00000037106HM2 2244227 00000037264HM9 2244227 00000037265HM6 2244227 00000037285HM4 2244227 00000037286HM2 2244227 00000037287HM0 2244227 00000037313HM4 2244227 00000037694HM7 2244227 00000037705HM1 2244227 00000037706HM9 2244227 00000037750HM7 2244227 00000037772HM1 2244227 00000037773HM9 2244227 00000037774HM7 2244227 00000037775HM4 2244227 00000037776HM2 2244227 00000037932HM1 2244227 00000037933HM9 2244227 00000037953HM7 2244227 00000037954HM5 2244227 00000037955HM2 2244227 00000038164HM0 2244227 00000038165HM7 2244227 00000038183HM0 2244227 00000039047HM6 2244227 00000039049HM2 2244227 00000039050HM0 2244227 00000039051HM8 2244227 00000039052HM6 2244227 00000039053HM4 2244227 00000039054HM2 2244227 00000039055HM9 2244227 00000039056HM7 2244227 00000039176HM3 2244227 00000039177HM1 2244227 00000039196HM1 2244227 00000039197HM9 2244227 00000039450HM2 2244227 00000039451HM0 2244227 00000039470HM0 2244227 00000039471HM8 2244227 00000039899HM0 2244227 00000039900HM6 2244227 00000039922HM0 2244227 00000039923HM8 2244227 00000039924HM6 2244227 00000039925HM3 2244227 00000039926HM1 2244227 00000040694HM2 2244227 00000040695HM9 2244227 00000040697HM5 2244227 00000040748HM6 2244227 00000040749HM4 2244227 00000040768HM4 2244227 00000040769HM2 2244227 00000041335HM1 2244227 00000041336HM9 2244227 00000041337HM7 2244227 00000041338HM5 2244227 00000041339HM3 2244227 00000041382HM3 2244227 00000041383HM1 2244227 00000041768HM3 2244227 00000041769HM1 2244227 00000041788HM1 2244227 00000041789HM9 2244227 00000041817HM8 2244227 00000041818HM6 2244227 00000041861HM6 2244227 00000041862HM4 2244227 00000042355HM8 2244227 00000042390HM5 2244227 00000042413HM5 2244227 00000042414HM3 2244227 00000042415HM0 2244227 00000042450HM7 2244227 00000042451HM5 2244227 00000042469HM7 2244227 00000042479HM6 2244227 00000042515HM7 2244227 00000042750HM0 2244227 00000042751HM8 2244227 00000042770HM8 2244227 00000042771HM6 2244227 00000042790HM6 2244227 00000042791HM4 2244227 00000043000HM9 2244227 00000043002HM5 2244227 00000043037HM1 2244227 00000043038HM9 2244227 00000043454HM8 2244227 00000043464HM7 2244227 00000043477HM9 2244227 00000043478HM7 2244227 00000043479HM5 2244227 00000043778HM0 2244227 00000043788HM9 2244227 00000043799HM6 2244227 00000043800HM2 2244227 00000043994HM3 2244227 00000043995HM0 2244227 00000043996HM8 2244227 00000044053HM7 2244227 00000044054HM5 2244227 00000044107HM1 2244227 00000044108HM9 2244227 00000044109HM7 2244227 00000044110HM5 2244227 00000044550HM2 2244227 00000044551HM0 2244227 00000044552HM8 2244227 00000044580HM9 2244227 00000044581HM7 2244227 00000044582HM5 2244227 00000045128HM6 2244227 00000045129HM4 2244227 00000045130HM2 2244227 00000045131HM0 2244227 00000045132HM8 2244227 00000045133HM6 2244227 00000045134HM4 2244227 00000045489HM2 2244227 00000045518HM8 2244227 00000045541HM0 2244227 00000045542HM8 2244227 00000045543HM6 2244227 00000045544HM4 2244227 00000045545HM1 2244227 00000045546HM9 2244227 00000045997HM4 2244227 00000045998HM2 2244227 00000046022HM0 2244227 00000046023HM8 2244227 00000046024HM6 2244227 00000046025HM3 2244227 00000046026HM1 2244227 00000046027HM9 2244227 00000046028HM7 2244227 00000046171HM5 2244227 00000046552HM6 2244227 00000046562HM5 2244227 00000046595HM5 2244227 00000046596HM3 2244227 00000046597HM1 2244227 00000046598HM9 2244227 00000046599HM7 2244227 00000046600HM3 2244227 00000046650HM8 2244227 00000047157HM3 2244227 00000047158HM1 2244227 00000047159HM9 2244227 00000047201HM9 2244227 00000047211HM8 2244227 00000047594HM7 2244227 00000047595HM4 2244227 00000047770HM3 2244227 00000047771HM1 2244227 00000047789HM3 2244227 00000047799HM2 2244227 00000047810HM7 2244227 00000047854HM5 2244227 00000047855HM2 2244227 00000048193HM7 2244227 00000048215HM8 2244227 00000048216HM6 2244227 00000048217HM4 2244227 00000048526HM8 2244227 00000048536HM7 2244227 00000048546HM6 2244227 00000048575HM5 2244227 00000048576HM3 2244227 00000048785HM0 2244227 00000048786HM8 2244227 00000048787HM6 2244227 00000048818HM9 2244227 00000048828HM8 2244227 00000049096HM1 2244227 00000049097HM9 2244227 00000049098HM7 2244227 00000049177HM9 2244227 00000049178HM7 2244227 00000049179HM5 2244227 00000049180HM3 2244227 00000049239HM7 2244227 00000049240HM5 2244227 00000049637HM2 2244227 00000049648HM9 2244227 00000049649HM7 2244227 00000049943HM4 2244227 00000049944HM2 2244227 00000049945HM9 2244227 00000049946HM7 2244227 00000049947HM5 2244227 00000049948HM3 2244227 00000049949HM1 2244227 00000049950HM9 2244227 00000050086HM8 2244227 00000050087HM6 2244227 00000050706HM1 2244227 00000050707HM9 2244227 00000050708HM7 2244227 00000051359HM8 2244227 00000052062HM7 2244227 00000052116HM1 2244227 00000052118HM7 2244227 00000052288HM8 2244227 00000052289HM6 2244227 00000052290HM4 2244227 00000052318HM3 2244227 00000052607HM9 2244227 00000052626HM9 2244227 00000052774HM7 2244227 00000052795HM2 2244227 00000052796HM0 2244227 00000052797HM8 2244227 00000052798HM6 2244227 00000052832HM3 2244227 00000053015HM4 2244227 00000053026HM1 2244227 00000053027HM9 2244227 00000053339HM8 2244227 00000053358HM8 2244227 00000053368HM7 2244227 00000053599HM7 2244227 00000053600HM3 2244227 00000053601HM1 2244227 00000053657HM3 2244227 00000053658HM1 2244227 00000053659HM9 2244227 00000053685HM4 2244227 00000054207HM6 2244227 00000054225HM8 2244227 00000054468HM4 2244227 00000054469HM2 2244227 00000054699HM4 2244227 00000054700HM0 2244227 00000054701HM8 2244227 00000054831HM3 2244227 00000054832HM1 2244227 00000054858HM6 2244227 00000054888HM3 2244227 00000054889HM1 2244227 00000054890HM9 2244227 00000054988HM1 2244227 00000054989HM9 2244227 00000054990HM7 2244227 00000055614HM2 2244227 00000055632HM4 2244227 00000055683HM7 2244227 00000055684HM5 2244227 00000055685HM2 2244227 00000055916HM1 2244227 00000055917HM9 2244227 00000055918HM7 2244227 00000056117HM5 2244227 00000056118HM3 2244227 00000056505HM1 2244227 00000056506HM9 2244227 00000056539HM0 2244227 00000056540HM8 2244227 00000056647HM1 2244227 00000056648HM9 2244227 00000056828HM7 2244227 00000056829HM5 2244227 00000056830HM3 2244227 00000056831HM1 2244227 00000057082HM0 2244227 00000057083HM8 2244227 00000057084HM6 2244227 00000057133HM1 2244227 00000057134HM9 2244227 00000057402HM0 2244227 00000057403HM8 2244227 00000057404HM6 2244227 00000057478HM0 2244227 00000057479HM8 2244227 00000057505HM0 2244227 00000057515HM9 2244227 00000057842HM7 2244227 00000057843HM5 2244227 00000057861HM7 2244227 00000058184HM3 2244227 00000058185HM0 2244227 00000058330HM2 2244227 00000058377HM3 2244227 00000058756HM8 2244227 00000058897HM0 2244227 00000058898HM8 2244227 00000058917HM6 2244227 00000058918HM4 2244227 00000059019HM0 2244227 00000059020HM8 2244227 00000059245HM1 2244227 00000059266HM7 2244227 00000059267HM5 2244227 00000059268HM3 2244227 00000059313HM7 2244227 00000059314HM5 2244227 00000059357HM4 2244227 00000059377HM2 2244227 00000059378HM0 2244227 00000059487HM9 2244227 00000059488HM7 2244227 00000059551HM2 2244227 00000059552HM0 2244227 00000059623HM9 2244227 00000059624HM7 2244227 00000059625HM4 2244227 00000059989HM4 2244227 00000059990HM2 2244227 00000060229HM2 2244227 00000060229HM2 2244227 00000060280HM5 2244227 00000060281HM3 2244227 00000060341HM5 2244227 00000060381HM1 2244227 00000060382HM9 2244227 00000060383HM7 2244227 00000060385HM2 2244227 00000060386HM0 2244227 00000060387HM8 2244227 00000060388HM6 2244227 00000061229HM1 2244227 00000061230HM9 2244227 00000061231HM7 2244227 00000061324HM0 2244227 00000061335HM6 2244227 00000061336HM4 2244227 00000061463HM6 2244227 00000061464HM4 2244227 00000061635HM9 2244227 00000061636HM7 2244227 00000061654HM0 2244227 00000061665HM6 2244227 00000061666HM4 2244227 00000061984HM1 2244227 00000061985HM8 2244227 00000061986HM6 2244227 00000062211HM8 2244227 00000062212HM6 2244227 00000062230HM8 2244227 00000062342HM1 2244227 00000062353HM8 2244227 00000062354HM6 2244227 00000062486HM6 2244227 00000062487HM4 2244227 00000062488HM2 2244227 00000062672HM1 2244227 00000062692HM9 2244227 00000063006HM1 2244227 00000063007HM9 2244227 00000063008HM7 2244227 00000063211HM7 2244227 00000063231HM5 2244227 00000063241HM4 2244227 00000063251HM3 2244227 00000063424HM6 2244227 00000063425HM3 2244227 00000063443HM6 2244227 00000063525HM0 2244227 00000063526HM8 2244227 00000063710HM8 2244227 00000063711HM6 2244227 00000063729HM8 2244227 00000063907HM0 2244227 00000064026HM8 2244227 00000064163HM9 2244227 00000064164HM7 2244227 00000064182HM9 2244227 00000064314HM8 2244227 00000064325HM4 2244227 00000064326HM2 2244227 00000064549HM9 2244227 00000064560HM6 2244227 00000064561HM4 2244227 00000064934HM3 2244227 00000064953HM3 2244227 00000064963HM2 2244227 00000065086HM1 2244227 00000065124HM0 2244227 00000065125HM7 2244227 00000065380HM8 2244227 00000065634HM8 2244227 00000065771HM8 2244227 00000065772HM6 2244227 00000065790HM8 2244227 00000066154HM6 2244227 00000066164HM5 2244227 00000066766HM7 2244227 00000066767HM5 2244227 00000066768HM3 2244227 00000066795HM6 2244227 00000066796HM4 2244227 00000066821HM0 2244227 00000066822HM8 2244227 00000067163HM6 2244227 00000067174HM3 2244227 00000067175HM0 2244227 00000067497HM8 2244227 00000067498HM6 2244227 00000068080HM1 2244227 00000068081HM9 2244227 00000068118HM9 2244227 00000068137HM9 2244227 00000068702HM0 2244227 00000068703HM8 2244227 00000068704HM6 2244227 00000068940HM6 2244227 00000069149HM3 2244227 00000069150HM1 2244227 00000069202HM0 2244227 00000069203HM8 2244227 00000069407HM5 2244227 00000069426HM5 2244227 00000069774HM8 2244227 00000069793HM8 2244227 00000070032HM8 2244227 00000070033HM6 2244227 00000070051HM8 2244227 00000070478HM3 2244227 00000070479HM1 2244227 00000070480HM9 2244227 00000070507HM9 2244227 00000070871HM9 2244227 00000070872HM7 2244227 00000071098HM8 2244227 00000071099HM6 2244227 00000071343HM8 2244227 00000071463HM4 2244227 00000071464HM2 2244227 00000071465HM9 2244227 00000071770HM2 2244227 00000071863HM5 2244227 00000071873HM4 2244227 00000071884HM1 2244227 00000071885HM8 2244227 00000072115HM9 2244227 00000072497HM1 2244227 00000072498HM9 2244227 00000074792HM3 2244227 00000079515HM3 2244227 00000745535YM5 2244227 00071863HM5000 2244227 UNK22442270001 2244227 UNK22442270002 2246999 00000054293HM6 2246999 00000054295HM1 2246999 00000054916HM2 2246999 00000055660HM5 2246999 00000061909HM8 2246999 00000061910HM6 2246999 00000061911HM4 2246999 00000062413HM0 2246999 00000065183HM6 2246999 00000068256HM7 2246999 00000069339HM0 2246999 00000069357HM2 2246999 00000069734HM2 2246999 00000070197HM9 2246999 00000070340HM5 2246999 00000070341HM3 2246999 00000070932HM9 2246999 00000071792HM6 2246999 00000072187HM8 2246999 00000072188HM6 2246999 00000072459HM1 2246999 00000072460HM9 2246999 00000072969HM9 2246999 00000072970HM7 2246999 00000073303HM0 2246999 00000074755YM3 2246999 00000076720HM2 2246999 00000077114HM7 2246999 00000078540HM2 2246999 00000078541HM0 2246999 00000078831HM5 2246999 00000078832HM3 2246999 00000079592HM2 2246999 00000079932HM0 2246999 00000079933HM8 2246999 00000083074HM5 2246999 00000083393HM9 2246999 00000084625HM3 2246999 00000084943HM0 2246999 00000084944HM8 2246999 00000089101HM0 2246999 00000090603HM2 2246999 00000093425HM7 2246999 00000695881YM3 2246999 00000695882YM1 2246999 00000695884YM7 2246999 00000695886YM2 2246999 00000695887YM0 2246999 00000695889YM6 2246999 00000695892YM0 2246999 00000695893YM8 2246999 00000695894YM6 2246999 00000695895YM3 2246999 00000695897YM9 2246999 00000695898YM7 2246999 00000695899YM5 2246999 00000695901YM9 2246999 00000695906YM8 2246999 00000695907YM6 2246999 00000695908YM4 2246999 00000695909YM2 2246999 00000695910YM0 2246999 00000695913YM4 2246999 00000695914YM2 2246999 00000695916YM7 2246999 00000695917YM5 2246999 00000695918YM3 2246999 00000695919YM1 2246999 00000695920YM9 2246999 00000695923YM3 2246999 00000695925YM8 2246999 00000695926YM6 2246999 00000695927YM4 2246999 00000695929YM0 2246999 00000695930YM8 2246999 00000695931YM6 2246999 00000695933YM2 2246999 00000695934YM0 2246999 00000695935YM7 2246999 00000695936YM5 2246999 00000695937YM3 2246999 00000695940YM7 2246999 00000695941YM5 2246999 00000695942YM3 2246999 00000695943YM1 2246999 00000695944YM9 2246999 00000695945YM6 2246999 00000695947YM2 2246999 00000695948YM0 2246999 00000695949YM8 2246999 00000695950YM6 2246999 00000695951YM4 2246999 00000695952YM2 2246999 00000695953YM0 2246999 00000695956YM3 2246999 00000695957YM1 2246999 00000695958YM9 2246999 00000747519YM7 2246999 00000747521YM3 2246999 00000747522YM1 2246999 00000747523YM9 2246999 00000747524YM7 2246999 00000747525YM4 2246999 00000747526YM2 2246999 00000747527YM0 2246999 00000747528YM8 2246999 00000747529YM6 2246999 00000747530YM4 2246999 00000747531YM2 2246999 00000747534YM6 2246999 00000747536YM1 2246999 00000747538YM7 2246999 00000747539YM5 2246999 00000747540YM3 2246999 00000747541YM1 2246999 00000747543YM7 2246999 00000747545YM2 2246999 00000747546YM0 2246999 00000747547YM8 2246999 00000747548YM6 2246999 00000747549YM4 2246999 00000747550YM2 2246999 00000747552YM8 2246999 00000747553YM6 2246999 00000747554YM4 2246999 00000747555YM1 2246999 00000747557YM7 2246999 00000747560YM1 2246999 00000747561YM9 2246999 00000747563YM5 2246999 00000747564YM3 2246999 00000747565YM0 2246999 00000747566YM8 2246999 00000747567YM6 2246999 00000747568YM4 2246999 00000747569YM2 2246999 00000747570YM0 2246999 00000747574YM2 2246999 00000747575YM9 2246999 00000747576YM7 2246999 00000747577YM5 2246999 00000747579YM1 2246999 00000747580YM9 2246999 00000747581YM7 2246999 00000747582YM5 2246999 00000747583YM3 2246999 00000747584YM1 2246999 00000747585YM8 2246999 00000747586YM6 2246999 00000747587YM4 2246999 00000747588YM2 2246999 00000747592YM4 2246999 00000747593YM2 2246999 00000747594YM0 2246999 00000747595YM7 2246999 00000747597YM3 2246999 00000747600YM5 2246999 00000747601YM3 2246999 00000747604YM7 2246999 00000747605YM4 2246999 00000747606YM2 2246999 00000747607YM0 2246999 00000747608YM8 2246999 00000747609YM6 2246999 00000747610YM4 2246999 00000747611YM2 2246999 00000747612YM0 2246999 00000747613YM8 2246999 00000747614YM6 2246999 00000747615YM3 2246999 00000747616YM1 2246999 00000747617YM9 2246999 00000747618YM7 2246999 00000747619YM5 2246999 00000747620YM3 2246999 00000747621YM1 2246999 00000747622YM9 2246999 00000747623YM7 2246999 00000747627YM8 2246999 00000747629YM4 2246999 00000747630YM2 2246999 00000747632YM8 2246999 00000747633YM6 2246999 00000747634YM4 2246999 00000747635YM1 2246999 00000747636YM9 2246999 00000747637YM7 2246999 00000747638YM5 2246999 00000747639YM3 2246999 00000747641YM9 2246999 00000747642YM7 2246999 00000747643YM5 2246999 00000747644YM3 2246999 00000747645YM0 2246999 00000747651YM8 2246999 00000747652YM6 2246999 00000747655YM9 2246999 00000747656YM7 2246999 00000747657YM5 2246999 00000747659YM1 2246999 00000747660YM9 2246999 00000747661YM7 2246999 00000747662YM5 2246999 00000747663YM3 2246999 00000747664YM1 2246999 00000747665YM8 2246999 00000747666YM6 2246999 00000747667YM4 2246999 00000747669YM0 2246999 00000747670YM8 2246999 00000747671YM6 2246999 00000747672YM4 2246999 00000747673YM2 2246999 00000747674YM0 2246999 00000747675YM7 2246999 00000747677YM3 2246999 00000747678YM1 2246999 00000747679YM9 2246999 00000747680YM7 2246999 00000747681YM5 2246999 00000747682YM3 2246999 00000747684YM9 2246999 00000747685YM6 2246999 00000747686YM4 2246999 00000747687YM2 2246999 00000747688YM0 2246999 00000747689YM8 2246999 00000747690YM6 2246999 00000747691YM4 2246999 00000747692YM2 2246999 00000747694YM8 2246999 00000747695YM5 2246999 00000747696YM3 2246999 00000747697YM1 