| | Class 1 Device Recall GE OEC 9800 |  |
| Date Initiated by Firm | February 22, 2008 |
| Date Posted | July 24, 2012 |
| Recall Status1 |
Terminated 3 on February 19, 2013 |
| Recall Number | Z-2038-2012 |
| Recall Event ID |
51222 |
| 510(K)Number | K024012 K021049 K022069 |
| Product Classification |
System, x-ray, fluoroscopic, image-intensified - Product Code JAA
|
| Product | OEC 9800; OEC FluoroTrak 9800 Plus; OEC 9800 Plus; OEC 9800 MD Motorized C-arm System, 892.1650 Image-intensified fluoroscopic x-ray system.
The OEC 9800 is designed to provide fluoroscopic and spot-film imaging of the patient during diagnostic, surgical and interventional procedures. The system includes features specifically designed for use in diagnostic and interventional cardiac imaging procedures and is also intended for cholangiography, endoscopic, urologic orthopedic, neurologic, vascular, critical care and emergency room procedures. It may be used for other imaging applications at the physician's discretion. |
| Code Information |
8S-1663 , 8S-2887 , 89-3119 , 89-0098 , 8S-0152-C , 89-0540 , 82-1252, 82-1519, 89-2326 , 89-1204 , 89-3272, 89-3273 , 8S-0258 , 82-0678, 82-1884 , 8S-2321 , 82-2042 , 82-3210 , 8S-0533 , 8S-1077 , 89-0910 , 82-0139, 82-3871, 89-1757 , 89-3298 , 89-3026 , 8S-2299, 8S-2857 , 89-2043 , 89-2533 , 89-2463 , 82-0356-RC , 8S-2120 , 89-1896 , 8S-7152-MH , 8S-7190-MH , 8S-3246 , 8S-0289 , 89-1626 , 8S-0625 , 8S-1301 , 8S-1920 , 89-1025 , 8S-1838 , 8S-1367 , 8S-1627 , 89-1112 , 8S-0076-RC , 8S-0508, 8S-1946 , 8S-3073, 8S-3105 , 8S-2324-C , 8S-0073 , 8S-0180 , 8S-2237 , 82-0126, 82-3040, 82-3710 , 8S-2647 , 89-2319 , 89-3105 , 82-2560, 89-0285, 8S-1310 , 8S-2124 , 8S-1626 , 89-1486 , 82-3426 , 82-0875 , 8S-2566 , 89-0955 , 82-1801 , 82-1568 , 82-3682 , 89-0073, 89-1118, 89-1147 , 8S-0706 , 8S-0622, 8S-0772 , 8S-1139 , 8S-2182, 8S-2183 , 8S-2184 , 8S-0567 , 8S-1974 , 89-0454, 82-1606, 82-1870, 82-2533, 82-7080-MH, 89-0039, 89-1607, 89-2108, 8S-1457 , 8S-7120-MH , 82-0956, 89-3006 , 89-3007 , 8S-7127-MH , 89-2585 , 82-0134, 82-3452, 82-3538, 89-0127 , 89-2725 , 82-0153, 82-0166, 82-1277, 82-2189, 89-0086, 8S-2413 , 89-1149 , 89-0738 , 82-2659, 82-3809 , 8S-1825 , 82-2934 , 89-0510 , 89-1087 , 82-2700, 8S-0832 , 8S-1736 , 82-2731, 82-3463 , 8S-1089, 8S-3099 , 82-0902 , 8S-2095 , 89-2575, 8S-1932-N , 8S-2856 , 8S-0663 , 82-2832 , 89-3337 , 8S-1682 , 82-2113, 89-0315 , 82-2828, 89-0865, 8S-1334 , 8S-2281, 8S-2630, 8S-2638 , 89-2064 , 8S-0658, 8S-1540 , 8S-2144, 8S-2894 , 8S-1005, 82-2782, 8S-2363 , 82-3010 , 82-2471 , 8S-0125 , 89-0096 , 89-0591 , 89-2813 , 82-7179-MH , 89-1000 , 82-0827 , 89-0493 , 82-3078, 82-3676 , 89-2081 , 82-1001, 82-1014, 89-0862 , 82-0711, 82-7109-MH, 8S-3254 , 89-2389 , 8S-0410 , 8S-3011 , 8S-2639 , 82-1474, 82-2779 , 89-0110 , 89-0873 , 89-2234, 8S-0317 , 8S-3049 , 89-3520 , 8S-0021-C , 8S-0883 , 8S-3096 , 89-0103 , 82-3810 , 8S-2525 , 8S-0274 , 82-0507 , 82-1098 , 82-1733 , 82-1972 , 82-2566 , 82-2730, 82-3027 , 82-3307 , 82-3315 , 82-3360 , 82-3470, 82-3509 , 82-3510 , 82-3577 , 82-3131 , 82-0894-RC , 82-1399 , 8S-2949 , 82-1663 , 8S-2938-C , 89-1153 , 82-2094 , 8S-1340 , 8S-2271 , 82-2626, 82-2630 , 8S-2484 , 82-0071 , 8S-3152 , 89-1563 , 82-1063 , 89-1194 , 82-0097, 82-0515 , 8S-7095-CMH , 82-2967, 89-2513 , 8S-2895 , 89-1800, 89-2589 , 8S-1525 , 89-0735 , 8S-0544 , 89-1239 , 82-0094 , 8S-0022-C , 89-3506 , 8S-0285, 8S-3176 , 8S-3147 , 89-3024, 8S-1530 , 8S-2931 , 82-1551 , 82-2844 , 82-0896-RC, 82-1002 , 8S-2688 , 8S-2312 , 8S-2309 , 82-0164, 82-0276, 82-2241, 82-3312 , 8S-2104 , 8S-2684 , 8S-1464 , 8S-1029, 8S-1034, 8S-1050 , 8S-1855 , 89-1709, 89-3269, 8S-0254 , 89-1841 , 8S-1083 , 8S-1511 , 82-2952, 89-3307 , 8S-7094-CMH , 89-0895-RC , 8S-2206 , 89-3163 , 82-7043-MH, 8S-0031 , 8S-0235-C , 82-2057 , 8S-2265 , 8S-0776 , 82-1957, 8S-0189 , 8S-2535 , 89-0448 , 89-0949 , 89-3372 , 89-0784 , 8S-2432 , 8S-1369 , 82-3872 , 82-2621, 89-2294 , 8S-1967 , 89-0297 , 89-2675 , 82-2633, 82-3564 , 82-7155-MH , 89-2615 , 8S-0413-C , 82-1178, 82-3351 , 89-0980 , 82-3175 , 82-3049 , 82-3473 , 8S-0110 , 8S-2075 , 82-2701 , 89-1465 , 89-2541 , 8S-2297 , 89-0987 , 8S-0807 , 82-1367 , 8S-1130 , 82-1013, 82-3072, 82-7174-MH , 8S-0157, 8S-1556 , 8S-3146 , 89-1550 , 89-1556 , 8S-1723 , 82-0493, 82-3089, 89-0599 89-2877, 89-2987 , 8S-1665 , 82-2403 , 89-2306 , 89-0626 , 8S-2238 , 82-1924 , 82-2975 , 89-0150 , 82-2227, 89-0444, 89-1904 , 82-0345, 82-3536 , 8S-1194-RC , 89-0556, 8S-1309 , 8S-1308 , 8S-2175 , 8S-2596 , 8S-1099 , 8S-2888 , 8S-0153 , 82-0437 , 89-2663 , 82-0317, 82-1037, 82-2331, 82-2699, 82-3643, 89-0878, 89-3226 , 89-2705, 89-2706 , 8S-2993 , 8S-2209 , 89-2628 , 89-1103, 89-3314 , 82-2800, 89-0389 , 8S-1921-RC , 82-0699, 82-1946, 82-3818, 8S-3046 , 82-1461, 89-0317, 89-0490 , 82-0997, 82-3207, 82-3474 , 82-3048, 8S-2407, 8S-2409 , 82-0842, 8S-0708-C, 8S-0901, 8S-1195 , 82-0647, 89-0491 , 82-0412, 8S-0542 , 89-1458 , 82-0977, 82-3562 , 82-0363, 82-1155, 82-1183, 82-1184, 82-2538, 82-3216, 82-3217 , 82-0511, 89-0445, 89-0741, 89-2523, 89-2795 , 82-0185, 89-1143, 89-1936, 89-2410, 8S-0692, 8S-0707, 8S-0745, 8S-0897, 8S , 82-0490, 82-1441, 82-3047, 82-3487 , 82-0158, 82-0335, 82-3791, 89-1266 , 8S-1566-C , 82-0934, 82-1246, 82-1787, 82-7045-MH , 89-0351 , 8S-0096 , 8S-0790-C, 8S-2045, 8S-2101 , 82-0305, 89-1812, 89-2629, 89-2685, 8S-3023, 8S-7014-CMI , 89-0609-RC , 82-0099 , 82-2008 , 89-0082, 89-0083, 89-1912 , 82-0067, 82-1132, 82-3837 , 8S-2232 , 82-3044, 89-2695, 8S-2419 , 8S-1904 , 89-0889 , 89-1680 , 82-2136, 82-2539 , 82-1964, 89-0149, 89-2068 , 89-1463 , 89-1441 , 82-3203, 8S-1605 , 82-0378, 8S-0181 , 89-2313 , 89-2558 , 8S-0095 , 8S-0234-C , 82-2369, 82-3336, 89-0782 89-1064, 89-1070 , 89-2104 , 8S-0137 , 8S-1190 , 89-0452 , 82-3404 , 82-0196, 82-1443, 82-3240, 82-3902, 82-7034-MH, 89-2203, 89-2205, 89-2207 , 8S-1517, 8S-3137 , 82-0165, 89-0558 , 89-2450 , 82-3070 , 89-1179 , 8S-2805 , 82-1107-RC , 89-0451 , 8S-0884 , 82-3191 , 82-0380 , 82-2881 , 82-2602 , 82-2254 , 82-0044 , 82-2175, 89-1732 , 89-0006, 89-2832 , 8S-0689 , 89-2739 , 82-1934 , 82-1940 , 82-1220, 89-1052, 8S-0712 , 82-3553 , 89-1434 , 82-1445, 82-2492, 82-3446, 82-3503, 8S-0256 , 8S-1376, 8S-2687 , 8S-2377-C , 82-1128 , 82-0873, 89-3336 , 82-3068, 82-3071, 89-2630 , 82-1215, 82-1457 , 82-1699 , 82-0508 , 82-1190 , 8S-0168 , 89-0353-RC , 89-0420 , 82-0425, 89-0826, 89-0887, 89-1919, 89-1920, 89-2361, 89-2363, 89-2373 , 89-0041 , 89-3239 , 82-2570 , 89-0106-RC , 89-3538 , 82-2698 , 82-1741, 82-3444 , 82-0523, 82-1868, 82-2712, 82-2714 , 8S-1163 , 82-1949 , 82-2110 , 8S-1641 , 8S-1423 , 8S-1390 , 8S-1698 , 8S-0402 , 82-1069 , 82-1768 , 89-0666, 8S-7145-MH , 82-1738-RC , 89-1435 , 8S-1470 , 8S-2475, 8S-2499 , 89-3215 , 89-1068 , 89-2395 , 82-2647 , 82-2458, 82-3370, 89-3521 , 82-2739 , 8S-1483 , 89-2391 , 82-2434-E, 82-2722, 89-0070, 89-0729, 89-1827 , 89-1813 , 82-1932 , 8S-0405-C , 82-0231, 82-1311 , 82-3162 , 82-2428, 82-3618 , 8S-3168-C , 82-1505, 82-2312, 82-2746, 82-2747 , 82-2314 , 8S-0155 , 82-1320, 89-1891 , 89-3290 , 89-0245 , 8S-1218 , 82-2903 , 89-1860 , 8S-2250-RC , 82-2964, 8S-7179-MH, 8S-7194-MH , 82-3575 , 89-1113, 89-1290 , 8S-2192 , 8S-0100 , 82-3081 , 82-3729 , 89-0991 , 89-1432-C, 8S-2011-C, 8S-2059-C, 8S-2133-C, 8S-2134-C, 8S-2228-C, 8S-258 , 8S-2787-C, 8S-2811-C, 8S-7161-CM , 8S-2749 , 89-2724 , 82-1247, 82-1248, 82-1249, 82-1250, 82-1251, 82-2723, 82-2893 , 89-2661 , 8S-3113 , 8S-0053-RC , 8S-3311 , 8S-2174 , 8S-0877 , 82-3088 , 8S-2562 , 82-2102 , 8S-1447 , 8S-0028 , 82-2026, 82-3855 , 89-1595 , 89-2429, 89-2431 , 89-2444 , 89-2103 , 89-2006 , 82-1706 , 82-0887 , 8S-0144-C , 82-2963 , 89-0904 , 8S-0507 , 82-3447 , 82-7083-MH , 82-2974 , 82-0314, 82-3168 , 82-3524, 89-3089 , 89-3197 , 89-0431, 8S-0365, 8S-1604 , 8S-2883 , 82-0804, 82-2836, 8S-0414 , 8S-1815 , 89-0742 , 89-1100 , 82-0716, 82-1297 , 89-2282 , 89-3246 , 82-0050, 82-0113, 82-3597 , 8S-1936 , 82-2553 , 8S-2699 , 82-1943, 82-2413, 82-3041, 82-3055 , 82-3056 , 82-0787, 8S-0337, 8S-2637 , 82-2306 , 8S-0695 , 8S-3245 , 8S-1801 , 82-2472, 89-0511, 89-1205 , 89-1940 , 89-0130 , 82-1211 , 82-2427 , 8S-2510 , 8S-2061 , 82-2618, 82-3340, 82-3343 , 8S-0917 , 82-2418 , 82-1966 , 82-1416 , 82-0105 , 89-1829 , 8S-3244 , 8S-1090 , 82-0329, 82-1842 , 8S-2030 , 82-2116, 82-2851, 82-2853, 82-2855 89-0752, 89-2446, 89-2448, 8S-2179 , 8S-7097-MH , 8S-2654 , 89-3501 , 82-0980 , 82-2156-RC , 8S-2930 , 82-0514, 82-1004 , 82-2873 , 8S-1616 , 82-0771 , 89-2031 , 82-1697, 82-3420 , 82-1743, 82-3852, 82-3853 , 82-7078-MH , 82-7033-MH , 82-7154-CMH , 82-0603 , 82-1902, 82-3241, 82-3534 8S-1006, 8S-1298 , 8S-2568 , 8S-1233 , 8S-2449 , 89-0918 , 8S-0389 , 8S-0651 , 89-3427 , 89-2590 , 8S-0850-C , 82-3448, 82-3451, 89-2534 89-2645, 89-3013, 8S-2776 , 8S-2779 , 82-0737 , 82-1675 , 82-1798 , 89-1421 , 82-1553 , 89-0622 , 89-0708 , 89-0091 , 82-0095, 89-0487 , 82-2612 , 82-3733 , 89-3066 , 8S-1452 , 8S-0023 , 8S-1777 , 82-0328 , 82-1458, 82-1822, 82-1970 82-2399, 82-3459, 82-3462 , 82-3878 , 89-2874 , 82-3823 , 89-0087-RC, 89-2997, 89-3393 , 89-2225 , 89-2607 , 82-0915, 89-2113 , 82-3169 , 82-1660 , 89-3102 , 82-1654, 8S-0249 , 82-0748, 82-3205 , 82-0045, 82-1470, 82-3425 , 82-1658 , 89-1040 , 8S-0246 , 82-2182 , 82-2846 , 8S-2672 , 89-1250 , 82-0461, 82-3147, 8S-0593 , 8S-0360 , 89-1202 , 89-1060 , 82-0414 , 8S-0671 , 8S-2064 , 8S-2200 , 8S-2544 , 82-0752 , 89-1279 , 89-2995, 8S-1772-C , 82-1527 , 8S-1708 , 8S-2829 , 82-1898, 82-3831, 82-3832 , 8S-0861 , 82-2359, 82-3347, 82-3519, 89-1519, 89-2028 , 82-2284 , 89-0268, 89-2354 , 8S-3177 , 89-2767 , 82-3356 , 82-1782, 82-2330, 8S-2929 , 89-0337 , 8S-2358 , 8S-1651-C , 89-1402 , 82-0364 , 82-0932 , 89-3084, 8S-1157, 8S-2560 , 82-2337 , 82-0263, 82-2947, 82-2958, 8S-2836 , 82-2938 , 8S-2057 , 8S-2717 , 8S-2750 , 82-1704, 89-2332, 8S-0140-C , 8S-2270 , 89-2163 , 82-1749 , 8S-3238 , 82-2892 , 8S-0088 , 8S-2501 , 89-1946 , 82-2046, 82-2049 , 89-2666 , 82-2045 , 89-3282 , 82-2299 , 82-2996 , 89-2579 , 82-2367, 8S-2170, 89-3328 , 89-3452 , 8S-1222 , 8S-1088 , 82-3600 , 89-1285 , 82-2627, 82-2628 , 89-2244 , 89-3512 , 8S-0940 , 8S-2345 , 8S-0078 , 8S-2226-N , 82-0125 , 89-3236 , 89-2348 , 89-2216 , 8S-1168 , 82-1504, 89-2196 , 82-0936, 82-3280 , 82-3572 , 82-1976, 89-0085 , 8S-2760 , 8S-2858 , 8S-0072 , 8S-2676 , 82-3900 , 82-0621, 82-2062, 8S-0361 , 8S-0177, 8S-0494 , 82-0250, 82-3116, 