| Date Initiated by Firm | July 29, 2009 |
| Date Posted | March 05, 2010 |
| Recall Status1 |
Terminated 3 on December 06, 2010 |
| Recall Number | Z-1014-2010 |
| Recall Event ID |
52851 |
| 510(K)Number | K040391 |
| Product Classification |
Diazo colorimetry, bilirubin - Product Code CIG
|
| Product | Pointe Scientific Direct Bilirubin Reagent Set, Pointe Scientific Inc., Canton, MI., Catalog # HB936-1250. |
| Code Information |
Lot 817103-337. |
| FEI Number |
1827821
|
Recalling Firm/ Manufacturer |
Pointe Scientific, Inc.
5449 Research Dr
Canton
MI
48188
|
| For Additional Information Contact | 800-445-9853 Ext. 110 |
Manufacturer Reason for Recall | R2 reagent component may fail because it may contain R1 reagent. The reagent was rendered inoperative and would not pass the required quality control (calibration and quality control testing) before beginning running the bilirubin assay. |
FDA Determined Cause 2 | Error in labeling |
| Action | Consignee was notified by letter dated 7/29/2009 and instructed to identify and notify them at once of this product recall. Product is to be returned or destroyed. Customers with questions should contact the firm directly. |
| Quantity in Commerce | 2 |
| Distribution | California. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|