• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Surgical Navigation System

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Surgical Navigation Systemsee related information
Date Initiated by FirmDecember 22, 2010
Date PostedMarch 04, 2011
Recall Status1 Terminated 3 on March 11, 2011
Recall NumberZ-1558-2011
Recall Event ID 57696
510(K)NumberK050438 
Product Classification Neurological stereotaxic instrument - Product Code HAW
ProductPolaris Spectra Position Sensor Unit, a component of the StealthStation S7 System, Medtronic Navigation, Louisville, CO 80027. The Polaris Spectra Position Sensor Unit (PSU) is the component of the StealthStation S7 System which detects optical markers on instruments and reference frames, determines their spatial positions, and continuously reports this information to the system computer. The computer uses this spatial information, in conjunction with information regarding the geometry of the instrument currently in use, to determine where the tip of the instrument is located on the patient anatomy.
Code Information Serial Numbers: P703048, P703049, P703050, P703054, P703055, P703060, P703061, P703063, P703066, P703068, P703070, P703076, P703077, P703079, P703114, P703140, P703141, P703143.
FEI Number 1000517638
Recalling Firm/
Manufacturer
Medtronic Navigation, Inc
826 Coal Creek Circle
Louisville CO 80027-9710
For Additional Information Contact
720-890-3409
Manufacturer Reason
for Recall
Position Sensor Unit may generate tracking errors or may stop tracking immediately.
FDA Determined
Cause 2
Component design/selection
ActionThe firm, Medtronic, sent a "Product Correction Notification" letter dated December 22, 2010, to all US Consignees/Customers on December 23, 2010 and foreign Consignees/Customers on December 27, 2010. The letter described the product, problem and actions to be taken. The customer were instructed to discontinue use of the impacted navigation system(s) until a replacement camera (PSU) is installed. A Medtronic Navigation Representative will visit each facility to remove the navigation camera and install a replacement. If you have any questions, contact your Medtronic Navigation Representative or Medtronic Navigation Technical Support services at 1-800-595-9709 or +1-720-890-3200.
Quantity in Commerce11 (US), 8 (Intnl)
DistributionWorldwide distribution: USA including states of: CA, CO, GA, SC, TN, and WI; and countries including: Australia, Canada, Chile, Russia, Saudi Arabia, and South Africa.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
510(K) Database510(K)s with Product Code = HAW
-
-