82-3527, 82-3569, 89-0740, 89-1053, 8S-1085 , 82-0559 , 8S-0873, 8S-1214-C-RC , 82-0021, 82-3378, 82-3481, 89-0323, 89-0541-RC, 89-0561 , 82-2777 , 8S-2123 , 8S-1009 , 82-7092-MH, 8S-1945 , 8S-3156 , 8S-0575 , 89-1085 , 82-1615 , 89-0014 , 89-0017 , 89-0350 , 89-0893 , 89-0938 , 89-1257 , 89-1470 , 89-1910 , 89-2989 , 8S-1494 , 8S-2269 , 89-1281 , 89-2567 , 89-1610 , 89-0276 , 8S-1690 , 82-1282-RC , 89-3514 , 8S-0216-RC , 8S-1188 , 8S-1612 , 8S-2362 , 8S-0657 , 89-1837, 89-3395 , 8S-0170 , 89-1437 , 8S-3015 , 8S-0523, 8S-7140-MH , 89-2041 , 89-2052 , 89-2392 , 8S-2768, 8S-3061 , 89-1581, 8S-0071, 8S-0602 , 89-1624 , 82-0108, 82-2770 , 8S-1306 , 82-0780 , 8S-0856 , 89-2279 , 82-7018-MH , 8S-1224-C-RC , 8S-3158 , 82-1810 , 89-1762 , 8S-1387 , 8S-1840 , 8S-0457 , 8S-2736 , 82-0867 , 82-0008-RC , 89-0928 , 89-2261 , 82-3692 , 8S-0713 , 8S-2137 , 89-1766 , 82-0631, 89-0393, 89-3132, 8S-2450 , 89-2099 , 8S-2081 , 8S-0063, 8S-3037 , 82-0244, 82-3159, 8S-1928 , 8S-2893 , 8S-2674 , 82-0340, 82-1187, 89-0044 , 8S-1091 , 82-2509 , 89-2118 , 82-0455, 82-2150 , 89-0286 , 8S-0915, 8S-1632 , 8S-1890-C , 82-7190-MH, 8S-2006 , 8S-2880 , 89-0021 , 82-2307, 8S-0037 , 82-3681 , 8S-1681 , 82-2557 , 82-2363 , 82-2962 , 82-3843 , 89-1003 , 8S-2512 , 8S-1575 , 8S-0292 , 89-0833 , 8S-0787 , 8S-2733 , 82-3011 , 82-1141 , 89-0733 , 89-1069 , 82-3106, 8S-2167 , 82-2585 , 8S-0156 , 8S-0332 , 8S-1171 , 82-7077-MH, 89-1419 , 89-2625 , 89-0554 , 89-0555 , 82-0880, 89-1006, 8S-2178-N , 89-0882, 89-2999 , 89-2743-N , 8S-1232, 8S-1509-C , 8S-2196 , 89-1678, 89-2593 , 89-0988 , 89-3224 , 82-1956, 82-3639, 89-0538 , 8S-0823 , 89-0263, 8S-1129, 8S-1625 , 89-2245, 8S-0007-C , 8S-0762 , 89-0919, 89-3550 , 89-1101 , 82-1859 , 89-0042, 89-0059 , 89-0537 , 89-1293, 89-1836, 8S-2556, 8S-2558, 8S-2559 , 89-1986, 89-1987, 8S-1642 , 89-2505, 89-3546 , 89-2667 , 89-0271, 89-1327, 8S-1926 , 82-1051, 82-2978 , 89-0022 , 82-3549 , 89-1201, 89-3559 , 89-3319 , 89-2277 , 89-1699 , 89-1661 , 82-3650 , 89-2785 , 82-0497, 82-1656, 8S-0057, 8S-1051, 8S-3267 , 82-0811, 82-1318, 82-7072-MH, 82-7125-MH 89-0690, 89-0691, 89-2350 , 89-2351 , 82-1467, 82-3734, 89-0646 , 82-3102, 89-1889 , 82-0462, 82-0528, 82-1280, 82-3276, 89-2413 , 8S-0738 , 82-0069, 82-1540, 89-1478 , 82-1836, 82-2593, 82-3085, 82-3310 , 82-3083, 89-0402 , 82-3433 , 8S-1630 , 82-1057-RC, 82-3491 , 82-3545, 8S-0726, 8S-2808 , 82-0846, 82-2263 , 8S-2615 , 8S-7048-MH , 82-0194 , 89-2030 , 82-3255, 8S-2278 , 82-0786, 82-2567, 82-3443 , 82-3690 , 89-1652 , 82-3608, 89-0377 , 8S-2122 , 8S-3240 , 8S-0421-C , 8S-2100 , 82-1298, 82-2905 , 89-2157 , 89-1056 , 82-0023-RC, 82-2247, 82-7076-MH , 89-0687 , 82-0689 , 8S-2291 , 82-0138-RC , 82-1196-RC, 89-1877 , 8S-1237, 8S-1757 , 89-1406 , 89-2668, 8S-1557 , 82-1258, 89-0978 , 89-0620 , 8S-3281 , 82-0410, 82-0838, 82-0947, 82-0948, 8S-0187 , 8S-0192-C , 8S-1508-N , 8S-2896 , 82-1099, 8S-0669, 8S-1439, 8S-1884, 8S-7147-MH , 8S-7119-MH , 8S-1544 , 82-1780 , 82-0318, 82-3742, 82-7022-MH , 8S-1491-C , 8S-2194 , 89-0463 , 89-1543 , 89-0846 , 8S-2801 , 8S-1836 , 8S-2466 , 8S-0650, 8S-1197 , 8S-0237 , 82-2815 , 89-2642 , 82-1804 , 8S-1238 , 8S-0537 , 8S-0331 , 82-3397 , 82-1848 , 8S-2728 , 82-0703 , 8S-0032 , 8S-2448 , 89-0174 , 89-0603 , 89-2588 , 89-2734 , 8S-0946 , 82-0486 , 82-0926, 82-1218-RC , 82-3303, 89-1801 , 82-0436, 82-3596 , 89-1263 , 82-2327, 8S-0330, 8S-0648 , 82-2448, 89-0422, 89-1094, 89-1508 , 89-3032, 89-3033 , 82-1109 , 82-0130 , 82-1886 , 82-0397, 89-0876 , 82-7124-MH , 89-0930, 89-0940 , 8S-1861 , 82-0704, 82-2803 , 8S-1688 , 82-0098, 82-2729, 89-1080, 89-2772 , 8S-2959 , 89-0089, 8S-0774 , 8S-0142-RC , 89-1379 , 89-0164 , 82-0372 , 89-0998 , 89-2555 , 82-3298 , 82-2397 , 8S-0829 , 82-0323, 82-3323 , 89-1751 , 8S-1817 , 82-2275, 82-2837 , 82-3718, 8S-0848, 8S-1964 , 8S-2204, 8S-2427 , 89-1162-RC , 82-0443 , 82-3251, 8S-0702 , 82-1102 , 82-1790, 89-0063 , 82-3062 , 8S-2498 , 89-0113 , 82-0417, 82-2362, 8S-3845 , 89-0825 , 8S-2114 , 82-0460, 82-1612 , 82-2060, 82-3176, 82-3179 , 82-0357, 8S-1705 , 89-3547 , 82-1593 , 82-0733, 82-1130 , 82-2865-RC, 82-3631 , 82-2945 , 82-0344 , 82-2569, 8S-0690, 8S-1519 , 8S-2594 , 82-3688 , 89-1814 , 82-0679, 82-2631, 89-0569 , 89-0453 , 82-1770, 89-1022, 89-1365, 8S-1748 , 89-0712 , 8S-0775 , 82-3921, 82-3921, 89-3578 , 89-2879, 8S-2438 , 8S-0373 , 82-0199, 89-1808 , 89-3104 , 82-0184, 89-1277, 89-2939 , 89-2480 , 82-2482 , 82-2789 , 8S-0919 , 89-3171 , 89-2304 , 82-2953 , 82-3642 , 82-0957 , 82-1521 , 82-1532 , 89-2109 , 82-2786 , 8S-0039, 8S-2632 , 8S-2172 , 82-1599-RC, 82-3053 , 82-3827 , 82-7064-MH , 89-3412 , 89-2621 , 8S-0876 , 8S-1832 , 89-1469 , 8S-1065-C , 89-1968 , 89-0208-RC , 89-0289 , 82-3348 , 8S-0486 , 89-3191 , 82-1144 , 82-3740 , 82-2348, 82-3266, 89-0318 , 89-1227 , 8S-0871, 8S-2264 , 82-1321, 8S-0833, 8S-1618, 8S-2693, 8S-2945 , 82-1905 , 8S-1292 , 89-0695, 89-0935 , 82-1356 , 8S-0579, 8S-1058 , 8S-0653 , 82-1456, 82-2302, 89-1089, 89-2409, 89-2755, 8S-0749 , 8S-1835 , 89-2408 , 8S-1788 , 82-0024, 82-1319, 8S-1013-C , 82-0680 , 82-3400, 82-3622, 89-2464, 8S-1427 , 82-2186 , 82-2339 , 82-3082, 89-1156 , 82-0391, 82-2237, 82-3456 , 89-0592, 8S-1017 , 82-0557 , 82-2134 , 82-0228, 82-1510, 82-3161, 8S-1004 , 82-0777, 89-0719, 89-2626 , 8S-0214 , 8S-2861 , 8S-1466 , 89-3459 , 8S-1372 , 89-2359, 8S-0938, 82-2588 , 89-0236 , 82-2717 , 8S-0784 , 89-0730, 89-1953, 89-3527 , 89-1012 , 8S-0736 , 89-1019 , 8S-2661 , 82-1046, 82-1538 , 89-0030 , 89-2331 , 8S-0664 , 82-2940 , 8S-3043 , 82-0666, 82-0667, 82-7173-MH , 8S-7125-MH , 89-0199, 89-0954-RC , 89-0597, 89-2515 , 82-1890 , 8S-3117 , 8S-2069 , 89-0116 , 8S-0385, 8S-2056 , 89-0993 , 82-1601, 89-0124, 89-1736, 89-2836, 8S-0134 , 82-1462 , 82-0310, 82-2266, 89-0943 8S-0843, 8S-1100 , 8S-7071-MH , 82-2470 , 82-0556, 8S-0255, 8S-1563 , 82-1452 , 82-1695 , 89-3225 , 89-1401 , 82-2370 , 82-0781 , 82-1684 , 82-0419, 82-1745, 89-0648 , 8S-0304 , 82-3607, 8S-2666 , 82-0792 , 82-0714 , 89-0334-RC , 82-3632, 8S-0412, 8S-2897 , 89-1854 , 8S-3219 , 82-0365 , 8S-2605 , 8S-2091 , 8S-1276 , 8S-1086 , 8S-2777 , 82-0034, 89-0864 , 8S-1903 , 8S-2424 , 82-1700 , 8S-1597 , 82-0911 , 89-3025 , 8S-2626 , 82-0916, 82-2657 , 8S-2966 , 82-2661 , 8S-2914 , 8S-1598 , 82-3551, 82-7181-MH, 8S-0530 , 8S-1880-N , 8S-2304 , 82-0562 , 82-7118-MH , 8S-1862 , 82-3644 , 82-0870, 82-1427, 82-1807, 82-2649 , 8S-0890 , 82-0629, 82-3390, 89-0296 , 82-1590, 82-1667, 8S, 2667-RC, 8S-2668-RC , 8S-1003 , 8S-1149, 8S-2181 , 89-1148 , 89-3194 , 8S-2248 , 82-1039 , 82-0541, 89-0468 , 89-0469 , 8S-3029 , 8S-0952 , 8S-1331 , 8S-0065 , 8S-2235 , 82-2088 , 82-3461 , 82-3458 , 82-2721 , 89-0868, 8S-1500 , 82-2840 , 8S-1502 , 8S-1726 , 82-2222 , 82-0984, 82-2414, 8S-0334 , 89-1644 , 89-1573 , 82-7096-MH , 82-0278, 82-3561 , 8S-0849-C , 89-0539 , 82-2379-RFB, 89-0352, 89-1943, 8S-1551-RC , 8S-2494 , 82-1053 , 82-0354 , 8S-1185 , 89-0123, 8S-1887 , 89-3496 , 82-0220, 82-2255, 82-3335 , 8S-0484-RC , 8S-2369, 8S-3106 , 82-1338 , 82-0230 , 82-7167-MH , 89-0224 , 89-3284 , 82-1227, 82-1717, 89-1964 , 82-0998, 82-2085, 8S-1589 , 8S-1569 , 8S-1183 , 8S-1541 , 82-1089 , 8S-0094 , 8S-0971, 8S-2950 , 8S-0678 , 89-1448, 89-2356 , 8S-2210 , 8S-0782, 8S-2249 , 82-3183 , 89-1137 , 8S-3024 , 82-2345 , 8S-1677 , 89-0739 , 82-0160 , 89-2466, 89-2602, 89-2612, 89-3012, 89-3015 , 8S-0005-C , 89-0137, 89-0753 , 89-2213 , 82-0832, 82-2012 , 82-3396 , 8S-2163 , 82-3332 , 82-2318 , 8S-3204 , 89-1499 , 8S-2087 , 8S-1388 , 89-2472 , 8S-2421 , 89-2757 , 89-2643 , 8S-1211 , 8S-3155 , 8S-1484 , 89-0441 , 8S-0171 , 89-1440, 89-2866 , 89-2330 , 82-2929, 89-1054 , 8S-0411 , 8S-1565 , 82-2262 , 82-3594 , 89-0256-RC , 89-0207 , 82-0658 , 8S-2763 , 8S-3291 , 82-0908 , 8S-0148-C , 89-0430 , 8S-3227 , 82-0527, 82-2111 , 89-3397 , 82-1111, 82-1979 , 82-3234 , 82-3614 , 82-0093, 82-3221, 89-1354 , 89-1730-RC , 82-2705, 89-1587, 89-3304 , 8S-1588 , 82-3357, 89-1783 , 82-1274, 89-1076 , 82-0564, 89-0120, 89-2174-N , 8S-1966, 8S-2035, 8S-2791-C , 82-2119 , 89-1126 , 8S-1394-C , 82-1152 , 82-2847, 89-2451, 8S-2998 , 82-3139 , 8S-0056 , 89-0018 , 82-1630 , 82-3260 , 82-0047 , 82-3476 , 82-7105-MH, 89-1750 , 89-1862 , 82-2525 , 89-0734 , 8S-2892 , 8S-2119 , 8S-2614 , 82-3275 , 8S-3010 , 82-1077 , 8S-2716 , 82-1316, 82-1925, 82-3143 , 82-2204, 82-2763 , 82-0973, 82-1450, 8S-2541, 8S-2542 , 89-3212 , 89-2069 , 89-3403 , 8S-0077 , 8S-2659 , 89-1021 , 82-0991, 82-3460, 89-0850 , 82-0472, 82-1451, 82-3226, 8S-0108 , 82-0544, 89-1197 , 89-2037 , 8S-1793, 8S-2600, 8S-3017 , 82-2616 , 8S-2467 , 8S-2470 , 8S-1377-C , 8S-0719 , 89-1122 , 8S-2293 , 89-1454, 89-1459, 89-3441, 8S-1126 , 8S-1547 , 82-0878, 82-3059 , 82-0772, 82-2274, 8S-1975 , 8S-2058 , 82-1414, 8S-2094-N , 8S-2113 , 8S-0059 , 89-3418 , 82-2603 , 8S-3016 , 8S-0990 , 82-1874 , 8S-2103, 8S-2106 , 8S-2107 , 82-0386 , 82-1664 , 8S-2105 , 8S-1386 , 82-2264, 89-3431 , 82-0985, 82-0986, 82-1303 , 82-3297 , 89-0467 , 89-1057 , 89-0257 , 89-1272 , 82-1018, 82-7132-MH, 89-0282 89-1326, 89-2477, 8S-7089-CMH, 89-3316 , 82-1261 , 82-3194 , 8S-2796 , 82-2995 , 89-0929 , 82-0549, 82-2769 , 8S-1957 , 82-1744 , 89-1452 , 82-3555 , 82-0290, 82-1832, 82-2248, 82-3152 82-3557, 82-3558, 82-3559 , 89-0495 , 82-3121 , 8S-1052 , 8S-3191 , 8S-2623 , 89-1586, 89-1653 , 8S-0781-C , 89-1695 , 89-1240, 8S-2921 , 82-0191, 82-0319-RC , 82-0187 , 82-3171 , 89-1588 , 8S-1215-RC , 82-3439 , 8S-2692 , 8S-1101 , 82-0136-RC , 8S-2416, 8S-2574 , 89-2502 , 82-1889 , 82-2309 , 89-1411 , 89-1971 , 82-2522 , 89-1451 , 8S-1291 , 8S-2332 , 82-1792, 82-3880, 89-1521, 89-2414 89-2777, 89-2779, 89-3487, 8S-1337 8S-3054, 89-3356 , 8S-1582 , 8S-2331 , 89-1412, 8S-1146 , 89-1623 , 89-3052 , 89-3234 , 89-1370, 89-2378 , 89-2792 , 89-2481 , 89-0455, 89-1883, 89-2497, 89-2499, 89-3097, 89-3291 , 82-0377-RC, 82-3501, 82-3745 , 8S-2826 , 89-2075 , 89-2521 , 82-0423, 82-3703, 89-0461 , 89-1965, 89-3180 , 89-1805 , 89-0156 , 89-1338 , 8S-2495 , 82-0325 , 89-2236 , 89-0020 , 82-1967 , 82-0859, 82-3218 , 89-3220 , 8S-0584 , 82-2191, 89-2401 , 8S-0286 , 8S-1098 , 82-2460, 89-1868, 8S-1804 , 8S-0448 , 8S-2354 , 8S-2454 , 89-0849 , 82-2930 , 82-0901, 82-3349, 8S-0348-C-RC , 89-0675 , 8S-2932 , 8S-3065 , 82-0090, 82-1758, 82-2982, 82-3828, 89-0376, 89-0814, 89-2402 , 89-1111 , 82-0801, 89-1679, 89-2453-N , 82-1145, 82-1724, 89-1893, 89-1923 , 89-2519 , 89-1319 , 89-2082 , 89-3175 , 82-3382, 8S-2108 , 89-1957 , 89-1649 , 82-1728 , 89-2073, 89-3121-N, 89-3404 , 82-2961 , 82-1958 , 8S-3033 , 89-1274 , 8S-0510-C-R, 8S-3057 , 89-3264 , 8S-0188 , 89-0834 , 82-1254, 82-3435, 8S-0161 , 89-1992, 89-2202, 89-3357 , 82-3033, 82-3353 , 82-2480 , 89-2853 , 8S-2618 , 82-0620 , 82-7141-MH , 82-1153, 8S-0589, 8S-2051 , 8S-1624 , 82-2446 , 8S-2092 , 82-3142 , 82-0336, 89-1139 , 89-2797 , 8S-0297-C , 82-3480 , 8S-1751 , 89-2353 , 8S-3192 , 8S-3050 , 82-7166-MH , 89-3168 , 8S-1336 , 89-1990, 8S-0666 , 8S-0585 , 8S-1531 , 89-0330 , 89-1074 , 89-3490 , 89-3165 , 82-1101, 82-3543 , 8S-2917 , 82-2645 , 89-3229 , 82-0758, 82-2599, 8S-1942-N, 8S-2300 , 89-2839 , 82-0794, 8S-2361 , 89-0379 , 82-0240 , 82-1075 , 8S-2368 , 82-0526, 82-1935 , 8S-2644 , 89-2876 , 89-3002 , 82-2862 , 89-0588 , 8S-2802 , 89-1120, 89-1861, 89-3167 , 89-1163 , 8S-0386 , 82-1720 , 8S-3285 , 82-1435-RC , 8S-1008, 89-3377 , 82-1078, 82-3624 , 82-2181 , 8S-3269-H, 8S-3270-H, 8S-3271-H , 8S-3272-H , 82-0107, 82-0132, 82-1108, 82-2160-N, 82-2278, 82-2371, 82-2907, 89-0549, 8 , 82-2280 , 82-3402 , 82-0570 , 89-0530 , 82-0429, 82-2106, 89-1650 , 82-1802, 89-1522 , 82-1803 , 8S-1192 , 8S-2079-C-RC , 82-3359 , 89-0034 , 8S-2187 , 89-0184 , 82-1260, 89-3375 , 8S-2068 , 89-1384 , 82-1993, 89-0356, 89-3259, 89-3260 , 8S-7144-MH , 89-1193 , 89-2781 , 82-0776, 89-0653, 8S-0394 , 8S-0395 , 82-3374 , 82-2485, 89-3339 , 8S-1317 , 89-3107 , 89-0343 , 89-1606 , 82-0275, 82-3224, 89-0997, 89-1851 , 8S-2295 , 89-1843, 89-2780 , 8S-0691 , 82-0518 , 89-3087 , 89-1925 , 89-3211 , 8S-0553 , 8S-0554 , 8S-0796 , 8S-1416 , 82-1887, 89-1629 , 89-1175 , 89-1928 , 89-2846 , 82-2906 , 89-2084 , 82-3830 , 82-1255, 82-3104, 82-3109, 89-1173, 89-1311, 89-1315, 89-1938, 8S-0785 , 89-3474 , 82-0552 , 82-2574-N , 89-0573, 8S-2885 , 82-3149, 8S-2459-C , 89-0623 , 8S-2513 , 89-1237 , 89-3090 , 82-3482 , 82-0101 , 82-1580-RC , 82-0161, 82-2017, 82-3173, 8S-2490 , 82-0504, 82-2716 , 89-0492, 89-2142 , 82-2757 , 82-3730, 89-0374, 89-1944, 89-1952 , 8S-1259 , 82-7005-MD , 8S-0240 , 89-0147 , 89-2275 , 89-3438 , 8S-0950 , 8S-0677 , 89-1731, 82-0672, 82-3532 , 82-7114-MH , 82-1678 , 82-0577, 89-3439 , 82-0708, 89-0589 , 8S-2431 , 82-0734, 82-3403 , 82-1680, 82-2404, 8S-7053-CMH , 82-2928, 82-3661 , 8S-2240 , 82-0513 , 89-1704 , 89-0720 , 89-2027, 89-3133 , 89-3139 , 82-0215, 82-1189, 89-1753, 8S-2609 , 82-1092, 89-1884 , 82-3122, 8S-0343 , 82-7019-MH , 89-0979 , 82-3612 , 8S-7036-CMH , 8S-1617 , 8S-0288, 8S-2982 , 82-3898 , 89-0181 , 82-0993, 82-1829, 8S-0069-C , 82-1739, 82-2927 , 82-1777 , 82-1436 , 8S-2074 , 82-1105 , 82-3022 , 82-3665, 82-3685, 89-0840 89-0841, 89-1866 , 89-1872 , 82-1330 , 89-2568 , 82-1542, 89-0077 , 8S-2986 , 8S-2219 , 89-1634 , 82-0484, 82-2732 , 82-0954 , 89-0423 , 82-0179, 82-1773, 82-2419 82-2422, 82-3232 , 82-3385 , 8S-1468 , 89-1562-N , 82-0533 , 89-1110 , 89-1177 , 8S-1883 , 89-3576 , 8S-2282 , 89-0183, 89-0602 , 8S-1459 , 82-1914, 82-3174, 82-3901 , 82-0170, 82-1464, 82-2324 , 82-7053-MH .8S-2918 , 8S-3282 , 8S-3283 , 89-0975 , 8S-1822 , 89-1798 , 89-0016 , 82-2527, 8S-0513, 8S-2597 , 8S-1916 , 8S-2785 , 89-1339 , 82-2636 , 89-0190 , 89-3295 , 82-7026-MH, 89-0697, 8S-2337-N, 8S-2341-N , 82-3712, 89-0027, 89-0028, 8S-0605 , 82-0019 , 8S-1591, 8S-1594 , 8S-2277 , 8S-2458, 8S-2602 , 8S-1535 , 89-3202, 8S-1889, 8S-1891 , 89-3196, 89-3205 , 82-3662 , 82-0788 , 82-0955-C-R, 8S-7021-CMH , 8S-1710 , 89-0336, 89-3077 , 82-0976, 82-3918 , 89-0043 , 8S-2827 , 89-1557 , 89-2707 , 82-3238, 89-3129, 89-3353 , 89-1976 , 82-0597, 89-0869, 89-0877, 89-1958, 8S-1643 , 89-2471 , 82-0512, 82-2084 , 82-3912 , 8S-2743 , 82-1151, 82-2024, 82-3387, 82-3824, 89-2488 , 89-0165, 89-1536 , 82-1150 , 8S-2682 , 8S-1787 , 89-0692 , 82-0446, 82-0447, 82-1278, 82-1279, 82-3058, 8S-1350-C , 82-0063, 82-2691 , 82-2445 , 82-1778 , 82-0128 , 89-0805-RC , 89-1116 , 82-2245, 8S-0859 , 89-1481 , 82-1324 , 82-2200 , 8S-2697 , 82-0683 , 82-1962 , 82-2437, 82-2784 , 89-0485-RC , 8S-3264 , 8S-0968 , 8S-1830 , 82-1525, 82-2360 , 82-2922 , 82-2288, 8S-7172-CMH , 82-3273, 82-3590, 8S-2463, 8S-2538 , 82-0999, 82-1194, 82-3021 , 89-2609 , 8S-2891 , 82-0542, 82-3445 , 89-0498 , 89-1180 , 82-0945, 82-2909 , 89-1708 , 8S-0596-C-RC , 82-3660, 8S-0634, 8S-1045 , 8S-2536 , 82-2352, 82-3004 , 82-3066 , 82-3601 , 8S-2634, 8S-3028 , 8S-0047 , 82-0697 , 82-0989, 8S-2201 , 82-0540, 82-2484, 82-3300 , 82-3301 , 89-0718 , 89-1114 , 89-0995 , 89-2100 , 82-0931, 82-1027, 82-1923, 82-2317 , 8S-2052 , 89-0557, 89-0707 , 89-3071 , 89-0832, 8S-1203 , 8S-3140 , 89-1564-RC , 89-1960 , 8S-1615 , 89-2011 , 8S-2314 , 82-3655, 8S-0128 , 8S-1397 , 8S-2860 , 8S-1498 , 82-0619-M, 89-2489 , 82-0580 , 8S-1167, 8S-1425 , 8S-2186 , 82-2402 , 89-2458 , 8S-0415 , 8S-0103 , 8S-1847 , 89-2770 , 89-0425, 8S-0667 , 89-1187 , 89-1190 , 8S-7137-MH , 8S-0949 , 89-1683 , 8S-3041 , 89-1780 , 8S-1122 , 89-0732 , 8S-2296 , 89-0662 , 8S-0409-RC , 8S-1430 , 8S-0324 , 89-1109 , 8S-0215 , 8S-0604 , 89-0136 , 8S-0744-RC, 8S-1198 , 8S-0992 , 89-1567 , 89-1425 , 8S-0358 , 8S-1036 , 89-0971 , 89-0999 , 8S-0724 , 89-0266 , 82-1496, 89-1414-RC, 89-2704, 89-2299, 89-2443, 89-2456 , 82-3002 , 8S-0154 , 82-2600, 82-3857 , 82-3001 , 82-2408, 82-2429, 82-2430, 82-2432, 82-2433, 8S-1758, 8S-1771, 8S-1773, 8S , 89-0831 , 89-2985 , 82-0176-RC, 82-3899, 89-3524 , 82-0157, 82-2797 , 89-1913 , 82-2356 , 89-1352, 89-1362, 89-2047 , 89-2858 , 89-1387 , 82-0294, 82-0487 , 89-0122 , 82-2391, 89-0786 , 89-2119 , 8S-1976 , 89-1576 , 89-0465 , 82-1688, 82-2384, 82-2385, 82-3599 , 8S-0265 , 82-0495, 82-2656 , 82-0753, 82-2646, 8S-7087-CMH , 82-2147 , 82-1755 , 8S-2821 , 8S-3012 , 82-0888, 82-1512, 82-2029, 82-2486, 89-0241, 89-2135, 89-2138 , 89-1942, 89-3120 , 82-2895 , 8S-2274 , 82-1147 , 82-3923 , 8S-0945 , 89-2298 , 8S-1456 , 82-2041, 82-7065-MH , 82-0109, 82-3869 , 82-3341, 82-7095-MH , 82-0424 , 89-3391 , 8S-1482 , 8S-3002 , 82-3028 , 8S-2126 , 89-0804-RC, 89-2188 , 89-0621, 8S-1475 , 82-1945 , 82-1328 , 8S-2552 , 82-0573, 82-3006 , 8S-3131 , 89-1079, 8S-0058 , 89-0963 , 82-0820, 82-3506, 89-1013, 89-2123, 89-2124, 8S-1156 , 8S-1587 , 82-1648, 82-7140-MH , 82-2203 , 82-3802 , 89-2469, 89-2516 ,89-3545 , 82-2694 , 89-0038-RC , 89-1246 , 89-1473 , 89-3014 , 8S-0551 , 8S-1984 , 82-2176, 89-3346 , 82-2838 , 89-2000 , 89-0163, 89-2587 , 89-2592 , 82-0988 , 89-0552 , 89-0647 , 89-0925, 89-3294 , 82-0203 , 82-0016 , 82-3188-N , 82-0975 , 82-1844 , 82-1854 , 8S-1824, 8S-2877 , 8S-2976 , 82-0929, 82-1709, 82-2165, 82-2358, 82-3064, 82-3319, 82-3383, 89-0111 , 82-0987, 82-2035, 89-1899 , 82-0200, 82-1471, 82-2650, 82-2651 , 89-0099 , 8S-3036 , 89-2096 , 82-0264, 82-2692 , 82-3613, 89-3250 , 82-0516, 89-2791 , 82-2791 , 82-7068-MH, 89-0270 , 82-7126-MH, 89-0015 , 89-1779 , 8S-2488 , 8S-2489 , 82-2808 , 82-0601 , 82-0322 , 8S-0750 , 89-1119 , 89-2102 , 89-2224, 89-2312 , 89-3183 , 82-3024 , 8S-1943 , 89-0427 , 8S-0401 , 82-3630 , 89-2368, 89-2483, 89-3108 , 89-0066 , 82-0140, 82-0700, 82-2236, 82-2238, 89-1969 , 82-1185, 89-1001 , 82-7037-MH , 82-7044-MH, 89-0060 , 8S-1985 , 8S-1490-C , 82-3464 , 82-3466 , 89-1230 , 89-1974 , 89-0026, 89-0886, 89-2440 , 89-1517 , 8S-3066, 8S-2669 , 82-3919 , 8S-0491 , 89-0228, 8S-0973 , 89-2436 , 82-1350 , 89-3138 , 82-7028-MH , 89-0293, 89-1507 , 89-0851, 89-1592, 8S-2260 , 89-1447 , 8S-1142 , 82-3287, 89-0138 , 8S-1737 , 8S-3274-CH , 89-1091, 89-1092 , 82-1919 , 82-2108, 82-2109 , 82-0715, 8S-1043 , 8S-2635, 8S-3181 , 8S-2063 , 8S-1040 , 8S-2756 , 89-1090 , 82-3135, 89-2708 , 82-7170-MH , 8S-0268 , 8S-1107 , 82-2034 , 8S-2176 , 8S-1166 , 82-2146, 89-0661 , 8S-2222 , 8S-2009-N , 8S-2029 , 8S-1534 , 82-0547 , 89-2094 , 89-0750 , 8S-0033 , 89-1271 , 8S-0091 , 8S-2193 , 8S-2118 , 89-0961 , 8S-2070 , 89-0134, 8S-2824 , 8S-2645, 8S-2890 , 8S-1068 , 89-3213 , 8S-0010, 8S-2726, 8S-3003 , 82-1276 , 8S-0436 , 89-0088 , 8S-3138 , 89-0885 , 8S-0207-C, 8S-0208-C , 8S-1574 , 89-1816 , 89-2864 , 8S-2631 , 8S-2085, 8S-2093 , 8S-2973 , 8S-2136 , 89-2838, 89-2840 , 8S-0472 , 89-2169 , 89-1026 , 8S-1752 , 82-1992 , 82-7145-MH , 89-2384 , 89-2388 , 82-1516, 82-3076, 8S-0670 , 89-1265 , 8S-1660-H , 8S-2764 , 89-0237, 8S-0488 , 8S-1138 , 82-0233 , 8S-1845-RC , 8S-2180 , 82-1746-RC , 89-0227 , 8S-0623 , 82-0311, 82-0500, 82-1036,82-2882, 82-3014 , 82-1296 , 82-1003, 82-7143-MH, 89-0532, 89-0870, 89-1538, 89-1540, 8S-0015-C , 82-1054, 8S-0661 , 82-2123, 89-2437 , 8S-1952 , 82-1301 , 8S-1462 , 8S-0720 , 82-2202 , 82-2795 , 82-2720 , 82-0641, 82-1228, 82-1314, 82-1439, 8S-3187 , 89-0068 , 89-2072 , 89-2504 , 89-2511 , 89-0223 , 89-3217 , 89-0460 , 8S-2653 , 82-3074, 89-0109 , 89-1457 , 89-1398 , 8S-1360 , 82-2799 , 82-2736 , 89-0644 , 82-1847 , 8S-1778 , 82-2548, 8S-2710 , 82-0724 , 82-1572 , 82-2265 , 82-1573, 89-3241 , 89-1445, 89-3179 , 82-0346, 82-0712, 82-2558 , 82-3887 , 89-0715 , 82-0718, 82-0952, 82-2767, 89-2098, 8S-0114, 8S-0993, 8S-1881 , 82-1559, 82-3178 , 89-3222 , 82-0326, 82-1156 , 82-1008 , 89-1317 , 82-3841 , 82-1023, 82-1410 , 82-2066, 82-2223, 8S-7075-CMH , 82-0884 , 82-1513 , 89-0239 , 82-0925, 8S-0024 , 82-3591, 8S-0501 , 82-1857, 82-1858, 89-1578 , 82-1465 , 82-1687 , 82-0061, 82-2479, 82-2842, 82-7082-MH , 8S-2348 , 89-1427 , 82-0393, 89-1329, 8S-0524 8S-1244, 8S-1865 , 8S-2371 , 82-1154, 89-2614 , 89-0615, 89-3530 , 82-1846, 89-0220 , 89-3297, 89-3358 , 82-1444 , 89-1301 , 8S-1112, 8S-2831 , 89-3068 , 8S-2189 , 82-2528, 8S-0085, 8S-1150 , 8S-2539, 8S-3218 , 82-1881 , 82-1604, 82-2643 , 89-0830-RC, 89-2375 , 89-0275, 89-0827, 8S-2245 , 82-2341 , 82-2994 , 82-2511, 82-3215, 82-3318 , 82-2597, 8S-3081 , 82-2742 , 82-2911, 8S-1956, 8S-2384, 8S-2934 , 82-3077, 8S-1961 , 82-3522, 82-7115-MH , 82-3884, 8S-3266 , 82-7184-MH, 8S-1954 , 8S-1960, 8S-1968, 8S-2383 , 8S-2090, 8S-2374 , 8S-3114, 8S-3118, 8S-3134, 8S-3235 , 89-3209, 8S-1001, 8S-2563 , 82-2811 , 8S-1971 , 82-3542 , 89-3457 , 8S-2983 , 89-2399 , 82-2601, 8S-2628 , 89-2008 , 82-2568, 8S-2387, 8S-2399, 8S-2412, 8S-2537, 8S-2601 , 82-2863, 82-3054, 82-3063, 89-2659, 8S-1962 , 82-3065, 8S-1965 , 82-3090, 8S-2636 , 82-3479, 8S-2373 , 8S-1948, 8S-2375 , 8S-1969, 8S-3125 , 82-3042, 82-3163, 8S-1963 , 82-1903 , 8S-2372 , 8S-2529, 8S-3108, 8S-3126 , 89-1809 , 8S-2202 , 8S-1389, 8S-2755 , 82-0432 , 82-1708, 89-0619 , 82-7027-MH, 89-0206 , 89-1756, 89-3552, 89-0815,89-1749, 89-0054, 89-0968, 89-0726, 89-1902, 89-2 , 89-3433, 89-0100, 89-0179, 89-1264, 89-1847, 89-2514, 89-3352 , 89-3004 , 8S-0262 , 82-0651 , 82-1415 , 89-0049 , 8S-0493-RC , 89-1134 , 82-2642, 8S-0487 , 8S-1816 , 89-1895 , 82-0899 , 8S-0489 , 8S-1827 , 82-0114, 82-0862-RC, 82-1671, 82-3399, 82-3687, 8S-1740 , 89-0005-RC , 8S-0754-C , 8S-2729 , 8S-7047-MH , 82-0060 , 82-1072 , 8S-0660 , 8S-2942 , 89-0812 , 82-0617, 89-1530, 89-2012, 89-2260-N, 89-3263 , 8S-2080 , 89-3543 , 82-0698, 82-1633 , 89-0700 , 8S-0107 , 82-1668 , 8S-0026 , 89-2231 , 89-3463 , 82-0757 , 82-1121 , 82-0742 , 89-1853 , 82-3743 , 82-2018 , 82-1566, 8S-1120 , 8S-2745 , 89-0985 , 8S-1667, 8S-1938 , 8S-2239 , 8S-1977 , 82-1017, 89-2223 , 89-0737 , 8S-3315 , 82-2999 , 8S-0983, 8S-1067 , 89-2674 , 89-1793, 8S-3214 , 89-2741 , 82-3546 , 82-0031-C, 82-2226, 82-3739, 89-3153 , 8S-1584 , 82-1560, 8S-2313 , 89-0696, 8S-0450 , 89-1959 , 8S-1927 , 82-3736, 8S-0118, 8S-1712 , 8S-1115 , 89-1107, 89-1108, 89-1785, 89-3216, 89-3219 , 82-1000 , 89-3156 , 8S-0347 , 82-1930, 82-3856 , 82-3101 , 89-1811, 89-3430 , 89-1926-RC, 8S-0662-RC, 8S-1145-N , 8S-1609 , 82-0251, 82-3079, 8S-0070 , 8S-2254 , 82-0032, 82-1313, 82-3132 , 89-0839 , 89-1081 , 8S-0974 , 89-2647, 89-3035 , 82-2771-N, 8S-1027 , 89-1306 , 82-1210, 82-2139 , 82-0453-RC, 82-1771, 82-2949, 82-2951 , 89-1667 , 8S-0767 , 89-0499 , 82-1825 , 89-1870 , 89-1065, 89-3509 , 82-2323 , 82-0381, 82-2138, 82-2697 , 82-0560, 82-1412, 82-1963, 8S-0173 , 89-1799, 89-2684, 89-2789 , 82-0962, 8S-0124 , 8S-2527 , 89-0544 , 8S-1716 , 8S-1923 , 8S-1747 , 89-0577 , 8S-1935 , 8S-0193-RC , 82-0662-RC , 82-2589, 82-3844 , 82-1020-RC , 8S-2702, 8S-1717 , 82-1730 , 82-1788, 82-3256, 89-2677, 8S-1162 , 82-3427 , 82-0584, 8S-2335 , 8S-0868 , 82-1998 , 8S-0943 , 89-0058, 89-2215 , 89-0548 , 8S-0562 , 82-2030 , 82-2805 , 82-0309, 89-2248 , 82-0426, 8S-2447 , 82-2586 , 89-3534 , 82-0600 , 8S-0079, 8S-1733 , 8S-3116 , 8S-2912 , 8S-1294 , 89-2034 , 89-0504, 89-1744 , 8S-2651, 8S-2679, 8S-3004 , 89-1980 , 8S-2301 , 82-1484 , 89-2273 , 89-1393 , 82-1585 , 82-3516-E , 82-0287 , 82-0676, 82-1306, 82-3009 , 89-0265 , 8S-2322 , 82-1430, 82-2925 , 82-2550 , 8S-0788, 8S-2573 , 8S-2971 , 82-1182, 82-2334, 82-3576 , 82-0039, 82-3258 , 82-7164-MH , 89-1018, 8S-0267 , 82-7165-MH, 82-0610, 82-7186-MH, 89-2852 , 82-0375, 89-1007, 8S-1199 , 82-2708-N , 82-0374, 82-3345, 8S-2401 , 82-2105 , 82-0388, 89-2209, 8S-0396-C , 82-2005 , 82-0755, 82-2077, 82-7039-MH, 82-7107-MH , 82-1007 , 89-1324 , 8S-3280 , 89-2756 , 82-1468, 89-2342 , 82-2431, 8S-0809 , 82-0874, 82-2115, 82-2562, 89-1138, 89-3532 , 89-0962 , 89-0187, 89-2723, 8S-3039 , 8S-1219 , 82-3144, 82-3199, 82-3200 , 89-1935 , 82-0271, 82-0411 , 8S-0099 , 89-1084 , 82-0598 , 82-0645 , 89-0946 , 89-2745 , 8S-1325 , 8S-2195 , 8S-2627 , 82-0219, 82-1828, 82-2794 , 82-3361, 8S-2276 , 89-0391 , 8S-2457 , 82-3530, 8S-1116 , 89-0860 , 89-2467 , 82-1562-RC, 89-2664 , 8S-1711-C , 82-2559 , 89-2327 , 89-1748 , 89-1059-RC, 89-1916-RC , 89-2281 , 89-3432 , 82-2610 , 82-3032 , 89-2039, 89-2040 , 89-3010 , 8S-3135 , 82-1569 , 89-0233 , 89-3037 , 82-0451, 8S-0257 , 82-0709, 82-1079 , 8S-1610 , 8S-2979 , 82-3410 , 82-0124 , 82-1750, 89-1477 , 82-2190, 89-1855 , 82-0767, 82-2095, 89-1570, 89-1580 , 82-2561 , 82-0190 , 8S-1996 , 8S-1670, 8S-2024 , 89-0915 , 8S-1621 , 89-2362 , 8S-2958 , 8S-3216 , 82-0266, 82-0267, 82-1048-RC, 89-1386, 89-2386, 89-2390, 8S-0068, 8S-3205 , 82-2188 , 89-3248 , 82-0477, 89-0400, 89-0401, 8S-0299 , 82-2766 , 89-2221 , 89-2733 , 89-2703, 8S-0381 , 8S-3260 , 82-1083, 82-1888, 82-2259, 82-3285, 8S-0251, 8S-1559, 8S-2579 , 82-0272, 89-1630, 89-3324, 8S-2844 , 8S-0323 , 82-0355, 82-2364, 82-3187 , 82-3190 , 8S-0441 , 8S-1911-C, 8S-2902-C , 82-1607, 8S-2813-C , 82-0688, 82-2711, 89-2340, 89-2542 , 8S-2168 , 82-0404 , 89-2818 , 89-1181 , 8S-0306 , 82-1922, 82-3018, 82-3647 , 89-2598 , 89-0178 , 89-2603 , 8S-3220 , 82-2349, 82-2888, 89-0344, 89-3495 , 8S-2517 , 89-0281 , 82-3320 , 89-0879 , 89-0705 , 82-0591-RC, 82-3026, 89-1568 , 89-1593 , 82-3411, 82-7017-MH, 8S-0715, 8S-2757 , 8S-2769 , 82-1954 , 89-1442 , 89-3127, 8S-2853 , 89-3130 , 89-3135 , 82-1163, 8S-0586 , 89-2608 , 82-1052 , 8S-0853 , 82-0918 , 89-1178 , 89-3100 , 89-0724, 8S-2864 , 8S-2041 , 89-1852 , 89-0253 , 89-3235 , 89-0151, 8S-0455 , 89-1282 , 89-0432 , 82-3145 , 89-1483 , 89-1292, 89-3519 , 82-2366, 82-2900, 8S-7226-CMH , 8S-0357 , 8S-0647 , 89-1966 , 8S-1841, 8S-1842, 89-1627 , 8S-0316-C, 8S-0844-RC , 82-0548 , 8S-0582, 8S-1401 , 82-0623 , 82-0072-RC , 89-2665 , 89-1701 , 89-0797-RC , 8S-0559-C, 8S-1554, 8S-2604, 8S-3186, 8S-3195 , 82-0418, 8S-0377 , 82-2357 , 89-3021 , 89-2020 , 89-0322 , 82-0303, 82-3641 , 82-3656, 89-3345, 89-3387 , 89-2601, 8S-0383, 8S-3178 , 82-3825, 82-3826, 89-3420, 89-3424, 89-3426 , 8S-0721-C , 8S-0654 , 82-0720, 82-2609 , 82-1879, 89-3254 , 82-1466, 89-0699 , 82-0593, 82-2860 , 8S-2378-C , 89-0755 , 89-1430 , 82-7059-MH , 8S-2330, 8S-2392 , 82-2015, 82-2884 , 82-0906, 82-1425, 82-2417, 8S-1076 , 82-2849, 89-2441 , 89-2442 , 8S-0384 , 8S-0498 , 8S-2408 , 89-2994 , 8S-0536 , 82-1348 , 8S-1471, 8S-1474, 8S-1477, 8S-1478 , 82-1791 , 82-0701 , 82-3505 , 8S-2287 , 82-1892, 8S-0164 , 89-0689, 89-0986-C , 82-0649 , 8S-0165 , 89-1529 , 89-0307, 8S-0217-C , 8S-2115 , 89-1532 , 89-3187 , 8S-1539, 8S-2727, 8S-2859 , 8S-3055 , 82-0529 , 82-3214 , 82-2864 , 89-1203, 89-1223, 89-1224, 8S-0780 , 8S-0143, 8S-0550 , 82-2541 , 89-3262 , 8S-0044 , 82-0532 , 82-1499 , 82-1103, 82-2001-RC , 89-0951 , 89-1247 , 82-2282, 82-3789, 89-1534 , 82-1714 , 82-2194-RC, 82-3925 , 89-1136 , 82-2902 , 82-3362 , 82-2436, 8S-3070 , 82-0646, 89-1391, 8S-2797 , 89-0781, 8S-2062 , 8S-2071, 8S-2794 , 89-3253 , 89-3147 , 82-1253, 82-3407 , 82-2552, 89-1256 , 82-0428 , 89-2525 , 82-1530, 89-0240 , 8S-0478 , 82-0198, 82-0904, 82-3606, 8S-1141 , 89-1049 , 82-0974, 82-2625 , 89-1579-N , 8S-0641 , 82-2103, 89-0310, 89-0416, 89-1334 , 82-3398 , 89-3096 , 8S-2343 , 89-0892, 89-2397, 8S-2116 , 82-2180 , 82-0367, 82-0749, 82-2100, 89-0394 , 8S-3142 , 82-0545, 82-0546, 82-1673, 82-3208 82-3835, 82-3836, 8S-0273, 8S-1343 , 8S-2781 , 82-2469 , 82-7006-MH , 89-2256 , 8S-7135-CMH , 8S-2662, 8S-3197 , 8S-0236-C , 8S-0362 , 8S-2775 , 89-0479, 8S-3193 , 8S-2220, 8S-2231 , 89-0528 , 82-1309 , 82-2300 , 82-2916, 89-1977, 89-1982, 89-1989 , 8S-2657 , 82-0141, 8S-1656 , 8S-0045 , 89-2419 , 8S-1012-C , 8S-7099-CMH , 82-3180, 89-0529 , 8S-3045 , 89-2727, 8S-1907 , 89-1474-N , 82-3468 , 82-1894 , 89-1765, 89-2422 , 82-3846 , 82-2545, 89-0486 , 8S-3292 , 82-1009, 82-2420, 82-3535 , 82-0286, 8S-1177 , 82-1478, 82-3454, 82-3678 89-0651, 89-0652, 89-2358 , 8S-0025-C , 89-2364 , 8S-0452 , 8S-0999, 8S-2015-N , 8S-1654 , 8S-1831 , 89-2140 , 89-2162 , 82-0481 , 82-3105 , 82-1100 , 82-2290 , 82-1878 , 82-0919, 82-1631, 89-3232 , 89-0185, 8S-0467 , 89-0290, 8S-2370 , 82-3195, 82-3196, 82-3197 , 82-3598 , 89-0424 , 82-0392 , 8S-1186 , 8S-1125-C , 8S-1269 , 89-2302 , 89-3078 , 82-0818, 82-1734, 89-3192, 89-3464 , 82-7050-MH, 89-2690, 89-3415, 89-3423 , 82-2859, 8S-0049-C, 82-1639 , 8S-2315, 8S-2754 , 82-0089 , 89-1488 , 82-0473, 82-1691, 89-0388, 89-2837 , 82-0581, 82-1271, 82-3368, 89-2775 , 82-0793, 89-3515 , 82-1290, 82-1701, 89-0303, 89-0363 , 82-2287, 8S-0363, 8S-2915 , 82-3914 , 89-2871 , 8S-0270 , 8S-2110 , 82-3114, 89-2679 , 89-1660, 89-1871 , 89-2754 , 82-1480 , 89-2315 , 82-0696 , 82-0358, 82-1358, 82-1473, 82-2924, 89-1071, 89-2683 , 82-2605 , 82-0078, 82-2003 , 82-0966, 89-0777 , 82-2760, 82-3000, 82-3037, 89-1229 , 89-1160 , 82-2768, 8S-1274 , 8S-3250 , 82-1751 , 8S-2762 , 82-0413, 82-1317, 82-1686, 89-2544, 8S-2109 , 82-7130-MH , 82-0237 , 89-0990 , 82-7012-MH , 82-2415 , 8S-1312, 8S-2385 , 82-1707, 82-2632 , 82-1895, 89-0246 , 89-0606 , 82-2518 , 82-0589 , 82-0648, 89-1378 , 82-1200 , 82-2281 , 89-0587 , 8S-2422 , 82-2809 , 89-3399 , 82-0534, 8S-0261, 8S-1432, 8S-2834, 8S-3248 , 89-0225, 8S-2952 , 82-1674 , 89-1632 , 82-2998 , 82-2592, 8S-0640 , 82-1702, 89-2341 , 82-0890, 82-2310, 89-1999 , 89-0071 , 89-1788 , 82-3015 , 82-0907, 89-3285, 8S-0931-C , 89-2051 , 82-1616, 89-2150 , 8S-2606 , 8S-2933 , 82-0189, 82-3206, 89-1449 89-1673, 8S-1837 , 8S-3183 , 82-0636, 82-2788, 8S-2701 , 89-1528 , 8S-1081 , 8S-1791 , 82-1696, 82-2009, 82-2128 , 82-2977 , 82-1661 , 8S-0298-C , 82-1947 , 82-0449 , 8S-0052, 8S-1833 , 82-0891, 82-1528 , 8S-1080 , 82-3737 , 89-0872 , 82-0517, 8S-1023 , 82-1882, 82-1882, 89-0061 , 82-0502, 82-2869 , 89-0793 , 89-2357 , 89-0197, 89-0197, 89-1135, 89-2015 , 82-0655, 89-0204, 89-2111 , 82-1446 , 8S-1802 , 89-0234 , 89-0748 , 89-2038 , 82-2464, 8S-1015 , 8S-0356, 8S-2303 , 89-2990 , 82-2629 , 89-2405 , 8S-3247 , 82-2321 , 89-2484 , 89-3160 , 8S-1173 , 8S-0104, 8S-0439 , 8S-1679 , 82-1852, 82-2126 , 82-0038 , 82-2014, 8S-2279, 8S-2870 , 82-1289 , 82-0218, 82-2197, 82-2260, 8S-1603 , 82-0301, 89-1428, 89-1743, 89-1806, 8S-0525, 8S-0526 , 82-0558, 82-3457 , 82-1420, 82-3670 , 82-1627, 89-0494, 8S-0722 , 8S-2283 , 82-1192 , 82-2130 , 82-1263, 82-3515 , 82-1085 , 82-2758 , 82-3666, 8S-0051, 8S-1356-N , 89-0521, 89-1500 , 82-0660 , 82-0273, 82-2400, 8S-1429 , 82-0853, 82-1160, 82-2551, 82-3326 , 82-3440 , 82-3475 , 82-0249, 8S-1187 , 82-0409, 82-1817 , 89-1154 , 89-0500, 89-2827 , 89-3169 , 82-1124, 82-2620, 89-3145, 8S-0338-C 8S-1988, 8S-1991, 8S-3305 , 8S-3309 , 8S-2658, 8S-2705 , 89-1839, 89-1846 , 82-2137, 89-1954 , 8S-1499 , 8S-0043-RC , 8S-1324 , 8S-0009 , 82-1434, 89-0524 , 82-2872, 89-1426 , 82-1977 , 89-3203 , 82-1349, 8S-2625, 8S-2648, 8S-2649, 8S-2650 , 8S-0428 , 8S-2570 , 8S-2940 , 82-2666 , 82-0470, 82-2688, 8S-3139 , 8S-0680-RC , 89-0754, 89-2411, 89-2847 , 82-7195-MH , 89-1951 , 8S-1007 , 82-0120, 89-1702, 89-2200 89-2201, 89-3069
