Class 2 Device Recall Merit One Step Centesis Drainage Catheter

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Merit One Step Centesis Drainage Cathetersee related information
Date Initiated by FirmNovember 17, 2010
Date PostedJanuary 07, 2011
Recall Status1 Terminated 3 on March 23, 2012
Recall NumberZ-0866-2011
Recall Event ID 57418
Product Classification Catheter, irrigation - Product Code gbx
ProductMerit One Step Centesis Drainage Catheter Intended for the percutaneous drainage of fluids
Code Information LOT NO.: F762656, F762657, F775508, F775516, F775524, F775533, F775552. . . . H101842, H101902, H108630, H110628, H110629, H110630, H110631, H113645, H113655, H115407, H115409, H118206 ,H118207, H118208, H118209, H118210, H118211, H121450, H121780, H124502, H124506, H124507, H124508, H124509, H124510, H124511, H134990, H134991, H134992, H134993, H134994, H137477, H139171, H141155, H141162, H144090, H144098, H152489, H161044, H168461, H176238, . . . . . T119693, T121646, T122832,T124548, T130679, T132868, T132953, T133542, T134201, T135053, T137437, T139844, T140413, T140414, T141401, T141403, T141404, T142352, T143776, T146337, T146360, T149794, T150330, T150862, T151354, T153082, T153417, T153437, T153683, T153684, T153987, T154640, T157268, T160213, T162756, T167168, T167454, T167544, T168498, T171102, T171125, T171265, T173628, T173860, T174223, T176638, T176932, T177808, T178914, T179018, T179791, T181437
FEI Number 1721504
Recalling Firm/
Manufacturer
Merit Medical Systems, Inc.
1600 West Merit Parkway
South Jordan UT 84095
For Additional Information ContactJackie Preece
801-253-1600
Manufacturer Reason
for Recall
The catheter's valved hub may fail resulting in an inability to drain fluid.
FDA Determined
Cause 2
Device Design
ActionMerit Medical Systems, Inc. sent a Product Safety Notice letter dated November 18, 2010, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to provide a copy of the Safety Notice form to their applicable customers. Advise them to immediately stop using the affected product and to discard it. Customers were asked to complete and sign the Safety Notice forms and fax them to 1-801-208-3378 and send the original Safety Notice Form with both signatures via Federal Express 2nd day air to: Customer Service Merit Medical Systems, Inc. 1600 West Merit Parkway South Jordan, UT 84095-2416 entirety and faxed them For questions regarding this recall call 801-316-4929 or Customer Service at 801-208-4365.
Quantity in Commerce20,863
DistributionWorldwide Distribution - USA including AL, AR, AZ, CA, CO, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NH, NJ, NM, NV, NY, OH, OR, PA, RI, SC, TN, TX, UT, VA, VT, WA, WI, AND WV and the countries of Canada, Finland, Germany, Netherlands, Singapore, Sweden, Switzerland, and the UK