2246999 00000747698YM9 2246999 00000747700YM3 2246999 00000747701YM1 2246999 00000747702YM9 2246999 00000747703YM7 2246999 00000747704YM5 2246999 00000747705YM2 2246999 00000747706YM0 2246999 00000747707YM8 2246999 00000747708YM6 2246999 00000747709YM4 2246999 00000747710YM2 2246999 00000747711YM0 2246999 00000747713YM6 2246999 00000747714YM4 2246999 00000747715YM1 2246999 00000747716YM9 2246999 00000747717YM7 2246999 00000812111YM3 2246999 00000812112YM1 2246999 00000812114YM7 2246999 00000812115YM4 2246999 00000812116YM2 2246999 00000812117YM0 2246999 00000812118YM8 2246999 00000812119YM6 2246999 00000812120YM4 2246999 00000812121YM2 2246999 00000812122YM0 2246999 00000812125YM3 2246999 00000812126YM1 2246999 00000812127YM9 2246999 00000812128YM7 2246999 00000812129YM5 2246999 00000812130YM3 2246999 00000812131YM1 2246999 00000812132YM9 2246999 00000812133YM7 2246999 00000812134YM5 2246999 00000812135YM2 2246999 00000812136YM0 2246999 00000812137YM8 2246999 00000812138YM6 2246999 00000812139YM4 2246999 00000817688YM5 2246999 00000817689YM3 2246999 00000817690YM1 2246999 00000817691YM9 2246999 00000817694YM3 2246999 00000817695YM0 2246999 00000817696YM8 2246999 00000817697YM6 2246999 00000817699YM2 2246999 00000817700YM8 2246999 00000817701YM6 2246999 00000817702YM4 2246999 00000817703YM2 2246999 00000817705YM7 2246999 00000817708YM1 2246999 00000817709YM9 2246999 00000817711YM5 2246999 00000817712YM3 2246999 00000817713YM1 2246999 00000817714YM9 2246999 00000817715YM6 2246999 00000821956YM0 2246999 00000821957YM8 2246999 00000821958YM6 2246999 00000821959YM4 2246999 00000821960YM2 2246999 00000821962YM8 2246999 00000821963YM6 2246999 00000821964YM4 2246999 00000821965YM1 2246999 00000821966YM9 2246999 00000821967YM7 2246999 00000821969YM3 2246999 00000821971YM9 2246999 00000821972YM7 2246999 00000821973YM5 2246999 00000821974YM3 2246999 00000821978YM4 2246999 00000821979YM2 2246999 00000821980YM0 2246999 00000821981YM8 2246999 00000821982YM6 2246999 00000821983YM4 2246999 00000821984YM2 2246999 00000821985YM9 2246999 00000829546YM1 2246999 00000829547YM9 2246999 00000829548YM7 2246999 00000829549YM5 2246999 00000829550YM3 2246999 00000829552YM9 2246999 00000829553YM7 2246999 00000829554YM5 2246999 00000829555YM2 2246999 00000829558YM6 2246999 00000829559YM4 2246999 00000829561YM0 2246999 00000829562YM8 2246999 00000829563YM6 2246999 00000829564YM4 2246999 00000829565YM1 2246999 00000829566YM9 2246999 00000829567YM7 2246999 00000829568YM5 2246999 00000829569YM3 2246999 00000829570YM1 2246999 00000829572YM7 2246999 00000829573YM5 2246999 00000829574YM3 2246999 00000829575YM0 2246999 00000835496YM1 2246999 00000835497YM9 2246999 00000835498YM7 2246999 00000835499YM5 2246999 00000835501YM8 2246999 00000835502YM6 2246999 00000835503YM4 2246999 00000835504YM2 2246999 00000835506YM7 2246999 00000835509YM1 2246999 00000835510YM9 2246999 00000835511YM7 2246999 00000835512YM5 2246999 00000835513YM3 2246999 00000835514YM1 2246999 00000835515YM8 2246999 00000835516YM6 2246999 00000835517YM4 2246999 00000835518YM2 2246999 00000835520YM8 2246999 00000835521YM6 2246999 00000835522YM4 2246999 00000835523YM2 2246999 00000835524YM0 2246999 00000835525YM7 2246999 00000852057YM9 2246999 00000852058YM7 2246999 00000852060YM3 2246999 00000852061YM1 2246999 00000852062YM9 2246999 00000852063YM7 2246999 00000852065YM2 2246999 00000852067YM8 2246999 00000852068YM6 2246999 00000852069YM4 2246999 00000852070YM2 2246999 00000852073YM6 2246999 00000852074YM4 2246999 00000852075YM1 2246999 00000852076YM9 2246999 00000852077YM7 2246999 00000852078YM5 2246999 00000852079YM3 2246999 00000852080YM1 2246999 00000852081YM9 2246999 00000852082YM7 2246999 00000852083YM5 2246999 00000852084YM3 2246999 00000852086YM8 2246999 00000870257YM3 2246999 00000870258YM1 2246999 00000870259YM9 2246999 00000870260YM7 2246999 00000870262YM3 2246999 00000870263YM1 2246999 00000870264YM9 2246999 00000870265YM6 2246999 00000870266YM4 2246999 00000870267YM2 2246999 00000870268YM0 2246999 00000870270YM6 2246999 00000870271YM4 2246999 00000870272YM2 2246999 00000870273YM0 2246999 00000870274YM8 2246999 00000870277YM1 2246999 00000870278YM9 2246999 00000870279YM7 2246999 00000870280YM5 2246999 00000870281YM3 2246999 00000870282YM1 2246999 00000870283YM9 2246999 00000870284YM7 2246999 00000870286YM2 2246999 00000879798YM7 2246999 00000879799YM5 2246999 00000879800YM1 2246999 00000879801YM9 2246999 00000879803YM5 2246999 00000879804YM3 2246999 00000879805YM0 2246999 00000879806YM8 2246999 00000879808YM4 2246999 00000879810YM0 2246999 00000879811YM8 2246999 00000879813YM4 2246999 00000879814YM2 2246999 00000879817YM5 2246999 00000879818YM3 2246999 00000879819YM1 2246999 00000879820YM9 2246999 00000879821YM7 2246999 00000879822YM5 2246999 00000879823YM3 2246999 00000879824YM1 2246999 00000879825YM8 2246999 00000879826YM6 2246999 00000884908YM5 2246999 00000884909YM3 2246999 00000884910YM1 2246999 00000884911YM9 2246999 00000884913YM5 2246999 00000884914YM3 2246999 00000884915YM0 2246999 00000884916YM8 2246999 00000884918YM4 2246999 00000884919YM2 2246999 00000884920YM0 2246999 00000884921YM8 2246999 00000884922YM6 2246999 00000884923YM4 2246999 00000884925YM9 2246999 00000884926YM7 2246999 00000884927YM5 2246999 00000884928YM3 2246999 00000884929YM1 2246999 00000884930YM9 2246999 00000884931YM7 2246999 00000884933YM3 2246999 00000884934YM1 2246999 00000884936YM6 2246999 00000887936YM6 2246999 00000893348YM3 2246999 00000893349YM1 2246999 00000893350YM9 2246999 00000893351YM7 2246999 00000893352YM5 2246999 00000893353YM3 2246999 00000893354YM1 2246999 00000893355YM8 2246999 00000893356YM6 2246999 00000893357YM4 2246999 00000893358YM2 2246999 00000893360YM8 2246999 00000893361YM6 2246999 00000893362YM4 2246999 00000893363YM2 2246999 00000893365YM7 2246999 00000893366YM5 2246999 00000893369YM9 2246999 00000893370YM7 2246999 00000893371YM5 2246999 00000893373YM1 2246999 00000893374YM9 2246999 00000893375YM6 2246999 00000893376YM4 2246999 00000906591YM3 2246999 00000906592YM1 2246999 00000906593YM9 2246999 00000906595YM4 2246999 00000906597YM0 2246999 00000906598YM8 2246999 00000906599YM6 2246999 00000906600YM2 2247000 00000053592HM2 2247000 00000062433HM8 2247000 00000062434HM6 2247000 00000062435HM3 2247000 00000068751HM7 2247000 00000068752HM5 2247000 00000696078YM5 2247000 00000696079YM3 2247000 00000696080YM1 2247000 00000696081YM9 2247000 00000696083YM5 2247000 00000696084YM3 2247000 00000696085YM0 2247000 00000696087YM6 2247000 00000747718YM5 2247000 00000747719YM3 2247000 00000747720YM1 2247000 00000747723YM5 2247000 00000747724YM3 2247000 00000747726YM8 2247000 00000747727YM6 2247000 00000747729YM2 2247000 00000747730YM0 2247000 00000747731YM8 2247000 00000747732YM6 2247000 00000747733YM4 2247000 00000747734YM2 2247000 00000747735YM9 2247000 00000747736YM7 2247000 00000747738YM3 2247000 00000747740YM9 2247000 00000747741YM7 2247000 00000747743YM3 2247000 00000747744YM1 2247000 00000747746YM6 2247000 00000747747YM4 2247000 00000747748YM2 2247000 00000747749YM0 2247000 00000747750YM8 2247000 00000747751YM6 2247000 00000747752YM4 2247001 00000054294HM4 2247001 00000056079HM7 2247001 00000060961HM0 2247001 00000060962HM8 2247001 00000062240HM7 2247001 00000065391HM5 2247001 00000065392HM3 2247001 00000067062HM0 2247001 00000071811HM4 2247001 00000073006HM9 2247001 00000073796HM5 2247001 00000074816HM0 2247001 00000075516HM5 2247001 00000075517HM3 2247001 00000076439HM9 2247001 00000076440HM7 2247001 00000081497HM0 2247001 00000081498HM8 2247001 00000081499HM6 2247001 00000081720HM5 2247001 00000081721HM3 2247001 00000086819HM0 2247001 00000088061HM7 2247001 00000088467HM6 2247001 00000090938HM2 2247001 00000091992HM8 2247001 00000092195HM7 2247001 00000092885HM3 2247001 00000696283YM1 2247001 00000696286YM4 2247001 00000696287YM2 2247001 00000696288YM0 2247001 00000696289YM8 2247001 00000696290YM6 2247001 00000696293YM0 2247001 00000696294YM8 2247001 00000696295YM5 2247001 00000696299YM7 2247001 00000696300YM3 2247001 00000696301YM1 2247001 00000696302YM9 2247001 00000696303YM7 2247001 00000696305YM2 2247001 00000696306YM0 2247001 00000696309YM4 2247001 00000696310YM2 2247001 00000696311YM0 2247001 00000696313YM6 2247001 00000696315YM1 2247001 00000696318YM5 2247001 00000696321YM9 2247001 00000696322YM7 2247001 00000696323YM5 2247001 00000696324YM3 2247001 00000696327YM6 2247001 00000696328YM4 2247001 00000696329YM2 2247001 00000696330YM0 2247001 00000696333YM4 2247001 00000696334YM2 2247001 00000696335YM9 2247001 00000696336YM7 2247001 00000696337YM5 2247001 00000696339YM1 2247001 00000696340YM9 2247001 00000696342YM5 2247001 00000696344YM1 2247001 00000696345YM8 2247001 00000696347YM4 2247001 00000696348YM2 2247001 00000696349YM0 2247001 00000696350YM8 2247001 00000696351YM6 2247001 00000696352YM4 2247001 00000696353YM2 2247001 00000696354YM0 2247001 00000696355YM7 2247001 00000696356YM5 2247001 00000696357YM3 2247001 00000696358YM1 2247001 00000696359YM9 2247001 00000696362YM3 2247001 00000696363YM1 2247001 00000696364YM9 2247001 00000696365YM6 2247001 00000696366YM4 2247001 00000696367YM2 2247001 00000696368YM0 2247001 00000696369YM8 2247001 00000696370YM6 2247001 00000696372YM2 2247001 00000696373YM0 2247001 00000696375YM5 2247001 00000696376YM3 2247001 00000696377YM1 2247001 00000696378YM9 2247001 00000696381YM3 2247001 00000696382YM1 2247001 00000696383YM9 2247001 00000696384YM7 2247001 00000696385YM4 2247001 00000696386YM2 2247001 00000696387YM0 2247001 00000696388YM8 2247001 00000696389YM6 2247001 00000696390YM4 2247001 00000696391YM2 2247001 00000696392YM0 2247001 00000696393YM8 2247001 00000696394YM6 2247001 00000696397YM9 2247001 00000696399YM5 2247001 00000696401YM9 2247001 00000696402YM7 2247001 00000696403YM5 2247001 00000696404YM3 2247001 00000696405YM0 2247001 00000696406YM8 2247001 00000696407YM6 2247001 00000696408YM4 2247001 00000696409YM2 2247001 00000696410YM0 2247001 00000696411YM8 2247001 00000696412YM6 2247001 00000696414YM2 2247001 00000696415YM9 2247001 00000696416YM7 2247001 00000696418YM3 2247001 00000696419YM1 2247001 00000696420YM9 2247001 00000696423YM3 2247001 00000696424YM1 2247001 00000696425YM8 2247001 00000696426YM6 2247001 00000696428YM2 2247001 00000696430YM8 2247001 00000696431YM6 2247001 00000696432YM4 2247001 00000696433YM2 2247001 00000696436YM5 2247001 00000696438YM1 2247001 00000696439YM9 2247001 00000696440YM7 2247001 00000696441YM5 2247001 00000696443YM1 2247001 00000696444YM9 2247001 00000696445YM6 2247001 00000696447YM2 2247001 00000696449YM8 2247001 00000696450YM6 2247001 00000696453YM0 2247001 00000696454YM8 2247001 00000696455YM5 2247001 00000696456YM3 2247001 00000696457YM1 2247001 00000696458YM9 2247001 00000696459YM7 2247001 00000696460YM5 2247001 00000696461YM3 2247001 00000747918YM1 2247001 00000747920YM7 2247001 00000747923YM1 2247001 00000747924YM9 2247001 00000747925YM6 2247001 00000747926YM4 2247001 00000747927YM2 2247001 00000747928YM0 2247001 00000747929YM8 2247001 00000747930YM6 2247001 00000747931YM4 2247001 00000747932YM2 2247001 00000747933YM0 2247001 00000747934YM8 2247001 00000747935YM5 2247001 00000747937YM1 2247001 00000747940YM5 2247001 00000747941YM3 2247001 00000747942YM1 2247001 00000747943YM9 2247001 00000747945YM4 2247001 00000747947YM0 2247001 00000747948YM8 2247001 00000747951YM2 2247001 00000747952YM0 2247001 00000747953YM8 2247001 00000747954YM6 2247001 00000747955YM3 2247001 00000747957YM9 2247001 00000747958YM7 2247001 00000747960YM3 2247001 00000747961YM1 2247001 00000747962YM9 2247001 00000747963YM7 2247001 00000747965YM2 2247001 00000747966YM0 2247001 00000747967YM8 2247001 00000747968YM6 2247001 00000747970YM2 2247001 00000747971YM0 2247001 00000747972YM8 2247001 00000747973YM6 2247001 00000747974YM4 2247001 00000747975YM1 2247001 00000747976YM9 2247001 00000747978YM5 2247001 00000747979YM3 2247001 00000747981YM9 2247001 00000747982YM7 2247001 00000747983YM5 2247001 00000747985YM0 2247001 00000747986YM8 2247001 00000747987YM6 2247001 00000747988YM4 2247001 00000747989YM2 2247001 00000747991YM8 2247001 00000747992YM6 2247001 00000747993YM4 2247001 00000747995YM9 2247001 00000747996YM7 2247001 00000748001YM5 2247001 00000748002YM3 2247001 00000748003YM1 2247001 00000748004YM9 2247001 00000748006YM4 2247001 00000748007YM2 2247001 00000748008YM0 2247001 00000748009YM8 2247001 00000748010YM6 2247001 00000748011YM4 2247001 00000748012YM2 2247001 00000748013YM0 2247001 00000748015YM5 2247001 00000748016YM3 2247001 00000748017YM1 2247001 00000748019YM7 2247001 00000748020YM5 2247001 00000748021YM3 2247001 00000748022YM1 2247001 00000748025YM4 2247001 00000748026YM2 2247001 00000748027YM0 2247001 00000748028YM8 2247001 00000748029YM6 2247001 00000748030YM4 2247001 00000748031YM2 2247001 00000748033YM8 2247001 00000748034YM6 2247001 00000748035YM3 2247001 00000748036YM1 2247001 00000748037YM9 2247001 00000748039YM5 2247001 00000748040YM3 2247001 00000748041YM1 2247001 00000748043YM7 2247001 00000748044YM5 2247001 00000748045YM2 2247001 00000748046YM0 2247001 00000748047YM8 2247001 00000748048YM6 2247001 00000748049YM4 2247001 00000748050YM2 2247001 00000748051YM0 2247001 00000748054YM4 2247001 00000748055YM1 2247001 00000748056YM9 2247001 00000748057YM7 2247001 00000748058YM5 2247001 00000748059YM3 2247001 00000748060YM1 2247001 00000748062YM7 2247001 00000748063YM5 2247001 00000748065YM0 2247001 00000748066YM8 2247001 00000748067YM6 2247001 00000748068YM4 2247001 00000748069YM2 2247001 00000748071YM8 2247001 00000748072YM6 2247001 00000748073YM4 2247001 00000748074YM2 2247001 00000748077YM5 2247001 00000748078YM3 2247001 00000748079YM1 2247001 00000748080YM9 2247001 00000748081YM7 2247001 00000748082YM5 2247001 00000748083YM3 2247001 00000748084YM1 2247001 00000748088YM2 2247001 00000748089YM0 2247001 00000748091YM6 2247001 00000748094YM0 2247001 00000748095YM7 2247001 00000748096YM5 2247001 00000748097YM3 2247001 00000748098YM1 2247001 00000748099YM9 2247001 00000748100YM5 2247001 00000748101YM3 2247001 00000748102YM1 2247001 00000748104YM7 2247001 00000748105YM4 2247001 00000748106YM2 2247001 00000748107YM0 2247001 00000748107YMO 2247001 00000748108YM8 2247001 00000748109YM6 2247001 00000748111YM2 2247001 00000748112YM0 2247001 00000748113YM8 2247001 00000748114YM6 2247001 00000748115YM3 2247001 00000748116YM1 2247001 00000748117YM9 2247001 00000818356YM8 2247001 00000818361YM8 2247001 00000818364YM2 2247001 00000818366YM7 2247001 00000818367YM5 2247001 00000818368YM3 2247001 00000818369YM1 2247001 00000818370YM9 2247001 00000818371YM7 2247001 00000818373YM3 2247001 00000818374YM1 2247001 00000818375YM8 2247001 00000818376YM6 2247001 00000818377YM4 2247001 00000818378YM2 2247001 00000818379YM0 2247001 00000818380YM8 2247001 00000818381YM6 2247001 00000818382YM4 2247001 00000818384YM0 2247001 00000818385YM7 2247001 00000826646YM2 2247001 00000826647YM0 2247001 00000826648YM8 2247001 00000826649YM6 2247001 00000826651YM2 2247001 00000826653YM8 2247001 00000826654YM6 2247001 00000826655YM3 2247001 00000826656YM1 2247001 00000826657YM9 2247001 00000826658YM7 2247001 00000826659YM5 2247001 00000826660YM3 2247001 00000826661YM1 2247001 00000826662YM9 2247001 00000826663YM7 2247001 00000826664YM5 2247001 00000826665YM2 2247001 00000826666YM0 2247001 00000826667YM8 2247001 00000826668YM6 2247001 00000826670YM2 2247001 00000826671YM0 2247001 00000826672YM8 2247001 00000826673YM6 2247001 00000826674YM4 2247001 00000826675YM1 2247001 00000826678YM5 2247001 00000826680YM1 2247001 00000826681YM9 2247001 00000826682YM7 2247001 00000826683YM5 2247001 00000826685YM0 2247001 00000826686YM8 2247001 