89-3497 , 8S-1446 , 8S-1998 , 8S-1981, 8S-1982, 8S-2020 , 8S-3053 , 8S-1792 , 8S-2505 , 8S-2157 , 82-7098-MH , 8S-1546 , 82-0905 , 89-3525 , 8S-2694 , 8S-1829 , 82-2655, 89-0305, 8S-3276 , 8S-3257 , 82-1690, 89-2729, 8S-0349 , 89-1820 , 82-0695, 82-2449 , 82-1454, 8S-0864, 8S-0865, 8S-3095 , 89-0852, 8S-0067-C , 8S-2837 , 89-0828, 89-2654 , 89-0215 , 89-0713 , 8S-2386-RC , 82-1118, 89-1512 , 89-2110 , 8S-2866-C , 8S-0064 , 8S-2972 , 89-0459-RC , 82-1417, 89-1399 , 82-1455 , 82-7069-MH , 82-1592 , 82-0910 , 89-3504 , 89-2636 , 8S-2066 , 89-2648 , 82-1757, 82-1838, 82-2866, 82-3227, 82-3651, 82-3657, 8S-1121, 8S-1527, 8S- , 8S-0109 , 8S-2723 , 82-2328 , 89-0657 , 82-0572 , 8S-1348 , 89-1575, 89-2599 , 89-3058 , 8S-1497, 8S-2117 , 82-1682 , 82-2347 , 89-1790 , 89-1165, 89-1797, 89-2851 , 82-1812, 8S-0263, 8S-3008 , 8S-3044 , 8S-0857 , 89-0333, 8S-0264 , 8S-1521 , 89-2854 , 8S-1374, 8S-1463 , 82-2734, 89-1618, 89-1663 , 82-1447 , 89-1372 , 8S-1779 , 89-3039, 89-3315, 89-3369 , 82-1485, 89-0635, 89-2732 , 89-3140 , 82-0760-RC, 82-1610, 82-3223, 82-3386 , 82-0142, 89-0717 , 82-0730, 82-2439, 82-3511, 89-1348, 89-1611, 89-3522 , 89-0152 , 89-1770 , 89-0981 , 82-0467, 82-2283, 89-0502, 8S-2292-N , 89-0182 , 82-1725, 82-2519, 82-3344 , 8S-0190 , 82-3939 , 8S-3243 , 89-1466, 89-1467 , 82-1539, 82-3337, 82-7180-MH, 89-2457 , 89-2883 , 82-0051, 82-0398-RC, 82-1400 , 89-2752 , 82-0405, 82-2690, 8S-0406 , 8S-0981 , 82-1074 , 82-1310 , 82-0087, 82-0431 , 89-2048, 89-0512-RC , 82-0334-RC, 82-1557-RC, 82-2114, 82-2658, 8S-0250 , 8S-3063 , 89-2173 , 82-0946, 82-2101 , 82-0731, 82-0732, 82-2441, 82-2521, 82-3124 , 8S-1220 , 8S-0527, 8S-2578 , 82-1931, 82-2514 , 82-3237 , 89-3173 , 82-1968, 89-1253, 89-1775, 89-3150 , 82-0869, 89-1967 , 89-2055 , 82-0498, 89-1932 , 89-1029 , 89-1670 , 89-2077 , 82-1740, 8S-0679-RC, 8S-1506-RC, 8S-1640, 8S-2389 , 82-3115, 8S-0301-C , 8S-3209 , 8S-1048 , 82-3123 , 82-3262 , 89-1582 , 89-2033 , 89-3529 , 89-1691, 89-2097 , 8S-2328 , 82-1672 , 8S-2468, 8S-2704 , 82-2686 , 82-1805, 82-2759, 8S-1191 , 8S-1346, 8S-2518-C, 8S-2519-C , 8S-2055 , 82-0123, 82-2286, 82-3228 , 82-3230 , 8S-2721 , 82-0606, 82-2201 , 82-2604 , 82-3485, 89-0758, 89-0764 , 82-0422 , 82-3008 , 82-0376 , 8S-0975, 8S-1720 , 82-0369, 82-7016-MH, 89-0973, 89-1948 , 82-3067 , 89-0447, 8S-0253 , 82-1952 , 8S-2038, 82-7011-MH, 8S-1581, 8S-2160, 8S-2161, 8S-2162 , 82-2171 , 89-3256, 89-3258, 89-3299 , 89-3181 , 82-0553, 82-2773, 89-0417, 8S-0388, 8S-2336-N , 8S-2155 , 82-3611, 89-1551 , 82-3746, 89-3434 , 89-2834 , 89-0624 , 8S-1426 , 82-3483, 82-3888 , 82-3267 , 82-3566, 8S-0407 , 89-1496 , 82-0707, 82-0883, 89-0947 , 82-3375, 89-1658 , 8S-0046-RC, 8S-0333-RC, 8S-1228-RC, 8S-2060 , 89-3117 , 82-1726, 82-2744 , 82-2208 , 8S-0233-C, 8S-7118-CMH , 89-1729 , 82-2468, 89-0192, 89-0210, 89-3400, 89-3450, 8S-3048 , 82-3379 , 89-1996 , 89-0320 , 8S-2451 , 8S-3236 , 89-3401 , 8S-2906 , 8S-3145 , 82-3645, 8S-3075 , 8S-3151, 8S-3153 , 89-2049, 8S-3286 , 89-0727, 89-2417 , 82-0877, 89-0802, 89-2300 , 89-0745, 89-3465 , 82-0450, 82-2336, 89-0497, 89-2086 , 82-0602, 89-1720, 89-2760, 89-3112, 89-3114, 8S-2848 , 89-2385 , 89-2763 , 8S-1315 , 8S-1749, 8S-1750 , 89-2983 , 8S-1781 , 89-1664 , 82-0049-rc , 82-2054 , 82-1960 , 82-2155 , 82-2157 , 82-1325, 8S-0817, 8S-0826 , 8S-0827 , 8S-0798 , 82-2233 , 89-2782, 89-2639, 89-2640 , 89-2641 , 8S-2621 , 8S-1868 , 82-3170, 89-2269, 89-2270, 89-2271 , 82-2019 , 82-2876 , 82-2883 , 89-0194, 8S-2999 , 8S-7072-MH , 82-0406, 82-1901, 82-1906, 82-2587 , 8S-0341, 8S-0824 , 89-0383, 89-3193 , 8S-0845 , 82-0468 , 8S-1341 , 82-0238 , 8S-3150 , 8S-0558 , 82-1106 , 89-2335 , 89-0514 , 82-1747 , 82-7048-MH , 82-7131-MH, 82-1038, 82-1598, 82-2453 , 89-2292 , 82-1951 , 82-0009, 82-1176, 82-2033-N, 82-2637, 82-2638, 82-2639, 82-2641, 82-3034, 8 , 82-0850, 82-0938, 82-0940, 82-0941, 82-0943, 82-1033, 82-1405, 82-2424, 82-2 , 82-0944, 82-7110-MH , 82-1541, 82-1830-N, 82-2818, 82-3649 , 82-3494 , 82-0615, 82-7089-MH , 89-1121 , 82-2140 , 82-0485,82-1849 , 82-3367 , 82-0150-RC , 82-0074 , 8S-0556, 8S-0773 , 82-0756 , 8S-2402 , 82-1978, 89-2875 , 82-2346 , 82-0568, 82-3633, 8S-1318, 8S-2546 , 89-0280, 89-0101, 89-0167, 89-0335, 89-0506, 89-0508, 89-1604, 89-1721, 89- , 89-0799 , 89-0072 , 82-1097 , 8S-2508 , 89-1856 , 89-0590-RC, 89-2423 , 89-2616 , 82-3815 , 82-2845, 8S-1929 , 8S-7160-MH , 89-2790 , 82-7135-MH, 89-0195 , 8S-0538 , 82-1198 , 82-1891 , 82-0277, 82-1994, 89-1345, 89-1823, 89-3020, 89-3027, 8S-0922, 8S-1655 , 8S-2664 , 82-2239, 89-0080, 89-2167, 89-2559 , 82-2173 , 8S-2633 , 8S-1011 , 8S-3013 , 8S-2696 , 8S-1707 , 82-0990, 89-0639, 89-3266, 8S-2981 , 89-0875 , 89-1672, 8S-2840 , 8S-2347 , 82-3680 , 89-1096 , 8S-0470 , 89-0927 , 82-0394, 82-7094-MH , 89-2130 , 89-3313, 8S-2140 , 89-3116 , 89-0659 , 89-1037 , 82-1125 , 82-0849-RC , 89-0033 , 89-2367 , 89-3210 , 8S-1579 , 82-2285, 89-0258, 8S-0284, 8S-0727, 8S-1243, 8S-1246, 8S-1623, 8S-2752, 8S , 89-2546 , 89-0866, 89-1651, 89-2769 , 89-1722 , 82-0121-RC, 82-3258, 89-0139, 89-0515, 89-2185, 8S-0492, 8S-1700, 8S-1799 , 8S-0093-C , 82-3728 , 82-0324, 8S-0779 , 82-1556, 8S-1201-RC , 82-3839 , 89-1151 , 89-0264 , 8S-2913 , 8S-1622 , 8S-2280, 8S-2919 , 82-0710 , 82-1449, 82-3431, 8S-1449-C, 8S-2053 , 82-7156-MH , 8S-3132, 8S-3136 , 89-2115 , 8S-2695 , 82-2214 , 82-1756, 82-3697, 8S-1570, 8S-1226-C , 82-1867, 8S-1250, 8S-2948 , 82-2225, 82-3204, 82-7093-MH, 8S-2151-N, 8S-3021 , 8S-1354, 8S-2481 , 82-3679, 89-1010 , 8S-2509 , 8S-2511 , 8S-1169 , 89-0238 , 8S-2138 , 8S-0881 , 8S-1983 , 82-0590, 89-0084, 89-2175 , 8S-0976 , 82-1062, 82-2040 , 82-3603 , 82-3036, 89-1584, 89-3029, 89-3267 , 8S-1706 , 89-1998 , 89-2219 , 82-2141 , 8S-2748 , 8S-2021, 8S-2022 , 8S-0465 , 82-2778 , 8S-2761 , 82-0057, 8S-0287 , 82-2989, 89-1930 , 89-2482 , 89-0180, 89-0252, 89-0277, 89-0295, 89-1608 , 89-2796 , 8S-1455 , 8S-2852 , 89-0260, 8S-1400 , 89-1879 , 82-1281, 89-0516 , 8S-3268 , 89-2134 , 89-0188, 89-0783 , 89-1489 , 82-0861 , 8S-0557 , 89-0269, 8S-2585 , 8S-1674 , 89-2591 , 82-1177, 8S-1637 , 82-0596, 8S-1614 , 89-1725 , 82-0630, 82-1292 , 82-2505, 89-0795, 89-2637, 89-3502 , 8S-3221 , 82-2169, 82-2459, 82-3409, 89-3472 , 82-0825, 82-3136 , 8S-0539 , 8S-2670 , 82-2234, 82-2447, 8S-0475 , 89-1034 , 82-1815, 89-2596 , 82-1735, 82-2465 , 89-3050 , 82-0300-C , 82-0855, 82-1920, 82-2534 , 82-1256 , 82-2991 , 82-2910, 82-3548-E, 89-1900 , 82-3198 , 82-2839 , 82-2634 , 89-2370, 89-3333 , 8S-2217 , 89-3425 , 8S-0476-RC , 8S-2839 , 8S-0327 , 89-0894 , 89-1258 , 8S-2707 , 89-0881, 89-3049 , 82-3634 , 82-2031 , 8S-2564 , 82-3193, 82-3913 , 82-0434, 82-0435, 89-0349 , 82-3851 , 82-0222-RC, 82-3605 , 82-0162 , 89-1594 , 8S-0842 , 82-1657 , 8S-2765 , 89-2046 , 89-2618 , 89-0143 , 8S-2492, 8S-2493, 8S-2956, 8S-2957, 8S-2964 , 89-1417, 89-3505 , 82-1413, 82-3874 , 89-2475, 8S-1596 , 8S-2730 , 8S-2289, 8S-3128, 8S-3129 , 82-3930 , 89-3245 , 82-1515 , 82-0368 , 89-1058, 89-2318, 89-3422 , 8S-0947 , 82-3128, 89-1616 , 89-2535 , 8S-1079 , 82-0257, 89-0216 , 82-0979, 82-2355, 8S-0709, 8S-3097 , 8S-0191-C , 82-2151 , 8S-1026 , 8S-0628 , 89-2263 , 89-0743, 89-2283, 89-2288 , 8S-7083-MH , 8S-1821 , 82-0618-M , 82-0366, 89-3468 , 82-3706, 89-0568 , 82-3847 , 89-0385, 89-1613 , 8S-1805 , 8S-1992 , 8S-1536 , 82-0935, 82-2319 , 82-3547 , 8S-2854-C , 82-1262, 82-1502, 82-2608, 82-2613, 82-2654 82-2804, 82-3412, 82-3421, 8S-1161 , 8S-2054-C , 8S-0591 , 82-1195 , 82-2104, 89-1955 , 89-1186, 8S-1020, 8S-1634 , 82-3346 , 8S-2916 , 89-1150 , 89-0040, 89-1380, 8S-1495 , 8S-2305 , 82-3432 , 82-3311 , 82-0075 , 82-3552 , 82-2270 , 8S-1193 , 82-2997, 89-0381 , 89-3123-N , 89-0327, 89-1819 , 89-3275 , 8S-0672 , 82-2761, 82-3669 , 89-0301 , 89-2080 , 8S-2943 , 82-0159, 82-0664-RC, 82-3141, 82-3719 , 8S-1947 , 8S-2446 , 8S-2969 , 8S-2622 , 8S-1229 , 89-2600 , 82-7147-MH, 89-2462, 89-3056, 89-3061, 89-3062, 8S-3094 , 89-3251 , 82-7119-MH, 82-7142-MH, 89-0701 , 82-2461, 89-3270, 89-3274 , 8S-2967 , 8S-2257 , 82-3915, 89-0470, 89-0471 , 89-2671 , 89-2672 , 8S-1124 , 8S-2557 , 8S-0303 , 8S-2391 , 8S-3056 , 8S-0090-RC, 8S-0218-RC , 89-1141, 8S-0535 , 89-1949 , 89-0177-RC, 89-0953-RC , 89-0166-RM , 8S-2586 , 8S-2247 , 89-3261 , 82-3012-N , 8S-0627, 8S-1979 , 8S-3062 , 89-1005 , 8S-3239 , 89-0853 , 89-1254 , 8S-2044 , 82-2904 , 89-1288 , 82-1833 , 89-1826, 89-3172 , 89-3174 , 82-0224 , 82-1715 , 8S-0225 , 89-1898 , 89-2178 , 89-1255, 89-1755, 89-2222 89-2540 , 89-0560-RC, 8S-0182 , 89-0751 , 82-2326 , 89-2863 , 82-0259, 82-2295-C , 89-0525, 89-0526 , 89-1947, 89-2002 , 8S-2819 , 82-2644, 8S-0329 , 89-1295 , 82-0595 , 82-0702, 82-2036, 8S-3032 , 8S-0141-C, 8S-0970, 8S-2656 , 89-3281 , 89-0903, 89-0338, 89-0763, 8S-0319, 8S-0369 , 82-2386 , 8S-0321, 8S-0325, 8S-0446, 8S-0777, 8S-1673 , 82-3430 , 82-0274, 82-2695-C , 8S-0271 , 8S-1117 , 89-0023 , 8S-1180 , 8S-2102 , 8S-2036 , 8S-2040 , 89-2712 , 8S-3206 , 8S-1024 , 8S-2308 , 8S-0368 , 8S-0886 , 8S-1958-RC , 8S-1780 , 89-2186 , 8S-2911 , 8S-0315 , 8S-1599 , 8S-0902 , 8S-0631 , 8S-0203 , 8S-1727 , 8S-0248, 8S-0573, 8S-0733 , 8S-2732, 8S-3164 , 8S-1590, 8S-3005 , 8S-0126 , 8S-0194-RC , 8S-1321 , 8S-3090, 8S-3091 , 8S-2082, 8S-2083, 8S-3225 , 8S-0614 , 89-0355, 89-0614 , 89-1908 , 89-2249 , 89-2527 , 8S-0372, 8S-1661 , 8S-0747 , 8S-1505 , 8S-2150-N , 8S-2267 , 8S-3059 , 8S-1487 , 8S-2425 , 8S-1818 , 8S-2607 , 8S-2923 , 8S-1082 , 8S-2198, 8S-3166 , 8S-2965 , 89-0974, 89-2620 , 89-3070 , 8S-0269, 8S-2485 , 82-2304 , 89-3493 , 8S-3112 , 89-1922 , 8S-0393 , 82-3488 , 89-1845 , 8S-0445 , 89-1574 , 8S-2415 , 8S-1087 , 82-3249 , 82-3372, 89-1332 , 89-3271 , 82-1159, 82-1723 , 82-3565 , 82-1869-RC, 89-0340 , 8S-1155 , 82-1797, 82-3868, 89-2560 , 8S-1843 , 89-1484 , 89-0031, 89-2606 , 8S-1686 , 82-1694 , 82-2793 , 82-1759, 8S-3302 , 89-1269-RC, 89-1674, 8S-0766 , 82-1104 , 8S-2350 , 8S-0694 , 89-0339 , 82-2713 , 82-0444, 82-1729 , 89-1498 , 89-0671, 8S-2400 , 89-2339 , 82-0856, 8S-2540 , 8S-1725, 8S-2013, 8S-2355 , 89-1381 , 82-2798 , 8S-2531 , 8S-0665 , 8S-1755 , 8S-1756 , 82-3023, 8S-0311 , 82-1716, 82-3455 , 82-0234, 8S-1718 , 8S-2875 , 89-3449 , 8S-2483 , 82-1526, 89-1248, 89-1761, 89-3038 , 89-3083 , 82-3894 , 82-0691 , 82-3084 , 8S-1901 , 8S-0867, 8S-1136 , 8S-2798 , 82-1676, 82-2856, 82-3904 , 82-2783, 82-3845 , 82-3035 , 89-0663, 89-1438 , 82-0430-RC , 89-2804 , 8S-3263 , 82-0070, 8S-0711 , 8S-0764 , 82-0129, 82-0747, 8S-0158 , 82-0116, 82-2899, 89-2509-N , 82-2380, 8S-0504 , 82-0399 , 82-0186, 8S-0626 , 8S-3253 , 8S-0588 , 89-3577 , 89-0736 , 89-3568 , 8S-0018-C , 82-0822, 82-2476 , 89-2856 , 89-2720 , 8S-1635 , 82-0216, 82-1180 , 82-0506, 82-0587, 82-1336, 82-3334, 8S-0668 , 82-2089, 82-7086-MH, 8S-1436 , 82-2268 , 82-1996 , 82-1586 , 8S-1689 , 82-1491, 82-7194-MH, 89-3448 , 8S-0874 , 89-0316-RC , 82-0471-RC , 89-2586 , 82-1563, 82-3380 , 89-2714 , 8S-2946 , 8S-3211-CH , 8S-0723-C , 82-1646, 82-2986, 89-2193, 89-2194, 89-2982 , 82-0256, 89-2365, 89-2554 , 82-1776, 89-0092, 89-1077 89-1078, 89-2303 , 8S-2879 , 82-0784 , 8S-1516 , 8S-2547 , 8S-0793-C , 8S-2793 , 82-2955 , 8S-2323 , 8S-2878 , 8S-2863 , 8S-3154 , 8S-0778 , 8S-1735 , 89-1988 , 8S-0675, 8S-2461, 8S-0979-RC , 8S-1578 , 8S-2456 , 8S-2142 , 89-0395, 89-1937 , 8S-1175 , 8S-2534 , 82-1305, 82-2426 , 8S-2430 , 8S-1323 , 8S-1734, 8S-2366 , 89-3214 , 89-1310 , 8S-1798 , 8S-2613 , 8S-2145 , 89-0309 , 8S-1790 , 8S-0937, 8S-2988 , 89-1639 , 82-1772-C, 89-2787, 8S-1611, 8S-1613-C , 89-2430 , 82-3150, 89-3034 , 82-1517 , 8S-1204 , 82-1024, 8S-0674 , 8S-1440 , 82-3916, 8S-3249 , 8S-2241 , 8S-3127 , 82-1440 , 82-1547 , 89-1685, 89-2678 , 89-1795, 8S-1653 , 8S-7141-MH , 82-0280 , 8S-0718, 8S-1016 , 82-0401, 89-1495 , 82-2750 , 82-3677, 89-1503, 89-3513 , 89-2058 , 82-2912, 82-3787, 89-3030 , 82-1179, 89-0055, 8S-2970 , 89-3554 , 8S-1179 , 82-2710 , 82-0448 , 8S-0870, 8S-1608 , 89-0570 , 89-1609 , 8S-2746 , 89-0638, 89-1984 , 82-0127, 89-0450 , 82-2707-N , 82-1959 , 89-1641 , 8S-1542 , 8S-3167 , 89-1675 , 8S-1645 , 89-2418 , 89-1125, 89-2673 , 89-0362 , 8S-1939 , 8S-0097-C-RC , 8S-0116, 8S-0117 , 8S-1849, 8S-1940 , 82-0338, 82-2161-N-RC, 82-3110 , 82-3113 , 82-1635, 89-1997 , 8S-2759 , 82-3299 , 89-2577 , 82-3354 , 82-7020-MH , 8S-0610, 8S-2039 , 8S-2871 , 82-0657 , 82-0551 , 82-2149 , 82-3209, 89-0125, 89-0855, 8S-0710 , 8S-1759-N , 89-1249, 89-1251, 89-1981 , 82-0403, 82-2396 , 8S-1803 , 8S-0518 , 8S-1722 , 8S-2714 , 8S-0199-RC , 89-3312, 89-3419 , 89-2717 , 89-0186, 89-0598, 89-0507, 89-1273, 89-1276, 89-1278, 89-1340, 89-1598, 89- , 82-0613, 82-0705, 82-2754 , 89-0148 , 8S-2603 , 82-0505-RC , 82-0800, 89-3166 , 8S-1333-C , 82-0341 , 8S-1769-C , 82-0857, 82-2679, 8S-2284 , 82-3377 , 89-2406 , 82-1364, 89-2981 , 82-1969, 89-1260, 89-1267 , 8S-0541, 8S-1178 , 89-1939, 8S-0696 , 8S-2782 , 89-3159 , 82-0154, 89-1093 , 82-0847, 89-0390, 89-1867 , 89-3101 , 82-1506 , 82-1591, 82-3589 , 89-2783 , 82-2703, 89-0249, 89-0250 , 89-1525 , 89-3182 , 82-0763, 82-1594, 82-3225, 82-3526 , 82-2540, 82-3621 , 89-2257, 89-2735 , 82-0628, 8S-2922 , 82-3886 , 82-7057-CMH , 82-2595 , 8S-1978 , 8S-1044, 8S-1789 , 82-2039 , 89-2106 , 82-3013, 82-3304, 89-0200, 89-0553, 89-1462, 89-1468, 89-3088, 8S-2719 , 8S-2256 , 89-0912 , 89-2753 , 82-2810 , 89-1789 , 82-0415, 82-7123-MH , 8S-1018 , 8S-1524-RC , 89-3081 , 89-0778 , 89-2309 , 82-1928, 8S-3102 , 82-1307, 82-1308 , 82-1493 , 82-2020, 82-2765, 82-7075-MH, 89-3176, 8S-0601, 8S-1629 , 8S-0202 , 8S-2832 , 82-0851-RC , 82-2724 , 89-0605 , 89-1548 , 8S-0201-C , 89-2114 , 89-2117 , 82-0147 , 82-1991 , 82-1419, 82-3521, 89-2250 , 82-2135, 82-3563 , 8S-0483 , 82-0223, 89-2649 , 82-0427, 82-1731 , 8S-3179, 8S-3308 , 82-0599, 8S-0244 , 8S-1453 , 89-3103 , 8S-2482 , 8S-0644 , 82-0192 , 8S-0786 , 82-3239 , 82-3284, 89-0384, 89-1128 , 8S-3009 , 82-3156 , 8S-1577 , 89-0601, 89-1004, 89-2063 , 82-1495, 89-0231 , 82-0491, 82-0879, 82-1030, 8S-1501-N , 82-7025-MH, 8S-1357 , 82-0421, 82-2880 , 82-7157-MH , 82-0501-RC, 8S-0061 , 8S-2744 , 89-1287 , 82-0440 , 82-0407-RC , 8S-0449 , 89-3494 , 89-2212 , 8S-0252 , 89-2398 , 89-2543 , 82-1137, 82-2087, 8S-0219 , 8S-1403 , 82-0299 , 82-0881, 82-0313, 82-0563, 82-2889 , 89-0989 , 82-1058 , 82-3721 , 8S-0564 , 82-1149, 89-1389, 8S-0300-C , 89-0586 , 82-2607 , 82-3167 , 82-3438, 8S-2884, 8S-2889 , 82-3652, 82-3653 , 89-1228 , 89-3540 , 89-2382 , 82-1543, 89-0807 , 82-1127, 89-3045 , 89-0678 , 89-0105, 89-1772 , 89-1995 , 8S-0763 , 82-0607, 82-1621, 8S-0200 , 89-0582 , 82-1016, 82-2932 , 8S-1307, 8S-2478 , 82-1548 , 82-2640 , 82-2727, 82-3715 , 82-3859 , 82-2112, 89-0347 , 89-0346 , 82-1148-RC , 82-0177, 82-0178, 82-1524, 82-2719 , 82-3080 , 8S-2655 , 89-3223, 8S-1724 , 89-0496 , 82-0605 , 8S-0810 , 8S-3231 , 82-0015 , 89-1232, 89-2343 , 82-0963, 82-2354, 89-2415, 8S-2272 , 8S-2712 , 89-0126 , 89-2060 , 82-1845 , 82-1494 , 82-0900 , 82-1323, 82-1666, 82-3578 , 82-1619 , 82-1809 , 82-0886, 82-2764 , 82-1637 , 82-3005-RC, 89-1231 , 82-1050 , 82-3233, 89-2377 , 8S-1972 , 82-0469, 82-1213, 82-7051-MH, 89-1142 , 8S-0561 , 8S-2067 , 82-3498 , 89-2184-N , 89-2581 , 8S-0151-RC1, 8S-2307 , 82-2920-N, 8S-0016-C, 8S-1293, 8S-1754 , 82-0010 , 8S-3241 , 89-2336 , 89-1810 , 89-3243 , 82-1040, 82-2565, 8S-0162 , 89-0952 , 89-0859 , 82-2596, 82-3882 , 89-0090 , 82-0488, 82-2375 , 89-1602 , 8S-0520-C , 8S-3208 , 8S-3213 , 82-0306 , 8S-1286 , 82-0133, 82-1202, 89-1405 , 89-2112 , 89-1002 , 89-2192 , 8S-1697 , 82-3151 , 82-1582, 8S-2767 , 8S-1856, 8S-1857 , 82-2052, 8S-0123, 8S-0998 , 82-2351, 8S-3107 , 82-0483, 82-2325, 82-2343-N , 8S-2616 , 8S-1227 , 8S-0818 , 8S-2073 , 8S-2962 , 89-0057 , 89-2259 , 8S-0649, 8S-0951 , 82-0531, 82-1011, 82-7036-MH, 89-0462 , 82-2228, 8S-0019-RC, 8S-0830 , 82-0035, 82-1843 , 82-0241, 89-0169, 89-0221, 89-0392, 89-1363, 89-1453, 89-3343 , 82-0616, 82-1025, 82-2032, 82-3339, 89-2217 , 82-0865, 89-2992 , 89-2539 , 8S-2660 , 8S-3104 , 82-3560, 8S-1558 , 82-0012-RC , 82-0535, 82-1984, 89-1157, 89-2689, 8S-0102, 8S-1353 , 8S-1381 , 82-1620, 89-3228 , 8S-0282 , 82-1546 , 82-3449, 89-2996 , 8S-1221 , 89-1571 , 89-0443, 89-1155 , 89-1328, 89-3451 , 89-0650 , 89-2771 , 89-2214 , 8S-1143 , 89-0342 , 8S-1094 , 89-0871 , 89-1931, 89-1933, 89-2500 , 82-2579 , 8S-1109, 8S-1147 , 89-2369 , 82-1850 , 82-3500 , 82-1087, 82-2091-C, 82-7111-MH, 89-2595, 89-3073 , 8S-0464, 8S-2978 , 82-1203 , 82-0201 , 89-2241 , 82-1080 , 8S-0038 , 8S-3174 , 8S-0011-C , 82-0994 , 82-0930, 89-0780, 89-1688, 89-1983 , 8S-2545 , 89-2807, 8S-2417 , 89-1881 , 89-3286 , 8S-0771, 8S-1075 , 82-0759, 82-2878, 82-3924, 89-2931 , 89-0321 , 8S-0283 , 8S-2833 , 89-3300, 89-3303, 8S-3034 , 82-0068, 89-3409, 89-3414 , 89-2744 , 8S-1364-C , 8S-2290 , 82-0212-C, 82-0252, 82-1939, 89-0867, 89-1553, 8S-1030 , 82-0650 , 89-1906 , 8S-0629 , 82-0893 , 8S-1776 , 82-0282-RC , 89-0744 , 89-0367 , 89-2101 , 8S-1937 , 82-2960 , 89-0945 , 89-0533 , 82-1173 , 82-1522, 82-3895, 82-3896, 89-3325, 89-3508, 8S-0918, 8S-2027 , 82-0062, 82-1535, 89-1261, 8S-2026, 8S-2439, 8S-2577 , 82-1800, 82-2830 , 82-2542, 82-7193-MH, 8S-2480 , 82-1942 , 8S-2333 , 8S-1786 , 89-2493 , 8S-0259 , 82-0543 , 89-3398 , 89-3435 , 89-3199 , 82-2153, 82-2983, 82-3321, 8S-0060 , 82-1808 , 82-1624 , 82-0836, 82-1865 , 82-0868, 82-2291, 89-0542 , 82-0576, 82-1122, 82-1861, 8S-1944 , 8S-0322 , 82-0442, 89-1972, 89-3268 , 82-0327, 82-3541 , 8S-0612 , 89-0102 , 89-0133 , 82-2696, 82-3405, 89-0972, 89-2311, 89-3340 , 8S-2849, 8S-3042 , 82-1689, 89-0522 , 8S-0624 , 89-0129-RC , 82-0017, 82-2198 , 8S-1460 , 82-0773, 89-1199, 89-3028 , 89-3003 , 82-2405 , 82-1411 , 82-0525, 8S-1671 , 82-1041, 89-0551, 89-0629, 89-1915 , 89-2317 , 82-0033, 82-1110, 82-1134 82-1327, 82-2664 , 89-1377 , 89-0914, 89-2237, 8S-1330-N , 82-0690 , 82-3629 , 89-3048 , 82-1076 , 8S-0040, 8S-1794 , 82-2489, 89-1098 , 82-2887, 82-7007-MH, 8S-1316 , 89-3161, 8S-1844, 8S-2253 , 82-1713 , 82-0536, 89-1655, 89-1656 , 8S-0280 , 8S-0920 , 89-1927 , 89-0716, 89-2605 , 89-0531 , 8S-1385 , 8S-0444, 8S-0546, 8S-1585 , 89-1694, 89-1696, 89-2530 , 82-2315, 82-3817, 8S-3143 , 82-1133 , 8S-2486, 8S-2487 , 8S-0081 , 89-0746, 89-2821 , 89-3278 , 8S-1593 , 8S-2031 , 89-2857 , 89-3170 , 8S-0997, 8S-2460 , 89-0219 , 8S-0875, 8S-3019 , 89-0757, 89-2855 , 89-0747 , 89-2762 , 82-7032-MH, 89-0348 , 89-0140-C , 82-0833 , 89-1041 , 89-1356 , 8S-1253-RC , 8S-0172-RC , 82-2063, 89-0616, 89-3407 , 82-3816, 82-3889, 89-3421 , 82-0978 , 8S-2753 , 8S-0549, 8S-2164 , 82-3389, 89-0863 , 89-1546 , 8S-2846 , 82-0058 , 82-3659 , 82-7191-MH , 82-2277, 82-3848, 89-0115, 89-0937 89-0941, 89-1028, 89-1233, 89-2508 , 8S-0645 , 82-1067, 89-3338 , 8S-0889, 8S-2334 , 82-2127 , 82-0858, 89-0053, 89-1792, 8S-0220-C , 89-0770 , 89-0135, 89-0994, 89-2276, 89-2278 , 82-0155, 82-3016 , 82-3220 , 8S-1583 , 82-2099 , 8S-1738 , 89-0612 , 82-2374, 82-2376, 82-2877, 8S-1713 , 89-2507 , 8S-2548 , 82-0604 , 8S-2867-C , 8S-2121 , 82-0119-RC , 89-2709 , 89-1934 , 82-2168, 82-3127 , 8S-2928 , 82-0782 , 82-0135 , 82-1224, 82-3465, 82-7038-MH , 8S-3217 , 8S-2843 , 8S-2817 , 82-1642 , 8S-2698 , 82-3593 , 82-2554, 82-3282 , 82-3408 , 82-7015-MH, 89-1042, 89-1047 , 89-1461 , 82-3713 , 89-3569 , 8S-2426 , 89-0773 , 82-2028-RC , 8S-0275 , 8S-3031 , 82-1162 , 82-1883 , 82-2215 , 82-2217 , 8S-2553 , 82-2870 , 89-1321 , 82-1285, 82-1537, 82-1545, 8S-0953, 8S-0954 , 89-1436, 89-2549 , 89-3471 , 82-1645, 82-7136-CMH, 8S-0858 , 89-1970 , 8S-1438 , 89-2532, 89-2860 , 82-0245 , 82-3020, 89-0013 , 8S-2207 , 8S-0245 , 8S-7126-MH , 82-1086 , 89-1200 , 89-3447 , 89-0104 , 82-1794, 8S-0159, 8S-0847 , 82-0524, 82-3467, 89-1531 , 8S-2685 , 89-0081, 89-2166 , 8S-0989 , 89-0456, 8S-3006 , 89-1275 , 89-2407 , 89-3237 , 82-0609, 89-1383, 89-2268 , 89-2274 , 89-1146 , 82-3892 , 8S-1235, 8S-2713 , 89-3106 , 82-3129, 82-3366, 8S-0027 , 89-2528 , 8S-0350 , 8S-1267 , 82-3201 , 82-2728, 89-2337 , 89-1115 , 82-0207, 82-1933 , 82-3257 , 82-0118 , 82-0083 , 89-0114 , 89-3053 , 8S-1894 , 89-2991 , 8S-2561 , 82-1139, 82-2230, 82-2235, 82-2487, 89-0948 , 82-2276-RC , 82-2107, 89-0045 , 89-1614 , 8S-1234 , 82-0137-RC, 82-1143 , 82-3615, 89-0284 , 89-2506 , 8S-1572 , 82-3944, 82-3945 , 8S-2050, 8S-2404 , 89-2479, 89-2631 , 8S-2128 , 82-2144 , 8S-2620, 8S-3103 , 82-0387, 8S-0048, 8S-1479, 8S-1481, 8S-2146, 8S-2158, 8S-2675, 8S-2680 , 89-0128 , 89-0607 , 89-0984, 8S-2165 , 89-1416 , 82-3202 , 8S-0656 , 8S-1414 , 82-1974, 8S-0693 , 8S-1486 , 82-0059, 82-2313, 89-0562, 89-1697, 89-1975, 8S-1338 , 8S-2190 , 82-0897-RC , 82-2976, 89-1132 , 8S-1302 , 82-3581 , 89-0711 , 82-1837 , 89-0617 , 89-2794, 8S-0716 , 89-1105 , 8S-2555 , 8S-2521 , 8S-1217 , 82-0390, 82-1877, 82-3181, 89-1599, 89-2125, 89-2126, 89-2128, 89-2129 , 8S-0087 , 8S-1264-RC , 8S-1620 , 8S-1592 , 8S-1039 , 82-1026 , 8S-1915 , 8S-2125 , 8S-2469 , 8S-0831 , 8S-1028, 8S-1037, 8S-1038, 8S-2275 , 8S-1110 , 8S-2049 , 89-2737 , 82-1659, 89-2518 , 82-0211 , 82-3219 , 8S-0728, 8S-2994 , 82-1431, 89-2230 , 82-0102 , 89-0369 , 8S-0984 , 82-0465 , 82-7071-MH , 8S-2532 , 89-0119 , 89-1566 , 82-0441, 82-1536, 82-3544, 89-2662 , 82-1732, 82-3333, 8S-0503 , 82-1135 , 82-2195 , 8S-1046 , 89-0191 , 82-1871 , 82-0895, 89-2381, 89-2884 , 8S-2968, 8S-1265 , 8S-0075 , 82-1826 , 89-0242 , 8S-1545 , 89-1539, 8S-1398 , 8S-3144 , 8S-1980 , 82-0270, 82-2311, 82-2342 , 89-1404 , 82-0433, 89-0996 , 82-1579-RC, 89-2701 , 8S-2610 , 89-2773 , 89-1333 , 82-0796, 82-0798, 82-2407, 82-3281, 82-3698, 82-3867 , 89-0581 , 82-2272, 82-3674 , 82-1712, 82-3093, 89-1244 , 89-1535 , 89-1446 , 82-2159, 82-3554 , 82-0082, 8S-1189 , 89-1786 , 82-2894, 8S-2734 , 8S-2984 , 89-3036 , 89-0923 , 82-3424, 8S-2683 , 82-3477, 89-1698, 89-2670 , 82-1705 , 82-0214, 82-2442 , 82-2250 , 89-1654 , 8S-0854-RC , 82-0260, 