 Class 2 Device Recall Symbiq Infusersee related information
Date Initiated by FirmFebruary 22, 2010
Date PostedSeptember 09, 2010
Recall Status1 Terminated 3 on April 01, 2011
Recall NumberZ-2390-2010
Recall Event ID 56464
510(K)NumberK041550 
Product Classification Pump, infusion - Product Code FRN
ProductHospira brand Symbiq One-Channel Infuser LIST No. 16026 Symbiq One-Channel Infuser, List Number Configurations: 16026-04-01/03/04,51/52, 53/54, 79/80, 81/82, 83/84, 87/88 and 89/90, 16026-13-25/26, 53/54, Distributed by: Hospira, Inc. Morgan Hill, CA
Code Information All units of these List Numbers.
FEI Number 2921482
Recalling Firm/
Manufacturer
Hospira Inc
755 Jarvis Dr
Morgan Hill CA 95037-2810
For Additional Information Contact
408-782-3200
Manufacturer Reason
for Recall
The clamp may not secure to the IV pole and the pump may slip or fall off. The pump may fall on a person or pull the tubing out of the IV container.
FDA Determined
Cause 2
Device Design
ActionHospira issued an "Urgent: Device Field Correction" Letter dated February 22, 2010 to all consignees, identifying the reason for the recall and affected product. Customers were informed that they will be contacted by a Hospira representative to replace the affected pole clamps. Consignees can contact Hospira Technical Support Operations at 1-800-241-4002.
DistributionWorldwide Distribution: USA (all states and Puerto Rico) and the countries of Australia, Canada, Colombia, Costa Rica, Malaysia, Mexico, Singapore. and the UAE.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
510(K) Database510(K)s with Product Code = FRN
-
-