00000826687YM6 2247001 00000826688YM4 2247001 00000826689YM2 2247001 00000826690YM0 2247001 00000826691YM8 2247001 00000826692YM6 2247001 00000826693YM4 2247001 00000826695YM9 2247001 00000826696YM7 2247001 00000826698YM3 2247001 00000826699YM1 2247001 00000826700YM7 2247001 00000826701YM5 2247001 00000826703YM1 2247001 00000826704YM9 2247001 00000842777YM5 2247001 00000842778YM3 2247001 00000842779YM1 2247001 00000842781YM7 2247001 00000842782YM5 2247001 00000842783YM3 2247001 00000842784YM1 2247001 00000842786YM6 2247001 00000842787YM4 2247001 00000842788YM2 2247001 00000842789YM0 2247001 00000842790YM8 2247001 00000842791YM6 2247001 00000842794YM0 2247001 00000842795YM7 2247001 00000842796YM5 2247001 00000842797YM3 2247001 00000842799YM9 2247001 00000842801YM3 2247001 00000842804YM7 2247001 00000866227YM2 2247001 00000866229YM8 2247001 00000866230YM6 2247001 00000866231YM4 2247001 00000866232YM2 2247001 00000866233YM0 2247001 00000866234YM8 2247001 00000866235YM5 2247001 00000866236YM3 2247001 00000866237YM1 2247001 00000866238YM9 2247001 00000866240YM5 2247001 00000866241YM3 2247001 00000866242YM1 2247001 00000866244YM7 2247001 00000866245YM4 2247001 00000866246YM2 2247001 00000866247YM0 2247001 00000866248YM8 2247001 00000866250YM4 2247001 00000866251YM2 2247001 00000866252YM0 2247001 00000866253YM8 2247001 00000866254YM6 2247001 00000866255YM3 2247001 00000866256YM1 2247001 00000877407YM7 2247001 00000877408YM5 2247001 00000877409YM3 2247001 00000877410YM1 2247001 00000877411YM9 2247001 00000877415YM0 2247001 00000877416YM8 2247001 00000877417YM6 2247001 00000877418YM4 2247001 00000877419YM2 2247001 00000877420YM0 2247001 00000877421YM8 2247001 00000877422YM6 2247001 00000877423YM4 2247001 00000877424YM2 2247001 00000877425YM9 2247001 00000877427YM5 2247001 00000877428YM3 2247001 00000877429YM1 2247001 00000877431YM7 2247001 00000877434YM1 2247001 00000877435YM8 2247001 00000877436YM6 2247001 00000894498YM5 2247001 00000894499YM3 2247001 00000894500YM8 2247001 00000894501YM6 2247001 00000894502YM4 2247001 00000894503YM2 2247001 00000894505YM7 2247001 00000894506YM5 2247001 00000894507YM3 2247001 00000894508YM1 2247001 00000894511YM5 2247001 00000894512YM3 2247001 00000894513YM1 2247001 00000894516YM4 2247001 00000894517YM2 2247001 00000894518YM0 2247001 00000894519YM8 2247001 00000894520YM6 2247001 00000894521YM4 2247001 00000894522YM2 2247001 00000894523YM0 2247001 00000894524YM8 2247001 00000894525YM5 2247001 00000894527YM1 2247001 00000924718YM0 2247001 00000924720YM6 2247002 00000064599HM4 2247002 00000066137HM1 2247002 00000078577HM4 2247002 00000748318YM3 2247002 00000748321YM7 2247002 00000748322YM5 2247002 00000748323YM3 2247002 00000748324YM1 2247002 00000748326YM6 2247002 00000748329YM0 2247002 00000748330YM8 2247002 00000748331YM6 2247002 00000748332YM4 2247002 00000748333YM2 2247002 00000748334YM0 2247002 00000748335YM7 2247002 00000748336YM5 2247002 00000748338YM1 2247002 00000748341YM5 2247002 00000748343YM1 2247002 00000748344YM9 2247002 00000748345YM6 2247002 00000748346YM4 2247002 00000748347YM2 2247002 00000748348YM0 2247002 00000748349YM8 2247002 00000748350YM6 2247002 00000748353YM0 2247002 00000748354YM8 2247002 00000748355YM5 2247002 00000748355YM7 2247002 00000748356YM3 2247002 00000748357YM1 2247002 00000748358YM9 2247002 00000748359YM7 2247002 00000748360YM5 2247002 00000748361YM3 2247002 00000748362YM1 2247002 00000748363YM9 2247002 00000748364YM7 2247002 00000748365YM4 2247002 00000748366YM2 2247002 00000748368YM8 2247002 00000748371YM2 2247002 00000748374YM6 2247002 00000748375YM3 2247002 00000748376YM1 2247002 00000748378YM7 2247002 00000748379YM5 2247002 00000748381YM1 2247002 00000748382YM9 2247002 00000748383YM7 2247002 00000748385YM2 2247002 00000748387YM8 2247002 00000748391YM0 2247002 00000748392YM8 2247002 00000748394YM4 2247002 00000748395YM1 2247002 00000748397YM7 2247002 00000748398YM5 2247002 00000748399YM3 2247002 00000748401YM7 2247002 00000748402YM5 2247002 00000748403YM3 2247002 00000748404YM1 2247002 00000748405YM8 2247002 00000748406YM6 2247002 00000748407YM4 2247002 00000748408YM2 2247002 00000748409YM0 2247002 00000748410YM8 2247002 00000748411YM6 2247002 00000748412YM4 2247002 00000748413YM2 2247002 00000748414YM0 2247002 00000748415YM7 2247002 00000748416YM5 2247002 00000748417YM3 2247002 00000748418YM1 2247002 00000748419YM9 2247002 00000748423YM1 2247002 00000748424YM9 2247002 00000748425YM6 2247002 00000748429YM8 2247002 00000748430YM6 2247002 00000748431YM4 2247002 00000748432YM2 2247002 00000748433YM0 2247002 00000748434YM8 2247002 00000748435YM5 2247002 00000748436YM3 2247002 00000748437YM1 2247002 00000748438YM9 2247002 00000748439YM7 2247002 00000748440YM5 2247002 00000748441YM3 2247002 00000748442YM1 2247002 00000748443YM9 2247002 00000748444YM7 2247002 00000748447YM0 2247002 00000748448YM8 2247002 00000696680YM8 2247002 00000696681YM6 2247002 00000696682YM4 2247002 00000696688YM1 2247002 00000696691YM5 2247002 00000696692YM3 2247002 00000696693YM1 2247002 00000696695YM6 2247002 00000696696YM4 2247002 00000696697YM2 2247002 00000696699YM8 2247002 00000696700YM4 2247002 00000696701YM2 2247002 00000696702YM0 2247002 00000696703YM8 2247002 00000696704YM6 2247002 00000696706YM1 2247002 00000696707YM9 2247002 00000696709YM5 2247002 00000696712YM9 2247002 00000696713YM7 2247002 00000696714YM5 2247002 00000696716YM0 2247002 00000696717YM8 2247002 00000696718YM6 2247002 00000696719YM4 2247002 00000696722YM8 2247002 00000696723YM6 2247002 00000696724YM4 2247002 00000696725YM1 2247002 00000696730YM1 2247006 00000056754HM5 2247006 00000061939HM5 2247006 00000064127HM4 2247006 00000064579HM6 2247006 00000066639HM6 2247006 00000068219HM5 2247006 00000068730HM1 2247006 00000069851HM4 2247006 00000069852HM2 2247006 00000070300HM9 2247006 00000070301HM7 2247006 00000070891HM7 2247006 00000070892HM5 2247006 00000071033HM5 2247006 00000071551HM6 2247006 00000071552HM4 2247006 00000073324HM6 2247006 00000073325HM3 2247006 00000073326HM1 2247006 00000073811HM2 2247006 00000075658HM5 2247006 00000075659HM3 2247006 00000076430HM8 2247006 00000078332HM4 2247006 00000079336HM4 2247006 00000079337HM2 2247006 00000079338HM0 2247006 00000079534HM4 2247006 00000079535HM1 2247006 00000079536HM9 2247006 00000079969HM2 2247006 00000082113HM2 2247006 00000082114HM0 2247006 00000082484HM7 2247006 00000083704HM7 2247006 00000084133HM8 2247006 00000085133HM7 2247006 00000085527HM0 2247006 00000086556HM8 2247006 00000087632HM6 2247006 00000088041HM9 2247006 00000089079HM8 2247006 00000089668HM8 2247006 00000089669HM6 2247006 00000090087HM8 2247006 00000090088HM6 2247006 00000090089HM4 2247006 00000091427HM5 2247006 00000091972HM0 2247006 00000092156HM9 2247006 00000092157HM7 2247006 00000095477HM6 2247006 00000099011HM9 2247006 00000100271HM6 2247006 00000104500HM4 2247006 00000105331HM3 2247006 00000106575HM4 2247006 00000107109HM1 2247006 00000109264HM2 2247006 00000111362HM0 2247006 00000120948HM5 2247006 00000126585HM9 2247006 00000128926HM3 2247006 00000131281HM8 2247006 00000134722HM8 2247006 00000137532HM8 2247006 00000766954YM2 2247006 00000766955YM9 2247006 00000766956YM7 2247006 00000766957YM5 2247006 00000766958YM3 2247006 00000766959YM1 2247006 00000766961YM7 2247006 00000766962YM5 2247006 00000766963YM3 2247006 00000766968YM2 2247006 00000766969YM0 2247006 00000766970YM8 2247006 00000766971YM6 2247006 00000766972YM4 2247006 00000766973YM2 2247006 00000766975YM7 2247006 00000766976YM5 2247006 00000766977YM3 2247006 00000766980YM7 2247006 00000766981YM5 2247006 00000766982YM3 2247006 00000766984YM9 2247006 00000766985YM6 2247006 00000766987YM2 2247006 00000766988YM0 2247006 00000766989YM8 2247006 00000766991YM4 2247006 00000766992YM2 2247006 00000766993YM0 2247006 00000766994YM8 2247006 00000766996YM3 2247006 00000766997YM1 2247006 00000766998YM9 2247006 00000766999YM7 2247006 00000767001YM1 2247006 00000767002YM9 2247006 00000767003YM7 2247006 00000767004YM5 2247006 00000767005YM2 2247006 00000767006YM0 2247006 00000767007YM8 2247006 00000767009YM4 2247006 00000767010YM2 2247006 00000767011YM0 2247006 00000767012YM8 2247006 00000767013YM6 2247006 00000767014YM4 2247006 00000767016YM9 2247006 00000767018YM5 2247006 00000767019YM3 2247006 00000767020YM1 2247006 00000767022YM7 2247006 00000767023YM5 2247006 00000767024YM3 2247006 00000767025YM0 2247006 00000767029YM2 2247006 00000767030YM0 2247006 00000767031YM8 2247006 00000767033YM4 2247006 00000767034YM2 2247006 00000767035YM9 2247006 00000767037YM5 2247006 00000767040YM9 2247006 00000767042YM5 2247006 00000767043YM3 2247006 00000767045YM8 2247006 00000767046YM6 2247006 00000767047YM4 2247006 00000767048YM2 2247006 00000767049YM0 2247006 00000767051YM6 2247006 00000767053YM2 2247006 00000767056YM5 2247006 00000767057YM3 2247006 00000767058YM1 2247006 00000767059YM9 2247006 00000767060YM7 2247006 00000767061YM5 2247006 00000767064YM9 2247006 00000767065YM6 2247006 00000767066YM4 2247006 00000767068YM0 2247006 00000767069YM8 2247006 00000767070YM6 2247006 00000767071YM4 2247006 00000767072YM2 2247006 00000767074YM8 2247006 00000767075YM5 2247006 00000767076YM3 2247006 00000767077YM1 2247006 00000767078YM9 2247006 00000767079YM7 2247006 00000767080YM5 2247006 00000767081YM3 2247006 00000767082YM1 2247006 00000767083YM9 2247006 00000767084YM7 2247006 00000767085YM4 2247006 00000767089YM6 2247006 00000767090YM4 2247006 00000767091YM2 2247006 00000767093YM8 2247006 00000767094YM6 2247006 00000767096YM1 2247006 00000767097YM9 2247006 00000767098YM7 2247006 00000767099YM5 2247006 00000767100YM1 2247006 00000767101YM9 2247006 00000767104YM3 2247006 00000767105YM0 2247006 00000767106YM8 2247006 00000767109YM2 2247006 00000767110YM0 2247006 00000767111YM8 2247006 00000767112YM6 2247006 00000767114YM2 2247006 00000767115YM9 2247006 00000767116YM7 2247006 00000767117YM5 2247006 00000767119YM1 2247006 00000767120YM9 2247006 00000767121YM7 2247006 00000767122YM5 2247006 00000767125YM8 2247006 00000767126YM6 2247006 00000767127YM4 2247006 00000767128YM2 2247006 00000767129YM0 2247006 00000767130YM8 2247006 00000767131YM6 2247006 00000767132YM4 2247006 00000767133YM2 2247006 00000767134YM0 2247006 00000767135YM7 2247006 00000767136YM5 2247006 00000767137YM3 2247006 00000767138YM1 2247006 00000767139YM9 2247006 00000767140YM7 2247006 00000767141YM5 2247006 00000767142YM3 2247006 00000767143YM1 2247006 00000767144YM9 2247006 00000767145YM6 2247006 00000767147YM2 2247006 00000767148YM0 2247006 00000767149YM8 2247006 00000767150YM6 2247006 00000767152YM2 2247006 00000767153YM0 2247006 00000852027YM2 2247006 00000852028YM0 2247006 00000852029YM8 2247006 00000852031YM4 2247006 00000852032YM2 2247006 00000852033YM0 2247006 00000852034YM8 2247006 00000852035YM5 2247006 00000852036YM3 2247006 00000852037YM1 2247006 00000852039YM7 2247006 00000852040YM5 2247006 00000852041YM3 2247006 00000852042YM1 2247006 00000852043YM9 2247006 00000852044YM7 2247006 00000852045YM4 2247006 00000852046YM2 2247006 00000852047YM0 2247006 00000852048YM8 2247006 00000852049YM6 2247006 00000852051YM2 2247006 00000852052YM0 2247006 00000852053YM8 2247006 00000852054YM6 2247006 00000852055YM3 2247006 00000852056YM1 2247006 00000865318YM0 2247006 00000865319YM8 2247006 00000865320YM6 2247006 00000865321YM4 2247006 00000865322YM2 2247006 00000865323YM0 2247006 00000865324YM8 2247006 00000865325YM5 2247006 00000865326YM3 2247006 00000865328YM9 2247006 00000865330YM5 2247006 00000865332YM1 2247006 00000865333YM9 2247006 00000865335YM4 2247006 00000865337YM0 2247006 00000865338YM8 2247006 00000865339YM6 2247006 00000865340YM4 2247006 00000865341YM2 2247006 00000865342YM0 2247006 00000865343YM8 2247006 00000865345YM3 2247006 00000865346YM1 2247006 00000873637YM3 2247006 00000873638YM1 2247006 00000873640YM7 2247006 00000873642YM3 2247006 00000873643YM1 2247006 00000873646YM4 2247006 00000873647YM2 2247006 00000873649YM8 2247006 00000873650YM6 2247006 00000873651YM4 2247006 00000873652YM2 2247006 00000873653YM0 2247006 00000873654YM8 2247006 00000873657YM1 2247006 00000873660YM5 2247006 00000873661YM3 2247006 00000873662YM1 2247006 00000873663YM9 2247006 00000873664YM7 2247006 00000873665YM4 2247006 00000873666YM2 2247006 00000886458YM9 2247006 00000886459YM7 2247006 00000886460YM5 2247006 00000886461YM3 2247006 00000886462YM1 2247006 00000886466YM2 2247006 00000886467YM0 2247006 00000886468YM8 2247006 00000886469YM6 2247006 00000886470YM4 2247006 00000886471YM2 2247006 00000886472YM0 2247006 00000886473YM8 2247006 00000886474YM6 2247006 00000886475YM3 2247006 00000886476YM1 2247006 00000886477YM9 2247006 00000886478YM7 2247006 00000886479YM5 2247006 00000886480YM3 2247006 00000886481YM1 2247006 00000886485YM2 2247006 00000886486YM0 2247006 00000901739YM3 2247006 00000901740YM1 2247006 00000901741YM9 2247006 00000901742YM7 2247006 00000901743YM5 2247006 00000901744YM3 2247006 00000901745YM0 2247006 00000901746YM8 2247006 00000901747YM6 2247006 00000901751YM8 2247006 00000901752YM6 2247006 00000901754YM2 2247006 00000901755YM9 2247006 00000901756YM7 2247006 00000901758YM3 2247006 00000901759YM1 2247006 00000901765YM8 2247006 00000901766YM6 2247006 00000930647YM3 2247010 00000056747HM9 2247010 00000058885HM5 2247010 00000059217HM0 2247010 00000060036HM1 2247010 00000060037HM9 2247010 00000060982HM6 2247010 00000062256HM3 2247010 00000062967HM5 2247010 00000062968HM3 2247010 00000063752HM0 2247010 00000064107HM6 2247010 00000064518HM4 2247010 00000065430HM1 2247010 00000065431HM9 2247010 00000065699HM1 2247010 00000065700HM7 2247010 00000066080HM3 2247010 00000066101HM7 2247010 00000066119HM9 2247010 00000066465HM6 2247010 00000066679HM2 2247010 00000066680HM0 2247010 00000066716HM2 2247010 00000066717HM0 2247010 00000066718HM8 2247010 00000067407HM7 2247010 00000067427HM5 2247010 00000067635HM3 2247010 00000068537HM0 2247010 00000068538HM8 2247010 00000068576HM8 2247010 00000070526HM9 2247010 00000071533HM4 2247010 00000073759HM3 2247010 00000073760HM1 2247010 00000074131HM4 2247010 00000074132HM2 2247010 00000074133HM0 2247010 00000074134HM8 2247010 00000074640HM4 2247010 00000074641HM2 2247010 00000074642HM0 2247010 00000074643HM8 2247010 00000075358HM2 2247010 00000075377HM2 2247010 00000075378HM0 2247010 00000075379HM8 2247010 00000075638HM7 2247010 00000076701HM2 2247010 00000076739HM2 2247010 00000077290HM5 2247010 00000077403HM4 2247010 00000077404HM2 2247010 00000077590HM8 2247010 00000077791HM2 2247010 00000078042HM9 2247010 00000078519HM6 2247010 00000078869HM5 2247010 00000078870HM3 2247010 00000079126HM9 2247010 00000079394HM3 2247010 00000080143HM1 2247010 00000080144HM9 2247010 00000080145HM6 2247010 00000080404HM7 2247010 00000080405HM4 2247010 00000080406HM2 2247010 00000080759HM4 2247010 00000081021HM8 2247010 00000081022HM6 2247010 00000082089HM4 2247010 00000082152HM0 2247010 00000082240HM3 2247010 00000082241HM1 2247010 00000082504HM2 2247010 00000082865HM7 2247010 00000082866HM5 2247010 00000083193HM3 2247010 00000083326HM9 2247010 00000083327HM7 2247010 00000083726HM0 2247010 00000083727HM8 2247010 00000083728HM6 2247010 00000084588HM3 2247010 00000084589HM1 2247010 00000085155HM0 2247010 00000085156HM8 2247010 00000085665HM8 2247010 00000085666HM6 2247010 00000085934HM8 2247010 00000086050HM2 2247010 00000086051HM0 2247010 00000086052HM8 2247010 00000086053HM6 2247010 00000086581HM6 2247010 00000086947HM9 2247010 00000086948HM7 2247010 00000087165HM7 2247010 00000087166HM5 2247010 00000087430HM5 2247010 00000087431HM3 2247010 00000087432HM1 2247010 00000087486HM7 2247010 00000087487HM5 2247010 00000087655HM7 2247010 00000087656HM5 2247010 00000087657HM3 2247010 