89-0503, 89-2132, 89-2133 , 82-3428-E , 8S-1507-N, 8S-2571 , 8S-2302 , 82-3627, 89-0670 , 8S-0547 , 89-1286, 89-1921 , 89-0649-RC , 82-0350-C , 82-0103-RC, 82-2383, 8S-0846, 8S-0994, 8S-2112, 8S-2799 , 82-0924 , 82-1983-RC , 82-2320, 89-0957 , 8S-0055-C , 82-2745 , 82-3177 , 82-3625, 8S-1431 , 8S-0828 , 89-2522 , 8S-2524-N , 82-2653, 89-3344 , 82-2857 , 89-2301 , 8S-0673 , 8S-2393 , 82-2662, 89-0535 , 89-2572 , 82-2129 , 89-2289 , 82-0112, 82-2027 , 82-2308 , 82-0762 , 89-0654 , 8S-1384 , 8S-3306 , 8S-2139 , 82-1129, 82-3133, 89-0709, 89-1597, 8S-0404 , 82-3429, 89-0749 , 82-1587-RC , 89-1133, 89-1914, 8S-2479 , 82-0091, 82-3910, 8S-0374, 8S-0714 , 89-0848, 89-0861, 89-1945, 89-2501 , 82-2072, 89-0267 , 82-0965, 89-2264 , 82-1222, 89-1097, 89-2573, 89-2574, 89-2578, 89-3507 , 82-0612, 82-3741, 89-0025 , 82-0262, 89-0976 , 89-0977 , 82-0496, 8S-0769, 8S-1304 , 82-1910, 89-0464, 8S-0030-C , 82-7062-MH , 89-0145 , 89-0306 , 89-0880, 89-1476 , 8S-2780 , 89-1349 , 82-3727, 89-0378 , 89-2210 , 89-3240 , 82-2246 , 89-1117, 89-2158 , 8S-1820 , 82-2819 , 8S-0105 , 82-3694 , 82-3388 , 8S-0098 , 82-2074, 89-0931 , 82-0104-RC, 89-1395 , 89-2026 , 89-2053 , 82-1918 , 82-0239 , 82-0248, 89-1738, 89-2882 , 82-3046 , 89-3551 , 8S-3255 , 89-0978 , 82-0382, 82-3165, 89-2749 , 82-1576, 89-2029, 89-2035, 8S-0481, 8S-1933 , 8S-3121 , 82-0769, 82-2984, 89-1184, 89-1684 , 82-0389 , 82-3436 , 82-0889, 89-3249, 89-3296, 89-3317 , 8S-1839 , 89-1409, 8S-3100 , 82-0687, 8S-2042 , 82-0819, 82-3286, 89-1067 , 82-1965, 82-2221, 82-3699, 89-2617 , 8S-0135 , 82-0283, 82-1784, 82-2623 , 82-0768, 82-2240 , 82-3146 , 82-1032, 8S-3076 , 82-1157 , 82-1402 , 82-1432, 8S-1879-N , 82-2002, 82-2172 , 89-3067 , 89-0600 , 89-2266 , 89-2349 , 8S-0378, 8S-0380, 8S-1019, 8S-1699, 8S-2411, 8S-2758, 8S-2955 , 82-1693, 82-2624-E, 82-2709, 82-2990, 8S-0034-C, 8S-0616, 8S-0655, 8S-1184 , 8S-0737 , 8S-1181 , 89-2045, 89-3233 , 89-0517, 8S-3074 , 8S-2445-N , 89-1168 , 89-0382, 8S-1371 , 82-1339, 82-2398, 89-0501 , 8S-0685 , 89-0354, 89-3043 , 82-0092 , 82-3291 , 82-0195 , 82-0213, 82-1049, 82-1577, 82-2133, 82-3038 , 82-0261, 8S-1567 , 82-1315 , 8S-2294 , 82-2806 , 89-3231 , 82-0492, 82-1482, 82-1796, 82-2948 , 8S-2881 , 82-0370 , 82-2267 , 8S-0516 , 8S-1680, 8S-2500 , 82-2395 , 89-0365, 89-1009 , 82-2293 , 82-0343-C-RC , 82-0281, 82-2801, 82-2802, 8S-0328, 8S-1619 , 82-0352, 89-1172, 8S-0416 , 82-0489, 82-7042-MH, 89-2233 , 82-0775, 82-1948, 82-2921-N, 82-3416 , 89-2788, 8S-0789 , 8S-0497 , 89-1961 , 89-1403, 8S-2464 , 8S-1239 , 82-0677, 89-1318, 89-3503 , 8S-1352 , 82-1864, 82-3134 , 82-1915 , 82-1907 , 82-0797, 82-2457 , 89-3428 , 8S-2208 , 89-0302, 8S-2496 , 89-2460 , 8S-2619 , 82-1911, 82-2444, 82-3189-N, 8S-1391 , 82-0265 , 82-2508 , 82-0494, 89-0287, 89-1330 , 82-7137-MH, 89-1322 , 89-2552 , 82-2249, 82-3587 , 82-1463 , 82-2007, 89-1066 , 82-2790, 89-1758 , 89-1817 , 82-0110, 82-2969, 8S-0138, 8S-1172 , 89-3322 , 82-0694, 8S-0872, 8S-2665 , 82-2353, 89-1055 , 8S-3148 , 8S-1955 , 82-0588, 82-3394, 89-1834, 8S-0242 , 82-0476, 82-1789 , 82-2843, 8S-2920 , 82-1721, 82-3471, 82-3822 , 89-1129 , 82-0269, 8S-0797, 8S-0840 , 89-1106, 8S-1290 , 89-2352, 89-3000 , 82-3017 , 82-0289, 82-3352, 82-3583, 8S-2944 , 8S-0792-C, 8S-1533, 8S-1925, 8S-2007-N, 8S-2008-N, 8S-2047 , 89-1828, 89-1863 , 8S-1056 , 89-0905 , 89-3075 , 82-1534, 82-2455 , 82-2748, 89-1245, 89-1907, 89-2738 , 89-2850 , 82-0073, 82-1823 , 82-0279, 82-1507, 89-0029 , 82-3324, 82-3325 , 82-2207 , 8S-0529, 8S-2678 , 8S-2691 , 89-2669, 8S-2640 , 82-2718 , 8S-1035 , 8S-0511, 8S-1111, 8S-2491 , 8S-1496 , 8S-2589-N , 8S-1469, 8S-2641 , 82-2340, 89-1824, 89-2394, 89-2412, 89-3009, 89-3016 , 89-3557 , 8S-7015-CMH , 89-0922 , 82-1662, 82-7112-MH , 82-3648 , 82-1783, 89-2183-N, 8S-1176 , 8S-2592 , 8S-1210 , 82-0608, 89-2597 , 89-2538 , 82-2423 , 82-0202 , 82-3570, 89-0643, 8S-1782 , 82-0171, 82-0561, 82-1529, 8S-0166, 8S-1672 , 89-1102 , 82-1355, 82-7081-MH, 89-0046 , 82-1761, 89-2054 , 8S-0863 , 8S-2977 , 82-1769 , 82-1164, 89-0248 , 82-0464 , 89-0888, 8S-1069-C , 82-1698, 89-0543 , 89-0796 , 89-0259, 89-1509, 8S-1888, 8S-2423 , 82-0046, 89-1705 , 82-2886, 82-3031-E, 8S-0320 , 82-2611, 82-3507, 89-0774, 89-3184, 89-3238, 8S-2960 , 82-0298, 89-0205, 89-1773, 89-1777 , 82-0569 , 89-0728, 89-1794 , 89-0959, 89-2380, 89-2865, 89-3011, 89-3122-N , 8S-0379, 8S-0794 , 8S-0896, 8S-0908, 8S-3157, 8S-3224 , 82-0371 , 82-2733 , 82-3125, 89-3394 , 8S-0765 , 82-3099, 8S-0204 , 8S-0036, 8S-0260 , 82-2438 , 89-3042, 89-3044 , 82-1875, 89-2177, 89-2551 , 82-2393 , 82-1204, 82-2301 , 8S-0224 , 89-0426 , 82-3052 , 82-3314 , 82-1781 , 89-0483-C , 89-3556 , 89-1364 , 82-0812, 82-1433 , 82-0445, 82-7120-MH, 82-7133-MH 89-0375, 89-2074 , 8S-7164-CMH , 82-1082, 82-2117 , 82-0206 , 82-0592, 82-1300, 82-3628 , 82-0385 , 82-2606 , 82-3908, 82-3909 , 82-0949, 82-1140 , 82-0204, 82-2143, 82-2162, 82-2170 , 89-2833 , 89-0078 , 8S-1216 , 82-0530, 82-0848, 82-1880, 82-3658, 8S-2963 , 89-0341 , 82-0643 , 82-0337 , 82-0594, 89-0446, 89-1270 , 82-1975 , 89-0047, 89-0048, 89-1130 , 8S-1580 , 8S-3237 , 8S-1704 , 8S-2023 , 82-2338 , 82-0302, 82-1509, 82-2392, 89-3023 , 82-0582, 8S-0581, 8S-2418 , 8S-2718 , 82-0632, 8S-0035 , 89-1061 , 8S-2502 , 82-0478-RC, 82-2401, 89-2494, 8S-0121-C, 8S-1770-N, 8S-2842 , 82-0814, 82-1655 , 82-3580 , 8S-2012 , 82-0852, 82-3635, 82-3673 , 82-0333, 82-0791, 82-2279, 8S-0534 , 8S-2197 , 89-3208 , 8S-2590-RC , 8S-3228 , 82-3029 , 82-1824 , 82-0152, 82-1209, 82-1841 , 89-1493, 8S-2504 , 89-1544 , 89-2537 , 8S-0555 , 8S-1814 , 89-3126 , 89-2722 , 82-2580, 89-2512 , 89-3178 , 82-2939, 8S-0688 , 89-1745, 89-2065 , 89-2131 , 82-1973 , 82-0693 , 8S-1095 , 8S-1504, 8S-2898 , 82-1683 , 82-1047 , 82-0106 , 8S-3038 , 82-2715, 89-0189, 8S-0442 , 89-0419 , 89-3242 , 89-2016 , 89-2003 , 89-1207 , 8S-3071 , 8S-0768 , 82-1929, 8S-2709 , 8S-2188 , 89-0950 , 8S-3299 , 8S-2543 , 89-0785 , 89-2127 , 8S-1200-RC , 8S-1867 , 89-0575 , 8S-3047 , 89-1291 , 82-1554 , 89-3001 , 82-3073, 82-3075, 89-3063 , 8S-0590 , 8S-2130 , 89-0909-RC , 82-2500, 82-2502 , 82-2504 , 82-0208, 82-2043, 82-2120, 8S-1552 , 82-2273, 8S-0266 , 82-0351, 82-1873, 82-3186, 82-3788, 89-2719 , 8S-0006-C-RC , 8S-2442 , 89-2187 , 89-0791 , 82-3166, 89-1415-RC , 89-2993 , 89-2445, 89-2461 , 8S-2677 , 8S-2909 , 8S-3069 , 8S-0086 , 89-0901 , 82-1421 , 82-3330 , 89-0056-RC , 8S-0243 , 89-1030 , 89-1241 , 89-1742 , 82-2187 , 89-2258 , 82-1677 , 8S-2306 , 89-0710, 8S-1054 , 8S-1485 , 8S-0183, 8S-2298 , 8S-3165 , 8S-2835 , 89-1561 , 89-0898, 89-2635 , 89-2137 , 8S-3020, 8S-3068 , 8S-3130 , 8S-1729, 8S-2554 , 8S-2507 , 8S-0568 , 89-0642 , 8S-1461, 8S-1800 , 8S-2724 , 82-2772 , 8S-0643 , 89-1803 , 8S-0174 , 8S-1864 , 82-0554 , 8S-2673, 8S-2865 , 8S-0074 , 89-0808 , 89-3489 , 8S-1595 , 82-0633 , 8S-1650-C , 8S-2246 , 89-3189 , 8S-3072 , 8S-1684, 8S-1685 , 8S-2617 , 89-2232 , 8S-2261-N , 82-1575 , 8S-2862 , 82-1570 , 8S-2731 , 8S-0783 , 8S-1339 , 82-0122 , 8S-1342 , 82-3514, 89-2653 , 8S-0089-C , 8S-0175, 8S-3200 , 82-3057 , 89-1243 , 8S-0495 , 89-0891 , 8S-0659 , 8S-1311-C, 8S-1678-C , 82-1893 , 8S-2528 , 82-0550 , 82-1010 , 89-3095 , 82-3663 , 82-2305 , 82-1055, 8S-3149 , 89-2393 , 82-1626, 89-2427, 89-3327 , 89-2172 , 82-0539, 82-1634, 82-3664, 89-2425 , 82-2931, 8S-2251 , 82-1031 , 82-0457, 89-0380, 8S-1753 , 8S-2551 , 8S-2233 , 89-1994 , 82-0146, 82-1181 , 82-0347, 82-2056, 89-1313 , 89-3074 , 89-0550, 89-2160 , 89-3462 , 8S-2715 , 82-0268-RC , 82-0673 , 89-1140 , 82-1511, 8S-1919 , 8S-2766 , 8S-0576, 8S-1454, 8S-2153 , 89-1558 , 89-2634 , 82-0353-RC, 82-7100-MH , 82-0583-RC, 82-1561, 82-2044, 82-2048, 89-1818 , 82-0585, 82-0586, 82-3489 , 82-3250 , 89-3134 , 89-2090 , 89-0536, 89-0819, 89-1343, 89-2366, 89-3488 , 82-2256 , 8S-2599 , 82-0379, 8S-2286 , 8S-2872 , 82-1814 , 82-0321, 82-2219, 89-0960, 89-1341, 8S-1370 , 82-3620 , 8S-0239, 8S-3287 , 8S-3052 , 89-2076 , 8S-0566, 8S-1047, 8S-1049 , 8S-0869 , 82-1813 , 8S-2506 , 8S-7131-CMH , 82-0006, 89-1734 , 8S-0717-C , 8S-1503, 8S-2784 , 8S-3189 , 89-0332 , 89-1784, 8S-1097, 8S-1513 , 8S-2025 , 89-0534 , 89-1549, 89-1807, 89-2229, 8S-1174, 8S-1196, 8S-1209 , 8S-2503 , 89-1909 , 89-2553 , 89-1439 , 89-1443 , 8S-2097, 8S-2147 , 8S-1489, 8S-2356 , 8S-0080 , 89-0563 , 89-0308-RC , 8S-3159 , 89-2845 , 8S-1399, 8S-2185 , 8S-3014 , 89-2849 , 89-2323 , 89-0478 , 89-1740 , 8S-0563 , 8S-1930-N , 8S-3222 , 89-1612 , 8S-1518, 89-2267 , 82-3309 , 82-3738, 8S-3232 , 89-1815 , 82-2090, 89-1781, 89-2198 , 89-1978 , 82-0416, 82-2762, 82-7134-MH, 89-2404, 8S-2593, 8S-2825 , 82-1955 , 82-0312 , 82-1212 , 82-0111, 82-2118 , 8S-0290 , 89-2492 , 89-1335, 89-1518 , 89-0706 , 82-0474-RC , 82-0503, 89-1491 , 89-3354 , 89-1735 , 89-2136 , 82-1623 , 89-3022 , 89-2831 , 8S-1606 , 8S-2903-C , 82-0439 , 82-2825-E , 82-0316, 82-1737, 82-2174, 89-0244, 89-0547, 89-1464 , 82-3567 , 82-1299 , 8S-0587, 8S-1628 , 82-0079, 82-1649, 82-3417, 82-3842, 82-3858 , 8S-0882 , 8S-2046-C , 8S-3119 , 82-1142 , 89-3301 , 89-0660, 89-2095 , 82-0308, 82-3384 , 82-2456 , 89-3086 , 8S-7006-CMH, 8S-7009-CMH , 82-1523, 82-3584 , 8S-1025 , 89-2195 , 89-1407 , 8S-1523, 8S-1586 , 89-2144 , 89-1262 , 89-2748 , 89-2986 , 8S-0473 , 8S-3275 , 8S-1465 , 8S-2533 , 89-0243, 89-1657 , 89-3080 , 8S-2177-N , 82-0036 , 82-1337, 82-1531 , 82-0480, 89-1048 , 82-2573 , 89-1918, 8S-2171 , 89-3364 , 89-1865, 89-2870, 89-2873 , 8S-1170 , 89-0429, 8S-2795 , 89-0024 , 82-1908 , 8S-2961 , 8S-1057 , 8S-1106 , 8S-1288 , 8S-0791 , 8S-2141, 8S-2340-N, 8S-2477 , 82-0923 , 82-0692, 82-3486, 82-3502 , 82-2807 , 82-0885, 82-2132, 82-3172 , 8S-1951 , 8S-1669 , 82-1597 , 82-0037, 82-1331, 89-0144, 8S-1379, 8S-1434, 8S-2173, 8S-3262 , 89-0112 , 89-1622 , 82-2253 , 89-1671, 89-2061 , 82-1158, 89-1320-RC, 89-1973-RC , 89-3548, 89-3549 , 8S-3173 , 8S-2311 , 8S-0980, 8S-1305, 8S-2191, 8S-2225-N , 8S-2690 , 89-0902 , 82-1816-RC , 89-1400-RC , 89-2632 , 89-1605 , 8S-3175 , 89-0415, 89-2240, 89-2786 , 8S-0770 , 82-0232 , 82-1544, 82-1748, 82-1793, 82-2037, 82-2835, 82-3107, 8S-0238 , 8S-1897, 8S-3279 , 82-3854, 89-0576 , 82-7009-MH, 89-1545, 8S-7179-CMH , 82-7010-MH, 89-1991, 89-1993, 89-2001 , 82-2755, 8S-2319, 8S-2497, 8S-2847, 8S-2989 , 82-1172, 82-2741, 8S-2317, 8S-2320 , 82-3497 , 8S-2886 , 89-2314 , 82-1644, 82-7102-MH , 82-0479 , 89-2116 , 82-1070, 82-7161-MH, 8S-1953, 8S-2259, 8S-2262-N , 8S-0272, 8S-1488 , 82-7003-MH , 82-0383, 82-2192, 82-3338 , 89-0513 , 82-3283, 