• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Stryker Reflex Hybrid Level Anterior Cervical Plate

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Stryker Reflex Hybrid Level Anterior Cervical Platesee related information
Date Initiated by FirmJune 14, 2006
Date PostedAugust 20, 2009
Recall Status1 Terminated 3 on August 20, 2009
Recall NumberZ-1958-2009
Recall Event ID 50783
510(K)NumberK040261 
Product Classification Appliance, fixation, spinal intervertebral body - Product Code KWQ
ProductStryker Reflex Hybrid Level Anterior Cervical Plate; Non Sterile; Manufactured by Stryker Spine SAS, Cestas, France. Distributed by Howmedica Osteonics Corp. Mahwah, NJ 07430.
Code Information All Product with the following Catalog Number/Description: 48650112, Reflex Hybrid one-level plate, size 12mm; 48650114m size 14 mm; 48650116, 16 mm; 48650118, size 18mm; 48650120, size 20 mm; 48650122, size 22 mm.  48650224, Reflex Hybrid Two-Level Plate, size 24 mm; 48650226, size 26 mm; 48650228, size 28 mm; 48650230, size 30 mm; 48650232, size 32 mm; 48650234, size 34 mm; 48650237, size 37 mm; 48650240, size 40 mm; 48650243, size 43 mm; 48650246, size 46 mm. 48650339, REflex Hybrid Three-Level Plate, Size 39 mm; 48650342, size 42mm; 48650345, suze 45 mm; 48650348, size 48 mm; 48650351, size 51 mm; 48650354, size 54 mm; 48650357, size 57 mm; 48650360, size 60 mm; 48650363, size 63 mm; 48650366, size 66 mm; 48650369, size 69 mm. 48650460, Reflex Hybrid four-Level Plate, size 60 mm; 48650464, size 64 mm; 48650468, size 68 mm; 48650472, size 72 mm; 48650476, size 76 mm; 48650489, size 80 mm; 48650484, size 84 mm. 
FEI Number 3004024955
Recalling Firm/
Manufacturer
Stryker Spine
2 Pearl Ct
Allendale NJ 07401-1611
For Additional Information ContactMichele Klein
201-760-8256
Manufacturer Reason
for Recall
Design Flaw: The design did not prevent over angulation of the screws when the recommended instrument, All in One Guide, was not used.
FDA Determined
Cause 2
Device Design
ActionStryker Spine issued an "Important Market Withdrawal" notice dated June 14, 2006. Consignees were advised to complete and return a Product Accountability Form via fax to 1-201-831-6069 prior to returning the affected product. Product return was to be performed during certain phase periods when upgrades were being shipped. For further questions, contact Stryker Spine at 1-201-831-5825.
Quantity in Commerce38987 plates
DistributionNationwide Distribution -- Including states of NC, OR, AZ, TN, CA, MA, MD, IL, GA, FL, NJ, NM, KY, LA, PA, MI, UT, MO, OK, KS, OH, MN, NE, IN, MT, TX, AL, CO and WI.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
510(K) Database510(K)s with Product Code = KWQ
-
-