00000087658HM1 2247010 00000089592HM0 2247010 00000089593HM8 2247010 00000090030HM8 2247010 00000090031HM6 2247010 00000090032HM4 2247010 00000090901HM0 2247010 00000090902HM8 2247010 00000091138HM8 2247010 00000091139HM6 2247010 00000091140HM4 2247010 00000091141HM2 2247010 00000091737HM7 2247010 00000091738HM5 2247010 00000091739HM3 2247010 00000092006HM6 2247010 00000092213HM8 2247010 00000092214HM6 2247010 00000092607HM1 2247010 00000092608HM9 2247010 00000092848HM1 2247010 00000092849HM9 2247010 00000093289HM7 2247010 00000093290HM5 2247010 00000093388HM7 2247010 00000093389HM5 2247010 00000093907HM4 2247010 00000093908HM2 2247010 00000093909HM0 2247010 00000093910HM8 2247010 00000094822HM4 2247010 00000095404HM0 2247010 00000095405HM7 2247010 00000095406HM5 2247010 00000095407HM3 2247010 00000096458HM5 2247010 00000096478HM3 2247010 00000096776HM0 2247010 00000096973HM3 2247010 00000096974HM1 2247010 00000097013HM7 2247010 00000097014HM5 2247010 00000098044HM1 2247010 00000098045HM8 2247010 00000098046HM6 2247010 00000098570HM5 2247010 00000098571HM3 2247010 00000098572HM1 2247010 00000098573HM9 2247010 00000099009HM3 2247010 00000099010HM1 2247010 00000100580HM0 2247010 00000102088HM2 2247010 00000102161HM7 2247010 00000102180HM7 2247010 00000102649HM1 2247010 00000102669HM9 2247010 00000103794HM4 2247010 00000103795HM1 2247010 00000104481HM7 2247010 00000104994HM9 2247010 00000105884HM1 2247010 00000106221HM5 2247010 00000107455HM8 2247010 00000107474HM9 2247010 00000107953HM2 2247010 00000107954HM0 2247010 00000108254HM4 2247010 00000108274HM2 2247010 00000108737HM8 2247010 00000108738HM6 2247010 00000109247HM7 2247010 00000109923HM3 2247010 00000109924HM1 2247010 00000110294HM6 2247010 00000110295HM3 2247010 00000110634HM3 2247010 00000110635HM0 2247010 00000110909HM9 2247010 00000111383HM6 2247010 00000111384HM4 2247010 00000112142HM5 2247010 00000112143HM3 2247010 00000112676HM2 2247010 00000112677HM0 2247010 00000114885HM7 2247010 00000115480HM6 2247010 00000115921HM9 2247010 00000116275HM9 2247010 00000116656HM0 2247010 00000116948HM1 2247010 00000116949HM9 2247010 00000117826HM8 2247010 00000117827HM6 2247010 00000117828HM4 2247010 00000118540HM4 2247010 00000119848HM0 2247010 00000119849HM8 2247010 00000119961HM1 2247010 00000119962HM9 2247010 00000120885HM9 2247010 00000120906HM3 2247010 00000120907HM1 2247010 00000121659HM7 2247010 00000121902HM1 2247010 00000122291HM8 2247010 00000122292HM6 2247010 00000122587HM9 2247010 00000123464HM0 2247010 00000123465HM7 2247010 00000123865HM8 2247010 00000124690HM9 2247010 00000124691HM7 2247010 00000125031HM5 2247010 00000125032HM3 2247010 00000125689HM0 2247010 00000125690HM8 2247010 00000126010HM8 2247010 00000126011HM6 2247010 00000126012HM4 2247010 00000126584HM2 2247010 00000127857HM1 2247010 00000127861HM3 2247010 00000127914HM0 2247010 00000128179HM9 2247010 00000128957HM8 2247010 00000129076HM6 2247010 00000129244HM0 2247010 00000129329HM9 2247010 00000129569HM0 2247010 00000129888HM4 2247010 00000130478HM1 2247010 00000130560HM6 2247010 00000130561HM4 2247010 00000130966HM5 2247010 00000131134HM9 2247010 00000131135HM6 2247010 00000131472HM3 2247010 00000131923HM5 2247010 00000131924HM3 2247010 00000131925HM0 2247010 00000131926HM8 2247010 00000132484HM7 2247010 00000132485HM4 2247010 00000133159HM4 2247010 00000134140HM3 2247010 00000134141HM1 2247010 00000134879HM6 2247010 00000134880HM4 2247010 00000135430HM7 2247010 00000135431HM5 2247010 00000136343HM1 2247010 00000136344HM9 2247010 00000136748HM1 2247010 00000136749HM9 2247010 00000137512HM0 2247010 00000766756YM1 2247010 00000766758YM7 2247010 00000766760YM3 2247010 00000766761YM1 2247010 00000766762YM9 2247010 00000766764YM5 2247010 00000766765YM2 2247010 00000766766YM0 2247010 00000766767YM8 2247010 00000766771YM0 2247010 00000766771YMO 2247010 00000766772YM8 2247010 00000766773YM6 2247010 00000766774YM4 2247010 00000766775YM1 2247010 00000766776YM9 2247010 00000766777YM7 2247010 00000766779YM3 2247010 00000766780YM1 2247010 00000766781YM9 2247010 00000766783YM5 2247010 00000766784YM3 2247010 00000766785YM0 2247010 00000766786YM8 2247010 00000766787YM6 2247010 00000766788YM4 2247010 00000766791YM8 2247010 00000766792YM6 2247010 00000766794YM2 2247010 00000766795YM9 2247010 00000766796YM7 2247010 00000766797YM5 2247010 00000766798YM3 2247010 00000766799YM1 2247010 00000766800YM7 2247010 00000766801YM5 2247010 00000766803YM1 2247010 00000766804YM9 2247010 00000766805YM6 2247010 00000766806YM4 2247010 00000766807YM2 2247010 00000766808YM0 2247010 00000766809YM8 2247010 00000766810YM6 2247010 00000766811YM4 2247010 00000766812YM2 2247010 00000766813YM0 2247010 00000766814YM8 2247010 00000766815YM5 2247010 00000766816YM3 2247010 00000766817YM1 2247010 00000766818YM9 2247010 00000766819YM7 2247010 00000766820YM5 2247010 00000766823YM9 2247010 00000766824YM7 2247010 00000766825YM4 2247010 00000766826YM2 2247010 00000766828YM8 2247010 00000766829YM6 2247010 00000766830YM4 2247010 00000766831YM2 2247010 00000766832YM0 2247010 00000766833YM8 2247010 00000766834YM6 2247010 00000766835YM3 2247010 00000766836YM1 2247010 00000766837YM9 2247010 00000766838YM7 2247010 00000766839YM5 2247010 00000766840YM3 2247010 00000766841YM1 2247010 00000766842YM9 2247010 00000766843YM7 2247010 00000766845YM2 2247010 00000766846YM0 2247010 00000766847YM8 2247010 00000766848YM6 2247010 00000766849YM4 2247010 00000766850YM2 2247010 00000766852YM8 2247010 00000766853YM6 2247010 00000766854YM4 2247010 00000766855YM1 2247010 00000766856YM9 2247010 00000766859YM3 2247010 00000766861YM9 2247010 00000766862YM7 2247010 00000766863YM5 2247010 00000766864YM3 2247010 00000766865YM0 2247010 00000766866YM8 2247010 00000766867YM6 2247010 00000766869YM2 2247010 00000766870YM0 2247010 00000766871YM8 2247010 00000766872YM6 2247010 00000766873YM4 2247010 00000766874YM2 2247010 00000766875YM9 2247010 00000766876YM7 2247010 00000766878YM3 2247010 00000766879YM1 2247010 00000766881YM7 2247010 00000766882YM5 2247010 00000766883YM3 2247010 00000766884YM1 2247010 00000766885YM8 2247010 00000766889YM0 2247010 00000766891YM6 2247010 00000766892YM4 2247010 00000766893YM2 2247010 00000766894YM0 2247010 00000766895YM7 2247010 00000766897YM3 2247010 00000766898YM1 2247010 00000766899YM9 2247010 00000766900YM5 2247010 00000766901YM3 2247010 00000766902YM1 2247010 00000766903YM9 2247010 00000766904YM7 2247010 00000766905YM4 2247010 00000766906YM2 2247010 00000766907YM0 2247010 00000766908YM8 2247010 00000766910YM4 2247010 00000766911YM2 2247010 00000766913YM8 2247010 00000766914YM6 2247010 00000766915YM3 2247010 00000766917YM9 2247010 00000766918YM7 2247010 00000766919YM5 2247010 00000766920YM3 2247010 00000766923YM7 2247010 00000766924YM5 2247010 00000766925YM2 2247010 00000766926YM0 2247010 00000766927YM8 2247010 00000766928YM6 2247010 00000766929YM4 2247010 00000766930YM2 2247010 00000766932YM8 2247010 00000766933YM6 2247010 00000766934YM4 2247010 00000766935YM1 2247010 00000766936YM9 2247010 00000766937YM7 2247010 00000766938YM5 2247010 00000766939YM3 2247010 00000766941YM9 2247010 00000766942YM7 2247010 00000766943YM5 2247010 00000766944YM3 2247010 00000766945YM0 2247010 00000766946YM8 2247010 00000766949YM2 2247010 00000766950YM0 2247010 00000766952YM6 2247010 00000766953YM4 2247010 00000847627YM7 2247010 00000847629YM3 2247010 00000847630YM1 2247010 00000847631YM9 2247010 00000847632YM7 2247010 00000847633YM5 2247010 00000847635YM0 2247010 00000847636YM8 2247010 00000847637YM6 2247010 00000847638YM4 2247010 00000847639YM2 2247010 00000847640YM0 2247010 00000847642YM6 2247010 00000847643YM4 2247010 00000847644YM2 2247010 00000847645YM9 2247010 00000847646YM7 2247010 00000847647YM5 2247010 00000847648YM3 2247010 00000847649YM1 2247010 00000847650YM9 2247010 00000847651YM7 2247010 00000847654YM1 2247010 00000847655YM8 2247010 00000847656YM6 2247010 00000864847YM9 2247010 00000864848YM5 2247010 00000864848YM7 2247010 00000864849YM5 2247010 00000864851YM1 2247010 00000864852YM9 2247010 00000864855YM2 2247010 00000864856YM0 2247010 00000864857YM8 2247010 00000864858YM6 2247010 00000864859YM4 2247010 00000864860YM2 2247010 00000864861YM0 2247010 00000864862YM8 2247010 00000864863YM6 2247010 00000864864YM4 2247010 00000864865YM1 2247010 00000864866YM9 2247010 00000864867YM7 2247010 00000864868YM5 2247010 00000864870YM1 2247010 00000864871YM9 2247010 00000864872YM7 2247010 00000864873YM5 2247010 00000864874YM3 2247010 00000864875YM0 2247010 00000864876YM8 2247010 00000873807YM2 2247010 00000873809YM8 2247010 00000873810YM6 2247010 00000873811YM4 2247010 00000873813YM0 2247010 00000873815YM5 2247010 00000873816YM3 2247010 00000873817YM1 2247010 00000873818YM9 2247010 00000873819YM7 2247010 00000873820YM5 2247010 00000873821YM3 2247010 00000873822YM1 2247010 00000873823YM9 2247010 00000873824YM7 2247010 00000873825YM4 2247010 00000873826YM2 2247010 00000873829YM6 2247010 00000873830YM4 2247010 00000873831YM2 2247010 00000873832YM0 2247010 00000873833YM8 2247010 00000873834YM6 2247010 00000873835YM3 2247010 00000873836YM1 2247010 00000884598YM4 2247010 00000884599YM2 2247010 00000884600YM8 2247010 00000884601YM6 2247010 00000884602YM4 2247010 00000884603YM2 2247010 00000884604YM0 2247010 00000884605YM7 2247010 00000884606YM5 2247010 00000884607YM3 2247010 00000884608YM1 2247010 00000884609YM9 2247010 00000884610YM7 2247010 00000884611YM5 2247010 00000884612YM3 2247010 00000884613YM1 2247010 00000884614YM9 2247010 00000884615YM6 2247010 00000884616YM4 2247010 00000884617YM2 2247010 00000884618YM0 2247010 00000884621YM4 2247010 00000884622YM2 2247010 00000884623YM0 2247010 00000884624YM8 2247010 00000884625YM5 2247010 00000884626YM3 2247010 00000884627YM1 2247010 00000898208YM4 2247010 00000898209YM2 2247010 00000898210YM0 2247010 00000898211YM8 2247010 00000898212YM6 2247010 00000898213YM4 2247010 00000898214YM2 2247010 00000898215YM9 2247010 00000898216YM7 2247010 00000898217YM5 2247010 00000898218YM3 2247010 00000898219YM1 2247010 00000898220YM9 2247010 00000898221YM7 2247010 00000898222YM5 2247010 00000898223YM3 2247010 00000898224YM1 2247010 00000898225YM8 2247010 00000898226YM6 2247010 00000898227YM4 2247010 00000898228YM2 2247010 00000898229YM0 2247010 00000898231YM6 2247010 00000898232YM4 2247010 00000898233YM2 2247010 00000898234YM0 2247010 00000898235YM7 2247010 00000898236YM5 2247010 00000898237YM3 2247010 00000898238YM1 2247010 00000898239YM9 2247010 00000898240YM7 2247010 00000898241YM5 2247010 00000898242YM3 2247010 00000898244YM9 2247010 00000898245YM6 2247010 00000898247YM2 2247010 00000914838YM8 2247010 00000914839YM6 2247010 00000914840YM4 2247010 00000914841YM2 2247010 00000914842YM0 2247010 00000914843YM8 2247010 00000914844YM6 2247010 00000914845YM3 2247010 00000914846YM1 2247010 00000914847YM9 2247010 00000914848YM7 2247010 00000914849YM5 2247010 00000914850YM3 2247010 00000914851YM1 2247010 00000914852YM9 2247010 00000914854YM5 2247010 00000914855YM2 2247010 00000914856YM0 2247010 00000914857YM8 2247010 00000914858YM6 2247010 00000914859YM4 2247010 00000914860YM2 2247010 00000914861YM0 2247010 00000914862YM8 2247010 00000914863YM6 2247010 00000914864YM4 2247010 00000914865YM1 2247010 00000914866YM9 2247010 00000914867YM7 2247010 00000935017YM4 2247010 00000935019YM0 2247010 00000935020YM8 2247010 00000935021YM6 2247010 00000935022YM4 2247010 00000935023YM2 2247010 00000935024YM0 2247010 00000935025YM7 2247010 00000935026YM5 2247010 00000935027YM3 2247010 00000935028YM1 2247010 00000935029YM9 2247010 00000935030YM7 2247010 00000935031YM5 2247010 00000935032YM3 2247010 00000935033YM1 2247010 00000935035YM6 2247010 00000935037YM2 2247010 00000935038YM0 2247010 0000873828YM8A 2247799 00000696478YM7 2247799 00000696479YM5 2247799 00000696480YM3 2247799 00000748118YM7 2247799 00000748121YM1 2247799 00000748122YM9 2247799 00000748123YM7 2247799 00000748124YM5 2247799 00000748126YM0 2247799 00000748127YM8 2247799 00000748128YM6 2247799 00000748130YM2 2247799 00000748132YM8 2247799 00000748133YM6 2247799 00000748137YM7 2247799 00000748138YM5 2247799 00000748139YM3 2247799 00000748140YM1 2247799 00000748141YM9 2247799 00000748142YM7 2247799 00000748143YM5 2247799 00000748145YM0 2247799 00000748148YM4 2247799 00000748149YM2 2247799 00000748151YM8 2247799 00000748152YM6 2249696 00000682378YM5 2249696 00000682379YM3 2249696 00000682380YM1 2249696 00000715318YM2 2249696 00000715319YM0 2249696 00000715320YM8 2249696 00000715321YM6 2249696 00000715322YM4 2249696 00000715323YM2 2249696 00000715325YM7 2307672 00000830376YM0 2307672 00000830377YM8 2307672 00000830378YM6 2307672 00000830379YM4 2307672 00000830381YM0 2320314 00000036479HM4 2320314 00000042493HM7 2320314 00000047895HM8 2320314 00000051339HM0 2320314 00000053647HM4 2320314 00000054197HM9 2320314 00000057110HM9 2320314 00000059590HM0 2320314 00000065801HM3 2320314 00000068669HM1 2320314 00000084484HM5 2320314 00000101233HM5 2320314 00000118046HM2 2320314 00000131327HM9 2320314 00000132452HM4 2320314 00000134516HM4 2320314 00000033983HM8 2320314 00000033984HM6 2320314 00000034202HM2 2320314 00000034327HM7 2320314 00000034328HM5 2320314 00000034329HM3 2320314 00000034330HM1 2320314 00000035650HM1 2320314 00000035651HM9 2320314 00000035652HM7 2320314 00000035672HM5 2320314 00000036480HM2 2320314 00000036481HM0 2320314 00000037142HM7 2320314 00000037158HM3 2320314 00000037159HM1 2320314 00000037173HM2 2320314 00000037329HM0 2320314 00000037332HM4 2320314 00000037333HM2 2320314 00000037373HM8 2320314 00000037375HM3 2320314 00000037658HM2 2320314 00000037659HM0 2320314 00000037660HM8 2320314 00000037661HM6 2320314 00000037726HM7 2320314 00000037727HM5 2320314 00000037728HM3 2320314 00000037987HM5 2320314 00000037988HM3 2320314 00000037989HM1 2320314 00000037990HM9 2320314 00000037991HM7 2320314 00000037992HM5 2320314 00000037993HM3 2320314 00000038198HM8 2320314 00000038199HM6 2320314 00000038200HM2 2320314 00000038201HM0 2320314 00000038202HM8 2320314 00000038203HM6 2320314 00000039143HM3 2320314 00000039569HM9 2320314 00000039570HM7 2320314 00000039852HM9 2320314 00000040676HM9 2320314 00000040677HM7 2320314 00000040732HM0 2320314 00000040733HM8 2320314 00000041402HM9 2320314 00000041403HM7 2320314 00000041840HM0 2320314 00000042367HM3 2320314 00000042368HM1 2320314 00000042369HM9 2320314 00000042491HM1 2320314 00000042492HM9 2320314 00000042708HM8 2320314 00000042709HM6 2320314 00000042710HM4 2320314 00000042711HM2 2320314 00000042960HM5 2320314 00000042961HM3 2320314 00000042962HM1 2320314 00000042964HM7 2320314 00000043627HM9 2320314 00000043628HM7 2320314 00000043629HM5 2320314 00000043630HM3 2320314 00000043763HM2 2320314 00000044075HM0 2320314 00000044076HM8 2320314 00000044077HM6 2320314 00000044078HM4 2320314 00000044512HM2 2320314 00000044513HM0 2320314 00000044514HM8 2320314 00000044515HM5 2320314 00000044516HM3 2320314 00000044608HM8 2320314 00000044619HM5 2320314 00000044620HM3 2320314 00000044621HM1 2320314 00000044622HM9 2320314 00000044848HM0 2320314 00000044849HM8 2320314 00000044850HM6 2320314 00000045193HM0 2320314 00000045194HM8 2320314 00000045210HM2 2320314 00000045211HM0 2320314 00000045212HM8 2320314 00000045478HM5 2320314 00000045597HM2 2320314 00000045598HM0 2320314 00000045616HM0 2320314 00000046089HM9 2320314 00000046090HM7 2320314 00000046092HM3 2320314 00000046118HM6 2320314 00000046128HM5 2320314 00000046129HM3 2320314 00000046660HM7 2320314 00000046668HM0 2320314 00000047187HM0 2320314 00000047221HM7 2320314 00000047229HM0 2320314 00000047615HM0 2320314 00000047616HM8 2320314 00000047617HM6 2320314 00000047637HM4 2320314 00000047829HM7 2320314 00000047838HM8 