89-1524 , 82-3870 , 82-2199 , 82-1022, 8S-1851-C , 89-2652 , 82-0086, 82-2454, 89-1924 , 89-1385, 8S-2608 , 82-1834, 8S-2221 , 89-0117 , 82-1835 , 8S-1576 , 8S-1066 , 8S-0293 , 82-3517 , 8S-2453 , 82-1019, 82-1397, 82-2154, 82-2530-C , 82-0247, 82-2058, 82-7138-MH, 89-1501, 89-3098, 8S-2152, 8S-7166-MH , 82-3512, 89-0373 , 82-1727, 82-3619, 89-1471, 8S-1528, 8S-1950, 8S-1959, 8S-2405, 8S-2414 , 82-0622, 82-3381, 89-0829, 89-1821, 8S-0014-C , 8S-7227-MH , 82-0519 , 82-3185, 89-0457, 89-0458, 8S-2792 , 82-0285, 89-0160-RC, 89-0312, 89-1480, 89-2711, 89-3064, 89-3533, 8S-0955, , 82-0670, 82-3821 , 82-1514, 82-3355 , 82-1926, 82-1953, 8S-0390, 8S-1313, 8S-1314, 8S-1445, 8S-3229, 8S-3230 , 82-1326, 82-7054-MH, 82-7056-MH , 89-1640 , 82-1481, 82-1811 , 89-2619 , 82-3492 , 89-1911 , 82-0567, 82-2261, 89-2434, 89-2438 89-2439, 89-3065, 89-3437, 89-3480 , 89-3518 , 82-1398, 89-1145 , 89-1360 , 82-2167, 8S-1366-C , 82-0221 , 89-0387, 89-0664 , 82-1219 , 82-1329, 8S-1480, 8S-2266, 8S-2364 , 82-1679, 89-1375 , 82-1692 , 82-2529 , 82-3709 , 8S-0748 , 82-0555, 82-0778, 89-1776, 89-2459 , 89-0051, 89-0545, 89-1323 , 89-1331 , 89-2825 , 89-3349, 89-3350 , 82-3126 , 82-0115 , 82-1351 , 82-1365 , 82-0022, 82-0840 , 82-1448-RC, 82-1937, 89-1642 , 8S-0426-C , 8S-1885 , 82-1786, 89-1541, 8S-0732, 8S-0916 , 8S-1212-C , 82-2635 , 82-7121-MH , 82-0671, 82-1827, 82-2152 , 8S-0346 , 82-0774 , 82-1588 , 82-2365 , 82-1304 , 8S-2771 , 82-0229 , 82-2879 , 82-0438, 82-3270, 8S-0139, 8S-1118, 8S-1395, 8S-1402 , 82-0920, 82-3556, 8S-7024-CMH , 8S-0020-C , 82-1622, 82-3373, 82-3903, 82-3917, 82-7197-MH, 89-2316, 89-3429 , 8S-7211-MH , 8S-7209-MH , 82-0349, 82-2178, 89-0292, 89-0856 89-0857, 8S-1701-RC, 8S-2288 , 8S-2452 , 89-0325 , 82-1608, 82-2443, 89-1195 , 82-2251 , 82-2821, 89-0418, 89-1802, 89-2629, 8S-0291 , 8S-1866 , 8S-1553 , 8S-0560-C , 82-3834 , 82-2021 , 82-3850 , 82-1094-C, 89-1589 , 82-1574 , 82-1650, 82-1652, 82-2450, 89-1382, 89-1621, 8S-1022 , 82-0456, 82-1564, 89-0762, 89-0803 , 89-2627, 8S-2388 , 8S-1262-C , 89-2093 , 89-0421, 89-2021, 89-3093, 8S-2549-N , 89-1011 , 89-3470 , 8S-1719 , 8S-1896 , 8S-1687 , 82-1806, 8S-0029 , 82-2532 , 82-3484-C , 8S-2800, 8S-3122 , 8S-0977 , 82-1294, 8S-2514 , 82-1936, 82-3540-C , 82-0395, 89-2465 , 82-0637, 82-0639, 82-0644 , 89-2059 , 8S-2580 , 82-2294 , 82-2093 , 82-3153 , 82-0898 , 89-2383 , 89-2613 , 82-0288 , 82-3911 , 82-2244 , 89-2548 , 89-0154 , 82-2406 , 82-3222 , 89-0155 , 82-1641 , 82-0783, 82-1071, 82-2648 , 89-3517 , 82-2536 , 82-2537 , 82-2075, 82-2582, 89-1769 , 8S-2273 , 89-1344 , 82-2257 , 89-1869, 8S-2360 , 82-0669 , 82-0348, 82-2382, 82-2416, 89-2042, 89-2044, 89-2078, 89-2083 , 82-2546, 82-2547 , 89-3500 , 8S-3030 , 82-0685 , 89-0520 , 82-1186 , 82-1897 , 89-2676 , 89-3442 , 89-3288 , 82-0706 , 82-0723 , 82-0736, 89-0594, 89-0595, 89-0596 , 82-0741, 82-3097, 89-2682 , 82-0761 , 82-0959 , 82-1043 , 82-1095, 82-2516, 82-3369 , 82-1489, 82-3192 , 82-1855, 82-3735, 82-3893 , 82-2481 , 82-2488 , 82-2491 , 82-2523 , 82-2535, 82-3100, 89-3287 , 8S-0370 , 82-2796 , 82-3087 , 82-3103 , 82-3154 , 82-3158, 82-7055-MH, 82-7061-MH, 89-2715 , 82-3722 , 82-3724 , 89-0618 , 89-1552, 89-1555 , 89-1617 , 89-1831 , 89-2750 , 8S-1858, 8S-1859 , 8S-2686 , 89-2728 , 82-0722, 82-0725, 82-0727, 82-3731 , 82-3732 , 82-1839 , 82-2435 , 8S-0734 , 82-2506 , 89-2584 , 82-0675 , 89-3207 , 82-2512, 8S-3022 , 82-0163-C, 82-0236-C , 89-2121 , 82-1885 , 82-1912 , 82-1221 , 82-7163-MH , 82-0315, 89-3292 , 8S-1658-CH , 82-0459 , 82-0210, 82-3725 , 8S-2681 , 82-1088 , 8S-0375 , 82-1068 , 89-2736 , 82-3098 , 82-1091 , 82-7049-MH , 82-0799, 82-0826, 89-2768 , 8S-0066 , 82-1138 , 8S-0642 , 82-1508, 82-3096 , 82-2394 , 82-1651 , 82-1760 , 82-7192-MH, 89-2176, 89-3476 , 82-0565, 82-2515 , 89-0992 , 82-1860 , 82-1611, 82-3094 , 82-2493, 82-3720, 8S-2810 , 82-2507 , 82-3707 , 82-0802, 89-3334, 89-3335 , 82-2490 , 82-1818, 82-3711, 89-3283, 89-3535, 8S-1652 , 8S-1797, 8S-2166 , 8S-1675, 8S-2211 , 82-1201 , 82-2483, 82-7144-MH, 89-2694 , 82-3876 , 82-0197 , 82-1257 , 82-3213 , 8S-3242-H , 8S-0101 , 8S-2526 , 82-2335 , 8S-1351 , 89-1226 , 89-1665 , 82-1357, 82-1710-RC, 89-3523 , 89-1985 , 82-3039, 89-0311 , 82-2501 , 82-1982, 8S-0050-C , 8S-0879 , 8S-2516-N , 82-0509, 8S-1644 , 89-0064 , 82-2177, 89-3544 , 82-0510, 82-0834, 82-2252 , 89-1444 , 89-2372 , 8S-0186, 8S-0878 , 89-0428 , 8S-2924 , 8S-1990 , 8S-2365, 8S-2845 , 89-3125 , 82-2499, 89-2148, 89-2155 , 89-3484 , 82-3391 , 82-0499, 89-0433, 89-0434 , 89-1373, 89-2774, 89-2776 , 8S-2624 , 82-2737 , 82-3130 , 82-2071 , 82-3814 , 82-1199 , 82-1146 , 82-3654 , 82-1259 , 82-1193 , 82-2663 , 89-0926 , 8S-0106 , 89-2742-N , 89-1804 , 8S-2612 , 82-0656, 82-3668, 89-3348 , 89-0518 , 89-1482 , 82-0096, 82-0624, 82-0950, 82-0951, 89-0801, 89-2486 , 82-1225, 82-3137, 89-2713 , 8S-0213 , 89-2360 , 82-0131 , 82-2992 , 82-2993 , 82-7030-MH , 89-2265 , 82-1774, 89-1497 , 8S-1828 , 8S-2818 , 82-0864, 82-2473 , 82-2463, 82-2467 , 82-3533 , 89-1374 , 82-0320 , 82-2752 , 8S-0367 , 8S-3259 , 8S-7188-MH , 8S-0735 , 8S-2339-N , 82-0854, 82-3602, 89-1516, 8S-0113 , 8S-1335 , 82-2591, 89-1123 , 82-2329, 8S-1668 , 8S-2581 , 82-1322 , 89-2583 , 82-2987, 89-0118, 89-2199, 8S-0972 , 8S-2072 , 82-2852, 89-0294 , 8S-2258 , 82-2584 , 8S-0517 , 89-2285 , 89-0845 , 82-1459, 89-0193 , 8S-0111 , 89-1888, 89-3109 , 82-1006 , 82-0342 , 82-3086 , 8S-0808 , 82-2373, 82-2378, 82-7079-MH, 89-0168, 89-0721, 89-0958, 8S-1882, 8S-1886 , 89-0835 , 82-0156, 82-3638 , 8S-2149 , 8S-2927 , 82-2361 , 89-0546 , 8S-2611 , 82-3043 , 82-1555 , 82-0928 , 8S-2926 , 82-7139-MH, 89-1355, 89-1979, 89-2218, 8S-7139-MH , 8S-1526 , 82-0482, 82-2520 , 82-7153-MH , 82-0537, 82-0538 , 82-3003, 89-1158, 89-2071, 89-2308 , 89-2211 , 89-2610 , 8S-2353 , 82-0007, 82-1613, 89-0291, 89-0324-RC, 8S-2127 , 82-0571 , 82-2368, 8S-2985 , 8S-7031-MH , 89-1565 , 89-1075 , 89-2473, 8S-0008-C , 8S-0522-C-RC , 8S-1973, 8S-2263 , 82-2372 , 8S-0509 , 89-1104 , 8S-2751, 8S-2770 , 89-2498 , 82-0876 , 8S-0697 , 8S-2820 , 82-2792, 89-3565 , 82-0882, 82-3295 , 82-1161 , 82-1851 , 8S-1846, 8S-2084 , 82-3140, 82-3441 , 82-0779, 8S-0342 , 8S-1380 , 82-3861 , 82-0373, 89-1390, 89-1392, 89-2085, 89-2087, 89-2089 , 82-1302 , 89-2032 , 89-0771 , 8S-0839 , 89-0571 , 8S-2017 , 82-0307 , 89-3188 , 89-0933 , 82-1012 , 8S-1571 , 82-1647, 89-0235, 89-0665 , 89-0484 , 89-1259 , 82-2053 , 82-2303 , 89-1494 , 82-1629, 89-3110, 89-3111 , 82-1742 , 89-3072 , 82-3211 , 89-1956 , 89-3469 , 8S-2904 , 8S-2255 , 89-2872, 8S-2223-N , 8S-0133 , 8S-3278 , 89-0283 , 8S-1715 , 82-1640, 89-2842, 8S-1021, 8S-2016, 89-0131 , 82-1437, 82-1438 , 82-3588, 8S-7078-MH , 89-3017 , 82-2812 , 82-2926 , 89-2149 , 89-1929, 82-1126, 82-2061-C, 89-1072 , 82-1999, 82-3437, 89-0806 , 89-0509, 89-1747 , 8S-1607 , 89-2139 , 8S-0646 , 82-0917, 8S-1906 , 82-0611, 8S-0012 , 82-3317, 89-0924 , 82-7188-MH , 82-2070, 89-3221 , 8S-0136 , 82-0209 , 89-3005 , 82-3550 , 89-2759 , 82-3623 , 89-1433, 8S-0496 , 89-0489 , 8S-2711 , 82-2296 , 82-3363, 89-1849, 8S-2420 , 82-3863 , 89-0097, 89-3059 , 82-2848 , 82-1614 , 82-0100-C, 82-3288, 89-0673, 89-1144 , 82-3717 , 82-1941, 82-3450 , 89-0162 , 82-0420 , 89-3396 , 89-2067, 8S-1182 , 8S-2850 , 82-0717, 82-0719, 82-7175-MH, 8S-3026, 8S-3027, 82-1533, 8S-0921, 82-0652 , 89-1647 , 8S-3060 , 89-1152 , 89-2228 , 82-1093 , 82-1921 , 8S-0185, 8S-1834 , 82-0205, 82-0995, 89-0314 , 82-0384, 89-0559, 89-0676, 89-2570, 89-2576, 8S-0041-C, 8S-3018 , 82-2025, 82-2901, 8S-0302 , 82-1442, 82-1578, 82-2298 , 82-2858 , 89-2529-N , 82-3350 , 89-0288, 89-0319 , 82-3376 , 8S-0403 , 8S-1319 , 82-7196-MH , 89-1082 , 8S-0241, 8S-1092 , 89-2844, 8S-2643 , 82-0217 , 89-0313, 89-2660 , 89-3440, 8S-1515 , 82-2753 , 82-2076 , 82-0996, 89-0466 , 82-2164 , 8S-2722 , 8S-0474 , 82-1853, 82-2377 , 82-2854 , 82-0921 , 82-1288, 89-0158, 8S-7040-CMH , 8S-0615-C, 8S-2671 , 8S-2572 .82-3308 , 89-1367 , 82-2988 , 89-1388 , 89-0944, 89-1308 , 82-3537 , 8S-0991, 8S-3120 , 89-2421 , 89-3185 , 89-3230 , 89-3252 , 89-2503 , 82-0816 , 82-3278, 89-3368 , 8S-0841 , 89-0157, 89-2416 , 82-3508, 89-0067 , 89-2878 , 8S-3058 , 82-0048, 82-2531, 89-2107 , 8S-2652 , 82-1944, 89-1645 , 82-1896, 82-1938, 89-1408 , 82-1665 , 8S-0569 , 82-0026 , 8S-2663 , 82-1005 , 8S-1666 , 82-3051 , 82-0903-RC , 82-3920 , 82-2578, 8S-1934 , 82-0246, 82-0682, 82-2957 , 89-0218, 8S-1010 , 82-3442, 89-0442, 8S-2037 , 89-3293 , 89-1590 , 89-1591 , 89-0212-RC , 82-0043, 82-1191, 82-7046-MH, 89-0608 , 8S-2642 , 82-0614 , 82-3313 , 8S-0996 , 89-3466 , 89-3342 , 8S-2515-RC , 89-3321 , 8S-1917-RC , 8S-3089 , 8S-3093 , 8S-1458 , 89-3341 , 8S-3085 , 89-3320 , 8S-3084 , 89-3311 , 89-3318 , 89-3310 , 82-2981 , 8S-3079-C , 89-3305 , 89-3360 , 8S-7220-CMH , 89-3306 , 8S-3086 , 89-3355 , 89-3308 , 8S-1908-N. |
| FEI Number |
1720753
|
Recalling Firm/ Manufacturer |
GE OEC Medical Systems, Inc
384 Wright Brothers Dr
Salt Lake City
UT
84116-2862
|
| For Additional Information Contact | Deena Pease 801-536-4516 |
Manufacturer Reason for Recall | Please be aware that this is not a new recall. The firm has taken action; but, due to administrative issues this recall is now being classified by the Agency. The vertical lift column power supply in the mainframe C-arm of the device is defective and subject to early life failure. |
FDA Determined Cause 2 | Software design |
| Action | Please be aware that this is not a new recall. The firm has taken action; but, due to administrative issues this recall is now being classified by the Agency. The firm, GE Healthcare, Surgery, sent an "URGENT RECALL NOTICE" dated November 8, 2006 via certified mail with return receipt request to consignees/customers with specific instructions. Customers were advised "to stop using the system WITH NAVIGATION until this replacement has been successfully completed. OEC@ FluoroTrak 9800 Plus system may be used as a standard C-Arm providing you review the other issues as listed below." GEHC OEC is actively working on solutions that will permanently resolve each of the noted issues.
If you have any questions or concerns regarding these issues, please do not hesitate to contact the service team for further information at 800-874-7378 option 8. Information is available at this number 24 hours per day, 7 days a week. |
| Quantity in Commerce | 8214 units |
| Distribution | Worldwide distribution. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|