• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Varian brand CSeries Ximatron Radiotherapy Treatment Stimulator,

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Varian brand CSeries Ximatron Radiotherapy Treatment Stimulator,see related information
Date Initiated by FirmJanuary 19, 2010
Date PostedJanuary 11, 2011
Recall Status1 Terminated 3 on February 24, 2012
Recall NumberZ-0892-2011
Recall Event ID 56350
510(K)NumberK853349 K964138 
ProductVarian brand C-Series Ximatron Radiotherapy Treatment Stimulator, Model H72, Product is manufactured and distributed by Varian Oncology Systems, Palo Alto, CA The product is indicated for use in radiation therapy simulation, using a fluoroscopic and/or radiographic x-ray system for visualizing the volume to be exposed during radiation therapy.
Code Information Listed are the affected system Serial Numbers. The last four digits represent the serial number of the device.  H720003 ,H720119 ,H720209 ,H720270 ,H720347 ,H720416 ,H720491 ,H720557 ,H720618, H720006 ,H720122 ,H720210 ,H720271 ,H720348 ,H720417 ,H720495 ,H720558 ,H720620, H720009 ,H720123 ,H720211 ,H720273 ,H720350 ,H720419 ,H720496 ,H720559 ,H720622, H720013 ,H720124 ,H720213 ,H720274 ,H720351 ,H720422 ,H720498 ,H720560 ,H720624, H720015 ,H720125 ,H720215 ,H720276 ,H720352 ,H720424 ,H720499 ,H720561 ,H720626, H720017 ,H720129 ,H720217 ,H720279 ,H720353 ,H720425 ,H720500 ,H720562 ,H720627, H720021 ,H720132 ,H720218 ,H720280 ,H720354 ,H720427 ,H720501 ,H720563 ,H720628, H720022 ,H720133 ,H720219 ,H720281 ,H720355 ,H720429 ,H720502 ,H720565 ,H720629, H720025 ,H720134 ,H720221 ,H720282 ,H720357 ,H720432 ,H720503 ,H720566 ,H720632, H720035 ,H720137 ,H720222 ,H720283 ,H720358 ,H720435 ,H720505 ,H720567 ,H720637, H720037 ,H720138 ,H720225 ,H720284 ,H720362 ,H720437 ,H720506 ,H720568 ,H720638, H720040 ,H720140 ,H720229 ,H720288 ,H720363 ,H720438 ,H720508 ,H720571 ,H720639, H720041 ,H720143 ,H720230 ,H720289 ,H720367 ,H720439 ,H720509 ,H720572 ,H720641, H720048 ,H720144 ,H720231 ,H720290 ,H720373 ,H720440 ,H720511 ,H720573 ,H720642, H720051 ,H720145 ,H720232 ,H720292 ,H720374 ,H720442 ,H720513 ,H720575 ,H720643, H720052 ,H720148 ,H720233 ,H720293 ,H720375 ,H720443 ,H720514 ,H720577 ,H720644, H720056 ,H720152 ,H720234 ,H720294 ,H720376 ,H720445 ,H720515 ,H720578 ,H720647, H720057 ,H720154 ,H720235 ,H720295 ,H720379 ,H720447 ,H720516 ,H720579 ,H720648, H720059 ,H720157 ,H720236 ,H720296 ,H720383 ,H720449 ,H720518 ,H720584 ,H720650, H720060 ,H720159 ,H720237 ,H720297 ,H720384 ,H720452 ,H720522 ,H720585 ,H720651, H720064 ,H720164 ,H720239 ,H720298 ,H720386 ,H720457 ,H720523 ,H720586 ,H720653, H720069 ,H720165 ,H720241 ,H720299 ,H720388 ,H720458 ,H720525 ,H720588 ,H720654, H720070 ,H720167 ,H720242 ,H720300 ,H720391 ,H720461 ,H720526 ,H720589 ,H720659, H720071 ,H720168 ,H720245 ,H720303 ,H720392 ,H720465 ,H720528 ,H720590 ,H720660, H720072 ,H720169 ,H720247 ,H720307 ,H720394 ,H720466 ,H720530 ,H720591 ,H720661, H720074 ,H720177 ,H720248 ,H720309 ,H720395 ,H720470 ,H720531 ,H720592 ,H720663, H720075 ,H720180 ,H720251 ,H720316 ,H720396 ,H720471 ,H720533 ,H720594 ,H720664, H720082 ,H720181 ,H720254 ,H720320 ,H720397 ,H720472 ,H720534 ,H720595 ,H720666, H720083 ,H720182 ,H720255 ,H720327 ,H720399 ,H720477 ,H720536 ,H720596 ,H720667, H720088 ,H720183 ,H720256 ,H720328 ,H720402 ,H720478 ,H720537 ,H720597 ,H720668, H720089 ,H720185 ,H720257 ,H720329 ,H720403 ,H720479 ,H720539 ,H720598 ,H720669, H720092 ,H720186 ,H720258 ,H720330 ,H720404 ,H720481 ,H720546 ,H720599 ,H720670, H720098 ,H720187 ,H720261 ,H720336 ,H720405 ,H720482 ,H720547 ,H720600 ,H720671, H720101 ,H720188 ,H720262 ,H720337 ,H720406 ,H720484 ,H720548 ,H720601 ,H720672, H720106 ,H720195 ,H720263 ,H720338 ,H720407 ,H720485 ,H720550 ,H720604 ,H720673, H720107 ,H720196 ,H720264 ,H720339 ,H720409 ,H720486 ,H720553 ,H720605 ,H720674, H720108 ,H720199 ,H720265 ,H720344 ,H720411 ,H720487 ,H720554 ,H720608 ,H720675, H720112 ,H720200 ,H720266 ,H720345 ,H720414 ,H720488 ,H720555 ,H720615 ,H720676, H720113 ,H720204 ,H720269 ,H720346 ,H720415 ,H720489 ,H720556 ,H720616 ,H720678,   H720679 ,H720743 ,H720805 ,H720858 ,H720905 ,H720955 ,H721005, H720680 ,H720744 ,H720806 ,H720859 ,H720906 ,H720956 ,H721006, H720681 ,H720746 ,H720807 ,H720860 ,H720907 ,H720957 ,H721007, H720682 ,H720747 ,H720808 ,H720861 ,H720908 ,H720958 ,H721008, H720683 ,H720748 ,H720809 ,H720862 ,H720909 ,H720959 ,H721009, H720684 ,H720749 ,H720810 ,H720863 ,H720910 ,H720960 ,H721010, H720686 ,H720750 ,H720811 ,H720864 ,H720911 ,H720961 ,H721011, H720688 ,H720752 ,H720812 ,H720865 ,H720913 ,H720962 ,H721012, H720689 ,H720754 ,H720813 ,H720866 ,H720914 ,H720963 ,H721013, H720692 ,H720756 ,H720814 ,H720867 ,H720915 ,H720964 ,H721014, H720693 ,H720759 ,H720815 ,H720868 ,H720916 ,H720965 ,H721015, H720694 ,H720760 ,H720816 ,H720869 ,H720918 ,H720966 ,H721016, H720697 ,H720761 ,H720817 ,H720870 ,H720919 ,H720967 ,H721017, H720699 ,H720763 ,H720818 ,H720871 ,H720920 ,H720968 ,H721018, H720701 ,H720766 ,H720819 ,H720872 ,H720921 ,H720969 ,H721019, H720702 ,H720767 ,H720820 ,H720873 ,H720922 ,H720974 ,H721020, H720704 ,H720769 ,H720822 ,H720874 ,H720923 ,H720975 ,H721022, H720705 ,H720770 ,H720823 ,H720875 ,H720925 ,H720976 ,H721023, H720706 ,H720773 ,H720824 ,H720876 ,H720926 ,H720977 ,H721024, H720707 ,H720774 ,H720825 ,H720877 ,H720927 ,H720978 ,H721025, H720711 ,H720775 ,H720826 ,H720878 ,H720928 ,H720979 ,H721026, H720712 ,H720776 ,H720827 ,H720879 ,H720929 ,H720980 ,H721027, H720714 ,H720777 ,H720829 ,H720880 ,H720930 ,H720981 ,H721028, H720715 ,H720778 ,H720830 ,H720881 ,H720931 ,H720982 ,H721029, H720716 ,H720780 ,H720831 ,H720882 ,H720932 ,H720983 ,H721030, H720718 ,H720781 ,H720832 ,H720883 ,H720933 ,H720984 ,H721031, H720719 ,H720782 ,H720833 ,H720884 ,H720935 ,H720985 ,H721032, H720720 ,H720783 ,H720834 ,H720885 ,H720936 ,H720986 ,H721033, H720721 ,H720784 ,H720835 ,H720886 ,H720937 ,H720987 ,H721034, H720722 ,H720786 ,H720836 ,H720888 ,H720938 ,H720988 ,H721035, H720723 ,H720787 ,H720840 ,H720889 ,H720940 ,H720989 ,H721036, H720724 ,H720788 ,H720841 ,H720890 ,H720941 ,H720990 ,H721037, H720725 ,H720790 ,H720842 ,H720891 ,H720942 ,H720991 ,H721038, H720726 ,H720792 ,H720843 ,H720892 ,H720943 ,H720992 ,H721039, H720727 ,H720795 ,H720844 ,H720894 ,H720945 ,H720993 ,H721040, H720728 ,H720796 ,H720847 ,H720895 ,H720946 ,H720995 ,H721041, H720729 ,H720797 ,H720848 ,H720896 ,H720947 ,H720996 ,H721042, H720732 ,H720798 ,H720849 ,H720897 ,H720948 ,H720997 ,H721043, H720733 ,H720799 ,H720851 ,H720898 ,H720949 ,H720998 ,H721044, H720734 ,H720800 ,H720852 ,H720899 ,H720950 ,H720999 ,H721045, H720737 ,H720801 ,H720854 ,H720901 ,H720951 ,H721000 ,H721046, H720739 ,H720802 ,H720855 ,H720902 ,H720952 ,H721001 ,H721047, H720741 ,H720803 ,H720856 ,H720903 ,H720953 ,H721002 ,H721048, H720742 ,H720804 ,H720857 ,H720904 ,H720954 ,H721004    
FEI Number 2916710
Recalling Firm/
Manufacturer
Varian Medical Systems Oncology Systems
911 Hansen Way
Palo Alto CA 94304-1028
For Additional Information ContactPatricia Erwin, RAC
650-424-5731
Manufacturer Reason
for Recall
There is a potential for failure of a lead screw drive system or image intensifier lead screw drive assembly, if normal maintenance inspections are not followed correctly.
FDA Determined
Cause 2
Other
ActionVarian sent an URGENT MEDICAL DEVICE CORRECTION letter via certified mail receipt on January 19, 2010, to all direct consignees. The letter identified the product, the problem, and the action the consignee should take. Consignees were instructed to have maintenance personnel perform the recommended measurements as described in the Varian Customer Technical Bulletin. Ensure that all measurements made as part of the procedure are recorded in the 'wear table' that is part of the procedure. Previous wear measurements should be compared to new measurements and any large increase in wear should be investigated immediately. Pay special attention to the "Recommendations based on wear measurements" table. Ensure that new wear measurements comply with the operating parameters described in the table. For any questions consignees are to contact Varian Medical Systems Customer Support Services at one of the locations described in the CTB. For questions regarding this recall call (650) 424-5731.
DistributionWorldwide Distribution including USA, Australia, New Zealand, Europe, Southeast Asia, China, Japan, Latin America, and Canada.

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
510(K) Database510(K)s with Product Code =
510(K)s with Product Code =
-
-