2320314 00000047839HM6 2320314 00000048134HM1 2320314 00000048135HM8 2320314 00000048136HM6 2320314 00000048137HM4 2320314 00000048169HM7 2320314 00000048177HM0 2320314 00000048185HM3 2320314 00000048566HM4 2320314 00000048810HM6 2320314 00000049126HM6 2320314 00000049127HM4 2320314 00000049128HM2 2320314 00000049149HM8 2320314 00000049150HM6 2320314 00000049164HM7 2320314 00000049216HM5 2320314 00000049217HM3 2320314 00000049218HM1 2320314 00000049583HM8 2320314 00000049599HM4 2320314 00000049600HM0 2320314 00000049601HM8 2320314 00000049622HM4 2320314 00000049623HM2 2320314 00000050023HM1 2320314 00000050040HM5 2320314 00000050041HM3 2320314 00000050042HM1 2320314 00000050043HM9 2320314 00000050070HM2 2320314 00000050071HM0 2320314 00000050106HM4 2320314 00000050107HM2 2320314 00000050675HM8 2320314 00000050676HM6 2320314 00000050677HM4 2320314 00000050678HM2 2320314 00000050933HM1 2320314 00000050941HM4 2320314 00000051337HM4 2320314 00000051338HM2 2320314 00000051849HM8 2320314 00000052054HM4 2320314 00000052072HM6 2320314 00000052080HM9 2320314 00000052144HM3 2320314 00000052580HM8 2320314 00000052852HM1 2320314 00000052853HM9 2320314 00000052867HM9 2320314 00000052961HM0 2320314 00000052972HM7 2320314 00000052973HM5 2320314 00000052974HM3 2320314 00000052975HM0 2320314 00000053001HM4 2320314 00000053379HM4 2320314 00000053380HM2 2320314 00000053394HM3 2320314 00000053632HM6 2320314 00000053633HM4 2320314 00000054036HM9 2320314 00000054037HM7 2320314 00000054038HM5 2320314 00000054183HM9 2320314 00000054489HM0 2320314 00000054490HM8 2320314 00000054660HM6 2320314 00000054661HM4 2320314 00000054662HM2 2320314 00000054663HM0 2320314 00000054689HM5 2320314 00000054871HM9 2320314 00000054872HM7 2320314 00000054889HM4 2320314 00000055624HM1 2320314 00000055719HM9 2320314 00000055727HM2 2320314 00000055898HM1 2320314 00000055906HM2 2320314 00000056523HM4 2320314 00000056524HM2 2320314 00000056622HM4 2320314 00000056631HM5 2320314 00000056632HM3 2320314 00000056782HM6 2320314 00000056783HM4 2320314 00000056799HM0 2320314 00000057118HM2 2320314 00000057286HM7 2320314 00000057288HM3 2320314 00000057653HM8 2320314 00000057667HM8 2320314 00000057677HM7 2320314 00000057971HM4 2320314 00000057980HM5 2320314 00000057981HM3 2320314 00000058203HM1 2320314 00000058212HM2 2320314 00000058213HM0 2320314 00000058314HM6 2320314 00000058315HM3 2320314 00000058388HM0 2320314 00000058389HM8 2320314 00000058404HM5 2320314 00000058405HM2 2320314 00000058589HM3 2320314 00000058590HM1 2320314 00000058785HM7 2320314 00000058786HM5 2320314 00000058938HM2 2320314 00000058939HM0 2320314 00000058940HM8 2320314 00000059039HM8 2320314 00000059040HM6 2320314 00000059060HM4 2320314 00000059333HM5 2320314 00000059334HM3 2320314 00000059348HM3 2320314 00000059589HM2 2320314 00000059605HM6 2320314 00000059606HM4 2320314 00000059811HM0 2320314 00000059818HM5 2320314 00000060248HM2 2320314 00000060249HM0 2320314 00000060264HM9 2320314 00000060265HM6 2320314 00000060300HM1 2320314 00000060301HM9 2320314 00000060455HM3 2320314 00000060456HM1 2320314 00000060470HM2 2320314 00000060920HM6 2320314 00000060921HM4 2320314 00000060938HM8 2320314 00000060939HM6 2320314 00000061219HM2 2320314 00000061258HM0 2320314 00000061259HM8 2320314 00000061454HM5 2320314 00000061686HM2 2320314 00000061687HM0 2320314 00000061974HM2 2320314 00000062648HM1 2320314 00000062656HM4 2320314 00000062664HM8 2320314 00000062848HM7 2320314 00000062849HM5 2320314 00000063187HM9 2320314 00000063195HM2 2320314 00000063203HM4 2320314 00000063270HM3 2320314 00000063497HM2 2320314 00000063498HM0 2320314 00000063514HM4 2320314 00000063552HM4 2320314 00000063783HM5 2320314 00000063784HM3 2320314 00000063798HM3 2320314 00000064056HM5 2320314 00000064064HM9 2320314 00000064269HM4 2320314 00000064277HM7 2320314 00000064285HM0 2320314 00000064345HM2 2320314 00000064346HM0 2320314 00000064360HM1 2320314 00000064682HM8 2320314 00000064683HM6 2320314 00000064697HM6 2320314 00000064705HM7 2320314 00000064913HM7 2320314 00000064914HM5 2320314 00000065145HM5 2320314 00000065146HM3 2320314 00000065147HM1 2320314 00000065348HM5 2320314 00000065349HM3 2320314 00000065350HM1 2320314 00000065648HM8 2320314 00000065649HM6 2320314 00000065650HM4 2320314 00000065651HM2 2320314 00000065802HM1 2320314 00000065876HM5 2320314 00000066250HM2 2320314 00000066251HM0 2320314 00000066253HM6 2320314 00000066293HM2 2320314 00000066294HM0 2320314 00000066295HM7 2320314 00000066979HM6 2320314 00000067214HM7 2320314 00000067215HM4 2320314 00000067216HM2 2320314 00000067217HM0 2320314 00000067244HM4 2320314 00000067245HM1 2320314 00000067472HM1 2320314 00000068148HM6 2320314 00000068149HM4 2320314 00000068164HM3 2320314 00000068165HM0 2320314 00000068670HM9 2320314 00000068671HM7 2320314 00000068896HM0 2320314 00000068908HM3 2320314 00000068909HM1 2320314 00000068929HM9 2320314 00000069170HM9 2320314 00000069171HM7 2320314 00000069172HM5 2320314 00000069376HM2 2320314 00000069377HM0 2320314 00000069378HM8 2320314 00000069804HM3 2320314 00000069805HM0 2320314 00000069825HM8 2320314 00000070001HM3 2320314 00000070002HM1 2320314 00000070221HM7 2320314 00000070222HM5 2320314 00000070455HM1 2320314 00000070456HM9 2320314 00000070641HM6 2320314 00000070826HM3 2320314 00000070838HM8 2320314 00000070839HM6 2320314 00000070859HM4 2320314 00000071128HM3 2320314 00000071129HM1 2320314 00000071130HM9 2320314 00000071373HM5 2320314 00000071492HM3 2320314 00000071493HM1 2320314 00000072025HM0 2320314 00000072026HM8 2320314 00000072027HM6 2320314 00000072125HM8 2320314 00000072142HM3 2320314 00000072152HM2 2320314 00000072155HM5 2320314 00000072156HM3 2320314 00000072157HM1 2320314 00000072632HM3 2320314 00000072633HM1 2320314 00000072838HM6 2320314 00000072856HM8 2320314 00000072857HM6 2320314 00000072909HM5 2320314 00000073212HM3 2320314 00000073230HM5 2320314 00000073231HM3 2320314 00000073252HM9 2320314 00000073253HM7 2320314 00000073516HM7 2320314 00000073517HM5 2320314 00000073959HM9 2320314 00000073960HM7 2320314 00000073961HM5 2320314 00000073962HM3 2320314 00000074316HM1 2320314 00000074317HM9 2320314 00000074318HM7 2320314 00000074319HM5 2320314 00000074596HM8 2320314 00000074597HM6 2320314 00000074598HM4 2320314 00000074599HM2 2320314 00000075049HM7 2320314 00000075062HM0 2320314 00000075063HM8 2320314 00000075064HM6 2320314 00000075999HM3 2320314 00000076527HM1 2320314 00000076587HM5 2320314 00000076637HM8 2320314 00000076638HM6 2320314 00000076658HM4 2320314 00000077229HM3 2320314 00000077231HM9 2320314 00000077442HM2 2320314 00000077443HM0 2320314 00000077444HM8 2320314 00000077644HM3 2320314 00000077645HM0 2320314 00000077874HM6 2320314 00000077875HM3 2320314 00000077942HM1 2320314 00000077953HM8 2320314 00000078259HM9 2320314 00000078271HM4 2320314 00000078272HM2 2320314 00000078610HM3 2320314 00000078949HM5 2320314 00000078950HM3 2320314 00000078951HM1 2320314 00000079450HM3 2320314 00000079451HM1 2320314 00000079519HM5 2320314 00000079991HM6 2320314 00000080120HM9 2320314 00000080121HM7 2320314 00000080461HM7 2320314 00000080462HM5 2320314 00000080482HM3 2320314 00000081464HM0 2320314 00000081465HM7 2320314 00000081466HM5 2320314 00000081860HM9 2320314 00000082174HM4 2320314 00000082175HM1 2320314 00000082196HM7 2320314 00000082197HM5 2320314 00000082218HM9 2320314 00000082219HM7 2320314 00000082311HM2 2320314 00000082312HM0 2320314 00000082714HM7 2320314 00000082936HM6 2320314 00000083214HM7 2320314 00000083234HM5 2320314 00000083449HM9 2320314 00000083460HM6 2320314 00000083606HM4 2320314 00000083607HM2 2320314 00000083781HM5 2320314 00000083938HM1 2320314 00000083950HM6 2320314 00000083951HM4 2320314 00000084160HM1 2320314 00000084161HM9 2320314 00000084191HM6 2320314 00000084430HM8 2320314 00000084463HM9 2320314 00000084464HM7 2320314 00000084760HM8 2320314 00000084761HM6 2320314 00000084811HM9 2320314 00000084903HM4 2320314 00000084914HM1 2320314 00000084931HM5 2320314 00000085017HM2 2320314 00000085086HM7 2320314 00000085316HM8 2320314 00000085327HM5 2320314 00000085338HM2 2320314 00000085777HM1 2320314 00000085778HM9 2320314 00000086133HM6 2320314 00000086151HM8 2320314 00000086152HM6 2320314 00000086460HM3 2320314 00000086461HM1 2320314 00000086481HM9 2320314 00000086499HM1 2320314 00000086500HM6 2320314 00000087243HM2 2320314 00000087244HM0 2320314 00000087729HM0 2320314 00000087730HM8 2320314 00000088573HM1 2320314 00000088574HM9 2320314 00000088594HM7 2320314 00000088667HM1 2320314 00000088668HM9 2320314 00000088688HM7 2320314 00000088755HM4 2320314 00000088756HM2 2320314 00000088776HM0 2320314 00000089002HM0 2320314 00000089003HM8 2320314 00000089023HM6 2320314 00000089987HM2 2320314 00000089988HM0 2320314 00000089989HM8 2320314 00000089990HM6 2320314 00000090446HM6 2320314 00000090608HM1 2320314 00000090609HM9 2320314 00000090750HM1 2320314 00000090751HM9 2320314 00000090752HM7 2320314 00000091261HM8 2320314 00000091448HM1 2320314 00000091449HM9 2320314 00000091663HM5 2320314 00000091664HM3 2320314 00000091684HM1 2320314 00000091888HM8 2320314 00000092289HM8 2320314 00000092347HM4 2320314 00000092348HM2 2320314 00000092644HM4 2320314 00000092686HM5 2320314 00000093105HM5 2320314 00000093245HM9 2320314 00000093262HM4 2320314 00000093326HM7 2320314 00000093833HM2 2320314 00000093834HM0 2320314 00000093835HM7 2320314 00000094729HM1 2320314 00000094741HM6 2320314 00000094742HM4 2320314 00000094876HM0 2320314 00000095499HM0 2320314 00000095500HM5 2320314 00000095501HM3 2320314 00000095549HM2 2320314 00000095550HM0 2320314 00000095572HM4 2320314 00000096485HM8 2320314 00000096503HM8 2320314 00000096504HM6 2320314 00000096670HM5 2320314 00000096671HM3 2320314 00000096672HM1 2320314 00000096673HM9 2320314 00000096714HM1 2320314 00000096715HM8 2320314 00000096717HM4 2320314 00000096755HM4 2320314 00000099534HM0 2320314 00000099535HM7 2320314 00000099536HM5 2320314 00000099959HM9 2320314 00000099960HM7 2320314 00000099994HM6 2320314 00000099995HM3 2320314 00000099996HM1 2320314 00000100303HM7 2320314 00000100304HM5 2320314 00000100324HM3 2320314 00000100966HM1 2320314 00000101910HM8 2320314 00000101990HM0 2320314 00000102224HM3 2320314 00000102938HM8 2320314 00000103143HM4 2320314 00000103427HM1 2320314 00000103696HM1 2320314 00000104145HM8 2320314 00000104305HM8 2320314 00000104726HM5 2320314 00000105184HM6 2320314 00000105320HM6 2320314 00000105581HM3 2320314 00000105582HM1 2320314 00000105583HM9 2320314 00000105935HM1 2320314 00000106435HM1 2320314 00000107198HM4 2320314 00000107199HM2 2320314 00000107584HM5 2320314 00000107585HM2 2320314 00000108056HM3 2320314 00000108057HM1 2320314 00000108545HM5 2320314 00000108546HM3 2320314 00000108652HM9 2320314 00000109096HM8 2320314 00000110714HM3 2320314 00000110715HM0 2320314 00000110716HM8 2320314 00000111460HM2 2320314 00000111926HM2 2320314 00000112568HM1 2320314 00000113451HM9 2320314 00000113504HM5 2320314 00000113505HM2 2320314 00000113506HM0 2320314 00000113892HM4 2320314 00000113924HM5 2320314 00000113925HM2 2320314 00000114661HM2 2320314 00000114662HM0 2320314 00000115083HM8 2320314 00000115084HM6 2320314 00000115085HM3 2320314 00000115399HM8 2320314 00000115400HM4 2320314 00000115401HM2 2320314 00000115880HM7 2320314 00000115881HM5 2320314 00000115882HM3 2320314 00000116324HM5 2320314 00000116852HM5 2320314 00000117410HM1 2320314 00000117411HM9 2320314 00000117412HM7 2320314 00000118591HM7 2320314 00000118592HM5 2320314 00000120179HM7 2320314 00000120180HM5 2320314 00000120492HM4 2320314 00000120493HM2 2320314 00000120494HM0 2320314 00000121255HM4 2320314 00000121256HM2 2320314 00000121257HM0 2320314 00000121524HM3 2320314 00000122065HM6 2320314 00000122066HM4 2320314 00000122636HM4 2320314 00000122637HM2 2320314 00000123079HM6 2320314 00000123080HM4 2320314 00000123614HM0 2320314 00000123615HM7 2320314 00000125003HM4 2320314 00000125004HM2 2320314 00000125530HM6 2320314 00000125531HM4 2320314 00000125532HM2 2320314 00000127555HM1 2320314 00000127556HM9 2320314 00000127883HM7 2320314 00000127884HM5 2320314 00000127885HM2 2320314 00000128548HM5 2320314 00000128549HM3 2320314 00000128641HM8 2320314 00000128642HM6 2320314 00000128643HM4 2320314 00000129095HM6 2320314 00000129096HM4 2320314 00000129298HM6 2320314 00000129299HM4 2320314 00000130215HM7 2320314 00000130216HM5 2320314 00000130516HM8 2320314 00000130517HM6 2320314 00000131192HM7 2320314 00000131193HM5 2320314 00000131328HM7 2320314 00000131329HM5 2320314 00000131330HM3 2320314 00000132450HM8 2320314 00000132451HM6 2320314 00000133074HM5 2320314 00000133075HM2 2320314 00000133076HM0 2320314 00000133524HM9 2320314 00000133525HM6 2320314 00000133526HM4 2320314 00000134196HM5 2320314 00000134197HM3 2320314 00000134513HM1 2320314 00000134514HM9 2320314 00000134515HM6 2320314 00000134917HM4 2320314 00000134918HM2 2320314 00000134919HM0 2320314 00000134920HM8 2320314 00000135224HM4 2320314 00000135608HM8 2320314 00000135609HM6 2369660 00000073480HM6 2369660 00000077855HM5 2369660 00000081402HM0 2369660 00000083972HM0 2369660 00000102631HM9 2369660 00000106775HM0 2369660 00000107853HM4 2369660 00000116865HM7 2369660 00000134856HM4 2369660 00000140604HM0 2369660 00000142090HM0 2369660 00000017441HM5 2369660 00000059781HM5 2369660 00000060315HM9 2369660 00000060384HM5 2369660 00000064046HM6 2369660 00000065370HM9 2369660 00000065887HM2 2369660 00000066175HM1 2369660 00000066176HM9 2369660 00000072879HM0 2369660 00000072880HM8 2369660 00000072881HM6 2369660 00000073181HM0 2369660 00000073182HM8 2369660 00000073183HM6 2369660 00000073478HM0 2369660 00000073479HM8 2369660 00000073481HM4 2369660 00000074275HM9 2369660 00000074277HM5 2369660 00000074278HM3 2369660 00000074554HM7 2369660 00000074555HM4 2369660 00000074556HM2 2369660 00000074557HM0 2369660 00000074764HM2 2369660 00000074765HM9 2369660 00000075116HM4 2369660 00000075117HM2 2369660 00000075118HM0 2369660 00000075119HM8 2369660 00000075444HM0 2369660 00000075445HM7 2369660 00000075446HM5 2369660 00000075477HM0 2369660 00000075478HM8 2369660 00000075479HM6 2369660 00000075813HM6 2369660 00000075814HM4 2369660 00000075815HM1 2369660 00000075816HM9 2369660 00000076499HM3 2369660 00000076500HM8 2369660 00000076501HM6 2369660 00000076909HM1 2369660 00000076920HM8 2369660 00000076921HM6 2369660 00000076940HM6 2369660 00000076941HM4 2369660 00000076959HM6 2369660 00000077052HM9 2369660 00000077062HM8 2369660 00000077260HM8 2369660 00000077271HM5 2369660 00000077272HM3 2369660 00000077473HM7 2369660 00000077615HM3 2369660 00000077616HM1 2369660 00000077617HM9 2369660 00000077834HM0 2369660 00000077835HM7 2369660 00000077854HM8 2369660 00000077993HM4 2369660 00000078003HM1 2369660 00000078305HM0 2369660 00000078306HM8 2369660 00000078621HM0 2369660 00000079414HM9 2369660 00000079415HM6 2369660 00000079616HM9 2369660 00000079617HM7 2369660 00000079635HM9 2369660 00000080736HM2 2369660 00000080778HM4 2369660 00000080788HM3 2369660 00000080878HM2 2369660 00000080879HM0 2369660 00000080897HM2 2369660 00000080912HM9 2369660 00000081307HM1 2369660 00000081317HM0 2369660 00000081430HM1 2369660 00000081440HM0 2369660 00000081452HM5 2369660 00000081690HM0 2369660 00000081691HM8 2369660 00000081709HM8 2369660 00000082059HM7 2369660 00000082070HM4 2369660 00000082071HM2 2369660 00000082532HM3 2369660 00000082533HM1 2369660 00000082725HM3 2369660 00000082736HM0 2369660 00000082737HM8 2369660 00000082830HM1 2369660 00000082831HM9 2369660 00000082958HM0 2369660 00000082959HM8 2369660 00000082960HM6 2369660 00000083246HM9 2369660 00000083247HM7 2369660 00000083429HM1 2369660 00000083439HM0 2369660 00000083577HM7 2369660 00000083578HM5 2369660 00000083579HM3 2369660 00000083845HM8 2369660 00000083861HM5 2369660 00000083862HM3 2369660 00000083973HM8 2369660 00000084001HM7 2369660 00000084202HM1 2369660 00000084213HM8 2369660 00000084214HM6 2369660 00000084326HM8 2369660 00000084327HM6 2369660 00000084498HM5 2369660 00000084516HM4 2369660 00000084824HM2 2369660 00000084825HM9 2369660 00000084826HM7 2369660 00000084988HM5 2369660 00000084989HM3 2369660 00000085007HM3 2369660 00000085097HM4 2369660 00000085244HM2 2369660 00000085245HM9 2369660 00000085246HM7 2369660 00000085350HM7 2369660 00000085351HM5 2369660 00000085369HM7 2369660 00000085464HM6 2369660 00000085465HM3 2369660 00000085466HM1 2369660 00000085493HM5 2369660 00000085494HM3 2369660 00000085512HM2 2369660 00000085761HM5 2369660 00000085799HM5 2369660 00000085800HM1 2369660 00000085819HM1 2369660 00000085820HM9 2369660 00000085886HM0 2369660 00000085887HM8 2369660 00000085888HM6 2369660 00000085889HM4 2369660 00000085890HM2 2369660 00000086123HM7 2369660 00000086173HM2 2369660 00000086174HM0 2369660 00000086193HM0 2369660 00000086194HM8 2369660 00000086734HM1 2369660 00000086750HM7 2369660 00000086751HM5 2369660 00000086771HM3 2369660 00000086772HM1 2369660 00000086773HM9 2369660 00000087091HM5 2369660 00000087101HM2 2369660 00000087119HM4 2369660 00000087120HM2 2369660 00000087121HM0 2369660 00000087122HM8 2369660 00000087205HM1 2369660 00000087206HM9 2369660 00000087207HM7 2369660 00000087208HM5 2369660 00000087544HM3 2369660 00000087545HM0 2369660 00000087546HM8 2369660 00000087547HM6 2369660 00000087753HM0 2369660 00000087754HM8 2369660 00000087755HM5 2369660 00000087756HM3 2369660 00000087946HM0 2369660 00000087956HM9 2369660 00000088154HM0 2369660 00000088155HM7 2369660 00000088156HM5 2369660 00000088184HM7 2369660 00000088185HM4 2369660 00000088186HM2 2369660 00000088251HM4 2369660 00000088613HM5 2369660 00000088614HM3 2369660 00000088615HM0 2369660 00000088700HM0 2369660 00000088701HM8 2369660 00000088719HM0 2369660 00000088794HM3 2369660 00000088795HM0 2369660 00000088813HM1 2369660 00000089100HM2 2369660 00000089250HM5 2369660 00000089251HM3 2369660 00000089252HM1 2369660 00000089280HM2 2369660 00000089281HM0 2369660 00000089282HM8 2369660 00000089517HM7 2369660 00000089518HM5 2369660 00000089536HM7 2369660 00000089554HM0 2369660 00000089555HM7 2369660 00000089556HM5 2369660 00000089557HM3 2369660 00000089648HM0 2369660 00000089876HM7 2369660 00000089877HM5 2369660 00000090299HM9 2369660 00000090300HM5 2369660 00000090329HM4 2369660 00000090330HM2 2369660 00000090709HM7 2369660 00000090710HM5 2369660 00000090729HM5 2369660 00000090730HM3 2369660 00000091213HM9 2369660 00000091232HM9 2369660 00000091233HM7 2369660 00000091470HM5 2369660 00000091471HM3 2369660 00000091634HM6 2369660 00000091635HM3 2369660 00000091636HM1 2369660 00000091696HM5 2369660 00000091697HM3 2369660 00000091725HM2 2369660 00000092307HM8 2369660 00000092308HM6 2369660 00000092336HM7 2369660 00000092478HM7 2369660 00000092479HM5 2369660 00000092480HM3 2369660 00000092661HM8 2369660 00000092671HM7 2369660 00000092698HM0 2369660 00000092699HM8 2369660 00000092923HM2 2369660 00000092924HM0 2369660 00000092925HM7 2369660 00000093273HM1 2369660 00000093339HM0 2369660 00000093340HM8 2369660 00000093368HM9 2369660 00000093369HM7 2369660 00000093864HM7 2369660 00000093875HM3 2369660 00000093876HM1 2369660 00000093894HM4 2369660 00000094764HM8 2369660 00000094765HM5 2369660 00000094766HM3 2369660 00000094793HM7 2369660 00000094794HM5 2369660 00000095308HM3 2369660 00000095820HM7 2369660 00000095821HM5 2369660 00000095822HM3 2369660 00000095848HM8 2369660 00000095859HM5 2369660 00000095860HM3 2369660 00000096175HM5 2369660 00000096176HM3 2369660 00000096205HM0 2369660 00000096206HM8 2369660 00000096224HM1 2369660 00000096247HM2 2369660 00000096248HM0 2369660 00000096249HM8 2369660 00000096766HM1 2369660 00000097252HM1 2369660 00000097253HM9 2369660 00000097254HM7 2369660 00000097281HM0 2369660 00000097282HM8 2369660 00000097623HM3 2369660 00000097624HM1 2369660 00000097625HM8 2369660 00000097626HM6 2369660 00000097792HM6 2369660 00000097793HM4 2369660 00000097794HM2 2369660 00000097821HM3 2369660 00000097822HM1 2369660 00000098103HM5 2369660 00000098104HM3 2369660 00000098105HM0 2369660 00000098131HM6 2369660 00000098141HM5 2369660 00000098152HM2 2369660 00000098153HM0 2369660 00000098658HM8 2369660 00000098929HM3 2369660 00000098930HM1 2369660 00000098931HM9 2369660 00000099323HM8 2369660 00000099324HM6 2369660 00000099325HM3 2369660 00000099326HM1 2369660 00000099362HM6 2369660 00000099363HM4 2369660 00000099364HM2 2369660 00000099392HM3 2369660 00000099393HM1 2369660 00000099394HM9 2369660 00000099807HM0 2369660 00000099808HM8 2369660 00000099809HM6 2369660 00000099810HM4 2369660 00000099811HM2 2369660 00000099812HM0 2369660 00000099813HM8 2369660 00000099939HM1 2369660 00000099940HM9 2369660 00000100203HM9 2369660 00000100219HM5 2369660 00000100220HM3 2369660 00000100461HM3 2369660 00000100462HM1 2369660 00000100480HM3 2369660 00000100495HM1 2369660 00000100496HM9 2369660 00000100497HM7 2369660 00000100498HM5 2369660 00000100499HM3 2369660 00000100500HM8 2369660 00000100552HM9 2369660 00000100553HM7 2369660 00000100554HM5 2369660 00000100673HM1 2369660 00000100721HM0 2369660 00000100722HM8 2369660 00000100740HM0 2369660 00000100951HM3 2369660 00000100985HM1 2369660 00000100986HM9 2369660 00000100987HM7 2369660 00000101508HM0 2369660 00000101509HM8 2369660 00000101510HM6 2369660 00000101598HM1 2369660 00000101599HM9 2369660 00000101900HM9 2369660 00000102096HM5 2369660 00000102108HM8 2369660 00000102109HM6 2369660 00000102110HM4 2369660 00000102136HM9 2369660 00000102146HM8 2369660 00000102236HM7 2369660 00000102237HM5 2369660 00000102511HM3 2369660 00000102512HM1 2369660 00000102601HM2 2369660 00000102602HM0 2369660 00000102603HM8 2369660 00000102630HM1 2369660 00000103124HM4 2369660 00000103125HM1 2369660 00000103390HM1 2369660 00000103391HM9 2369660 00000103392HM7 2369660 00000103393HM5 2369660 00000103452HM9 2369660 00000103452HM9 2369660 00000103453HM7 2369660 00000103707HM6 2369660 00000104156HM5 2369660 00000104188HM8 2369660 00000104316HM5 2369660 00000104451HM0 2369660 00000104462HM7 2369660 00000104921HM2 2369660 00000104933HM7 2369660 00000104934HM5 2369660 00000104935HM2 2369660 00000105258HM8 2369660 00000105259HM6 2369660 00000105277HM8 2369660 00000105569HM8 2369660 00000105772HM8 2369660 00000105874HM2 2369660 00000105903HM9 2369660 00000106130HM8 2369660 00000106131HM6 2369660 00000106150HM6 2369660 00000106151HM4 2369660 00000106193HM6 2369660 00000106194HM4 2369660 00000106195HM1 2369660 00000106345HM2 2369660 00000106773HM5 2369660 00000106774HM3 2369660 00000106824HM6 2369660 00000106843HM6 2369660 00000107535HM7 2369660 00000107536HM5 2369660 00000107537HM3 2369660 00000107564HM7 2369660 00000107565HM4 2369660 00000107692HM6 2369660 00000107693HM4 2369660 00000107712HM2 2369660 00000107713HM0 2369660 00000107854HM2 2369660 00000108045HM6 2369660 00000108079HM5 2369660 00000108097HM7 2369660 00000108401HM1 2369660 00000108402HM9 2369660 00000108422HM7 2369660 00000108423HM5 2369660 00000108424HM3 2369660 00000108450HM8 2369660 00000108917HM6 2369660 00000108928HM3 2369660 00000108929HM1 2369660 00000108948HM1 2369660 00000108949HM9 2369660 00000108969HM7 2369660 00000108971HM3 2369660 00000109512HM4 2369660 00000109513HM2 2369660 00000109542HM1 2369660 00000109543HM9 2369660 00000109640HM3 2369660 00000110011HM4 2369660 00000110022HM1 2369660 00000110023HM9 2369660 00000110206HM0 2369660 00000110207HM8 2369660 00000110225HM0 2369660 00000110235HM9 2369660 00000110250HM8 2369660 00000110388HM6 2369660 00000110389HM4 2369660 00000110408HM2 2369660 00000110409HM0 2369660 00000110493HM4 2369660 00000110494HM2 2369660 00000110495HM9 2369660 00000110672HM3 2369660 00000110691HM3 2369660 00000110796HM0 2369660 00000110797HM8 2369660 00000110817HM4 2369660 00000110818HM2 2369660 00000110819HM0 2369660 00000110846HM3 2369660 00000110847HM1 2369660 00000111420HM6 2369660 00000111432HM1 2369660 00000111433HM9 2369660 00000111434HM7 2369660 00000111471HM9 2369660 00000111887HM6 2369660 00000111888HM4 2369660 00000111907HM2 2369660 00000111908HM0 2369660 00000111961HM9 2369660 00000111962HM7 2369660 00000111963HM5 2369660 00000112111HM0 2369660 00000112122HM7 2369660 00000112123HM5 2369660 00000112580HM6 2369660 00000112581HM4 2369660 00000112605HM1 2369660 00000112606HM9 2369660 00000112607HM7 2369660 00000112608HM5 2369660 00000112609HM3 2369660 00000112610HM1 2369660 00000112611HM9 2369660 00000113130HM9 2369660 00000113140HM8 2369660 00000113465HM9 2369660 00000113466HM7 2369660 00000113467HM5 2369660 00000113468HM3 2369660 00000113536HM7 2369660 00000113537HM5 2369660 00000113555HM7 2369660 00000113734HM8 2369660 00000113735HM5 2369660 00000113859HM3 2369660 00000113860HM1 2369660 00000113878HM3 2369660 00000114534HM1 2369660 00000114545HM7 2369660 00000114546HM5 2369660 00000114632HM3 2369660 00000114633HM1 2369660 00000114634HM9 2369660 00000114683HM6 2369660 00000114684HM4 2369660 00000114703HM2 2369660 00000114704HM0 2369660 00000114904HM6 2369660 00000115238HM8 2369660 00000115239HM6 2369660 00000115257HM8 2369660 00000115431HM9 2369660 00000115432HM7 2369660 00000115452HM5 2369660 00000115453HM3 2369660 00000115454HM1 2369660 00000115576HM1 2369660 00000115577HM9 2369660 00000115797HM3 2369660 00000115798HM1 2369660 00000115799HM9 2369660 00000115826HM0 2369660 00000115827HM8 2369660 00000116295HM7 2369660 00000116296HM5 2369660 00000116314HM6 2369660 00000116664HM4 2369660 00000116665HM1 2369660 00000116866HM5 2369660 00000116867HM3 2369660 00000116937HM4 2369660 00000116985HM3 2369660 00000117213HM9 2369660 00000117214HM7 2369660 00000117443HM2 2369660 00000117444HM0 2369660 00000117445HM7 2369660 00000117471HM3 2369660 00000117689HM0 2369660 00000117690HM8 2369660 00000117772HM4 2369660 00000117773HM2 2369660 00000117918HM3 2369660 00000117929HM0 2369660 00000117930HM8 2369660 00000118027HM2 2369660 00000118028HM0 2369660 00000118080HM1 2369660 00000118081HM9 2369660 00000118346HM6 2369660 00000118347HM4 2369660 00000118562HM8 2369660 00000118563HM6 2369660 00000118564HM4 2369660 00000118614HM7 2369660 00000118615HM4 2369660 00000118616HM2 2369660 00000119031HM3 2369660 00000119032HM1 2369660 00000119052HM9 2369660 00000119352HM3 2369660 00000119353HM1 2369660 00000119354HM9 2369660 00000119420HM8 2369660 00000119421HM6 2369660 00000119464HM6 2369660 00000119819HM1 2369660 00000119820HM9 2369660 00000120107HM8 2369660 00000120108HM6 2369660 00000120140HM9 2369660 00000120141HM7 2369660 00000120142HM5 2369660 00000120143HM3 2369660 00000120471HM8 2369660 00000120472HM6 2369660 00000120835HM4 2369660 00000120836HM2 2369660 00000120837HM0 2369660 00000121142HM4 2369660 00000121143HM2 2369660 00000121164HM8 2369660 00000121165HM5 2369660 00000121166HM3 2369660 00000121167HM1 2369660 00000121244HM8 2369660 00000121579HM7 2369660 00000121580HM5 2369660 00000121581HM3 2369660 00000121582HM1 2369660 00000121583HM9 2369660 00000122045HM8 2369660 00000122054HM0 2369660 00000122150HM6 2369660 00000122151HM4 2369660 00000122171HM2 2369660 00000122172HM0 2369660 00000122173HM8 2369660 00000122596HM0 2369660 00000122597HM8 2369660 00000122616HM6 2369660 00000122617HM4 2369660 00000122875HM8 2369660 00000122906HM1 2369660 00000123039HM0 2369660 00000123040HM8 2369660 00000123059HM8 2369660 00000123060HM6 2369660 00000123103HM4 2369660 00000123104HM2 2369660 00000123501HM9 2369660 00000123502HM7 2369660 00000123503HM5 2369660 00000123970HM6 2369660 00000123971HM4 2369660 00000124111HM6 2369660 00000124112HM4 2369660 00000124113HM2 2369660 00000124916HM8 2369660 00000124927HM5 2369660 00000125968HM8 2369660 00000125969HM6 2369660 00000125986HM0 2369660 00000125987HM8 2369660 00000126769HM9 2369660 00000126770HM7 2369660 00000126791HM3 2369660 00000126792HM1 2369660 00000126793HM9 2369660 00000126794HM7 2369660 00000127481HM0 2369660 00000127482HM8 2369660 00000127526HM2 2369660 00000127527HM0 2369660 00000127528HM8 2369660 00000128048HM6 2369660 00000128049HM4 2369660 00000128111HM2 2369660 00000128360HM5 2369660 00000128361HM3 2369660 00000128362HM1 2369660 00000128363HM9 2369660 00000128570HM9 2369660 00000128571HM7 2369660 00000128780HM4 2369660 00000128789HM5 2369660 00000129127HM7 2369660 00000129319HM0 2369660 00000129588HM0 2369660 00000129589HM8 2369660 00000129859HM5 2369660 00000129860HM3 2369660 00000129878HM5 2369660 00000130538HM2 2369660 00000130539HM0 2369660 00000131067HM1 2369660 00000131068HM9 2369660 00000131086HM1 2369660 00000131377HM4 2369660 00000131519HM1 2369660 00000131530HM8 2369660 00000131540HM7 2369660 00000131541HM5 2369660 00000131888HM0 2369660 00000131889HM8 2369660 00000131890HM6 2369660 00000131994HM6 2369660 00000132420HM1 2369660 00000132421HM9 2369660 00000132422HM7 2369660 00000133044HM8 2369660 00000133045HM5 2369660 00000133046HM3 2369660 00000133138HM8 2369660 00000133148HM7 2369660 00000133348HM3 2369660 00000133349HM1 2369660 00000133482HM0 2369660 00000133492HM9 2369660 00000133503HM3 2369660 00000133504HM1 2369660 00000134089HM2 2369660 00000134090HM0 2369660 00000134091HM8 2369660 00000134120HM5 2369660 00000134121HM3 2369660 00000134175HM9 2369660 00000134185HM8 2369660 00000134818HM4 2369660 00000134867HM1 2369660 00000135236HM8 2369660 00000135237HM6 2369660 00000135730HM0 2369660 00000135754HM0 2369660 00000135755HM7 2369660 00000135756HM5 2369660 00000135826HM6 2369660 00000138002HM1 2369660 00000138036HM9 2369660 00000138037HM7 2369660 00000138038HM5 2369660 00000138856HM0 2369660 00000138857HM8 2369660 00000139136HM6 2369660 00000139137HM4 2369660 00000139138HM2 2369660 00000139139HM0 2369660 00000139612HM6 2369660 00000139613HM4 2369660 00000139631HM6 2369660 00000139910HM4 2369660 00000139919HM5 2369660 00000140574HM5 2369660 00000140575HM2 2369660 00000140594HM3 2369660 00000140605HM7 2369660 00000142235HM1 2369660 00000142236HM9 2369660 00000143081HM8 2369660 00000143809HM2 2369660 00000143810HM0 2369660 00000143811HM8 2369660 00000143837HM3 2369660 00000144125HM2 2369660 00000144126HM0 2369660 00000144127HM8 2369660 00000144128HM6 2369660 00000144739HM0 2369660 00000144740HM8 2369660 00000144741HM6 2369660 00000145087HM3 2369660 00000145298HM6 2369660 00000145299HM4 2369660 00000146060HM9 2369660 00000146061HM7 2369660 00000146148HM2 2369660 00000146555HM8 2369660 00000146556HM6 2369660 00000146557HM4 2369660 00000146557HM4 2369660 00000146558HM2 2369660 00000147658HM9 2369660 00000147659HM7 2369660 00000147660HM5 2369660 00000147661HM3 2369660 00000147722HM3 2369660 00000147881HM7 2369660 00000147882HM5 2369660 00000148226HM4 2369660 00000149227HM1 2369660 00000149238HM8 2369660 00000149239HM6 2369660 00000149257HM8 2369660 00000149519HM1 2369660 00000149530HM8 2369660 00000149543HM1 2369660 00000149544HM9 2369660 00000149545HM6 2369660 00000149546HM4 2369660 00000149547HM2 2369660 00000150559HM3 2369660 00000150560HM1 2369660 00000150561HM9 2369660 00000150562HM7 2369660 00000150597HM3 2369660 00000151437HM1 2369660 00000151448HM8 2369660 00000151449HM6 2369660 00000151468HM6 2369660 00000151469HM4 2369660 00000152227HM5 2369660 00000152238HM2 2369660 00000153121HM9 2369660 00000153122HM7 2369660 00000153154HM0 2369660 00000153155HM7 2369660 00000153961HM8 2369660 00000153962HM6 2369660 00000153963HM4 2369660 00000153964HM2 2369660 00000153965HM9 2369660 00000154004HM6 2369660 00000154005HM3 2369660 00000154006HM1 2369660 00000154083HM0 2369660 00000154084HM8 2369660 00000154135HM8 2369660 00000155174HM6 2369660 00000155175HM3 2369660 00000155730HM5 2369660 00000155731HM3 2369660 00000158931HM6 2369660 00000159551HM1 2369660 00000159552HM9 2369660 00000160034HM5 2369660 00000160035HM2 2369660 00000160503HM9 2369660 00000160504HM7 2369660 00000162312HM3 2369660 00000162323HM0 2369660 00000162324HM8 2369660 00000162999HM7 2369660 00000163691HM9 2369660 00000164390HM7 2369660 00000164715HM5 2369660 00000165328HM6 2369660 00000165329HM4 2369660 00000165330HM2 2369660 00000165640HM4 2369660 00000165641HM2 2369660 00000165662HM8 2369660 00000165663HM6 2369660 00000166090HM1 2369660 00000166091HM9 2369660 00000166092HM7 2369660 00000166116HM4 2369660 00000166117HM2 2369660 00000166985HM2 2369660 00000167327HM6 2369660 00000167468HM8 2369660 00000168506HM4 2369660 00000170342HM0 2369660 00000170343HM8 2369660 00000171782HM6 2369660 00000171792HM5 2369660 00000172023HM4 2369660 00000172032HM5 2369660 00000173211HM4 2369660 00000173212HM2 2369660 00000173213HM0 2369660 00000173214HM8 2369660 00000173215HM5 2369660 00000173216HM3 2369660 00000173266HM8 2369660 00000174296HM4 2369660 00000174297HM2 2369660 00000174739HM3 2369660 00000175439HM9 2369660 00000175440HM7 2369660 00000175441HM5 2369660 00000175442HM3 2369660 00000176328HM3 2369660 00000177666HM5 2369660 00000179319HM9 2369660 00000179780HM2 2369660 00000179966HM7 2369660 00000180441HM8 2369660 00000180450HM9 2369660 00000180464HM0 2369660 00000180534HM0 2369660 00000180535HM7 2369660 00000180536HM5 2369660 00000180537HM3 2369660 00000181595HM0 2369660 00000181608HM1 2369660 00000181609HM9 2369660 00000182200HM6 2369660 00000182201HM4 2369660 00000182202HM2 2369660 00000182681HM7 2369660 00000182682HM5 2369660 00000185084HM1 2369660 00000185797HM8 2369660 00000186052HM7 2369660 00000186824HM9 2369660 00000186867HM8 2369660 00000187426HM2 2369660 00000187645HM7 2369660 00000187646HM5 2369660 00000187664HM8 2369660 00000188744HM7 2369660 00000189542HM4 2369660 00000189543HM2 2369660 00000190230HM3 2369660 00000190231HM1 2369660 00000191489HM4 2369660 00000192286HM3 2369660 00000193647HM5 2369660 00000194065HM9 2369660 00000194155HM8 2369660 00000196077HM2 2369660 00000196078HM0 2369660 00000197034HM2 2369660 00000197493HM0 2369660 00000197700HM8 2369660 00000197941HM8 2369660 00000197942HM6 2369660 00000199507HM5 5114671 00000141141HM2 5114671 00000160446HM1 5114671 00000096796HM8 5114671 00000096797HM6 5114671 00000097312HM3 5114671 00000097313HM1 5114671 00000097314HM9 5114671 00000098369HM2 5114671 00000098370HM0 5114671 00000098371HM8 5114671 00000098669HM5 5114671 00000098670HM3 5114671 00000098788HM3 5114671 00000098789HM1 5114671 00000098790HM9 5114671 00000101628HM6 5114671 00000101629HM4 5114671 00000102204HM5 5114671 00000102532HM9 5114671 00000102533HM7 5114671 00000102534HM5 5114671 00000102950HM3 5114671 00000102951HM1 5114671 00000103440HM4 5114671 00000104167HM2 5114671 00000104168HM0 5114671 00000104737HM2 5114671 00000104963HM4 5114671 00000104964HM2 5114671 00000104965HM9 5114671 00000105289HM3 5114671 00000105290HM1 5114671 00000105291HM9 5114671 00000106170HM4 5114671 00000106171HM2 5114671 00000106802HM2 5114671 00000106803HM0 5114671 00000107731HM2 5114671 00000108567HM9 5114671 00000108568HM7 5114671 00000108891HM3 5114671 00000109109HM9 5114671 00000109283HM2 5114671 00000109284HM0 5114671 00000109367HM3 5114671 00000109368HM1 5114671 00000109369HM9 5114671 00000109370HM7 5114671 00000109490HM3 5114671 00000109491HM1 5114671 00000109960HM5 5114671 00000109962HM1 5114671 00000109963HM9 5114671 00000109964HM7 5114671 00000110356HM3 5114671 00000110357HM1 5114671 00000110358HM9 5114671 00000110701HM0 5114671 00000110868HM7 5114671 00000110869HM5 5114671 00000110870HM3 5114671 00000110871HM1 5114671 00000111938HM7 5114671 00000111939HM5 5114671 00000112070HM8 5114671 00000112071HM6 5114671 00000112072HM4 5114671 00000112073HM2 5114671 00000113064HM0 5114671 00000113065HM7 5114671 00000113066HM5 5114671 00000113067HM3 5114671 00000113068HM1 5114671 00000113890HM8 5114671 00000113891HM6 5114671 00000114915HM2 5114671 00000114916HM0 5114671 00000115269HM3 5114671 00000115270HM1 5114671 00000115271HM9 5114671 00000115596HM9 5114671 00000115597HM7 5114671 00000115943HM3 5114671 00000115944HM1 5114671 00000115945HM8 5114671 00000116336HM9 5114671 00000116337HM7 5114671 00000116685HM9 5114671 00000116686HM7 5114671 00000116687HM5 5114671 00000116896HM2 5114671 00000116897HM0 5114671 00000116898HM8 5114671 00000116899HM6 5114671 00000117387HM1 5114671 00000117388HM9 5114671 00000117483HM8 5114671 00000117484HM6 5114671 00000117485HM3 5114671 00000118058HM7 5114671 00000118059HM5 5114671 00000118314HM4 5114671 00000118315HM1 5114671 00000118316HM9 5114671 00000119164HM2 5114671 00000119165HM9 5114671 00000119166HM7 5114671 00000119382HM0 5114671 00000119383HM8 5114671 00000119384HM6 5114671 00000119440HM6 5114671 00000119441HM4 5114671 00000120126HM8 5114671 00000120181HM3 5114671 00000120864HM4 5114671 00000120865HM1 5114671 00000121287HM7 5114671 00000121288HM5 5114671 00000121627HM4 5114671 00000121627HM4 5114671 00000121628HM2 5114671 00000121629HM0 5114671 00000122201HM7 5114671 00000122202HM5 5114671 00000122203HM3 5114671 00000122885HM7 5114671 00000122886HM5 5114671 00000123121HM6 5114671 00000123122HM4 5114671 00000123532HM4 5114671 00000123533HM2 5114671 00000123534HM0 5114671 00000123535HM7 5114671 00000124144HM7 5114671 00000124145HM4 5114671 00000124146HM2 5114671 00000124147HM0 5114671 00000124148HM8 5114671 00000124149HM6 5114671 00000124750HM1 5114671 00000124941HM6 5114671 00000124942HM4 5114671 00000124943HM2 5114671 00000124944HM0 5114671 00000124945HM7 5114671 00000124946HM5 5114671 00000125487HM9 5114671 00000125488HM7 5114671 00000125489HM5 5114671 00000125490HM3 5114671 00000125608HM0 5114671 00000125609HM8 5114671 00000125610HM6 5114671 00000125611HM4 5114671 00000125903HM5 5114671 00000125904HM3 5114671 00000125905HM0 5114671 00000125906HM8 5114671 00000125907HM6 5114671 00000125908HM4 5114671 00000126530HM5 5114671 00000126531HM3 5114671 00000126532HM1 5114671 00000126533HM9 5114671 00000126534HM7 5114671 00000127580HM9 5114671 00000127581HM7 5114671 00000127583HM3 5114671 00000127584HM1 5114671 00000128068HM4 5114671 00000128069HM2 5114671 00000128090HM8 5114671 00000128091HM6 5114671 00000128758HM0 5114671 00000128759HM8 5114671 00000128946HM1 5114671 00000129610HM2 5114671 00000129611HM0 5114671 00000130239HM7 5114671 00000130240HM5 5114671 00000130241HM3 5114671 00000130242HM1 5114671 00000130865HM9 5114671 00000130866HM7 5114671 00000130867HM5 5114671 00000130868HM3 5114671 00000131260HM2 5114671 00000131261HM0 5114671 00000131747HM8 5114671 00000131803HM9 5114671 00000131804HM7 5114671 00000132521HM6 5114671 00000132522HM4 5114671 00000132523HM2 5114671 00000133107HM3 5114671 00000133108HM1 5114671 00000133109HM9 5114671 00000133326HM9 5114671 00000133327HM7 5114671 00000133558HM7 5114671 00000133559HM5 5114671 00000133560HM3 5114671 00000133561HM1 5114671 00000133561HM1 5114671 00000134295HM5 5114671 00000134469HM6 5114671 00000134470HM4 5114671 00000134471HM2 5114671 00000134472HM0 5114671 00000134787HM1 5114671 00000134788HM9 5114671 00000134789HM7 5114671 00000135266HM5 5114671 00000135267HM3 5114671 00000135366HM3 5114671 00000135367HM1 5114671 00000135368HM9 5114671 00000135369HM7 5114671 00000135370HM5 5114671 00000135371HM3 5114671 00000135785HM4 5114671 00000135786HM2 5114671 00000135787HM0 5114671 00000135788HM8 5114671 00000135839HM9 5114671 00000135840HM7 5114671 00000135841HM5 5114671 00000135842HM3 5114671 00000136435HM5 5114671 00000136436HM3 5114671 00000136437HM1 5114671 00000136438HM9 5114671 00000136487HM6 5114671 00000136488HM4 5114671 00000136489HM2 5114671 00000136490HM0 5114671 00000136492HM6 5114671 00000136493HM4 5114671 00000136494HM2 5114671 00000137070HM9 5114671 00000137097HM2 5114671 00000137098HM0 5114671 00000137099HM8 5114671 00000137100HM4 5114671 00000137101HM2 5114671 00000137102HM0 5114671 00000137103HM8 5114671 00000137104HM6 5114671 00000137105HM3 5114671 00000137106HM1 5114671 00000137636HM7 5114671 00000137637HM5 5114671 00000137712HM6 5114671 00000137713HM4 5114671 00000137714HM2 5114671 00000137715HM9 5114671 00000137716HM7 5114671 00000137717HM5 5114671 00000137718HM3 5114671 00000137719HM1 5114671 00000138013HM8 5114671 00000138014HM6 5114671 00000138270HM4 5114671 00000138271HM2 5114671 00000138272HM0 5114671 00000138273HM8 5114671 00000138274HM6 5114671 00000138330HM6 5114671 00000138331HM4 5114671 00000138332HM2 5114671 00000138333HM0 5114671 00000138334HM8 5114671 00000138759HM6 5114671 00000138760HM4 5114671 00000138761HM2 5114671 00000138762HM0 5114671 00000138763HM8 5114671 00000138764HM6 5114671 00000138822HM2 5114671 00000138823HM0 5114671 00000138824HM8 5114671 00000139080HM6 5114671 00000139081HM4 5114671 00000139082HM2 5114671 00000139083HM0 5114671 00000139356HM0 5114671 00000139368HM5 5114671 00000139590HM4 5114671 00000139591HM2 5114671 00000139668HM8 5114671 00000139669HM6 5114671 00000139934HM4 5114671 00000139935HM1 5114671 00000139936HM9 5114671 00000139937HM7 5114671 00000139938HM5 5114671 00000139939HM3 5114671 00000140725HM3 5114671 00000140738HM6 5114671 00000140739HM4 5114671 00000140740HM2 5114671 00000141105HM7 5114671 00000141106HM5 5114671 00000141107HM3 5114671 00000141140HM4 5114671 00000141163HM6 5114671 00000141502HM5 5114671 00000141522HM3 5114671 00000141523HM1 5114671 00000141524HM9 5114671 00000141525HM6 5114671 00000141526HM4 5114671 00000141527HM2 5114671 00000141528HM0 5114671 00000141529HM8 5114671 00000141530HM6 5114671 00000142049HM6 5114671 00000142050HM4 5114671 00000142051HM2 5114671 00000142189HM0 5114671 00000142190HM8 5114671 00000142191HM6 5114671 00000142192HM4 5114671 00000142256HM7 5114671 00000142257HM5 5114671 00000142258HM3 5114671 00000143018HM0 5114671 00000143019HM8 5114671 00000143020HM6 5114671 00000143021HM4 5114671 00000143022HM2 5114671 00000143180HM8 5114671 00000143181HM6 5114671 00000143182HM4 5114671 00000143183HM2 5114671 00000143545HM2 5114671 00000143546HM0 5114671 00000143547HM8 5114671 00000143548HM6 5114671 00000143549HM4 5114671 00000143550HM2 5114671 00000143551HM0 5114671 00000143776HM3 5114671 00000143777HM1 5114671 00000143778HM9 5114671 00000143909HM0 5114671 00000143910HM8 5114671 00000143911HM6 5114671 00000143912HM4 5114671 00000143915HM7 5114671 00000143916HM5 5114671 00000144666HM5 5114671 00000144678HM0 5114671 00000144815HM8 5114671 00000144816HM6 5114671 00000144817HM4 5114671 00000144818HM2 5114671 00000144819HM0 5114671 00000144820HM8 5114671 00000144821HM6 5114671 00000145424HM8 5114671 00000145425HM5 5114671 00000145426HM3 5114671 00000145427HM1 5114671 00000145656HM5 5114671 00000145657HM3 5114671 00000145658HM1 5114671 00000145659HM9 5114671 00000145660HM7 5114671 00000145661HM5 5114671 00000146048HM4 5114671 00000146367HM8 5114671 00000146590HM5 5114671 00000146591HM3 5114671 00000146592HM1 5114671 00000146764HM6 5114671 00000146765HM3 5114671 00000146790HM1 5114671 00000146791HM9 5114671 00000146792HM7 5114671 00000146793HM5 5114671 00000147317HM2 5114671 00000147318HM0 5114671 00000147319HM8 5114671 00000147320HM6 5114671 00000147629HM0 5114671 00000147772HM8 5114671 00000147773HM6 5114671 00000148017HM7 5114671 00000148018HM5 5114671 00000148019HM3 5114671 00000148020HM1 5114671 00000148180HM3 5114671 00000148181HM1 5114671 00000148182HM9 5114671 00000148183HM7 5114671 00000148184HM5 5114671 00000148240HM5 5114671 00000148241HM3 5114671 00000148242HM1 5114671 00000148243HM9 5114671 00000148244HM7 5114671 00000148651HM3 5114671 00000148652HM1 5114671 00000148653HM9 5114671 00000148687HM7 5114671 00000148688HM5 5114671 00000148689HM3 5114671 00000149469HM9 5114671 00000149470HM7 5114671 00000149471HM5 5114671 00000149472HM3 5114671 00000149473HM1 5114671 00000149998HM7 5114671 00000149999HM5 5114671 00000150000HM8 5114671 00000150035HM4 5114671 00000150036HM2 5114671 00000150037HM0 5114671 00000150102HM2 5114671 00000150103HM0 5114671 00000150104HM8 5114671 00000150105HM5 5114671 00000150393HM7 5114671 00000150394HM5 5114671 00000150395HM2 5114671 00000150396HM0 5114671 00000150397HM8 5114671 00000150525HM4 5114671 00000150526HM2 5114671 00000150527HM0 5114671 00000150974HM4 5114671 00000150975HM1 5114671 00000151113HM8 5114671 00000151114HM6 5114671 00000151115HM3 5114671 00000151116HM1 5114671 00000151160HM9 5114671 00000151161HM7 5114671 00000151162HM5 5114671 00000151191HM4 5114671 00000151804HM2 5114671 00000151805HM9 5114671 00000151806HM7 5114671 00000151807HM5 5114671 00000151808HM3 5114671 00000151932HM1 5114671 00000151933HM9 5114671 00000151934HM7 5114671 00000151935HM4 5114671 00000151936HM2 5114671 00000151937HM0 5114671 00000151938HM8 5114671 00000151939HM6 5114671 00000152688HM8 5114671 00000152689HM6 5114671 00000152690HM4 5114671 00000152691HM2 5114671 00000152692HM0 5114671 00000152693HM8 5114671 00000152694HM6 5114671 00000152695HM3 5114671 00000152696HM1 5114671 00000152697HM9 5114671 00000152698HM7 5114671 00000152699HM5 5114671 00000153327HM2 5114671 00000153399HM1 5114671 00000153400HM7 5114671 00000153401HM5 5114671 00000153402HM3 5114671 00000153403HM1 5114671 00000153901HM4 5114671 00000153902HM2 5114671 00000153903HM0 5114671 00000153904HM8 5114671 00000153905HM5 5114671 00000154041HM8 5114671 00000154042HM6 5114671 00000154043HM4 5114671 00000154044HM2 5114671 00000154145HM7 5114671 00000155062HM3 5114671 00000155063HM1 5114671 00000155064HM9 5114671 00000155065HM6 5114671 00000155096HM1 5114671 00000155117HM5 5114671 00000155118HM3 5114671 00000155119HM1 5114671 00000155120HM9 5114671 00000155121HM7 5114671 00000155276HM9 5114671 00000155639HM8 5114671 00000155640HM6 5114671 00000155641HM4 5114671 00000155642HM2 5114671 00000155643HM0 5114671 00000155644HM8 5114671 00000155708HM1 5114671 00000155709HM9 5114671 00000156232HM1 5114671 00000156233HM9 5114671 00000156234HM7 5114671 00000156267HM7 5114671 00000156268HM5 5114671 00000156356HM8 5114671 00000156357HM6 5114671 00000156358HM4 5114671 00000156359HM2 5114671 00000156360HM0 5114671 00000156361HM8 5114671 00000156648HM8 5114671 00000156649HM6 5114671 00000156674HM4 5114671 00000156675HM1 5114671 00000156676HM9 5114671 00000156677HM7 5114671 00000156727HM0 5114671 00000156728HM8 5114671 00000156729HM6 5114671 00000156987HM0 5114671 00000156988HM8 5114671 00000156989HM6 5114671 00000157079HM5 5114671 00000157080HM3 5114671 00000157081HM1 5114671 00000157082HM9 5114671 00000157083HM7 5114671 00000157568HM7 5114671 00000157569HM5 5114671 00000157570HM3 5114671 00000157571HM1 5114671 00000157614HM9 5114671 00000157762HM6 5114671 00000157763HM4 5114671 00000158102HM4 5114671 00000158103HM2 5114671 00000158104HM0 5114671 00000158105HM7 5114671 00000158415HM0 5114671 00000158428HM3 5114671 00000158429HM1 5114671 00000158453HM1 5114671 00000158454HM9 5114671 00000158455HM6 5114671 00000158880HM5 5114671 00000158881HM3 5114671 00000158882HM1 5114671 00000158883HM9 5114671 00000158884HM7 5114671 00000159509HM9 5114671 00000159510HM7 5114671 00000159511HM5 5114671 00000159512HM3 5114671 00000159952HM1 5114671 00000159953HM9 5114671 00000159954HM7 5114671 00000159955HM4 5114671 00000159956HM2 5114671 00000159957HM0 5114671 00000159958HM8 5114671 00000159959HM6 5114671 00000160399HM2 5114671 00000160400HM8 5114671 00000160401HM6 5114671 00000160402HM4 5114671 00000160447HM9 5114671 00000160448HM7 5114671 00000160449HM5 5114671 00000160450HM3 5114671 00000160522HM9 5114671 00000161380HM1 5114671 00000161381HM9 5114671 00000161382HM7 5114671 00000161383HM5 5114671 00000161427HM0 5114671 00000161428HM8 5114671 00000161429HM6 5114671 00000161465HM0 5114671 00000161466HM8 5114671 00000161467HM6 5114671 00000161468HM4 5114671 00000161469HM2 5114671 00000162079HM8 5114671 00000162080HM6 5114671 00000162081HM4 5114671 00000162118HM4 5114671 00000162119HM2 5114671 00000162120HM0 5114671 00000162121HM8 5114671 00000162122HM6 5114671 00000162123HM4 5114671 00000162346HM1 5114671 00000162347HM9 5114671 00000162348HM7 5114671 00000162349HM5 5114671 00000162350HM3 5114671 00000162906HM2 5114671 00000162921HM1 5114671 00000162922HM9 5114671 00000162923HM7 5114671 00000162924HM5 5114671 00000162968HM2 5114671 00000162969HM0 5114671 00000162970HM8 5114671 00000163146HM4 5114671 00000163147HM2 5114671 00000163301HM5 5114671 00000163302HM3 5114671 00000163303HM1 5114671 00000163304HM9 5114671 00000163349HM4 5114671 00000163350HM2 5114671 00000163351HM0 5114671 00000163352HM8 5114671 00000163630HM7 5114671 00000163657HM0 5114671 00000163658HM8 5114671 00000163680HM2 5114671 00000164257HM8 5114671 00000164258HM6 5114671 00000164299HM0 5114671 00000164300HM6 5114671 00000164301HM4 5114671 00000164302HM2 5114671 00000164303HM0 5114671 00000164304HM8 5114671 00000164305HM5 5114671 00000164306HM3 5114671 00000164307HM1 5114671 00000164308HM9 5114671 00000164724HM7 5114671 00000164739HM5 5114671 00000164740HM3 5114671 00000164741HM1 5114671 00000164742HM9 5114671 00000165361HM7 5114671 00000165362HM5 5114671 00000165363HM3 5114671 00000165364HM1 5114671 00000165365HM8 5114671 00000165366HM6 5114671 00000165582HM8 5114671 00000165583HM6 5114671 00000165605HM7 5114671 00000165618HM0 5114671 00000166024HM0 5114671 00000166026HM5 5114671 00000166027HM3 5114671 00000166028HM1 5114671 00000166029HM9 5114671 00000166899HM5 5114671 00000166900HM1 5114671 00000166901HM9 5114671 00000166902HM7 5114671 00000166903HM5 5114671 00000167240HM1 5114671 00000167241HM9 5114671 00000167242HM7 5114671 00000167272HM4 5114671 00000167273HM2 5114671 00000167294HM8 5114671 00000167295HM5 5114671 00000167315HM1 5114671 00000167364HM9 5114671 00000167365HM6 5114671 00000167393HM8 5114671 00000167394HM6 5114671 00000167395HM3 5114671 00000167396HM1 5114671 00000167397HM9 5114671 00000167398HM7 5114671 00000167399HM5 5114671 00000168409HM1 5114671 00000168410HM9 5114671 00000168411HM7 5114671 00000168412HM5 5114671 00000168413HM3 5114671 00000168414HM1 5114671 00000168472HM9 5114671 00000168473HM7 5114671 00000168474HM5 5114671 00000169227HM6 5114671 00000169228HM4 5114671 00000169229HM2 5114671 00000169335HM7 5114671 00000169336HM5 5114671 00000169337HM3 5114671 00000169338HM1 5114671 00000169339HM9 5114671 00000169396HM9 5114671 00000169849HM7 5114671 00000169850HM5 5114671 00000169851HM3 5114671 00000169852HM1 5114671 00000169853HM9 5114671 00000169903HM2 5114671 00000169904HM0 5114671 00000169905HM7 5114671 00000169906HM5 5114671 00000170252HM1 5114671 00000170253HM9 5114671 00000170254HM7 5114671 00000170290HM1 5114671 00000170291HM9 5114671 00000170292HM7 5114671 00000170293HM5 5114671 00000170294HM3 5114671 00000170901HM3 5114671 00000170902HM1 5114671 00000170903HM9 5114671 00000171098HM7 5114671 00000171099HM5 5114671 00000171100HM1 5114671 00000171101HM9 5114671 00000171102HM7 5114671 00000171149HM8 5114671 00000171887HM3 5114671 00000171888HM1 5114671 00000171889HM9 5114671 00000171890HM7 5114671 00000171891HM5 5114671 00000171892HM3 5114671 00000171893HM1 5114671 00000171894HM9 5114671 00000172010HM1 5114671 00000172383HM2 5114671 00000172446HM7 5114671 00000172447HM5 5114671 00000172448HM3 5114671 00000172449HM1 5114671 00000172450HM9 5114671 00000172451HM7 5114671 00000172452HM5 5114671 00000172770HM0 5114671 00000172771HM8 5114671 00000173438HM3 5114671 00000173439HM1 5114671 00000173440HM9 5114671 00000173441HM7 5114671 00000173442HM5 5114671 00000173443HM3 5114671 00000173444HM1 5114671 00000173445HM8 5114671 00000173446HM6 5114671 00000174318HM6 5114671 00000174319HM4 5114671 00000174320HM2 5114671 00000174321HM0 5114671 00000174360HM8 5114671 00000174361HM6 5114671 00000174362HM4 5114671 00000174363HM2 5114671 00000174364HM0 5114671 00000174365HM7 5114671 00000174366HM5 5114671 00000174367HM3 5114671 00000174368HM1 5114671 00000174369HM9 5114671 00000175355HM7 5114671 00000175356HM5 5114671 00000175357HM3 5114671 00000175358HM1 5114671 00000175402HM7 5114671 00000175403HM5 5114671 00000175404HM3 5114671 00000175484HM5 5114671 00000175485HM2 5114671 00000175486HM0 5114671 00000175487HM8 5114671 00000175488HM6 5114671 00000175489HM4 5114671 00000175490HM2 5114671 00000175491HM0 5114671 00000175492HM8 5114671 00000176185HM7 5114671 00000176186HM5 5114671 00000176187HM3 5114671 00000176188HM1 5114671 00000176189HM9 5114671 00000176240HM0 5114671 00000176257HM4 5114671 00000176258HM2 5114671 00000176259HM0 5114671 00000176260HM8 5114671 00000176261HM6 5114671 00000176262HM4 5114671 00000176263HM2 5114671 00000177172HM4 5114671 00000177173HM2 5114671 00000177174HM0 5114671 00000177175HM7 5114671 00000177176HM5 5114671 00000177257HM3 5114671 00000177258HM1 5114671 00000177259HM9 5114671 00000177260HM7 5114671 00000177261HM5 5114671 00000177262HM3 5114671 00000177263HM1 5114671 00000177328HM2 5114671 00000177678HM0 5114671 00000177679HM8 5114671 00000177680HM6 5114671 00000177681HM4 5114671 00000177728HM3 5114671 00000177729HM1 5114671 00000177730HM9 5114671 00000177731HM7 5114671 00000177732HM5 5114671 00000177733HM3 5114671 00000178621HM9 5114671 00000178622HM7 5114671 00000178623HM5 5114671 00000178624HM3 5114671 00000178673HM0 5114671 00000178674HM8 5114671 00000178675HM5 5114671 00000178716HM7 5114671 00000178717HM5 5114671 00000178718HM3 5114671 00000178719HM1 5114671 00000178720HM9 5114671 00000178721HM7 5114671 00000178722HM5 5114671 00000178723HM3 5114671 00000178724HM1 5114671 00000178725HM8 5114671 00000179237HM3 5114671 00000179238HM1 5114671 00000179239HM9 5114671 00000179240HM7 5114671 00000179241HM5 5114671 00000179242HM3 5114671 00000179243HM1 5114671 00000179244HM9 5114671 00000179799HM2 5114671 00000179800HM8 5114671 00000179801HM6 5114671 00000179802HM4 5114671 00000179803HM2 5114671 00000179804HM0 5114671 00000179805HM7 5114671 00000179806HM5 5114671 00000179807HM3 5114671 00000179808HM1 5114671 00000179809HM9 5114671 00000180580HM3 5114671 00000180581HM1 5114671 00000180582HM9 5114671 00000180583HM7 5114671 00000180584HM5 5114671 00000180585HM2 5114671 00000180586HM0 5114671 00000180587HM8 5114671 00000180588HM6 5114671 00000180589HM4 5114671 00000181154HM6 5114671 00000181155HM3 5114671 00000181156HM1 5114671 00000181157HM9 5114671 00000181158HM7 5114671 00000181159HM5 5114671 00000181160HM3 5114671 00000181161HM1 5114671 00000181162HM9 5114671 00000181572HM9 5114671 00000181573HM7 5114671 00000181626HM3 5114671 00000181627HM1 5114671 00000181654HM5 5114671 00000181655HM2 5114671 00000181656HM0 5114671 00000181657HM8 5114671 00000181658HM6 5114671 00000181659HM4 5114671 00000182295HM6 5114671 00000182296HM4 5114671 00000182297HM2 5114671 00000182298HM0 5114671 00000182299HM8 5114671 00000182300HM4 5114671 00000182619HM7 5114671 00000182620HM5 5114671 00000182645HM2 5114671 00000182646HM0 5114671 00000182668HM4 5114671 00000182756HM7 5114671 00000182757HM5 5114671 00000182758HM3 5114671 00000182964HM7 5114671 00000182965HM4 5114671 00000182966HM2 5114671 00000182967HM0 5114671 00000183012HM4 5114671 00000183013HM2 5114671 00000183014HM0 5114671 00000183015HM7 5114671 00000183569HM3 5114671 00000183570HM1 5114671 00000183571HM9 5114671 00000183758HM2 5114671 00000183759HM0 5114671 00000183812HM7 5114671 00000183813HM5 5114671 00000183814HM3 5114671 00000183815HM0 5114671 00000183816HM8 5114671 00000184091HM7 5114671 00000184092HM5 5114671 00000184093HM3 5114671 00000184130HM3 5114671 00000184131HM1 5114671 00000184132HM9 5114671 00000184133HM7 5114671 00000184134HM5 5114671 00000184135HM2 5114671 00000184192HM3 5114671 00000184193HM1 5114671 00000184626HM0 5114671 00000184627HM8 5114671 00000184628HM6 5114671 00000184629HM4 5114671 00000184667HM4 5114671 00000184682HM3 5114671 00000184683HM1 5114671 00000184684HM9 5114671 00000184685HM6 5114671 00000184994HM2 5114671 00000184995HM9 5114671 00000185022HM1 5114671 00000185023HM9 5114671 00000185024HM7 5114671 00000185025HM4 5114671 00000185026HM2 5114671 00000185027HM0 5114671 00000185450HM4 5114671 00000185462HM9 5114671 00000185480HM1 5114671 00000185481HM9 5114671 00000185482HM7 5114671 00000185483HM5 5114671 00000185484HM3 5114671 00000185485HM0 5114671 00000185486HM8 5114671 00000186040HM2 5114671 00000186071HM7 5114671 00000186072HM5 5114671 00000186073HM3 5114671 00000186074HM1 5114671 00000186075HM8 5114671 00000186076HM6 5114671 00000186077HM4 5114671 00000186078HM2 5114671 00000186079HM0 5114671 00000186080HM8 5114671 00000186606HM0 5114671 00000186628HM4 5114671 00000186629HM2 5114671 00000186630HM0 5114671 00000186631HM8 5114671 00000186632HM6 5114671 00000186633HM4 5114671 00000186634HM2 5114671 00000186635HM9 5114671 00000186636HM7 5114671 00000186637HM5 5114671 00000186638HM3 5114671 00000186800HM9 5114671 00000186812HM4 5114671 00000186836HM3 5114671 00000186837HM1 5114671 00000187677HM0 5114671 00000187678HM8 5114671 00000187679HM6 5114671 00000187680HM4 5114671 00000187723HM2 5114671 00000187724HM0 5114671 00000187760HM4 5114671 00000187761HM2 5114671 00000188098HM8 5114671 00000188099HM6 5114671 00000188100HM2 5114671 00000188101HM0 5114671 00000188102HM8 5114671 00000188149HM9 5114671 00000188161HM4 5114671 00000188653HM0 5114671 00000188654HM8 5114671 00000188655HM5 5114671 00000188688HM6 5114671 00000188689HM4 5114671 00000188713HM2 5114671 00000188714HM0 5114671 00000188715HM7 5114671 00000189462HM5 5114671 00000189463HM3 5114671 00000189464HM1 5114671 00000189496HM3 5114671 00000189510HM1 5114671 00000189511HM9 5114671 00000189512HM7 5114671 00000189704HM0 5114671 00000189705HM7 5114671 00000189730HM5 5114671 00000189731HM3 5114671 00000189732HM1 5114671 00000189733HM9 5114671 00000190249HM3 5114671 00000190250HM1 5114671 00000190251HM9 5114671 00000190288HM1 5114671 00000190289HM9 5114671 00000190290HM7 5114671 00000190291HM5 5114671 00000190292HM3 5114671 00000190293HM1 5114671 00000190868HM0 5114671 00000190869HM8 5114671 00000190892HM0 5114671 00000190893HM8 5114671 00000191160HM1 5114671 00000191161HM9 5114671 00000191162HM7 5114671 00000191163HM5 5114671 00000191164HM3 5114671 00000191165HM0 5114671 00000191166HM8 5114671 00000191167HM6 5114671 00000191443HM1 5114671 00000191444HM9 5114671 00000191445HM6 5114671 00000191477HM9 5114671 00000191818HM4 5114671 00000191819HM2 5114671 00000191820HM0 5114671 00000192234HM3 5114671 00000192235HM0 5114671 00000192236HM8 5114671 00000192237HM6 5114671 00000192238HM4 5114671 00000192423HM2 5114671 00000192442HM2 5114671 00000192443HM0 5114671 00000192444HM8 5114671 00000192445HM5 5114671 00000192446HM3 5114671 00000192447HM1 5114671 00000192448HM9 5114671 00000192449HM7 5114671 00000192938HM9 5114671 00000192939HM7 5114671 00000192940HM5 5114671 00000192941HM3 5114671 00000192942HM1 5114671 00000192943HM9 5114671 00000193000HM7 5114671 00000193001HM5 5114671 00000193635HM0 5114671 00000193664HM0 5114671 00000193665HM7 5114671 00000193666HM5 5114671 00000193667HM3 5114671 00000193668HM1 5114671 00000193669HM9 5114671 00000193670HM7 5114671 00000193671HM5 5114671 00000194081HM6 5114671 00000194082HM4 5114671 00000194083HM2 5114671 00000194084HM0 5114671 00000194085HM7 5114671 00000194086HM5 5114671 00000194087HM3 5114671 00000194545HM0 5114671 00000194708HM4 5114671 00000194709HM2 5114671 00000194710HM0 5114671 00000194844HM7 5114671 00000194845HM4 5114671 00000194846HM2 5114671 00000194847HM0 5114671 00000194848HM8 5114671 00000194849HM6 5114671 00000194850HM4 5114671 00000194851HM2 5114671 00000194852HM0 5114671 00000194941HM1 5114671 00000194942HM9 5114671 00000194943HM7 5114671 00000194944HM5 5114671 00000194945HM2 5114671 00000194946HM0 5114671 00000194947HM8 5114671 00000196039HM2 5114671 00000196040HM0 5114671 00000196041HM8 5114671 00000196129HM1 5114671 00000196130HM9 5114671 00000196131HM7 5114671 00000196132HM5 5114671 00000196675HM3 5114671 00000196676HM1 5114671 00000196677HM9 5114671 00000196678HM7 5114671 00000196679HM5 5114671 00000196685HM2 5114671 00000196686HM0 5114671 00000196687HM8 5114671 00000196688HM6 5114671 00000196689HM4 5114671 00000196739HM7 5114671 00000196976HM5 5114671 00000196977HM3 5114671 00000196978HM1 5114671 00000196979HM9 5114671 00000196980HM7 5114671 00000197042HM5 5114671 00000197530HM9 5114671 00000197531HM7 5114671 00000197532HM5 5114671 00000197533HM3 5114671 00000197534HM1 5114671 00000197535HM8 5114671 00000197536HM6 5114671 00000197893HM1 5114671 00000197894HM9 5114671 00000197895HM6 5114671 00000197896HM4 5114671 00000198051HM5 5114671 00000198052HM3 5114671 00000198053HM1 5114671 00000198054HM9 5114671 00000198055HM6 5114671 00000198056HM4 5114671 00000198121HM6 5114671 00000198122HM4 5114671 00000198167HM9 5114671 00000198168HM7 5114671 00000198169HM5 5114671 00000198170HM3 5114671 00000198171HM1 5114671 00000198172HM9 5114671 00000198835HM1 5114671 00000198836HM9 5114671 00000198933HM4 5114671 00000198934HM2 5114671 00000198935HM9 5114671 00000198936HM7 5114671 00000199462HM3 |
| FEI Number |
2126677
|
Recalling Firm/ Manufacturer |
GE Medical Systems, LLC
3000 N Grandview Blvd
Waukesha
WI
53188-1615
|
| For Additional Information Contact | 262-544-3894 |
Manufacturer Reason for Recall | During internal testing, GE Healthcare has recently determined that the collimator used on our CT/e and HiSpeed X/i families of Computed Tomography systems does not comply with the specific requirements in United States Title 21 of the Code of Federal Regulations, and IEC 60601-1-3, related to leakage radiation from the diagnostic source assembly, which includes the collimator.
The excess leak |
FDA Determined Cause 2 | Radiation Control for Health and Safety Act |
| Action | GE Healthcare issued an "Important - Electronic Product Radiation Warning" to affected customers. GE Healthcare field service representative will perform a field corrective action on all affected units. All potential users of the device should be made aware of the notification and the information contained.
For further information, contact GE Healthcare at 1-800-437-1171 (US) or by contacting your local GE Healthcare service manager. |
| Quantity in Commerce | 7005 units (787 US, 6218 OUS) |
| Distribution | Worldwide Distribution -- UNITED STATES (AL, AZ, AR, CA, CT, DC, FL, GA, IL, IN, KS, KY, LA, MD, MA, MI, LMN, MS, MO, NE, NJ, NM, NY, NC, OH, OK, PA, PR,SC, SD, TN, TX, VA, WA, WV and WI), ALGERIA, ANTIGUA, ARGENTINA, AUSTRALIA, AUSTRIA, AZERBAIJAN, BELGIUM, BENIN, BOLIVIA, BOSNIA & HERZEGOVINIA, BRAZIL BULGARIA, BYELORUSSIAN, CAMERON, CANADA, CHILE, CHINA, COLOMBIA, CONGO, COSTA RICA, CROATIA, CYPRUS, CZECH REPUBLIC, DENMARK , DOMINICAN REPUBLIC, ECUADOR, EGYPT, EL SALVADOR, ESTONIA, FINLAND, FRANCE, GEORGIA, GERMANY, GREECE, GUADELOUPE, GUATEMALA, GUYANA, HONDURAS, HONG KONG, HUNGARY, ICELAND, INDIA, INDONESIA, IRAN, IRELAND, ISRAEL, ITALY, IVORY COST, JAPAN, JORDAN, KAZAKHSTAN, KOREA, KUWAIT, LATVIA, LEBANON, LIBYAN ARAB JAMAHIRIYA, LITHUANIA, , MACEDONIA, MALAYSIA, MALTA, MARTINIQUE, MAURITANIA, MAURITIUS, MEXICO, MOROCCO, NETHERLANDS, NEW ZEALAND, NICARAGUA, NIGER, NIGERIA, NORWAY, OMAN, PAKISTAN, PANAMA, PARAGUAY, PERU, PHILIPPINES, POLAND, PORTUGAL QATAR, ROMANIA, RUSSIAN F=FEDERATION, SAUDI ARABIA, SINGAPORE, SLOVAK IA, SOUTH AFRICA, SPAIN, SUDAN, SWEDEN, SWITZERLAND , SYRIA ARAB REPUBLIC, TAIWAN, THAILAND, TRINIDAD & TOBAGO TUNISIA, TURKEY,, UGANDA, UKRAINE, UNITED KINGDOM, URUGUAY, VENEZUELA, VIETNAM, YEMEN and YUGOSLAVIA. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|